manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. AEMC
  6. •
  7. Measuring Instrument
  8. •
  9. AEMC JM830A User manual

AEMC JM830A User manual

User Manual
AC Current Probe
Model JM830A
DESCRIPTION
The JM830A (Catalog #2110.83) is designed for use in industrial environments. The “squared” jaws permit
multiple conductor or bus bar positioning. The current output makes it the perfect tool for measurement with
DMMs, recorders, power and harmonic meters. The Model JM830A is compatible with any AC ammeter,
multimeter, or other current measuring instruments with an input impedance lower than 5Ω. To achieve the
stated accuracy, use the JM830A with a DMM having an accuracy of 0.75%.
WARNING
These safety warnings are provided to ensure the safety of personnel and proper operation of the instrument.
• Read the instruction manual completely and follow all the safety information before attempting
to use or service this instrument.
• Use caution on any circuit: Potentially high voltages and currents may be present and may pose a
shock hazard.
• Read the Safety Specifications section prior to using the current probe. Never exceed the maximum
voltage ratings given.
• Safety is the responsibility of the operator.
• ALWAYS connect the current probe to the display device before clamping the probe onto the sample
being tested.
• ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use. Replace any
defective parts immediately.
• NEVER use the current probe on electrical conductors rated above 600V in overvoltage Category III
(CAT III). Use extreme caution when clamping around bare conductors or bus bars.
INTERNATIONAL ELECTRICAL SYMBOLS
This symbol signifies that the current probe is protected by double or reinforced insulation. Use only
factory specified replacement parts when servicing the instrument.
This symbol signifies CAUTION! and requests that the user refer to the user manual before using the
instrument.
This is a type A current sensor. This symbol signifies that application around and removal from
HAZARDOUS LIVE conductors is permitted.
DEFINITION OF MEASUREMENT CATEGORIES
Cat. II: For measurements performed on circuits directly connected to the electrical distribution system.
Examples are measurements on household appliances or portable tools.
Cat. III: For measurements performed in the building installation at the distribution level such as on
hardwired equipment in fixed installation and circuit breakers.
Cat. IV: For measurements performed at the primary electrical supply (<1000V) such as on primary
overcurrent protection devices, ripple control units, or meters.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your
distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the
carrier and notify your distributor at once, giving a detailed description of any damage.
2
ELECTRICAL SPECIFICATIONS
Current Range:
1 to 2400A AC, continuous cycle for
the full temperature range
(3000A if temperature < 35°C or 95°F)
Transformation Ratio:
3000:1
Output Signal:
0.333mA AC/A AC (1A at 3000A)
Accuracy and Phase Shift*:
Primary current 150A 600A 3000A
Accuracy % 1.5% 0.75% 0.5%
Phase shift 1.5° 0.75° 0.5°
(*Reference conditions: 23°C±3°K, 20 to 85%
RH, 48 to 65 Hz, external magnetic field < 40
A/m, no DC component, no external current
carrying conductor, test sample centered.)
Load impedance 5Ω.
3
Overload: 3600A during 5 mn
Open Circuit Voltage:
< 46V by limiting circuit
Accuracy: Per IEC 185-26-27
Class 0.5 from 48 to 1000 Hz
Frequency Range:
30 Hz to 5 kHz; current derating above
1 kHz for continuous use
Ampere Second Product: 90 AA
Load Impedance: ≤ 5Ω
Working Voltage: 600V AC
Common Mode Voltage: 600V AC
Influence of Adjacent Conductor:
0.005A/A AC
Influence of Conductor in Jaw Opening:
1% ± 0.1A of reading
MECHANICAL SPECIFICATIONS
Operating Temperature:
14° to 122°F (-10° to 50°C)
Storage Temperature:
-13° to 176°F (-25° to 80°C)
Influence of Temperature:
< 0.1% per 10°K
Jaw Opening: 3.54" (90mm)
Maximum Conductor Size:
Cable: 2.52" Ø max. (64mm)
Bus bar: 1.97 x 5.31" (50 x 135mm)
2.52 x 3.94" (64 x 100mm)
Envelope Protection:
IP 20 (IEC 529)
Drop Test:
500mm (IEC 68-2-32)
Mechanical Shock:
100g (IEC 68-2-27)
Vibration:
10/55/10Hz, 0.15mm (IEC 68-2-6)
Polycarbonate Material:
Handles: 10% fiberglass charged
polycarbonate UL 94 V0
Jaws: ABS UL 94 V2
Dimensions:
4.72 x 12.40 x 1.89" (120 x 315 x 48 mm)
Weight:
2.65 lbs. (1200 g)
Colors:
Dark gray handles with red jaws
Output:
5 ft. (1.5m) lead with safety 4mm banana plug
SAFETY SPECIFICATIONS
Electrical Conformity to International Standards:
This instrument is compliant with IEC 61010-2-032, 300V
category IV and 600V category III.
Double or reinforced insulation .
Type of current sensor per IEC 61010-2-032:
type A .
Electromagnetic Compatibility:
The device is in conformity with standard IEC
61326-1.
ORDERING INFORMATION
Current Probe JM830A ............................. Cat #2110.83
Includes a user manual and a product warranty and
registration card
Accessories:
Banana plug adapter
(to non-recessed plug)................................ Cat #1017.45
4
OPERATION
Please make sure that you have already read and fully understand the WARNING section on page 1.
Making Measurements with the AC Current Probe Model JM830A
• Connect the black and red terminals to the Ampere AC range of your DMM or current measuring
instrument. Select the appropriate current range (2A AC). Clamp the probe around the conductor to be
tested. If the reading is less than 200mA, select the lower range until you obtain the best resolution.
Read the value display on the DMM and multiply it by the probe ratio (3000/1 = 3000). (If reading =
0.592A AC, the current flowing through the probe is 0.592A x 3000 = 1776A).
• For best accuracy: carefully center the conductor inside the probe jaw, avoid if possible, the proximity
of other conductors which may create noise.
Tips for Making Precise Measurements
• When using a current probe with a meter, it is important to select the range that provides the best
resolution. Failure to do this may result in measurement errors.
• Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for
power measurement.
MAINTENANCE
Warning:
• For maintenance use only original replacement parts.
• To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so.
• To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents
into the probe.
Cleaning:
To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times.
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 600)
to avoid scratching the jaw, then gently clean with a soft, oil cloth.
REPAIR AND CALIBRATION
You must contact our Service Center for a Customer Service Authorization number (CSA#). This will ensure
that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the
outside of the shipping container.
Chauvin Arnoux®, Inc. d.b.a. AEMC®Instruments
15 Faraday Drive • Dover, NH 03820 USA
(800) 945-2362 (Ext. 360) • (603) 749-6434 (Ext. 360) • [email protected]
(Or contact your authorized distributor)
NOTE: All customers must obtain a CSA# before returning any instrument.
TECHNICAL AND SALES ASSISTANCE
If you are experiencing any technical problems, or require any assistance with the proper use or application
of this instrument, please contact our technical hotline:
(800) 945-2362 (Ext. 351) • (603) 749-6434 (Ext. 351) • [email protected]
LIMITED WARRANTY
The current probe is warranted to the owner for a period of two years from the date of original purchase
against defects in manufacture. This limited warranty is given by AEMC®Instruments, not by the distributor
from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the defect
is related to service not performed by AEMC®Instruments.
Full warranty coverage and product registration is available on our website at: www.aemc.com.
Please print the online Warranty Coverage Information for your records.
99-MAN 100113 v9 08/21

Other AEMC Measuring Instrument manuals

AEMC MN05 User manual

AEMC

AEMC MN05 User manual

AEMC MN114 User manual

AEMC

AEMC MN114 User manual

AEMC PowerPad III 8336 User manual

AEMC

AEMC PowerPad III 8336 User manual

AEMC OX 5022-CK Setup guide

AEMC

AEMC OX 5022-CK Setup guide

AEMC 5070 User manual

AEMC

AEMC 5070 User manual

AEMC 1000N Use and care manual

AEMC

AEMC 1000N Use and care manual

AEMC CA7027 User manual

AEMC

AEMC CA7027 User manual

AEMC 721 User manual

AEMC

AEMC 721 User manual

AEMC powerpad 3945 User manual

AEMC

AEMC powerpad 3945 User manual

AEMC 193-24-BK User manual

AEMC

AEMC 193-24-BK User manual

AEMC 1015 User manual

AEMC

AEMC 1015 User manual

AEMC 6292 Use and care manual

AEMC

AEMC 6292 Use and care manual

AEMC CA7027 User manual

AEMC

AEMC CA7027 User manual

AEMC 1227 User manual

AEMC

AEMC 1227 User manual

AEMC 8500 User manual

AEMC

AEMC 8500 User manual

AEMC MD305 User manual

AEMC

AEMC MD305 User manual

AEMC 5050 User manual

AEMC

AEMC 5050 User manual

AEMC MN379 User manual

AEMC

AEMC MN379 User manual

AEMC 205 User manual

AEMC

AEMC 205 User manual

AEMC CA846 User manual

AEMC

AEMC CA846 User manual

AEMC FlexProbe 24-3001 User manual

AEMC

AEMC FlexProbe 24-3001 User manual

AEMC 5070 User manual

AEMC

AEMC 5070 User manual

AEMC powerpad 3945 User manual

AEMC

AEMC powerpad 3945 User manual

AEMC SL261 User manual

AEMC

AEMC SL261 User manual

Popular Measuring Instrument manuals by other brands

SellEton Scales OP-902 user manual

SellEton Scales

SellEton Scales OP-902 user manual

Omron H7CX - datasheet

Omron

Omron H7CX - datasheet

ICT International HFD8 quick start guide

ICT International

ICT International HFD8 quick start guide

FT Technologies FT702LT manual

FT Technologies

FT Technologies FT702LT manual

Bender ISOMETER iso175 quick start guide

Bender

Bender ISOMETER iso175 quick start guide

Winters WinSMART TY52 Series Installation and maintenance manual

Winters

Winters WinSMART TY52 Series Installation and maintenance manual

Lufft XC200 manual

Lufft

Lufft XC200 manual

CARLO GAVAZZI WM30-96 - PROGRAMMING instruction manual

CARLO GAVAZZI

CARLO GAVAZZI WM30-96 - PROGRAMMING instruction manual

EKO Instruments MS-01 instruction manual

EKO Instruments

EKO Instruments MS-01 instruction manual

LINSHANG LS171 user manual

LINSHANG

LINSHANG LS171 user manual

Mitsubishi Electric SS Series manual

Mitsubishi Electric

Mitsubishi Electric SS Series manual

Dairymaster MOOMONITOR + installation manual

Dairymaster

Dairymaster MOOMONITOR + installation manual

Lancol CTE-4000 instruction manual

Lancol

Lancol CTE-4000 instruction manual

Dakota Digital MLX-9X04 Series manual

Dakota Digital

Dakota Digital MLX-9X04 Series manual

Agilent Technologies X Series user guide

Agilent Technologies

Agilent Technologies X Series user guide

PT PT200P Operator's manual

PT

PT PT200P Operator's manual

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Amberg Technologies AG Tunnel 2.0 - Tunnelscan user manual

Amberg Technologies AG

Amberg Technologies AG Tunnel 2.0 - Tunnelscan user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.