Ametek LAND LSP HD Series User manual

LSPHD System
LSPHD Linescanner Head
User Guide
Publication Nº 805233
Issue 7
Language: English
25 May 2016
© Land Instruments International, 2016

Health and Safety Information
Read all of the instructions in this booklet - including all the WARNINGS and CAUTIONS
- before using this product. If there is any instruction which you do not understand.
DO NOT USE THE PRODUCT.
Safety Signs
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or
personal injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in minor or
moderate injury to the user or users, or result in damage to the product or to property.
CAUTION
Warning laser radiation.
Embedded laser (diode) is Class 1M, <5mW. Do not view directly with optical magniers.
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
1mW MAX OUTPUT AT 635nm
IEC 60825-1:1994
IMPORTANT: Please note radiation safety warning, where this symbol appears.
NOTE
Indicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of
data.
Equipment Operation
Use of this instrument in a manner not specied by Land Instruments International may be hazardous.
Read and understand the user documentation supplied before installing and operating the equipment.
Protective Clothing, Face and Eye Protection
It is possible that this equipment is to be installed on, or near to, machinery or equipment operating at
high temperatures and high pressures. Suitable protective clothing, along with face and eye protection
must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing
this product. If in doubt, contact Land Instruments International.
Electrical Power Supply
Before working on the electrical connections, all of the electrical power lines to the equipment must
be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these
operating instructions. If in doubt, contact Land Instruments International.
Storage
The instrument should be stored in its packaging, in a dry sheltered area.
Unpacking
Check all packages for external signs of damage. Check the contents against the packing note.
Lifting Instructions
Where items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the
Technical Specication for weights. All lifting should be done as stated in local regulations.
IMPORTANT INFORMATION - PLEASE READ

Health and Safety Information
Read all of the instructions in this booklet - including all the WARNINGS and CAUTIONS
- before using this product. If there is any instruction which you do not understand.
DO NOT USE THE PRODUCT.
Safety Signs
WARNING
Indicates a potentially hazardous situation which, if not avoided, could result in death or
personal injury.
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in minor or
moderate injury to the user or users, or result in damage to the product or to property.
CAUTION
Warning laser radiation.
Embedded laser (diode) is Class 1M, <5mW. Do not view directly with optical magniers.
LASER RADIATION
DO NOT STARE INTO BEAM
CLASS 2 LASER PRODUCT
1mW MAX OUTPUT AT 635nm
IEC 60825-1:1994
IMPORTANT: Please note radiation safety warning, where this symbol appears.
NOTE
Indicates a potentially hazardous situation which, if not avoided, could result in damage or the loss of
data.
Equipment Operation
Use of this instrument in a manner not specied by Land Instruments International may be hazardous.
Read and understand the user documentation supplied before installing and operating the equipment.
Protective Clothing, Face and Eye Protection
It is possible that this equipment is to be installed on, or near to, machinery or equipment operating at
high temperatures and high pressures. Suitable protective clothing, along with face and eye protection
must be worn. Refer to the health and safety guidelines for the machinery/equipment before installing
this product. If in doubt, contact Land Instruments International.
Electrical Power Supply
Before working on the electrical connections, all of the electrical power lines to the equipment must
be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these
operating instructions. If in doubt, contact Land Instruments International.
Storage
The instrument should be stored in its packaging, in a dry sheltered area.
Unpacking
Check all packages for external signs of damage. Check the contents against the packing note.
Lifting Instructions
Where items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the
Technical Specication for weights. All lifting should be done as stated in local regulations.
IMPORTANT INFORMATION - PLEASE READ
Contact Us
UK - Droneld
Land Instruments International
Tel: +44 (0) 1246 417691
Email: [email protected]
Web: www.landinst.com
USA - Pittsburgh
AMETEK Land, Inc.
Tel: +1 412 826 4444
Web: www.ametek-land.com
For further details on all AMETEK Land ofces, distributors and representatives, please visit
our websites.
Return of Damaged Goods
IMPORTANT If any item has been damaged in transit, this should be reported to the carrier and to the
supplier immediately. Damage caused in transit is the responsibility of the carrier not the supplier.
DO NOT RETURN a damaged instrument to the sender as the carrier will not then consider a claim. Save
the packing with the damaged article for inspection by the carrier.
Return of Goods for Repair
If you need to return goods for repair please contact our Customer Service Department. They will be
able to advise you on the correct returns procedure.
Any item returned to Land Instruments International should be adequately packaged to prevent damage
during transit.
You must include a written report of the problem together with your own name and contact information,
address, telephone number, email address etc.
Design and Manufacturing Standards
The Quality Management System of Land Instruments International is
approved to BS EN ISO 9001 for the design, manufacture and on-site
servicing of combustion, environmental monitoring and non-contact
temperature measuring instrumentation.
Approvals apply in the USA
Operation of radio transmitters, telephones or other electrical/electronic devices in close proximity
to the equipment while the enclosure doors of the instrument or its peripherals are open, may cause
interference and possible failure where the radiated emissions exceed the EMC directive.
The protection provided by this product may be invalidated if alterations or additions are made to the
structural, electrical, mechanical or pneumatic parts of this system. Such changes may also invalidate the
standard terms of warranty.
Copyright
This manual is provided as an aid to owners of Land Instruments International’s products and contains
information proprietary to Land Instruments International. This manual may not, in whole or part, be
copied, or reproduced without the expressed written consent of Land Instruments International Ltd.
Copyright © 2016 Land Instruments International.

Blank

Contents
User Guide LSPHD Linescanner Head
Contents
Preface
1 Introduction 1
1.1 Nomenclature 2
1.2 Specications-LSPHD Series 3
2 Installation 8
2.1 Introduction 8
2.2 Changing the focus distance 9
2.3 Positioning The Scanner 10
2.4 Scan Widths and Target Size 11
2.5 Alignment and Scan Direction 14
2.6 Installation Precautions 15
3 Services 16
3.1 Electrical connections: Cable requirements 16
3.2 WaterRequirements(Water-CooledUnitsOnly) 17
3.3 AirRequirements(Air-Purged/CooledUnitsOnly) 18
4 Laser Alignment 19
4.1 UsingtheLaserAlignmentUnit 20
5 Emissivity-TypicalValues 21
5.1 SettingtheEmissivityvalue 21
5.2 LSPHD 1 Series 23
5.4 LSPHD 5 Series 23
5.3 LSPHD 2 Series 23
5.5 LSPHD 6 Series 24
5.6 LSPHD7Series 24
6 SettingtheScanFrequency 25
6.1 ChangingtheScanFrequencysettingusingtheinternalswitch 25
6.2 ScanFrequencyTable 26
6.3 ScanLineSeparation 27
6.4 DataStorageRequirements 27
6.5 SkewAngle 28
7 InternalAmbientTemperatureSensor 29
7.1 Introduction 29
8 Maintenance 30
8.1 Air Purge Services 30
8.2 Water And Air Cooling Services 30
8.3 PlateMountedBlowerUnit 30
9 Installation Accessories 31
10 SoftwareVersionInformation 32
10.1 Reading the WCA program version 32
10.2 ReadingtheLSP-HDScannerversioninformation 33
10.3 ReadingtheIOversioninformation 34

Preface
User GuideLSPHD Linescanner Head
Preface
Fig. A LSPHD Head
ThisUserGuideisapplicabletothefollowingLSPHD variants:
LSPHD 10 LSPHD 5FL LSPHD 60 LSPHD71
LSPHD 11 LSPHD 50 LSPHD 61
LSPHD 20 LSPHD 52 LSPHD 62
LSPHD 21
LSPHD 22
TheUserGuidecoverstheinstallation,operationandmaintenanceoftheLSPHD
Linescanning Thermometer.
TheinstructionsgiveninthisUserGuideguideshouldbeusedinconjunctionwiththe
documentslistedbelow:
• LSPHDLinescanningThermometerInstallationGuide
• LSPHDSystemMountings&AccessoriesInstallationGuide

Page 1
User Guide LSPHD Linescanner Head
12
34
5
Fig. 1 Typical LSPHD System Overview
LSPHD SYSTEM
LSPHD Linescanner Head
Power Supply Unit
Ethernet
Ethernet
PROCESS MONITORING (optional)
Landscan
for Windows
Conguration
Software
Power over Ethernet
I/O Processor
1 Introduction
The LANDSCAN LSPHD SYSTEM uses high speed infrared detectors and precision
electronics to provide accurate linescan temperature measurement at high scan rates.
Thesystemiscompact,yetrugged,providingreliableoperationinthemostsevere
industrial environments.
The LANDSCAN LSPHD SYSTEM comprises:
• LSPHD Linescanner Head
• LSPHD Ethernet Cable
• LANDSCANforWindowsCongurationSoftware
• LANDSCANWCASoftware(Optional)
• I/OModules(Optional)
Landscan
WCA
Control Inputs & Outputs
PROCESS CONTROL
Up to four copper Ethernet
(RJ45) connections and one
optical Ethernet connection may
be made to the scanner using
the LSPHD Power Supply Unit

Page 2
User GuideLSPHD Linescanner Head
System Name: LSP LSPSystem
High Denition: HD Fast scan rate and High resolution
Series: 6 OperatingWavelength
Temperature Range: 1 SeeSpecications
LSP HD 6
SystemName
Series Number
Temperature Range
1.1 Nomenclature
The LSP Linescanning Thermometer detail label can be found on the black scanner
bodysection.Thelabeldisplaysthescannerserialnumber,themodelnumberandthe
'CE'conformitymark.Themodelnumberindicatesspecicinformationabouttheunit,
including 'Series Number' and 'Temperature Range'.
Example:
1
HighDenition

Page 3
User Guide LSPHD Linescanner Head
1.2 Specications-LSPHD Series
1.2.1 Common specications
Ambient temperature range 5to60°C/41to140°F(inspecication)
5to70°C/41to158°F(operating)
Alignment Class 2 Laser indicating scan plane and extent
Laser scan rate 1 Hz
Internal Ambient Temperature Sensor
(IATS)
-10to90°C/14to194°F
Emissivityrange
(remoteorinternallyset)
0.20 to 1.00
Sealing IP65
EMC
LVD
EN61326ClassA[emissions&immunity]
EN61010-1
Vibration(Anyaxis) 2g[10to30Hz]
3g[30to300Hz]
1.2.2 LSPHD 1 series specications
LSPHD 10 LSPHD 11
Operatingwavelength 1µm 1µm
Temperature range 600 to 1400 °C
1112 to 2552 °F
700to1500°C
1292to2732°F
SystemAccuracy(1)
(5to95%ofrange)
±2°C/3.6°F
Focus distance (2) 1m/39.7intoinnity(continuously
adjustable)
FieldofView
(staticto95%ofradiance)
500:1(0.002radians)
Repeatability <0.5°C/0.9°F
Temperature Resolution (3)
< 5% of range
(noise) < 15% of range
< 30% of range
Above 50% of range
≤6°C/10.8°F
≤6°C/10.8°F
≤4°C/7.2°F
≤1°C/1.8°F
Scan Angle (4) 80°
Scan Rate Factoryset
Useradjustable
50 Hz
10 to 150 Hz in 1 Hz steps
Speed of response
(Tau-linearoutputonrisingedge)
≤1µs
AmbientStability
deltaT[indicated]/deltaT[ambient]
≤1°/10°
(1)Includeserrorcorrection
(2)Focusisfactorysetto2m
(3)4lineaveragingbelow15%ofrange
(4)Scananglemaybesetintherange40to80°bythePCsoftware

Page 4
User GuideLSPHD Linescanner Head
1.2.3 LSPHD 2 series specications
LSPHD 20 LSPHD 21 LSPHD 22
Operatingwavelength 2.2 µm 1.9 µm 1.9 µm
Temperature range 200 to 850 °C
392 to 1562 °F
300 to 1000 °C
572to1832°F
400 to 1200 °C
752to2192°F
SystemAccuracy(1)
(5to95%ofrange)
±2°C/3.6°F
Focus distance (2) 1m/39.7intoinnity(continuouslyadjustable)
FieldofView
(staticto95%ofradiance)
300:1
(0.0033
radians)
500:1
(0.002radians)
500:1
(0.002radians)
Repeatability <0.5°C/0.9°F
Temperature Resolution (3)
< 5% of range
(noise) < 15% of range
< 30% of range
Above 50% of range
≤6°C/10.8°F
≤6°C/10.8°F
≤2°C/3.6°F
≤1°C/1.8°F
≤6°C/10.8°F
≤6°C/10.8°F
≤4°C/7.2°F
≤1°C/1.8°F
≤6°C/10.8°F
≤6°C/10.8°F
≤4°C/7.2°F
≤1°C/1.8°F
Scan Angle (4) 80°
Scan Rate Factoryset
Useradjustable
50 Hz
10 to 150 Hz in 1 Hz steps
Speed of response
(Tau-linearoutputonrisingedge)
≤1.5µs ≤1µs ≤1µs
AmbientStability
deltaT[indicated]/deltaT[ambient]
≤1°/10°
(1)Includeserrorcorrection
(2)Focusisfactorysetto2m
(3)4lineaveragingbelow15%ofrange
(4)Scananglemaybesetintherange40to80°bythePCsoftware

Page 5
User Guide LSPHD Linescanner Head
1.2.4 LSPHD 5 series specications
LSPHD 5FL LSPHD 50* LSPHD 52*
Operatingwavelength 5 µm 5 µm 5 µm
Temperature range 150to750°C
302 to 1382 °F
150to750°C
302 to 1382 °F
500 to 1100 °C
932 to 2012 °F
SystemAccuracy(1)
(5to95%ofrange)
±2°C/3.6°F ±2°C/3.6°F ±3°C/6.4°F
Focus distance Fixedat1200mm/47in
Targetwidth Target distance
(staticto95%radiance) <1200mm/47in
Target distance
1200 to 3000mm
47to118in
12mm/0.5in
Fieldofviewof100:1(0.01radians)
Repeatability <0.5°C/0.9°F
Temperature Resolution (2)
Below15%ofrange
(noise) Above 15% of range
≤1.5°C/2.7°F
≤1°C/1.8°F
≤1.5°C/2.7°F
≤1°C/1.8°F
≤2.5°C/4.5°F
≤2.5°C/4.5°F
Scan Angle (3) 80°
Scan Rate Factoryset 20 Hz 50 Hz 50 Hz
Useradjustable 10 to 150Hz in 1Hz steps
Speed of response
(Tau-linearoutputonrisingedge)
≤5µs
AmbientStability
deltaT[indicated]/deltaT[ambient]
≤2°/10° ≤2°/10° ≤3°/10°
Settling Time (4) 2 Hours
*LSPHD 50 and LSPHD 52 as per LSPHD 5FLunlessotherwisestated
(1)Includeserrorcorrection.Appliesonly'inapplication'
(2)Nolineaveraging.4LAreducesnoisebyfactor2
(3)Scananglemaybesetintherange40to80°bythePCsoftware
(4)Foroptimumperformance,thethermoelectriccoolertakestimetostabilisethedetectorthermally.Thisprocesstakeslongerathigherambient
temperatures.

Page 6
User GuideLSPHD Linescanner Head
1.2.5 LSPHD 6 series specications
LSPHD 60 LSPHD 61* LSPHD 62* LSPHD 63*
Operating
Wavelength
3 to 5 µm 3 to 5 µm 3 to 4.2 µm
Temperature range 20 to 250 °C
68 to 482 °F
50 to 400 °C
122to752°F
100 to 600 °C
212 to 1112 °F
SystemAccuracy (1)
(5to95%ofrange)
±2°C/3.6°F
Focus distance Fixedat1200mm/47in
Targetwidth
(staticto95%radiance)
Targetdistance<1200mm/47in 12mm/0.5in
Targetdistance1200to3000mm/
47to118in Fieldofviewof100:1(0.01radians)
Repeatability <0.5°C/0.9°F
Temperature
Resolution (2)
Below15%ofrange ≤1.5°C/2.7°F ≤1.5°C/2.7°F ≤2.5°C/4.5°F ≤2°C/3.6°F
(noise)Above15%ofrange ≤1°C/1.8°F ≤1°C/1.8°F ≤2.5°C/4.5°F
Scan Angle (3) 80 °
Scan Rate 10 to 150Hz in 1 Hz steps
Factoryset 20 Hz 50 Hz 50 Hz 50 Hz
Useradjustable
Speed of response
(Tau-linearoutputonrisingedge)
≤5µs ≤10µs
AmbientStability
deltaT[indicated]/deltaT[ambient]
<1°/10° <1°/10° <2°/10° <1°/10°
Settling Time (4) 2 Hours
*LSPHD 61, LSPHD 62, and LSPHD 63 as per LSPHD 60unlessotherwisestated
(1)Includeserrorcorrection
(2)Nolineaveraging.4LAreducesnoisebyfactor2
(3)Scananglemaybesetintherange40to80°bythePCsoftware
(4)Foroptimumperformance,thethermoelectriccoolertakestimetostabilisethedetectorthermally.Thisprocesstakeslongerathigherambient
temperatures.

Page 7
User Guide LSPHD Linescanner Head
1.2.6 LSPHD 7 series specications
LSPHD 71
Operatingwavelength 3.4 µm
Temperature range 50 to 350 °C
122 to 662 °F
SystemAccuracy(1)
(5to95%ofrange)
±2°C/3.6°F
Focus distance Fixedat1200mm/47in
Targetwidth Target distance
(staticto95%radiance) <1200mm/47in
Target distance
1200 to 3000 mm
47to118in
12mm/0.5in
Fieldofviewof100:1(0.01radians)
Repeatability <0.5°C/0.9°F
Temperature Resolution (2)
Below15%ofrange
(noise) Above 15% of range
≤4°C/7.2°F
≤2°C/3.6°F
Scan Angle (3) 80°
Scan Rate Factoryset 50 Hz
Useradjustable 10 to 150 Hz in 1 Hz steps
Speed of response
(Tau-linearoutputonrisingedge)
≤10µs
AmbientStability
deltaT[indicated]/deltaT[ambient]
≤1°/10°
Settling Time (4) 2 Hours
(1)Includeserrorcorrection
(2)Nolineaveraging.4LAreducesnoisebyfactor2
(3)Scananglemaybesetintherange40to80°bythePCsoftware
(4)Foroptimumperformance,thethermoelectriccoolertakestimetostabilisethedetectorthermally.Thisprocesstakeslongerathigherambient
temperatures.

Page 8
User GuideLSPHD Linescanner Head
2 Installation
2.1 Introduction
TheequipmentspeciedinthisUserGuidemustbeoperated,maintainedandserviced
bysuitablytrainedpersonnel,capableofcarefullyfollowingtheprocedures,warnings
andguidelinesgiven.ItisadvisablethatallUserGuidesandInstallationGuides
associatedwithLSPHDSystemequipmentshouldbereadthoroughlyandbekeptreadily
available.
CAUTION
Ensure that the LSPHDHeadismountedsecurely,inalocationwheretheeffects
of vibration, dust and fumes are kept to a minimum. Ensure that the scanning
sight path is not obscured.
The LSPHD10,20&21haveafocusdistancefactorysetat2000mm/79in.
RefertoSection2.2fordetailsofhowtochangethefocusdistance.
Chooseamountinglocationwhichminimisesexternalheatingofthescanner.
Iftheambienttemperatureoftheunitexceedsthespeciedmaximum,a
coolingsystem(availablefromAMETEKLand)mustbetted.
Iftheinstrumentistobeusedinenvironmentswheregasesresultingfrom
combustionarepresent,contactAMETEKLandsothattheapplicationcanbe
fullyassessedbeforeyouinstallthescanner.
Fig.2FocusadjustmentfortheLSPHD 1, LSPHD 11, LSPHD 20, LSPHD 21 and LSPHD 22
Focusing ring
Clamping plate

Page 9
User Guide LSPHD Linescanner Head
2.2 Changing the focus distance
(LSPHD 10,LSPHD 11,LSPHD 20,LSPHD 21&LSPHD 22 only)
The focus distance of the LSPHD 10, LSPHD 11, LSPHD 20, LSPHD 21 and LSPHD 22,
linescannerheadisfactorysetto2m/79in.Thefocusdistancecanbeuser-adjusted
intherange1m/40intoinnity.
1) Removethefrontcoverofthescanner(thepanelnearestthelaserswitch).
RefertoFig.2.ThefocuscanbeadjustedbyturningtheFocusing ring,which
isheldinplacebytheClamping plate.
2) RemovethetopscrewfromtheClamping plate to free the Focusing ring.
3) Inordertofocustheinstrumentaccurately,thefocusingmechanismmustrstbe
'zeroed'. Rotate the Focusing ringinthedirectionofthe'Focuslong'arrow(i.e.
totheright)untilitiswoundinfully.
4) Refertothetablebelow.RotatetheFocusing ring to the left for the number of
turnsrequiredtomatchthefocusdistanceyouwanttoset.
Thetablebelowgivesthefocusdistancesthatcanbeachievedbyturningthe
focusingring.Eachscrewholeonthefocusingringisat0.05ofaturn.
Focus distance
m / in
LSPHD 10
LSPHD 11
LSPHD 20 LSPHD 21
LSPHD 22
Numberofturnsfromfullywoundin
1.0/40 7.70 8.95 8.45
1.2/4.7 6.80 8.00 7.50
1.5/59 5.85 7.05 6.60
2.0/79 4.95 6.15 5.65
3.0/118 4.05 5.20 4.70
4.0/157 3.60 4.70 4.25
5.0/197 3.35 4.45 4.00
6.0/236 3.15 4.25 3.80
7.0/276 3.00 4.15 3.70
8.0/315 2.95 4.05 3.60
10.0/394 2.80 3.90 3.45
Innity 2.30 3.40 2.95
5) Toadjustthefocustoadistanceotherthanthatquotedintheabovetable,rotate
the focusing ring to the required intermediate position.
6) Whenyouhavefocusedthescannertotherequireddistance,therewillbeahole
in the Focusing ring thatalignswiththeholeintheClamping plate(youmay
needtoadjusttheFocusingringslightly).Insertandtightenthescrewthrough
the Clamping plate to secure the Focusing ring in position.
7) Replacethecoverpanel.
Thescannerisnowfocused.

Page 10
User GuideLSPHD Linescanner Head
Fig.3Scanwidthandtargetsizedimensions
40°
80°
TargetPath(TP)
Target
Distance
(TD)
ScanWidth(SW)
x axis
yaxis
x axis
yaxis
TP = TD
cos 40
Targetspotbecomesmoreellipticalwith
increased distance from scan centre line.
Seetargetsizetablesforxandyaxis
sizes at ±40°
2.3 Positioning The Scanner
Before installing the scanner, refer to the cautions given in Section 2.1.
The scanner must be installed so that the target plane is parallel to the scanner
window.
RefertoFig.3(below).Thewidthofthescan(SW)isproportionaltothedistance
(TD)betweenthescannerandthetargetplane.Chooseatargetdistancethatwillgive
ascanlinethatjustexceedsthewidthoftheproduct,asthisgivesmaximumspatial
resolution.

Page 11
User Guide LSPHD Linescanner Head
2.4 Scan Widths and Target Size
2.4.1 Scan Width
ThefullscananglefortheLSPScanneris80°,andtheScanWidth(SW)canbe
calculated as:
SW=(TD+35)x1.678mm
or
SW=(TD+1.38)x1.678inches
Where‘TD’istheperpendicularTargetDistancebetweentheLSPwindowandthe
target plane.
Atthescanangleextremes(±40°)thelengthoftheTargetPath(TP)betweenthe
scanner and the target is:
TP=(TD+35)/0.766mm
or
TP=(TD+1.38)/0.766inches
Note
Theactivescananglemaybesetintherange40to80°.SeetheLandscan
CongurationorWCAsoftwareguidesforfurtherinformation.
2.4.2 Target Size
ThecalculatedTargetSizeisstrictlyonlycircularatascanangleofzerodegrees,i.e.
perpendicular to the instrument.
Atallotherscanangles,thetargetspotwillbeanellipse.Theellipticaltargetsizeis
showninthefollowingtablesforthescanangleextremes(±40°).

Page 12
User GuideLSPHD Linescanner Head
2.4.3 LSPHD 1 & 2 Target size table
The nominal Target Size (TS)isgivenby:
TS = TD / FoV
Where‘FoV’istheinstrument’sFieldofView
• LSPHD 10, LSPHD 11, LSPHD21&LSPHD 22 FoV=500:1
• LSPHD20 FoV=300:1
TD SW TP LSPHD 10, LSPHD 11, LSPHD 21 & LSPHD 22
Target Size
LSPHD 20-TargetSize
Target Distance Scan Width Target Path 0° Scan angle ±40°(xaxis) ±40°(yaxis) 0° Scan angle ±40°(xaxis) ±40°(yaxis)
mm in mm in mm in mm in mm in mm in mm in mm in mm in
1000 39.4 1737 68.4 1305 51.4 2.0 0.08 7.8 0.31 6.0 0.24 3.3 0.13 10.0 0.39 7.7 0.30
1250 49.2 2156 84.9 1632 64.3 2.5 0.10 8.7 0.34 6.7 0.26 4.2 0.17 11.5 0.45 8.8 0.35
1500 59.1 2576 101.4 1958 77.1 3.0 0.12 9.6 0.38 7.4 0.29 5.0 0.20 13.0 0.51 10.0 0.39
1750 68.9 2996 118.0 2284 89.9 3.5 0.14 10.5 0.41 8.0 0.31 5.8 0.23 14.4 0.57 11.0 0.43
2000 78.7 3415 134.4 2611 102.8 4.0 0.16 11.4 0.45 8.7 0.34 6.7 0.26 16.0 0.63 12.2 0.48
2500 98.4 4254 167.5 3264 128.5 5.0 0.20 13.1 0.52 10.0 0.39 8.3 0.33 18.7 0.74 14.4 0.57
3000 118.1 5093 200.5 3916 154.2 6.0 0.24 14.8 0.58 11.4 0.45 10.0 0.39 21.6 0.85 16.5 0.65
3600 141.7 6100 240.2 4700 185.0 7.2 0.28 17.0 0.67 13.0 0.51 12.0 0.47 25.0 0.98 19.2 0.76
4000 157.5 6772 266.6 5222 205.6 8.0 0.31 18.3 0.72 14.0 0.55
6000 236.2 10128 398.7 7832 308.3 12.0 0.47 25.1 0.99 19.2 0.76
8000 315.0 13484 530.9 10443 410.7
Notes:
1) Theindicatedtargetsizesassumethatthefocusdistanceoftheinstrumentequalsthetargetdistance.
2) Iftheinstrumentisfocusedatadistancebetweenthe‘targetdistance’andthe‘targetpath’,thetargetsizeatthescan
angleextremeswillreduceattheexpenseofanincreaseatzerodegreesscanangle.

Page 13
User Guide LSPHD Linescanner Head
2.4.4 LSPHD 5, 6 & 7 Target size table
The nominal Target Size (TS)isgivenby:
• At target distances less than1200mm/47in:
TS = 12mm / 0.47in
or
• At target distances greater than1200mm/47in:
TS = TD / FoV
• LSPHD 5, LSPHD6&LSPHD7 FoV=100:1
TD SW TP LSPHD 5, LSPHD 6 & LSPHD 7
Target Size
Target Distance Scan Width Target Path 0° Scan angle ±40°(xaxis) ±40°(yaxis)
mm in mm in mm in mm in mm in mm in
200 7.9 394 15.5 261 10.3 12.0 0.47 15.7 0.62 12.0 0.47
500 19.7 898 35.4 653 25.7 12.0 0.47 15.7 0.62 12.0 0.47
750 29.5 1317 51.9 979 38.5 12.0 0.47 15.7 0.62 12.0 0.47
1000 39.4 1737 68.4 1305 51.4 12.0 0.47 17.0 0.67 13.1 0.52
1200 47.2 2073 81.6 1566 61.7 12.0 0.47 20.4 0.80 15.7 0.62
1500 59.1 2576 101.4 1958 77.1 15.0 0.59 25.5 1.00 19.5 0.77
2000 78.7 3415 134.5 2611 102.8 20.0 0.79 34.0 1.34 26.0 1.02
3000 118.1 5093 200.5 3916 154.2 30.0 1.18 51.0 2.00 39.0 1.54
4000 157.5 6772 266.6 5222 205.6 40.0 1.57 68.0 2.68 52.0 2.05

Page 14
User GuideLSPHD Linescanner Head
Scan Direction Arrowhead
Scan Direction
Product Pass Direction
Product
LSP Scanner
2.5 Alignment and Scan Direction
EachAMETEKLandLSP
HD
Head has an integral laser unit that can be used to help align
thescannerduringinstallation.Forinstructionsonhowtousethelaseralignmentunit,
refertoSection4.0ofthisUserGuide.
Tointerpretthescanproleproducedbythescanner,thescandirectionoftheunitmust
beknown.Thescandirectionisindicatedbyanembossedarrowheadcastintothetop
ofthescannercasing(SeeFig.4overleaf).
Thedesignofthescannermountingsensuresthatthescannerwillalwaysbereplacedin
the correct orientation, should the scanner be removed from the application for servicing
etc.
Fig. 4 Scan Direction
This manual suits for next models
12