Advanced Energy Impac 600 Series User manual

Impac® Series 600 Pyrometer
User Manual
May 2022 57010228-00C

COPYRIGHT
This manual and the information contained herein are the proprietary property of
Advanced Energy Industries, Inc.
No part of this manual may be reproduced or copied without the express written
permission of Advanced Energy Industries, Inc. Any unauthorized use of this manual
or its contents is strictly prohibited. Copyright © 2020-2022 Advanced Energy
Industries, Inc. All Rights Reserved.
DISCLAIMER AND LIMITATION OF LIABILITY
The information contained in this manual is subject to change by Advanced Energy
Industries, Inc. without prior notice. Advanced Energy Industries, Inc. makes no
warranty of any kind whatsoever, either expressed or implied, with respect to the
information contained herein. Advanced Energy Industries, Inc. shall not be liable in
damages, of whatever kind, as a result of the reliance on or use of the information
contained herein.
PRODUCT USAGE STATEMENT
WARNING:
Read this entire manual and all other publications pertaining to the work
to be performed before you install, operate, or maintain this equipment.
Practice all plant and product safety instructions and precautions. Failure
to follow instructions can cause personal injury and/or property damage. If
the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment might be impaired. All personnel who
work with or who are exposed to this equipment must take precautions to
protect themselves against serious or possibly fatal bodily injury.
Advanced Energy Industries, Inc. (AE) provides information on its products
and associated hazards, but it assumes no responsibility for the after-sale
operation of the equipment or the safety practices of the owner or user.
NEVER DEFEAT INTERLOCKS OR GROUNDS.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C ii

TRADEMARKS
All Advanced Energy trademarks are the property of Advanced Energy
Industries, Inc. For the list of Advanced Energy trademarks, visit: https://
www.advancedenergy.com/legal/. Any unauthorized use of Advanced Energy
trademarks is prohibited.
All other trademarks are the property of their respective owners.
CUSTOMER FEEDBACK
Advanced Energy’s technical writing staff has carefully developed this manual using
research-based document design principles. However, improvement is ongoing, and
the writing staff welcomes and appreciates customer feedback. Please send any
comments on the content, organization, or format of this user manual to:
To order a manual, please contact Technical Support:
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C iii

Table of Contents
Chapter 1. Safety and Product Compliance Guidelines
Important Safety Information ................................................................................. 1-1
Danger, Warning, and Caution Boxes ................................................................... 1-1
Safety Guidelines .................................................................................................. 1-1
Rules for Safe Installation and Operation ....................................................... 1-2
Interpreting Product Labels ................................................................................... 1-2
Product Compliance .............................................................................................. 1-3
Product Certification ................................................................................ ....... 1-3
Safety and EMC Regulations, Directives, and Standards ....................... ....... 1-3
Conditions of Use .................................................................................... ....... 1-3
Environmental Compliance ............................................................................. 1-4
Chapter 2. Product Overview
General Description ....................................................................................... ....... 2-1
Unit Components ........................................................................................... ....... 2-1
Pyrometer ................................................................................................ ....... 2-1
Sensor Heads ................................................................................................. 2-1
Sensor Cables ......................................................................................... ....... 2-2
Converter Box ................................................................................................. 2-2
Analog Box .............................................................................................. ....... 2-2
Multi-Sensor Box ..................................................................................... ....... 2-2
Accessories ............................................................................................. ....... 2-2
Unit Connectors ............................................................................................. ....... 2-3
Sensor Head Connectors ........................................................................ ....... 2-3
Extension Cable Connectors ................................................................... ....... 2-3
Sensor Cable Connectors ....................................................................... ....... 2-4
Converter Box Connectors ...................................................................... ....... 2-5
Analog Box Connectors .................................................................................. 2-5
Multi-Sensor Box Connectors ......................................................................... 2-6
Software ................................................................................................................ 2-7
Chapter 3. Installation, Setup, and Operation
Preparing to Install the Unit ................................................................................... 3-1
Unpacking the Unit .................................................................................. ....... 3-1
Measuring Through External Windows ........................................................... 3-1
Verifying the Spot Size and Installation Location .................................... ....... 3-1
Installation at Any Measuring Distance .................................................... 3-1
Installing the Unit ................................................................................................... 3-2
Mounting the Pyrometer .......................................................................... ....... 3-2
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C ivTable of Contents

Grounding the Pyrometer ........................................................................ ....... 3-3
Connecting the Sensor Cable ......................................................................... 3-3
Connecting the Sensor Head .................................................................. ....... 3-4
Connecting Input Power to the Unit ................................................................ 3-5
Connecting Serial Communications ........................................................ ....... 3-6
Baud Rate ......................................................................................... ....... 3-6
Cable Length ............................................................................................ 3-7
Connecting to an RS-232 Interface .......................................................... 3-7
Connecting to an RS-485 Interface .......................................................... 3-7
Connecting to PROFINET ........................................................................ 3-8
Connecting to USB ........................................................................... ....... 3-9
Connecting Additional Devices ....................................................................... 3-9
Connecting Multiple Analog Outputs ....................................................... ....... 3-9
First Time Operation ...................................................................................... ..... 3-10
Installing the Infra600 Software ............................................................... ..... 3-10
Setting the Converter Box Address ......................................................... ..... 3-11
Setting the Sensor Head Address ........................................................... ..... 3-12
Normal Operation ................................................................................................ 3-13
Setting Pyrometer Parameters ................................................................ ..... 3-13
Parameterizing with the Universal Pyrometer Protocols (UPP)
Commands ...................................................................................... ..... 3-13
Parameterizing Devices Using the Converter Box Display ............... ..... 3-15
Parameterization Flowchart .............................................................. ..... 3-16
Exposure Time ........................................................................................ ..... 3-18
Emissivity ...................................................................................................... 3-18
Clear Time Interval for the Maximum Value Storage .................................... 3-19
Clear Time Interval ................................................................................. 3-19
Wait Time ................................................................................................ ..... 3-20
Temperature Display ............................................................................... ..... 3-20
Internal Pyrometer Temperature ............................................................. ..... 3-20
Limit Switch ............................................................................................. ..... 3-21
Hysteresis ..................................................................................................... 3-21
Maintenance ........................................................................................................ 3-21
Replacing the Sensor Head .......................................................................... 3-21
Periodic Calibration ................................................................................. ..... 3-22
Inspection and Cleaning .......................................................................... ..... 3-22
Refurbishment Schedule ......................................................................... ..... 3-22
Chapter 4. Troubleshooting and Technical Support
Organization
Troubleshooting ............................................................................................. ....... 4-1
Technical Support Organization ............................................................................ 4-3
Service, Repairs, and Upgrades .................................................................... ....... 4-3
Appendix A. Unit Dimensional Drawings
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C vTable of Contents

Appendix B. Unit Specifications
Physical Specifications ......................................................................................... B-1
Temperature Measurement Specifications .................................................... ...... B-2
Electrical Specifications ................................................................................. ...... B-4
Cooling Specifications .......................................................................................... B-5
Environmental Specifications ............................................................................... B-5
Appendix C. Spot Size Specifications
IN 600 and IN 600/5 Pyrometer Spot Sizes ................................................... ...... C-1
IS 600 Pyrometer Spot Sizes ............................................................................... C-2
IGA 600 and IGA 600/23 Pyrometer Spot Sizes .................................................. C-2
Appendix D. Factory Settings
Appendix E. Pyrometer Parameters (Converter Box with
Display)
Appendix F. UPP Commands
Appendix G. Emissivity of Common Materials
Appendix H. Wiring Diagrams for Multiple Analog Outputs
Possible wiring for differential voltage inputs ................................................. ...... H-1
Possible wiring for simple, non-differential voltage inputs .................................... H-2
Possible wiring for galvanically separated analog inputs ............................... ...... H-3
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C viTable of Contents

List of Tables
Table 2-1. Sensor head connector pin assignments (male) .................................. 2-3
Table 2-2. Sensor cable connector pin assignments (female) ....................... ....... 2-4
Table 3-1. Soft key functions ............................................................................... 3-15
Table 3-2. Clear time interval settings ................................................................. 3-20
Table 4-1. Troubleshooting checklist ............................................................. ....... 4-1
Table B-1. Physical specifications ........................................................................ B-1
Table B-2. Temperature measurement specifications ................................... ...... B-2
Table B-3. Electrical specifications ................................................................ ...... B-4
Table B-4. Air cooling specifications .............................................................. ...... B-5
Table B-5. Standard environmental specifications ............................................... B-5
Table B-6. Climatic specifications ................................................................. ...... B-5
Table C-1. Optics 1N (2:1), −40°C to 700°C (−40°F to 1292°F) (MB 7) ........ ...... C-1
Table C-2. Optics 2N (10:1), −40°C to 700°C (−40°F to 1292°F) (MB 7) ...... ...... C-1
Table C-3. Optics 3N (20:1), −40°C to 700°C (−40°F to 1292°F) (MB 7) ...... ...... C-1
Table C-4. Optics (65:1) ...................................................................................... C-2
Table C-5. Optics (70:1) ....................................................................................... C-2
Table D-1. Default factory settings ....................................................................... D-1
Table E-1. Converter box parameters .................................................................. E-1
Table E-2. Sensor head parameters .................................................................... E-2
Table F-1. Sensor head UPP commands ...................................................... ...... F-1
Table F-2. Converter box and analog box UPP commands ................................. F-4
Table G-1. Emissivities for spectral range: 0.7 μm to 1.1 μm ........................ ...... G-1
Table G-2. Emissivities for spectral range: 1.45 μm to 1.8 μm ...................... ...... G-1
Table G-3. Emissivities for spectral range: 2 μm to 2.6 μm ........................... ...... G-2
Table G-4. Emissivities for spectral range: 8 μm to 14 μm ............................ ...... G-3
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C viiList of Tables

List of Figures
Figure 2-1. Standard sensor head connector (male) ..................................... ....... 2-3
Figure 2-2. Extension cable connectors ................................................................ 2-4
Figure 2-3. Sensor cable connector (female) ........................................................ 2-4
Figure 2-4. Converter box connectors ................................................................... 2-5
Figure 2-5. Single-analog box connectors ..................................................... ....... 2-6
Figure 2-6. Multi-analog box connectors ............................................................... 2-6
Figure 2-7. Multi-sensor box connectors ............................................................... 2-7
Figure 3-1. Spot size from any measuring distance (ratio) ............................ ....... 3-2
Figure 3-2. Ground connection for the converter box .................................... ....... 3-3
Figure 3-3. Connecting the sensor cable ....................................................... ....... 3-3
Figure 3-4. Connected sensor cable ..................................................................... 3-4
Figure 3-5. Converter box display and soft keys ................................................. 3-15
Figure 3-6. Parameterization process flowchart .................................................. 3-17
Figure A-1. Sensor head dimensions with sensor cable ................................ ...... A-1
Figure A-2. High temperature sensor head dimensions with sensor cable .......... A-1
Figure A-3. Sensor head dimensions with sensor cable to USB .......................... A-1
Figure A-4. Converter box and multi-analog box dimensions ........................ ...... A-2
Figure A-5. Single-Analog box dimensions .......................................................... A-3
Figure A-6. Multi-sensor box dimensions ............................................................. A-4
Figure H-1. Wiring example for differential voltage inputs ............................. ...... H-2
Figure H-2. Wiring example for simple, non-differential voltage inputs ................ H-3
Figure H-3. Wiring example for galvanically separated input modules .......... ...... H-4
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C viiiList of Figures

Safety and Product Compliance
Guidelines
IMPORTANT SAFETY INFORMATION
To ensure safe installation and operation of the Advanced Energy Impac unit, read
and understand this manual before attempting to install and operate this unit. At a
minimum, read and follow the safety guidelines, instructions, and practices.
DANGER, WARNING, AND CAUTION BOXES
This symbol represents important notes concerning potential harm to people, this
unit, or associated equipment. Advanced Energy includes this symbol in danger,
warning, and caution boxes to identify specific levels of hazard seriousness.
DANGER:
DANGER indicates an imminently hazardous situation that, if not avoided,
will result in death or serious injury. DANGER is limited to the most extreme
situations.
WARNING:
WARNING indicates a potentially hazardous situation that, if not avoided,
could result in death or serious injury, and/or property damage.
CAUTION:
CAUTION indicates a potentially hazardous situation that, if not avoided,
could result in minor or moderate injury, and/or property damage. CAUTION
is also used for property-damage-only accidents.
SAFETY GUIDELINES
Review the following information before attempting to install and operate the
product.
Impac® Series 600 Pyrometer Chapter
1
57010228-00C Safety and Product Compliance Guidelines 1‑1

Rules for Safe Installation and Operation
Please note the following rules:
• Ensure that this unit is properly grounded.
• Ensure that all cables are properly connected.
• Verify that input voltage and current source capacity are within specifications
before turning on the unit.
• Use proper electrostatic discharge (ESD) precautions.
• Maintenance and service must be performed only by AE-trained service
personnel.
INTERPRETING PRODUCT LABELS
The following labels might appear on your unit:
or
CE label
Complies with applicable European
directives.
UK Conformity Assessed label
Complies with applicable United
Kingdom regulations.
Chassis ground terminal
This terminal is available to connect
the chassis to a specific ground. It
is not required to be connected to
an external ground to meet safety
requirements. It is commonly used for
electromagnetic compatibility (EMC)
grounding of the chassis. Refer to the
installation instructions.
or Refer to manual for more information
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Safety and Product Compliance Guidelines 1‑2

PRODUCT COMPLIANCE
The following sections include information about unit compliance and certification,
including the conditions of use required to be in compliance with the standards and
directives.
Product Certification
Certain options of this product may be certified according to the list below.
For more information, refer to the Certificate or Letter of Conformity (US) or
Declaration of Conformity (EU) accompanying the product.
• CE Marking – Self-declaration, assessed by AE Corporate Compliance
• EMC measurements – Verified by AE Corporate Compliance
☞ Important
The calibration/adjustment of the pyrometer was carried out in accordance
with VDI/VDE directive “Temperature measurement in industry, Radiation
thermometry, Calibration of radiation thermometers,” VDI/VDE 3511, Part
4.4.
The terminology used in this manual corresponds to VDI/VDE 3511, Part 4.
For additional details on this directive, visit https://www.advancedenergy.com/
resource-library, locate the entry for Pyrometer Calibration Geometry, and
click View Resource. You can also order the directive from Beuth Verlag
GmbH in D-10772 Berlin Germany (https://www.beuth.de/en).
Safety and EMC Regulations, Directives, and Standards
For information concerning compliance to applicable EU requirements, refer
to the EU Declaration of Conformity shipped with this unit. For information
concerning compliance to applicable UK requirements, refer to the UK Declaration
of Conformity for this unit. The Declaration of Conformity might also include a
supplementary section covering compliance to other EU regulatory requirements
and/or industry standards or guidelines.
Conditions of Use
To comply with the stated directives and standards, you must meet the following
conditions of use:
• Install and operate this unit in a pollution degree environment according to
environmental specifications.
• Operate this device within the ambient temperature declared in the
specifications.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Safety and Product Compliance Guidelines 1‑3

• Use only a shielded cable for communications and/or control connections.
• Use only a shielded cable for the input power connections.
• Use only a shielded cable for the output process power connections.
• To prevent electromagnetic interference, connect the product to ground.
• Dispose of this product as directed by applicable laws and regulations.
• For the input and output power connections, use wires that are suitable for at
least the maximum allowed ambient temperature of the unit.
• Do not allow condensation of any liquids or accumulation of conductive dust on
the instrument. Doing so will have unpredictable results, possibly including, but
not limited to, loss of accuracy.
• Advanced Energy requires that AE-trained personnel perform all maintenance
and service on the unit.
Environmental Compliance
•EU REACH – European Union Regulation (EC) No. 1907/2006
Registration, Evaluation, Authorization and Restriction of Chemicals
Advanced Energy manufactures articles subject to Article 33 of REACH
and, upon request, will provide information regarding Substances of Very
High Concern (SVHC) currently identified by the European Chemical Agency
(ECHA) that are contained in this product, at concentrations greater than 0.1%
by weight.
•EU RoHS – European Union Directive 2011/65/EU (RoHS 2)
Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment
This product is EU 2011/65/EU (RoHS 2) compliant, containing no more than
the maximum concentration of hazardous substances listed in Annex II, but
might utilize application exemptions in Annex III or IV.
European Delegated Directive (EU) 2015/863 (RoHS 3) – Amendment
to Annex II of Directive 2011/65/EU (RoHS 2) Regarding the List of
Restricted Substances
This product is also compliant to the 2015/863 amendment, containing no
more than the maximum concentration of four additional hazardous substances
(phthalates) listed in amended Annex II.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Safety and Product Compliance Guidelines 1‑4

Product Overview
GENERAL DESCRIPTION
All substances with a temperature above absolute zero emit radiant energy. Thermal
radiation increases with an increase in temperature. An optical pyrometer measures
the radiation at a given wavelength, and from that radiant flux, calculates the absolute
temperature of the target. Practically, an optical pyrometer is a light intensity meter
calibrated to read out in temperature units.
Each Impac pyrometer has customized optics that measure thermal radiation from a
target object at a distance and determine the temperature in real time. Your pyrometer
was factory-calibrated to ensure that it operates within specifications. For more
information, refer to the work certificate or inspection sheet provided with your
pyrometer.
UNIT COMPONENTS
Pyrometer
The configuration and accessories of your pyrometer will vary depending on the
needs of your application. It has a modular design allowing you to connect and swap
multiple sensor heads. With the exception of USB sensor heads, your configuration
will include various sensor heads and sensor cables that connect to either a converter
box or an analog box (for communications), with the addition of optional multi-
sensor boxes (for connecting more sensor heads) and accessories.
For more information about the options available for your pyrometer, see the online
product data sheet.
1. Navigate to https://www.advancedenergy.com/resource-library.
2. Select the Data Sheets resource type.
3. Search for your pyrometer model.
4. Click View Resource.
Sensor Heads
A variety of sensor heads are available for non-contact temperature measurement at
various wavelengths. All sensor heads can be interchanged with any other standard
sensor head without restriction. When connecting more than one sensor head to the
converter box, analog box, or multi-sensor box, you can use any combination of the
available sensor heads.
Impac® Series 600 Pyrometer Chapter
2
57010228-00C Product Overview 2‑1

Sensor heads are attached to a 300 mm (11.8″) cable with a male connector on the
end for connecting them to a sensor cable or extension cable. The only exception
are the high temperature sensor heads, which have an additional 3 m (9′10″) cable
attached behind the sensor head to separate the electronics.
Sensor Cables
Sensor cables are not included with your pyrometer. You can order sensor cables and
extension cables of various lengths from AE for connecting any sensor head to a
converter box, multi-sensor box, analog box, or for connecting a multi-sensor box to
a converter box or multi-analog box. A variety of electrical cables of different lengths
are also available to order from AE for connecting signal output, power supply, and
serial communications.
Converter Box
The converter box has two sensor head connections and is equipped with one of the
following communication modules:
• RS-232
• RS-485
• PROFINET
The converter box can also be equipped with a display for temperature reading and
parameter setting via an integrated keyboard.
Analog Box
The analog box is an analog-output only alternative to the converter box and is
available in two options. The single-analog box has one analog output and one sensor
head connection. The multi-analog box has eight analog outputs and two sensor head
connections. No display or keys are equipped on the analog box.
Multi-Sensor Box
The multi-sensor box can be connected to a sensor head connection on the converter
box or the multi-analog box (instead of a sensor head) to enable you to use more
than two sensor heads at a time. Up to four sensor heads can be connected to
each multi-sensor box, allowing the connection of up to eight sensor heads to one
converter box or one multi-analog box.
Accessories
Electrical and mechanical accessories are available from AE for your pyrometer.
Only use accessories ordered from AE. If you use third party accessories, AE cannot
guarantee compatibility and will not service them.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Product Overview 2‑2

UNIT CONNECTORS
Depending on the setup for your application, the pyrometer might use one or several
of the following connectors.
Sensor Head Connectors
Sensor heads connect to a sensor cable or an extension cable using a 3-pin connector
(male) as shown in the following figure.
Figure 2‑1. Standard sensor head connector (male)
Table 2‑1. Sensor head connector pin assignments (male)
Pin Wire Color
1 Brown
4 Black
3 Blue
Extension Cable Connectors
Extension cables for connecting a sensor head to a sensor cable have two 3-pin
connectors as shown in the following figure.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Product Overview 2‑3

Figure 2‑2. Extension cable connectors
Sensor Cable Connectors
Sensor cables for connecting a sensor head or an extension cable to a converter box,
multi-sensor box, or analog box have one 3-pin connector (female) as shown in the
following figure.
Sensor cables for connecting a sensor head directly to a PC via USB are also
available. These cables will not work with extension cables.
Sensor cables for connecting a multi-sensor box to a converter box do not have any
connectors.
Figure 2‑3. Sensor cable connector (female)
Table 2‑2. Sensor cable connector pin assignments (female)
Pin Wire Color
3 Blue
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Product Overview 2‑4

Table 2‑2. Sensor cable connector pin assignments (female) (Continued)
Pin Wire Color
4 Black
1 Brown
Converter Box Connectors
Wires for each connection to the converter box must be terminated and connected
individually to the appropriate pins in the terminal blocks as shown in the following
figure.
Micro USB
Serial communications
Signal output and power supply
Sensor heads
Figure 2‑4. Converter box connectors
Analog Box Connectors
Wires for each connection to the analog box must be terminated and connected
individually to the appropriate pins in the terminal blocks as shown in the following
figures.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Product Overview 2‑5

Micro USB
Sensor head
Relay contact, analog output, power supply
Figure 2‑5. Single-analog box connectors
Micro USB
Sensor headsAnalog outputs
Relay contacts and power supply
Figure 2‑6. Multi-analog box connectors
Multi-Sensor Box Connectors
Wires for each connection to the multi-sensor box must be terminated and connected
individually to the appropriate pins in the terminal blocks as shown in the following
figure.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Product Overview 2‑6

Converter box
Sensor heads
Figure 2‑7. Multi-sensor box connectors
SOFTWARE
Though your pyrometer will work without software, you must use the Infra600
software if you want to connect it to a PC and perform certain tasks (for example,
adjusting the parameters, visualizing the temperature measurements, or analyzing the
data).
• The Infra600 software is included as a free download with your pyrometer.
• You can use the optional MultiTempLogger software for multi-channel logging,
storage, and display of temperatures and their temporal development, and for
setting parameters of multiple pyrometers.
• You can use the optional MultiTempLogger Trigger software (a more advanced
version of the MultiTempLogger) software for an additional feature to define
trigger thresholds from an I/O module with up to eight digital inputs.
• You can use the optional Temperature Quality Control System software to
record and archive process temperatures from multiple pyrometers for order-
related, manipulation-safe temperature recordings of every single work piece in
applications such as forging, pouring, and induction heating.
Advanced Energy®Impac® Series 600 Pyrometer
57010228-00C Product Overview 2‑7

Installation, Setup, and
Operation
PREPARING TO INSTALL THE UNIT
Unpacking the Unit
☞ Important
The labels on the packaging provide important safety and handling
information. Follow your company procedures when unpacking the unit.
1. Unpack and inspect the unit carefully, looking for obvious physical damage.
2. If no damage is apparent, proceed with the unit installation and setup.
3. If you do see signs of shipping damage, contact AE Technical Support
Organization and the carrier immediately.
Save the shipping container for submitting necessary claims to the carrier.
Measuring Through External Windows
If your application requires that you measure through an external window that is
between the pyrometer and the target object, ensure that the window is clean and
made of a material that is suitable for the wavelength of your pyrometer. See
Appendix G, “Emissivity of Common Materials”.
Verifying the Spot Size and Installation Location
The spot size of your pyrometer is determined by the measuring distance (distance
from the front of the pyrometer to the target), aperture, and temperature range. To
ensure accurate temperature measurement values when using your pyrometer, the
target object needs to be as big as or bigger than the spot size of the pyrometer.
INSTALLATION AT ANY MEASURING DISTANCE
Your pyrometer can be used at any measuring distance; however, the greater the
measuring distance, the larger the spot size will be. You can estimate the spot size by
using the distance ratio of the sensor head optics.
Impac® Series 600 Pyrometer Chapter
3
57010228-00C Installation, Setup, and Operation 3‑1
Table of contents