AquaMaster Tools S300 Pro 2 User manual

User manual
Handleiding
Benutzerhandbuch
Substrate pH/Temp meter
S300 Pro 2

About Aqua Master Tools
Established in 2018
Aqua Master Tools is the Dutch supplier of user-friendly
and reliable pH, EC, and temperature meters. Our meters
are used for swimming pools, aquariums, Jacuzzi’s,
ponds, and liquid plant nutrition.
We understand the importance of convenient, accurate
and high-quality meters - and we are on a continuous
quest to innovate our products. Aqua Master Tools’ wide
range of products are customised to our user’s needs.
Built to last products
Fully calibrated meters
Unprecedented quality
Satisfaction guaranteed
Easily replaceable electrodes

Table of
Contents
Handleiding Nederlands 4 - 7
User manual English 8 - 11
Benutzerhandbuch Deutsch 12 - 15

4NL
1. Dop
2. Display
3. Aan/uit knop
4. Temp/CAL knop
5. HOLD knop
6. pH elektrode
7. BNC connector
8. Beschermkapje
Nederlands
6
1
2
3
4
5
7
8

5NL
Bediening en gebruik
1. Sluit de BNC connector van de elektrode aan op de pH-meter
en haal de beschermkap met bewaarvloeistof van de elektrode.
2. Spoel de elektrode af met water en droog deze met schoon
keukenpapier.
3. Druk op de toets on/off om de meter in te schakelen.
4. Plaats de elektrode in het te meten (zachte) substraat (zorg dat
het substraat goed nat/vochtig is anders werkt de meter niet naar
behoren) of vloeistof.
5. Wacht tot de waarde stabiel wordt. De meting is nu klaar.
6. Druk op de toets HOLD om de meting op te slaan. Vergeet niet om
de toets HOLD opnieuw in te drukken wanneer de meter opnieuw
gebruikt wordt.
7. Spoel de elektrode na gebruik af met schoon water, zodat er geen
residuen achterblijven op de elektrode. Plaats de elektrode terug
in de beschermkap met bewaarvloeistof.
8. Druk op de toets on/off om de meter uit te schakelen.
Wij adviseren om de meter met KCI bewaarvloeistof in het beschermkapje
te bewaren.
pH meter kalibreren
1. Druk op de Aan/uit toets zodat de meter ingeschakeld is.
2. Reinig de elektrode zorgvuldig met gedistilleerd water en droog deze af.
3. Plaats de elektrode rechtop in pH bufferoplossing pH 7,01 (zorg dat
deze 25°C is) en roer zachtjes.
4. Wacht tot de waarde op het display stabiel is.
5. Houd de CAL toets 5 seconden lang ingedrukt. CAL pH 7.0 zal nu op het
scherm verschijnen. Zodra er een vinkje op het scherm verschijnt is de
kalibratie succesvol verlopen.
6. Reinig de elektrode zorgvuldig met gedistilleerd water en droog deze af.
7. Herhaal het proces met pH 4,01 bufferoplossing.
8. Reinig na dit proces de meter zorgvuldig voor u deze opbergt.
Het kalibreren is nu klaar.
• Als de elektrode is vervangen.
• Als de meter langer dan een maand geleden voor het laatst
is gekalibreerd.
• Als de meter heel vaak gebruikt wordt.
• Als er hoge precisie nodig is.
Wanneer kalibreren?

6NL
Elektrode vervangen
1. Koppel de BNC connector van de oude pH elektrode los van de meter.
2. Sluit de BNC connector van de nieuwe pH elektrode aan op de meter.
3. Kalibreer de meter zoals beschreven staat in deze handleiding.
4. De meter is nu weer klaar voor gebruik.
Specificaties
Bereik pH: 0,0 ~ 14,0 pH
Temp.: 0°C ~ 50°C (32°F ~ 122°F)
Resolutie pH: ± 0.1 pH
Temp.: 0.1 ºC (1.0 ºF)
Accuraatheid
pH: ±0,1 pH
Temperatuur: ± 0.5°C (32°F)
Temperatuurcompensatie: 0°C to 50°C
(32°F to 122°F)
Kalibratie pH: 2 point auto 4.0 - 7.0 pH
Kenmerken Batterij indicator & automatische
temperatuur compensatie
Voeding 3 × 1,5 Volt AG-13 knoopcel (meegeleverd)
Automatisch uitschakelen Na 8 minuten
Kalibratie indicatie Bij correcte kalibratie
Waterproef IP65 (elektroden zijn IP67 waterproef)
Afmeting 220 x 40 x 40 mm
Gewicht 158 gram

7NL
Waarschuwing
• Zorg ervoor dat er tijdens het kalibreren geen luchtbelletjes aanwezig
zijn in de vloeistof. Dit kan de metingen beïnvloeden.
• Zorg ervoor dat de kalibratie vloeistof 25oC is tijdens het kalibreren
van de meter. Indien dit niet het geval is kan dit het resultaat van de
kalibratie beïnvloeden.
• Bewaar de elektrode nooit in gedistilleerd water.
• De elektrode NIET onderdompelen in olie, eiwitten of gesuspendeerde
vaste stoffen die een laag op de glazen bol achterlaten.
• De elektrode NIET weken of spoelen in OO (Omgekeerde Osmose),
gedistilleerd of gedeïoniseerd water. Zuiver water wijzigt de chemie
in de referentievloeistof, waardoor deze niet meer werkt.
• Een koude elektrode NIET in hete vloeistof (of vice versa) plaatsen.
Door plotselinge temperatuursveranderingen kan het glas breken
en kan de pen permanent beschadigd raken. Gebroken glas valt niet
onder de garantie.
• De glaselektrode nergens tegenaan stoten. Anders kan de glazen
punt aan de buitenkant of het interne glazen buisje breken.
Gebroken glas valt niet onder de garantie.
• Plaats de meter bij voorkeur niet in de buurt van voorschakel-
of andere elektrische apparatuur. Spanningsvelden kunnen de
meetresultaten beïnvloeden.
• Als tijdens het kalibreren de pH waarde langzaam blijft stijgen of
dalen en niet stabiel wordt, vervangt u de batterijen en kalibreert u
opnieuw. Slechte batterijen of een laag spanningsniveau kunnen de
meetresultaten beïnvloeden.
Garantie
Deze meter heeft garantie voor alle materiële defecten en productiefouten
gedurende een periode van één jaar vanaf de datum van aankoop. De
elektrode heeft garantie voor alle materiële defecten en productiefouten
gedurende een periode van zes maanden vanaf de datum van aankoop. Als
tijdens deze periode de reparatie of vervanging van onderdelen vereist is
en de schade niet te wijten is aan een foutieve bediening door de gebruiker,
stuur dan de onderdelen naar de dealer of naar ons en de reparatie zal
kosteloos plaatsvinden. Zie de waarschuwing voor uitsluitingen.
Batterijen vervangen
Als de batterij-indicator rechtsboven in het scherm aan gaat, of als het
scherm vaag wordt, moeten de batterijen worden vervangen. Draai hiervoor
de bovenkant los en vervang de 3 x 1,5 Volt AG-13 knoopcelbatterijen.

8EN
English
1. Cap
2. Display
3. On/off button
4. Temp/CAL button
5. HOLD button
6. pH electrode
7. BNC connector
8. Protective cap
6
1
2
3
4
5
7
8

9EN
Operation and use
1. Connect the BNC connector of the electrode to the pH meter and
remove the protective cap containing storage solution from the
electrode.
2. Rinse the electrode with running water and dry it using clean
kitchen paper.
3. Press the on/off button to turn on the meter.
4. Place the electrode in the (soft) substrate (make sure the substrate
is well wet/damp, otherwise the meter will not work properly) or
liquid to be measured.
5. Wait for the value to stabilise. The measurement is now complete.
6. Press the HOLD button to save the measurement. Do not forget to
press the HOLD button again when the meter is used once more.
7. Rinse the electrode with clean water after use, so that no residue
is left on the electrode. Replace the electrode in the protective cap
containing the storage solution.
8. Press the on/off button to turn off the meter.
We recommend storing the meter with KCI storage solution in the
protective cover.
Calibrating the pH meter
1. Press the On/off button to start the meter.
2. Gently clean the electrode with distilled water and dry it off.
3. Insert the meter in an upright position in pH buffer solution pH 7.01
(make sure this is 25°C) and stir gently.
4. Wait until the value on the display is stable.
5. Press and hold the CAL button for 5 seconds. CAL pH 7.0 will now
appear on the display. Calibration is complete as soon as a tick sign
appears on the screen.
6. Gently clean the electrode with distilled water and dry it off.
7. Repeat the process with pH 4.01 buffer solution.
8. Gently clean the meter before storing it. Calibration is now complete.
• If the electrode has been replaced.
• If the meter was last calibrated over a month ago.
• If the meter is used on a regular basis.
• If high precision is needed.
When is calibration required?

10EN
Replacing the electrode
1. Detach the BNC connector of the old pH electrode from the meter.
2. Connect the BNC connector of the new pH electrode to the meter.
3. Calibrate the meter as described in this manual.
4. The meter is now ready for use.
Specifications
Range pH: 0,0 ~ 14,0 pH
Temperature: -50°C ~ 70°C (-58°F ~ 158°F)
Resolution pH: 0,1 pH
Temperature: 0,1 ºC (0,2 ºF)
Accuracy
pH: ±0,1 pH
Temperature: ± 0.5°C (32°F)
Temperature compensation: 0°C - 50°C
(32°F - 122°F)
Calibration pH: 2 point auto 4.0 - 7.0 pH
Features Battery indicator & automatic temp.
compensation
Power supply 3 × 1.5 Volt AG-13 button cell battery
(included)
Auto. power-off After 8 minutes
Calibration indicator With correct calibration
Waterproof IP65 (electrodes are IP67 waterproof)
Dimensions 220 x 40 x 40 mm
Weight 158 grams

11EN
Warning
• Always make sure that there are no air bubbles present in the solution
during calibration. This could affect the measurements.
• Make sure the calibration solution is 25°C when calibrating the meter.
Failure to do so may affect the result of the calibration.
• Never store the electrode in distilled water.
• DO NOT immerse the electrode in oil, proteins or suspended solids that
leave a layer on the glass bulb.
• DO NOT soak or rinse in RO (Reverse Osmosis), distilled or de-ionised
water. Pure water alters the chemistry of the reference liquid so that it
no longer works.
• DO NOT place a cold electrode in hot liquid (or vice versa). Sudden
changes in temperature may break the glass and permanently damage
the pen. Broken glass is not covered by the warranty.
• Make sure not to bump the glass electrode into anything, because this
might break the glass point on the outside of the internal glass tube.
Broken glass is not covered by the warranty
• Avoid placing the meter near ballasts or any other electrical appliances.
Force fields may affect the readings.
• If during calibration the pH value keeps rising or dropping slowly and
does not stabilise, please replace the batteries and calibrate the
meter. Bad batteries or low power levels may influence the
measurement results.
Replacing the batteries
If the battery indicator at the top right of the display is activated, or if the
display dims, please replace the batteries. To do so, unscrew the top and
replace the 3 x 1.5 Volt AG-13 button cell batteries.
Warranty
This meter is warranted from all defects in material and manufacturing
for a period of one year from the date of purchase. The electrode is
warranted from all defects in material and manufacturing for a period
of six months from the date of purchase. If during this period any parts
need reparation or replacement and the damage is not due to incorrect
operation by the user, please send the parts to the dealer or to us and the
repair will be free of charge. Please refer to the exclusion warning.

12DE
Deutsch
1. Verschluss
2. Display
3. Ein-/Aus-Schalter
4. Temp/CAL-Taste
5. HOLD-Taste
6. pH Elektrode
7. BNC-Stecker
8. Schutzkappe
6
1
2
3
4
5
7
8

13DE
Bedienung und Betrieb
1. Verbinden Sie den BNC-Stecker der Elektrode mit dem pH-Meter
und entfernen Sie die Schutzkappe mit der Lagerungsflüssigkeit der
Elektrode.
2. Spülen Sie die Elektrode unter fließendem Wasser ab und trocknen Sie
sie mit sauberem Küchenpapier.
3. Drücken Sie die on/off-Taste, um das Messgerät einzuschalten.
4. Legen Sie die Elektrode in das (weiche) Substrat (stellen Sie sicher, dass
das Substrat gut befeuchtet ist, sonst funktioniert das Messgerät nicht
richtig) oder in die zu messende Flüssigkeit.
5. Warten Sie, bis sich der Wert stabilisiert hat. Die Messung ist nun
abgeschlossen.
6. Drücken Sie die HOLD-Taste, um die Messung zu speichern. Vergessen
Sie nicht, erneut die HOLD-Taste zu drücken, wenn Sie das Messgerät
ein weiteres Mal verwenden.
7. Spülen Sie die Elektrode nach dem Gebrauch mit sauberem Wasser ab,
damit keine Rückstände auf der Elektrode verbleiben. Stecken Sie die
Elektrode wieder in die Schutzkappe, welche die Lagerungsflüssigkeit
enthält.
8. Drücken Sie die Ein/Aus-Taste, um das Messgerät auszuschalten.
Wir empfehlen, das Messgerät mit der KCI-Lagerlösung in der Schutzkappe
aufzubewahren.
pH-Meter kalibrieren
1. Drücken Sie zum Einschalten des Meters auf den Ein-/Aus-Schalter.
2. Reinigen Sie die Elektrode sorgfältig mit destilliertem Wasser und
trocknen Sie diese.
3. Halten Sie das Messgerät senkrecht in die pH-Pufferlösung pH 7,01
(diese muss 25 oC warm sein) und lassen Sie es leicht kreisen.
4. Warten Sie, bis der Wert auf dem Display stabil bleibt.
5. Drücken und halten Sie die CAL-Taste 5 Sekunden lang gedrückt.
Auf dem Display erscheint nun CAL pH 7.0. Die Kalibrierung ist
abgeschlossen, sobald ein Häkchen auf dem Bildschirm erscheint.
6. Reinigen Sie die Elektrode sorgfältig mit destilliertem Wasser und
trocknen Sie diese.
7. Wiederholen Sie dieses Verfahren mit der Pufferlösung pH 4,01.
8. Reinigen Sie das Messgerät nach dem Kalibrieren gründlich, bevor Sie
es wieder wegräumen.

14DE
Elektrode ersetzen
1. Ziehen Sie den BNC-Stecker der alten pH-Elektrode vom Messgerät ab.
2. Schließen Sie den BNC-Stecker der neuen pH-Elektrode an das
Messgerät an.
3. Kalibrieren Sie das Multimeter gemäß der Beschreibung in diesem
Handbuch.
4. Das Multimeter ist nun wieder einsatzbereit.
Technische Daten
Messbereich pH: 0.0 ~ 14.0 pH
Temp.: 0°C ~ 50°C (32°F ~ 122°F)
Auflösung pH: ± 0.1 pH
Temp.: 0.1 ºC (1.0 ºF)
Messgenauigkeit pH: ± 0.1 pH
Temperatur: ±0.5°C (32°F)
Temperaturausgleich: 0°C - 50°C (32°F - 122°F)
Kalibrierung pH: 2 point auto 4.0 - 7.0 pH
Eigenschaften Batterie anzeige & Autom. Temperaturausgleich
Energieversorgung 3 × 1,5 Volt AG-13 Knopfzelle (einschließlich)
Abschaltautomatik 8 minuten
Kalibrierung anzeige Bei korrekter Kalibrierung
Wasserdicht IP65 (Elektroden IP67 wasserdicht)
Maße 220 x 40 x 40 mm
Gewicht 171 Gramm
• Wenn die Elektrode ausgetauscht wurde.
• Wenn die letzte Kalibrierung des Messgeräts mehr als einen Monat
zurückliegt.
• Wenn das Messgerät sehr oft gebraucht wird.
• Wenn höchste Präzision erforderlich ist.
Wann kalibriert werden muss:

15DE
Warnhinweis
• Achten Sie darauf, dass während des Kalibriervorgangs keine Luftbläschen
in der Flüssigkeit sind. Dies kann die Messungen beeinflussen.
• Sorgen Sie dafür, dass die Temperatur der Kalibrierflüssigkeit während
des Kalibriervorgangs bei 25°C liegt. Ist dies nicht der Fall, kann dies das
Resultat der Kalibrierung beeinflussen.
• Bewahren Sie die Elektrode nicht in destilliertem Wasser auf.
• Tauchen Sie die Elektrode NICHT in Öl oder Lösungen mit Eiweißen oder
Schwebstoffen, die einen Film auf der Glasmembran hinterlassen.
• VERMEIDEN Sie es, die Elektrode in Umkehrosmosewasser, destilliertem
oder deionisiertem Wasser zu spülen oder einzutauchen. Reines Wasser
verändert die Chemie in der Bezugsflüssigkeit, die dadurch nicht mehr
funktioniert.
• Bringen Sie eine kalte Elektrode NICHT in Kontakt mit heißer Flüssigkeit
(oder umgekehrt). Durch abrupte Temperaturschwankungen kann das
Glas brechen und der Elektrodenstift unwiderruflich beschädigt werden.
Zerbrochenes Glas fällt nicht unter die Garantie.
• Achten Sie darauf, mit der Glaselektrode nicht irgendwo dagegen zu
stoßen, weil dadurch die Glasspitze an der Außenseite des inneren
Glasrohrs brechen könnte. Zerbrochenes Glas ist nicht durch die Garantie
abgedeckt.
• Installieren Sie das Messgerät vorzugsweise nicht in der Nähe von
Vorschaltgeräten oder anderen Elektrogeräten. Spannungsfelder können
die Messresultate beeinflussen.
• Wenn sich der pH-Wert während des Kalibrierens nicht stabilisiert,
sondern langsam weiter steigt oder sinkt, ersetzen Sie die Batterien und
führen Sie den Kalibriervorgang erneut durch. Minderwertige Batterien
oder eine zu geringe Spannung können die Messresultate beeinflussen.
Batterien ersetzen
Wenn die Batterieanzeige oben rechts im Bildschirm aufleuchtet oder wenn das
Display schwächer wird, müssen die Batterien ersetzt werden. Drehen Sie dazu
die Oberseite ab und ersetzen Sie die drei 1,5-Volt-Knopfzellbatterien AG-13.
Garantie
Für dieses Multimeter gilt eine einjährige Garantie auf alle Material- und
Produktionsfehler ab Kaufdatum. Für dieses Elektrode gilt eine sechs Monate
Garantie auf alle Material- und Produktionsfehler ab Kaufdatum. Wenn in
diesem Zeitraum Komponenten repariert oder ersetzt werden müssen und
der Schaden nicht auf fehlerhafte Verwendung des Bedieners zurückzuführen
ist, schicken Sie die entsprechenden Komponenten zu einem Vertragshändler
oder zu uns ein. Die Reparatur erfolgt dann kostenlos. Siehe den Warnhinweis
bezüglich ausgeschlossener Garantieleistungen.

www.aquamastertools.com
Aqua Master Tools
Ambachtsweg 55c
1271 AL Huizen
The Netherlands
This manual suits for next models
1
Table of contents
Languages:
Other AquaMaster Tools Measuring Instrument manuals

AquaMaster Tools
AquaMaster Tools P50 Pro User manual

AquaMaster Tools
AquaMaster Tools E300 Pro User manual

AquaMaster Tools
AquaMaster Tools Combo pen P110 Pro User manual

AquaMaster Tools
AquaMaster Tools E5O Pro User manual

AquaMaster Tools
AquaMaster Tools Combo Pen P100 Pro User manual

AquaMaster Tools
AquaMaster Tools Combo Pen P150 Pro User manual

AquaMaster Tools
AquaMaster Tools Combo pen P160 Pro User manual

AquaMaster Tools
AquaMaster Tools Combo meter P700 Pro 2 User manual