Amsino iReceptal 3 Operating manual

SL 5107 Rev. D
1
iReceptal Digital Surgical Suction System
Site Preparation, Installation, and Maintenance Guide
100~240 VAC Docking Station
REF iRD301
Instructions For Use RxONLY

SL 5107 Rev. D
2
Contents
Introduction.........................................................................................................................................3
Conventions..............................................................................................................................3
Contact Information..................................................................................................................3
Indications For Use...........................................................................................................................3
Contraindications For Use................................................................................................................3
Intended patient.................................................................................................................................3
For Use With......................................................................................................................................4
Accessories........................................................................................................................................4
Description..........................................................................................................................................4
User/Patient Safety...........................................................................................................................5
General......................................................................................................................................5
Electrical Safety........................................................................................................................5
Environmental/Biological.........................................................................................................6
Features..............................................................................................................................................7
Symbols..............................................................................................................................................8
Instructions.........................................................................................................................................9
To Install the Docker................................................................................................................9
1. To Install the Docker and Connect Utilities...........................................................10
2. To Apply Power, Water, and Detergent ................................................................11
To Test the Docker.................................................................................................................11
1. To Prepare the Rover...............................................................................................12
2. To Prepare the Docker.............................................................................................12
3. To Perform a Wash Cycle.......................................................................................13
To Shut Down the Docker.....................................................................................................14
Inspection and Cleaning Maintenance.........................................................................................14
Add the Detergent..................................................................................................................15
To Wipe Down the Docker....................................................................................................16
Storage and Handling.....................................................................................................................16
Disposal/Recycle.............................................................................................................................17
Troubleshooting ...............................................................................................................................17
Error Messages ...............................................................................................................................18
Specifications...................................................................................................................................18
Electromagnetic Compatibility.......................................................................................................20

SL 5107 Rev. D
3
Introduction
This Instructions For Use manual is the most comprehensive source of information for your product. Keep and
consult this reference manual during the life of the product.
The iReceptal Digital Surgical Suction System Manifold is non-patient contacting, non-sterile, disposable device
that provide a fluid path from the suction tubing lines of a iReceptal Digital Surgical Suction System to the
receiving collection canisters of this system.
Conventions
The following conventions are used in this manual:
WARNING: A warning highlights a safety-related issue. ALWAYS comply with this information to prevent
patient or healthcare staff injury.
CAUTION: A caution highlights a product reliability issue. ALWAYS comply with this information to prevent
product damage.
NOTE: A note supplements and/or clarifies procedural information.
Contact Information
For additional information, including safety information, or in-service training, contact your AMSINO sales
representative or call AMSINO iReceptal Customer Service at XXXXXX.
Indications For Use
The iReceptal Digital Surgical Suction System is intended to be used in the operating room, pathology, surgical
centers, and doctor’s offices to collect and dispose of surgical fluid waste as well as collect smoke generated
from electrocautery or laser devices.
Contraindications For Use
The iReceptal Digital Surgical Suction System is contraindicated against:
▪Connection directly to chest tubes.
▪Connection to closed wound drainage systems.
Intended patient
The general patient population is anyone excluding newborn subject to general surgery or diagnostic procedures
where the device can be used.

SL 5107 Rev. D
4
For Use With
The following components are required to be used with the equipment described in this manual to create a
complete system:
Description
REF
iReceptal 3 Rover
iRR301 100~120V
iRR302220~240V
iReceptal 3 Docking Detergent
iRC003
Accessories
This section describes system components that may be ordered to replace original equipment that is damaged,
worn, or must be replaced. This section may also contain optional components used with the system.
The following AMSINO-approved accessories are sold separately:
Description
REF
iReceptal 3 Docking Detergent
iRC003
Power Cord (US)
iRA105-01
Power Cord (UK)
iRA106-01
Power Cord (EU)
iRA107-01
NOTE: For a complete list of accessory information, contact your AMSINO sales representative or call AMSINO
iReceptal Digital Surgical Suction System Customer Service.
Description
The iReceptal Docking Station (docker) is a component of the iReceptal Digital Surgical Suction System. The
Stryker Rover (rover), another component of the system, is a mobile unit used to suction and collect fluid waste
and surgical smoke from a surgical site in an operating room.
After collection, the rover is relocated and mated to the docker. Once the rover is connected to the docker, the
emptying of the fluid waste and cleaning of the canisters occurs automatically (Figure 1).
The rover chambers are rinsed with clean water and iReceptal Docking Detergent REF iRC003 to clean the
chambers of any residual fluid waste (Figure 2).

SL 5107 Rev. D
5
Figure 1 To Empty the Canister Figure 2 To Clean the Canister
User/Patient Safety
⚠WARNINGS:
General
•Before using any system component, or any component compatible with this system, read and understand
the instructions. Pay particular attention to WARNING information. Become familiar with the system
components prior to use.
•Only trained and experienced healthcare professionals may use this equipment.
•Healthcare professionals should be thoroughly familiar with the instructions for use, handling
characteristics, and the indicated and intended uses of this equipment. Contact your AMSINO sales
representative or AMSINO iReceptal Digital Surgical Suction System Customer Service for in-service
training.
•DO NOT disassemble, modify, service, or repair any system component or accessory, unless otherwise
specified. Call AMSINO iReceptal Digital Surgical Suction System Customer Service.
•Upon initial receipt and before each use, inspect each component for damage. DO NOT use any
equipment if damage is apparent or the inspection criteria are not met. See the Inspection and
Maintenance section for inspection criteria.
•ALWAYS operate the equipment within the specified environmental condition values. See the
Specifications section.
•MANUFACTURER will provide circuit diagrams, component part lists, descriptions, calibration instructions
to assist to SERVICE PERSONNEL in parts repair.
•Do not use iReceptal Digital Surgical Suction System in the presence of magnetic resonance imaging
(MRI) devices.
Electrical Safety
•Use only AMSINO-approved system components and accessories, unless otherwise specified. Using other
electronic components and accessories may result in increased electromagnetic emissions or decreased
electromagnetic immunity of the system.
•Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical
equipment like this system. Install and place this system into service according to the EMC information
contained in this manual. See the Specifications section. Portable and mobile radio frequency (RF)
communications equipment can affect the function of this system.
•ELECTRICAL SHOCK HAZARD-ALWAYS connect this equipment to a hospital-grade, facility power
receptacle with protective earth (ground). Failure to comply may cause electrical shock and result in patient

SL 5107 Rev. D
6
or healthcare staff injury.
•Please take attention that changes or modification not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
•This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference that may cause undesired
operation.
•This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to the
following two conditions:
(1) this device may not cause interference, and
(2) this device must accept any interference, including interference that may cause undesired operation of
the device.
•Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radioexempts de
licence. L'exploitation est autorisée aux deux conditions suivantes :
•(1) l'appareil ne doit pas produire de brouillage, et
•(2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d'en compromettre le fonctionnement.
•This equipment complies with FCC/IC RSS-102 radiation exposure limits set forth for an uncontrolled
environment. This equipment should be installed and operated with minimum distance 20cm between the
radiator & your body.
•ce matériel est conforme aux limites de dose d'exposition aux rayonnements, FCC / CNR-102 énoncée dans
un autre environnement.cette eqipment devrait être installéet exploitéavec distance minimale de 20 entre
le radiateur et votre corps.
•Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type and
maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio
interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically
radiated power (e.i.r.p.) is not more than that necessary for successful communication.
•Conformément àla réglementation d'Industrie Canada, le présent émetteur radio peutfonctionner avec une
antenne d'un type et d'un gain maximal (ou inférieur) approuvépour l'émetteur par Industrie Canada. Dans
le but de réduire les risques de brouillage radioélectrique àl'intention des autres utilisateurs, il faut choisir le
type d'antenne et son gain de sorte que la puissance isotrope rayonnée équivalente (p.i.r.e.) ne dépasse
pas l'intensiténécessaire àl'établissement d'une communication satisfaisante.
Environmental/Biological
•FIRE HAZARD –DO NOT use this equipment in areas in which flammable anesthetics or flammable agents
are mixed with air, oxygen or nitrous oxide. Failure to comply may cause a fire and result in burn injury or
property damage.
•BLOODBORNE PATHOGEN HAZARD
•The Bloodborne Pathogens Standard provided by the United States Occupational Safety and Health
Association (US OSHA 29 CFR 1910.1030) requires those with employees having occupational
exposure to potentially infectious materials to establish a written Exposure Control Plan. The
Exposure Control Plan is designed to eliminate or minimize employee exposure through use of
personal protective equipment (PPE), appropriate vaccinations (e.g. hepatitis B), and other control
measures.

SL 5107 Rev. D
7
•ALWAYS wear PPE when operating or handling this equipment.
•ALWAYS follow local regulations regarding proper handling and disposal of biohazard waste.
Failure to comply may cause infection and result in healthcare staff injury.
•CONTAMINATION HAZARD-ALWAYS follow local regulations for safe handling, recycling, and disposal of
biohazard fluid waste and equipment. See Disposal/Recycle section. Failure to comply may cause
environmental contamination and result in injury.
Features
Figure 3 Docking Station Front view Figure 4 Docking Station Back view
A
Magnet–Provide for the automatic connection of the rover to the docker.
B
Infrared Communication Port –Allow infrared data transfer between the docker and rover. Data
transfer is necessary during the docking procedure.
C
Detergent canister lid –Allows add detergent into the canister
D
Detergent canister- Stores the detergent
E
Fluid Connectors (two) –One connector allows fresh water to enter the rover. The other connector
provides for the disposal of waste water from the rover.
F
Guides (two) –Facilitate the alignment of the rover to the docker.
G
Waste Outlet Port –Allows for the disposal of fluid waste from the rover when the rover is connected
to the docker.
H
Water Inlet Port –Allows fresh water to enter the rover when the rover is connected to the docker.
I
Fixing Casters (four)--Allows adjust the height of each docking station to align with the rover.
Increase caster height by turning the dial counterclockwise. Decrease caster height by turning the
dial clockwise.
J
Power Cord –Allows for the connection of facility electrical power to the docker. Power cord
configurations may vary. See the Accessories section for options.
K
Water Inlet Hose –Allows fresh water to flow from the facility water source to the docker water inlet
port.
L
Waste Outlet Hose –Allows fluid waste to flow from the docker waste outlet port to the drain of the
facility waste disposal system.

SL 5107 Rev. D
8
M
Power Cord Receptacle –Allows for the connection of facility power using the docker power cord.
O
Power Switch –Allows for the application or removal of facility power.
N
Specification Label –
Symbols
The following symbols appear on the device and/or its labeling. For more information, see “Technical
specifications”
Symbol
Description
ON (POWER)
OFF (POWER)
ALTERNATING CURRENT (AC)
General warning
Consult instructions for use
Follow instructions for use
This symbol is located near the protective ground locations
on this device
DETERGENTADD PORT
WASTE OUTLET PORT
WATER INLET PORT

SL 5107 Rev. D
9
Instructions
To Install the Docker
⚠WARNING:HEAVY EQUIPMENT –ALWAYS have more than one person unpack and move this
equipment from the shipping pallet. See the Specifications section for rover weight. Failure to comply may result
in personal injury.
Figure 5 Docker Lift Points
CAUTIONS:
•Make sure the pressure and temperature values of the facility water supply are within the required specified
ranges. See the Specifications section.
•If mounting the docker to a wall, make sure no gap exists between the docker mounting bracket and the
wall. ALWAYS align the mounting hardware (not supplied) with the wall studs to make sure the docker is
mounted to the wall securely. Failure to comply may cause inadvertent docker movement and result in wall
or product damage.
•ALWAYS use the correct power cord. Configurations may vary. See the Accessories section for power cord
options
NOTES:
•Only individuals trained and experienced in the maintenance of reusable medical devices should install,
inspect, and test this equipment.
•The docker is installed in a utility closet or disposal area with access to electrical power, a water supply,
and a fluid waste drain. The healthcare facility is responsible for the preparation of the installation site and
the availability of utilities. See the Specifications section for electrical power, water, and drainage
requirements. Make sure the installation area meets utility and space requirements.
•If the installation site does not meet local ventilation requirements, obtain and install a ventilation device
that will meet the necessary local requirements.
•The docker is equipped with an internal backflow prevention device. See the Specifications section for
details. Refer to local plumbing codes to determine whether an external backflow prevention device is also

SL 5107 Rev. D
10
required.
•Make sure the plumbing configuration is NOT susceptible to water hammer conditions.
•Make sure the waste outlet hose is connected properly to minimize the escape of noxious fumes and
odors.
Figure 6 Minimum Floor Space Requirements
A
Left-Side Clearance
No minimum requirement
B
Right-Side Clearance
15 cm
C
Front Clearance
120cm
1. To Install the Docker and Connect Utilities
a. Place the docker on the floor and against a wall with access to electrical power, water, and a fluid waste
disposal drain.
b. Recommended: Install mounting screws (not supplied) through the mounting bracket of the docker and secure
the docker to the wall.
NOTE: If allowed by current local building and electrical codes, mounting the docker to the wall is strongly
recommended. The docker is not to be installed in a patient environment.
c. Connect the water inlet hose between the water inlet port of the docker and the facility water supply.
d. Connect the waste outlet hose to the waste outlet port of the docker and the drain emptying into the fluid
waste disposal system.

SL 5107 Rev. D
11
Figure 7 To Connect the Waste Outlet Hose
e. Connect the power cord between the electrical receptacle of the docker and the facility electrical power
source.
2. To Apply Power, Water, and Detergent
a. Press the power switch ON. Make sure the power switch illuminates.
b. Open the facility water valve to allow water to flow to the docker. Inspect the water supply connections for any
leaks. Repair any plumbing to stop leakage if necessary.
c. Unscrew the lid of the detergent canister and pour the detergent into the detergent canister. see the
instructions for use supplied with the Docking Detergent REF iRC003. See the Accessories section
Figure 8 To Apply Power, Water, and Detergent
To Test the Docker
⚠WARNING:ALWAYS keep hands out and away from the mating surfaces of the rover and docker during
the docking procedure to avoid a pinch point hazard.
CAUTIONS:

SL 5107 Rev. D
12
•DO NOT put any objects, including hoses, tubes, towels or detergent bottles, on or over the docker
enclosure. Failure to comply may cause corrosion.
•DO NOT allow fluid of any kind to spill directly onto the exterior surface of the electrically-powered docker.
Failure to comply may cause corrosion or product failure.
NOTES:
•Only individuals trained and experienced in the maintenance of reusable medical devices should install,
inspect, and test this equipment.
•Make sure the rover is operating properly. See the instructions for use supplied with the rover.
•While the rover is docked, DO NOT lock the rover casters.
1. To Prepare the Rover
a. Connect the rover to facility electrical power using the power cord.
b. Push the power switch to the ON position.
c. Read the WARNING message on the user interface, then tap the OK key to access the CONTROL screen.
d. Install a disposable manifold into manifold port.
e. Attach a suction tube to one manifold port on manifold. Make sure all the other manifold ports are capped.
f. Place the attached suction tube into a sink filled with water.
g. Push the rover SUCTION button to start the vacuum pump.
h. Adjust the SUCTION SETTING to initiate suction and transfer about two liters of water into the canister.
i. After the canister is filled with two liters of water, push the rover power switch to the OFF position.
j. Disconnect the rover from facility electrical power. Wrap the power cord around the cord bracket.
k. Remove the disposable manifold and suction tube.
Figure 9 To Prepare the Rover
2. To Prepare the Docker
a. Make sure the power switch is in the ON position and illuminated.

SL 5107 Rev. D
13
b. Make sure the canister of Detergent has enough detergent to perform a wash cycle.
Figure 10 To Prepare the Docker
3. To Perform a Wash Cycle
a.Push the rover toward the docker and between the guides until the rover and docker attach automatically.
b. As the rover and docker attached together, the Wash Cycle View will display automatically and “Docking in
place” voice message will be heard. From the Wash Cycle View select a cycle by tapping the related
BUTTON. For the wash cycle options, refer to the wash cycle options table.
Figure 11 To perform a wash cycle
NOTES:
•During the first rover docking procedure, the information on the user interface display may appear
inconsistent. The procedure may also take a few more minutes than specified in the Wash Cycle Options
table. Both conditions are normal and temporary.
•If a specific wash cycle is not selected within ten seconds, the “Wash” cycle will be performed
automatically. See the Wash Cycle Options table.

SL 5107 Rev. D
14
•To exit the current wash cycle, push the “Stop Suction/Stop Docking”button to terminate the current a
wash cycle and the rover detaches from the docker automatically. Pull the rover away from the docker. And
dock again, select the needed wash cycle.
Figure 12 To cease the current wash Cycle
Wash Cycle Options
CYCLES
DESCRIPTION
TIME (approx.)
Wash
Cycle drains the contents, applies detergent to the
interior walls of both chambers, and rinses the
detergent with water.
6 minutes
Quick Drain
Cycle drains the contents of both chambers.
3minutes
Extended Wash
Cycle drains the contents, applies detergent to the
interior walls of both chambers and rinses the
detergent with water. Intermittent periods of soaking
occur during the cycle.
50 minutes
c. During the wash cycle, inspect all the docker plumbing and connections for any leakage. If leakage occurs,
repair as required.
d. After the cycle is complete, the rover detaches from the docker automatically. Pull the rover away from the
docker.
NOTE: After successfully testing, the docker is ready for use.
To Shut Down the Docker
The docker does not need to be shut down between uses. To remove facility power from the docker, push the
power switch to the OFF position.
Inspection and Cleaning Maintenance
WARNINGS:
•Upon initial receipt and before each use, inspect each component for damage. DO NOT use any
equipment if damage is apparent or the inspection criteria are not met.

SL 5107 Rev. D
15
•DO NOT disassemble, modify, service, or repair any system component or accessory, unless otherwise
specified. Call AMSINO iReceptal Customer Service.
NOTES:
•Only individuals trained and experienced in the maintenance of reusable medical devices should install,
inspect, and test this equipment.
•For service, contact your AMSINO sales representative or call AMSINO iReceptal Customer Service.
Outside the US, contact your nearest AMSINO subsidiary.
•Maintenance documentation for this equipment is available upon request to AMSINO-authorized service
personnel only.
INTERVAL
INSPECTION CRITERIA
ACTION
Before initial use
Make sure the equipment has
been tested before first use.
See the To Test the Docker
section.
Before each use and after each
cleaning
Check equipment for damage or
missing components.
If damage is apparent, replace the
equipment.
Check power cord for cuts.
Check power cord receptacle for
bent pins or bent contacts.
Six months
Check the plumbing connections
and hoses for leaks.
Repair any plumbing to stop
leakage as required. Replace
leaking hoses as required.
As required
Check the two infrared
communication ports for any
obstructions.
Remove any obstruction covering
the infrared communication ports,
including hoses, tubes, and
towels.
Check the level of the detergent in
the detergent canister.
Add the detergent into the
detergent canister as required.
NOTE: If any component must be discarded, see the Disposal/ Recycle section.
Add the Detergent
WARNINGS:
•The iReceptal Detergent REF iRC003 is an irritant to eyes and skin. ALWAYS wear protective gloves and
eye protection to avoid contact with skin and eyes.
•In case of contact with eyes, rinse eyes immediately with plenty of water. See the Material Safety Data
Sheet (MSDS) supplied with the iReceptal Docking Detergent for first aid information.
•DO NOT spill the detergent. Spills will be slippery and may result in a slip/fall hazard.
•ALWAYS follow the current local regulations governing environmental protection to recycle or dispose of
the bottle. DO NOT reuse the empty container.

SL 5107 Rev. D
16
CAUTION: Use only AMSINO iReceptal Detergent REF iRC003 with the docker. Failure to comply will result in
damage to rover and docker internal components.
1. Remove the detergent canister lid tube from the docker.
2. Pour the detergent into the detergent canister from the detergent bottle and dispose of the container properly.
Cleaning Maintenance
CAUTIONS:
•DO NOT immerse any system component in liquid. DO NOT allow liquids or moisture to enter any electrical
connection.
•DO NOT sterilize any system component.
•DO NOT use solvents, lubricants, or other chemicals, including glutaraldehyde or similar chemical
cleaners, unless otherwise specified.
•Use of unapproved disinfectants may cause system damage.
Recommended Equipment
•Personal Protective Equipment (PPE) as recommended by the disinfectant supplier (minimum: gown,
gloves, face/eye shield)
•Soft, lint-free cloth
•Environmental Protection Agency (EPA) registered disinfectant with a claim for activity against Hepatitis B.
The following disinfectants have been validated for use with the AMSINO iReceptal 3 Waste Management
System:
◼Sodium Hypochlorite Based - Clorox. Clean-Up. Disinfectant Cleaner with Bleach (EPA Reg. #67619-
1)
◼Quaternary Ammonium Based - CaviCide. (EPA Reg. #46781-6)
To Wipe Down the Docker
1. Wipe the external surfaces of the docker with a soft, lint-free cloth moistened with a non-abrasive, hospital
disinfectant prepared according to the manufacturer’s instructions. Make sure all surfaces remain visibly wet
at room temperature.
2. Remove any excess disinfectant solution using a soft, lint-free cloth moistened with water if required by the
instructions supplied by the disinfectant manufacturer.
3. Thoroughly clean the docker’s infrared communication ports to make sure the rover and docker can
communicate and function properly.
4. Inspect the docker. See the Inspection and Maintenance section.
Storage and Handling
CAUTIONS:

SL 5107 Rev. D
17
•ALWAYS store and transport the equipment within the specified environmental condition values throughout
its useful life. See the Specifications section.
•ALWAYS call AMSINO iReceptal Customer Service before transporting or storing this equipment in
freezing conditions. Failure to comply will cause the expansion of frozen internal fluid to damage the
equipment.
To ensure the longevity, performance and safety of this equipment, use of the original packaging material is
recommended when storing or transporting this equipment.
Disposal/Recycle
WARNING:
BLOODBORNE PATHOGEN AND CONTAMINATION HAZARDS –
•ALWAYS follow local regulations for safe handling, recycling, and disposal of biohazardous fluid waste and
Neptune equipment.
•Call iReceptal Customer Service for docker decontamination procedures.
•Discarded electromedical equipment must not be disposed together with waste but must be collected
separately to guarantee ecologically correct disposal to prevent dispersion of potential pollutants into the
environment. Please hand in the device at the end of its useful life to the local collection and recycling point
for electrical and electronic devices.
Failure to comply may cause environmental contamination or infection and result in personal injury.
Per the European Community (EC) Waste Electrical and Electronic Equipment (WEEE) Directive
2012/96/EC, product must be collected separately. DO NOT dispose of as unsorted municipal
waste. Contact local distributor for disposal information. Ensure infected equipment is
decontaminated prior to recycling.
Troubleshooting
NOTE: For service, contact your AMSINO sales representative or call AMSINO iReceptal Customer Service.
Outside the US, contact your nearest Stryker subsidiary.
PROBLEM
CAUSE
SOLUTION
Power switch does not illuminate
in the ON position.
Power cord is not connected or is
loosely connected.
Make sure the power cord is
connected securely.
The rover will not dock or an error
has occurred during the docking
procedure.
The facility water valve is shut off.
Turn on the facility water valve.
The docker is damaged.
Contact Customer Service.
Water inlet hose is leaking.
Water inlet hose connection is not
secure.
Make sure the water inlet hose
connection is secure.
Water inlet hose is damaged.
Replace the water inlet hose.
Waste outlet hose is leaking.
Waste outlet hose connection is
not secure.
Make sure the waste outlet hose
connection is secure.

SL 5107 Rev. D
18
Waste outlet hose is damaged.
Replace the waste outlet hose.
The docker does not dispense
detergent during the cleaning
cycle.
The canister of detergent is
empty.
Add detergent
If the problem persists, the docker
detergent pump may be damaged.
Contact Customer Service.
Sporadic electrical interference is
experienced.
Electrical noise is present.
Turn off all the electrical
equipment not in use in the room.
Relocate the electrical equipment
to maximize the distance between
the equipment. Increase spatial
distance.
Plug equipment into different
outlets.
Error Messages
NOTE: For error message information, see the instructions for use supplied with the rover.
Specifications
Model:
iRD301
REF
iRD301-01
Electrical Power Requirements:
100V-240V~,50/60 Hz,3 A
Power inlet module
Power switch with 250V fuses on neutral and line connection
European Conformity:
CE
Product Safety Certification:
EU Medical Device Directive 93/42/EEC and Directive 2007/47/EC
IEC 60601-1:2005, IEC 60601-1:2005/AMD1:2012 Medical electrical
equipment –Part 1:
General requirements for basic safety and essential performance
IEC 60601-1-2:2014 Collateral Standard: Electromagnetic disturbances
–Requirements and tests
Dimensions:
Width:63.3cm
Height:86.3cm
Depth:43.7cm
Mode of Operation:
Continuous
Mass
46kg—detergent empty
56kg—detergent full
Equipment Classification:
Class I Medical Electrical (ME) Equipment
Ingress Protection (IP):
IPX0
Light Emitting Diode (LED)
Classification (infrared
communication windows):
WARNING: INVISIBLE LED RADIATION
DO NOT VIEW DIRECTLY WITH OPTICAL INSTRUMENTS CLASS
1M LED PRODUCT —Viewing the laser output with certain optical
instruments (for example, eye loupes, magnifiers and microscopes)
within a distance of 100 mm may pose an eye hazard.

SL 5107 Rev. D
19
Ground Type:
Protective Earth (ground); when connected to facility power
Water Requirements:
Pressure Range:
Temperature Range:
Fitting Connection:
Water Quality:
Water Usage:
Facility Backflow Prevention
Device:
345 kPa to 827 kPa [50 to 120 psi]
4.4 to 43.3 °C [40 to 110 °F] NOTE: For optimal cleaning of the rover
canisters, use 37.8 to 43.3 °C [100 to 110 °F].
Facility source is equipped with a 3/4” Male (garden) Hose Thread
(MHT) fitting and has a dedicated shutoff valve.
Potable tap water
Approximately 34 liters [9 gallons] per rinse cycle at default settings on
standard cycle; water usage fluctuates due to selected cycle and facility
flow.
Refer to local plumbing codes to determine whether an external
backflow prevention device is required.
Drainage Requirements::
Floor drain or permanent service connection per local plumbing codes;
2.44 m [8 feet] connection distance (maximum)
Water Inlet Hose:
inner diameter:25 cm [0.50 inch]
length:3m
Waste Outlet Hose:
inner diameter:1.27 cm [0.50 inch]
length:3m
Waste Pump Outlet Flow:
23L/min
Environmental Conditions:
Temperature Limitation
Humidity Limitation
Atmospheric Pressure Limitation:
Operation Storage and Transportation Storage and
Transportation (before initial use) (after initial use)
(初次使用前) (初次使用后)

SL 5107 Rev. D
20
Electromagnetic Compatibility
Guidance and Manufacturer’s Declaration
Below cables information are provided for EMC reference.
Cable
Max. cable length,
Shielded/unshielded
Number
Cable
classification
AC Power Line
3.0m
Unshielded
1 Set
AC Power
Important information regarding Electro Magnetic Compatibility (EMC)
This electrical medical equipment needs special precautions regarding EMC and put into service according to
the EMC information provided in the user manual; The equipment conforms to this IEC 60601-1-2:2014
standard for both immunity and emissions. Nevertheless, special precautions need to be observed:
➢ The equipment with no ESSENTIAL PERFORMANCE is intended used in Professional healthcare facility
environment
➢ WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally”.
➢ The use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or decreased
electromagnetic immunity of this equipment and result in improper operation.
➢ WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of this product,
including cables specified by the manufacturer. Otherwise, degradation of the performance of this
equipment could result.”
➢ WARNING: If the use location is near (e.g. less than 1.5 km from) AM, FM or TV broadcast antennas,
before using this equipment, it should be observed to verify that it is operating normally to assure that
the equipment remains safe with regard to electromagnetic disturbances throughout the expected
service life.
EMI Compliance Table (Table 1)
Table 1 - Emission
Phenomenon
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1, Class A
Professional healthcare facility environment
Harmonic distortion
IEC 61000-3-2
Class A
Professional healthcare facility environment
This manual suits for next models
2
Table of contents
Other Amsino Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Zeiss
Zeiss CT LUCIA Series Handling instructions

Sunrise Medical
Sunrise Medical Jay 1200 Series User instruction manual & warranty

Atos Medical
Atos Medical Provox Dilator 17 manual

Pari
Pari VORTEX Tracheo Instructions for use

EASYHALER
EASYHALER Budesonide Instructions for using

GE
GE Marquette Responder 3000 Operator's manual