AEMC 1050 User manual

MEGOHMMETER 1050
1060
ENGLISH User Manual
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certifies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met its published specifications.
An NIST traceable certificate may be
requested at the time of purchase, or obtained
by returning the instrument to our repair and
calibration facility, for a nominal charge.
The recommended calibration interval for this
instrument is 12 months and begins on the date of
receipt by the customer. For recalibration, please
use our calibration services. Refer to our repair
and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2130.01 / 2130.03
Model #: 1050 / 1060
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

2
Megohmmeter Models 1050 /1060
Table of Contents
1. INTRODUCTION............................................................................... 5
1.1 International Electrical Symbols...........................................................5
1.2 DenitionofMeasurementCategories.................................................6
1.3 ReceivingYourShipment.....................................................................6
1.4 OrderingInformation ............................................................................6
1.4.1 AccessoriesandReplacementParts......................................7
1.4.2 AccessoryInformation ............................................................7
2. PRODUCT FEATURES ...................................................................... 8
2.1 ControlFeatures ................................................................................10
2.2 DigitalDisplayFeatures .....................................................................11
2.3 Bargraph ............................................................................................11
2.4 Symbols .............................................................................................12
2.5 ButtonFunctions ................................................................................13
2.5.1 2ndButton .............................................................................13
2.5.2 V-TIME / Button ..............................................................13
2.5.3 R-DAR-PI/R(t)Button..........................................................14
2.5.4 /ALARMButton ..............................................................16
2.5.5 /SMOOTHButton ............................................................17
2.5.6 Button ..............................................................................18
2.5.7 SET-UPButton(ConguringtheInstrument)........................18
2.5.8 ClearingtheMemory.............................................................20
2.5.9 CommunicationRate(RS-232).............................................20
2.5.10 LeadResistanceCompensation...........................................20
2.5.11 DefaultDeviceConguration ................................................21
2.5.12 Blocking(Disabling)TestVoltages........................................21
2.6 MeasurementFunctions.....................................................................22
2.6.1 AC/DCVoltage......................................................................22
2.6.2 InsulationMeasurement........................................................22
2.6.3 Continuity(40Ω )/Resistance(400kΩ)..........................23
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
3
3. SPECIFICATIONS........................................................................... 25
3.1 ReferenceConditions.........................................................................25
3.2 Voltage ...............................................................................................25
3.3 InsulationResistance.........................................................................25
3.4 Continuity ...........................................................................................29
3.5 Resistance .........................................................................................29
3.6 PowerSupply.....................................................................................30
3.7 MechanicalSpecications..................................................................31
3.8 EnvironmentalSpecications.............................................................31
3.9 SafetySpecications..........................................................................31
4. OPERATION .................................................................................. 32
4.1 MeasurementProcedure....................................................................32
4.2 InsulationMeasurement.....................................................................32
4.3 ContinuityMeasurement ....................................................................33
4.4 ResistanceMeasurement ..................................................................33
4.5 CapacitanceMeasurement ................................................................33
5. MEMORY / RS-232 (Model 1060)........................................... 34
5.1 RS-232Specications........................................................................34
5.2 Saving/RecallingValues(MEM/MRButton)..............................34
5.3 PrintingMeasuredValues ..................................................................36
5.4 InstantaneousPrintingofMeasurements(PRINTbutton).............37
5.5 PrintingDatainMemory(PRINTMEMbutton)...........................38
5.6 PrintingwiththeSerial-to-ParallelAdapter.........................................39
6. APPLICATION EXAMPLES ............................................................. 40
6.1 InsulationMeasurementsonElectricalInstallations...............................40
6.2 MeasurementsonElectricalorTelecomCable..................................40
6.3 InsulationMeasurementsonMotors..................................................41
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

4
Megohmmeter Models 1050 /1060
7. USING DATAVIEW®........................................................................ 42
7.1 InstallingDataView®...........................................................................42
7.2 ConnectingtheModel1060toyourComputer ..................................46
7.3 UsingDataView®................................................................................46
7.3.1 ConguringtheInstrument....................................................47
7.3.2 RunningtheTest ...................................................................49
8. MAINTENANCE ............................................................................. 52
8.1 BatteryReplacement(Model1050) ...................................................52
8.2 RechargingtheBattery(Model1060) ................................................52
8.3 FuseReplacement.............................................................................53
8.4 Cleaning.............................................................................................53
8.5 Storage...............................................................................................53
RepairandCalibration........................................................................................54
TechnicalandSalesAssistance .........................................................................54
Limited Warranty ................................................................................................55
WarrantyRepairs................................................................................................55
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
5
CHAPTER 1
INTRODUCTION
WARNING
These safety warnings are provided to ensure the safety of personnel and
properoperationoftheinstrument.
• Do not attempt to perform any tests with these instruments until you
havereadtheinstructionmanual.
• Safetyistheresponsibilityoftheoperator!
• Testsaretobecarriedoutonlyonnon-energizedcircuits!Checkforlive
circuitsbeforemakingresistancemeasurements(safetycheck).
• High voltage is present, as is the sample connected to it. Anyone
performing or assisting in testing must follow all safety precautions to
preventelectricalshocktothemselvesandtoothers.
• AEMC
®
considers the use of rubber gloves to be an excellent safety
practice, even if the equipment is properly operated and correctly
grounded.
• When testing samples with a capacitive component, make sure they
havebeenproperlydischargedandaresafetotouch.Dielectricinsula-
tionsamplesshouldbeshort-circuitedforatleastvetimestheamount
oftimetheywereenergized.
• Megohmmetersshouldneverbeusedinanexplosiveenvironment.
• Usetheleadssuppliedwiththemegohmmeters.Iftheyaredefectiveor
worn,replacebeforetesting.
•
Thisinstrumentcanbeusedoninstallationsratedfor600V,CategoryIII.
1.1 International Electrical Symbols
This symbol signifies that the instrument is protected by double or reinforced
insulation.
This symbol on the instrument indicates a WARNING and that the operator must refer
to the user manual for instructions before operating the instrument. In this manual,
the symbol preceding instructions indicates that if the instructions are not followed,
bodily injury, installation/sample and product damage may result.
Risk of electric shock. The voltage at the parts marked with this symbol may be
dangerous.
In conformity with WEEE 2002/96/EC
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

6
Megohmmeter Models 1050 /1060
1.2 Definition of Measurement Categories
CAT I: FormeasurementsoncircuitsnotdirectlyconnectedtotheACsupply
wall outlet such as protected secondaries, signal level, and limited
energycircuits.
CAT II: For measurements performed on circuits directly connected to the
electricaldistributionsystem.Examplesaremeasurementsonhousehold
appliancesorportabletools.
CAT III: Formeasurementsperformedinthebuildinginstallationatthedistribution
level such as on hardwired equipment in xed installation and circuit
breakers.
CAT IV: Formeasurementsperformedattheprimaryelectricalsupply(<1000V)
suchasonprimaryovercurrentprotectiondevices,ripplecontrolunits,
or meters.
1.3 Receiving Your Shipment
Uponreceivingyourshipment,makesurethatthecontentsareconsistentwiththe
packinglist.Notifyyourdistributorofanymissingitems.Iftheequipmentappears
tobedamaged,leaclaimimmediatelywiththecarrierandnotifyyourdistributor
atonce,givingadetaileddescriptionofanydamage.Savethedamagedpacking
containertosubstantiateyourclaim.
NOTE: Charge the instrument fully before use (Model 1060).
1.4 Ordering Information
Megohmmeter Model 1050.............................................................Cat. #2130.01
Includes detachable accessory pouch (one red, one blue test lead, one black shielded lead,
three color-coded (black, red and blue) alligator clips, one black test probe), batteries, spare
fuses and a user manual.
Megohmmeter Model 1060.............................................................Cat. #2130.03
Includes detachable accessory pouch (one red, one blue test lead, one black shielded lead,
three color-coded (black, red and blue) alligator clips, one black test probe; one RS-232 DB9
F/F 6 ft null modem cable, rechargeable battery, DataView®software, US 120V power cord,
spare fuses and a user manual.
NOTE: Sparefuseandcliparelocatedinsidetheinstrument’scase.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
7
1.4.1 Accessories and Replacement Parts
Remote Test Probe........................................................................... Cat. #2118.97
Cable, PC RS-232, DB9 F/F 6 ft Null Modem Cable (1060)............. Cat. #2119.45
Fuse, set of 5, 0.1A, 660V................................................................ Cat. #2119.56
Fuse, set of 1, 2.5A, 1200V.............................................................. Cat. #2119.57
Lead, Replacement Set (1 red, 1 blue test lead, 1 black shielded
lead, 3 color-coded alligator clips, 1 black test probe, no RS cable)
..... Cat. #2119.58
Inverter – 12VDC to 120VAC 200 Watt for vehicle us ........................Cat. #2135.43
Lead, set of 3 Color-coded 10 ft Safety Leads .................................Cat. #2951.70
Replacement 9.6V Rechargeable Battery Pack (Model 1060).........Cat. #2960.21
115V Power Cord .............................................................................Cat. #5000.14
Order Accessories and Replacement Parts Directly Online
Check our Storefront at www.aemc.com/store for availability
1.4.2 Accessory Information
Remote Test Probe
• Theyellowtestbuttongeneratesthetestvoltagewhenpressed.
• The push-button on the back of the probe allows you to light the test point
(approx 500 lux of light). This function is very useful, since insulation testing is
performed on de-energized installations, which can be in a dark area.
DataView®Software (Model 1060)
• Retrievedatafrommemoryandplotgraphsofthechangesininsulationasa
function of the time over which the test voltage is applied, R(t).
• Printoutprotocolsofpersonalizedtests(dependingontheuser’sneeds).
• Createtextlesforuseonspreadsheets.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

8
Megohmmeter Models 1050 /1060
CHAPTER 2
PRODUCT FEATURES
TheMegohmmetersModels1050and1060areportableinstrumentshousedin
ruggedcasing.
TheModel1050uses1.5V(Ccell)alkalinebatteries.
TheModel1060usesarechargeablebatteryandACpower.
Thesemegohmmetersaredesignedtocheckthesafetyofelectricalinstallations
andequipment.
Units Measure:
• Voltage
• InsulationResistance
• Continuity
• Resistance
• Capacitance
Advantages:
• Digitallteringofinsulationmeasurements
• Measuringwiththeremotecontrolprobe
• Automaticvoltagemeasurementinallfunctions
• AutomaticdetectionofexternalACorDCvoltagesontheterminals
• Thresholdprogrammingineachfunction,triggeringaudiblealarms
• Timedcontrolofmeasurementduration
• Fuseprotection,withdefectivefusedetectionandindication
• Automaticdischargeofresidualhighvoltageonthetesteddevice
• Automaticshut-offtosavethebatteries
• Batterylevelindicator(1050)orbatterychargestatusindicator(1060)
• Largeeasy-to-readback-litLCD
Additional Features (Model 1060 only):
• Integralrechargeablebattery
• 128kBmemory,real-timeclockandserialinterface
• ControlandprogrammingoftheinstrumentfromaPC
(withDataView®software)
• PrintinginRS-232orCentronicsmode
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
9
MODEL 1050
!
V/TIME
R-DAR-PI
MEGOHMMETER
MODEL 1050
START/STOP
400 KΩ
SET-UP
G
600V CAT III
OFF
40
Ω
MΩ-1000
V
MΩ- 500V
MΩ-250V
MΩ-100V
MΩ-50V
2nd
SMOOTH ALARM
R(t)
5
3
2
4
1
8.8.8.8
0
.
1
M
1
M
1
0
M
1
0
0
M
1
G
1
0
G
1
0
0
G
1
T
µA nA µF
COM
P
GΩ
PI
DAR
ALARM
TΩ
MΩkΩ
MEM MR
SMOOTH REMOTE
OBJ.
Ω
min sec
TEST
V
DC
AC
2nd
8.8.8.8
MODEL 1060
!
!
!
V/TIME
R-DAR-PI
MEGOHMMETER
MODEL 1060
START/STOP
RS-232
85-256 V
50/60 Hz
MEM PRINT
400 KΩ
SET-UP
G
600V CAT III
OFF
40
Ω
MΩ-1000
V
MΩ- 500V
MΩ-250V
MΩ-100V
MΩ-50V
2nd
SMOOTH ALARM
R(t)
MR PRINT MEM
5
3
2
6
4
1
7
8.8.8.8
0
.
1
M
1
M
1
0
M
1
0
0
M
1
G
1
0
G
1
0
0
G
1
T
µA nA µF
COM
P
GΩ
PI
DAR
ALARM
TΩ
MΩkΩ
MEM MR
SMOOTH REMOTE
OBJ.
Ω
minsec
TEST
V
DC
AC
2nd
8.8.8.8
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

10
Megohmmeter Models 1050 /1060
2.1 Control Features
1. Safetyterminals:“+”,“G”and“-”.
Nexttothe“-”terminal,thereare2additionalterminalcontactsfortheremote
controlprobeconnection(3contactconnector).
2. Rotaryselectorswitchwith9positions:
• OFF-Instrumentisoff
• MΩ- 50V-Insulationmeasurement(2kΩto 200GΩ)
• MΩ- 100V-Insulationmeasurement(4kΩto 400GΩ)
• MΩ- 250V-Insulationmeasurement(10kΩto 1TΩ)
• MΩ- 500V-Insulationmeasurement(20kΩto 2TΩ)
• MΩ- 1000V-Insulationmeasurement(40kΩto 4TΩ)
• 400KΩ-Resistancemeasurement
• 40Ω-Continuitymeasurement
• SET-UP-Set-upofthemegohmmeter
3. START/STOPbutton
4. Back-litliquidcrystaldisplay
5. 6buttons(1050)or8buttons(1060),eachwitharstandsecondfunction.
Thesecondfunctionsarehighlightedinyellowbeloweachbutton:
• 2nd-Selectsthesecondfunctiononeachbutton
• R-DAR-PI-Letsyouchoosethetypeofmeasurementtobeperformed.
InstantaneousResistance(R),DielectricAbsorption(DAR),orPolariza-
tionIndex(PI).
• R(t)-Show/hidetheintervalvaluesofinsulationresistance,testvoltage
andtime-stamping,followingatime-controlledtest.
• V/TIME-Whenmeasuring insulation,pressingthis buttondisplaysthe
time elapsed since the beginning of the measurement, then the exact
voltagegenerated.Whenperformingresistanceorcapacitancemeasure-
ment,thisbuttonhasnoeffect.InMR(memoryrecall),itdisplaysthedate
andtimeatwhichthemeasurementwasstored,theexacttestvoltage
andtheOBJ:TESTnumber.
•-Activatesordeactivatesthe“time-controlled”testmode
• -TurnsthedisplaybacklightONorOFF
• ALARM-ActivatesordeactivatesthealarmsprogrammedinSET-UP
•-Selectsaparametertobemodied
• SMOOTH-Stops/Startssmoothingofdisplayedvaluesduringinsulation
testing
• - Increases the ashing parameter being displayed. To move
about the list of interval insulation measurements, in the R(t) function.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
11
• - Decreases the ashing parameter being displayed. To move
about the list of interval insulation measurements in the R(t) function.
Iftheand buttonsarehelddown,themovementbetweenparam-
etersisincreasedtoafasterrate.
Model 1060 only:
• MEM-Savesmeasuredvalues
• MR - Recalls saved data
• PRINT-Printsmeasurementresults
• PRINT MEM-Printsmemorycontents
6. ACpowerplug(directoperationonACandbatteryrecharge).
7. RS-232serialinterfacemaleplug(9-pin)forconnectiontoaPCorprinter.
2.2 Digital Display Features
Main Display Indicates:
• Insulationmeasurement(resistance,DARandPI,capacitance)
• Continuitymeasurement
• Resistancemeasurement
Small Display Indicates:
• Voltagemeasuredorappliedbytheinstrument
• Elapsedtimeortheoutputvoltage,duringinsulationmeasurement
Afterrecordingdata(1060),thesmalldisplayalsoindicatesthetimeanddateinMR
(memoryrecall)mode,andthememoryaddresswiththeOBJ:TESTnumber.
2.3 Bargraph
• Activeduringinsulationmeasurement(0.1MΩto 1TΩ).
• Indicatesthebatterychargeatstart-up.
• Indicatesfreememoryspace-onesegmentrepresentingapproximately
100groupsofsavedvalues,witheachgroupholdingapproximately50
recordings(1OBJand1TEST).
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

12
Megohmmeter Models 1050 /1060
2.4 Symbols
MEM/MR:Memoryaddress(1060)-thenumberisdisplayedonthesmalldigital
display.
OBJ:TEST: Memoryaddress(1060)-thenumberisdisplayedonthesmalldigital
display.
COM:Flashesonthescreenwhendataistransmittedtotheserialinterface(1060)
orremainspermanentlydisplayedifthereisaproblemduringtransfer.
DAR/PI: Indicatesthemodechosenbeforeinsulationmeasurementortheresults
ofthesemeasurements.
Dangerousvoltagegenerated;V>120V.
External voltage present, symbol is activated after pressing START, if
V>25VAC±3Vor>35VDC.
Activationof“time-controlledtest”modeorclockadjustmentwheninSET-
UP(1060).
Indicatesthatthesecondfunctionofabuttonistobeused.
Flashes if the battery voltage is low, indicating that the batteries should
bereplacedorrecharged(See§8).Thevoltageisdisplayedonthesmall
displayfor2secondswhentheinstrumentisturnedon.Themaindisplay
indicates“bat”whenthetestiscarriedon.
Signalthatleadcompensationisineffect.
Warningbuzzerisactivated.
Indicatesthattheautomaticswitch-offfunctionhasbeendeactivated.
SMOOTH:Smoothstheinsulationmeasurementsdisplayed.
REMOTE: Remotecontrol via an interface (1060). In this mode, all the buttons
and the rotary switch on the instrument are non-functional, except for the OFF
position.
FUSE HI: Flashesifthe“+”inputfuseisdefective.
FUSE -G-: Flashesifthe“G”inputfuseisdefective.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
13
2.5 Button Functions
2.5.1 2nd Button
• Selectsthesecondfunction(highlightedinyellow)onthebuttons.
• The symbolappears.Thissymboldisappearsuponpressingthefunction
buttonchosen,exceptifthebuttonisactivated.Inthiscase,itonlydisap-
pearswhen the button is pressed again,or if other functionbuttons are
pressed. This allows you to rapidly decrease parameters with the button,
withouthavingtopressthe buttoneverytime.
2.5.2 V-TIME / Button
First Function - V/TIME
Displaysallthesecondaryinformationavailableonthesmalldisplay.
For Insulation Measurement:
• Thetimeelapsedsincethestartofthemeasurement
• Thevoltagebetweentheinstrument’s“+”and“–”terminals
• Date,time,testvoltageandOBJ:TESTnumberinmemoryrecallmode(MR)
For Resistance or Continuity Measurement:
• Voltagebetweentheinstrument’s“+”and“–”terminals
• Date,time,testvoltageandOBJ:TESTnumberinmemoryrecallmode(MR)
Second Function - Time-controlled Test
DisplaysthemeasurementdurationprogrammedinSET-UPonthesmalldisplay.
• The symbollightsup.
• PressingtheSTARTbuttonbeginsthemeasurementprocess.Themeasurement
durationissetto15minutesbydefault.
• Assoonasthemeasurementstarts,thesmalldisplaycountsdowntheremaining
durationofthemeasurement.Onceitreacheszero,themeasurementstops.
• Asthetime-controlledtestiscarriedout,intervalsamples(resistance/voltagevalues
asafunctionoftime)areautomaticallysaved.Thetimebetweeneachsampleisset
to30sbydefault,butitcanbechangedintheSET-UPmenu.
• ThesamplesaredisplayedwiththeR(t)function(see§2.5.3)aslongasanew
measurement has not been started. With each new measurement, the previous
samplevalueiserasedfromthememory.Thesesamplevaluesaresavedwiththe
lastvalueoftheresistanceifusedwiththeMEM (memorystorage)function.
NOTE: If the selector switch’s position is altered, or if the STOP button is pressed
during measurement, the measurement is interrupted.
This function is only active for insulation measurement.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

14
Megohmmeter Models 1050 /1060
2.5.3 R-DAR-PI / R(t) Button
First Function - R-DAR-PI
TheR-DAR-PIbuttonallowsautomaticmeasurementof:
• PolarizationIndex(PI)
• DielectricAbsorptionRatio(DAR)
Thesetwoparametersareusefulonlongcables,orwhenmonitoringtheagingof
insulationonelectricalmachines.
Onthismaterial,measurementisadverselyaffectedatthebeginning,duetointer-
fering currents (from capacitive charges, dielectric absorption) which eventually
canceleachotherout.Therefore,inordertogetanaccuratemeasurementofthe
leakagecurrent,itisnecessarytocarryoutthemeasurementoveralongperiod
oftime,toreducetheinuenceofanyinterferingcurrentsatthebeginningofthe
measurement.
NextwecalculatethePIortheDAR ratios:
PI=R10 min / R1 min(2valuestorecordduringa10minmeasurement)
DAR=R1 min / R30 s(2valuestorecordduringa1minmeasurement)
Thequalityofinsulationdependsontheresultsfound.
DAR PI State of the insulation
< 1.25 < 1 Inadequate, even dangerous
< 2
< 1.6 > 2 to < 4 Good
> 1.6 > 4 Excellent
Using the R-DAR-PI Function:
Duringoraftermeasurement,the R-DAR-PI buttonallows youto scrollthrough
thefollowingvalues:
• DAR(formeasurementslasting>1min).
• PI(formeasurementslasting>10min).
• CapacitanceinµF(onlyafterthemeasurementhasstoppedandthecir-
cuithasbeendischarged).
• InsulationresistanceinMΩ,GΩor TΩ.
Automatic Measurement of DAR or PI:
• Ifthebuttonispressedwheninvoltagemeasurementmodebeforepro-
ceedingwithameasurement,thefollowingisdisplayed:
MΩ
DAR
PI
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
15
Dependingonthechoices(DARorPI),themeasurementgoesasfollows:
a) DAR: PressSTART-theDARsymbolashesandthedisplayindicates“----”
aslongasthecalculationofthecoefcientisnotpossible(t<1mn).
e.g.:
DAR
After1min,themeasurementstopsandthemaindisplayautomaticallyshowsthe
DARvalue.Duringoraftermeasurement,theR-DAR-PIbuttoncanbeusedtosee
theinsulationmeasurementcarriedout,butitdoesnotgivethePIvaluesincethe
measurementhasnotlastedlongenough.
b) PI: Press START - the PI symbol ashes and the display indicates “- - - -”
aslongasthecalculationofthecoefcientisnotpossible(t<10mn).
e.g.:
PI
After 10 minutes, the measurement stops and the main display automatically
showsthePIvalue.Duringandaftermeasurement,theR-DAR-PIbuttonmakes
itpossibletodisplaytheDAR(after1min),thePI(after10min)andtheinsulation
measurement.
Note: IfduringDARorPImeasurement(automaticornot),ahighexternalinter-
ferencevoltage becomes present, or the insulation resistance goesbeyondthe
device’smeasurementrange,thentheDARorPImeasurementsareinterrupted
andthescreenindicates:
DAR or
MΩPI MΩ
Note: The10min/1minmeasurementdurationsforPIcalculationcanbemodi-
edintheSET-UPmenu(see§2.5.7),allowingchangestostandardsortosuita
particularapplication.
Second Function - R(t)
Usedtoaccesstheintervalinsulationresistancevaluesasafunctionoftimeafter
ameasurementin“time-controlledtest”mode(see§2.5.2).
ThetimebetweeneachsavedsampleisprogrammedintheSET-UPmenu.
ThisfunctionisalsoavailableontheModel1050,whichhasneitheraread-write
memory for saving measured data, nor an interface for data retrieval from the
instrumentbyaPC.
Model 1050:
Upto20samplescanberecordedduringmeasurementatthesampleratechosen
in SET-UP(thedefaultvalueis30seconds).Itispossibletosavemorethan20
samplesdependingonavailablememory.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

16
Megohmmeter Models 1050 /1060
Model 1060:
Thenumberofsamplesthatcanbesavedisonlylimitedbythememoryspace
available.
Toenterthedisplaymode,presstheR(t)button:
• Thesmalldisplayindicatesthetime00:30(e.g.thesamplingfrequencyis
every30s).
• ThemaindisplayshowsthecorrespondingRvalue.
UsetheV/TIMEbuttontoalternatebetweenthevoltageandtime(onthesmall
display),associatedwiththeRvalueonthemaindisplay.
The button is used to scroll down through the all the samples saved during
measurement,allowingyoutoreadtheinformationinordertodrawR(t)andV(t)
graphs.
ThismakesitpossibletocarryoutR(t)analysison-site,withoutaprinteroraPC.
Toexitthisfunction,presstheR(t) or R-DAR-PIbuttonagain.
2.5.4 / ALARM Button
First Function -
Turnsthedisplayback-lightON or OFF.
Second Function - ALARM
Activates/deactivates the ALARM function. The “ALARM” symbol is displayed
whenactivated.
Ifactivated,andthehighorlowthresholdvaluesareprogrammed,the“ALARM”
symbolashesandthebuzzer(ifactivated)soundscontinuouslywhenthethresh-
old is crossed.
Itispossibletoprogramadifferentlimitineachfunction.Theselimitswillbestored
inmemoryevenafterthedeviceisturnedoff.
Activating the Alarm Indication before a Test
ToactivatetheAlarmIndicationbeforeperformingatestyoumustgothroughthe
followingsteps:
• ProgramtheAlarmSetPointResistanceValuefortheTestVoltageyou
willbeusing.
• TurntheRotaryswitchtotheSetupposition.
• PresstheYellow2ndbuttonandthentheAlarmbuttontosetAlarms.The
rsttimeyoupress2nd&Alarmyouwillseethealarmsettingfor50volt
tests.
• Press2nd&Alarmbuttonsrepeatedlyuntilyouseethetestyouwishtoset
alarmsforinthetoplineofthedisplay.Thesequencingwillbe50,100,
250,500,1000,ResandCont.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

Megohmmeter Models 1050 /1060
17
• UsethecursorkeystoadjusttheBlinkingvalue.FirstselectKΩ, MΩ, GΩ
or TΩusingtheUpArrowKey().
• Next,presstheRightArrowKey()tomovetothenextselectionwhich
isthe greaterthan(>) /lessthan (<)choice.Use theto makeyour
selection.
• PresstheagaintomovetotheResistanceValueSetPointSelection.
You can change this value digit by digit using the / combination
workingfromthehighesttolowestdigit.
• Whennishedprogrammingthealarmsettings,placetherotaryswitchto
thedesiredtestfunction.
• Beforeinitiatingatest,presstheYellow2ndbuttonandtheAlarmbutton
toactivatetheAlarmNotication.Theword“ALARM”willappearinthe
lowerleftsideofthedisplay.
• Next,presstheroundyellowStart/Stopbuttontobeginthetest.TheAlarm
symbolwillashandthebuzzerwillsoundcontinuouslyiftheAlarmset
pointistripped.
NOTE: If the rotary switch is turned off prior to the next test, the alarm notification
will be deactivated. You will need to press the Yellow 2nd button and the Alarm
button to activate the Alarm Notification when you turn the unit back on before
starting the next test, if alarm notification is desired.
2.5.5 / SMOOTH Button
First Function -
Selectsthedesiredparametertobemodied-theselectedparameterashes.
Itismodiedusingthe button(see§2.5.6).
Second Function - SMOOTH
Activatesadigitallterforinsulationmeasurement.Itonlyhasaneffecton the
displayedvalues(whicharesmoothed)andnottheactualmeasurements.
Forexample,thisfunctionisusefulwhenthedisplayedinsulationvaluesarehighly
unstable,broughtaboutduetoacapacitivecomponentinthetestedelement.
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412

18
Megohmmeter Models 1050 /1060
2.5.6 Button
Changestheashingparametersdisplayed,orviewsR(t)values(see§2.5.3).
As a general rule, two gures (day, month, hour, min., sec., and OBJ:TEST)
ash.
Theand functionshavea“follow-on”mode.(e.g.assoonasthehighorlow
programlimitisreached,theparametertobemodiedswitchesautomaticallyto
thefollowingloworhighlimit.)
First Function
• Pressbuttonbrieytoincreasethedisplayednumberslowly
• Pressbuttonforalongertimetoincreaseatafasterrate
Second Function
• Pressbuttonbrieytodecreasethedisplayednumberslowly
• Pressbuttonforalongertimetodecreaseatafasterrate
Unlikeallthesecondfunctionsofotherbuttons,itisnotnecessarytopressthe
buttoneachtimetogettothefunction.The symbolremainsdisplayed
andisvalidforthefunction(only),aslongastheuserdoesnotdeactivateitby
pressing oranotherbutton.
2.5.7 SET-UP Button (Configuring the Instrument)
AfterselectingSET-UPwiththerotaryselectorswitch:
• Allthedisplaysegmentsarelitupfor1second
• Thesoftwareversionnumberisdisplayed
• Theinstrumentserialnumberisdisplayed
• “PUSH”comesuponthesmalldisplayand“btn”onthemaindisplay,
invitingtheoperatortopressabutton
The SET-UP function is used to directly access the parameters to be pro-
grammed, by pressing the corresponding button:
• After having pressed a button, the corresponding gures or symbols
appearonthescreen
• Theguresorthesymbolsthatcanbemodiedashonthescreen
• Usethe
and buttons
• Alltheparametersareimmediatelyandpermanentlysaved
www.ShopAEMC.com
Shop for AEMC products online at: 1.877.766.5412
This manual suits for next models
1