DC Power The ML2400A Series Power Meter can also operate
from a nominal external 12-24 VDC input in the ab-
sence of AC line power. DC line power is protected
by a fuse mounted inside the unit, on the main
board. A grounding terminal is provided on the rear
panel to ground the unit during operation from a
DC supply.
Battery
Power The ML2400A Series Power Meter can be operated
using the optional internal battery pack. During
battery operation, an icon will be displayed on
measurement screens indicating the state of charge.
When the remaining capacity reaches less than
10%, the icon will flash, indicating that charging
will soon be required. When running from battery
power, an estimate of typical-use running time
remaining can be viewed using the System menu
(see Chapter 4, Front Panel Operation). Note that,
due to power consumption considerations, GPIB and
serial remote operation are not available when the
power meter is running from the battery.
The AUTO POWER OFF feature is also available
through the System menu, and can be used to
automatically switch the unit to standby after a
specified period of inactivity to save battery power.
The timer can be set for 10 to 240 minutes, and any
key press will restart the timer. This same feature
will automatically switch the unit to standby when
the battery is fully discharged in order to minimize
the risk of over-discharge.
For optimum battery life, store the battery pack
at –20 to +50° C for short periods and –20 to +35° C
for long term storage.
The ML2400A Series Power Meter will operate from
AC or DC main power with this battery removed.
This battery is not used for the retention of nonvola-
tile memory functions. Refer to Section 2-7, Battery
Charging, Removal and Replacement, for further in-
formation.
Fuses The ML2400A Series Power Meter AC and DC input
lines are protected by internally mounted fuses.
These fuses should only be changed by qualified
service personnel. Replace only with fuses of the
same type and rating (AC fuse is 2A, 250V,
slow-blow; DC fuse is 3A, 125V, slow-blow).
2-2 ML2400A OM
POWER REQUIREMENTS INSTALLATION
NOTE
The ML2400A Series Power
Meter uses a high-capacity
Ni-MH battery (option
ML2400A-11). Over-discharge
canresultinapermanentlossof
battery capacity of as much as
20%.Ifthe unitistobestoredfor
anextendedperiod (longerthan
one week), remove the optional
battery pack so as to preclude
over-discharge.