AEMC 6611 User manual

WITH AEMC
®
INSTRUMENTS
ELECTRICAL TEST TOOLS
HERRAMIENTAS PARA PRUEBAS ELÉCTRICAS
Phase & Motor Rotation Meter
Model 6611
Medidor de Secuencia de Fases y
Rotación de Motor
Modelo 6611
ESPAÑOL
Manual de Usuario
ENGLISH
User Manual

Copyright©Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments. All rights
reserved.
No part of this documentation may be reproduced in any form or by
any means (including electronic storage and retrieval or translation into
any other language) without prior agreement and written consent from
Chauvin Arnoux®, Inc., as governed by United States and International
copyright laws.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive • Dover, NH 03820 USA
Tel: (603) 749-6434 or (800) 343-1391 • Fax: (603) 742-2346
This documentation is provided as is, without warranty of any kind,
express, implied, or otherwise. Chauvin Arnoux®, Inc. has made every
reasonable eort to ensure that this documentation is accurate; but
does not warrant the accuracy or completeness of the text, graphics,
or other information contained in this documentation. Chauvin Arnoux®,
Inc. shall not be liable for any damages, special, indirect, incidental,
or inconsequential; including (but not limited to) physical, emotional or
monetary damages due to lost revenues or lost prots that may result
from the use of this documentation, whether or not the user of the
documentation has been advised of the possibility of such damages.

Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
certies that this instrument has been calibrated
using standards and instruments traceable to
international standards.
We guarantee that at the time of shipping your
instrument has met the instrument’s published
specications.
The recommended calibration interval for this
instrument is 12 months and begins on the date
of receipt by the customer. For recalibration,
please use our calibration services.
Refer to our repair and calibration section at
www.aemc.com/calibration.
Serial #: _________________________
Catalog #: 2121.90
Model #: 6611
Please ll in the appropriate date as indicated:
Date Received: _____________________
Date Verication Due: _________________
Chauvin Arnoux®, Inc.
d.b.a AEMC®Instruments
www.aemc.com

TABLE OF CONTENTS
1. INTRODUCTION............................................................6
1.1 International Electrical Symbols................................. 6
1.2 Denition of Measurement Categories (CAT) ............ 7
1.3 Precautions for Use ................................................7
1.4 Receiving Your Shipment........................................... 8
1.5 Ordering Information.................................................. 8
1.5.1 Accessories and Replacement Parts ..................8
2. PRODUCT FEATURES..................................................9
2.1 Description................................................................. 9
2.2 Control Features ...................................................... 10
3. OPERATION ................................................................ 11
3.1 Determine Rotary Field Direction..............................11
3.2 Instrument Front........................................................11
3.2.1 Faceplate .......................................................... 11
3.3 Instrument Back....................................................... 12
3.3.1 Instruction Label/Safety Information .................12
3.4 Non-Contact Rotary Field Indication........................ 12
3.5 Determine The Motor Connection............................ 14
3.6 Magnetic Field Detection ......................................... 14
4. SPECIFICATIONS........................................................15
4.1 Determine Rotary Field Direction............................. 15
4.2 Non-Contact Rotary Field Indication........................ 15
4.3 Determine The Motor Connection............................ 15
4.4 Electrical .................................................................. 15

4.5 Mechanical............................................................... 15
4.6 Environmental.......................................................... 15
4.7 Safety....................................................................... 16
5. MAINTENANCE...........................................................17
5.1 Battery Replacement ............................................... 17
5.2 Cleaning................................................................... 17
5.3 Repair and Calibration ............................................. 18
5.4 Technical Assistance................................................ 18
5.5 Limited Warranty...................................................... 19
5.5.1 Warranty Repairs ..............................................20

6
Phase & Motor Rotation Meter Model 6611
1. INTRODUCTION
Thank you for purchasing an AEMC®Instruments Phase & Motor
Rotation Meter Model 6611.
For the best results from your instrument and for your safety,
you must read the enclosed operating instructions carefully and
comply with the precautions for use. Only qualied and trained
operators should use this product.
1.1 International Electrical Symbols
Signies that the instrument is protected by double or
reinforced insulation.
CAUTION - Risk of Danger! Indicates a WARNING.
Whenever this symbol is present, the operator must
refer to the user manual before operation.
Indicates a risk of electric shock. The voltage at the
parts marked with this symbol may be dangerous.
Indicates Important information to acknowledge
Battery
Ground/Earth
AC or DC
This product complies with the Low Voltage &
Electromagnetic Compatibility European directives.
In the European Union, this product is subject to a
separate collection system for recycling electrical and
electronic components in accordance with directive
WEEE 2012/19/EU.

Phase & Motor Rotation Meter Model 6611
7
1.2 Denition of Measurement Categories (CAT)
CAT IV: Corresponds to measurements performed at the primary
electrical supply (< 1000 V).
Example: primary overcurrent protection devices, ripple
control units, and meters.
CAT III: Corresponds to measurements performed in the building
installation at the distribution level.
Example: hardwired equipment in xed installation and
circuit breakers.
CAT II: Corresponds to measurements performed on circuits
directly connected to the electrical distribution system.
Example: measurements on household appliances and
portable tools.
1.3 Precautions for Use
This instrument complies with safety standard IEC 61010-1.
For your own safety, and to prevent any damage to your
instrument, you must follow the instructions given in this
manual.
This instrument can be used on CAT IV electrical circuits not
exceeding 600 V with respect to earth. It must be used indoors,
in an environment not exceeding pollution level 2, at an
altitude of not more than 6562 ft (2000 m). The instrument can
therefore be used in complete safety on (40 to 850) V three-
phase networks in an industrial environment.
For safety reasons, you must use only measurement leads
having a voltage rating and category at least equal to those of
the instrument and compliant with standard IEC 61010-031.
Do not use if the housing is damaged or not correctly closed.
Do not place your ngers near unused terminals.
If the instrument is used other than as specied in this manual,
the protection provided by the instrument may be impaired.

8
Phase & Motor Rotation Meter Model 6611
Do not use this instrument if it seems to be damaged.
Check the integrity of the insulation of the leads and of the
housing. Replace damaged leads.
Be prudent when working in the presence of voltages
exceeding 60 VDC or 30 VRMS and 42 Vpp; such voltages can
cause a risk of electrocution. The use of individual protections
is recommended in some cases.
Always keep your hands behind the physical guards of the
probe tips or alligator clips.
Always disconnect all leads from the measurement and from
the instrument before opening the housing.
1.4 Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are
consistent with the packing list. Notify your distributor of any
missing items. If the equipment appears to be damaged, le a
claim immediately with the carrier and notify your distributor at
once, giving a detailed description of any damage. Save the
damaged packing container to substantiate your claim.
1.5 Ordering Information
Phase and Motor Rotation Meter Model 6611 ...... Cat. #2121.90
Includes meter, (3) color-coded test leads (red, black, blue),
(3) alligator clips (black), soft carrying case and a user manual.
1.5.1 Accessories and Replacement Parts
Soft carrying case......................................................Cat. #2117.73
Set of (3) color-coded leads with
(3) black alligator clips CAT III 1000 V 10 A.............. Cat. #2121.55

Phase & Motor Rotation Meter Model 6611
9
2. PRODUCT FEATURES
2.1 Description
This three-in-one test tool is a must for any plant maintenance
sta and will identify proper sequencing for three phase power
very quickly and easily.
This is also an ideal tool for measuring the proper rotation
of motors, conveyors, pumps and other electrical devices
interconnected on the power line system before installation.
NOTE: The Model 6611 does not require fusing because
the inputs are protected by a high impedance circuit
which limits the current to a safe value.
This meter provides the following functions:
determination of the direction of phase rotation
presence or absence of phase
determination of the direction of rotation of a motor
with or without connection
determination of the activation of a solenoid valve
without connection

10
Phase & Motor Rotation Meter Model 6611
2.2 Control Features
1Test Lead Input Terminals
2L1 Phase Indicator
3L2 Phase Indicator
4L3 Phase Indicator
5Clockwise Rotation Indicator
6Counterclockwise Rotation Indicator
7ON/OFF Indicator
8ON/OFF Button
9Back Label
10 Battery Compartment & Cover Screw
2
3
5
6
1
4
7
8
10
9

Phase & Motor Rotation Meter Model 6611
11
3. OPERATION
3.1 Determine Rotary Field Direction
On a three-phase electrical network:
1. Connect one end of the test leads to the Phase & Motor
Rotation Meter, make sure the L1, L2 and L3 test leads are
connected to the corresponding input jacks.
2. Connect the alligator clips to the other end of the test leads.
3. Connect the alligator clips to the three mains phases, PRESS
the ON/OFF Button, the green ON indicator shows that the
instrument is ready for testing.
4. Either the Clockwise or Counterclockwise Rotary indicator
illuminates showing the Type of rotary eld direction present.
5. The rotary indicator lights even if the neutral conductor, N, is
connected instead of the Test lead input jacks.
6. Refer to Figure 2 shown in § 3.3.1 (also shown on the back of
the Phase & Motor Rotation Meter) for more information.
3.2 Instrument Front
3.2.1 Faceplate
Figure 1
CAT III 1000 V
CAT IV 600 V
PHASE & MOTOR ROTATION
MODEL 6611

12
Phase & Motor Rotation Meter Model 6611
3.3 Instrument Back
3.3.1 Instruction Label/Safety Information
Chauvin Arnoux
®
, Inc.
d.b.a AEMC
®
Instruments
15 Faraday Dr. Dover NH 03820 - USA
www.aemc.com
Un=400/230 VAC Ume=1...400 VAC
fn=2...400 Hz
I
L1=I
L2=I
L3 3.5 mA/400 V
1x9 V
Figure 3
WARNING: The wrong direction of rotation may
be displayed if a lead is connected in error to the
neutral conductor. Refer to the instrument’s back
label (see Figure 2 above) for a summary of the
various display possibilities.
3.4 Non-Contact Rotary Field Indication
1. Disconnect all test leads from the Phase & Motor Rotation
Meter.
2. Position the Indicator on the motor so that it is parallel to the
length of the motor shaft, the Indicator should be one inch or
close to the motor.
3. PRESS the ON/OFF Button, the green ON indicator shows
that the instrument is ready for testing.
4. Either the Clockwise or Counterclockwise Rotary indicator
illuminates showing the Type of rotary eld direction present.
Figure 2

Phase & Motor Rotation Meter Model 6611
13
NOTE: The indicator will not operate with engines
controlled by frequency converters. The bottom of
the Phase & Motor Rotation Meter should be oriented
towards the drive shaft. See the Orientation Symbol on
the Phase & Motor Rotation Meter.
Figure 4
See the table below for the minimum motor diameter and number
of pole pair to obtain a reliable test result.
Number
of Pole
Pair
Rotary Number of Rotary Field
(1/mm) at Frequency (Hz)
Angle
Between
Poles
Minimum Ø
of Motorcase
16 2/3 50 60 ºcm
11000 3000 3600 60 5.3
2500 1500 1800 30 10.7
3333 1000 1200 20 16.0
4250 750 900 15 21.4
5200 600 720 12 26.7
6167 500 600 10 3 2 .1
8125 375 450 7.5 42.8
10 100 300 360 653.5
12 83 250 300 564.2
16 62 188 225 3.75 85.6

14
Phase & Motor Rotation Meter Model 6611
3.5 Determine The Motor Connection
1. Connect one end of the test leads to the Phase & Motor
Rotation Meter, make sure the L1, L2 and L3 test leads are
connected to the corresponding jack.
2. Connect the alligator clips to the other end of the test leads.
3. Connect the alligator clips to the motor connections, L1 to U,
L2 to V, L3 to W.
4. PRESS the ON/OFF Button, the green ON indicator shows
that the instrument is ready for testing.
5. Turn the motor shaft half a revolution towards the right.
NOTE: The bottom of the Phase & Motor Rotation
Meter should be oriented towards the drive shaft. See
the Orientation Symbol on the Phase & Motor Rotation
Meter.
NOTE: Either the Clockwise or Counterclockwise Rotary
indicator illuminates showing the type of rotary eld
direction present.
3.6 Magnetic Field Detection
To detect a magnetic eld, place the Phase & Motor Rotation
Meter to a solenoid valve.
A magnetic eld is present if either the Clockwise or the
Counterclockwise.

Phase & Motor Rotation Meter Model 6611
15
4. SPECIFICATIONS
4.1 Determine Rotary Field Direction
Nominal Voltage Rotary Direction (1 to 400) VAC
Nominal Voltage Phase Indirection (120 to 400) VAC
Frequency Range (fn) (2 to 400) Hz
Test Current (In per phase) Less than 3.5 mA
4.2 Non-Contact Rotary Field Indication
Frequency Range (fn) (2 to 400) Hz
4.3 Determine The Motor Connection
Nominal Test Voltage (U me) (1 to 400) VAC
Nominal Test Current (In per phase) Less than 3.5 mA
Frequency Range (fn) (2 to 400) Hz
4.4 Electrical
Battery 9 V Alkaline, IEC 6LR61
Current Consumption Max 20 mA
Battery Life Minimum 1 year for average use
4.5 Mechanical
Dimensions (5.3 x 2.95 x 1.22) in (135 x 75 x 31) mm
Weight 4.83 oz (137 g)
4.6 Environmental
Operating Temperature (32 to 104) °F (0 to 40) °C
Storage Temperature (-4 to 122) °F (-20 to 50) °C; RH < 80 %
Operating Humidity (15 to 80) % RH
Operating Altitude 6562 ft (2000 m)
Pollution Degree 2

16
Phase & Motor Rotation Meter Model 6611
4.7 Safety
Safety Rating
CAT IV 600 V, 1000 V CAT III
IEC 61010-1, IEC 61557-7,
Tightness : IP40 (as per IEC 60529 Ed.92)
Double Insulation Yes
CE Mark Yes

Phase & Motor Rotation Meter Model 6611
17
5. MAINTENANCE
5.1 Battery Replacement
WARNING: Always disconnect all leads before
replacing a battery or fuse.
The Phase & Motor Rotation Meter uses a 9 V battery (supplied).
To replace the battery, follow these steps.
1. Place the instrument face down on a nonabrasive surface
and loosen the battery compartment cover screw with a
screwdriver.
2. Lift the battery access lid away from the instrument.
3. Remove battery and replace with new 9 V battery. Observe
the battery polarity shown in the battery compartment.
4. Secure the battery access lid back in position with the screw.
NOTE: Do not treat spent alkaline batteries as ordinary
household waste. Take them to the appropriate
collection facility for recycling.
5.2 Cleaning
WARNING: To avoid electrical shock or damage to
the instrument, do not allow water to get inside of
the case.
The instrument should be cleaned periodically to keep the LCD
clear and prevent the buildup of dirt and grease around the
instrument’s buttons.
Wipe the case with a soft cloth lightly moistened with mild,
soapy water.
Dry completely with a soft, dry cloth before using again.
Do not allow water or other foreign substances into the case.
Never use alcohol, abrasives, solvents or hydrocarbons.

18
Phase & Motor Rotation Meter Model 6611
5.3 Repair and Calibration
To ensure that your instrument meets factory specications,
we recommend that the instrument be sent back to our factory
Service Center at one-year intervals for recalibration or as
required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service
requesting a CSA#, you will be provided a CSA Form and other
required paperwork along with the next steps to complete the
request. Then return the instrument along with the signed CSA
Form. This will ensure that when your instrument arrives, it will be
tracked and processed promptly. Please write the CSA# on the
outside of the shipping container.
Ship To: Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone: (800) 945-2362 (Ext. 360) / (603) 749-6434 (Ext. 360)
Fax: (603) 742-2346
(Or contact your authorized distributor.)
Contact us for the costs for repair and standard calibration.
NOTE: You must obtain a CSA# before returning any
instrument.
5.4 Technical Assistance
If you are experiencing any technical problems or require any
assistance with the proper operation or application of your
instrument, please call, e-mail or fax our technical support team:
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
Phone: (800) 343-1391 (Ext. 351)
Fax: (603) 742-2346

Phase & Motor Rotation Meter Model 6611
19
5.5 Limited Warranty
The instrument is warrantied to the owner for a period of two years
from the date of original purchase against defects in manufacture.
This limited warranty is given by AEMC®Instruments, not by the
distributor from whom it was purchased. This warranty is void if
the unit has been tampered with, abused, or if the defect is related
to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available
on our website at www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for
your records.
What AEMC®Instruments will do:
If a malfunction occurs within the warranty period, you may
return the instrument to us for repair, provided we have your
warranty registration information on le or a proof of purchase.
AEMC®Instruments will repair or replace the faulty material at our
discretion.
REGISTER ONLINE AT: www.aemc.com/warranty.html

20
Phase & Motor Rotation Meter Model 6611
5.5.1 Warranty Repairs
What you must do to return an Instrument for Warranty
Repair:
Service Authorization Number (CSA#) from our Service
Department. You will be provided a CSA Form and other required
paperwork along with the next steps to complete the request.
Then return the instrument along with the signed CSA Form.
Please write the CSA# on the outside of the shipping container.
Return the instrument, postage or shipment pre-paid to:
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive, Dover, NH 03820 USA
Phone: (800) 945-2362 (Ext. 360)
(603) 749-6434 (Ext. 360)
Fax: (603) 742-2346
Caution: To protect yourself against in-transit loss, we
recommend that you insure your returned material.
NOTE: You must obtain a CSA# before returning any
instrument.