BAIR HUGGER OPERATOR’SMANUAL
Bair Hugger*Model 505
Temperature Management Unit
Betriebshandbuch für Bair Hugger
Temperaturmanagement-Geräte Modell 505
Manuel de l’utilisateur pour les appareils
de gestion de la température Bair Hugger
modèle 505
Manuale d’uso delle unità di gestione della
temperatura Bair Hugger modello 505
Manual del usuario para las unidades
de control de temperatura Bair Hugger,
modelo 505
Manual do utilizador das unidades de gestão
de temperatura Bair Hugger Modelo 505
Bair Hugger Temperatuurregelunit model 505
Gebruikshandleiding
Brugsanvisning til Bair Hugger temperatur-
reguleringsenheder, Model 505
Bruksanvisning till Bair Hugger modell 505
temperaturstyrenhet
Brukerhåndbok for Bair Hugger-temperatur-
behandlingsenhet, modell 505
Bair Hugger -lämpötilansäätölaitteen mallit
505 käyttäjän opas
Operator’sManual
Check the Bair Hugger therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #200977C

Bair Hugger Model 505 Temperature Management Unit
Operator’sManual
English
Deutsch
Français
Italiano
Español
Português
Nederlands
Dansk
Svenska
Norsk
Suomi
1
13
25
37
49
61
73
85
97
109
121

2 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
Table of Contents
Introduction
Description of the Total Temperature Management* System
Indications for Use
Contraindications
Warnings
Precautions
Important Information
Read Before Servicing Equipment
Setup and Operation
Temperature Management Unit
Control Panel Features of the Model 505 Temperature Management Unit
Mounting the Model 505 Temperature Management Unit
Attaching and Storing the Model 505 Hose
Blanket Dispenser
General Maintenance
Cabinet Cleaning
Technical Service and Order Placement
In-Warranty Repair and Exchange
When You Call for Technical Support
Specifications
Definition of Symbols
3
3
3
3
3
3
3
4
4
5
6
7
8
8
9
9
9
9
9
10
12

Bair Hugger Model 505 Temperature Management Unit Operator’s Manual English 3
Introduction
Description of the Total Temperature Management* System
The Bair Hugger brand Total Temperature Management system consists of a Bair Hugger forced-air
temperature management unit (with available rolling stand and sheet clip) and disposable compo-
nents, including Bair Hugger forced-air blankets, Bair Paws* patient warming gowns, and the 241*
blood/fluid warming set. The system and disposables are intended for use in clinical settings including
the operating room.
This manual includes operating instructions and unit specifications for the Model 505 forced-air
temperature management unit. For information about using Bair Hugger blankets, Bair Paws gowns,
or the 241 blood/fluid warming set with Bair Hugger units, refer to the “Instructions for Use”
included with each of these disposable components.
Indications for Use
The Bair Hugger brand Total Temperature Management system is intended to prevent and treat
hypothermia and to provide warmth to cold or shivering patients. In addition, the system should be
used whenever conditions exist that could cause patients to become cold.
Contraindication
Do not apply heat to lower extremities during aortic cross-clamping. Thermal injury may occur if
heat is applied to ischemic limbs.
Warnings
• Do not use Bair Hugger units with any forced-air blanket or cover other than Bair Hugger blankets
or Bair Paws gowns. Thermal injury may result.
• Do not use a forced-air warming device over transdermal medication. Increased drug delivery and
patient injury or death may occur.
• Do not provide warming therapy when in intra-aortic balloon pump is in use; thermal injury may
occur if heat is applied to ischemic limbs.
• Do not warm patients with the Bair Hugger unit hose alone. Thermal injury may result. Always attach
the hose to a Bair Hugger blanket or Bair Paws gown before providing skin surface warming therapy.
• Do not continue therapy if the Over Heat warning light illuminates and the audible alarm sounds.
Thermal injury may result. Turn the temperature management OFF and contact qualified technical
personnel. If using 241 fluid warming, immediately stop fluid flow and discard the fluid warming set.
• Do not initiate therapy unless the Model 505 temperature management unit is securely mounted or
injury may result.
Precautions
• Monitor the patient’s temperature and vital signs regularly according to institutional protocol.
Reduce air temperature or discontinue therapy when the therapeutic goal is reached or if vital sign
instability occurs. Notify physician immediately of vital sign instability.
• To prevent suffocation from misuse, do not leave children or infants unattended when administering
Bair Hugger therapy.
• See CONTRAINDICATION and WARNINGS before administering therapy. Read this Operator’s
Manual, blanket and gown instructions and 241 fluid warming set package instructions before use.
Important Information
EXPLOSION HAZARD
Do not use in the presence of flammable anesthetics.

4 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
ELECTRICAL SHOCK HAZARD
Do not disassemble the temperature management unit; refer to authorized technical personnel. There
are electrically live parts within the unit when it is connected to the power source, even when the
switches are in the OFF position.
ELECTRICAL INTERFERENCE
If radio frequency interference with monitoring equipment should occur, connect the temperature
management unit to a different power source.
Read Before Servicing Equipment
The repair, calibration, and servicing of the temperature management unit requires the skill of quali-
fied technical personnel who are familiar with good practice for medical device repair. If service is
designated as not requiring the manufacturer’s attention, the technical information is provided in the
Service Manual or will be provided upon request by Arizant Healthcare Inc.
REFER TO SERVICE MANUAL
Perform all repairs and maintenance in accordance with the instructions in the Service Manual.
SAFETY INSPECTION
Perform a safety inspection after making repairs to the temperature management unit and before
returning the unit to service. A safety inspection must include a test of the operating temperatures
(described in the Service Manual), the Over Heat Alarm system, as well as a leakage current test.
PROPER USE AND MAINTENANCE
Arizant Healthcare Inc. assumes no responsibility for the reliability, performance, or safety of the
equipment if:
• Modifications or repairs are performed by unauthorized personnel.
• The equipment is used in a manner other than that described in the Operation or Service Manuals.
• The equipment is installed in an environment that does not meet the relevant grounding requirements.
Setup and Operation
The Bair Hugger brand Total Temperature Management system is easy to set up and to use. Follow
the instructions provided with each Bair Hugger blanket or Bair Paws gown for specific information.
1) Place the blanket or gown on the patient with the perforated side (the side with small holes)
against the patient’s skin.
2) Insert the hose of the temperature management unit in the hose port on the blanket or gown. Use
a twisting motion to ensure a snug fit (see Figure A).
Figure A.

Bair Hugger Model 505 Temperature Management Unit Operator’s Manual English 5
3) Connect the unit to a properly grounded power source.
4) Press the System ON/OFF button to turn the unit ON and select the appropriate temperature
setting.
5) You may place a cotton blanket over the blanket or gown for maximum effectiveness.
6) Monitor the patient’s temperature at regular intervals according to institutional protocol and
adjust the temperature setting of the unit as required.
Temperature Management Unit
The temperature management unit uses a high-efficiency motor, a heating element, and a solid-state
temperature control to create a continuous flow of warm air to the blanket or gown. It is designed
for safe use in all areas, including the operating room.
The Model 505 temperature management unit can be attached to an IV pole or to the railing on a bed.
Figure C. Model 505 unit
attached to an IV pole
Figure B. Model 505 unit attached
to a bedrail

6 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
Control Panel Features of the Model 505 Temperature Management Unit
TEMPERATURE IN RANGE INDICATOR
The temperature in range indicator illuminates when the output air temperature is within the range
of the selected level.
MAIN POWER INDICATOR
The main power indicator illuminates when the unit is connected to a power source. This indicator
must be illuminated for any functions to operate.
SYSTEM ON/STANDBY
Push this button to turn the unit either ON or OFF. The indicator directly above the switch
illuminates when the unit is ON.
OVER HEAT INDICATOR
The Over Heat Indicator illuminates and an audible alarm sounds when an over-temperature condition
is detected. To reset, turn the unit OFF and then ON, using the System ON/STANDBY button.
(Also refer to the Warnings section of this manual.)
TEMPERATURE INDICATORS
The temperature indicators illuminate up to the selected temperature level. When the unit is initially
turned on, none of these indicators are illuminated and ambient air will be delivered.
TEMPERATURE SELECT
Push this button to increase the temperature setting level by level to the desired setting. When the
temperature setting is at HIGH, push the button again to return to delivery of ambient air.
Figure D. Control panel of the Model 505 unit

Bair Hugger Model 505 Temperature Management Unit Operator’s Manual English 7
Mounting the Model 505 Temperature Management Unit
USING AN IV POLE
The Model 505 unit clamps easily to an IV pole (see Figure E). Simply turn the handle clockwise to
tighten the clamp onto an IV pole, counterclockwise to release.
Warning: To prevent tipping, clamp the Model 505 unit to an IV pole at a height that provides
stability. We recommend clamping the unit no higher than 44" (112 cm) on an IV pole with a minimum
14" (35.6 cm) radius wheelbase. Failure to do so may result in IV pole tipping, catheter site trauma, and
patient injury.
USING ABEDRAIL
The Model 505 unit can also hang on the edge of a bed. The safety strap is designed to loop around
the bedrail, keeping the Model 505 unit safely suspended even if the unit is inadvertently dislodged
from the bedrail (see Figure F).
Figure E. Model 505 unit attached to
an IV pole
Figure F. Model 505 unit attached to
a bedrail
Safety strap
≤44 inches
>14 inches
_

8 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
Attaching and Storing the Model 505 Unit Hose
The Model 505 unit has a unique “snap-fit” hose. This extended-length swivel hose, adapted for
241 fluid warming, attaches by inserting the flange end at a 45° angle in the grooved blower outlet
and “snapping” the hose into place.
Press the white tab on the blower outlet to release the hose.
When storing the Model 505 unit, insert the hose clip tab in the hanger slot near the blower outlet.
Blanket Dispenser
The blanket dispenser attaches to the unit and can store several Bair Hugger blankets. See the
package instructions for details.
Figure G. Attaching the Model 505
unit hose
Insert flange in
groove and snap
into place
Press this tab
to release hose
Figure H. Storing the Model 505 unit hose
Hang hose clip tab
in this slot for storage
Figure I. Temperature management
unit with blanket dispenser
Blanket Dispenser

Bair Hugger Model 505 Temperature Management Unit Operator’s Manual English 9
General Maintenance
Cabinet Cleaning
1) Disconnect the temperature management unit from the power source before cleaning.
2) Use a damp soft cloth and a mild detergent to clean the unit cabinet. Dry with a separate
soft cloth.
Caution
• Do not use a dripping wet cloth to clean the cabinet. Moisture may seep into the electrical contacts,
damaging the components.
• Do not use alcohol or other solvents to clean the cabinet. Solvents may damage the labels and other
plastic parts.
Technical Service and Order Placement
USA, WORLDWIDE
TEL: 1-952-947-1200 FAX: 1-952-947-1400
1-800-733-7775
WITHIN EUROPE
TEL: +49-4154-9934-0 FAX: +49-4154-9934-20
0800-100-1236 (Toll-free in Germany) 0800-100-1324 (Toll-free in Germany)
In-Warranty Repair and Exchange
Replacement parts to correct a problem are delivered at no charge. To return a device to Arizant
Healthcare Inc. for service, first obtain a Return Authorization (RA) number from a technical service
representative. Please use this number on all correspondence when returning a device for service. A
shipping carton will be delivered to you at no charge, if needed. Call your local supplier or sales rep-
resentative to inquire about loaner devices while your device is being serviced.
When You Call for Technical Support
Remember, we will need to know the serial number
of your unit when you call us. On Model 505 units,
the serial number label is affixed to the rear panel.
Figure J. Serial number label on Model 505
Serial number label

10 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
Specifications
Model 505
Physical Characteristics
DIMENSIONS 13 in. high x 10 in. deep x 11 in. wide
33 cm high x 25 cm deep x 28 cm wide
WEIGHT 13.6 lb; 6.2 kg
MOUNTING IV pole clamp, bedrail hook with safety strap;
can be placed on hard surface
RELATIVE NOISE LEVEL 53 decibels
HOSE Detachable, flexible, washable; compatible
with the 241 fluid warming system
FILTRATION SYSTEM 0.2µM level
RECOMMENDED FILTER CHANGE Every 500 hours of use
Temperature Characteristics
TEMPERATURE CONTROL Electronically controlled using a thermocouple
sensor
HEAT GENERATED 1800 BTUs (average)
SYSTEM TIME TO 100°F (37.7°C) ~17 secs
OPERATING TEMPERATURES Air temperatures reaching the patient are
approximately 2°C lower than the listed
temperatures.
Average temperatures at the end of the hose:
HIGH: 43° ± 3°C 109.4° ± 5.4°F
MED: 38° ± 3°C 100.4° ± 5.4°F
LOW: 32° ± 3°C 89.6° ± 5.4°F
Safety System
THERMOSTAT Independent bulb and capillary
OVERCURRENT PROTECTION Dual fused input lines
ALARM SYSTEM Over-heat: flashing red light with audible
alarm; heater shuts down
CERTIFICATIONS UL 544, CSA C22.2 No. 125, IEC 601-1,
IEC 601-1-2, EN55014, AS 3200.1990
CLASSIFICATION Classified under IEC 601-1 Guidelines as Class I,
Type BF, Ordinary equipment, Continuous operation

Bair Hugger Model 505 Temperature Management Unit Operator’s Manual English 11
Electrical Characteristics
BLOWER MOTOR Operating speed: 3150 rpm
Airflow: 28-30 cfm
POWER CONSUMPTION Peak: 1000W
Average: 450W
LEAKAGE CURRENT <100µA
HEATING ELEMENT 850W Resistive
POWER CORD 15-foot, SJT, 3 cond., 10A
4.6 meter, HAR, 3 cond., 10A
DEVICE RATINGS 110-120 VAC, 60 Hz, 9.5 Amperes, or
220-240 VAC, 50 Hz, 4.5 Amperes, or
100 VAC, 50/60 Hz, 10 Amperes
FUSES 10A, 200mA (110 - 120 VAC Units)
6.3A, 100mA (220 - 240 VAC Units)
15A, 160mA (100 VAC Units)
DIAGNOSTICS Over-heat test can be performed by the
biomedical group.

12 English Bair Hugger Model 505 Temperature Management Unit Operator’s Manual
Definition of Symbols
!
VAC
V
ON/STANDBY
ON (used on isolation switch)
OFF (used on isolation switch)
ON/OFF push button switch
Temperature Control
Equipotentiality plug (Ground)
Fuse
Attention (see appropriate documents)
Non Explosion-Proof
Dangerous Voltage
Type BF Equipment (patient applied)
Voltage, Alternating Current (AC)

Bair Hugger Temperaturmanagement-Geräte Modell 505
Betriebshandbuch
English
Deutsch
Français
Italiano
Español
Português
Nederlands
Dansk
Svenska
Norsk
Suomi
1
13
37
55
73
91
109
127
145
163
181

Bair Hugger Temperaturmanagement-Geräte Modell 505 Betriebshandbuch
14 Deutsch
Inhaltsverzeichnis
Einleitung
Bair Hugger Komplett-Temperaturmanagement-System
Indikationen
Kontraindikationen
Warnhinweise
Vorsichtsmaßnahmen
Wichtige Informationen
Wartungsanweisungen
Vorbereitung und Betrieb
Temperaturmanagement-Gerätes
Bedienungsfeld des Wärmegeräts Modell 505 Temperaturmanagement-Gerät
Befestigung des Temperaturmanagement-Gerätes Modell 505
Befestigung und Aufbewahrung des Schlauchs des Modells 505
Deckenbehälter
Allgemeine Wartung
Gehäusereinigung
Technischer Kundendienst und Bestellannahme
Reparaturen und Produktrückgabe im Rahmen der Garantie
Technische Hilfe per Telefon
Technische Angaben
Symbole
15
15
15
15
15
15
15
16
16
17
18
19
20
20
21
21
21
21
21
22
24

Bair Hugger Temperaturmanagement-Geräte Modell 505 Betriebshandbuch Deutsch 15
Einleitung
Bair Hugger Komplett-Temperaturmanagement-System
Das Komplett-Temperaturmanagement-system der Marke Bair Hugger besteht aus der konvektiven
Temperaturregelungseinheit (lieferbar mit Rollgestell und Deckenclips) und Einwegkomponenten,
einschließlich konvektiver Wärmedecken und dem 241* Blut-/Flüssigkeitswärmeset. Das System ist
zur Anwendung in allen klinischen Einsatzbereichen, einschließlich dem Operationssaal, geeignet.
Dieses Betriebshandbuch enthält die Betriebsanleitung und die technischen Gerätedaten für die kon-
vektiven Temperaturmanagementgeräte Modell 505. Informationen zur Benutzung der Bair Hugger
Decken oder des 241 Blut-/Flüssigkeitswärmeset mit Bair Hugger Geräten, entnehmen Sie bitte der
„Gebrauchsanweisung“, die jeder dieser Einwegkomponenten beiliegt.
Indikationen
Das Komplett-Temperaturmanagement-system der Marke Bair Hugger dient dazu, Hypothermie zu
verhindern und zu behandeln, und kalte oder an Kältezittern leitende Patienten, mit Wärme zu ver-
sorgen. Außerdem, sollte das System immer dann eingesetzt werden, wenn Umstände vorliegen, auf-
grund derer Patienten hypotherm werden könnten.
Kontraindikationen
Bei Aortenabklemmung darf den unteren Gliedmaßen des Patienten keine Wärme zugeführt werden.
Die Wärmezufuhr zu ischämischen Gliedmaßen kann zu Verbrennungen führen.
Warnhinweise
• Bair Hugger Geräte niemals mit anderen Wärmedecken oder sonstigen Abdeckungen verwenden.
Bair Hugger Geräte ausschließlich mit Bair Hugger Wärmedecken verwenden, sonst könnte es zu
Verbrennungen kommen.
• Verwenden Sie kein Warmluft-Wärmegerät bei transdermaler Medikation. Dies könnte eine
erhöhte Wirkstofffreisetzung, Verletzungen oder den Tod des Patienten verursachen.
• Verabreichen Sie keine Wärmetherapie, wenn eine intraaortale Ballonpumpe im Einsatz ist. Die
Wärmezufuhr zu ischämischen Gliedmaßen kann zu Verbrennungen führen.
• Patienten nicht nur mit dem Schlauch des Bair Hugger Gerätes Wärme verabreichen, sonst könnte
es zu Verbrennungen kommen. Vor der Wärmebehandlung den Schlauch immer zuerst an eine Bair
Hugger Decke anschließen.
• Wärmebehandlung abbrechen, wenn die Überhitzungsanzeige aufleuchtet und der Alarmton zu hören
ist, sonst könnte es zu Verbrennungen kommen. Das Temperaturmanagementgerät ausschalten,
Netzstecker herausziehen und qualifiziertes Fachpersonal benachrichtigen. Bei Anwendung der 241
Flüssigkeitswärmung, sofort den Flüssigkeitsstrom unterbrechen und das Flüssigkeitswärmeset entsorgen.
• Eine Behandlung mit dem Temperaturmanagement-Gerät Modell 505 immer erst dann beginnen,
wenn es sicher montiert worden ist, da es sonst zu Verletzungen kommen kann.
Vorsichtsmaßnahmen
• Überwachen Sie entsprechend der institutionellen Verfahren regelmäßig die Temperatur und
Vitalfunktionen des Patienten. Die Temperatur verringern oder die Behandlung abbrechen, wenn das
Behandlungsziel erreicht wurde oder wenn es zu einer Instabilität der Vitalfunktionen kommt.
Unverzüglich einen Arzt bei Instabilität der Vitalfunktionen unterrichten.
• Kinder und Säuglinge bei der Anwendung der Bair Hugger Therapie niemals unbeaufsichtigt lassen,
um ein Ersticken durch unsachgemäße Benutzung des Wärmesystems zu vermeiden.
• Vor Behandlungsbeginn alle KONTRAINDIKATIONEN und WARNHINWEISE lesen. Vor
Benutzung dieses Betriebshandbuch und die Anleitungen für die Wärmedecken sowie für die 241-
Flüssigkeitswärmesets lesen.

Bair Hugger Temperaturmanagement-Geräte Modell 505 Betriebshandbuch
16 Deutsch
Wichtige Informationen
EXPLOSIONSGEFAHR
Nicht im Bereich von entflammbaren Anästhetika verwenden.
GEFÄHRLICHE ELEKTRISCHE SPANNUNG
Temperaturmanagement-Gerät nicht auseinander nehmen; dies ist die Aufgabe von qualifiziertem
Fachpersonal. In einem an das Netz angeschlossenen Gerät stehen bestimmte elektrische
Komponenten auch bei ausgeschalteten Schaltern unter Strom.
ELEKTRISCHE STÖRUNGEN
Bei Auftreten von Frequenzstörungen in Überwachungsgeräten, das Temperaturmanagement-Geräte
an eine andere Netzquelle anschließen.
Wartungsanweisungen
Reparatur, Kalibrierung und Wartung des Temperaturmanagement-Gerätes darf nur von qualifiziertem
Fachpersonal, das für die Arbeit mit Medizingeräten ausgebildet worden ist, durchgeführt werden. Für alle
Wartungsarbeiten, die nicht vom Hersteller ausgeführt werden müssen, sind tech-nische Informationen im
Wartungshandbuch zu finden oder können bei Arizant Healthcare Inc. angefordert werden.
WARTUNGSHANDBUCH
Alle Reparaturen und Wartungsarbeiten sind gemäß den Anleitungen im Wartungshandbuch
durchzuführen.
SICHERHEITSPRÜFUNG
Nach Reparaturen am Temperaturmanagement-Gerät und vor seiner erneuten Inbetriebnahme muss das
Gerät einer Sicherheitsprüfung unterzogen werden. Hierzu gehören die Überprüfung der Betriebstemper-
aturen (wie im Wartungshandbuch beschrieben) und des Überhitzungswarnsystems sowie ein Verluststromtest.
SACHGERECHTER BETRIEB UND WARTUNG
Arizant Healthcare Inc. übernimmt keine Verantwortung für die Zuverlässigkeit, Funktion oder
Sicherheit des Geräts, wenn:
• Veränderungen oder Reparaturen von nicht autorisiertem Personal durchgeführt werden.
• das Gerät nicht gemäß den Anleitungen im Betriebs- oder Wartungshandbuch verwendet wird.
• das Gerät in einer Umgebung verwendet wird, in der es nicht vorschriftsmäßig geerdet werden kann.
Vorbereitung und Betrieb
Das Bair Hugger Komplett-Temperaturmanagement-system ist einfach vorzubereiten und zu verwen-
den. Bitte befolgen Sie die den einzelnen Wärmedecken beigelegten ausführlichen Anweisungen.
1) Die Bair Hugger Wärmedecke so auf den Patienten legen, dass die perforierte Seite (die Seite mit
den kleinen Löchern) zum Patienten hin zeigt.
2) Den Schlauch des Temperaturmanagement-Gerätes in den Schlauchanschluss der Decke einsteck-
en. Mit Hilfe einer Drehbewegung einen festen Sitz gewährleisten (siehe Abbildung A).
Abbildung A.

3) Das Gerät an eine vorschriftsmäßig geerdeten Stromquelle anschließen.
4) Das Gerät mit dem EIN/AUS-Schalter einschalten und die entsprechende Temperatureinstellung
wählen.
5) Um eine optimale Wirkung zu erzielen, eine Baumwolldecke auf die Bair Hugger Decke legen.
6) Überwachen Sie entsprechend der institutionellen Verfahren regelmäßig die Temperatur und die
Temperatureinstellung am Bair Hugger Gerät je nach Bedarf anpassen.
Temperaturmanagement-Geräte
Die Temperaturmanagement-Geräte sind mit einem Hochleistungsmotor, einem Heizelement und
einer Temperaturregelung ausgestattet und erzeugen so einen kontinuierlichen Warmluftstrom zur
Decke. Das Modell 505 wurden zur Anwendung auf allen Stationen, einschließlich dem
Operationssaal, entwickelt.
Das Temperaturmanagement-Gerät kann an einem Infusionsständer oder an der Bettschiene montiert
werden.
Bair Hugger Temperaturmanagement-Geräte Modell 505 Betriebshandbuch Deutsch 17
Abbildung C. Gerät Modell 505
an einer Infusionsstange befestigt
Abbildung B. Temperaturmanagement-
Gerät Modell 505

Bair Hugger Temperaturmanagement-Geräte Modell 505 Betriebshandbuch
18 Deutsch
Bedienungsfeld des Wärmegeräts Modell 505 Temperaturmanagement-Gerät
„TEMPERATUR IM BEREICH“-ANZEIGE
Diese Anzeige leuchtet, wenn die Temperatur der abgegebenen Warmluft sich im Bereich der
eingestellten Temperatur befindet.
NETZSTROMANZEIGE
Die Netzstromanzeige leuchtet, wenn das Gerät an eine Stromquelle angeschlossen ist. Das Gerät ist
nur bei Aufleuchten der Netzstromanzeige funktionsbereit.
SYSTEM EIN/BEREITSCHAFTSMODUS
Durch Drücken dieser Taste das Gerät entweder ein- oder ausschalten. Bei eingeschaltetem Gerät
leuchtet die Anzeige direkt über dem Schalter.
ÜBERHITZUNGSANZEIGE
Bei Überhitzung leuchtet die Überhitzungsanzeige auf und ein Alarmton ist zu hören. Zum
Rückstellen, das Gerät am „System EIN/BEREITSCHAFTSMODUS“- Schalter aus- und dann wieder
einschalten. (Siehe auch Abschnitt Warnhinweise in diesem Handbuch).
TEMPERATURANZEIGEN
Die Anzeigen leuchten entsprechend der gewählten Temperatur auf. Beim Einschalten des Geräts
leuchtet keine dieser Anzeigen und es wird Luft mit Umgebungstemperatur abgegeben.
TEMPERATURWAHL
Auf diese Taste drücken, um die Temperatur stufenweise bis zur gewünschten Einstellung zu
erhöhen. Wenn die Temperatureinstellung HOCH anzeigt, die Taste erneut drücken, um auf Luft
mit Umgebungstemperatur zurückzustellen.
Abbildung D: Bedienungsfeld des Geräts Modell 505

Bair Hugger Temperaturmanagement-Geräte Modell 505 Betriebshandbuch Deutsch 19
Befestigung des Temperaturmanagement-Gerätes Modell 505
BEFESTIGUNG AN EINEM INFUSIONSSTÄNDER
Das Gerät Modell 505 lässt sich leicht an einem Infusionsständer befestigen (siehe Abbildung E). Die
Flügelschraube zum Festklemmen am Infusionsständer im Uhrzeigersinn bzw. zum Lösen gegen den
Uhrzeigersinn drehen.
WARNUNG: Um ein Umkippen zu verhindern, das Gerät Modell 505 am Infusionsständer auf einer
Höhe anbringen, die Stabilität gewährleistet. Es wird empfohlen, das Gerät nicht höher als 112 cm zu
befestigen. Der Radius der Basis sollte nicht kleiner als 36 cm sein. Nichtbeachtung kann zum Umkippen
des Infusionsständers, Trauma am Kathetereinlass und Verletzung des Patienten führen.
BEFESTIGUNG AN DER BETTSCHIENE
Das Gerät Modell 505 kann auch an der Bettschiene befestigt werden. Der Sicherheitsriemen wird
zur zusätzlichen Absicherung um die Bettschiene geschlungen, so dass das Gerät auch bei zufälligem
Verschieben an der Schiene bleibt (siehe Abbildung F).
Abbildung E: Gerät Modell 505 an
einer Infusionsstange befestigt
Abbildung F: Gerät Modell 505 an
der Bettschiene befestigt
Sicherheitsriemen
≤112 cm
>36 cm
_