AquaMaster Tools P50 Pro User manual

User manual
Handleiding
Benutzerhandbuch
pH Temp meter
P50 Pro

About Aqua Master Tools
Established in 2018
Aqua Master Tools is the Dutch supplier of user-friendly
and reliable pH, EC, and temperature meters. Our meters
are used for swimming pools, aquariums, Jacuzzi’s,
ponds, and liquid plant nutrition.
We understand the importance of convenient, accurate
and high-quality meters - and we are on a continuous
quest to innovate our products. Aqua Master Tools’ wide
range of products are customised to our user’s needs.
Built to last products
Fully calibrated meters
Unprecedented quality
Satisfaction guaranteed
Easily replaceable electrodes

Table of
Contents
Handleiding Nederlands 4 - 7
User manual English 8 - 11
Benutzerhandbuch Deutsch 12 - 15

4NL
1. Dop
2. Aan/uit knop
3. Celsius/Fahrenheit knop
4. Display
5. CAL knop
6. Elektrode
7. Draaidop
8. Beschermkapje
Nederlands
1
2
5
3
6
7
4
8

5NL
Bediening en gebruik
1. Laat uw pH meter voor het eerste gebruik 12 uur weken in
kraanwater.
2. Verwijder het beschermkapje.
3. Reinig de elektrode zorgvuldig met gedistilleerd water en droog
deze af.
4. Druk op de Aan/uit toets zodat de meter opstart.
5. Steek de meter rechtop in de oplossing tot de rand en roer zachtjes.
Wacht enkele seconden tot de juiste waarde op het display verschijnt.
6. Druk op de C/F toets om te schakelen tussen Celsius en Fahrenheit.
7. Spoel na gebruik de elektrode schoon met water en druk op de Aan/
uit toets om de meter uit te schakelen.
Wij adviseren om de meter rechtop te bewaren met KCI bewaarvloeistof in het
beschermkapje.
pH meter kalibreren
1. Druk op de Aan/uit toets zodat de meter ingeschakeld is.
2. Reinig de elektrode zorgvuldig met gedistilleerd water en droog
deze af.
3. Steek de meter rechtop in pH bufferoplossing pH 7,01 (zorg dat
deze 25°C is) en roer zachtjes.
4. Wacht tot de waarde op het display stabiel is.
5. Houd de CAL toets 5 seconden lang ingedrukt en laat deze daarna
los. De waarde pH 7,0 zal nu gaan knipperen op het display.
Wanneer de meter weer overspringt naar de normale modus is het
kalibreren klaar.
6. Reinig de elektrode zorgvuldig met gedistilleerd water en droog
deze af.
7. Herhaal het proces met pH 4,01 bufferoplossing.
8. Reinig na dit proces de meter zorgvuldig voor u deze opbergt.
• Als de elektrode is vervangen.
• Als de meter langer dan een maand geleden voor het laatst
is gekalibreerd.
• Als de meter heel vaak gebruikt wordt.
• Als er hoge precisie nodig is.
Wanneer kalibreren?

6NL
Elektrode vervangen
1. Draai het schroefdraad van de afdichtingsring van het onderste deel
van de meter goed los.
2. Verwijder de kapotte/oude elektrode door er zachtjes aan te trekken.
3. Plaats de nieuwe elektrode zorgvuldig met de uitsparing op de juiste
plaats zodat de connectoren goed aansluiten en de meter of de
elektrode niet beschadigen.
4. Controleer ook of het afdichtingsrubber goed aansluit.
5. Draai het schroefdraad van de afdichtingsring goed aan.
6. Kalibreer de meter zoals beschreven staat in deze handleiding.
7. De meter is nu weer klaar voor gebruik.
Specificaties
Bereik pH: 0,0 ~ 14,0 pH
Temperatuur: -50°C ~ 70°C (-58°F ~ 158°F)
Resolutie pH: 0,1 pH
Temperatuur: 0,1 ºC (0,2 ºF)
Accuraatheid
pH: ±0,1 pH
Temperatuur: ±1°C (±2°F)
Temperatuurcompensatie: 0°C tot 50°C
(32°F tot 122°F)
Werktemperatuur 0°C ~ 50°C (32°F ~ 122°F)
Batterij indicator Ja
Automatisch uitschakelen Na 8 minuten
Voeding 3 × 1,5 Volt AG-13 knoopcel (meegeleverd)
Afmeting 220 x 40 x 40 mm
Gewicht 102 gram

7NL
Waarschuwing
• Zorg ervoor dat er tijdens het kalibreren geen luchtbelletjes aanwezig
zijn in de vloeistof. Dit kan de metingen beïnvloeden.
• Zorg ervoor dat de kalibratie vloeistof 25 oC is tijdens het kalibreren
van de meter. Indien dit niet het geval is kan dit het resultaat van de
kalibratie beïnvloeden.
• Bewaar de elektrode nooit in gedistilleerd water.
• De elektrode NIET onderdompelen in olie, eiwitten of gesuspendeerde
vaste stoffen die een laag op de glazen bol achterlaten.
• De elektrode NIET weken of spoelen in OO (Omgekeerde Osmose),
gedistilleerd of gedeïoniseerd water. Zuiver water wijzigt de chemie in
de referentievloeistof, waardoor deze niet meer werkt.
• Een koude elektrode NIET in hete vloeistof (of vice versa) plaatsen.
Door plotselinge temperatuursveranderingen kan het glas breken en
kan de pen permanent beschadigd raken. Gebroken glas valt niet onder
de garantie.
• De glaselektrode nergens tegenaan stoten. Anders kunnen de glazen
bol aan de buitenkant of het interne glazen buisje breken. Gebroken glas
valt niet onder de garantie.
• Plaats de meter bij voorkeur niet in de buurt van voorschakel- of andere
elektrische apparatuur. Spanningsvelden kunnen de meetresultaten
beïnvloeden.
• Als tijdens het kalibreren de pH waarde langzaam blijft stijgen of
dalen en niet stabiel wordt, vervangt u de batterijen en kalibreert u
opnieuw. Slechte batterijen of een laag spanningsniveau kunnen de
meetresultaten beïnvloeden.
Garantie
Deze meter heeft garantie voor alle materiële defecten en productiefouten
gedurende een periode van één jaar vanaf de datum van aankoop. Als tijdens
deze periode de reparatie of vervanging van onderdelen vereist is en de
schade niet te wijten is aan een foutieve bediening door de gebruiker, stuur
dan de onderdelen naar de dealer of naar ons en de reparatie zal kosteloos
plaatsvinden. Zie de waarschuwing voor uitsluitingen.
Batterijen vervangen
Als de batterij-indicator rechtsboven in het scherm aan gaat, of als het
scherm vaag wordt, moeten de batterijen worden vervangen. Draai hiervoor
de bovenkant los en vervang de 3 x 1,5 Volt AG-13 knoopcelbatterijen.

8EN
1. Cap
2. On/off button
3. Celsius/Fahrenheit button
4. Display
5. CAL button
6. Electrode
7. Screw cap
8. Protective cover
1
2
5
3
6
7
4
8
English

9EN
Operation and use
1. Prior first usage - soak your pH meter in tap water for 12 hours.
2. Remove the protective cover.
3. Gently clean the electrode with distilled water and dry it off.
4. Press the On/off button to start the meter.
5. Insert the meter upright into the solution up to the edge and stir
gently. Wait a few seconds for the correct value to appear on the
display.
6. Press the C/F button to switch between Celsius and Fahrenheit.
7. Rinse the electrode with water after use and press the On/off
button to turn off the meter.
We recommend storing the meter in an upright position with KCI storage
solution in the protective cover.
Calibrating the pH meter
1. Press the On/off button to start the meter.
2. Gently clean the electrode with distilled water and dry it off.
3. Insert the meter in an upright position in pH buffer solution pH 7.01
(make sure this is 25 oC) and stir gently.
4. Wait until the value on the display is stable.
5. Press and hold the CAL button for 5 seconds. Then release the
button. The value pH 7.0 will now blink on the display. As soon as the
meter switches back to the standard mode, calibration is complete.
6. Gently clean the electrode with distilled water and dry it off.
7. Repeat the process with pH 4.01 buffer solution.
8. Gently clean the meter before storing it.
• If the electrode has been replaced.
• If the meter was last calibrated over a month ago.
• If the meter is used on a regular basis.
• If high precision is needed.
When is calibration required?

10EN
Replacing the electrode
1. Properly unscrew the sealing ring in the lower part of the meter.
2. Remove the defected/old electrode by gently pulling it.
3. Carefully place the new electrode with the recess in the correct
position so that the connectors properly connect and do not damage
the meter or electrode.
4. Check that the rubber seal is properly placed.
5. Tightly tighten the sealing ring.
6. Calibrate the meter as described in this manual.
7. The meter is now ready for use.
Specifications
Range pH: 0,0 ~ 14,0 pH
Temperature: -50°C ~ 70°C (-58°F ~ 158°F)
Resolution pH: 0,1 pH
Temperature: 0,1 ºC (0,2 ºF)
Accuracy
pH: ±0,1 pH
Temperature: ±1°C (±2°F)
Temperature compensation: 0°C - 50°C
(32°F - 122°F)
Operation temperature 0°C ~ 50°C (32°F ~ 122°F)
Battery indicator Yes
Auto. power-off After 8 minutes
Battery 3 × 1,5 Volt AG-13 button cell (included)
Dimensions 220 x 40 x 40 mm
Weight 102 grams

11EN
Warning
• Always make sure that there are no air bubbles present in the solution
during calibration. This could affect the measurements.
• Make sure the calibration solution is 25°C when calibrating the meter.
Failure to do so may affect the result of the calibration.
• Never store the electrode in distilled water.
• DO NOT immerse the electrode in oil, proteins or suspended solids that
leave a layer on the glass bulb.
• DO NOT soak or rinse in RO (Reverse Osmosis), distilled or de-ionised
water. Pure water alters the chemistry of the reference liquid so that it
no longer works.
• DO NOT place a cold electrode in hot liquid (or vice versa). Sudden
changes in temperature may break the glass and permanently damage
the pen. Broken glass is not covered by the warranty.
• Make sure not to bump the glass electrode into anything, because this
might break the glass bulb on the outside of the internal glass tube
Broken glass is not covered by the warranty.
• Avoid placing the meter near ballasts or any other electrical appliances.
Force fields may affect the readings.
• If during calibration the pH value keeps rising or dropping slowly and
does not stabilise, please replace the batteries and calibrate the meter.
Bad batteries or low power levels may influence the measurement
results.
Replacing the batteries
If the battery indicator at the top right of the display is activated, or if the
display dims, please replace the batteries. To do so, unscrew the top and
replace the 3 x 1.5 Volt AG-13 button cell batteries.
Warranty
These instruments are warranted from all defects in material and
manufacturing for a period of one year from the date of purchase. If during
this period any parts need reparation or replacement and the damage is
not due to incorrect operation by the user, please send the parts to the
dealer or to us and the repair will be free of charge. Please refer to the
exclusion warning.

12DE
1. Verschluss
2. Ein-/Aus-Schalter
3. Celsius/Fahrenheit-Taste
4. Display
5. CAL Taste
6. Elektrode
7. Drehverschluss
8. Schutzkappe
Deutsch
1
2
5
3
6
7
4
8

13DE
Bedienung und Betrieb
1. Weichen Sie Ihr pH-Messgerät vor dem ersten Gebrauch 12 Stunden
lang in Leitungswasser ein.
2. Entfernen Sie die Schutzkappe.
3. Reinigen Sie die Elektrode sorgfältig mit destilliertem Wasser und
trocknen Sie diese.
4. Drücken Sie zum Einschalten des Messgeräts auf den Ein-/Aus-
Schalter.
5. Halten Sie das Messgerät senkrecht und bis zum Rand in die Lösung
und lassen Sie es leicht kreisen. Warten Sie einige Sekunden, bis die
Werte korrekt auf dem Display angezeigt werden.
6. Mit der C/F-Taste wechseln Sie zwischen den Einheiten Celsius und
Fahrenheit.
7. Spülen Sie die Elektrode nach Gebrauch mit Wasser ab und drücken Sie
zum Ausschalten des Messgeräts auf den Ein-/Aus-Schalter.
Wir empfehlen, das Messgerät aufrecht und mit KCl-Lösung in der Schutzkappe
aufzubewahren.
pH-Meter kalibrieren
1. Drücken Sie zum Einschalten des Meters auf den Ein-/Aus-Schalter.
2. Reinigen Sie die Elektrode sorgfältig mit destilliertem Wasser und
trocknen Sie diese.
3. Halten Sie das Messgerät senkrecht in die pH-Pufferlösung pH 7,01
(diese muss 25 oC warm sein) und lassen Sie es leicht kreisen.
4. Warten Sie, bis der Wert auf dem Display stabil bleibt.
5. Halten Sie die CAL-Taste 5 Sekunden lang gedrückt und lassen Sie
diese dann los. Auf dem Display blinkt dann der Wert pH 7.0. Sobald
das Messgerät wieder in den normalen Modus umschaltet, ist der
Kalibriervorgang abgeschlossen.
6. Reinigen Sie die Elektrode sorgfältig mit destilliertem Wasser und
trocknen Sie diese.
7. Wiederholen Sie dieses Verfahren mit der Pufferlösung pH 4,01.
8. Reinigen Sie das Messgerät nach dem Kalibrieren gründlich, bevor Sie
es wieder wegräumen.
• Wenn die Elektrode ausgetauscht wurde.
• Wenn die letzte Kalibrierung des Messgeräts mehr als einen Monat
zurückliegt.
• Wenn das Messgerät sehr oft gebraucht wird.
• Wenn höchste Präzision erforderlich ist.
Wann kalibriert werden muss:

14DE
Elektrode ersetzen
1. Lösen Sie die Elektrode durch Drehen am Gewinde des
Dichtungsrings am unteren Teil des Messgeräts.
2. Ziehen Sie die defekte/alte Elektrode vorsichtig ab.
3. Setzen Sie die neue Elektrode sorgfältig und mit richtig
ausgerichtetem Spalt ein, sodass die Kontaktstellen gut anschließen
und weder der Meter noch die Elektrode beschädigt wird.
4. Überprüfen Sie auch, ob die Gummidichtung gut sitzt.
5. Drehen Sie das Gewinde am Dichtungsring wieder gut fest.
6. Kalibrieren Sie das Multimeter gemäß der Beschreibung in diesem
Handbuch.
7. Das Multimeter ist nun wieder einsatzbereit.
Technische Daten
Messbereich pH: 0,0 ~ 14,0 pH
Temperatur: -50°C ~ 70°C (-58°F ~ 158°F)
Auflösung pH: 0,1 pH
Temperatur: 0,1 ºC (0,2 ºF)
Messgenauigkeit
pH: ±0,1 pH
Temperatur: ±1°C (±2°F)
Temperaturausgleich: 0°C - 50°C
(32°F - 122°F)
Betriebstemperatur 0°C ~ 50°C (32°F ~ 122°F)
Batterie anzeige Ja
Abschaltautomatik 8 minuten
Batterie 3 × 1,5 Volt AG-13 Knopfzelle
(einschließlich)
Maße 220 x 40 x 40 mm
Gewicht 102 Gramm

15DE
Warnhinweis
• Achten Sie darauf, dass während des Kalibriervorgangs keine
Luftbläschen in der Flüssigkeit sind. Dies kann die Messungen
beeinflussen.
• Sorgen Sie dafür, dass die Temperatur der Kalibrierflüssigkeit während
des Kalibriervorgangs bei 25°C liegt. Ist dies nicht der Fall, kann dies das
Resultat der Kalibrierung beeinflussen.
• Bewahren Sie die Elektrode nicht in destilliertem Wasser auf.
• Tauchen Sie die Elektrode NICHT in Öl oder Lösungen mit Eiweißen oder
Schwebstoffen, die einen Film auf der Glasmembran hinterlassen.
• VERMEIDEN Sie es, die Elektrode in Umkehrosmosewasser, destilliertem
oder deionisiertem Wasser zu spülen oder einzutauchen. Reines Wasser
verändert die Chemie in der Bezugsflüssigkeit, die dadurch nicht mehr
funktioniert.
• Bringen Sie eine kalte Elektrode NICHT in Kontakt mit heißer Flüssigkeit
(oder umgekehrt). Durch abrupte Temperaturschwankungen kann das
Glas brechen und der Elektrodenstift unwiderruflich beschädigt werden.
Zerbrochenes Glas fällt nicht unter die Garantie.
• Stoßen Sie mit der Glaselektrode nirgends an. Sonst zerbricht unter
Umständen die äußere Glasmembran oder das innere Glasröhrchen.
Zerbrochenes Glas fällt nicht unter die Garantie.
• Installieren Sie das Messgerät vorzugsweise nicht in der Nähe von
Vorschaltgeräten oder anderen Elektrogeräten. Spannungsfelder können
die Messresultate beeinflussen.
• Wenn sich der pH-Wert während des Kalibrierens nicht stabilisiert,
sondern langsam weiter steigt oder sinkt, ersetzen Sie die Batterien und
führen Sie den Kalibriervorgang erneut durch. Minderwertige Batterien
oder eine zu geringe Spannung können die Messresultate beeinflussen.
Batterien ersetzen
Wenn die Batterieanzeige oben rechts im Bildschirm aufleuchtet oder wenn
das Display schwächer wird, müssen die Batterien ersetzt werden. Drehen Sie
dazu die Oberseite ab und ersetzen Sie die drei 1,5-Volt-Knopfzellbatterien
AG-13.
Garantie
Für dieses Multimeter gilt eine einjährige Garantie auf alle Material- und
Produktionsfehler ab Kaufdatum. Wenn in diesem Zeitraum Komponenten
repariert oder ersetzt werden müssen und der Schaden nicht auf
fehlerhafte Verwendung des Bedieners zurückzuführen ist, schicken Sie die
entsprechenden Komponenten zu einem Vertragshändler oder zu uns ein.
Die Reparatur erfolgt dann kostenlos. Siehe den Warnhinweis bezüglich
ausgeschlossener Garantieleistungen.

www.aquamastertools.com
Aqua Master Tools
Ambachtsweg 55c
1271 AL Huizen
The Netherlands
This manual suits for next models
1
Table of contents
Languages:
Other AquaMaster Tools Measuring Instrument manuals

AquaMaster Tools
AquaMaster Tools Combo pen P160 Pro User manual

AquaMaster Tools
AquaMaster Tools S300 Pro 2 User manual

AquaMaster Tools
AquaMaster Tools Combo pen P110 Pro User manual

AquaMaster Tools
AquaMaster Tools E300 Pro User manual

AquaMaster Tools
AquaMaster Tools Combo Pen P150 Pro User manual

AquaMaster Tools
AquaMaster Tools Combo meter P700 Pro 2 User manual

AquaMaster Tools
AquaMaster Tools E5O Pro User manual

AquaMaster Tools
AquaMaster Tools Combo Pen P100 Pro User manual