Anritsu Company Site Master S312D User manual

Maintenance Manual
Site Master Model S312D
Cable, Antenna, and Spectrum Analyzer
Anritsu Company
490 Jarvis Drive
Morgan Hill, CA 95037-2809
USA
P/N: 10580-00195
Revision: B
Printed: July 2009
Copyright 2009 Anritsu Company


S312D MM PN: 10580-00195 Revision B Safety-1
Safety Symbols
To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Company uses the
following symbols to indicate safety-related information. For your own safety, please read the information
carefully before operating the equipment.
Symbols Used in Manuals
Safety Symbols Used on Equipment and in Manuals
The following safety symbols are used inside or on the equipment near operation locations to provide
information about safety items and operation precautions. Ensure that you clearly understand the meanings of
the symbols and take the necessary precautions before operating the equipment. Some or all of the following
five symbols may or may not be used on all Anritsu equipment. In addition, there may be other labels attached
to products that are not shown in the diagrams in this manual.
This indicates a prohibited operation. The prohibited operation is indicated symbolically in or near
the barred circle.
This indicates a compulsory safety precaution. The required operation is indicated symbolically in or
near the circle.
This indicates a warning or caution. The contents are indicated symbolically in or near the triangle.
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
Danger
This indicates a very dangerous procedure that could result in serious
injury or death, and possible loss related to equipment malfunction, if not
performed properly.
Warning This indicates a hazardous procedure that could result in light-to-severe
injury or loss related to equipment malfunction, if proper precautions are
not taken.
Caution
This indicates a hazardous procedure that could result in loss related to
equipment malfunction if proper precautions are not taken.

Safety-2 PN: 10580-00195 Revision B S312D MM
For Safety
Warning Always refer to the operation manual when working near locations at which
the alert mark, shown on the left, is attached. If the operation, etc., is
performed without heeding the advice in the operation manual, there is a
risk of personal injury. In addition, the equipment performance may be
reduced.
Moreover, this alert mark is sometimes used with other marks and
descriptions indicating other dangers.
Warning
When supplying power to this equipment, connect the accessory 3-pin
power cord to a 3-pin grounded power outlet. If a grounded 3-pin outlet is
not available, use a conversion adapter and ground the green wire, or
connect the frame ground on the rear panel of the equipment to ground. If
power is supplied without grounding the equipment, there is a risk of
receiving a severe or fatal electric shock.
Warning This equipment cannot be repaired by the operator. Do not attempt to
remove the equipment covers or to disassemble internal components.
Only qualified service technicians with a knowledge of electrical fire and
shock hazards should service this equipment. There are high-voltage parts
in this equipment presenting a risk of severe injury or fatal electric shock to
untrained personnel. In addition, there is a risk of damage to precision
components.
Caution
Electrostatic Discharge (ESD) can damage the highly sensitive circuits in
the instrument. ESD is most likely to occur as test devices are being
connected to, or disconnected from, the instrument’s front and rear panel
ports and connectors. You can protect the instrument and test devices by
wearing a static-discharge wristband. Alternatively, you can ground
yourself to discharge any static charge by touching the outer chassis of the
grounded instrument before touching the instrument’s front and rear panel
ports and connectors. Avoid touching the test port center conductors
unless you are properly grounded and have eliminated the possibility of
static discharge.
Repair of damage that is found to be caused by electrostatic discharge is
not covered under warranty.

S312D MM PN: 10580-00195 Rev. B Contents-i
Table of Contents
Chapter 1—General Information
1-1 Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1-3 Recommended Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1-4 Replaceable Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Important Information Regarding Service of Anritsu Equipment . . . . . . . . . . . . . . . . . . . . . 1-3
Chapter 2—Performance Verification
2-1 Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2-2 VNA Frequency Accuracy. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-3 VNA Return Loss Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2-4 InstaCal Module Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2-5 Spectrum Analyzer Frequency Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7
2-6 Spectrum Analyzer Phase Noise Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
2-7 Input Related Spurious Response Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2-8 Spectrum Analyzer Resolution Bandwidth Accuracy Verification. . . . . . . . . . . . . . . . . . . . . . 2-12
1 MHz RBW Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
300 kHz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
100 kHz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
30 kHz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
10 kHz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
3 kHz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
1 kHz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
300 Hz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
100 Hz RBW Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
2-9 Spectrum Analyzer Level Accuracy Verification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Level Accuracy with Frequency Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
Level Accuracy with Power Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
2-10 Spectrum Analyzer Residual Spurious Response Verification. . . . . . . . . . . . . . . . . . . . . . . . 2-22
2-11 Spectrum Analyzer Displayed Average Noise Level (DANL) . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2-12 Power Monitor Verification (Option 5). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2-13 Bias Tee Verification (Option 10A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Bias Tee Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
Fault Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2-14 Transmission Measurement Verification (Option 21) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2-15 Power Meter Verification (Option 29) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28
Chapter 3—Removal and Replacement
3-1 Battery Pack Removal and Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3-2 Battery Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Battery Testing Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3

Contents-ii PN: 10580-00195 Rev. B S312D MM
Table of Contents (Continued)
3-3 Removal and Replacement Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Important Information Regarding Service of Anritsu Equipment . . . . . . . . . . . . . . . . . . . . . 3-4
3-4 Opening the Case. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-5 Removal of the LCD and Keypad Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3-6 Removal of the Main PCB, Option 5 PCB, Spectrum Analyzer Assembly . . . . . . . . . . . . . . . . 3-6
3-7 RTC Battery Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Appendix A—Test Records
Spectrum Analyzer Frequency Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Spectrum Analyzer Level Accuracy Verification (continued) . . . . . . . . . . . . . . . . . . . . . . . A-6
Appendix B—Test Fixture Schematics

S312D MM PN: 10580-00195 Revision B 1-1
Chapter 1 — General Information
1-1 Introduction
This manual provides maintenance instructions for the Site Master Model S312D Cable, Antenna and
Spectrum Analyzer. It describes the product and provides performance verification procedures, parts
replacement procedures, and a replaceable parts list.
1-2 Description
The Site Master is a handheld SWR/RL (standing wave ratio/return loss), Distance-To-Fault, spectrum
analysis, and power meter (optional) measurement instrument. It combines a synthesized source,
VSWR bridge, receiver, and spectrum analyzer circuitry in a compact instrument.
Throughout this manual, the term “VNA” (Vector Network Analyzer) denotes Return Loss, SWR, Cable Loss,
and DTF modes, and the term “SPA” denotes Spectrum Analyzer mode. All other modes are referenced
individually.

1-3 Recommended Test Equipment General Information
1-2 PN: 10580-00195 Revision B S312D MM
1-3 Recommended Test Equipment
Table 1-1. Required Equipment for Performance Verification
Instrument Critical Specification Recommended
Manufacturer/Model
Synthesized Signal Generator Frequency: 0.1 Hz to 20 GHz Anritsu Model MG3692A or B with
Option 2A and Option 4
Power Meter Power Range: –70 dBm to +20 dBm Anritsu Dual Channel Power Meter
Model ML2438A
Power Sensor Frequency: 10 MHz to 18 GHz
Power Range: –67 dB to +20 dB Anritsu Model MA2442D or equivalent
(quantity 2)
Frequency Reference Frequency: 10 MHz Symmetricom Model RubiSource T&M
50 MHz Lowpass Filter N(m) to N(f) Anritsu Part Number 1030-96
Power Splitter Frequency: DC to 18 GHz Aeroflex/Weinschel Model 1870A
Adapter N(m)-N(m), 50 Ohm Anritsu Model 34NN50A
Adapter Frequency: DC to 20 GHz,
K(m)-N(f), 50 Ohm Anritsu Model 34RKNF50
50 Ohm Termination Frequency: DC to 18 GHz Anritsu Model 28N50-2
RF Coaxial Cable N(m)-N(f), 50 Ohm Anritsu Model 15NN50-1.5C or
equivalent
Spectrum Analyzer 9 kHz to 21.2 GHz Anritsu Model MS2665C or equivalent
Open/Short DC to 18 GHz N(m) Anritsu Model 22N50
InstaCal Module 2 MHz to 6.0 GHz, N(m), 50 Ohm Anritsu Part Number ICN50
Fixed Attenuator 10 dB Aeroflex/Weinschel Model 44-10
Fixed Attenuator 20 dB Aeroflex/Weinschel Model 44-20
Fixed Attenuator 30 dB Aeroflex/Weinschel Model 44-30
Offset Termination 20 dB, 4 GHz Anritsu Model SC5270
Offset Termination 6 dB, 4 GHz Anritsu Model SC5237
RF Detector (for Option 5) 10 MHz to 20 GHz Anritsu Model 560-7N50B
Low Current Load Fixture
(for Option 10A) 105 Ohm, 1 Watt Anritsu Part Number T3377
High Current Load Fixture
(for Option 10A) 40 Ohm, 5 Watt Anritsu Part Number T2904

General Information 1-4 Replaceable Parts
S312D MM PN: 10580-00195 Revision B 1-3
1-4 Replaceable Parts
Important Information Regarding Service of Anritsu Equipment
Only qualified service personnel should attempt to perform repairs on this instrument. During the warranty
period, opening of the case by non-Anritsu Service personnel will void the warranty.
Extreme care must be used when handling internal assemblies. Careless handling will cause damage.
Unless authorized by the factory, no attempts should be made to repair a defective assembly. (Discharged RTC
batteries on the main PCB may be replaced). Exchange assemblies that are determined to have been damaged
by improper handling will not be accepted for credit.
Table 1-2. List of Replaceable Assemblies
Part Number Description
ND67965 S312D Main/SPA Assembly (without Option 10A, without Option 31)
ND67967 S312D Main/SPA Assembly (with Option 31, without Option 10A)
ND67969 S312D Main/SPA Assembly (with Option 10A, without Option 31)
ND67971 S312D Main/SPA Assembly (with both Option 31 and Option 10A)
ND66432 Option 5 PCB
3-633-26 Coin Battery for RTC
15-123 LCD Display
46649-6 Keypad Membrane
67199-3 Keypad PCB
633-27 Rechargeable battery
790-523 Speaker
48231-1 Battery Door
61440-1 Case Top
58214 Case Bottom
46655 Corner Bumper
40-168-R AC Adapter

1-4 Replaceable Parts General Information
1-4 PN: 10580-00195 Revision B S312D MM

S312D MM PN: 10580-00195 Revision B 2-1
Chapter 2 — Performance Verification
2-1 Performance Verification
The following sections contain tests that can be used to verify the performance of the Site Master S312D.
Before performing any of the tests in these sections, ensure that the S312D has had a minimum of 30 minutes
warm up and that the test equipment has been warmed up to the specifications of its manufacturer. Table 1-1,
“Required Equipment for Performance Verification” on page 1-2 lists the required equipment for performance
verification.
Note Anritsu Company recommends that you use an AC/DC power adapter during performance
verification of the S312D.

2-2 VNA Frequency Accuracy Performance Verification
2-2 PN: 10580-00195 Revision B S312D MM
2-2 VNA Frequency Accuracy
The following test can be used to verify the VNA CW frequency accuracy of the Site Master. Measurement
calibration of the Site Master is not required for this test.
Required Equipment (refer to Table 1-1):
•Spectrum Analyzer, Anritsu Model MS2665C
•Frequency Reference Symmetricom Rubisource T&M
•Site Master S312D
Procedure:
1. Connect a 10 MHz Reference signal to the 10 MHz STD Ref In of the Spectrum Analyzer as shown in
Figure 2-1.
2. Press and hold the ESCAPE/CLEAR key, then press the ON/OFF key to turn on the Site Master. (This
sets the instrument to the factory preset state.)
3. Press the FREQ/DIST key, then press the F1 soft key and set F1 to 1000 MHz.
4. Press the F2 soft key, set F2 to 1000 MHz.
5. Press the MEAS/DISP key, then press the Fixed CW soft key to turn on Fixed CW.
6. Connect the RF cable from the Site Master Reflection Test Port to the RF Input on the Spectrum
Analyzer.
7. Set up the Spectrum Analyzer as follows:
a. Press the Preset key, then press Preset All (F1).
Figure 2-1. VNA Frequency Accuracy Test Setup
3MS2665C Spectrum Analyzer
9kHz–21.2GHz
Preset
Remote
Local
Copy Cont
Copy
Stby On
Memory Card
Menu On/Off
F1
F2
F3
F4
F5
F6
More
Freq/Ampl Marker
Multi Mkr
Marker
Frequency
Span
Amplitude
Peak
Search
User
Marker
Continuous
Single
CF RLV
Peak
Save
Recall Measure
A.B Time Trig/Gate
RBW VBW Sweep Time Atten
A/Time
A/B.A/BG
Display
Coupled Function
5),QSXWȍ
+30dBm
0V DC Max !
ȝ9
ȝVHF
mV
msec
kHz
Hz+/–ƒ
0
123
Sound System Title
4 5 6 MHz V
sec
Cal Interface Mem Card Enter
7 8 9 GHz dBm
dB
PTA Define Hold
CE
BS
Shift
Entry
S332D Site Master SiteMaster
1
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT MARKER
SAVE
DISPLAY
RECALL
DISPLAY
PRINT
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
12
4
56
78
09
3
MODE
FREQ/DIST
AMPLITUDE MEAS/DISP
MS2665C
10 MHz Reference
S312D
to
rear
connector

Performance Verification 2-2 VNA Frequency Accuracy
S312D MM PN: 10580-00195 Revision B 2-3
b. Press the Frequency key.
c. Set the Center Frequency to 1 GHz.
d. Press the Span key.
e. Set the frequency span to 750 kHz.
f. Press the RBW key.
g. Set the RBW to 10 kHz.
h. Press the VBW key.
i. Press the Filter Off soft key (F3) to turn off the VB filter.
j. Press the Amplitude key.
k. Set the Reference Level to 0 dBm.
l. Press the Log Scale soft key (F5).
m. Press 2dB/Div(F3) and then press the Return soft key.
8. When a peak response appears on the Spectrum Analyzer, press the Marker Peak Search key on the
Spectrum Analyzer. Record the frequency in Table A-1, “VNA Frequency Accuracy” on page A-2.
9. On the Site Master, press the MEAS/DISP key then the Fixed CW soft key to turn off Fixed CW.
Note If the Site Master has entered into the hold mode, then press the RUN/HOLD key to return to normal
mode.

2-3 VNA Return Loss Verification Performance Verification
2-4 PN: 10580-00195 Revision B S312D MM
2-3 VNA Return Loss Verification
The following test can be used to verify the accuracy of return loss measurements. Measurement calibration of
the Site Master is required for this test.
Required Equipment (refer to Table 1-1):
•Site Master S312D
•Open/Short, Anritsu Model 22N50
•50 ohm termination, Anritsu Model 28N50-2
Procedure:
1. Press and hold the ESCAPE/CLEAR key, then press the ON/OFF key to turn on the Site Master. (This
sets the instrument to the factory preset state.)
2. Press the MODE key.
3. Use the Up/Down arrow key to highlight Return Loss, then press ENTER.
4. Press the START CAL key.
5. Follow the instructions on the screen to perform a calibration using a 22N50 Open/Short and
28N50-2 Termination.
6. Connect the 20 dB offset to the RF Out port and record the minimum and maximum values in Table A-2,
“VNA Return Loss Verification” on page A-2.
7. Connect the 6 dB offset to the RF Out port and record the minimum and maximum values in Table A-2.
Figure 2-2. VNA Return Loss Verification Test Setup
28N50-2
S312D
22N50
SC5270
SC5237
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT MARKER
SAVE
DISPLAY
RECALL
DISPLAY
PRINT
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
12
4
56
78
09
3
MODE
FREQ/DIST
AMPLITUDE MEAS/DISP
SiteMaster
S332D Site Master
1

Performance Verification 2-4 InstaCal Module Verification
S312D MM PN: 10580-00195 Revision B 2-5
2-4 InstaCal Module Verification
This test verifies the performance of the Anritsu Site Master InstaCal Calibration Module. The InstaCal
Module (part number ICN50) is an optional accessory for the S312D.
Required Equipment (refer to Table 1-1):
•Site Master S312D
•InstaCal Module
•Offset Termination, 6 dB, Anritsu Model SC5237
•Offset Termination, 20 dB, Anritsu Model SC5270
Procedure:
1. Press and hold the ESCAPE/CLEAR key, then press the ON/OFF key to turn on the Site Master. (This
sets the instrument to the factory preset state.)
2. Press the MODE soft key.
3. Use the Up/Down arrow key to highlight RETURN LOSS, then press ENTER.
4. Press the START CAL key. The message “CONNECT OPEN or InstaCal TO RF Out PORT” appears in
the display.
5. Connect the InstaCal Module to the RF Out port (as shown in Figure 2-3) and press the ENTER key.
Figure 2-3. InstaCal Module Verification Test Setup
Note Before continuing, allow a five-minute warm-up for the internal circuitry to stabilize.
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT MARKER
SAVE
DISPLAY
RECALL
DISPLAY
PRINT
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
12
4
56
78
09
3
MODE
FREQ/DIST
AMPLITUDE MEAS/DISP
SiteMaster
S332D Site Master
1
S312D
ICN50
SC5270
SC5237

2-4 InstaCal Module Verification Performance Verification
2-6 PN: 10580-00195 Revision B S312D MM
6. Verify that the calibration has been properly performed by checking that the CAL ON! message is
displayed in the upper-left corner of the display.
7. Remove the InstaCal Module from the RF Out port and connect the 20 dB Offset (SC5270) to the RF Out
port.
8. Measure the return loss of the 20 dB Offset. The level should be 20 dB ± 2 dB across the calibrated
frequency range.
9. Remove the 20 dB Offset from the RF Out port and connect the 6 dB Offset (SC5237) to the RF Out port.
10. Measure the return loss of the 6 dB Offset. The level should be 6 dB ± 1.2 dB across the calibrated
frequency range.
Note
If this particular InstaCal Module has been used previously to calibrate this Site Master, then the
Site Master senses the familiar InstaCal Module and automatically calibrates the instrument by using
the OSL procedure.
If the Site Master senses that the characterization data for the InstaCal Module that is connected to
this Site Master is different than the characterization data that is currently stored, then it displays soft
key options to keep or to replace the InstaCal characterization data.
Pressing the YES soft key transfers all of the characterization data from this InstaCal Module to the
Site Master. The transfer may take up to three minutes. This option is preferred if this InstaCal
Module is to stay with this particular Site Master. After calibration is completed, the data need not be
transferred again for this combination of Site Master and InstaCal Module.
Pressing the NO soft key temporarily transfers only the portion of the characterization data that is
necessary for this particular calibration. This transfer takes approximately 30 minutes to 60 minutes
and must be repeated every time that a calibration is performed by using this combination of
Site Master and InstaCal Module.

Performance Verification 2-5 Spectrum Analyzer Frequency Accuracy
S312D MM PN: 10580-00195 Revision B 2-7
2-5 Spectrum Analyzer Frequency Accuracy
The following test can be used to verify the CW frequency accuracy of the Site Master Spectrum Analyzer.
Required Equipment (refer to Table 1-1):
•Anritsu Model MG3692x Synthesized Signal Generator
•Frequency Reference Symmetricom Rubisource T&M
•Site Master S312D
Procedure:
1. Connect the 10 MHz reference source to the Synthesized Signal Generator.
2. Connect the output of the source to the Spectrum Analyzer RF Input of the Site Master.
3. Press and hold the ESCAPE/CLEAR key, then press the ON/OFF key to turn on the Site Master. (This
sets the instrument to the factory preset state.)
4. Set the Synthesized Signal Generator output to 1000 MHz CW, with an RF output level of 0 dBm.
5. On the Site Master, press the MODE key. Use the Up/Down arrow key to highlight Spectrum Analyzer and
press ENTER to select spectrum analyzer mode.
6. Press the AMPLITUDE key and the Ref Level soft key.
7. Enter 20 and press the ENTER key to set the Reference Level to 20 dBm.
8. Press the FREQ/DIST key and the Center soft key.
9. Enter 1000 and press ENTER to set the center frequency to 1000 MHz.
10. Press the Span soft key, enter 20, and press the kHz soft key to set the span to 20 kHz.
11. Confirm that the RBW is 100 Hz, and the VBW is 30 Hz. If adjustment of the RBW and VBW are
required:
a. Press the MEAS/DISP key and the Bandwidth soft key.
Figure 2-4. Spectrum Analyzer Frequency Accuracy Test Setup
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT MARKER
SAVE
DISPLAY
RECALL
DISPLAY
PRINT
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
12
4
56
78
09
3
MODE
FREQ/DIST
AMPLITUDE MEAS/DISP
SiteMaster
S332D Site Master
1
MG369XA Source
10 MHz Reference
to rear panel
connector
S312D

2-5 Spectrum Analyzer Frequency Accuracy Performance Verification
2-8 PN: 10580-00195 Revision B S312D MM
b. Press the RBW Manual soft key and use the Up/Down arrow key to select 100 Hz. Press ENTER to
set the resolution bandwidth to 100 Hz.
c. Press the VBW Manual soft key and use the Up/Down arrow key to select 30 Hz. Press ENTER to
set the video bandwidth to 30 Hz.
12. Press the MARKER key, then the M1 soft key.
13. Press the Edit soft key and use the Up/Down arrow key to center the marker on the waveform. Record the
frequency in Table A-3, “Spectrum Analyzer Frequency Accuracy” on page A-3.

Performance Verification 2-6 Spectrum Analyzer Phase Noise Verification
S312D MM PN: 10580-00195 Revision B 2-9
2-6 Spectrum Analyzer Phase Noise Verification
This test can be used to verify the phase noise of the Site Master Spectrum Analyzer.
Required Equipment (refer to Table 1-1):
•Anritsu Model MG3692x Synthesized Signal Generator
•Frequency Reference Symmetricom Rubisource T&M
•Site Master S312D
Procedure:
1. Connect the output of the Synthesized Signal Generator to the Site Master RF Input. (Refer to
Figure 2-4.)
2. Press and hold the ESCAPE/CLEAR key, then press the ON/OFF key to turn on the Site Master. (This
sets the instrument to the factory preset state.)
3. Set the Synthesized Signal Generator output to 1000 MHz CW, with an RF output level of –30 dBm.
4. On the Site Master, press the MODE key. Use the Up/Down arrow key to highlight Spectrum Analyzer,
then, press the ENTER key to select spectrum analyzer mode.
5. Press the MEAS/DISP key and the Bandwidth soft key.
6. Press the RBW Manual soft key and use the Up/Down arrow key to select 1kHz. Press ENTER to set the
resolution bandwidth to 1 kHz.
7. Press the VBW Manual soft key and use the Up/Down arrow key to select 30 Hz. Press ENTER to set the
video bandwidth to 30 Hz.
8. Press the FREQ/DIST key and the Center soft key.
9. Enter 1000 and press the ENTER key to set the center frequency to 1000 MHz.
10. Press the Span soft key and enter 0.1. Press the ENTER key to set the span to 0.100 MHz.
11. Press the AMPLITUDE key.
12. Press the Ref Level soft key and enter –27. Press ENTER to set the reference level to –27 dBm.
13. Press the MARKER key, then the M1 soft key.
14. Press EDIT and enter 1000. Press ENTER to set the M1 marker frequency to 1000 MHz.
15. Press the BACK soft key and the M2 soft key.
16. Press EDIT and enter 1000.03. Press ENTER to set the M2 marker frequency to 1000.03 MHz (30 kHz
higher than the center frequency).
17. Press the DELTA (M2–M1) soft key.
18. Press the RUN/HOLD key and read and record the Δ2 reading into Table A-4, “Spectrum Analyzer Phase
Noise Verification” on page A-3.
19. Press the RUN/HOLD key to read and record a total of five values into Table A-4, then calculate the
average of the five recorded values, and enter the average into Table A-4.
20. To convert to dBc/Hz, subtract 30 dB from the average value and record the calculated value into
Table A-4.Verify that the result is ≤–75 dBc/Hz.
For example: –45 dBc measured – 30 dB = –75 dBc/Hz
21. Press the BACK soft key and the M3 soft key.
22. Press the EDIT key and enter 999.97. Press MHz to set the M3 marker frequencyto999.97MHz(30kHz
lower than the center frequency).
23. Press the DELTA (M3–M1) soft key.
24. Press the RUN/HOLD key. Read and record the Δ3 reading into Table A-4.

2-7 Input Related Spurious Response Verification Performance Verification
2-10 PN: 10580-00195 Revision B S312D MM
25. Press the RUN/HOLD key to read and record a total of five values into Table A-4, then calculate the
average of the five recorded values, and enter the average into Table A-4.
26. Subtract 30 dB from the average value and record the calculated value into Table A-4, Verify that the
result is ≤–75 dBc/Hz.
2-7 Input Related Spurious Response Verification
The following test can be used to verify the input related spurious response of the Site Master Spectrum
Analyzer.
Required Equipment (refer to Table 1-1):
•Anritsu Model MG3692x Synthesized Signal Generator
•Frequency Reference Symmetricom Rubisource T&M
•Lowpass Filter, Anritsu Part Number 1030-96 (RLC Electronics)
•Site Master S312D
Procedure:
1. Connect the 10 MHz reference source to the Synthesized Signal Generator as shown in Figure 2-5.
2. Connect one end of the 50 MHz Low Pass Filter to the output of the source and the other end to the
Site Master Spectrum Analyzer RF Input with the coaxial cable.
3. On the Site Master, press and hold the ESCAPE/CLEAR key, then press the ON/OFF key to turn on the
Site Master. (This sets the instrument to the factory preset state.)
4. Set the Synthesized Signal Generator output to 40 MHz CW, with an RF output level of –30 dBm.
5. On the Site Master, press the MODE key. Use the Up/Down arrow key to highlight Spectrum Analyzer,
then, press the ENTER key to select spectrum analyzer mode.
6. Press the AMPLITUDE key and the Ref Level soft key.
7. Enter –27and press the ENTER key to set the Reference Level to –27 dBm.
Figure 2-5. Input Related Spurious Response Verification Test Setup
HOLD
RUN
START
CAL
AUTO
SCALE
SAVE
SETUP
RECALL
SETUP
LIMIT MARKER
SAVE
DISPLAY
RECALL
DISPLAY
PRINT
SYS
ENTER
CLEAR
ESCAPE
ON
OFF
12
4
56
78
09
3
MODE
FREQ/DIST
AMPLITUDE MEAS/DISP
SiteMaster
S332D
Site Master
1
MG369XA Source
10 MHz Reference
to rear panel
connector
S312D
Low Pass
Filter
Table of contents
Other Anritsu Company Measuring Instrument manuals

Anritsu Company
Anritsu Company ME7838A4X User manual

Anritsu Company
Anritsu Company MS2760A-0032 User manual

Anritsu Company
Anritsu Company ATEK S251B User manual

Anritsu Company
Anritsu Company VectorStar ME7828A User manual

Anritsu Company
Anritsu Company ShockLine MS46122B Series User manual

Anritsu Company
Anritsu Company VectorStar ME7838A4 User manual

Anritsu Company
Anritsu Company VectorStar ME7848A/E-0210 User manual

Anritsu Company
Anritsu Company VectorStar ME7848A-0240 User manual