Anyload DD-KC1 User manual

Page 2 of 11 Anyload-manual-indicator-DD-KC1-en-101
1. Structure and Installation
2. Packing List
①Indicator 1 piece
②ACAdapter 1 piece
③Instruction 1 copy
④Guarantee Statement 1 copy
3. Buttons’ Functions
Table 1
Key
Function
Menu Function
Serial port order code
Turn on/off
Null
O
Switch the unit
Go Down on menu or
parameters
U
Hold / Print
function
Go Up on menu or
parameters
H
Tare
Enter menu or confirm the
change of parameters
T
4.Operation Method
4.1 Turn on and off
Adapter: Insert the plug of AC Adapter into the power supply hub which is on the
left of the Display Indicator, and insert the AC Adapter into AC socket.
Battery:Open the battery cover which is on the right of the Display Indicator, load 4
xAAA alkaline cells into the cell box (note the polarity) and then close the battery cover.
Empty the platform, press button once, it will show all the characters then

Page 3 of 11 Anyload-manual-indicator-DD-KC1-en-101
the Maximum Capacity and Division. After the screen displays “0”, you can begin to
measure the object.
When power-on, press button to turn off the scale.
4.2 Instruction for load cell connection
4.2.1 Connection of the load cell
The connection of the four wire load cell and the indicator:
Connect +EXC with the red wire of the indicator’s cable.
Connect –EXC with the black wire of the indicator’s cable.
Connect +SIG with the green wire of the indicator’s cable.
Connect –SIG with the white wire of the indicator’s cable.
The indicator provides driving voltage through +EXC wire and –EXC wire. +SIG
wire and –SIG wire will transmit the change of signal voltage to the indicator.
*To avoid short circuit, please make sure the welding parts of all the wires
covered by the heat shrinking tube.
Figure:Connection of the Load Cell and the indicator
4.2.2 Load cell output sensitivity:
1~2m V/V
4.2.3 Factory parameter setting.
1. Turn off the scale, then hold the three buttons: , and at the
same time until the screen displays “------”;
2. Press or button to go through the menu;
3. Press button to enter the display menu.;

Page 4 of 11 Anyload-manual-indicator-DD-KC1-en-101
4. After entering the display menu, Press or key to select
parameters;Press button to verify the parameters and return to the
menu;
5. After all the parameters are set, go to “-End-”in the menu, short press
button to save and exit to weighing or counting mode
4.3 Weighing
When power-on, with nothing on the platform, press button once, the
instrument will then display “0 ”. Place an object on the platform. After the “kg”, “lb”
or “pcs” character stops flashing, which means that the data is stable, you can
begin to read it.
4.4 Calibration Methods
4.4.1 Random Single-point Calibration
a) At weighing mode, press and hold button until the window
displays“------”;
b) Press or button to enter calibration mode;
c) Press button to make sure enter the calibration mode;
d) After entering the calibration mode, the window will display“Load”and“0.0kg”
alternately,which means that the weight“0.0kg”should be loaded on the
platform;
e) After making sure that there is nothing on the platform, press button to
confirm so that the scale begins“0.0kg”calibration. In the process of
“0.0kg”calibration, “0.0kg” will flicker rapidly on the display window;keep the
platform calm, stable, and without other interference factors which will influence
Menu
Meaning of Menu
Selective Parameter
1.dP
Decimal Point Setting
0/0.0/0.00/0.000/0.0000
2.CAP
Capacity setting
3000,6000,7500,8000,10000,12000,
15000,20000,30000,50000,60000
75000,80000,100000,120000,150000
,200000,300000
3.InC
Division setting
d=1/2/5
-End-
Save and Exit

Page 5 of 11 Anyload-manual-indicator-DD-KC1-en-101
the weighing performance. If the weighing remains stable for longer than 2s,
the scale will finish “0.0kg”calibration automatically, it will then begin the next
calibration point; otherwise, “0.0kg” will keep flickering to wait for the scale to
the stablize.
f) Suppose that the next calibration point should be“xx.xkg”,the window will
display “Load”and“xx.xkg” alternately,which means that the weight“xx.xkg”
(WEIGHT)should be loaded onthe platform. If you have to change this weight,
you can press or button to enter the modify menu; then press
button to change the flashing position and press button to make
the flashing position’s number plus 1, and press button to make sure the
modification is finished;
g) After making sure the weight on the platform is“xx.xkg”, press button to
confirm so that the scale begins the “xx.xkg”calibration. In the process of“xx.xkg”
calibration, “xx.xkg” flickers rapidly on the display window;keep the platform
calm, stable, and without other interference factors which will influence the
weighing performance. If the weighing remains stable for more than 2s, the
scale will finish “xx.xkg”calibration and save the calibrated parameters, then will
return to the weighing mode; otherwise, “xx.xkg”will keep flickering until the
scale stable.
4.4.2 Standard Two Point Weight Calibration
a) At weighing mode, press and hold button until the window displays“------”;
b) Press or button cycle the display until the window displays“Line”;
c) Press button to begin calibration procedure;
d) After the scale begins to calibrate, the window will display“Load”and“0.0kg”
alternately;
e) After you confirm that there is nothing on the platform, press button to
confirm and begin the zero calibration when the window displays“0.0kg”. In the
process of zero calibration,“0.0kg”flickers rapidly on the display window;please
keep the platform calm, stable, and without other interference factors which will
influence the weighing performance. If the weighing keeps stable exceeding 2s,
the scale will finish zero calibration automatically, and then it will begin the next

Page 6 of 11 Anyload-manual-indicator-DD-KC1-en-101
calibration point;otherwise, “0.0kg” will keep flickering to wait for the stable
weighing.
f) Repeat the step d) and step e) above,calibrate the half capacity range point
and the Full capacity range point orderly, save the calibrated parameters
automatically and return to the weighing mode.
4.5 Switching the Unit
Press the button to switch the unit of measurement, kg --- lb--- pcs--- kg.
4.6 Keeping Function “Hold”
Once the key is pressed:
4.6.1 The CPU will send the weight data into RS232 socket.
4.6.2 The weight will remain on the display for 120 seconds after the item has been
removed from the scale so the weight can easily be read.
a) Place the item on the scale platform.
b) Wait for the stable weights to be displayed.
c) Press the key. The “►” hold indicators will turn-on.
d) Remove the item from the scale platform. The item’s weight reading will
remain on the display for 120 seconds.
e) Quit to hold function: Once an item is weighed and the hold function is
enabled; when you want to quit to hold function, press the key again
to cancel the “►” hold indicator and return the scale to weight value.
4.7 Print output function
Press and hold the
for 2 seconds, then the buzzer will sound, and the weight data
will output to the computer or the serial printer.
4.8 Counting Sample
a. During counting mode, remove unrelated objects on the platform, then press
button to revert to zero;
b. Press button to switch the unit into “PCS”, and the window will display

Page 7 of 11 Anyload-manual-indicator-DD-KC1-en-101
“ADD10”.
c. “ADD10”means that 10 objects to be weighed should be added onto the
platform. Attention: The weight of each sample should be higher than the
scale’s minimum division value; otherwise, the window will display wrong
cue“no”.
d. If you want to change the sampling quantity, please press the button to
change the quantity; if you do not need to sample again, press the
button to cancel the sampling, and the scale will adopt last sampling value to
count ;
e. After you confirm the quantity of objects on the platform and the data displayed
on the LCD are the same, short press to confirm, and then begin to
sample and count;
f. During the process of sampling and counting, the screen will display the
sampling data by rapidly flickering; keep the platform calm, stable, and without
other interference factors which will influence the weighing performance. If the
weighing stays stable longer than 2s, the sampling and counting will be
automatically finished, the parameter will be saved automatically and the scale
will return to counting mode; if not, the screen will keep flickering the data and
waiting for the stability of measuring.
4.9 Serial Port Printout
a.Connect the serial port with the scale according to the drawing below. Baud
Rate=9600,without parity digit, sent stop bit 1 (The serial connection diagram
is omitted);
b. Open the serial communication software, for example, “HyperTerminal” in

Page 8 of 11 Anyload-manual-indicator-DD-KC1-en-101
windows. Select the connected portal number “COM” and set the baud rate as
9600bps, the parity bit as “NONE”, the data bit as “8Bit”, and the stop bit as
“1Bit”.
c. You can select the serial output modes according to your own needs. The
output modes include 3 types: output by pressing button, automatic
output after the weight keeps stable, and continuous output. You can set
though the “out” in “Parameter setting”.
d. Output Data Form: 8 bytes data bit + 3 bytes unit code + 2 bytes ending code.
The above-mentionedcharacters are all ASCII codes, and the unit can be kg, lb
or pcs.
e. The serial ports order can be f0und in Chapter 2 “button function”.
Example of Output Data:
Display
Data Bit
Unit Code
Ending Code
1
2
3
4
5
6
7
8
9
1
0
11
12
13
123.45kg
Blank
Blank
1
2
3
.
4
5
k
g
Blank
CR
LF
-123.45lb
Blank
-
1
2
3
.
4
5
l
b
Blank
CR
LF
123.0pcs
Blank
Blank
Blank
1
2
3
.
0
p
c
s
CR
LF
4.10 When the scale is on, press and hold key until the display "0FF" or "ON",
then release. That means open or close the function of animal weighing. When the
action symbol which is above the units symbol flashes, that means the current mode is
animal mode. in this state, once the weighing is locked, it will automatically remain
locked weighs for about 10 seconds.
5.Set User Parameter
a) Wheb the scale is off, please press button together with the button
to turn on the scale. Do not release the button until the screen displays
“------”;
b) Press the or button to switch the menu;
c) Press the button to enter the display menu;
d) After entering the display menu, press the or button to select the

Page 9 of 11 Anyload-manual-indicator-DD-KC1-en-101
parameter;
e) Press the button to confirm the parameter, and return to the menu;
f) After all the parameters have been set, switch the menu to “-End-”,and press the
button to finish the setting and save the parameters, then return to the
weighing or counting mode.
Table 3
Menu
Meaning of
Menu
Selectable
Parameter
Meaning of Parameter
1.codE
Check the
internal code
null
Check the internal code for debugging by
manufacturer.
2.ASt
Zero tracking
range
0.5/1/2/3d
Set automatic zero tracking range.
3.out
Printout
method
kEy/Auto/Co
nti
kEy:printout while press the key,Auto:
printout automatically ,Conti :printout
continuously
4.LEd
Backlight
control method
OFF/Auto/on
OFF:backlight is turned off,Auto:
backlight is automatic,On:backlight is
turned on.
5.OFF
Automatic
shutdown
control
no/3/510/15/
30/60 min
No: no automatic shutdown; 3~60min: if
no action of weighing objects or pressing
buttons, the instrument will automatically
turn off.
6.-LL-
Set Lower limit
alarm
Xxxx (lower
limit value)
If the lower limit value is set to“0”, the
lower limit alarm function is canceled.
7.-HH-
Set upper limit
alarm
xxxx(upper
limit value)
If the upper limit value is set to“0”, the
upper limit alarm function is canceled.
-End-
Finish
measuring
and save the
parameters
null
6.Meaning of Cue Code Table 4
Display Code
Meaning of Code
Solution
no
1. The calibration weight value is
not loaded according to the display
when calibrating.
2. The quantity of objects is not
loaded according to the quantity of
1. Load the weight value
according to the show value.
2. Load the quantity
according to the show value.

Page 10 of 11 Anyload-manual-indicator-DD-KC1-en-101
sampling.
AC
The batteries are brownout.
1. Use new batteries.
2. Use power adapter.
Err-0
1. Lost the calibration data.
2. The load cell or main board
maybe damaged.
1. Re-calibrate.
2. Change the Loadcell or
main board.
Err-1
1.The weight is over the max
capacity.
2. Lost the calibration data.
3. The load cell or main board
maybe damaged.
1. Remove the objects on the
platform.
2. Re-calibrate.
3. Change the load cell or
main board.
--L--
Lower limit alarm
1.cancel lower limit alarm
--H--
Upper limit alarm
1.cancel upper limit alarm
7. Maintenance of Scale
In order to keep the scale’s normal operating, the crust of indicator and the platform
of scale should be kept clean and prevented from corrosion by foreign substance. If
necessary, you should use a piece of cloth with mild washing agent to clean them.
When you don’t use the scale, pull out the A/C plug. If not being used in a long time,
please take out the batteries.
8. Troubleshooting
Table 5
Trouble
Cause
Solution
No display
The power plug is not
inserted in the right place.
The batteries are expired.
Insert the plug in right place.
Use the new batteries.
Low-voltage
indication
The batteries power is low.
Use the new batteries.
Error in weighing
data-reading
The scale didn’t back to
zero before using.
The scale is not calibrated.
Do not put objects on the platform,
press button, and then begin
to weigh.
Re-calibrate.
False calibration
The calibration weight is not
precise.
Use the precise calibration weight.

Page 11 of 11 Anyload-manual-indicator-DD-KC1-en-101
9.Specification
www.anyloadgroup.com
North America Toll Free: 1-855-ANYLOAD (269 5623)
Fax: 1-866-612 9088 info@anyloadgroup.com
© 2015 ANYLOAD All rights reserved. Dimensions and specifications subject to change without notice
Model
DD-KC1
Capacity x Readability
3000e
Parts counting
Standard
Repeatability (std.dev)
0.03%
Linearity
0.03%
T range
To capacity by subtraction
Over range capacity
Capacity + 9d
Stabilization time
2sec
Sensitivity drift (%)
0.01
Operation
temperature
10~40℃
Operation Humidity
≤85%
Power requirements
110 or 220V 50/60HzAC adapter (included); Or 4 AAAbatteries
Other manuals for DD-KC1
1
Table of contents
Other Anyload Accessories manuals