OPERATION
Making Measurements with the AC Current Probe Model MN185
•Connect the terminals to the 200mA range of your DMM or instrument. The MN185 has a ratio of 1000:1.
This means that for 100A AC in the conductor around which the probe is clamped, 100mA AC will come out
of the probe leads to your DMM or instrument. The output is 1mA AC per Amp. Select the range on your
DMM or instrument which best corresponds to the measured current. If the magnitude is unknown, start with
the highest range (200mA AC) then work down until the appropriate range and resolution is reached. Clamp
the probe around the conductor. Take the reading on the meter and multiply it by 1000 to obtain the
measured current (e.g., 59mA reading: 59 x 1000 = 59,000mA or 59A).
•For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
•When using a current probe with a meter, it is important to select the range that provides the best resolution.
Failure to do this may result in measurement errors.
•Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air gaps
between the jaws, increasing the phase shift between primary and secondary. It is very critical for power
measurement.
MAINTENANCE
Warning:
•For maintenance use only original factory replacement parts.
•To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
•To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into the
probe.
Cleaning: To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at
all times. Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper
(fine 600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure
that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the
outside of the shipping container.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive •Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • repair@aemc.com
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or application
of this instrument, please contact our technical hotline:
(800) 343-1391 • (508) 698-2115 • techsupport@aemc.com
LIMITED WARRANTY
The current probe is warranted to the owner for a period of one year from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC®Instruments, not by the distributor
from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the
defect is related to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at:
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
AC Current Probe
Model MN185
User Manual
DESCRIPTION
The MN185 (Catalog #100.185) is a high accuracy current probe for tight areas such as crowded wiring.
Extends DMM AC measurements to 120A AC.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting to
use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may pose a
shock hazard.
• Read the safety specifications section prior to using the current probe. Never exceed the maximum
voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use. Replace any
defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 250V. Use extreme caution
when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use only
factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using
the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your
distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of any damage.
INSTRUMENT COMPATIBILITY
The Model MN185 is compatible with any ammeter, multimeter, or other current measuring instruments with
an input impedance of less than 5Ω. To achieve the stated accuracy, use the MN185 with an ammeter
having an accuracy of 0.75% or better.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
ELECTRICAL SPECIFICATIONS
Current Range:
50mA to 120A AC, continuous
Output Signal:
1mA AC / A AC (120mA @ 120A)
Accuracy and Phase Shift*:
Accuracy:
50mA to 100A: 1% ± 0.01A
(with non-inductive load)
100A to 120A: 1.5% ± 0.01A
Phase Shift:
≤3.5°from 1 to 100A, 50 to 60Hz
(*Reference conditions: 23°C ± 3°K, 20 to
85% RH, external magnetic field < 40A/m,
no DC component, no external current
carrying conductor, test sample centered.)
Load impedance 1Ω.
Overload:
170A continuously
Frequency Range:
30Hz to 10kHz
Load Impedance:
5Ωmax
Working Voltage:
250V AC
Common Mode Voltage:
30V AC
MECHANICAL SPECIFICATIONS
Operating Temperature:
-13° to 122°F (-25° to 50°C)
Storage Temperature:
-40° to 176°F (-40° to 80°C)
Temperature Influence:
< 0.2% per 10°K
Maximum Conductor Size:
0.47” Ø max. (12mm)
Dimensions:
1.26 x 4.53 x 0.87"
(32 x 115 x 22mm)
Weight: 5.6 oz (160g)
Colors:
Dark gray handles with red cover
Polycarbonate Material:
Handle: 10% Fiberglass charged
polycarbonate UL 94 V0
Output:
Two standard safety banana jacks (4mm)
SAFETY SPECIFICATIONS
Electrical:
3kV 50/60 Hz dielectric
30V max common mode between output
and ground
ORDERING INFORMATION
AC Current Probe MN185...Cat #100.185
Accessories:
Lead - Set of 2, 3 ft color-coded
non-safety leads...................Cat #1007.41
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
ELECTRICAL SPECIFICATIONS
Current Range:
50mA to 120A AC, continuous
Output Signal:
1mA AC / A AC (120mA @ 120A)
Accuracy and Phase Shift*:
Accuracy:
50mA to 100A: 1% ± 0.01A
(with non-inductive load)
100A to 120A: 1.5% ± 0.01A
Phase Shift:
≤3.5°from 1 to 100A, 50 to 60Hz
(*Reference conditions: 23°C ± 3°K, 20 to
85% RH, external magnetic field < 40A/m,
no DC component, no external current
carrying conductor, test sample centered.)
Load impedance 1Ω.
Overload:
170A continuously
Frequency Range:
30Hz to 10kHz
Load Impedance:
5Ωmax
Working Voltage:
250V AC
Common Mode Voltage:
30V AC
MECHANICAL SPECIFICATIONS
Operating Temperature:
-13° to 122°F (-25° to 50°C)
Storage Temperature:
-40° to 176°F (-40° to 80°C)
Temperature Influence:
< 0.2% per 10°K
Maximum Conductor Size:
0.47” Ø max. (12mm)
Dimensions:
1.26 x 4.53 x 0.87"
(32 x 115 x 22mm)
Weight: 5.6 oz (160g)
Colors:
Dark gray handles with red cover
Polycarbonate Material:
Handle: 10% Fiberglass charged
polycarbonate UL 94 V0
Output:
Two standard safety banana jacks (4mm)
SAFETY SPECIFICATIONS
Electrical:
3kV 50/60 Hz dielectric
30V max common mode between output
and ground
ORDERING INFORMATION
AC Current Probe MN185...Cat #100.185
Accessories:
Lead - Set of 2, color-coded (red/black)
5 ft Safety Leads
(1000V CAT IV) ...................Cat #2152.24
Adapter - 4 mm non-insulated for
Safety Leads....................... Cat #1017.45
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
99-MAN 100008.v13 03/19
OPERATION
Making Measurements with the AC Current Probe Model MN185
•Connect the terminals to the 200mA range of your DMM or instrument. The MN185 has a ratio of 1000:1.
This means that for 100A AC in the conductor around which the probe is clamped, 100mA AC will come out
of the probe leads to your DMM or instrument. The output is 1mA AC per Amp. Select the range on your
DMM or instrument which best corresponds to the measured current. If the magnitude is unknown, start with
the highest range (200mA AC) then work down until the appropriate range and resolution is reached. Clamp
the probe around the conductor. Take the reading on the meter and multiply it by 1000 to obtain the
measured current (e.g., 59mA reading: 59 x 1000 = 59,000mA or 59A).
•For best accuracy, avoid if possible, the proximity of other conductors which may create noise.
Tips for Making Precise Measurements
•When using a current probe with a meter, it is important to select the range that provides the best resolution.
Failure to do this may result in measurement errors.
•Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air gaps
between the jaws, increasing the phase shift between primary and secondary. It is very critical for power
measurement.
MAINTENANCE
Warning:
•For maintenance use only original factory replacement parts.
•To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
•To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into the
probe.
Cleaning: To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at
all times. Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper
(fine 600) to avoid scratching the jaw, then gently clean with a soft oiled cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure
that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the
outside of the shipping container.
(Or cont
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or application
of this instrument, please contact our technical hotline:
LIMITED WARRANTY
The current probe is warranted to the owner for a period of two yearsfrom the date of original purchase
against defects in manufacture. This limited warranty isgiven by AEMC®Instruments, not by the distributor
from whom itwas purchased. This warranty isvoid if the unit has been tampered with, abused or ifthe
defect isrelated to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at:
Please print the online Warranty Coverage Information for your records.
AC Current Probe
Model MN185
User Manual
DESCRIPTION
The MN185 (Catalog #100.185) is a high accuracy current probe for tight areas such as crowded wiring.
Extends DMM AC measurements to 120A AC.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the
instrument.
• Read the instruction manual completely and follow all the safety information before attempting to
use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may pose a
shock hazard.
• Read the safety specifications section prior to using the current probe. Never exceed the maximum
voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the
sample being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use. Replace any
defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 250V. Use extreme caution
when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use only
factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using
the instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your
distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of any damage.
INSTRUMENT COMPATIBILITY
The Model MN185 is compatible with any ammeter, multimeter, or other current measuring instruments with
an input impedance of less than 5Ω. To achieve the stated accuracy, use the MN185 with an ammeter
having an accuracy of 0.75% or better.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Other AEMC Measuring Instrument manuals

AEMC SR634 User manual

AEMC

AEMC SR634 User manual

AEMC SL261 User manual

AEMC

AEMC SL261 User manual

AEMC JM813 User manual

AEMC

AEMC JM813 User manual

AEMC 565 User manual

AEMC

AEMC 565 User manual

AEMC 514 User manual

AEMC

AEMC 514 User manual

AEMC 4500 User manual

AEMC

AEMC 4500 User manual

AEMC 203 User manual

AEMC

AEMC 203 User manual

AEMC DTR 8510 User manual

AEMC

AEMC DTR 8510 User manual

AEMC 1110 User manual

AEMC

AEMC 1110 User manual

AEMC MN254 User manual

AEMC

AEMC MN254 User manual

AEMC PowerPad III 8336 User manual

AEMC

AEMC PowerPad III 8336 User manual

AEMC 193-24-BK User manual

AEMC

AEMC 193-24-BK User manual

AEMC Simple Logger CL600 User manual

AEMC

AEMC Simple Logger CL600 User manual

AEMC 193-24-BK User manual

AEMC

AEMC 193-24-BK User manual

AEMC 8230 User manual

AEMC

AEMC 8230 User manual

AEMC PowerPad 3945-B User manual

AEMC

AEMC PowerPad 3945-B User manual

AEMC 6611 User manual

AEMC

AEMC 6611 User manual

AEMC K100 User manual

AEMC

AEMC K100 User manual

AEMC MR6292 User manual

AEMC

AEMC MR6292 User manual

AEMC MN379 User manual

AEMC

AEMC MN379 User manual

AEMC 8220 User manual

AEMC

AEMC 8220 User manual

AEMC 511 User manual

AEMC

AEMC 511 User manual

AEMC SR701 User manual

AEMC

AEMC SR701 User manual

AEMC 566 User manual

AEMC

AEMC 566 User manual

Popular Measuring Instrument manuals by other brands

Endress+Hauser Prosonic FDU80 Safety instructions

Endress+Hauser

Endress+Hauser Prosonic FDU80 Safety instructions

Monacor DMT-2505 instruction manual

Monacor

Monacor DMT-2505 instruction manual

SONBEST SM2120B user manual

SONBEST

SONBEST SM2120B user manual

Pasco Scientific OS-8537 Instruction manual and experiment guide

Pasco Scientific

Pasco Scientific OS-8537 Instruction manual and experiment guide

Würth WDM 6-22 manual

Würth

Würth WDM 6-22 manual

Rigol DSA1000A Series quick guide

Rigol

Rigol DSA1000A Series quick guide

Mastech MS2201 manual

Mastech

Mastech MS2201 manual

F&F LE-03d CT400 quick start guide

F&F

F&F LE-03d CT400 quick start guide

Stahl 8404/2 Series operating instructions

Stahl

Stahl 8404/2 Series operating instructions

Vega VEGAFLEX FX-E.80H operating instructions

Vega

Vega VEGAFLEX FX-E.80H operating instructions

Keithley 196 instruction manual

Keithley

Keithley 196 instruction manual

Temperature Guard Input Guard VM500-6 Manual and installation instructions

Temperature Guard

Temperature Guard Input Guard VM500-6 Manual and installation instructions

Sensitech TempTale Ultra Instructions for use

Sensitech

Sensitech TempTale Ultra Instructions for use

Scaletron AccuPro-5000 Operator's manual

Scaletron

Scaletron AccuPro-5000 Operator's manual

Honeywell TWM 9000 Supplementary instructions

Honeywell

Honeywell TWM 9000 Supplementary instructions

NEO TOOLS 75-100 instruction manual

NEO TOOLS

NEO TOOLS 75-100 instruction manual

REMA GGP-01 owner's manual

REMA

REMA GGP-01 owner's manual

TFT 3 quick start guide

TFT

TFT 3 quick start guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.