Table of Contents
1. INTRODUCTION............................................................................... 2
1.1 International Electrical Symbols................................................2
1.2 Receiving Your Shipment..........................................................3
1.3 Ordering Information.................................................................3
1.3.1 Accessories and Replacement Parts............................3
2. PRODUCT FEATURES ...................................................................... 4
2.1 Indicators and Buttons ..............................................................4
2.2 Inputs and Outputs ...................................................................5
2.3 Mounting ...................................................................................5
3. SPECIFICATIONS............................................................................. 6
3.1 ElectricalSpecications ............................................................6
3.2 MechanicalSpecications ........................................................6
3.3 EnvironmentalSpecications....................................................7
3.4 SafetySpecications ................................................................7
4. OPERATION .................................................................................... 8
4.1 Software Installation .................................................................8
4.1 Recording Data.........................................................................9
4.3 Using the Software .................................................................10
5. MAINTENANCE ............................................................................. 11
5.1 Battery Installation ..................................................................11
5.2 Cleaning.................................................................................. 11
APPENDIX A ....................................................................................... 12
Importing .TXT Files into a Spreadsheet.........................................12
OpeningaSimpleLogger.TXTleinExcel ...........................12
Formatting the Date and Time ................................................13
Repair and Calibration...........................................................................15
Technical and Sales Assistance ............................................................15
Limited Warranty ...................................................................................16
Warranty Repairs...................................................................................16

2
Thermocouple Simple Logger®Models L610 / L620 / L630
CHAPTER 1
INTRODUCTION
Warning
These safety warnings are provided to ensure the safety of per-
sonnel and proper operation of the instrument.
• Read the instruction manual completely and follow all safety
information before operating this instrument.
• Use caution on any circuit: Potentially high voltages and cur-
rents may be present and may pose a shock hazard.
• Readthespecicationssectionpriortousingthedatalogger.
Neverexceedthemaximumvoltageratingsgiven.
• Safety is the responsibility of the operator.
• For maintenance, use only original replacement parts.
• NEVER open the back of the instrument while connected to
any circuit or input.
• ALWAYS connect the leads to the logger before inserting the
leads to the test voltage
• ALWAYS inspect the instrument and leads prior to use.
Replace any defective parts immediately.
• NEVER use the Simple Logger®Models L610, L620 and
L630 on electrical conductors rated above 30V in overvolt-
age category III (CAT III).
1.1 International Electrical Symbols
Thissymbolsigniesthattheinstrumentisprotectedbydoubleor
reinforcedinsulation.Useonlyspeciedreplacementpartswhen
servicing the instrument.
This symbol on the instrument indicates a WARNING and that
the operator must refer to the user manual for instructions before
operating the instrument. In this manual, the symbol preceding
instructions indicates that if the instructions are not followed, bodily
injury, installation/sample and product damage may result.
Risk of electric shock. The voltage at the parts marked with this
symbol may be dangerous.

Thermocouple Simple Logger®Models L610 / L620 / L630
3
1.2 Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are consistent
with the packing list. Notify your distributor of any missing items. If the equip-
mentappearstobedamaged,leaclaimimmediatelywiththecarrierand
notify your distributor at once, giving a detailed description of any damage.
Save the damaged packing container to substantiate your claim.
1.3 Ordering Information
Simple Logger®Thermocouple Model L610.................... Cat. #2116.15
(Type J)
Simple Logger®Thermocouple Model L620.................... Cat. #2116.16
(Type K)
Simple Logger®Thermocouple Model L630.................... Cat. #2116.17
(Type T)
All models are supplied with software (CD-ROM), 6 ft DB-9 RS-232 serial cable, 9V Alkaline
battery and user manual.
1.3.1 Accessories and Replacement Parts
One 6 ft RS-232 cable with DB9F ....................................... Cat. #2114.27
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com for availability

4
Thermocouple Simple Logger®Models L610 / L620 / L630
CHAPTER 2
PRODUCT FEATURES
2.1 Indicators and Buttons
The Simple Logger®has one button and one indicator. Both are located
on the front panel. The PRESS button is used to start and stop recordings
and to turn the logger on and off.
The Red LED indicates the status of the logger:
•Single Blink: STAND-BY mode
•Double Blink: RECORD mode
•Continuously On: OVERLOAD condition
•No Blinks: OFF mode
1
2
34
(1) Start/Stop Button
(2) Indicator Light
(3) Miniature T.C. Connecto
r
(4) RS-232 Interface
SIMPLE LOGGER
®
THERMOCOUPLE MODULE L620
®
INSTRUMENTS
30 V
CAT III
K
!RS-232
!INPUT
PRESS
LOGGER OPERATION:
1. Press to Start Logging
(Logging = Two Blinks)
2. Press to Stop Logging
(Stand-By = One Blink)
3. Press & Hold 5 sec. to Shut Off
(Off = No Blinks)
Warning: All Data will be Deleted!
!
Continuous LED indicates OUT OF RANGE

Thermocouple Simple Logger®Models L610 / L620 / L630
5
2.2 Inputs and Outputs
The left side of the Simple Logger®incorporates a color-coded miniature
T.C. connector that is compatible with the thermocouple for which your
logger was designed.
The right side of the logger has a female 9-pin “D” shell serial connector
used for data transmission from the logger to your computer.
2.3 Mounting
Your Simple Logger®is equipped with clearance holes in the base plate
tabs for mounting. For less permanent mounting, the Velcro®pads (sup-
plied loose) can be attached to the logger and the surface to which the
logger will be mounted.

6
Thermocouple Simple Logger®Models L610 / L620 / L630
CHAPTER 3
SPECIFICATIONS
3.1 Electrical Specifications
Number of Channels: 1
Measurement Range: L610: 32° to 1380°F (0° to 750°C)
L620: -325°to 2280°F (-200°to 1250°C)
L630: -325°to 660°F (-200°to 350°C)
Input: L610: J Thermocouple
L620: K Thermocouple
L630: T Thermocouple
Input Connection: Miniature color-coded thermocouple jacks
Black- J, Yellow-K, Blue-T
Resolution: 12 Bit (better than 0.5°C)
Sample Rate: 4096/hrmax;decreasesby50%eachtimememoryisfull
Data Storage: 16,384 readings
Data Storage Technique: TXR™TimeExtensionRecording™
Power: 9V Alkaline NEDA 1604, 6LF22, 6LR61
Battery Life Recording: Up to 1 year of recording @ 77°F (25°C)
Output: RS-232 via DB9 connector (4800 Baud)
Reference condition: 23°C ± 3K, 20 to 70% RH, Frequency 50/60Hz, No AC external mag-
netic eld, DC magnetic eld ≤ 40A/m, battery voltage 9V ± 10%.
Accuracy: ±0.5% of Reading + 1°C + T/C Accuracy
3.2 Mechanical Specifications
Size: 2-7/8x2-5/16x1-5/8"(73x59x41mm)
Weight (with battery): 5 oz (140g)
Mounting: Base plate mounting holes or Velcro®pads
Case Material: Polystyrene UL V0

Thermocouple Simple Logger®Models L610 / L620 / L630
7
3.3 Environmental Specifications
Operating Temperature: -4 to 158°F (-20 to 70°C)
Storage Temperature: -4 to 176°F (-20 to 80°C)
Relative Humidity: 5to95%non-condensing
Temperature Inuence: 0.1%/°Coverfulltemperaturerange
3.4 Safety Specifications
Working Voltage: EN 61010, 30V, Cat III
*All specications are subject to change without notice

8
Thermocouple Simple Logger®Models L610 / L620 / L630
CHAPTER 4
OPERATION
4.1 Software Installation
Minimum Computer Requirements
• Windows®7, 8, 8.1, 10
• Processor - 486 or higher
• 8MB of RAM
•8MBofharddiskspaceforapplication,400Kforeachstoredle
•One9-pinserialport;oneparallelportforprintersupport
• CD-ROM drive
1. Insert the Simple Logger®CD into your CD-ROM drive.
If auto-run is enabled, the Setup program will start automatically. If
auto-run is not enabled, select Run from the Start menu and type in
D:\SETUP (if your CD-ROM drive is drive D. If this is not the case,
substitute the appropriate drive letter).
2. The Set-up window will appear.
Figure 1

Thermocouple Simple Logger®Models L610 / L620 / L630
9
There are a several options to choose from. Some options(*)require an
internet connection.
• Simple Logger, Version 6.xx - Installs the Simple Logger® soft-
ware to the computer.
• *Acrobat Reader - Links to the Adobe®web site to download the
most recent version of Adobe®Acrobat Reader.
Acrobat Reader is
required for viewing PDF documents supplied on the CD-ROM.
• *Check for Available Software Updates - Opens the AEMC
Software update web site, where updated software versions are
available for downloading, if necessary.
• View User Guide and Manuals - Opens Windows®Explorerfor
viewingofdocumentationles.
3. To install the software, select Simple Logger Software Setup in
the top section of the Set-up window, then select Simple Logger,
Version 6.xx in the Options section.
4. Click the Install button and follow the on-screen prompts to install the
software.
4.1 Recording Data
• Connect the temperature probe to the logger.
NOTE: Make sure that the temperature sensor is set where you want
to measure before starting a recording session.
• Press the PRESS button on the front of the logger to begin the record-
ing session. The LED indicator will double-blink to indicate that the
recording session has started.
• When the recording session has been completed, press the PRESS
button to end the recording. The LED indicator will single-blink to indi-
cate that the recording session has ended and the unit is in Stand-by.
• Connect the logger to the computer for data downloading. See the
User Guide on the CD-ROM for downloading.
NOTE: If no T.C. element is connected to the logger, an Overload Condi-
tion will occur when in record mode or connected to the PC. To eliminate
the Overload Condition when in record mode, press the PRESS button
once to place the instrument into Stand-by.

10
Thermocouple Simple Logger®Models L610 / L620 / L630
4.3 Using the Software
Launch the software and connect the RS-232 cable from your computer
to the logger.
NOTE:Thersttimetheprogram islaunchedyouwillneed toselecta
language.
Select Port from the menu bar and select the Com port (COM 1, 2 3 or 4)
you will be using (see your computer manual). Once the software automati-
cally detects the baud rate, the logger will communicate with the computer.
(ID number of the logger and number of points recorded displayed).
Select Function from the menu bar to select °F or °C.

Thermocouple Simple Logger®Models L610 / L620 / L630
11
CHAPTER 5
MAINTENANCE
5.1 Battery Installation
Under normal conditions, the battery will last up to a year of continuous
recording unless the logger is restarted very frequently.
In the OFF mode, the logger puts almost no load on the battery. Use the
OFF mode when the logger is not in use. Replace the battery once a year
in normal use.
If the logger will be used at temperatures below 32°F (0°C) or is frequently
turnedonandoff,replacethebatteryeverysixtoninemonths.
1. Make sure your logger is turned off (no blinking light) and all inputs
are disconnected.
2. Turn the logger upside down. Remove the four Phillips head screws
from the base plate, then take off the base plate.
3. Locate the two-wire (red/black) battery connector and attach the 9V
battery to it. Make sure that you observe polarity by lining up the
battery posts to the proper terminals on the connector.
4. Once the connector is plugged onto the battery, insert the battery
into the holding clip on the circuit board.
5. If the unit is not in record mode after installing the new battery, dis-
connect it and press the button twice then reinstall the battery.
6. Reattach the base plate using the four screws removed in Step 2.
Your logger is now recording (LED blinking). Press the PRESS button for
vesecondstostoptheinstrument.
NOTE: For long-term storage, remove the battery to prevent discharge
effects.
5.2 Cleaning
The body of the logger should be cleaned with a cloth moistened with
soapy water. Rinse with a cloth moistened with clean water. Do not use
solvent.

12
Thermocouple Simple Logger®Models L610 / L620 / L630
APPENDIX A
Importing .TXT Files into a Spreadsheet
Opening a Simple Logger .TXT file in Excel
ThefollowingexampleusedwithExcelVer.7.0orhigher.
1. After opening the Excel program, select “File” from the main
menu and then select “Open”.
2. In the dialog box that appears, browse and open the folder
where your logger .TXT les are stored. This will be located
in C:\Program Files\Simple Logger 6.xx if you accepted the
default choice offered by the logger installation program.
3. Next,changetheletypeto“Text Files”intheeldlabeledFiles
of Type.Allthe.TXTlesintheloggerdirectoryshouldnowbe
visible.
4. Double-clickonthedesiredletoopentheTextImportWizard.
5. Reviewtheselectionsintherstwizardscreenandmakesure
that the following choices are selected:
Original Data Type: Delimited
Start Import at Row: 1
File Origin: Windows (ANSI)
6. Click the “NEXT”buttonatthebottomoftheWizarddialogbox.
The second wizard screen will appear.
7. Click on “Comma”intheDelimitersbox.Acheckmarkshould
appear.
8. Click the “NEXT” buttonatthebottomoftheWizarddialogbox.
The third wizard screen will appear.
9. A view of the actual data to be imported should appear in the
lower section of the window. Column 1 should be highlighted. In
the Column Data Format window, select “Date”.
10. Next,clickon“Finish” to complete the process and import the
data.
11. The data will now appear in your spreadsheet in two columns (A
and B) and will look similar to that shown in Figure A-1.

Thermocouple Simple Logger®Models L610 / L620 / L630
13
A B
8 Arms
35401.49 3.5
35401.49 5
35401.49 9
35401.49 13.5
35401.49 17
35401.49 20
35401.49 23.5
35401.49 27.5
35401.49 31
35401.49 34.5
35401.49 38
Figure A-1. Sample Data Imported into Excel.
Formatting the Date and Time
Column ‘A’ contains a decimal number that represents both date and time.
Excelcanconvertthisnumberdirectlyasfollows:
1. Click on column ‘B’ at the top of the column to select the data,
then click on “Insert” from the main menu and select “Columns”
from the drop-down menu.
2. Next,clickoncolumn‘A’atthetopofthecolumntoselectthe
data, then click on “Edit” from the main menu and select “Copy”
to copy the entire column.
3. Click on cell 1 of column ‘B’ and then click on “Edit” and select
“Paste” to insert a duplicate of column ‘A’ into column ‘B’. This is
necessary if you want to show the date and time in two separate
columns.
4. Next,clickonthetopofcolumn‘A’,thenclickon“Format” and
select “Cells” from the drop-down menu.

14
Thermocouple Simple Logger®Models L610 / L620 / L630
5. Inthedialogboxthatopens,selectthe“Date” option from the
category list on the left. Select the date format you desire and
click on “OK” to format the column.
6. Click on the top of column ‘B’, then click on “Format” and select
“Cells” from the drop-down menu.
7. Inthedialogboxthatopens,selectthe“Time” option from the
category list on the left. Select the time format you desire and
click on “OK” to format the column.
FigureA-2 shows a typical spreadsheet with date, time and value displayed.
It may be necessary to change the column width to see all the data.
A B C
12/02/04 11:45 AM 17
12/02/04 11:45 AM 20
12/02/04 11:45 AM 23.5
12/02/04 11:45 AM 27.5
12/02/04 11:45 AM 31
12/02/04 11:45 AM 34.5
12/02/04 11:45 AM 38
12/02/04 11:45 AM 41.5
12/02/04 11:45 AM 45.5
12/02/04 11:46 AM 49
12/02/04 11:46 AM 52
Figure A-2. Shows Date, Time and Value

Thermocouple Simple Logger®Models L610 / L620 / L630
15
Repair and Calibration
Toensurethatyourinstrumentmeetsfactoryspecications,werecommend
that it be scheduled back to our factory Service Center at one-year intervals
for recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization
Number (CSA#). This will ensure that when your instrument arrives, it will be
tracked and processed promptly. Please write the CSA# on the outside of the
shipping container. If the instrument is returned for calibration, we need to
know if you want a standard calibration, or a calibration traceable to N.I.S.T.
(Includescalibrationcerticateplusrecordedcalibrationdata).
Ship To:ChauvinArnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive
Dover, NH 03820 USA
Phone:(800)945-2362(Ext.360)
(603) 749-6434(Ext.360)
Fax: (603)742-2346or(603)749-6309
E-mail: [email protected]
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are
available.
NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewith
theproperoperationorapplicationofyourinstrument,pleasecall,mail,faxor
e-mail our technical support team:
ChauvinArnoux®, Inc. d.b.a. AEMC®Instruments
200FoxboroughBoulevard
Foxborough,MA02035USA
Phone: (800) 343-1391
(508) 698-2115
Fax: (508)698-2118
E-mail: [email protected]
www.aemc.com
NOTE: Do not ship Instruments to our Foxborough, MA address.

16
Thermocouple Simple Logger®Models L610 / L620 / L630
Limited Warranty
The Simple Logger®Model L610/L620/L630 is warranted to the own-
er for a period of two yearsfrom the date of original purchase against
defects in manufacture. This limited warranty is given by AEMC®In-
struments, not by the distributor from whom it was purchased. This
warranty is void if the unit has been tampered with, abused or if the defect is
related to service not performed by AEMC®Instruments.
For full and detailed warranty coverage, please read the Warranty
Coverage Information, which is attached to the Warranty Registration
Card (if enclosed) or is available at www.aemc.com. Please keep the
Warranty Coverage Information with your records.
What AEMC®Instruments will do:
If a malfunction occurs within the warranty period, you may return the in-
strument to us for repair, provided we have your warranty registration in-
formation on le or a proof of purchase. AEMC®Instruments will, at its
option, repair or replace the faulty material.
REGISTER ONLINE AT:
www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone
orbyfaxfromourServiceDepartment(seeaddressbelow),thenreturnthe
instrument along with the signed CSA Form. Please write the CSA# on the
outside of the shipping container. Return the instrument, postage or shipment
pre-paid to:
Ship To: ChauvinArnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone:(800)945-2362(Ext.360)
(603) 749-6434(Ext.360)
Fax: (603)742-2346or(603)749-6309
E-mail: [email protected]
Caution: To protect yourself against in-transit loss, we recommend you insure
your returned material.
NOTE: You must obtain a CSA# before returning any instrument.

08/18
99-MAN 100214 v11
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346
www.aemc.com