ARIES USER’S MANUAL
This manual contains details on installation and operation of your Aries Lab Water system. For optimum performance
and safety please read and follow these instructions carefully.
You have a quality, world class water treatment system that will provide ultrapure water for many years of trouble-free
operation. Years of engineering and experience went into the design and manufacture of this instrument which will
dispense water exceeding Type I water purity standards.
In order that we may keep you informed regarding your unit, please take a moment to complete the information below
that will assist with support should it be needed.
Model Number:
Serial Number:
Contact:
Cartridge #1:
Cartridge #2:
Cartridge #3:
Options Installed:
Date:
Dealer:
Phone:
Part Number:
Part Number:
Part Number:

Copyright Notice and Trademarks
© 2020 ResinTech, Inc.
This manual or parts thereof may not be reproduced in any form without the express written permission of Aries
Filterworks, Inc. Aries Filterworks® is a registered trademark of ResinTech, Inc. Camden, NJ USA. For more
information, please visit our website: ^^^HYPLZÄS[LY^VYRZJVT

TABLE OF CONTENTS
1.......... Health and Safety
1.......... System Overview
2.......... Flow Diagram
2.......... Model Types
3.......... Influent & Effluent Specifications
3.......... Installation
4.......... Dimensions
5.......... System Check List & Unpacking Guide
5.......... Main Check List
6.......... System Wall Mount Instructions
6.......... Setup
6.......... Installing the Cartridges
6.......... Installing the UV Bulb (Option)
7.......... Start up Procedure
8.......... Display and Key Pad
8.......... Resistivity Monitor Operations
9.......... Maintenance
10........ Sanitization
11........ Consumables and Spare Parts
12........ Troubleshooting
13........ Warranty
14........ Water Purification Maintenance Log Form

HEALTH AND SAFETY
Installation and service of the equipment should
be performed by authorized and trained personnel.
Safety measures and local guidelines must be
followed. Use good practices wearing personal
protective equipment (PPE) during the installation
and service of the equipment.
Electricity – Aries Filterworks lab water equipment
should be powered by a grounded GFI electrical
connection. Do not open the system unless the
system is unplugged. Before any installation or service
is performed on the equipment, make sure the power
is isolated. To unplug the system, do not pull directly
on the cord but grip the plug and remove it from
the outlet.
Ultraviolet Light – Aries Filterworks lab water systems
may include an ultraviolet light lamp in a stainless-steel
housing. Make sure all power is removed prior to
accessing the side panel to perform any UV
maintenance. Make sure gloves are used when
handling the UV bulb. (Gloves must be worn when
handling the UV bulb. Which protects the integrity of
the bulb) Do not look directly at the UV lamp while the
system is operating.
Water Supply – Good piping practices should be
followed feeding water to the lab equipment. Isolation
valves, pressure gauges and bleed valves are recom-
mended on the main water supply. Pressure should
be regulated to meet the minimum and maximum
ZWLJPÄJH[PVUZZ[H[LKMVY[OLLX\PWTLU[
Sanitization – Make sure PPE is worn during
any sanitization process. The Aries system uses
chlorine as a disinfectant. Chlorinated water will be
KPZJOHYNLKK\YPUNZHTWSPUNHUKÅ\ZOPUNZLX\LUJLZ
Avoid breathing fumes and ensure the area is
Z\ѝJPLU[S`]LU[PSH[LK
SYSTEM OVERVIEW
;OL(YPLZ3HI^H[LYZ`Z[LTKLSP]LYZ;`WL4£
X\HSP[`^H[LYVUKLTHUK>H[LYPZW\YPÄLKPUHZ[HNLK
process consisting of high-purity ion exchange resins
to remove dissolved minerals and internal recirculation
[VTHPU[HPUW\YP[`([KPZJOHYNLHÄUHSTÄS[LY
removes particulates and bacteria to attain CLSI /
(:;4;`WL^H[LYZWLJPÄJH[PVUZ(KKP[PVUHSVW[PVUHS
technologies are incorporated depending on model.
9LZPZ[P]P[`TVUP[VYWYV]PKLZMLLKIHJRVM[OLÄUHS
water quality.
The Aries system is designed to recirculate through
multiple technologies to maintain purity. A timer
controller runs system automatically for 30 minutes
every 2 hours to maintain the integrity of the system.
As the system transitions from “Operating” mode
to “Stand By” mode, the recirculation pump,
solenoid, resistivity display and UV (if installed) will be
H\[VTH[PJHSS`[\YULKVќ]PH[OL[PTLY(>(2,I\[[VU
on the side of the system will reset the cycle timer
back to “Operating” condition.
Options available with the Aries lab system are UV,
recirculating dispensing gun, secondary direct feed
port, and wall mounting bracket.
1 / ARIES HIGH PURITY WATER SYSTEM

INSTALLATION & OPERATION MANUAL / 2
MODEL TYPES*
ARS-102 Aries Lab Water System with 0.2 Micron Outlet Final Filter
ARS-105 (YPLZ3HI>H[LY:`Z[LT^P[O4PJYVU6\[SL[<S[YHÄS[LYMVY,UKV[V_PU9LTV]HS
ARS-102-UV Aries Lab Water System with installed UV and 0.2 Micron Outlet Final Filter
ARS-105-UV (YPLZ3HI>H[LY:`Z[LT^P[OPUZ[HSSLK<=4PJYVU6\[SL[<S[YHÄS[LYMVY,UKV[V_PU9LTV]HS
:\ITPJYVUÄS[LYPUJS\KLK:[HY[<W*HY[YPKNLZ:VSK:LWHYH[LS`
MODEL
OPTIONS*
ARA-UV* Combination UV for TOC and Bacteria Reduction (installed)
ARA-DG Dispensing Gun with Final Filter – Factory Installed option. Recirculating gun mounted on the side of the unit with
approximately 5 feet of span. User manually dispenses water.
ARA-DF
Direct Feed – Factory installed option. Secondary Port on the side of the system. Used to feed ancillary equipment such as
HUHS`aLYZHUKZJPLU[PÄJLX\PWTLU[0U[LYUHSWYLZZ\YLZ^P[JOH\[VTH[PJHSS`RLLWZ[OLZ`Z[LTVWLYH[PUNK\YPUNH^H[LYKYH^
JVUKP[PVU4PUPT\TÅV^VMSWTYLX\PYLK
5V[L[OLYLPZUVÄUHSÄS[LYVU[OPZWVY[(KV^UZ[YLHTÄUHSÄS[LYTH`ILJVUZPKLYLK
ARA-WB Wall Mounting Bracket – Factory Installed option. Wall mounting bracket supplied with additional support base plate to
the main system.
<=WYL]PV\ZS`VќLYLKHZHUVW[PVU^P[OHZLWHYH[LWHY[U\TILY
START UP
CARTRIDGES*
VPK-4010 RO Feed Start Up Cartridges – (3) High Purity Series Mixed Bed Cartridges
VPK-3805 Tap Water Feed / Service DI – (1) Organics Pretreatment & (2) Mixed Bed Cartridges
* Sold Seperately.
FLOW DIAGRAM

3 / ARIES HIGH PURITY WATER SYSTEM
PARAMETER REQUIREMENT
Source/Type 9L]LYZLVZTVZPZ96VY:LY]PJLKLPVUPaH[PVU:+0WYLMLYYLKTWHY[PJ\SH[LWYLÄS[LYPUNPZYLJVTTLUKLK
for SDI and tap water feeds.
Conductivity 96!#:JT
:+0!%4£9LZPZ[P]P[`
TOC
(Using appropriate conditioning
as necessary)
RO: < 50 ppb
SDI: < 200 ppb
Temperature 20° F to 100°F (5° C to 38°C)
Pressure 30 psig minimum/90 psig maximum (1.38 to 6.21 bar)
Fouling Index Silt Density Index: < 3
Dissolved Carbon Dioxide < 30 ppm
Free Chlorine < .05 ppm
INFLUENT WATER SPECIFICATIONS
PARAMETER REQUIREMENT
Resistivity %4£JT9LZPZ[P]P[`
Microorganism < 10 CFU/mL
#*-<T3^P[OTPJYVU\S[YHÄS[LYV\[SL[
TOC #WWI^P[O<=VW[PVUWWIPUÅ\LU[
Chlorides < 1 ppb
Sodium < 1 ppb
Endotoxin #,<T3^P[OTPJYVU\S[YHÄS[LYV\[SL[
EFFLUENT WATER SPECIFICATIONS
CONNECTIONS
- Inlet - 3/8” OD tubing
- Outlet – ¼ FNPT
- Direct Feed Port – 3/8” CPC Female Bulkhead with
CPC Male x 3/8” OD Tube Compression
INSTALLATION
It is recommended to review the pre-site requirements
before proceeding with the installation. Read through
the entire manual prior to the start of the installation.
Review all paperwork of items that have been ordered
and compare to the items that have shipped. If
any parts are missing, notify the factory or your
sales contact.

INSTALLATION & OPERATION MANUAL / 4
POWER REQUIREMENTS
- 110 VAC 60 Hz (1.0 amp), 220 VAC / 50 Hz
(0.5 amp)
- Dedicated GFCI protected outlet six feet within
the system
FEED WATER PRESSURE
30 – 80 psig regulated. If gravity feed water is used,
HUL_[LYUHSKLSP]LY`W\TW^P[OH\[VZO\[Vќ^PSSIL
required (Part number: 1200109)
DIMENSIONS

SYSTEM CHECK LIST AND UNPACKING GUIDE
Special care should be used for shipping and
delivery of these systems. Similar care should be
considered when removing the system from the box.
Items should be inspected and itemized upon
removal to ensure they have not been damaged in
transit. Immediately report any noticeable damage
to shipping carrier.
MAIN CHECKLIST
:`Z[LT
ARS-102 – Main System with 0.2-micron
JHWZ\SLÄS[LY
Or
ARS-102-UV – Main System with UV module
installed. Bulb shipped loose. 0.2 micron
ÄS[LYPUJS\KLK
ARS-105 – Main System with 0.05-micron
\S[YHÄS[LY
ARS-105-UV – Main System with UV module
installed. Bulb shipped loose. 0.05 micron
\S[YHÄS[LYPUJS\KL
6W[PVUZ
Wall Mounting Bracket Plate (ARA-WB).
Installed Dispensing Gun (ARA-DG). Dispensing
Gun Tubing connected to side system with
IYHJRL[TPJYVUÄS[LY7-PUJS\KLK
with ARA-DG
Direct Feed Port (ARA-DF). Side port with CPC
3/8” Female panel mount
0UJS\KLK(JJLZZVYPLZ
-PUHS-PS[LYHZVќLYLK^P[OZWLJPÄJTVKLS
6\[SL[HUNSL]HS]L^P[O;LÅVU[HWL
*LY[PÄJH[LVM*VUMVYTHUJL
/VZLUPWWSLHKHW[LYMVY(9:\S[YHÄS[LY
UV bulb for UV option
-PUHSÄS[LYMVY(9(+.KPZWLUZPUNN\UVW[PVU
Quick connect CPC 3/8” Male x 3/8”
*VTWYLZZPVUÄ[[PUNMVYKPYLJ[MLLKVW[PVU
9LJVTTLUKLK0[LTZULLKLKMVY0UZ[HSSH[PVU
(2) 3/8” Quick Connection Isolation Valve
for upstream isolation and depressurization
(not included)
0-100 psi pressure gauge (300 grade stainless
steel) for diagnostics (not included)
Tube Cutter
#2 Phillips Head Screw Driver
;LÅVU;HWL
Basic Hand tools (adjustable wrench, pliers,
socket wrench, etc.)
Tie-wraps
6W[PVUHS0[LTZ
Drill for Wall Mount
Level for Wall Mount
Measuring Tape
:THSSLSLJ[YVUPJZZPaLÅH[ZJYL^KYP]LY
(Electronics diagnostics)
6 feet of 3/8” Hose for commissioning
5 / ARIES HIGH PURITY WATER SYSTEM

MOUNT INSTRUCTIONS
Wall mount must be capable of supporting a minimum
weight of 50 lbs (22.7 kg). The wall mount bracket has
holes on 16” centers to be secured to studs. If studs
are not available, a plywood sheet mounted to the wall
is recommended. Make sure holes are level.
Mount bracket with small slots in tabs pointing up.
SETUP
Select a location for the unit which is near the water
supply and electrical supply connections. If the water
supply does not already have an isolation valve for
servicing the Aries Lab System, install one ahead of
the unit connection.
Install right-angle valve into the front of the unit outlet.
+VUV[PUZ[HSS[OLÄUHSJHWZ\SLÄS[LYPU[V[OLV\[SL[]HS]L
\U[PS[OLÄUHSZ[LWVMZ[HY[\W
<ZPUN[OL[\ILÄ[[PUNZWYV]PKLK^P[O[OL\UP[JVUULJ[
the unit inlet to the water supply.
+656;VWLU[OLZ\WWS`PZVSH[PVU]HS]L
H[[OPZ[PTL
INSTALLING THE CARTRIDGES
;OLYLHYL[OYLLJHY[YPKNLWVZP[PVUZ;OL^H[LYÅV^Z
MYVTSLM[[VYPNO[(U`WYL[YLH[TLU[ÄS[LYZT\Z[IL
installed in position #1 (furthest from the display) and a
WVSPZOPUNÄS[LYT\Z[ILPUZ[HSSLKPUWVZP[PVUJSVZLZ[
to the display).
Locate the water purifying cartridges selected for your
Z`Z[LT<U^YHW[OLÄS[LYZYLTV]L[OLZOPWWPUNWS\NZ
HUKHѝ_[OLSHILSZ
7SHJL[OLSHYNLVWLUPUNH[[OLIV[[VTVM[OLÄS[LYVU[V
the plunger. Push the cartridge down and tilt back
\U[PS[OLÄS[LYPZHSPNULK^P[O[OL[VWWVZ[(SSV^[OL
ÄS[LY[VZSPKL\WV]LY[OL\WWLYÄ[[PUNHUKNLU[S`YV[H[L
left and right to insure a good seal with the top and
bottom O-rings.
9LWLH[HIV]LWYVJLK\YLMVYÄS[LYZHUK
Filters should always be installed in the 1,2,3 order
and removed in reverse – 3,2,1.
05:;(3305.;/,<=)<3)67;065
WARNING – ELECTRICAL SHOCK HAZARD.
4HRLZ\YLHSSWV^LYPZYLTV]LKMYVT[OLZ`Z[LT
Remove the top cover by removing the phillips head
screws that attach it. Gently lift the top cover up and
set it aside. The UV chamber will be visible in the top
of the lab unit.
Unclasp the clamps that hold the stainless-steel body
of the UV light. Gently lift the body up about 2 inches
to allow clearance for bulb removal.
Unscrew the blue collar which is threaded into the
stainless-steel body. Remove O-ring gasket and
stainless-steel washer.
Tabs pointing up
INSTALLATION & OPERATION MANUAL / 6

Hold the UV lamp by the plastic ends ONLY. To
H]VPKVPSZHUKÄUNLYWYPU[ZP[PZYLJVTTLUKLK[V\ZL
gloves when handling the UV bulb. DO NOT touch
the glass. If the lamp should become oily or dirty,
clean it with a clean soft tissue or cloth wetted with
isopropyl alcohol.
First, slide the O-ring gasket of the male, pronged
end of the UV bulb until it seats. Second, place
washer on top of gasket. Third, place the blue collar
on top of the gasket.
Connect the female wire connection to UV male prong
until it is fully seated.
Install the bulb into the stainless-steel chamber. Gently
thread the blue collar until hand tight.
Gently push UV body back into the bracket. Squeeze
JSHTWZ\U[PS[OLJSPJRÄYTS`HIV\[[OL<=IVK`
Replace the top cover and turn the unit on.
START UP PROCEDURE
Plug in the unit. The electrical outlet should be prop-
erly grounded and fused. It is recommended that the
system run on its own dedicated electrical circuit.
Open outlet valve and place a container underneath.
Slowly turn on the water supply to the unit and correct
any leaks at the supply connections.
Turn unit on at the power switch and allow water
[VÅV^>H[LY^PSSIL]PZPISLÄSSPUN[OLJHY[YPKNLZ\W
ÅV^MYVT[OLSLM[[V[OLYPNO[(SSV^[OLHPY[VISLLK
from the system through the outlet valve into the
open container.
When all air has been purged from the system close
the outlet valve.
0UZ[HSS[OLÄUHSJHWZ\SLÄS[LYVU[OLV\[SL[]HS]L6WLU
[OLV\[SL[]HS]LHUKISLLKVќHU`YLZPK\HSHPYPU[OL
ÄS[LYI`VWLUPUN[OL[O\TIZJYL^VU[OLZPKLVM[OL
ÄS[LY*SVZL[O\TIZJYL^^OLU^H[LYPZWYLZLU[HUK
close the outlet valve.
Check connections and cartridges for leaks.
“INITIAL POLISHING” OF NEW CARTRIDGES
Cartridge may require up to 24 hours to rinse up to
18 Megohms.
SP[LYZVM^H[LYZOV\SKILY\U[OV\NO[[OLÄS[LYZHZHU
initial purge of the system.
We recommend a sample be collected in an
appropriate sterile sample container and tested for
water purity. Periodic laboratory water testing to
JVUÄYT[OLLќLJ[P]LULZZVM[OL[YLH[TLU[WYVJLZZPZ
the owners’ responsibility.
7 / ARIES HIGH PURITY WATER SYSTEM

DISPLAY AND KEYPAD FUNCTIONS
PRIMARY BUTTONS
>HRL - Wakes the unit up from the Standby mode
and initiates the 30 minute polishing cycle.
(NOTE: The Wake key must be pressed longer than 3 second
or unit will not wake up.)
7V^LY- Turns on the power, starts the timer controller.
7VSPZOPUN - Green LED is on when the unit is running,
Solenoid, pump and meter are on.
:[HUK)` - Amber LED is on when the unit is in
Stand By mode. Solenoid, pump and meter are
powered down.
RESISTIVITY MONITOR
MONITOR OPERATION AND PROGRAMMING
TO CHECK WATER TEMPERATURE
Press and Hold TEMP
- The water temperature is displayed while the
button is pressed.
- The monitor returns to normal operation
when released.
- The displayed temperature can be toggled from
C to F by repeated pressing.
TO CHECK SET-POINT
Press and Release SET-POINT
-LO- will be displayed momentarily, followed by the
current stored value. The Low set-point value is
the point where the relay will energize.
-HI- will then be displayed followed by the High set-
point value. The High set-point value is the point
where the relay will de-energize.
- The monitor will automatically return to
normal operation.
INSTALLATION & OPERATION MANUAL / 8

9 / ARIES HIGH PURITY WATER SYSTEM
TO ADJUST SET-POINT
Press and Hold SET-POINT then also press ADJUST
Release both buttons
-LO- will be displayed momentarily, followed by the
current Low set-point value with the LCD’s
OPNOLZ[KPNP[ÅHZOPUN
- To skip this digit, Press and Release ADJUST again
OR, Press SET-POINT repeatedly until the desired
number is achieved.
Then Press and Release ADJUST
;OLUL_[SV^LZ[KPNP[^PSSILNPUÅHZOPUN
- Press SET-POINT to adjust the digit
- Press ADJUST to Skip.
Continue this sequence until all the digits are set and
press ADJUST
-HI- will be displayed momentarily, followed by the
current High set-point value with the LCD’s high-
LZ[KPNP[ÅHZOPUN
- Press SET-POINT to adjust the digit
- Press ADJUST to Skip.
Continue this sequence until all the digits are set and
press.
- The Monitor will automatically return to its normal
operating mode.
MAINTENANCE
This unit was designed and manufactured to have
the ability to produce the purest water that can be
VI[HPULKMVYSHIVYH[VY`\ZHNL0[W\YPÄLZ^H[LY[OH[
TLL[ZVYL_JLLKZ[OLZWLJPÄJH[PVUZYLJVTTLUKLK
by CLSI, ASTM, for Type I Reagent Grade Water
when it is operated and maintained according to
these instructions.
• The ultraviolet lamp (if installed) must be changed
once each year.
• The resin cartridges must be replaced when the
purity meter indicates the water purity is below your
requirements or every 6 months.
• The exterior surfaces may be wiped clean with a
mild detergent and a soft cloth to keep the unit
clean and new looking.
REPLACING THE FILTER CARTRIDGES AND
FINAL FILTER
The cartridges are expendable items which have a
ÄUP[LJHWHJP[`[VKLPVUPaLHUKWVSPZO[OL^H[LY>OLU
the digital display indicates the purity is less than your
minimum requirements, the DI resin cartridges have
exhausted their capacity to perform. Cartridge should
be replaced every 6 months as a matter of routine
THPU[LUHUJL;OLÄUHSÄS[LYZOV\SKILJOHUNLKL]LY`
time the cartridges are replaced.
;\YU[OL\UP[Vќ\ZPUN[OLWV^LYZ^P[JO
• Close feed water valve to the system.
• Place a container under the outlet valve to drain
water from the system, which releases pressure.
Open the outlet valve and leave open until water
stops draining from the system.

INSTALLATION & OPERATION MANUAL / 10
• Before removing DI cartridges, wrap the bottom
of each plunger with paper towels or a towel to
absorb the small water that will drain from the
IV[[VTVM[OLÄS[LYVUJLP[PZKPZLUNHNLKMYVT[OL
WS\UNPUNHWWHYH[\Z8\PJRS`WSHJLJHY[YPKNLÄSSLK
with water into a local basin or sink. Covering the
bottom cartridge with your hand will help minimized
the excess water.
• Remove the three cartridges starting with Cartridge
#3. Push down on the cartridge gently and tilt
MVY^HYK8\PJRS`YLTV]L[OLÄS[LYHUKKPZJHYKPU[OL
trash. There will be some residual water inside the
ÄS[LY>H[LY^PSSJVTLV\[[OLV\[SL[]HS]LZ[PSSVWLU
when the cartridge is depressed.
• Remove cartridges #2 and #1 in this order using
the same procedure above. When the plunger is
depressed some water will come out of the upper
Ä[[PUNUL_[[V[OLÄS[LY
• Dry the unit and replace the cartridges as described
in Section 5.2, Installing the Cartridge.
• The used cartridges may be disposed of in the same
THUULYHZHU`^H[LYÄS[LYJHY[YPKNL
• Replace the Final Filter with a new one each time the
ÄS[LYJHY[YPKNLZHYLYLWSHJLK
SANITIZATION
• Use Sanitization Kit HPA-010.
;\YUVќZ`Z[LTHUKKPZJVUULJ[WV^LYWS\N
• Open dispense valve.
• Remove cartridges.
9LTV]LTPJYVUÄUHSÄS[LY
• Insert sanitization cartridges in the following order:
CARTRIDGE 1 - SANITIZER, VP-17-0001
CARTRIDGE 2 - BLANK, VP-17-0000
CARTRIDGE 3 - BLANK, VP-17-0000
• Insert power plug into electrical outlet and turn
on system.
+YH^VќHIV\[[^V[VSP[LYZVM^H[LYMYVTZ`Z[LT
then close dispense valve.
• Allow the system to remain in recirculation for 45
to 60 minutes. *Note: The unit must remain in the
Polishing Mode while sanitizing. If the Aries HP water
system was manufactured after the year 2005, the
“WAKE” button will need to be pressed once within
[OLÄYZ[TPU\[LZVM[OLZHUP[PaH[PVUPU[LY]HSZV
that the water remains in recirculation. Pressing the
“WAKE” button interrupts system standby to start
30-minute run cycle.
(M[LYTPU\[LZVWLUKPZWLUZL]HS]LKYH^Vќ
approximately 250 ml and use test strips to verify a
minimum chlorine residual of 5 ppm.
• After 45 to 60 minutes, dispense about 20 liters
to drain.
(M[LYÅ\ZOPUN[OLZ`Z[LTVWLU[OLKPZWLUZL
]HS]LKYH^VќTSHUK\ZL[LZ[Z[YPWZ[VLUZ\YL
chlorine is removed.
;\YUVќZ`Z[LTHUKKPZJVUULJ[WV^LYWS\N
• Open dispense valve.
• Remove cartridges.
• Please refer to “Installing the Cartridge” and
“Start Up Procedure” sections

CONSUMABLES & SPARE PARTS
11 / ARIES HIGH PURITY WATER SYSTEM
PART NUMBER MODEL DESCRIPTION
1200014 Outlet Angle Valve
1200102 Resistivity Calibration Cell Block
ARA-WB Wall Mounting Bracket
HPA-003 UV Bulb for Combination UV System
HPA-008 220 Volt Operation Step Down Converter
HPA-010 Sanitization Kit (Includes Blank Filters, Sanitization Cartridge, Gloves & Test Strips)
HPA-011 O-ring Repair Kit for Upper Outlet and Plunger (3-sets)
HPA-013 3V^LY-P[[PUN9LWHPY2P[0UJS\KLZZL[ZVMWS\UNLYJ\WZWYPUNZVYPUNZHUKÄ[[PUNZ
HPA-014 <WWLY9LWHPY2P[0UJS\KLZZL[ZVM\WWLYÄ[[PUNZ
PF-00-6402 TPJYVUJHWZ\SLÄS[LY¹457;_¹OVZLIHYI
PF-00-6505-HN TPJYVU\S[YHÄS[LY¹457;_¹OVZLIHYI
PF-00-7101 POU 0.2-micron Endotoxin Removal Capsule Filter - 1/4” MNPT x 1/4” hose barb
PX-11-0001 Sanitization Cartridge Only
VP-17-0001 Single Sanitization Cartridge
VPK-3805 VP Series Cartridge Kit for Tap Water Feed (1-Organics Pretreatment & 2 Mixed Beds)
VPK-4010 VP Series Cartridge Kit for RO Feed (3 Mixed Beds)

TROUBLESHOOTING
SYMPTOM CAUSE REMEDY
Unit will not WAKE up Wake button pushed too quickly. WAKE button must be pushed at least 3
second to initiate the polishing mode.
+PZWSH`PZVќ No power to the unit or unit in
Stand By mode.
Check the unit is plugged into a live outlet.
Check the unit’s fuse - replace it if it is blown.
Press the Wake switch to initiate the polish
cycle.
Unit blows fuses. An electrical short exists in the unit. Return the unit to the factory for repairs or
have unit serviced by an authorized dealer.
Unit will not dispense water.
Supply water is not on.
Bad Solenoid.
(ÄS[LYJHY[YPKNLPZSVHKLK^P[OWHY[PJSLZ
and plugged.
5VÅV^
*VUÄYT[OH[[OLZ\WWS`^H[LYPZ[\YULKVU
Have unit serviced by an authorized dealer or
return to factory.
9LWSHJL[OLÄS[LYJHY[YPKNL
*HWZ\SLÄS[LYPZIV\UK^P[OHPY6WLU
bleed valve.
Display indicates water purity is below
set point.
The supply water is improperly treated.
The resin cartridges are not seated.
Resin is exhausted.
Test the supply water TDS add additional
treatment.
Open cartridge housing and reseat cartridge.
Install fresh cartridges.
Resin cartridges exhaust quickly. The supply water is improperly treated. Test the supply water TDS. Add additional
pretreatment to obtain higher quality.
:\ITPJYVUÄS[LYWS\NZYHWPKS` The supply water has colloids and is
improperly treated.
Perform a Silt Density Test. If it is higher than 1
add additional pretreatment.
INSTALLATION & OPERATION MANUAL / 12

WARRANTY
Systems and options have a one-year warranty from the date of purchase accompanied by system registration.
0MHU`JVTWVULU[MHPSZ[VM\UJ[PVUUVYTHSS`^P[OPU[OLÄYZ[`LHYYL[\YU[OL\UP[[V[OLMHJ[VY`WYLWHPK0MPU[OLVWPUPVU
of the factory, failure was due to defects in materials or workmanship, repair or replacement will be made without
charge. A reasonable service charge will be made for diagnosis or repairs not covered by this warranty. Aries
Filterworks / ResinTech, Inc. assumes no other responsibility or liability.
This warranty does not cover any product which has been subject to misuse, neglect, accident or used in violation
of operating instructions nor does it extend to any units altered or repaired for warranty defects by anyone other than
Aries Filterworks / ResinTech, Inc.
IN NO EVENT SHALL ARIES FILTEWORKS / RESINTECH, INC. BE LIABLE FOR ANY INCIDENTAL,
CONSEQUENTIAL, SPECIAL, INDIRECT, PUNITIVE OR EXEMPLARY DAMAGES OR LOST PROFITS FROM ANY
BREACH OF THIS WARRANTY OR OTHERWISE.
(YPLZ-PS[LY^VYRZ9LZPU;LJO0UJYLZLY]LZ[OLYPNO[[VTHRLJOHUNLZPUKLZPNUZWLJPÄJH[PVUZHUKWYPJLZ
without notice.
13 / ARIES HIGH PURITY WATER SYSTEM

ARIES WATER PURIFICATION MAINTENANCE LOG
Serial Number: Date Installed:
Asset # or ID: Location:
DATE MAINTENANCE PERFORMED BY COMMENTS
INSTALLATION & OPERATION MANUAL / 14

DATE MAINTENANCE PERFORMED BY COMMENTS