7 Maintenance and cleaning
TheGermlyser®AMrequiresnomaintenancewithinthespeciedoperatingtime.Theoutsideofthelter
canbecleanedusingcommonsurfacedisinfectants.
WhenreplacingtheGermlyser®AM,asmallamountofunlteredwaterwillowout.Overtime,thiswill
leadtodepositsonandintheadapters.Aquafreerecommendsreplacementoftheadaptersafteraperiod
ofuseevery6to12months.
Inthecaseofadeviationfromthespecicationsoradamageofthepackage,theproductshouldbe
returnedtoAquafree,alongwiththeDefectReportformdeliveredwiththeproduct(formdownloadsatthe
servicepageunderwww.aqua-free.com).
GEBR-ANW-AM-GB-08/2021-V004
Germlyser®AM
Headquarters:
Aqua free GmbH
Winsbergring31
22525Hamburg
Germany
Phone: +49(0)40-46899990
Fax: +49(0)40-46899999
www.aqua-free.com
Aqua free Solutions Ltd.
TheMediaCentre
7NorthumberlandStreet
Hudderseld,HD11RL,GreatBritain
Phone: +44(0)1484483045
9 Returning
TheGermlyser®reusableltersmustbesentbacktoAquafreewithintheagreedtimeforreprocessing
(formdownloadsattheservicepageunderwww.aqua-free.com).
8 Storage
TheGermlyser®ltermustbeproperlystoredinadryandhygienicenvironmentatatemperaturebetween
10and50°C.
Table of contents
1Product
2Certicates/Declarationofconformity
3Labelling
4 Preparationofcommissioninganduseofproduct
5 Disassemblyofusedlter/lterchange
6 Accessories
7 Maintenance/cleaning
8 Storage
9 Return
1 Product
1.1Productdescription
The Germlyser® AM is a -marked, point-of-use, sterile tap filter for use as a mobile shower in ICU
wards to prevent bacterial nosocomial infectious diseases caused by Pseudomonas aeruginosa,
Legionella pneumophila, Stenotrophomonas maltophilia or Staphylococcus aureus among
immunosuppressed and immunocompromised patients. To achieve this intended use, a specific
combination of retention of these bacteria in the filter membrane structure and specific measurements to
prevent retrogarde contamination with these bacteria on the filter outlet, are necessary.
1.2Technicaldata
Size: L113mmxØ43mm
7logstepsBrev. dim.
0.2µm
Retention*:
Poresize:
Operatingtime: Dependingontheapplication
Chlorineresistance**:<10ppm
Minimumowrate: Adjustableupto5l/min.
at5bar,at25°C
Max.operatingpressure***: 5barat50°C
Max.operatingtemperature: 60°C(at70°C<10min.
overoperatingtime)
Connection: Quick-releasecoupling
**Continuousdosageof≤10ppmoveroperatingtime;short-term(1h)highdosage(400,000ppm)
forchemicaldisinfection.
***TheGermlyser®AMisnotsuitableforconnectiontolow-pressuresystems(e.g.boilers).
* Measured in accordance with the international ASTM F838-15a standard - Standard Test Method for
Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration