8. Connection o the Transducers
8. CO ECTIO OF THE TRA SDUCERS
Any ALMEMO® sensors can be connected to the ALMEMO® input sockets M0
to M8 o the measuring instrument (2). For connecting existing sensors it is
only necessary to connect a corresponding ALMEMO® connector.
8.1 Transducers
A detailed description o the comprehensive ALMEMO® sensor range (see
manual section 3) and the connection o existing sensors (see manual section
4) to the ALMEMO® instruments are provided in the ALMEMO® manual. All
standard sensors with ALMEMO® connector usually have the measuring range
and dimension already programmed and can be immediately connected to any
input socket. A mechanical coding ensures that sensor and output modules
can only be connected to the correct sockets. Furthermore, each ALMEMO®
connector has two locking levers that snap in when the insertion into the socket
is established and that prevent a disconnection caused by pulling the cable.
Both levers must be pressed on the sides or disconnecting the connector.
8.2 Measuring inputs and Additional Channels
The measuring instrument ALMEMO® 2890-9 has 9 input sockets (2) that the
measuring channels M0 to M8 are initially allocated to. However, ALMEMO®
sensors can, i required, provide up to 4 channels so that 36 channels are
available with 5 input sockets. The additional channels can be especially used
with humidity sensors with 4 measuring variables (temperature/humidity/dew
point/mixture ratio) or used or unction channels. I required, the sensor can
also be programmed with several ranges or scaling or, depending on the pin
assignment, 2 or 3 sensors can be combined in one connector (e.g. rH/Ntc,
mV/V, mA/V etc.). The additional measuring channels o a connector are in-
creased in steps o 10 (e.g. the irst sensor has the channels M0, M10, M20,
M30, the second sensor has the channels M1, M11, M21, M31 etc.).
Device-internal channels:
An innovation on this device is its our extra additional channels. The irst o
these M5 is programmed by de ault as di erential channel M1 – M0. This only
happens, however, i there are two sensors with the same units and same dec-
imal point position connected at measuring points M0 and M1. However, all
our channels can be programmed with any other unction channels (e.g. U-
Bat, CJ, average, volume low, etc.); (s. 11.3.10, man. 6.3.4). Mb1 = M1 and
Mb2 = M0 are always used as re erence channels by de ault.
Advantage o the device-internal channels: I several sensors are being used
or the same application, they do not have to be reprogrammed and can be
reely exchanged without losing their unction channels. I , however, the whole
application operates with just one sensor, then programming in the sensor it-
sel makes more sense.
On the measuring instrument this gives the ollowing channel assignment:
18 ALMEMO® 2890-9