AlkaWay HIM-19 User manual

Hydrogen Inhaler User manual
OPERATING INSTRUCTIONS
Model: HIM-19 Version 6.0

2
Introduction
Congratulations on purchasing the AlkaWay Hydrogen Inhaler. We
hope you enjoy many hours of beneficial hydrogen over the years to
come.
And thank you for supporting our family business.
If you have any questions, please don’t hesitate to contact us on
1800 268 469. We are here to help.
Why AlkaWay?
We have been in business since 2000, and we are an Australian-owned and operated company. In our
opinion, a positive client experience starts with providing good service. Our goal is to maintain a high
level of customer service, with an emphasis on continual improvement. Each customer's situation is
different, and we recognise this, so we work hard to find the ideal solution for each one.
Why this Inhaler?
This hydrogen inhaler has been scrutinised extensively with inhouse testing, before we put our name
to it. We consider this unit to be the best Hydrogen inhaler on the market today. It is intelligently
designed with safety and user in mind.
300 to
900ml/min
TDS detection
Hydrogen
Gas
Low Noise
Water Level
detection
Hydrogen
Rich water
bubbler
Hydrogen
Purity
99.9%
Smart
Indicator
Lamp
Security
Settings
SPE
Technology

3
Table of contents
Introduction 2
How to operate 3
1. Unboxing 3
2. Using the machine for the first time 4
3. How to use the machine (Machine operation) 4
4. Precautions 5
5. Creating Hydrogen water with the nano rod bubbler 5
6. Inhaling Hydrogen 6
7. Inhaling oxygen and hydrogen together 6
Cleaning 7
Basic parameters 7
Disclaimer 7
Trouble shooting 8
(Please read this user manual carefully before using the Hydrogen Inhaler)
How to Operate
1. Unboxing
1. Parts list
a. Hydrogen generator machine
b. TDS meter (LR44 button batteries not included)
c. Stainless steel nano-rod bubbler
d. Cannulas
e. White connection tubes
f. Y connectors
2. First follow the instructions under the “Using the machine for the first time”section, and
3. Read the Precautions section (Very important).
Drain Outlet
Distilled Water Tank
Tank Water
Level
Observation
Window
Display Screen
Hydrogen Outlet
Oxygen Outlet
Humidification
Bottle
Front of unit
Rear of unit
Power switch
and socket

4
Drain
Outlet
Tank Cap
Tank Water
Level
Observation
Window
Power
Switch and
Socket
Hydrogen
Outlet
Oxygen
Outlet
2. Using the machine for the first time
1. Remove any caps on the Hydrogen and Oxygen outlets.
2. Unscrew and remove the tank cap, then pour distilled
water (TDS<3) into the tank. Make sure the tank is at
least ¾ full.
3. The drain outlet is located on the back of the machine.
Position the back of the machine over a sink. Unscrew
the drain cap. Water will pour out as soon as the cap is
removed. Drain out about a third of the water. Then
replace the drain cap.
This is to ensure that the system is flushed fully. Leave
overnight to fully soak the membrane and to ensure that
the electrolytic cell is fully wet.
4. Connect the power cord and turn the power on (power
switch and cord socket is on the back of the machine). The
machine will beep and the screen will turn on.
5. Unscrew and remove the tank cap and top up the
water tank with distilled water (TDS<3). To check
the water level, use the water level observation
window on the back of the machine. Then screw the
tank cap back on.
6. If the water level is low an alarm will sound until the
water level is back above the minimum level.
3. How to use the machine (Machine operation)
1. To turn the machine on, the main power switch is located on the
back of the machine (red switch).
2. Press the button to set the flow rate. Pre-set flow rates, 300, 450, 600, 750 and
900 ml/min. For initial use, start with a lower flow rate until you get use to the hydrogen.
3. Press the button key to set the countdown clock. Pre-set times, 1,2,3,5 and 8 hours.
4. After setting the flow rate and the time, press the start button to start the machine.
5. Press the button to turn OFF and ON the background light and the display.
6. When the machine is working, press the button to stop the machine and reset the
timer.
7. Once the countdown timer reaches zero the machine will beep and stop generating hydrogen.
8. When you have finished using the machine, turn it off at the power switch, (the red switch on
the back of the machine.

5
4. Precautions
1. For adult use only.
2. Use only distilled water with a TDS of less than 3 to avoid voiding your warranty.
Water with a TDS greater than 3 WILL damage the Cell. For the recommended brand
of distilled water contact our Sales Support Team - 1800 268 469.
3. Always use in a well-ventilated or spacious room.
4. Operating temperature 5 to 30°C with a relative humidity of between 25 to 75°.
5. Do NOT use near an open flame (including smoke, burning incense or cigarettes) or a
heater, or anything that sparks or has an open flame (within 3m).
6. Do NOT use when going to sleep. Remain awake at all times when using the
machine.
7. Keep the power plug away from water.
8. Do NOT block the hydrogen and oxygen outlet holes when using the machine.
Blocking these outlets will void the warranty.
9. Do NOT connect the bubbler to the Oxygen outlet, as the back pressure will cause
damage to the cell and will void the warranty.
10. To maximise and maintain the optimal efficiency, the hydrogen generator should be
rested for 2-3 hours after more than 2 consecutive hours of use. After each shorter
time session, it is ideal to rest the machine off for 1-2 hours.
11. Don’t use the machine without water in it, this too voids the warranty.
5. Creating Hydrogen water with the bubbler
1. Push one end of the white connection tube onto
the small end of the nano-rod bubbler.
2. Push the other end of the white connection tube
onto the Hydrogen outlet on the machine.
3. Keep the Oxygen outlet open (not covered).
4. Place the larger end of the bubbler into a glass of
water and start the machine.
5. Set the flow rate to 150ml/min.
6. Hydrogen-rich water will take between 5-10
minutes. So, stop the machine after 10 minutes.
Hint: The colder the water the more Hydrogen can be
absorbed by the water.
Caution: Only connect the bubbler assembly to the Hydrogen outlet. DO NOT connect it to the
Oxygen outlet, as the back pressure will cause damage to the Cell and will void the warranty.
Stainless steel nano-rod bubbler
White connection tube

6
Hydrogen Outlet
Y-Connector
A
B
C
6. Inhaling Hydrogen
1. Push the white connection tube onto the Hydrogen
outlet.
2. Keep the Oxygen outlet open (not blocked).
3. Fit the cannula onto your nose as shown below.
4. Turn on the machine.
Hints: Initially, start with 30 minutes a day at a flow
rate of 300ml/min until the body gets used to the
Hydrogen (1 week).
Short sessions are more beneficial than longer
sessions (1 hour ideally). Also rest the unit for 1-2
hours between each 1-hour session.
Caution: Do NOT use near an open flame (including
smoke, burning incense or cigarettes) or a heater or
anything that sparks within 3 meters.
7. Inhaling Hydrogen and Oxygen
1. Push one end of 1 white connection tube onto the Oxygen outlet and the other end onto the Y-
Connector (see image below), connect to the end indicated by the letter A.
2. Push one end of another white connection tube onto the Hydrogen outlet and the other end
onto the Y-Connector (see image below), connect to the end indicated by the letter B.
3. Push the connection end of the Cannula onto the Y-Connector (see image below), connect to
the end indicated by the letter C
4. Fit the Cannula onto your nose as shown in the Inhaling Hydrogen section above.
5. Then, turn on the machine.
Oxygen Outlet
Cannula
Water Chamber
Cannula vapour chamber
captures excess water vapour
and water droplets from the
tube. Empty after when needed.
Cannula
White connection tube

7
Drain Outlet
Distilled Water Tank
Water Level
Observation
Window
Cleaning
1. If the machine is used for more than 3 hours a day, it is
recommended to replace the water in the machine every 7
days.
2. Otherwise, it is recommended to replace the
distilled water after every 20 hours of use.
3. To clean, simply position the back of the
machine over a sink. Unscrew the tank cap, then
unscrew the drain cap. Water will pour out as
soon as the drain cap is removed.
4. Refill the tank with distilled water (TDS<3). Allow
to soak for 60 minutes, then drain out the water.
Refill again with fresh distilled water (TDS<3).
5. If transporting the machine, drain out all the
distilled water. Once you're ready to use the
machine again, refill the tank with distilled water
(TDS<3).
6. If storing the machine for more than 2 weeks, drain out
the distilled water. Once you're ready to use the machine
again, flush the machine as above and then refill the tank with
distilled water (TDS<3).
7. Periodically empty out the water in the humidification bottle (refer to images on page 3).
Caution: Use only distilled water with a TDS of less than 3 to avoid voiding your warranty. And
don’t use the machine without water in it, this too voids the warranty.
Basic parameters
Product Name
Hydrogen Generator
Model
HIM-19
Hydrogen Output
0-900mL/min
Output Pressure
200kPa
Hydrogen Purity
99.9%
Input Power
<300W
Voltage
(AC)100V-240V 60Hz/50Hz
Water Tank Capacity
2L
Weight
10.6KG
Size
375mm (H) x 355mm (L) x 250mm (W)
Disclaimer
These products have not been medically evaluated and are not intended to diagnose, cure, treat or
prevent disease. Seek medical advice if seriously ill or attempting to cure or prevent disease.
This product is not a medical device and is not advertised for any therapeutic effects.

8
Trouble Shooting
Issue
Possible Reason
Solutions
The machine is
not starting
1.
Power plug is not plugged in
properly
2.
ON/OFF button has not been
pressed
3.
Is in screen saver mode
1.
Check that the plug is plugged in
properly
2. Check the Power is ON at the wall
3. Press the setting icon
There is no
Hydrogen gas
1. Not enough water in tank
2. The cannula is not properly
connected
3. The bubbler is not properly
connected
4. TDS value of the distilled water is
too high
1. Fill the water tank to the top line on
the water viewing panel.
2. Ensure the cannula is properly
connected and there are no tangles,
kinks or that the tube is blocked with
water
3. Replace and flush with distilled water
(TDS<3) water in the tank and start
it again
4. Contact your Sales Support Team
Imbalance error
1. Unit has been tilted more than 30
degrees and will return an
imbalance error
1. Place machine on a level surface and
restart the machine.
Abnormal display
value
1. Scale formation in electrolytic cell
2. The display is abnormal
3. The TDS value is too high
1. Contact your Sales Support Team -
1800 268 469
Alarm sound
1. Water level is low, or
2. Water level too high
1. Add more distilled water (TDS<3) or
2. Drain out water from the tank until
the water level is below the top
(Max) mark on the water viewing
panel
Screen blank
1.
Screen saver mode
2.
Machine is not turned on
1. Press the setting icon
2. Turn on power
If you run into any other problems, contact your AlkaWay Sales Support Team on 1800 268 469 or