Contraindications—Warnings—Precautions
THIS INSTRUMENT OPERATES ON 120 VOLTS AC, 60 Hz. (unless otherwise indicated on the
unit) AND MUST BE PROPERLY GROUNDED FOR SAFETY. The three wire power cord with
"hospital grade" plug should be connected to a GROUNDEDAC wall receptacle. It is the personal
responsibility and obligation of the user to insure that this instrument is properly connected to
the AC POWER source before use.
Important
OPTIONAL FEATURE (1.0/3.3) AMREX Ultrasound Intensity Reset Circuit: The Ultrasound
modality incorporates a unique reset circuit feature as a part of the intensity control. The
ultrasound output will be disabled in the event that:
•The Ultrasound Intensity control is not set to the 0position at power on
•The ac power is interrupted
•The treatment period has been completed and the bell has sounded
•The the Ultrasound Frequency switch setting is changed during treatment
The Ultrasound Intensity control must be returned to the 0position before ultrasound output
can be activated.
Ultrasound—Contraindications
Ultrasound should not be used in the following areas: transcerebrally; near the heart or reproductive
organs; over viscera, eyes, ears, the spinal column, malignancies, the joint capsule in arthritic
conditions either acute or subacute, or over total joint replacements. Ultrasound should not be used
in cases of: cardiac pacemakers, pregnancy, implants, malignant or benign tumors, multiple sclerosis,
arteriosclerosisorweakenedbloodvessels,hemophilia, thrombosisandthrombophlebitiseither acute
or subacute. Neither should ultrasound be used over bone in growing children, where the skin
suffers from any sensory impairment, nor in the area of the carotid sinus nerves, visceral plexus or
large autonomous ganglion.
Ultrasound—Precautions
Excessive doses of ultrasound can cause damage to tissue. Periosteal pain is an indication of excess
intensity. Should this occur, reduce the power and/or move the transducer more quickly over the
area being treated.
Any bleeding tendency is increased by heating because of the increase in blood flow and
vascularity of
the heated tissues. Care should be used in treating patients who have bleeding disorders.
Warning–Risk of burns and fire. DO NOT use near conductive
material such as metal bed parts or innerspring mattresses. Renew
electrode cables upon evidence of deterioration. Use of controls,
adjustment,or performance of procedures,other than those specified
herein, may result in hazardous exposure to ultrasonic energy.