ASD 220 Slide Dryer
Service Manual
1、
The Service Manual of the instrument is compiled according to its structure and
characteristic.
2、
Power supply of the instrument is AC110V/220V.
Slide Drier can heat up and control temperature.
3、
A、The screen has no display when the machine turns on.
1): The instrument is not plugged in or the fuse is blow out, please connect the
power supply again or replace the fuse.
2): power module is damaged
Maintenance:Use Philip screwdriver to loosen
the six screw in the base-plate,and take
down the base-plate,and there will be a
power source module on the main board in
the instrument as it shows on the picture.
And take down the power source module
from the main board by hand,if it is difficult
to take down the power source module with
hand,you can use the flat apparatus as flat
Heating film Figure sensor
Power source module Solid state relay