alre JDS-211 Instruction manual

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 1 of 16
Manual for connection and operation of
JDS-211
(item number: G8000409 00)
as of version 2.3
Fa. ALRE-IT Regeltechnik GmbH • Richard-Tauber-Damm 10 • D-12277 Berlin • Phone: (030) 39984-0

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 2 of 16
I N D E X
1. Safety regulations..........................................................................................................3
2. Introduction....................................................................................................................4
3. Electric Connection .......................................................................................................4
3.1. Terminal assignment................................................................................................................................................ 4
3.2. Connection data ....................................................................................................................................................... 5
3.3. Probe connection...................................................................................................................................................... 5
3.4. Connecting switching outputs.................................................................................................................................. 5
4. General notes to the operating of the menu .........................................................6
5. Configuration .................................................................................................................6
5.1. Output function........................................................................................................................................................ 6
5.1.1. 2-point-controller, 3-point-controller, 2-point-controller with alarm................................................................ 7
5.1.2. Min-/Max-Alarm (getrennt oder gemeinsam).................................................................................................. 8
5.2. Offset- and slope-adjustment: Menu calling and adjustment................................................................................... 8
6. Operation........................................................................................................................9
6.1. Min-/max-value memory: ........................................................................................................................................ 9
6.2. Switching points and alarm-boundaries................................................................................................................... 9
6.2.1. 2-point-controller, 3-point-controller................................................................................................................ 9
6.2.2. 2-point-controller with alarm.......................................................................................................................... 10
6.2.3. Minimum/maximum-alarm (individual or common)..................................................................................... 10
6.3. Manual adjustment of the manipulating variable (manual control) ....................................................................... 11
7. Serial interface .............................................................................................................11
8. Error codes...................................................................................................................12
9. Specification ................................................................................................................13
10. Disposal notes ...........................................................................................................13
Addendum A: PID-control definition ..............................................................................14
Addendum B: Tips for using the JDS-211 as heating controller .................................15

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 3 of 16
1. Safety regulations
This device was designed and tested considering the safety regulations for electronic measuring devices.
Faultless operation and reliability in operation of the measuring device can only be assured if the General
Safety Measures and the devices specific safety regulations mentioned in this users manual are considered.
1. Faultless operation and reliability in operation of the measuring device can only be assured if the device
is used within the climatic conditions specified in the chapter “Specifications“
2. Always disconnect the device from its supply before opening it. Take care that nobody can touch any of
the unit‘s contacts after the device has been installed.
3. Standard regulations for operation and safety for electrical, light and heavy current equipment have to be
observed, with particular attention paid to the national safety regulations (e.g. VDE 0100).
4. When connecting the device to other devices (e.g. the PC) the interconnection has to be designed most
thoroughly, as internal connections in third-party devices (e.g. connection of ground with protective earth)
may lead to undesired voltage potentials.
5. The device must be switched off and must be marked against using again, in case of obvious malfunc-
tions of the device which are e.g.:
-visible damage.
-no prescripted working of the device.
-storing the device under inappropriate conditions for longer time.
When not sure, the device should be sent to the manufacturer for repairing or servicing.
ATTENTION: When running electric devices, parts of them will always be electrically
live. Unless the warnings are observed serious personal injuries or damage to prop-
erty may result. Skilled personnel only should be allowed to work with this device. For
trouble-free and safe operation of the device please ensure professional transport,
storage, installation and connection as well as proper operation and maintenance.
SKILLED PERSONNEL
Are persons familiar with installation, connection, commissioning and operation of the product and have
professional qualification relating to their job.
For example:
• Training and instruction or qualifications to switch on or off, isolate, ground and mark electric circuits and
devices or systems.
• Training or instruction according to the state.
• First-aid training.
ATTENTION:
Do NOT use this product as safety or emergency stopping device, or in any other ap-
plication where failure of the product could result in personal injury or material dam-
age.
Failure to comply with these instructions could result in death or serious injury and
material damage.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 4 of 16
2. Introduction
The JDS-211 is a microprocessor controlled displaying,
monitoring and controlling device.
The devices supports an interface for the connection of Pt100
temperature probes.
Additionally the JDS 211 features two switching outputs.
The state of the switching outputs (relays) is displayed with the
LED’s “1” and “2”.
Furthermore all devices supports a serial interface for communicating with a host computer.
When leaving our factory the JDS-211 has been subjected to various inspection tests and is completely cali-
brated.
Before the JDS-211 can be used, it has to be configured for the customer’s applica-
tion.
Hint: In order to avoid undefined input states and unwanted or wrong switching processes, we
suggest to connect the device’s switching outputs after You have configured the device
properly.
Hint: By calling a configuration menu for the offset- and slope-adjustment the measurement and
regulation of the device will be deactivated.
By leaving the menu the device will be reinitialised and the measuring/regulation will be
started again.
3. Electric Connection
Wiring and commissioning of the device must be carried out by skilled personnel only.
In case of wrong wiring the device may be destroyed. We can not assume any warranty in case of
wrong wiring of the device.
3.1. Terminal assignment
15 Interface
14 Interface
13
12 PT100
11 PT100
10 PT100
7Relay 2, break contact, *1
6Relay 2, make contact, *1
5Relay 2, input, *1
4Relay 1, make contact, *1
3Relay 1, input, *1
2Supply voltage 230VAC, *1
1Supply voltage 230VAC, *1
*1 = or the corresponding designation on the label on the housing
15 14 13 12 11 10 765 43 21
Taster 5
SET
1234
SET Cal
T
SP/Tn
min.
Xp/Tv
max.
Scale
Analog
Alarm
1
2

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 5 of 16
3.2. Connection data
typical limitations
between
terminals min. max. min. max. notes
Supply voltage 1 and 2 207 VAC 244 VAC 0VAC 253 VAC
or corresponding des-
ignation on the type
plate
Output 1 (make contact) 3 and 4 253 VAC
5(1) A
or corresponding des-
ignation on the type
plate
Output 2 (change-over
contact) 5, 6 and 7 253 VAC
10(2) A
or corresponding des-
ignation on the type
plate
Input PT100 10 - 12 0 Ω∞Ωactive signal not al-
lowed
Interface 14 and 15 12 V 36 V 0 V 42 V
These limits must not be exceeded (not even for a short time) !
3.3. Probe connection
Please take care not to exceed the limitations of the inputs when connecting the device as this may lead
to the destruction of the device.
Note: The Pt100 probe should be connected as 3-wire system, because there the resistivity is mostly
compensated. 4-wire probes should be connected as 3-wire system, too. If a 2-wire system is essential,
the potential deviation can be compensated by the offset adjustment (chapter 5.2).
Pt100-temperature probe (3-wire) Pt100- temperature probe (4-wire) Pt100- temperature probe (2-wire)
3.4. Connecting switching outputs
Hint: In order to avoid unwanted or wrong switching processes we suggest to connect the
device’s switching outputs after you have configured the device’s switching outputs
properly.
The device features two switching outputs (relays) by default.
Please take care that you must not exceed the limits of the voltage and of the maximum current
of the switching outputs (not even for a short period of time). Please take extreme care when
switching inductive loads (like coils or relays, etc.). Their high voltage peaks have to be limited
by adequate protective measures (e.g. RC-element).
Note: In case of configuring one output as an alarm output, the output will be active in idle state (no
alarm present). The output relay opens when an alarm condition occurred.
Pt100
10
11
12
Pt100
10
11
12
Pt100
10
11
12
set link
not connected

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 6 of 16
4. General notes to the operating of the menu
By means of button 1 you can go to the next parameter.
Additionally a given changing in the parameter setting can be con-
firmed by this button and the new value will be saved. Afterwards it
will be changed to the parameter view again
By means of button 2 or button 3 you can go from the parameter
view to the parameter setting and adjust its value there.
Hint: The buttons 2 and 3 are featured with a ‘roll-function‘. When pressing the button once the value will be raised
(button 2) by one or lowered (button 3) by one. When holding the button pressed for longer than 1 sec. the
value starts counting up or down, the counting speed will be raised after a short period of time.
The device also features a ‘overflow-function‘, when reaching the upper limit of the range, the device switches
to the lower limit, vice versa.
By means of button 4 a given changing will be cancelled in the parameter setting. The changing will be
discarded and the former parameter value will be preserved.
In the parameter view button 4 closes the menu.
Please note: If you don’t press any button for more than 10 sec. in the parameter setting, the adjustment will
be cancelled, the changing discarded and it will be changed to the parameter view. If you don’t
press any button for more than 60 sec. in the menu, the menu will be automatically closed.
5. Configuration
5.1. Output function
- Turn the device on and wait until it completed its built-in segment test.
- At the same time press the pushbutton on the rear side (button 5) and button 1 for >2 seconds.
- The device displays “outP“ ('output').
- Use button 2 or button 3 to select the output function.
- Press button 1 to validate the selected output function. The display shows “outP“ again.
- According to the selected output function further adjustments (see table below) have to be done.
“outP“Output function.
According to the selection the outputs will be allocated as the following table shows.
Description To select
as output
Output 1
(make contact)
Output 2
(change-over
contact)
See chapter
No output, device is
used as display unit no off off --
2-point-controller
2P
PID switching func-
tion
motorised valve
control open
off
motorised valve
control close
5.1.1
3-point-controller 3P PID switching func-
tion switching function 2 5.1.1
2-point-controller with
Min-/Max-alarm 2P.AL PID switching func-
tion
Min-/Max-alarm,
inverse 5.1.1
Min-/Max-alarm,
common AL.F1 off Min-/Max-alarm,
inverse 5.1.2
Min-/Max-alarm,
individual AL.F2 Min-/Max-alarm,
inverse
Min-/Max-alarm,
inverse 5.1.2
SET
1234
SET Cal
T
SP/Tn
min.
Xp/Tv
max.
Scale
Analog
Alarm
1
2

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 7 of 16
5.1.1. 2-point-controller, 3-point-controller, 2-point-controller with alarm
This chapter describes device how to configurate the switching function and how to adjust the switching values
of the device used as a 2-point-controller or 3-point-controller. This instruction demands that you selected
“2P“, “3P“ or “2P.AL“ as your desired output function like it is explained in chapter 5.1.
By means of button 1 you validate the set parameter or go to the next parameter.
By means of button 2 or button 3 you choose your desired value.
By means of button 4 you exit the output configuration menu.
Depending on the selected output function and type of control some parameters may not be displayed and can
not be set.
For an explanation of the single terms see addendum A
“1.rEG“ type of control
possible settings:
Pid.H PID-switching function heating
Pid.C PID- switching function cooling
3Pt.H 3point motorised valve control heating (only at “2P“)
3Pt.C 3point motorised valve control cooling (only at “2P“)
“1.SP“ set point value
possible settings: min. ... max. display range
“1.Pb“ Proportional band
possible settings: 1...9999 (position of the dot depends on display range)
“1.Int“ Integral time in sec. (I-action)
possible settings: off, 1...9999.
“1.dEr“ Derivative time in sec. (D-action)
possible settings: off, 1...9999.
“1.CyC“Cycle time in sec.
possible settings: 0,1 ...320,0.
“1. dur“Propagation time of propulsion in sec (only at “ 3Pt.H“ or “ 3Pt.C“)
possible settings:: 0,1 ... 999,9.
“1. thr“Minimum value of manipulating variable in % (only at “ 3Pt.H“ or “ 3Pt.C“)
possible settings: 0,0..20,0
“1. Err“Preferred output state in case of an error
possible settings:
Type of control value effect
“Pid.H“ or “ Pid.C““on“Output 1 is active in case of an error.
“Pid.H“ or “ Pid.C““off“Output 1 is inactive in case of an error.
“3Pt.H“ or “3Pt.C““on“In case of an error is:
output 1 on !100%
output 2 off
“3Pt.H“ or “3Pt.C““off“In case of an error is:
output 1 off
output 2 on !0%
“2.0n“Turn-on-point of switching function 2 ( only at “3P“)
possible settings: min. ... max. display range.
“2.0ff“Turn-off-point of switching function 2 ( only at “3P“)
possible settings: min. ... max. display range
“2. dEL“Delay of switching function 2 ( only at “3P“)
possible settings: 0,00 .. 2,00

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 8 of 16
“2. Err“Preferred state in case of error ( only at “3P“)
possible settings:
„On“ output in case of error active (relay on)
„Off“ output in case of error inactive (relay off)
“AL.Hi“Maximum alarm-value (only at“2P.AL“ )
possible settings: AL.Lo .. max. display range
“AL.Lo“ Minimum alarm-value (only at“2P.AL“)
possible settings: min. display range .. AL.Hi
“A.dEL“Delay of the alarm-function in sec (only at “2P.AL“)
possible settings: 0 .. 9999
When pressing button 1, the display shows “outP“ again. Now you have finished the output configuration.
Press now button 4 to exit the menu.
5.1.2. Min-/Max-Alarm (individual or common)
This chapter describes how to adjust the device‘s alarm boundaries for min-/max-alarm-monitoring.
This instruction demands that you selected “AL.F1“ or “AL.F2“ as your desired output function like it is ex-
plained in chapter 5.1.
By means of button 1 you validate the set parameter or go to the next parameter.
By means of button 2 or button 3 you choose your desired value.
By means of button 4 you exit the output configuration menu.
Please note that the alarm-outputs are inverted! This means, that the output will be active when there is no alarm!
“AL.Hi“ maximum alarm-value
possible settings: AL.Lo ... max. display range
“AL.Lo“ minimum alarm-value
possible settings: min. display range ... AL.Hi
“A.dEL“alarm delay in sec
possible settings: 0 .. 9999
When pressing button 1, the display shows “outP “ again. Now you have finished the configuration. Press
now button 4 to exit the menu.
5.2. Offset- and slope-adjustment: Menu calling and adjustment
- Turn on the device and wait after it finished its built-in segment test.
- At the same time press the pushbutton on the rear side (button 5) and button 3 for >2 seconds. The
device displays “OFFS“ (offset).
- Use button 2 or button 3 for setting the desired zero point offset-value.
The input of the offset will be in digit or °C.
The value that had been set will be subtracted from the measured value.. (for further information see “scale”)
- Press button 1 to validate your selection. The display shows “OFFS“ again.
- When pressing button 1 again, the device will be displaying “SCAL“. (scale = slope)
- Use button 2 or button 3 to select the desired slope-adjustment.
The slope adjustment will be entered in %. The displayed value is calculated depending on the input sig-
nal according to the following formula.
Displayed value = (measured value – zero point offset) * (1 + slope adjustment [% / 100])
Example: The setting is 2.00 => the slope has risen 2.00% => slope = 102%.
When measuring a value of 1000 (without slope-adjustment) the device would display 102 (with slope
adjustment of 102%)
- Press button 1 to validate the selection of the slope-adjustment. The display shows “SCAL“ again.
- When pressing button 1 again, the display shows “OFFS“ again.
Now you have finished the offset- and slope-adjustment.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 9 of 16
- Press now button 4 to exit the offset- and slope-adjustment menu.
Example for offset- and slope-adjustment:
The device displays the following values (without offset- or slope-adjustment): 2°C at 0°C and 102°C at 100°C
Therefore you calculated: zero point: 2
slope: 102 – 2 = 100 (deviation = 0)
You have to set: offset = 2 (= zero point-deviation)
scale = 0.00
6. Operation
6.1. Min-/max-value memory:
The device features a minimum/maximum-value memory. In this memory the highest and lowest performance
data is saved.
Calling of the minimum-value press button 3 shortly the device will display “Lo“ briefly, after that
the min-value is displayed for about 2 sec.
Calling of the maximum-value press button 2 shortly the device will display “Hi“ briefly, after that
the max-value is displayed for about 2 sec.
Erasing of the min/max values press button 2 and 3 for 2 sec. the device will display “CLr“ briefly, after
that the min/max-values are set to the
current displayed value.
6.2. Switching points and alarm-boundaries
- When pressing button 1 for >2sec. the menu to select the switching points and alarm-boundaries will be called.
- Depending on the configuration you have made in the output configuration menu you will get different
display values. Please follow the specific chapter for further information.
Description Selected
as output
Go on in
chapter Note
No output, device is used as displaying
device no -- Calling of the menu not possi-
ble
2-point-controller 2P 6.2.1
3-point-controller 3P 6.2.1
2-point-controller with min-/max-alarm 2P.AL 6.2.2
min-/max-alarm, common AL.F1 6.2.3
min-/max-alarm, individual AL.F2 6.2.3
6.2.1. 2-point-controller, 3-point-controller
This chapter describes how to adjust the switching values of the device used as a 2-point-controller or 3-
point-controller. This instruction demands that you selected “2P“ or “3P“ as your desired output function like
it is explained in chapter 5.1.
By means of button 1 you validate the set parameter or go to the next parameter.
By means of button 2 or button 3 you choose your desired value.
By means of button 4 you exit the output configuration menu.
“1.SP“set point value
possible settings: min. ... max. display range
“1.Pb“proportional band
possible settings:: 1...9999 (position of the dot depends on display range)
“1.Int“integral time in sec. (I-action)
possible settings:: off, 1...9999.
“1.dEr“ derivative time in sec. (D-action)
possible settings:: off, 1...9999.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 10 of 16
“1.CyC“cyclic time in sec
possible settings:: 0,1 ...320,0.
“2.0n“turn-on-point of switching function 2 ( only at “3P“)
possible settings:: min. ... max. display range
“2.0ff“ turn-off-point of switching function 2 ( only at “3P“)
possible settings:: min. ... max. display range
When pressing button 1, the display shows “1.SP “ again. Now you have finished the configuration.
Press now button 4 to exit the menu.
6.2.2. 2-point-controller with alarm
This chapter describes how to adjust the switching values of the device used as a 2-point-controller with
alarm. This instruction demands that you selected “2P.AL as your desired output function like it is explained
in chapter 5.1. .
By means of button 1 you validate the set parameter or go to the next parameter.
By means of button 2 or button 3 you choose your desired value.
By means of button 4 you exit the output configuration menu.
“1.SP“set point value
possible settings: min. ... max. display range
“1.Pb“proportional band
possible settings:: 1...9999 (position of the dot depends on display range)
“1.Int“integral time in sec. (I-action)
possible settings:: off, 1...9999.
“1.dEr“ derivative time in sec. (D-action)
possible settings:: off, 1...9999.
“1.CyC“cyclic time in sec
possible settings:: 0,1 ...320,0.
“AL.Hi“Maximum alarm-value
possible settings: AL.Lo .. max. display range
“AL.Lo“Minimum alarm-value
possible settings: min. display range .. AL.Hi
“A.dEL“Delay of the alarm-function in sec
possible settings: 0 .. 9999
Example: You want to have a temperature alarm-monitoring of a cold room. The alarm should start when
the temperature rises above -15°C or falls below -30°C. Therefore your settings will be -15°C for
the maximum alarm-value “AL.HI“ and -30°C for the minimum alarm-value “AL.Lo“.
=> The alarm will be starting after the temperature rises above -15°C and stays above -15°C for
the entered delay time or after it had been falling below -30°C and stays below -30°C for the en-
tered delay time.
Please note that the alarm-outputs are inverted! This means, that the output will be active when
there is no alarm!
When pressing button 1, the display shows “1.SP “ again. Now you have finished the configuration. Press now
button 4 to exit the menu.
6.2.3. Minimum/maximum-alarm (individual or common)
This chapter describes how to adjust the device‘s alarm boundaries for min-/max-alarm-monitoring.
This instruction demands that you selected “AL.F1“ or “AL.F2“ as your desired output function like it is ex-
plained in chapter 5.1.
By means of button 1 you validate the set parameter or go to the next parameter.
By means of button 2 or button 3 you choose your desired value.
By means of button 4 you exit the output configuration menu.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 11 of 16
“AL.Hi“Maximum alarm-value
possible settings: AL.Lo .. max. display range
“AL.Lo“Minimum alarm-value
possible settings: min. display range .. AL.Hi
“A.dEL“Delay of the alarm-function in sec
possible settings: 0 .. 9999
Example: You want to have a temperature alarm-monitoring of a greenhouse. The alarm should start when
the temperature rises above 50°C or falls below 15°C.
Therefore your settings will be 50°C for the maximum alarm-value “AL.HI“ and 15°C for the mini-
mum alarm-value “AL.Lo“.
=> The alarm will be starting after the temperature rises above 50°C and stays above 50°C for
the entered delay time or after it had been falling below 15°C and stays below 15°C for the en-
tered delay time.
Please note that the alarm-outputs are inverted! This means, that the output will be active when
there is no alarm!
When pressing button 1, the display shows “1.SP “ again. Now you have finished the configuration. Press
now button 4 to exit the menu.
6.3. Manual adjustment of the manipulating variable (manual control)
This chapter describes how to adjust manually the manipulating variable. This allows to adjust the power
supplied to the process by hand.
The manually adjusted manipulating variable is active till the menu is closed. Then the automatic regulation
will be activated again.
- The menu for the manual adjustment of the manipulating variable is called by pressing button 4 for >2
seconds.
-The device displays “1.Set“.
-“1.SEt“manipulating variable in %
possible settings: 0,0. ... 100,0
-Depending on the configurated type of control the manual adjustment of the manipulating variable is different:
-PID-control:
By means of button 2 or button 3 you choose your desired manipulating variable.
The manipulating variable is immediately active and will persist till the menu is closed or it is changed
again.
-3point-motorised valve control:
By means of button 2 or button 3 you choose your desired manipulating variable and validate it by
button 1.
The controller calculates the needed the activation of the actuator to get the desired manipulating vari-
able and actuates the motor accordingly.
Please note: If the minimum change of manipulating variable is not reached, there will be no activation.
-Press button 4 to exit the manual adjustment of the manipulating variable.
7. Serial interface
The serial interface allows the device to communicate with a host computer. Data polling and data transfer is
done in master/slave mode, so the device will only send data on demand. Every device has a unique ID-
number that makes exact identification of each device possible. With the help of a software (like EbxKonfig –
freeware version available via internet at alre.de) you are able to reassign an address to the device.
Additional accessories needed for the interface mode:
- Level converter: interface of device "PC: i.e. EBW1
- Software for communication with the device
EBS9M: 9-channel-software for displaying a measured value.
EASYCONTROL: universal multi-channel software for real-time-recording and displaying measure-values
of a device in ACCESS®-database-format.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 12 of 16
8. Error codes
When detecting an operating state which is not permissible, the device will display an error code
The following error codes are defined:
Err.1: Exceeding of the measuring range
Indicates that the valid measuring range of the device has been exceeded.
Possible causes: - Input signal to high.
- Sensor broken
Remedies: - The error-message will be reset if the input signal is within the limits.
- Check sensor
Err.2: Values below the measuring range
Indicates that the values are below the valid measuring range of the device.
Possible causes: - Input signal is to low
- Sensor shorted
Remedies: - The error-message will be reset if the input signal is within the limits.
- Check sensor
Err.3: Display range has been exceeded
Indicates that the valid display range (9999 digit) of the device has been exceeded.
Err.4: Values below display range
Indicates that display value is below the valid display range of the device (-1999 digit).
Err.7: System-error
The device features an integrated self-diagnostic-function which checks essential parts of the de-
vice permanently. When detecting a failure, error-message Err.7 will be displayed.
Possible causes: - Valid operating temperature has exceeded o has fallen below the
valid temperature range
- Device defective
Remedies: - Stay within valid temperature range
- Exchange the defective device
Err.9: Sensor defective
The device features an integrated diagnostic-function for the connected sensor or transmitter.
When detecting a failure, error-message Err.9 will be displayed.
Possible causes: - Sensor broken or shorted (Pt100)
Remedies: - Check sensor or exchange defective sensor
Er.11: Value could not be calculated
Indicates a measuring value, needed for calculation of the display value, is faulty or out of range.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 13 of 16
9. Specification
Absolute maximum ratings: see chapter 3.2. (Connection data)
Measuring inputs: PT100, 3-wire
Measuring range: -50.0 ... + 200.0°C
Resolution: 0.1°C
Accuracy: < 0.3% FS ±1Digit (at nominal temperature)
Temperature drift: < 0.015% FS / K
Max. permissible resistivity: 20 Ohm
Measuring freq.: ca. 4 measures / sec.
Display: approx. 13 mm height, 4-digit red LED-display
Operating: 4 push-buttons or by interface
Interface: serial interface, electrically isolated, connectable to RS232 via level converter
Outputs: 2 volt-free Relay-outputs
Output 1: make contact, breaking capacity: 5A (1A), 250 VAC
Output 2: change-over contact, breaking capacity: 10A (2A), 250 VAC
Response Time: < 0.5 sec.
Output functions:2-point-controller, 3-point-controller, alarm
Control function: PID, 3point-motorised valve control, on/off, min-/max-alarm
Power supply: 230 VAC , 50/60 Hz (standard) or the corresponding designation on the label on the housing
Nominal temp.: 25°C
Operating ambient: -20 to +50°C
Relative humidity: 0 to 80% RH (non condensing)
Storage temp.: -30 to +70°C
Housing:
Dimensions: 48 x 96 mm (front-panel dimensions).
Installation depth: approx. 115 mm (incl. screw-in/plug-in clamps)
Panel Mounting: with brackets
Panel cut-out: 43.0 +0.5 x 90.5 +0.5 mm (H x W)
Connection: by screw-in/plug-in clamps
Conductor cross-selection from 0.14 to 1.5 mm²
Protection class: front IP54, with optional mounting seals IP65
EMC: EN61326 +A1 +A2 (appendix A, class B),
additional errors: < 1% FS
If long leads are connected take adequate measures against voltage surges.
That specifications are found out under laboratory conditions according to normal test specification (espe-
cially the DIN norms). The specification are only assured in this respect. The costumer has to assure on his
own that the device is adequate to the considered application and under normal conditions. Therefore we
can not take on responsibility.
Subject to change without notice.
10. Disposal notes
This device must not be disposed as ‘residual waste’.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 14 of 16
Addendum A: PID-control definition
Set point value:
The temperature on which the controlling shall be done.
Manipulating variable:
The power the controller passes on the process.
0 = 0% power; 1 = 100% power.
This is done at
PID-controller by clocking switching function 1 or the continuous output.
Motorised valve control: by switching “switching output 1“ and “switching output 2“.
output 1 = open actuator
output 2 = close actuator
„1.Pb“ (Proportional band):
Bandwidth around the set point value, where the proportional control responds linear.
A small proportional proportion has the effect that a little divergence from the set point value leads to a
big change of the heating power.
A big Proportional proportion has the effect, that the controller intervene weakly. The control gets inert.
„1.Int“ (Integral time, I-action):
Integral time [in sec.] of the PID control algorithms.
The higher the set value, the weaker the effect. The control tends to swinging if integral time is too low.
„1.dEr“ (Derivative time, D-action):
Derivative time [in sec.] of the PID control algorithms. The smaller the set value, the effect.
The derivative time reacts on the change of the manipulating variable.
„1.CyC“ (Cyclic time):
Cyclic time of the control [in sec.]. The control state is detected in the set cycle and the manipulating vari-
able is output accordingly.
At PID-control is this the cyclic time of output 1, too.
„1.dur“ (Propagation time of propulsion):
Time [in sec.] the propulsion needs to get the actuator from “entire close“ to “entire open“.
„1.thr“ (Minimum change of manipulating variable):
The minimum value of the manipulating variable in %, that must be reached, before there is an effect on
the output. By this parameter you can avoid that the actuator actuated too frequently or for a too short pe-
riod of time.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 15 of 16
Addendum B: Tips for using the JDS-211 as heating controller
1. Set point value „1.SP“
The temperature on which the controlling shall be done. Please set here the desired temperature.
2. Proportional band „1.Pb“
The proportional band defines how strongly the device reacts on divergence between the actual and the set
point temperature . If the divergence equals the proportional band, the heating power is 100%. For a diver-
gence of 0°C the heating power is 0%.
Example 1: „1.Sp“=200.0, „1.Pb“=50.0
actual temperature=150°C !divergence=50°C !heating power=100%
actual temperature=180°C !divergence=20°C !heating power =40%
Example 2: „1.Sp“=200.0, „1.Pb“=100.0
actual temperature =150°C !divergence=50°C !heating power =50%
actual temperature =180°C !divergence=20°C !heating power =20%
A small proportional proportion has the effect that the device reacts to a little divergence with a big change of
the heating power.
If the proportional action is too small, there will be a over-reaction. The control gets instable.
!! Attention : A too small proportional band can lead too big excess temperature!!
A big Proportional proportion has the effect, that the controller intervene weakly. The control gets inert.
Tips for identification of the proportional band „1.Pb“.
Please set the following values by the configuration of your device:
Set point value “1.SP“: desired temperature
Proportional band “1.Pb“: 150.0
Integral time “1.Int“: 0 (off)
Derivative time “1.dEr“: 0 (off)
Cyclic time “1.CyC“: 2s (10s for inert control system)
Integral time „1.Int“ and derivative time „1.dEr“ are switched off. The device works as P-controller.
Start the control and wait until the temperature gets constant. This temperature , although much below the set point, has to be stable.
Now shorten the “1.Pb“ value, the divergence from the set point gets smaller. Go on with the scaling down until the temperature is not
stable any more, but oscillates continuously (about ±1°C). If the oscillation is too big re-raise the “1.Pb“ value a little bit.
Please consider the long settling time of some control systems you have to wait for.
Your “1.Pb“ value is the double of the value identified by that procedure.
3. Integral time „1.Int“
In chapter A.2 the device was a P-controller, who reacts only on divergences between actual and set point
temperature. Because there is no heating power at the divergence 0°, the actual temperature will always be
below the set point temperature. That changes with the configuration of the device as a PI-controller. The PI-
controller is additionally a integral controller, i.e. that it not only reacts on divergences, but considers also how
long this divergences have existed. The longer a divergence have existed, the bigger the heating power is. The
heating power changes while a divergence exists. Therefore even small divergences can be corrected in the
long run.
The strength of this effect is adjusted by the integral time „1.Int“. The device regulates the temperature quickly to
the set point temperature.
If “1.Int“ is too small, there will be a overshoot. The control gets instable and the temperature oscillates around
the set point.
!! Attention : A too small integral time “1.Int“ band can lead too big excess temperature!!
If 1.Int“ is too big, it may last very long until the device regulates the temperature to the set point.

E32.0.A3.6C-01 Manual for connection and operation of the JDS-211 page 16 of 16
Tips for identification of the integral time „1.Int“
Please set the following values by the configuration of your device:
Set point value “1.SP“: desired temperature
Proportional band “1.Pb“: value identified in chapter A.2
Integral time “1.Int“: 3600
Derivative time “1.dEr“: 0 (off)
Cyclic time “1.CyC“: 2s (10s for inert control system)
Derivative time “1.dEr“ is switched off. The device works as PI-controller. Start the control and wait until the
temperature gets to the set point. If this takes too much time shorten the value of “1.Int“. To avoid overshoots
please watch how fast the temperature rises. Is the rise too small, shorten “1.Int“ again. Is the rise already
very big, raise “1.Int“. If “1.Int“ is not too small the temperature should be stable at the set point.
Now there has to be an external interference on the control system (i.e. fill in cold water, put a cold object
onto the controlled metal block, etc.). This interference should be a quite realistic to that under operating
conditions. If no external interference can be carried out, you can also change the set point value.
Watch how the device regulates the temperature to the set point. If there is a overshoot the “1.Int” value has
to be raised. If the temperature approach the set point only very slowly the “1.Int” value can be shorten.
Please consider to wait the settling time (up to several minutes for some control systems) after each change
of “1.Int”.
Your “1.Int” value is 1.2 times the value identified by that procedure.
4. Derivative time „1.dEr“
If cold water is filled in a heated water tank the temperate falls rapidly. A human operator, who regulates the
temperature of the tank manually, will intuitively turn immediately the full heating power on, then cut it down to
approach the set point by small changes of the heating power.
The D-action (derivative action) of the PID-controller is responsible for the intervention at that big by big tem-
perature changes. The D-action doesn’t react on the divergence between actual and set point temperature, but
on temperature changes. If the temperature falls very quickly there will be a big heating power due to the D-
action. If the actual temperature approaches the set point very fast the D-action reduces the heating power
calculated by the P- and I-action. If there is no temperature change the D-action is zero.
The strength of this effect is adjusted by the derivative time “1.dEr“. A small “1.dEr“ value means, that there is
only a small reaction on temperature changes. A big one means, that there is a big reaction on changes.
In a lot of application the use of the device as a PI-controller is entirely satisfactory. In this case set the deriva-
tive time “1.dEr“ to 0.
If the PID-control is needed, “1.dEr“ has to be identified. This requires exact information about the controlled
process and knowledge of control engineering. However, a derivative time “1.dEr“=Integral time “1.Int”/5 has
delivered an optimal performance in practice.
Your “1.dEr“ value is 0.2 times the “1.Int“ value.
5. Cyclic time “1.CyC“
The device regulates the heating power by switching the existing heating on and off. If only 50% of the ex-
isting heating power are needed the heating is only half the time switched on. The frequency switching on
and off is adjusted by the cyclic time “1.CyC“.
Example: existing heating power 1000 W, 600 W heating power are needed
At a period time T=10s: the device switches the heating on for 6s and then off for 4s
At a period time T=200s: the device switches the heating on for 120s and then off for 80s
If the cyclic “1.CyC“ is too high the temperature of the heated object will get too high during the switch-on
period (although „1.SP“, „1.Int“, „1.dEr“ are correct) , only to cool down in the following switch-off period.
A very small “1.CyC“ value means that the relays switches a lot of times and that shortens the durability.
Therefore the cyclic time is ideally set to that value that is as high as possible, but where the effect during the
switch-on and –off periods is just not perceptible.
Tips for identification of the Cyclic time “1.CyC“:
Raise the cyclic time as long as the of the controlling deterioration is just not perceptible.
Your „1.CyC“ is 08. times the value identified by that procedure.
Table of contents