TEC 2602 Slide Dryer AMOSTEC2602SM20140309
---1---
TEC 2602 Slide Drier
Service Manual
1、
The Service Manual of the instrument is compiled according to its structure and characteristic.
2、
Power supply of the instrument is AC110V/220V.
Slide Drier can heat up and control temperature.
3、
A、The screen has no display when the machine turns on.
1):The instrument is not plugged in or the fuse is blow out, please connect
the power supply again or replace the fuse.
2):power module is damaged
Maintenance:Use Philip screwdriver to
loosen the six screw in the base-plate,
and take down the base-plate,and there
will be a power source module on the main
board in the instrument as it shows on
the picture. And take down the power
source module from the main board by
hand,if it is difficult to take down the
power source module with hand,you can
use the flat apparatus as flat tip
Heating film Figure sensor
IC Power source module Solid state relay
General Provisions
Confi
uration
Troubleshootin