Quick Start Guide 5
The maintenance screen has a number of useful features as described
below.
Field Description
Prime Wicks Whenever wicks are replaced or the unit is fully
shut off for prolonged periods of time (> 1 day),
wicks need to be primed. Click “Prime Wicks”
and the instrument will go through a priming cycle
that injects water to all three wicks. Make sure the
fill and drain bottles are connected to the unit. Unit
will go through the prime cycle on its own.
Drain Wicks The instrument needs to be drained before
transporting it to a different location. Make sure
the drain bottle is connected and the fill bottle is
disconnected. Note that if the fill bottle is
connected, you will never be able to drain the
wicks. Unit will go through the drain cycle on its
own.
Enable Flow Turns the pump on or off. See information below
to purge the system.
New Wicks Reset
This is for TSI use only.
Bottle Counter
Reset If you refill the bottle with ultrapure, de-ionized, or
distilled water, click “Bottle Counter Reset.” This
allows the instrument to keep track of the amount
of water used and notify you if the fill bottle is
getting empty.
IMPORTANT
When shipping make sure you drain the water from the unit by clicking on
“DrainWicks”. First, you must disconnect the water fill bottle from the unit
and keep the drain bottle connected to the device.