Ambul D-Tiger 6000S User manual

密级:★★★★ ABMOBAN0001(A1.0)
保密文件,知识产权归安保科技公司所有
文件类型
技术文件
项目编号
606
文件编号
H-1.601.00070
打印要求
彩色 □黑白
6000S User Manual
参考资料
文件编号
说明
修订记录
版本 ECR/PCN
TCN
更改内容
制订
批准日期
A1
/
首次发布
苑振家
2014-09-03
A2
ECR201320
15100053
更新公司 Logo
苑振家
2015-10-22
A3.0
ECR201801
07
更新地址及联系方式
发放部门
生产部 □采购部 质量部 □市场部 □研发部
存档方式
电子文档 纸文档 □其它:
制订
审核
批准
日期
日期
日期
2018.1.22
2018.1.22
2018.1.22

D-Tiger
6000S USER MANUAL
Instructions and Safety Considerations of Ambulanc Transport Ventilator
FILE NO.: H-1.601.00070
VERSION: A3.0
ISSUE DATE: 2018-01-22

H-1.601.00070-A3.0
I
© COPYRIGHT AMBULANC (SHENZHEN) TECH.CO. ,LTD.
2017 ALL RIGHTS RESERVED
Intellectual Property Statement
AMBULANC (SHENZHEN) TECH.CO., LTD. (hereinafter called Ambulanc) owns the
intellectual property rights to this Ambulanc product and this manual. This manual
may refer to information protected by copyrights or patents and does not convey any
license under the patent rights of Ambulanc, nor the rights of others.
Ambulanc intends to maintain the contents of this manual as confidential information.
Disclosure of the information in this manual in any manner whatsoever without the
written permission of Ambulanc is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaption and translation of
this manual in any manner whatsoever without the written permission of Ambulanc is
strictly forbidden.
are the registered trademarks or trademarks owned by Ambulanc
in China and other countries. All other trademarks that appear in this manual are used
only for editorial purposes without the intention of improperly using them. They are the
property of their respective owners.
Responsibility Statement on the Manufacturer Party
Contents of this manual are subject to changes without prior notice.
All information contained in this manual is believed to be correct. Ambulanc shall not
be liable for errors contained herein nor for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.
Ambulanc is responsible for the effects on safety, reliability and performance of this
product, only if:
All installation operations, expansions, changes, modifications and repairs of this
product are conducted by Ambulanc authorized personnel;
The electrical installation of the relevant room complies with the applicable
national and local requirements;
The product is used in accordance with the instructions for use.
Warning
This equipment must be operated by skilled/trained clinical professionals.
It is important for the hospital or organization that employs this
equipment to carry out a reasonable service/maintenance plan. Neglect
of this may result in machine breakdown or personal injury.

H-1.601.00070-A3.0
II
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY
OR FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Ambulanc's obligation or liability under this warranty does not include any
transportation or irresistible natural disaster or other charges or liability for direct,
indirect or consequential damages or delay resulting from the improper use or
application of the product or the use of parts or accessories not approved by
Ambulanc or repairs by people other than Ambulanc authorized personnel.
This warranty shall not extend to
Malfunction or damage caused by improper use or man-made failure.
Malfunction or damage caused by unstable or out-of-range power input.
Malfunction or damage caused by force majeure such as fire and earthquake.
Malfunction or damage caused by improper operation or repair by unqualified or
unauthorized service people.
Malfunction of the instrument or part whose serial number is not legible enough.
Others not caused by instrument or part itself.
Safety Information
Danger
Indicate an imminent hazard that, if not avoided, will result in death or
serious injury.
Warning
Indicate a potential hazard or unsafe practice that, if not avoided, could
result in death or serious injury.
Caution
Indicate a potential hazard or unsafe practice that, if not avoided, could
result in minor personal injury or product/property damage.
Note
Provide application tips or other useful information to ensure that you get the
most from your product.

H-1.601.00070-A3.0
III
Product Information
Product Name: Transport Ventilator
Model: 6000S
Manufacturer: Ambulanc (Shenzhen) Tech. Co., Ltd.
Manufacturer address:3rd Floor, Block C, Building #5, Skyworth Innovation Industry
Park, Tang Tou 1st Road, Shiyan Town , Baoan District, Shenzhen 518108, China.
Warning
This device is not intended for home using!
Company Contact
Manufacturer: Ambulanc (Shenzhen) Tech. Co., Ltd.
Address: 3rd Floor, Block C, Building #5, Skyworth Innovation Industry Park, Tang
Tou 1st Road, Shiyan Town, Baoan District, Shenzhen 518108, China.
Tel.: +86-755-26072210
Fax: +86-755-23016012
Website: http://www.ambulgroup.com
E-mail: manager@ambu-lanc.com
EC-Representative: Shanghai International Holding Corp. GmbH(Europe)
Eiffestrasse 80,20537 Hamburg, Germany
Contact Pers.: Qiming Cheng
Telephone: +49-40-2513175
Fax: +49-40-255726
Preface
Manual Purpose
This manual contains the instructions necessary to operate the product safely and in
accordance with its function and intended use. Observance of this manual is a
prerequisite for proper product performance and correct operation and ensures
patient and operator safety.
This manual is based on the maximum configuration and therefore some contents
may not apply to your product. If you have any question, please contact us.
This manual is an integral part of the product. It should always be kept close to the
equipment so that it can be obtained conveniently when needed.
Intended Audience
This manual is geared for clinical professionals who are expected to have a working
knowledge of medical procedures, practices and terminology as required for
emergency rescue.
Illustrations
All illustrations in this manual serve as examples only. They may not necessarily
reflect the setup or data displayed on your ventilator.

H-1.601.00070-A3.0
IV
Contents
1. Overview.............................................................................................................................1
2. Description..........................................................................................................................4
2.1 Intended use..................................................................................................................4
2.2 Applications..................................................................................................................4
2.3 Users qualification........................................................................................................4
2.4 Ventilation function ......................................................................................................5
2.5 Controlled ventilation (IPPV).......................................................................................5
2.6 Assisted ventilation (Assist).........................................................................................5
2.7 Manual operation mode................................................................................................6
2.8 Patient ventilation Hose System...................................................................................6
3. Safety instructions..................................................................................................................7
3.1 General .........................................................................................................................7
3.2 Safe Using of Oxygen ..................................................................................................8
3.3 Ventilation/Operation...................................................................................................8
3.4 Patient Ventilation Hose System..................................................................................9
3.5 Software........................................................................................................................9
3.6 Accessories/spare parts.................................................................................................9
4. Installation............................................................................................................................10
4.1 Connecting the oxygen cylinder.................................................................................10
4.2 Removing the empty cylinder.....................................................................................10
4.3 Connecting the new cylinder......................................................................................10
4.4 Ventilation Hose System.............................................................................................11
5. Using the ventilator ..............................................................................................................13
5.1 Switching on/ self test ................................................................................................13
5.2 Selecting the ventilation settings................................................................................14
5.3 Selecting ventilation mode ........................................................................................15
5.4 Performing ventilation................................................................................................16
5.5 Monitoring ventilation................................................................................................17
5.6 Ending ventilation ......................................................................................................18
5.7 Alarm signals..............................................................................................................18
5.8 Selecting language for user guidance.........................................................................22
5.9 Audio response messages...........................................................................................22
5.10 Calculation of oxygen content/remaining operating time ........................................23

H-1.601.00070-A3.0
V
5.11 Battery management.................................................................................................23
6. Hygienic preparation............................................................................................................25
6.1 6000S..........................................................................................................................25
6.2 Patient valve ...............................................................................................................25
6.3 Ventilation hose..........................................................................................................25
6.4 Pressure gauge tube....................................................................................................26
6.5 Components and accessories......................................................................................26
6.6 Fittings........................................................................................................................26
6.7 Cleaning, disinfecting and sterilizing.........................................................................27
7. Functional checks.................................................................................................................29
7.1 Intervals......................................................................................................................29
7.2 Checking for leaks in the system................................................................................30
7.3 Checking the patient valve .........................................................................................31
7.4 Checking the alarm systems .......................................................................................31
8. Trouble shooting...................................................................................................................33
9. Servicing...............................................................................................................................35
9.1 Intervals......................................................................................................................35
9.2 Performing technical safety check and servicing .......................................................35
9.3 Battery........................................................................................................................36
9.4 Change Filter..............................................................................................................36
9.5 Storage........................................................................................................................37
9.6 Disposal......................................................................................................................37
10. Product and accessories......................................................................................................38
10.1 Standard product.....................................................................................................38
10.2 Optional Accessories................................................................................................38
10.3 Parts List of Ventilator Bag I....................................................................................39
11. Technical data.....................................................................................................................39
11.1 Specifications ...........................................................................................................39
11.2 Product structure diagram.........................................................................................42
11.3 O2content when using Air Mix................................................................................42
12. Warranty .............................................................................................................................43
13. Storage and transportation..................................................................................................44
14. Electromagnetic Capability................................................................................................45

H-1.601.00070-A3.0
1 –48
1. Overview
1. Pressure monitoring hose
connection
Connect with breather hose and
monitoring of airway pressure.
2. Connection for ventilation hose
Connecting with patient valve and
breather hose
3. Safety valve
When the airway pressure overruns, it
will automatically open the pressure
safety valve.
4. Air filter
Fresh air suction passage.
5. Battery and maintenance interface
Built-in 3400 mAh high-capacity
rechargeable battery, non-professional
couldn’t replacement it; Maintenance
interface can be used for the machine
maintenance, non-professional couldn’t
use it.
6. Speaker
Voice prompt and alarm speaker.
7. O2inlet
Connect with oxygen cylinder.
8. DC interface
Connect the power adapter, or DC
power supply of an ambulance and
rescue helicopter.
9. Adapter
To convert the 100-240 V alternating
current (AC) for 12 V DC, provide power
for breathing machine.
1. Pressure monitoring hose connection
2. Connection for ventilation
hose
空气进口
4. Air filter
7. O2inlet
5. Rechargeable battery
3.Safety valve
8. DC interface
6. Speaker
9. Adapter

H-1.601.00070-A3.0
2 –48
Control of 6000S
1.LED display
Display respiration parameter
2. Alarm panel
Display Alarm message type
3. AC adapter indicators light
Whether it is switched on or off, the power
converter are shown in green, do not
display when no external power supply.
Note: when breathing machine by an
external power supply switch to the
internal power, breathing machine still
work normally.
4. Battery indicators light
Instructs the battery status, Low battery
will show red light, when the battery is
being charged, Flash yellow light, indicator
will be in green once the battery has been
charged completely.
5. Voice alarm mute key
This button could turn off the voice alarm function
for a period of time (2 minutes), when the alarm
mute indicator.
6. Air Mix/No Air Mix switch
Switch between Air Mix/No Air Mix.
7. Minute volume regulator
Set up minute respiratory volume.
8. Manual operation key
Presses the button continually, it will supply
gas to the patients at constant velocity, until
loosen Manual button to stop gas supply.
9. ON/OFF switch assisted ventilation
Press the assisted mode; patient breath under
assisted ventilation mode, Yellow light is
bright.
10. ON/OFF button
Press this button switch on, long press the
power off after 3 S
11. Regulator of Ventilation frequency
Set the patients breathing rate
12. Regulator of Max ventilation pressure
Set the airway pressure limit, if the airway
pressure exceeds this value, will be sounded
the alarm
13. Color code
infant(yellow,10—30Kg)
child(orange,30—60Kg)
Adult(blown,60—110K)
12.Max. ventilation pressure
regulator
11. Regulator for ventilation setting
13.Colour code
2. Alarm panel
1. LED display
3. AC adapter indicators ligth
8. Manual operation key
7. Minute volume regulator
10.ON/Off button
机键
6. Air Mix/No Air Mix switch
9. ON/OFF switch assisted
ventilation
5. Voice alarm mute key
4. Battery indicators light

H-1.601.00070-A3.0
3 –48
Display of 6000S
1. Battery status indicators
when battery is low display red to give a
alarm, rolling green bar when
charging ,more than two compartment
display green.
2. Line operation indicators
Show red fork means no external power
supply.
3. Ventilation modes indicators
Indicate a real-time breathing pattern.
4. Measurement parameters display
mage display the real-time airway
pressure, monitor the airway pressure
within the set maximum pressure, Arc
area filled with green, Paw and mbar
letters show green, more than set the
maximum pressure Arc area filling red,
Paw and mbar letter turn to red to give an
alarm.
5. Mean airway pressure
Show the Mean airway pressure value of
monitoring.
6. Peak airway pressure
Show highest airway pressure value of
monitoring.
Symbols used on the display
Symbol
Meaning
Emergency ventilation mode —Infant
Emergency ventilation mode —Child
Emergency ventilation mode —Adult
Battery status indicator
1. Battery status indicators
3. Ventilation modes indicators
4. Measurement parameters display
6. Peak airway
pressure
5. Mean airway pressure
2. Line operation indicators

H-1.601.00070-A3.0
4 –48
2. Description
2.1 Intended use
The 6000S is an automatic oxygen respiration device(short-term ventilator) with
additional inhalation facility.6000S can control and assist Patient whose weight is more
than 15Kg, as well as non-invasive and invasive mechanical ventilation.
6000S must only be operated when it is security installed or when it is on approved
carrying platforms.
Warning
6000S Should not be used on neonatal patients.
2.2 Applications
You can use 6000S:
To revive patients at the site of the emergency;
For longer periods in more protracted emergencies, e.g. fire;
For short-term O2inhalation using a respiration masks.
You can use 6000S while transporting patients:
Between the various rooms and departments of a hospital;
Between the hospital and other premises;
In emergencies;
When transport over considerable distances is planned.
6000S:
Is designed to provide controlled ventilation to persons of 15kg body weight or
more;
Is used to treat respiration arrest;
Can be preset to parameters that ensure evenly balanced ventilation, provided
that the selected Maximum ventilation pressure Pmax is not exceeded;
Permits respiration-controlled oxygen inhalation in Demand mode.
2.3 Users qualification
6000S must only be used by persons who can verify that they have the following
qualifications:
A medical qualification and training in ventilation techniques.
Training in the use of the 6000S by a person authorized by Ambulanc
Improper use may lead to serious physical injury.

H-1.601.00070-A3.0
5 –48
2.4 Ventilation function
6000S operates within a pressure range of 2.7 to 6 bar and at a flow rate of not less
than 70l/min O2. It has a built-in power pack.
It uses high-pressure, medicinal-grade oxygen. An external pressure reducer brings
this down to the required operating pressure. The oxygen supply is fed in at input valve.
The ventilation settings are infinitely variable. Both this and the inspiration/expiration
ratio of 1:1.67 are regulated by internal electronic control mechanisms.
The gas for inspiration flows along the hose and through the patient valve and either
the mask or the tracheal tube into the patient’s airways. The patient valve is fitted with a
lip membrane that enables expired gas to be conducted away through the expiration
tube.
At the normal Air Mix setting atmospheric air is admixed to give an O2 concentration of
between 55% and 85% at 10 mbar ventilation pressures (note “10.2 O2 content when
using Air Mix”).
In certain indications and in cases where the surrounding atmosphere is contaminated,
you can switch to No Air Mix and ventilate with pure oxygen.
The injector unit is switched off when switching from Air Mix to No Air Mix. This
increases minute volume which can result in the set pressure limit being exceeded and
a stenosis alarm (Stenosis) being triggered. In this case, set minute volume
correspondingly lower.
In the opposite instance, in other words when switching from NO Air Mix to Air Mix, the
injector unit is switch on. This reduces minute volume which can lead to the set
pressure limit being undershot. In this case, set minute volume correspondingly higher.
2.5 Controlled ventilation (IPPV)
After being switched on, the 6000S is automatically in Controlled Ventilation
mode. This administers mandatory ventilation strokes to the intubated patient
according to the ventilation values set on the device.
2.6 Assisted ventilation (Assist)
In addition to Controlled Ventilation mode, the 6000S also has an Assisted Ventilation
mode. After you have switched on Assisted Ventilation mode by pressing the Assist key,
a flashing green LED indicates this mode.
The patient now has the option of triggering a triggered ventilation stroke within a time
window of 40% of expiration. To do so, the patient must generate a flow of over 5 l/min
by his own breathing efforts.
If the breathing efforts of the patient are not sufficient to trigger, the patient
automatically receives a mandatory ventilation stroke at the end of the time window, so
that the set minute volume is guaranteed.

H-1.601.00070-A3.0
6 –48
With this function, the ventilation strokes of the device can be synchronized with the
breathing in air from the surrounding atmosphere via the patient valve.
If the patient does not trigger the device, an alarm is triggered. The patient continues to
receive controlled ventilation.
2.7 Manual operation mode
When 6000S in the IPPV ventilation mode, press the Manually ventilation button (M),
6000S would supply gas to the patients at constant velocity, until loosen manually
ventilation button to stop gas supply. The constant flow rate is consistent with the
parameters which set up under IPPV ventilation mode.
Loosen the button manually ventilation, If the controller did not press the M key within 3 s,
it will automatically switch back to the breathing pattern
2.8 Patient ventilation Hose System
The ventilation gas is supplied to the patient via the Patient ventilation hose system,
comprising the ventilation hose and all leads necessary for comprehensive ventilation
and monitoring. It includes the ventilation hoses and patient valve.
Ventilation hose component is designed to permit spontaneous respiration even if
the 6000S malfunctions.
Ventilation hose is available in two versions:
a) Reusable ventilation hose;
b) Disposable ventilation hose

H-1.601.00070-A3.0
7 –48
3. Safety instructions
For your own safety and that of your patients, please observe the following points.
3.1 General
Always carry out a functional check before using the unit. (See "7. Function
check").
Please read the directions for use carefully. They are an integral part of the
ventilator and should be available for reference at all times.
Before starting to work with 6000S,you must understand how to operate it.
Please comply with section “5. Hygienic preparation”on page 30 to prevent
infection or bacterial contamination.
6000S should be used only by medically qualified personnel who have had
training in ventilation techniques. Incorrect use can cause severe physical
injury.
It is advisable for you to have servicing and repairs carried out only by the
manufacturer Ambulanc, or by qualified technicians expressly authorized by
Ambulanc.
Malfunctions and a lack of biocompatibility may result if third-party articles are
used. Please bear in mind that in these cases any warranty entitlement and
liability shall lapse where the accessories recommended in the instructions for
use or original spare parts are not utilized.
6000S should be used only for the purposes for which it is designed (note “2.1
Uses”on page4).
6000S is not designed for use under hyperbaric conditions (pressure
chamber).
6000S should never be used with flammable anesthetics.
In the case of use in poisoned or low-oxygen atmospheres, do not operate the
6000S with the “Air Mix”setting or in Assist mode.
A back-up ventilator should always be available in case of technical failure.
Risk of injury. Never leave the patient or the ventilator unattended during
ventilation. Only then can you respond quickly if the patient's condition
deteriorates or in the event of an alarm or malfunction. Delayed response on
the part of medical personnel may lead to serious physical injury.
Do not place a switched-on cellular phone or radio closer than 1m from the
6000S, as this could cause malfunctions.
When operating the unit with the power supply unit, make sure that the power
cord cannot cause anyone to trip or cause any obstruction.

H-1.601.00070-A3.0
8 –48
3.2 Safe Using of Oxygen
Highly compressed oxygen can lead to spontaneous explosive reactions in
combination with flammable substances (grease, oil, alcohol, disinfectants, etc,):
Warning
All screw connections and other components of the ventilator must be
kept absolutely free of oil and grease.
Always wash your hands before starting to work on the oxygen supply.
Smoking and open flames are strictly prohibited in the vicinity of all fittings
containing or transporting oxygen.
During assembly and when changing the oxygen cylinder, only hand
pressure should be used when tightening the screw connections to the
cylinder and to the pressure reducer. Never use tools for this purpose.
Excessive tighten damages the screw threads and seals and can cause
leaks.
Protect oxygen cylinders from accidental falls. If a cylinder falls, the
pressure reducer or the valve may break off and cause a violent
explosion.
Caution
Always open the valve of the oxygen cylinder slowly to prevent pressure
damage to the other fittings.
The oxygen cylinder should never be completely emptied as this may
allow moisture-containing air to enter the cylinder and cause corrosion.
Secure the oxygen cylinders so that they cannot fall over. If a cylinder falls
on the pressure reducer or valve, these could break off, causing a violent
explosion.
3.3 Ventilation/Operation
Patient and transport ventilator must be kept under continuous observation
during ventilation.
Prolonged ventilation can lead to atrophy of the muscles (dependency of the
patient on ventilation).
Prolonged ventilation may lead to the airway drying out. Ensure adequate
conditioning of the respiratory air.
Only apply high ventilation pressures for short periods and only if medically
indicated. Permanently applied high ventilation pressures can be injurious to the
patient.

H-1.601.00070-A3.0
9 –48
Make sure that the patient valve is not covered or its function impaired, e.g. by
the patient's position. In the process of ventilation, it must be continuously
observed patient and ventilator.
3.4 Patient Ventilation Hose System
Warning
Risk of injury. Only use the Patient Hose System if you are a qualified
medical professional and have received training in respiration techniques.
Improper use may lead to serious physical injury.
The Patient Hose System must be subjected to a functional check and
visual inspection by the user before use. For this, refer to the instruction
manual for the Patient Hose System.
When connecting the patient valve, check that the direction of flow of the
respiratory gas is correct. Make sure that the expiration opening of the
patient valve is not covered or prevented from functioning, e.g., by the
patient's position.
Only use the Patient Hose System for the purpose described. For this,
refer to the instruction manual for the Patient Hose System.
The Patient Hose System is not suitable for hyperbaric use (pressure
chamber).
Also refer to the instruction manual for the Patient Hose System.
3.5 Software
Risks due to software errors have been minimized by means of extensive qualification
measures.
3.6 Accessories/spare parts
Caution
Protect silicone/rubber parts against UV light and prolonged direct
exposure to sunlight to prevent them becoming brittle.
We recommend that work such as inspections and repairs should be
carried out by the manufacturer, Ambulanc, or by a technician expressly
authorized by Ambulanc.
Malfunctions and a lack of biocompatibility may result if products from
other manufacturers are used. Please bear in mind that in these cases all
warranty rights and liability shall become void if the accessories
recommended in the instruction manual or original replacement parts are
not used.

H-1.601.00070-A3.0
10 –48
4. Installation
As a rule, 6000S only has to be installed for stationary use in rescue vehicles,
helicopters or aircraft. In this case, fastening sets can be supplied as accessories.
If 6000S is supplied complete on a stretcher board, or mounted on an Ambulanc
Emergency Bag, then the device is ready for operation and no further installation work
is required.
Warning
After installation, you must make function check to ensure reliable operation.
(See "7. Function Check")
4.1 Connecting the oxygen cylinder
Warning
Risk of explosion! Wash your hands thoroughly before doing any work on
the oxygen supply. Hydrocarbon compounds (e.g. oil, grease, cleaning
alcohol, hand cream or adhesive plasters) can cause explosive reactions
if they come into contact with highly compressed oxygen.
Never use wrenches or other tools to tighten or loosen the screw
connections.
4.2 Removing the empty cylinder
1. Close the valve on the oxygen cylinder.
2. Switch on 6000S at On/Off switch. This allows the remaining oxygen to
escape and the unit is pressure-free. Wait until the pressure gauge on the
pressure reducer shows an oxygen content of zero before undoing the screw
connection by hand.
3. Switch off 6000S at On/Off switch.
4. Then loosen the screw connection to the cylinder by hand.
4.3 Connecting the new cylinder
1. First open the valve of the new oxygen cylinder, then shut it again quickly. This
will blow away any particles matter.
2. Connect the pressure reducer to the valve on the oxygen cylinder. Tighten the
connecting nut by hand.
3. Screw the pressure hose onto the outlet of the pressure reducer (if not already
connected) using the 9/16’’-18UNF union nut.
4. Screw the other end of the pressure hose to the compressed gas connection
of the 6000S if this has not yet been done.

H-1.601.00070-A3.0
11 –48
Caution
Make sure that the patient is not connected up to the 6000S when you
are establishing the gas supply. Otherwise, the automatic self-test of
machine will lead to incorrect results.
When doing this, hold the valve opening away from your body in such a
way that any flying particles cannot injure yourself or other people.
4.4 Ventilation Hose System
A reusable hose system is supplied with the 6000S. Optionally, a disposable hose
system is also available. The procedure for connecting both systems is as follows:
1. Connect pressure-measurement tube onto the corresponding connection on the
front of the unit. Make sure that the connected tubes are not kinked.
2. Connect the ventilation hose onto the corresponding connection on the front of the
unit. Make sure that the connected hoses are not kinked.
3. Connect the patient valve with the ventilation hose and pressure measurement tube
as below indicated drawing of patient valve.
4. If performing mask ventilation, attach the ventilation mask to the patient valve, or if
the patient is incubated, attach the tube to the patient valve.
Caution
Only grasp the ventilation hose by its ends. Otherwise the hose may be
damaged.
Patient valve

H-1.601.00070-A3.0
12 –48
Warning
We strongly recommend to use the original Accessories supplied by
Ambulanc only, malfunctions and a lack of biocompatibility may result if
products from other manufacturers are used. Please bear in mind that in
these cases all warranty rights and liability shall become void if the
accessories recommended in the instruction manual or original replacement
parts are not used.
1
Outlet
Φ30mm Convex cone Interface
2
Intake
Φ15mm/20mm Coaxial Interface
3
Patients connector (Connect to
mask / intubations)
Φ15mm/20mm Coaxial Interface
4
Emergency air intake
Non-standard interfaces
5
Pressure measurement tube
connecting point
Φ5mm Non-standard cylindrical
interfaces

H-1.601.00070-A3.0
13 –48
5. Using the ventilator
5.1 Switching on/ self test
1. Open the valve of the oxygen cylinder slowly. The
pressure gauge will now show the pressure in the
cylinder.
2. Calculate the remaining operating time (please refer to
“5.9 Calculation the oxygen level/operating time”). Always
change the cylinder in good time, e.g. when the pressure
falls below 50 bar, to ensure a sufficiently long operating
time.
3. Select the desired settings for the ventilation (see “5.2
Selecting the ventilation settings”).
4. Press the ON/OFF button to switch on 6000S, an
automatic self-test runs. The screen will show the
process percentage of self-test, after finishing self-test,
the screen will change to the main interface with voice
prompts “Open oxygen cylinder.” “Check respiration and
select mode, adjust settings.”
If the self-test is unsuccessful, it will be shown in the
screen as the left. You should hear the message “Device
malfunction! Administer alternative ventilation”If this
happens, 6000S must not be used for ventilation.
After the self-test, the ventilator repeatedly checks the oxygen cylinder pressure until
adequate pressure is detected. Otherwise an alarm is sounded. The 6000S will then
start to function with the selected ventilation settings.
Warning
The automatic self-test is not a substitute for the functional check. Before
using the unit, always carry out a functional check, as described in Section "7.
Function check”. That is the only way to ensure that the unit is fully functional.
5. When the screen switches into the main display interface, indicating that the unit is
able to work promptly. All the settings will be the same at the time when you turn off
the unit.
6. Connect to the patient after finish the setting.
7. During the ventilation, may need to adjust the respiratory parameters based on
patient condition. Please refer to following contents of this chapter for details.
Table of contents
Other Ambul Medical Equipment manuals
Popular Medical Equipment manuals by other brands

C-Aire
C-Aire NewLife Elite Technical manual

Life form
Life form CRiSis LF03953U instruction manual

ARJO HUNTLEIGH
ARJO HUNTLEIGH Airtech Instructions for use

Pro-tec
Pro-tec PRS 500 C Instructions for use

Hillrom
Hillrom Welch Allyn 52400 Mounting instructions

Etac
Etac immedia PediTurn Soft Short Instruction