ARAG Visio Quick start guide

Software rel. 2.0.x
INSTALLATION, uSe ANd mAINTeNANce
4670610
APPLICATION RATE MONITOR VERSION

=Generic danger
=Warning
This manual is an integral part of the equipment to which it refers and must accompany the equipment in case of sale
or change of ownership. Keep it for any future reference; ARAG reserves the right to modify product specications and
instructions at any moment and without notice.
LEGEND OF SYMBOLS

3
CONTENTS
• Productdescription........................................................................................................................................................... 4
1 Intendeduse ......................................................................................................................................................................4
2 Contentofthepackage.....................................................................................................................................................4
3 Precautions ........................................................................................................................................................................4
4 Risksandprotectionsbeforeassembly.......................................................................................................................... 4
4.1 Positioning..................................................................................................................................................................4
4.2 Power supply and sensor connection.........................................................................................................................5
5 Assemblydiagrams...........................................................................................................................................................5
6 Controlsinthemenu......................................................................................................................................................... 6
6.1 First switch-on ............................................................................................................................................................6
6.2 Controls in the menu ..................................................................................................................................................6
6.3 Menu structure ...........................................................................................................................................................7
7 Preliminarysetupforuse..................................................................................................................................................8
7.1 Alarms ........................................................................................................................................................................8
7.1.1 Flowrate alarms.......................................................................................................................................................8
7.1.2 Speed alarms ..........................................................................................................................................................8
7.1.3 Application rate alarms........................................................................................................................................... 9
7.2 Sensors ......................................................................................................................................................................9
7.3 Flowmeter calibration ...............................................................................................................................................10
7.3.1 Automatic calibration ............................................................................................................................................. 10
7.3.2 Manual calibration ................................................................................................................................................. 10
7.4 Speed sensor calibration.......................................................................................................................................... 11
7.4.1 Automatic calibration ............................................................................................................................................ 11
7.4.2 Manual calibration................................................................................................................................................. 12
7.4 Language .................................................................................................................................................................13
7.5 Units of measurement ..............................................................................................................................................13
7.5.1 Rate units of measurement ................................................................................................................................... 13
7.5.2 Volume units of measurement............................................................................................................................... 13
7.5.3 Speed units of measurement ................................................................................................................................ 14
7.5.4 Length units of measurement................................................................................................................................ 14
7.5.5 Surface units of measurement .............................................................................................................................. 14
7.5.6 Application rate units of measurement .................................................................................................................. 15
7.5.7 Distance units of measurement............................................................................................................................. 15
7.6 Display contrast........................................................................................................................................................16
7.7 Alarm tones ..............................................................................................................................................................16
7.8 Keytones...................................................................................................................................................................16
7.9 Display settings ........................................................................................................................................................17
7.10 Operating mode........................................................................................................................................................18
7.11 Setup management..................................................................................................................................................18
7.12 Test menu.................................................................................................................................................................19
7.12.1 Display test............................................................................................................................................................ 19
7.12.2 Keys test................................................................................................................................................................ 19
7.12.3 Sensors test ..........................................................................................................................................................20
8 Use.................................................................................................................................................................................... 20
8.1 Partial totalizer reset ..............................................................................................................................................20
9 Maintenance/Diagnostics/Repairs .............................................................................................................................21
9.1 Troubleshooting ........................................................................................................................................................21
10Technicaldata..................................................................................................................................................................21
10.1 Technical data ..........................................................................................................................................................21
11 Guaranteeterms .............................................................................................................................................................. 22
12Endoflifedisposal .........................................................................................................................................................22

4
VISIO is a very compact and accurate top-notch multifunction display, able to display any kind of information concerning agricultural treatments.
Operator can select the required function via software.
It can display several types of values, which change according to set operating mode and type of connected sensors.
Thisdeviceisdesignedtoworkonagriculturalmachineryforsprayingandcropsprayingapplications.
ThemachineisdesignedandbuiltincompliancewithENISO14982standard(Electromagneticcompatibility-Forestryand
farmingmachines),harmonizedwith2004/108/ECDirective.
The table below indicates the components that you will nd in the VISIO package:
Legend:
1VISIO
2Fixing kit
3Instruction manual (on CD-ROM)
4Installation sheet
•Donotaimwaterjetsattheequipment.
•Donotusesolventsorfueltocleanthecaseoutersurface.
•Donotcleanequipmentwithdirectwaterjets.
•Complywiththespeciedpowervoltage(12VDC).
•Incaseofvoltaicarcwelding,removeconnectorsfromVISIOanddisconnectthepowercables.
•OnlyuseARAGgenuinesparepartsandaccessories.
Allinstallationworksmustbedonewithbatterydisconnected,usingsuitabletoolsandanyindividualprotectionequipment
deemednecessary.
4.1 Positioning
1) Set mounting rail in cabin and fasten it with the relevant screws
(Fig. 2), in a position where VISIO can be easily seen and at hands'
reach, but away from any moving organs.
2) Secure VISIO to rail and push down until locked in place.
3) Fasten wiring so that it does not interfere with any moving parts.
Power cable, sensors and
cables to be connected to
VISIO must be ordered
separately.
INTRODUCTION
Fi g. 2
Fi g. 1

5
4.2 Powersupplyandsensorconnection
Sensorsandpowersupplymustbeinstalledandconnectedbyqualiedpersonnel.
VISIOmustbeexclusivelyconnectedtoARAGequipment.
WHENARCWELDINGISREQUIRED,MAKESURETHATEQUIPMENTPOWERISSWITCHEDOFF;DISCONNECTPOWERCABLESIFNEEDED.
ARAGisnotliablefordamagetothesystem,persons,animalsorpropertycausedbyVISIOwrongorunsuitableassembly.
Failuretoobservetheaboveinstructionsautomaticallyvoidsthewarranty.
INSTALLATION
sensors
Fi g. 3
power supply Wire color
(power cable) Connection
red positive
black negative
green counting abort
Tab. 1
Legend:
1) Power cable
2) Connection cable for double sensor
3) Speed sensor
4) Flowmeter
Fi g. 4

6
6.1 Firstswitch-on
Atrstswitch-on,VISIOwillrunaguidedprocedureallowingusertosetthedevice'sbasicsettings.
Press to scroll through items, OK to save and move on to next setting, or ESC to go back to previous setting.
WARNING:
Beforechangingoperatingmode,makesurethatallsensorsareDISCONNECTEDfromthedevice.
6.2 Controlsinthemenu
Inthefollowingpages,accordingtothesetoperatingmode,somemenuitemscouldslightlydifferfromtheshownones.
Press for1second.
Press the key a few times to view the various values in extended mode, (on display central
part).
Everytimethedeviceisswitchedon,itwillshortlyshowapagewiththenameof
deviceandsoftwareversion.
Press for2seconds.
From the main page, press keys at the same time for2seconds to open the Setup Menu.
Press a few times to scroll through items (selected item is indicated by a black line)
Press to open the selected menu item
Thethreedotsunderanitemindicatepresenceofanothersetupmenu.
Press to move through digits
Press a few times to edit the highlighted digit
Press to conrm. The display goes back to previous page.
Press to exit page without conrming modication.
Editedvaluemustfallwithintherangeshown.
Fi g. 5
Fig. 6
Fig. 7
Fig. 8
CONTROLS

7
6.3 Menustructure
MENU

8
7.1 Alarms
7.1.1 Flowrate alarms
Set minimum and maximum owrate thresholds for alarm message.
1) Open Alarm menu (Setup menu > Alarms).
Minimumandmaximumratealarmsaresetinthesameway.
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to edit the selected menu item.
2) To activate the alarm, press
and
at the same time until message OFF goes off
and rate alarm value is displayed instead.
Carry out the same procedure to disable alarm again.
3) Set alarm value:
A) Press to move through digits
B) Press a few times to edit the highlighted digit
C) Press to save changes, or D) Press to quit the page without conrming changes.
7.1.2 Speed alarms
Set minimum and maximum speed thresholds for alarm message.
1) Open Alarm menu (Setup menu > Alarms).
Minimumandmaximumspeedalarmsaresetinthesameway.
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to edit the selected menu item.
2) To activate the alarm, press
and
at the same time until message OFF goes off
and speed alarm value is displayed instead.
Carry out the same procedure to disable alarm again.
3) Set alarm value:
A) Press to move through digits
B) Press a few times to edit the highlighted digit
C) Press to save changes, or D) Press to quit the page without conrming changes.
Fig. 9
Fi g. 10
Fi g. 11
Fi g. 12
Fi g. 13
Fi g. 14
SETUP

9
7.1.3 Application rate alarms
Set the desired alarm display thresholds for minimum and maximum application rate.
1) Open Alarm menu (Setup menu > Alarms).
Minimumandmaximumapplicationratealarmsaresetinthesameway.
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to edit the selected menu item.
2) To activate the alarm, press
and
at the same time until message OFF goes off
and application rate alarm value is displayed instead.
Carry out the same procedure to disable alarm again.
3) Set alarm value:
A) Press to move through digits
B) Press a few times to edit the highlighted digit
C) Press to save changes, or D) Press to quit the page without conrming changes.
7.2 Sensors
1) Open Sensors menu (Setup menu > Sensors).
2) Press OK to edit the selected menu item.
3) Press to open the desired calibration mode.
Fig. 15
Fi g. 16
Fi g. 17
Fi g. 18
Fi g. 19
Fi g. 20
SETUP

10
7.3 Flowmetercalibration
Duetothedifferentsystemcongurations(tubes,valves,etc.)ratereadingcouldnotbecorrect.
Itisthereforerecommendedtoperformasprayingtest.Ifmeasuredvalueisdifferentfromactualvalue,changetherateconstant
throughanautomaticcalibrationprocedureormanuallycalculatetheconstant.
7.3.1 Automatic calibration
Let a previously measured quantity of uid go through the system or a quantity that could be measured through another system. Thehigherthe
quantityofuidusedtoperformthecalibrationprocedure,themoreaccuratethecalibration.
1) Open automatic calibration menu (Setup menu > Sensors > Flowmeter > Auto Calibration).
Assoonasmenuisopen,theequipmentisreadytostartmeasuringwithnofurther
controlsbeingrequired.
2) Start liquid ow in the system. The display will start showing the increasing value of the
measured uid quantity. As soon as uid ow is over, the displayed counter will stop.
3) Now press OK. At the bottom of the display Stabilization message will turn on
and equipment will then show the page on the side.
4) Enter previously measured quantity of uid:
A) Press to move through digits
B) Press a few times to edit the highlighted digit
C) Press to complete the calibration procedure, or D) Press for 1 second to cancel the
calibration procedure.
1
Setting the uid quantity actually passed through owmeter during the calibration procedure.
2
Viewing the uid quantity read by owmeter during the calibration procedure.
Ifequipmentdoesnotdetectanyowaftercalibrationisstarted(anddisplayedvalueremains0),pressOKtoquitthecalibrationprocedure
withoutsaving.
IfthedevicecontinuestodetectowafterOKwaspressed,theerrormessageStop flow!willbedisplayedafterafewseconds.
Assoonasowisstopped,readingwillstabilize,asperstandardprocedure.
IncaseVISIOcalculatesavalueoutoftherangethatcanbesetbymeansofthemanualcalibration,thecalculatedvaluewillnot
beset.
7.3.2 Manual calibration
In order to manually set rate constant, calculate and set suitable constant using the following formula:
[quantity measured by equipment]
[actually sprayed quantity]
1) Open manual calibration menu (Setup menu > Calibration > Man. Calibration).
Under Calibrationmenu,selectMan. Calibration,toviewcurrentlysetconstantvalue
belowtheitem.
PressOKtoeditthevalue.
Forowmeterconstantvalue,refertothecorrespondingmanual.
2) Set owmeter constant value:
A) Press to move through digits
B) Press a few times to edit the highlighted digit
C) Press to save changes, or D) Press to quit the page without conrming changes.
Fig. 21
Fig. 22
Fig. 23
Fig. 24
x [constant indicated on flowmeter body]
SETUP

11
SETUP
7.4 Speedsensorcalibration
VISIO calculates the information concerning the speed thanks to pulses received by the sensor installed on the wheel.
To perform calibration, proceed as follows:
- Measure a straight path at least 100 m (300 feet) long.
The longer the distance traveled, the more accurate the wheel constant calculation.
- Take measurements with tyres at the operating pressure.
This test must be performed on medium-hard terrain; for application to very soft or very hard terrain, rolling diameter may vary, leading to inaccurate
output calculation; when this is the case, repeat the procedure.
During the test, cover the distance with the tank lled up to half capacity with water.
7.4.1 Automatic calibration
Calculate and save the wheel constant according to the procedure below:
1) Open automatic calibration menu (Setup menu > Sensors > Speed sensor > Auto
Calibration).
Assoonasmenuisopen,theequipmentisreadytostartmeasuringwithnofurther
controlsbeingrequired.
Cover the requested distance: the number of pulses will increase during the path and the
bottom area will indicate instant speed reading.
Stop the tractor at the end of the distance.
If a malfunction occurs, the message Check sensor! will be shown in the top
part of the display.
Press OK to stop the counting. The display will go back to the previous menu and will show
the acquired value.
In case of measurement errors, or if it is necessary to stop the calibration, press ESC for 2
seconds to quit the calibration procedure without saving.
In this case, the value will be the one previously measured, or the default value.
Fi g. 25
Fi g. 26
Fi g. 27
Fig. 28

12
7.4.2 Manual calibration
Manual calibration allows to enter the wheel constant value calculated with the suitable formula:
Kwheel= distance traveled (cm)
no. of detection points x wheel rpm
<distancetraveled>distance expressed in cm covered by the wheel along measurement travel;
<no.ofmeasurementpoints>number of measurement points (e.g., magnets, bolts, etc.), mounted on wheel;
<no.ofwheelrevolutions>number of wheel revolutions required to travel measurement distance.
The wheel constant can be calculated with a good approximation by detecting the distance traveled by the wheel with the speed sensor.
1) Open manual calibration menu (Setup menu > Sensors > Speed sensor > Man.
Calibration).
Cover the requested distance. Stop the tractor at the end of the distance and calculate wheel
constant (Kwheel).
2) Set wheel constant value:
A) Press to move through digits
B) Press a few times to edit the highlighted digit
C) Press to save changes, or D) Press to quit the page without conrming changes.
Fig. 29
Fig. 30
SETUP

13
7.4 Language
Set the desired language.
Open language setting menu (Setup menu > Options > Language).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select the language.
1) Select a language through key.
2) Press OK to save, or ESC to quit without saving.
7.5 Unitsofmeasurement
Set unit of measurement for the values detected by the device.
7.5.1 Rate units of measurement
Open instant rate unit of measurement setting menu (Setup menu > Options > Units of
measurem. > Flowrate).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select owrate unit of measurement.
1) Select a unit through key.
2) Press OK to save, or ESC to quit without saving.
7.5.2 Volume units of measurement
Open volume unit of measurement setting menu (Setup menu > Options > Units of
measurem. > Volume).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select volume unit of measurement.
1) Select a unit through key.
2) Press OK to save, or ESC to quit without saving.
Fi g. 31
Fi g. 32
Fi g. 33
Fi g. 34
Fi g. 35
Fig. 36
SETUP

14
SETUP
7.5.3 Speed units of measurement
Open speed unit of measurement setting menu (Setup menu > Options > Units of measurem. >
Speed).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select speed unit of measurement.
1) Select a unit through .
2) Press OK to save, or ESC to quit without saving.
7.5.4 Length units of measurement
Open length unit of measurement setting menu (Setup menu > Options > Units of measurem. >
Length).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select length unit of measurement.
1) Select a unit through key.
2) Press OK to save, or ESC to quit without saving.
7.5.5 Surface units of measurement
Open surface unit of measurement setting menu (Setup menu > Options > Units of
measurem. > Surface).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select surface unit of measurement.
1) Select a unit through key.
2) Press OK to save, or ESC to quit without saving.
Fi g. 37
Fi g. 38
Fi g. 39
Fi g. 40
Fi g. 41
Fi g. 42

15
SETUP
7.5.6 Application rate units of measurement
Open application rate unit of measurement setting menu (Setup menu > Options > Units of
measurem. > Application rate).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select application rate unit of measurement.
1) Select a unit through key.
2) Press OK to save, or ESC to quit without saving.
7.5.7 Distance units of measurement
Open distance unit of measurement setting menu (Setup menu > Options > Units of
measurem. > Distance).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to select distance unit of measurement.
1) Select a unit through key.
2) Press OK to save, or ESC to quit without saving.
Fi g. 43
Fi g. 44
Fi g. 45
Fig. 46

16
7.6 Displaycontrast
Set display contrast.
Open display contrast menu (Setup menu > Options > Display contrast).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to edit the selected menu item.
1) Set a value through key. Every time you press it, value will increase by 5% up to
100%. Use key to decrease value by 5%.
2) Press OK to save, or ESC to quit without saving.
7.7 Alarmtones
Enable/disable the alarm tones.
Open Alarm tones menu (Setup menu > Options > Alarm tones).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to edit the selected menu item.
1) Set status through key.
2) Press OK to save, or ESC to quit without saving.
7.8 Keytones
Enable/disable keytones.
Open keytones menu (Setup menu > Options > Keytones).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to edit the selected menu item.
1) Set status through key.
2) Press OK to save, or ESC to quit without saving.
Fi g. 47
Fig. 48
Fi g. 49
Fi g. 50
Fi g. 51
Fi g. 52
SETUP

17
7.9 Displaysettings
The main page shows the display divided into three horizontal parts.
Every sector can be assigned the desired value.
Open Display settings menu (Setup menu > Options > Display settings).
Press OK to edit the selected menu item.
1) Select value through key.
2) Press OK to edit, or ESC to quit without saving.
1) Set sector to the required value through key.
2) Press OK to save, or ESC to quit without saving.
Carry out the same procedure for the other 2 values.
Fig. 53
Fig. 54
Fig. 55
SETUP

18
7.10 Operatingmode
Set required operating mode.
Open operating mode menu (Setup menu > Options > Operating mode).
Thedisplaywillshowthecurrentsettingbelowtheselecteditem.
Press OK to edit the operating mode.
1) Select the required operating mode through key.
2) Press OK to conrm change, or ESC to quit without saving.
WARNING:
OnceOKispressed,thepageonthesidewillbedisplayed.
Beforechangingoperatingmode,makesurethatallsensorsareDISCONNECTED
fromthedevice.
Press OK toconrm.
ConnectthesensorsREQUIREDFORTHESETOPERATINGMODE.
7.11 Setupmanagement
VISIO settings can be loaded from or saved on a USB pen drive in order to restore them if required, x problems or set another VISIO with no need
to repeat all manual operations.
Onceinstallationiscompleted,checkVISIOoperationanditisrecommendedtosaveallsettingsontoaUSBpendrive.
TobeabletousethefollowingfunctionsitisnecessarytoinsertaUSBpendriveintherelevantportatthebottomofVISIO.
Open Setup management menu (Setup menu > Setup management).
Press OK to edit the selected menu item.
Load setup
Allows to select a conguration le saved in the USB pen drive and to set VISIO again.
WARNING:ByloadingtheSETUP.BINlecontainedintheUSBpendriveonto
theVISIO,allcurrentsettingswillbelost.
1) Select control through key.
2) Press OK to conrm loading, or ESC to quit without saving.
TheSETUP.BINlecanbeloadedonlyifitissavedintheUSBpendriveroot
directory.
Ifsetupdownloadinvolveschangingoperatingmodeandusingdifferentsensors
thantheonesinuse,makesurethatallsensorsareDISCONNECTEDfromthe
device.
Press OK toconrmloading.
Reconnectsensors.
Save setup
Allows saving VISIO conguration le on the USB pen drive: it will be possible to load it again
any time the same settings need to be retrieved.
1) Select control through key.
2) Press OK to conrm saving, or ESC to quit without saving.
IfaSETUP.BINleisalreadypresentintheUSBpendriverootdirectory,thele
willbeoverwritten.
Fig. 56
Fig. 57
Fig. 58
Fig. 59
Fig. 60
Fig. 61
SETUP

19
7.12 Testmenu
This menu allows user to view some data and carry out an operation test of VISIO:
-Firmwareversion:
the display shows the rmware version installed.
-Batteryvoltage:
the display shows the power voltage of the device.
7.12.1 Display test
Display test checks the device display correct operation.
Open display test menu (Setup menu > Test > Display).
Press OK to perform the test.
All pixels on display are turned on.
Press ESC to go back to previous page.
7.12.2 Keys test
Keys test checks the device keys correct operation.
Open keys test menu (Setup menu > Test > Keys).
Press OK to perform the test.
1) Press any key and the corresponding display area will turn on.
2) Press ESC to quit: as soon as you acknowledge the switch-on of the corresponding area
on the display, device will go back to previous page.
Fig. 62
Fig. 63
Fig. 64
Fig. 65
SETUP

20
7.12.3 Sensors test
Sensors test checks correct operation of the sensors connected to the device.
Open sensors test menu (Setup menu > Test > Sensors).
Press OK to perform the test.
The display will show the current sensor reading below the selected item.
1) Several sensors could be displayed, depending on the set operating mode. In this case,
select required sensor through .
2) Press ESC to quit.
In the 3 areas of the monitor, VISIO will display the parameters set in the Display settings
menu.
Press
key several times to view a value in extended mode until value is on display
central part.
8.1 Partial totalizer reset
To reset a totalizer you must view it in extended mode.
Press key several times until value to be reset is at the central area of the display.
Press key for two seconds. Totalizer resets.
Settingcanorcannotbereset,dependingonsetvalue.
Symbol------indicatesthatowrateortotalizerexceedmaximumvaluethatcanbedisplayed.
Totalizershaveaoatingpointandshowamaximumof5digits.Twodecimalsareshownupto999.99.Onedecimalisshownafterthatand
0decimalsareshownwhenvaluereaches10,000(Fig.71).
Fig. 66
Fig. 67
Fig. 68
Fig. 69
Fig. 70
Fig. 71
USE
Other manuals for Visio
3
Table of contents
Other ARAG Monitor manuals