Apogee MI-210 User manual

APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: (435) 792-4700 | FAX: (435) 787-8268 | WEB: APOGEEINSTRUMENTS.COM
Copyright © 2016 Apogee Instruments, Inc.
OWNER’S MANUAL
INFRARED TEMPERATURE METER
Models MI-210, MI-220, MI-230, and MI-2H0

2
CONTENTS
Owner’s Manual ................................................................................................................................................................................................................ 1
Certificate of Compliance .................................................................................................................................................................................... 3
Introduction ............................................................................................................................................................................................................. 4
Sensor Models ......................................................................................................................................................................................................... 5
Specifications........................................................................................................................................................................................................... 6
Deployment and Installation.............................................................................................................................................................................. 7
Operation and Measurement............................................................................................................................................................................. 9
Maintenance and Recalibration.......................................................................................................................................................................12
Troubleshooting and Customer Support......................................................................................................................................................13
Return and Waranty Policy................................................................................................................................................................................15

3
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the manufacturer:
Apogee Instruments, Inc.
721 W 1800 N
Logan, Utah 84321
USA
for the following product(s):
Models: MI-210, MI-220, MI-230, MI-2H0
Type: Infrared Radiometer
The object of the declaration described above is in conformity with the relevant Union harmonization legislation:
2014/30/EU Electromagnetic Compatibility (EMC) Directive
2011/65/EU Restriction of Hazardous Substances (RoHS 2) Directive
Standards referenced during compliance assessment:
EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use – EMC requirements
EN 50581:2012 Technical documentation for the assessment of electrical and electronic products with respect to the
restriction of hazardous substances
Please be advised that based on the information available to us from our raw material suppliers, the products manufactured
by us do not contain, as intentional additives, any of the restricted materials including cadmium, hexavalent chromium, lead,
mercury, polybrominated biphenyls (PBB), polybrominated diphenyls (PBDE).
Further note that Apogee Instruments does not specifically run any analysis on our raw materials or end products for the
presence of these substances, but rely on the information provided to us by our material suppliers.
Signed for and on behalf of:
Apogee Instruments, May 2016
Bruce Bugbee
President
Apogee Instruments, Inc.

4
INTRODUCTION
All objects with a temperature above absolute zero emit electromagnetic radiation. The wavelengths and intensity of
radiation emitted are related to the temperature of the object. Terrestrial surfaces (e.g., soil, plant canopies, water, snow) emit
radiation in the mid infrared portion of the electromagnetic spectrum (approximately 4-50 µm).
Infrared radiometers are sensors that measure infrared radiation, which is used to determine surface temperature without
touching the surface (when using sensors that must be in contact with the surface, it can be difficult to maintain thermal
equilibrium without altering surface temperature). Infrared radiometers are often called infrared thermometers because
temperature is the desired quantity, even though the sensors detect radiation.
Typical applications of infrared radiometers include plant canopy temperature measurement for use in plant water status
estimation, road surface temperature measurement for determination of icing conditions, and terrestrial surface (soil,
vegetation, water, snow) temperature measurement in energy balance studies.
Apogee Instruments MI series infrared meters consist of a handheld meter and a dedicated infrared radiometer, which
includes a thermopile detector, germanium filter, precision thermistor (for detector reference temperature measurement),
and signal processing circuitry mounted in an anodized aluminum housing. All infrared meters also come with a radiation
shield designed to minimize absorbed solar radiation, but still allowing natural ventilation. The radiation shield insulates the
radiometer from rapid temperature changes and keeps the temperature of the radiometer closer to the target temperature.
The separate sensors are potted solid with no internal air space and are designed for continuous temperature measurement
of terrestrial surfaces in indoor and outdoor environments. The meters provide a real-time temperature reading on the LCD
display and come equipped with manual and automatic data logging features for making spot-check measurements or
monitoring temperature over time.

5
SENSOR MODELS
Apogee MI series infrared meters covered in this manual are self-contained and come complete with handheld meter, sensor,
and radiation shield.
The four FOV options and associated model numbers are shown below:
MI-230 MI-220 MI-210 MI-2H0
Sensor model number and serial number are located on a label
on the backside of the handheld meter.

6
SPECIFICATIONS
Calibration Traceability
Apogee MI series infrared temperature meters are calibrated to the temperature of a custom blackbody cone held at multiple
fixed temperatures over a range of radiometer (detector/sensor body) temperatures. The temperature of the blackbody cone
is measured with replicate precision thermistors thermally bonded to the cone surface. The precision thermistors are
calibrated for absolute temperature measurement against a platinum resistance thermometer (PRT) in a constant
temperature bath. The PRT calibration is directly traceable to the NIST.
Spectral Response
MI-210
MI-220
MI-230
MI-2H0
Calibration Uncertainty (-20 to 65 C),
when target and detector temperature
are within 20 C
± 0.3 C ± 0.3 C ± 0.4 C ± 0.3 C
Calibration Uncertainty (-40 to 80 C),
when target and detector temperate are
different by more than 20 C
(see Calibration Traceability below)
± 0.5 C ± 0.5 C ± 0.6 C ± 0.5 C
Measurement Repeatability ± 0.1 C
Stability (Long-term Drift) less than 2 % change in slope per year when germanium filter is maintained in a
clean condition (see Maintenance and Recalibration section below)
Response Time
0.6 s, time for detector signal to reach 95% following a step change; meter
firmware averaging results in a 3 s response for digital output on meter screen
Field of View 22º half angle 18º half angle 14º half angle
32º horizontal
half angle; 13º
vertical half angle
Spectral Range 8 to 14 µm; atmospheric window (see Spectral Response below)
Operating Environment
0 to 50 C; less than 90 % non-condensing relative humidity up to 30 C; less than 70
% non-condensing-relative humidity from 30 to 50 C
Meter Dimensions 126 mm length; 70 mm width; 24 mm height
Sensor Dimensions 23 mm diameter; 60 mm length
Mass 270 g (with radiation shield)
Cable
2 m of four conductor, shielded, twisted-pair wire; additional cable available;
santoprene rubber jacket (high water resistance, high UV stability, flexibility in cold
conditions)
Spectral response of MI series infrared
radiometers. Spectral response (green
line) is determined by the germanium
filter and corresponds closely to the
atmospheric window of 8-14 µm,
minimizing interference from
atmospheric absorption/emission
bands (blue line) below 8 µm and
above 14 µm. Typical terrestrial
surfaces have temperatures that yield
maximum radiation emission within
the atmospheric window, as shown by
the blackbody curve for a radiator at
28 C (red line).

7
DEPLOYMENT AND INSTALLATION
The mounting geometry (distance from target surface, angle of orientation relative to target surface) is determined by the
desired area of surface to be measured. The field of view extends unbroken from the sensor to the target surface. Sensors
must be carefully mounted in order to view the desired target and avoid including unwanted surfaces/objects in the field of
view, thereby averaging unwanted temperatures with the target temperature (see Field of View below). Once mounted, the
green cap must be removed. The green cap can be used as a protective covering for the sensor, when it is not in use.
An Apogee Instruments model AM-210 mounting bracket is recommended for mounting the sensor to a cross arm or pole.
The AM-210 allows adjustment of the angle of the sensor with respect to the target and accommodates the radiation shield
designed for all MI series infrared radiometers.

8
Field of View
The field of view (FOV) is reported as the half-angle of the apex of the cone formed by the target surface (cone base) and the
detector (cone apex), as shown below, where the target is defined as a circle from which 98 % of the radiation detected by
the radiometer is emitted.
Sensor FOV, distance to target, and sensor mounting angle in relation to the target will determine target area. Different
mounting geometries (distance and angle combinations) produce different target shapes and areas, as shown below.
A simple FOV calculator for determining target dimensions based on infrared radiometer model, mounting height, and
mounting angle, is available on the Apogee website: http://www.apogeeinstruments.com/using-your-apogee-instruments-
infrared-radiometer/.

9
OPERATION AND MEASUREMENT
MI series infrared temperature meters are designed with a user-friendly interface allowing quick and easy measurements.
To power the meter, slide the included battery (CR2320) into the battery holder, after removing the battery door
from the meter’s back panel. The positive side (designated by a “+” sign) should be facing out from the meter circuit board.
Press the power button to activate the LCD display. After two minutes of non-activity the meter will revert to sleep
mode and the display will shut off to conserve battery life.
Press the mode button to access the main menu, where the appropriate logging (manual or automatic) is selected
and where the meter can be reset.
Press the sample button to log a reading while taking manual measurements.
Press the up button to make selections in the main menu. This button is also used to view and scroll through the
logged measurements on the LCD display.
Press the down button to make selections in the main menu. This button is also used to view and scroll through the
logged measurements on the LCD display.
Logging: To choose between manual or automatic logging, push the mode button once and use the up/down buttons to
make the appropriate selection (SMPL or LOG). Once the desired mode is blinking, press the mode button two more times to
exit the menu. When in SMPL mode press the sample button to record up to 99 manual measurements (a counter in the
upper right hand corner of the LCD display indicates the total number of saved measurements). When in LOG mode the
meter will power on/off to make a measurement every 30 seconds. Every 30 minutes the meter will average the sixty 30
second measurements and record the averaged value to memory. The meter can store up to 99 averages and will start to
overwrite the oldest measurement once there are 99 measurements. Every 48 averaged measurements (making a 24 hour
period), the meter will also record an average daily value.
Reset: To reset the meter, in either SMPL or LOG mode, push the mode button twice (RUN should be blinking), then while
pressing the down button, press the mode button once. This will erase all of the saved measurements in memory, but only
The LCD display consists of the total number of logged
measurements in the upper right hand corner, the real-time
temperature reading in the center, and the selected menu options
along the bottom.

10
for the selected mode. That is, performing a reset when in SMPL mode will only erase the manual measurements and
performing a reset when in LOG mode will only erase the automatic measurements.
Review/Download Data: Each of the logged measurements in either SMPL or LOG mode can be reviewed on the LCD
display by pressing the up/down buttons. To exit and return to the real-time readings, press the sample button. Note that the
average daily values are not accessible through the LCD and can only be viewed by downloading to a computer.
Downloading the stored measurements will require the AC-100 communication cable and software (sold separately). The
meter outputs data using the UART protocol and requires the AC-100 to convert from UART to USB, so standard USB cables
will not work. Set up instructions and software can be downloaded from the Apogee website
(http://www.apogeeinstruments.com/ac-100-communcation-cable/).
Sensor Calibration
Apogee MI series infrared temperature meters are calibrated in a temperature controlled chamber that houses a custom-built
blackbody cone (target) for the radiation source. During calibration, infrared radiometers (detectors) are held in a fixture at
the opening of the blackbody cone, but are thermally insulated from the cone. Detector and target temperature are
controlled independently. At each calibration set point, detectors are held at a constant temperature while the blackbody
cone is controlled at temperatures below (12 C), above (18 C), and equal to the detector temperature. The range of detector
temperatures is -15 C to 45 C (set points at 10 C increments). Data are collected at each detector temperature set point, after
detectors and target reach constant temperatures.
Emissivity Correction
Appropriate correction for surface emissivity is required for accurate surface temperature measurements. The simple (and
commonly made) emissivity correction, dividing measured temperature by surface emissivity, is incorrect because it does not
account for reflected infrared radiation.
The radiation detected by an infrared radiometer includes two components: 1. radiation directly emitted by the target
surface, and 2. reflected radiation from the background. The second component is often neglected. The magnitude of the
two components in the total radiation detected by the radiometer is estimated using the emissivity (ε) and reflectivity (1 - ε)
of the target surface:
( )
Backgroundet
argTSensor
E1
E
E⋅ε
−+⋅
ε=
(1)
where ESensor is radiance [W m-2 sr-1] detected by the radiometer, ETarget is radiance [W m-2 sr-1] emitted by the target surface,
EBackground is radiance [W m-2 sr-1] emitted by the background (when the target surface is outdoors the background is generally
the sky), and ε is the ratio of non-blackbody radiation emission (actual radiation emission) to blackbody radiation emission at
the same temperature (theoretical maximum for radiation emission). Unless the target surface is a blackbody (ε = 1; emits
and absorbs the theoretical maximum amount of energy based on temperature), Esensor will include a fraction (1 – ε) of
reflected radiation from the background.
Since temperature, rather than energy, is the desired quantity, Eq. (1) can be written in terms of temperature using the
Stefan-Boltzmann Law, E = σT4(relates energy being emitted by an object to the fourth power of its absolute temperature):

11
()
4
Background
4
etargT
4
Sensor T1TT ⋅σ⋅ε−+⋅σ⋅ε=⋅σ
(2)
where TSensor [K] is temperature measured by the infrared radiometer (brightness temperature), TTarget [K] is actual temperature
of the target surface, TBackground [K] is brightness temperature of the background (usually the sky), and σ is the Stefan-
Boltzmann constant (5.67 x 10-8 W m-2 K-4). The power of 4 on the temperatures in Eq. (2) is valid for the entire blackbody
spectrum.
Rearrangement of Eq. (2) to solve for TTarget yields the equation used to calculate the actual target surface temperature (i.e.,
measured brightness temperature corrected for emissivity effects):
( )
4
4
Background
4
Sensor
etargT
T
1T
Tε
⋅
ε−−
=
. (3)
Equations (1)-(3) assume an infinite waveband for radiation emission and constant ε at all wavelengths. These assumptions
are not valid because infrared radiometers do not have infinite wavebands, as most correspond to the atmospheric window
of 8-14 µm, and ε varies with wavelength. Despite the violated assumptions, the errors for emissivity correction with Eq. (3) in
environmental applications are typically negligible because a large proportion of the radiation emitted by terrestrial objects
is in the 8-14 µm waveband (the power of 4 in Eqs. (2) and (3) is a reasonable approximation), ε for most terrestrial objects
does not vary significantly in the 8-14 µm waveband, and the background radiation is a small fraction (1 – ε) of the measured
radiation because most terrestrial surfaces have high emissivity (often between 0.9 and 1.0). To apply Eq. (3), the brightness
temperature of the background (TBackground) must be measured or estimated with reasonable accuracy. If a radiometer is used
to measure background temperature, the waveband it measures should be the same as the radiometer used to measure
surface brightness temperature. Although the ε of a fully closed plant canopy can be 0.98-0.99, the lower ε of soils and other
surfaces can result in substantial errors if ε effects are not accounted for.

12
MAINTENANCE AND RECALIBRATION
Blocking of the optical path between the target and detector, often due to moisture or debris on the filter, is a common
cause of inaccurate measurements. The filter in the separate sensor of the MI series infrared temperature meters is inset in an
aperture, but can become partially blocked in four ways:
1. Dew or frost formation on the filter.
2. Salt deposit accumulation on the filter, due to evaporating irrigation water or sea spray. This leaves a thin white film
on the filter surface. Salt deposits can be removed with a dilute acid (e.g., vinegar). Salt deposits cannot be
removed with solvents such as alcohol or acetone.
3. Dust and dirt deposition in the aperture and on the filter (usually a larger problem in windy environments). Dust
and dirt are best removed with deionized water, rubbing alcohol, or in extreme cases, acetone.
4. Spiders/insects and/or nests in the aperture leading to the filter. If spiders/insects are a problem, repellent should
be applied around the aperture entrance (not on the filter).
Clean inner threads of the aperture and the filter with a cotton swab dipped in the appropriate solvent. Never use an
abrasive material on the filter. Use only gentle pressure when cleaning the filter with a cotton swab, to avoid scratching
the outer surface. The solvent should be allowed to do the cleaning, not mechanical force.
It is recommended that infrared radiometers be recalibrated every two years. See the Apogee webpage for details regarding
return of sensors for recalibration (http://www.apogeeinstruments.com/tech-support-recalibration-repairs/).

13
TROUBLESHOOTING AND CUSTOMER SUPPORT
Verify Functionality
Pressing the power button should activate the LCD and provide a real-time temperature reading. Direct the sensor toward a
surface with a temperature significantly different than the detector. The real-time reading should respond accordingly
(decreased reading when pointing at a low temperature surface and increased reading when pointing at a high temperature
surface). Placing a piece of tinfoil over the front of the sensor should force the reading to ambient temperature.
Battery Life
When the meter is maintained properly the coin cell battery (CR2320) should last for many months, even after continuous
use. The low battery indicator will appear in the upper left hand corner of the LCD display when the battery voltage drops
below 2.8 V DC. The meter will still function correctly for some time, but once the battery is drained the pushbuttons will no
longer respond and any logged measurements will be lost.
Pressing the power button to turn off the meter will actually put it in sleep mode, where there is still a slight amount of
current draw. This is necessary to maintain the logged measurements in memory. Therefore, it is recommended to remove
the battery when storing the meter for many months at a time, in order to preserve battery life.
Low-Battery Error after Battery Replacement
A master reset will usually correct this error, please see the master reset section for details and cautions. If a master reset does
not remove the low battery indicator, please double check that the voltage of your new batter is above 2.8 V, this is the
threshold for the indicator to turn on.
Master Reset
If a meter ever becomes non-responsive or experiences anomalies, such as a low battery indicator even after replacing the
old battery, a master reset can be performed that may correct the problem. Note that a master reset will erase all logged
measurements from memory.
First press the power button so that the LCD display is activated. While still powered, slide the battery out of the holder,
which will cause the LCD display to fade out. After a few seconds, slide the battery back into the holder. The LCD display will
flash all segments and then show a revision number (e.g. “R1.0”). This indicates the master reset was performed and the
display should return to normal.
Error Codes and Fixes
Error codes will appear in place of the real-time reading on the LCD display and will continue to flash until the problem is
corrected. Contact Apogee if the following fixes do not rectify the problem.
Err 1: battery voltage out of range. Fix: replace CR2320 battery and perform master reset.
Err 2: sensor voltage out of range. Fix: perform master reset.
Err 3: not calibrated. Fix: perform master reset.
Err 4: CPU voltage below minimum. Fix: replace CR2320 battery and perform master reset.
Modifying Cable Length
Although it is possible to splice additional cable to the separate sensor of the MI series, note that the cable wires are soldered
directly into the circuit board of the meter. Care should be taken to remove the back panel of the meter in order to access the
board and splice on the additional cable, otherwise two splices would need to be made between the meter and sensor head.

14
See Apogee webpage for further details on how to extend sensor cable length: (http://www.apogeeinstruments.com/how-
to-make-a-weatherproof-cable-splice/).

15
RETURN AND WARANTY POLICY
RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as the product is in new condition (to be
determined by Apogee). Returns are subject to a 10 % restocking fee.
WARRANTY POLICY
What is Covered
All products manufactured by Apogee Instruments are warranted to be free from defects in materials and craftsmanship for a
period of four (4) years from the date of shipment from our factory. To be considered for warranty coverage an item must be
evaluated either at our factory or by an authorized distributor.
Products not manufactured by Apogee (spectroradiometers, chlorophyll content meters) are covered for a period of one (1)
year.
What is Not Covered
The customer is responsible for all costs associated with the removal, reinstallation, and shipping of suspected warranty
items to our factory.
The warranty does not cover equipment that has been damaged due to the following conditions:
1. Improper installation or abuse.
2. Operation of the instrument outside of its specified operating range.
3. Natural occurrences such as lightning, fire, etc.
4. Unauthorized modification.
5. Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine recalibration of sensors/meters is considered part of
proper maintenance and is not covered under warranty.
Who is Covered
This warranty covers the original purchaser of the product or other party who may own it during the warranty period.
What We Will Do
At no charge we will:
1. Either repair or replace (at our discretion) the item under warranty.
2. Ship the item back to the customer by the carrier of our choice.
Different or expedited shipping methods will be at the customer’s expense.
How To Return An Item
1. Please do not send any products back to Apogee Instruments until you have received a Return Merchandise Authorization
(RMA) number from our technical support department by calling (435) 792-4700 or by submitting an online RMA form at
www.apogeeinstruments.com/tech-support-recalibration-repairs/. We will use your RMA number for tracking of the service
item.

16
2. Send all RMA sensors and meters back in the following condition: Clean the sensor’s exterior and cord. Do not modify the
sensors or wires, including splicing, cutting wire leads, etc. If a connector has been attached to the cable end, please include
the mating connector – otherwise the sensor connector will be removed in order to complete the repair/recalibration.
3. Please write the RMA number on the outside of the shipping container.
4. Return the item with freight pre-paid and fully insured to our factory address shown below. We are not responsible for any
costs associated with the transportation of products across international borders.
5. Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of
operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will
repair or replace the items free of charge. If it is determined that your product is not covered under warranty, you will be
informed and given an estimated repair/replacement cost.
Apogee Instruments, Inc.
721 West 1800 North Logan, UT
84321, USA
OTHER TERMS
The available remedy of defects under this warranty is for the repair or replacement of the original product, and Apogee
Instruments is not responsible for any direct, indirect, incidental, or consequential damages, including but not limited to loss
of income, loss of revenue, loss of profit, loss of wages, loss of time, loss of sales, accruement of debts or expenses, injury to
personal property, or injury to any person or any other type of damage or loss.
This limited warranty and any disputes arising out of or in connection with this limited warranty ("Disputes") shall be
governed by the laws of the State of Utah, USA, excluding conflicts of law principles and excluding the Convention for the
International Sale of Goods. The courts located in the State of Utah, USA, shall have exclusive jurisdiction over any Disputes.
This limited warranty gives you specific legal rights, and you may also have other rights, which vary from state to state and
jurisdiction to jurisdiction, and which shall not be affected by this limited warranty. This warranty extends only to you and
cannot by transferred or assigned. If any provision of this limited warranty is unlawful, void or unenforceable, that provision
shall be deemed severable and shall not affect any remaining provisions. In case of any inconsistency between the English
and other versions of this limited warranty, the English version shall prevail.
This warranty cannot be changed, assumed, or amended by any other person or agreement.
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA
TEL: (435) 792-4700 | FAX: (435) 787-8268 | WEB: APOGEEINSTRUMENTS.COM
Copyright © 2016 Apogee Instruments, Inc.
This manual suits for next models
3
Table of contents
Other Apogee Measuring Instrument manuals

Apogee
Apogee SI-111 User manual

Apogee
Apogee SL-510 User manual

Apogee
Apogee Quantum User manual

Apogee
Apogee MQX Series User manual

Apogee
Apogee SP-110 User manual

Apogee
Apogee SN-500 User manual

Apogee
Apogee SI-411 User manual

Apogee
Apogee MC-100 User manual

Apogee
Apogee SI-411 User manual

Apogee
Apogee SI series User manual

Apogee
Apogee MC-100 User manual

Apogee
Apogee 2349 DLTM4 User manual

Apogee
Apogee MQ-100 User manual

Apogee
Apogee SP-110 User manual

Apogee
Apogee MU-100 User manual

Apogee
Apogee SP-420 User manual

Apogee
Apogee 21054010 User manual

Apogee
Apogee SN-500 User manual

Apogee
Apogee MU-100 User manual

Apogee
Apogee MC-100 User manual