2. Check list
This material is included in your shipment.
One “Heater Jacket Control Unit Communication”(HJCU-COM).
One “Heater Jacket” (HJSA ...-...or one HJSB ...-...)
One “Extension cable kit”, includes two cables each 2,5 meter (Temp & Heat cable)
One “Alarm” connector, “Alarm output” 4-way female
One “Mains” connector, Mains input power 3-way female
One “Communication” connector, Communication RS-485 4-way male
Two Spare fuses 5X20mm, one T 4,0A L/250VAC and one T 8,0A L/250VAC
One “Heater Jacket Control Unit” manual, model HJCU-COM
One “Heater Jacket” manual, model HJSA & HJSB
Three “E5CN Temperature Controller” manuals, six languages (ENG - JPN, DE - IT, FR - ES)
One “E5CN Temperature Controller Communication Event Input” manuals, six languages (ENG, DE, FR, JPN, IT, ES)
One “E5CN Temperature Controller Option Units” manuals (E53-CNQO3N), six languages (ENG, DE, FR, JPN, IT, ES)
IF TCViews are ordered this also includes
One TCViews Software
One TCViews Manual
3. Introduction Heater Jacket Control Unit
Heater Jacket Control Unit (HJCU) is designed for use together with one Heater Jacket (HJ).
HJCU monitors and controls the Heater Jacket temperature. The temperature on HJ1can easily
be altered. HJCU has two alarm outputs, upper limit and lower limit.
HJCU-COM communicates with RS-485.
4. Installing Heater Jacket Control Unit
Check the main fuse, make sure it’s of the correct value in relation to the mains voltage and conforms to
specifications for the HJ. See “Technical specification”.
Mount the HJCU on a vertical wall with screws intended for the supporting surface. Use the screw holes
under the cover screws.
Mount the HJCU so that the mains switch and regulator are not blocked.
Connect the HJ to the HJCU using “Extension cable kit”.
If used, connect the alarm cable to the HJCU. See “Alarm description” & “Technical specification”.
Fit the HJ as described in the manual “Heater Jacket HJSA & HJSB”
Connect Mains to the HJCU using the correct supply cable to an earthed electrical outlet socket.
Requisite knowledge is required.
Press the power switch. The equipment is now operational.
The default factory setting of SV2is 80ºC. SV can be changed within the range 30ºC to 140ºC
Should an error message be displayed on the regulator see the manual Omron E5CN Temperature controller.
In the event of a fault see “Fault tracing”
1. HJ = Heater Jacket, HJSA...-..., HJSB...-…
2. SV = Set value, Set working temperature.