Arizant Ranger 245 User manual

OPERATOR’SManual
Blood/FluiBlood/Fluid Wd Warming Systemarming SystemBlood/Fluid Warming System
Ranger
Blood/Fluid Warming System
Ranger
Blut-/Flüssigkeitswärmesystem
Betriebshandbuch
Ranger
Dispositif de réchauffement
du sang et des liquides
Manuel de l’utilisateur
Ranger
Sistema per il riscaldamento
di sangue e di fluidi
Manuale d’uso
Ranger
Sistema de calentamiento
para sangre y fluidos
Manual del usuario
Ranger
Sistema de aquecimento de sangue/fluidos
Manual do utilizador
Ranger
Bloed/vloeistofverwarmingssysteem
Gebruiksaanwijzing
Ranger
Varmesystem til blod og væsker
Brugsanvisning
Ranger
Uppvärmningssystem för blod/vätska
Användarhandbok
Ranger
Oppvarmingssystem for blod/væske
Bruksanvisning
Ranger
Veren/nesteiden lämmitysjärjestelmä
Käyttäjän opas
Ranger
™‡ÛÙËÌ· ı¤ÚÌ·ÓÛ˘ ·›Ì·ÙÔ˜/ ˘ÁÚÒÓ
∂Á¯ÂÈÚ›‰ÈÔ ÃÂÈÚÈÛÙ‹
Ranger
*
*
Check the Ranger system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #200750C
*
Check the Ranger system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #200750C
Check the Ranger system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #241434A

OPERATOR’SManual
English
Deutsch
Français
Italiano
Español
Português
Nederlands
Dansk
Svenska
Norsk
Suomi
∂ÏÏËÓÈο
1
13
27
41
55
69
83
97
111
125
139
153
167

Table of Contents
Technical Service and Order Placement
Introduction
Indications for Use
Warnings and Precautions
Product Description
Ranger Warming Unit
Ranger Disposable Set
Distal Fluid Temperatures
Product Safety Features
Instructions for Use
Preparation and Setup of the Ranger Warming Unit
Removing the Disposable Set From the Ranger Warming Unit
Transferring the Disposable Set From One Ranger Warming Unit to Another
Troubleshooting
Maintenance and Storage
Specifications
Definition of Symbols
Ranger Blood/Fluid Warming System
English
1
3
3
3
4
4
4
5
5
6
6
7
7
8
10
11
12
2

Technical Service and Order Placement
Technical service
USA
TEL: +1-952-947-1200
FAX: +1-952-947-1400
GERMANY
TEL: +49-4154-9934-0
FAX: +49-4154-9934-20
FRANCE
TEL: +33 4 42 98 14 00
Indigo (in France only): 0 820 33 90 40
FAX: +33 4 42 98 14 01
Order placement
USA
TEL: +1-952-947-1200
FAX: +1-952-947-1400
In-warranty repair and exchange
Replacement parts to correct a problem are delivered at no charge. To return a device to
Arizant Healthcare Inc. for service, first obtain a Return Authorization (RA) number from a
customer service representative. Please use the (RA) number on all correspondence when
returning a device for service. A shipping carton will be delivered to you at no charge, if
needed. We will service and ship your device within five (5) working days of our receipt.
Call your local supplier or sales representative to inquire about loaner devices while your
device is being serviced.
When you call for technical support
Remember, we will need to know the serial number of your unit when you call us. The serial
number label is located on the bottom of the warming unit.
Ranger Blood/Fluid Warming System
2
English

Introduction
The Ranger blood/fluid warming system includes a warming unit and a disposable IV
blood/fluid administration set. The Ranger warming system is designed to warm blood, blood
products, and liquids and deliver these at flow rates from KVO to 500 ml/min. At these flow
rates, the device maintains fluid output temperatures ranging from 33°C to 41°C (Note: This
is for room temperature fluids only). It takes less than 2 minutes to warm up to the 41°C set
point temperature.
Two disposable blood/fluid sets are available: one for standard flow applications and one for
high flow applications. Disposable sets are sterile, latex-free, single-use-only items and are
designed to be used with the reusable warming unit.
The Ranger warming unit is designed to be mounted to an IV pole. A handle located on the
top of the unit makes transport easy. When mounted to the IV pole, the unit fits easily above
the Bair Hugger model 505 warming unit or above a model 500/OR.
Indications for use
The Ranger warming system is intended to warm blood, blood products, and liquids.
Warnings and precautions
DANGER
Explosion hazard. Do not use in the presence of flammable anesthetics.
WARNING
•Electrical shock hazard. Do not open the warming unit case.
•Do not substitute other devices for the Ranger warming unit or Ranger disposable sets.
Thermal or electrical injury or device damage may occur.
•Never infuse fluids if air bubbles are present in the fluid line, as air embolism may result.
•Do not continue use of the unit if the over-temperature alarm continues to sound and the
temperature does not return to the set point temperature. Immediately stop fluid flow and
discard the disposable set. Have the warming unit tested by a biomedical technician or call
Arizant Healthcare Customer Service.
CAUTION
Do not immerse the blood/fluid warming unit. Wipe with a clean, slightly damp cloth.
Ranger Blood/Fluid Warming System
English
3

Product Description
The Ranger blood/fluid warming system consists of a warming unit and a sterile disposable IV
blood/fluid delivery set.
The RANGER warming unit
The warming unit is a compact, lightweight, liquid-resistant device with a clamp located on
the side for attachment to an IV pole. A carrying handle on the top of the unit makes it easy to
transport. Located on the front panel you will find:
•Alphanumeric display that indicates the heater temperature during normal operation. In an
over temperature condition, the display alternately flashes a temperature of 43˚C or higher
and the word “HI.” In an under-temperature condition, the display alternately flashes a
temperature of 33˚C or lower and word “LO.”
•Alarm indicator light that comes on when either an over- or under-temperature
condition occurs.
RANGER disposable set
Blood/fluid administration sets include a warming cassette (heat exchanger), tubing and
connectors compatible with hospital standard IV blood/fluid administration sets, bubble trap,
and injection port. The High Flow set also includes 2 blood/fluid bag spikes and a 150 micron
filter located in a central drip chamber.
Disposable irrigation sets deliver 30 L/hr flow rates and 142 ml priming volume. They
include spike sets and a 183 cm patient line.
Refer to the instructions provided with each disposable set for information on use.
Ranger Blood/Fluid Warming System
4
English
Slot for warming
cassette
Alphanumeric
display Bubble trap holder
Power switch
IV pole clamp
Warming Cassette
Alarm indicator light
Flow Rate Priming Volume Patient Line Length
Standard Flow Set 150 ml/mn 39-44 ml 76 cm
152 cm (76 cm patient line
plus 76 cm extension)
High Flow Set 100-150 ml/min 150 ml 152 cm (76 cm patient line
plus 76 cm extension)
Pediatric/Neonate 6 L/hr 20 cc 46 cm
Set

Distal fluid temperatures
Nominal measurements only. Temperatures taken at 18°C room temperature.
Product safety features
The following chart describes the safety alarm features of the Ranger warming unit.
*Transient conditions may cause an over-temperature alarm condition. These conditions include:
•There was an extreme change in flow rates (e.g., from 500 mL/min to stop flow).
•The unit was turned on and reached the set point temperature of 41˚C before the warming
cassette was inserted in the device.
•Fluids were prewarmed to above 42˚C before being infused.
FLOW RATE (L/HR)
0.5
1.0
2.0
3.0
4.5
6.0
9.0
15.0
20.0
30.0
USING 10°C INLET FLUID
32˚C
36˚C
39˚C
39˚C
39˚C
39˚C
38˚C
-
-
-
USING 20°C INLET FLUID
32˚C
36˚C
38˚C
40˚C
40˚C
39˚C
38˚C
36˚C
35˚C
33˚C
Ranger Blood/Fluid Warming System
English
5
ALARM TYPE
Over-temperature
alarm - 43˚C
Under-temperature
alarm - 33˚C
Independent backup
safety alarm - 46ºC
WHAT TO LOOK FOR
Alarm indicator light
illuminates and alarm
sounds, alphanumeric
display alternately
flashes a temperature
of 43˚C or higher and
the word “HI.”
Alarm indicator light
illuminates and alarm
sounds, alphanumeric
display alternately
flashes a temperature
of 33˚C or lower and
the word “LO.”
•Alarm indicator light
illuminates and alarm
sounds, alphanumeric
display alternately flashes
a temperature of 43˚C or
higher and the word
“HI.”
•Alphanumeric dis-
play is dark, alarm
sounds (backup safety
alarm still works even
if display is dark).
DESCRIPTION/CAUSE
Heater temperature
rose to 43˚C because of
transient conditions.*
Heater temperature
has fallen to 33˚C.
Heater temperature
rose to 46˚C. Safety
backup system is acti-
vated at 46˚C and
unit shuts off power
to the heating plates.
ACTION
Observe alphanumeric
display. If tempera-
ture does not drop to
41˚C (may take a few
minutes), discontinue
use of unit. Call
Arizant Healthcare
Customer Service.
Alarms stop when temper-
ature rises above 33ºC.
Continue use of unit.
If the temperature does
not rise above 33ºC,
unplug the unit and
call Arizant Healthcare
Customer Service.
TURN OFF THE
UNIT AND
UNPLUG IT. Do not
use the unit. Discard
the disposable set.
Call Arizant Healthcare
Customer Service.

Instructions for Use
Preparation and setup of the Ranger warming unit
DANGER
Explosion hazard. Do not use in the presence of flammable anesthetics.
1. Attach the Ranger warming unit to the IV pole. Tighten the pole clamp securely.
CAUTION:To prevent tipping, clamp the Ranger warming unit to an IV pole with a
minimum 14” (35.6 cm) radius wheelbase and at a height no higher than 44” (112 cm). Failure
to do so may result in damage to the product or catheter site trauma.
2. Slide the fluid warming cassette into the slot in the warming unit. The cassette can only fit
into the device one way.
3. Prime the disposable set. For more information about priming the set, refer to instructions
provided with the disposable sets.
4. Place the bubble trap in the holder.
5. Plug the power cord into an appropriate outlet. Turn the unit ON. In a few seconds the
alphanumeric display will illuminate. It takes less than 2 minutes to warm up to the 41ºC
set point temperature.
6. Begin infusion. When infusion is completed, remove the disposable set and discard
according to institutional protocol.
Ranger Blood/Fluid Warming System
6
English
44” (112 cm)
14” (35.6 cm)

Removing the disposable set from the Ranger warming unit
1. Close the inlet clamp proximal to the fluid warming cassette and open all clamps distal
to the cassette.
2. Disconnect the disposable set from the fluid source, if applicable.
3. Allow fluid to flow to the patient (this may take 2-3 seconds). Close a distal clamp.
4. Remove the fluid warming cassette from the warming unit and discard according to
institutional protocol.
5. Reconnect the patient IV line to the fluid source to continue the infusion without warming.
Transferring the disposable set from one Ranger warming unit to another
1. Follow steps 1-3 above, then remove the disposable set from the first warming unit.
2. During transport, keep the clamps closed and do not infuse fluids while the warming
cassette is outside of the warming unit.
3. Slide the warming cassette into the second warming unit.
4. Ensure that air is removed from the tubing.
5. Open the clamps and continue the infusion.
Ranger Blood/Fluid Warming System
English
7

Condition
Nothing illuminates on the
warming unit panel.
Condition
Alarm indicator illuminates
and alarm sounds, alpha-
numeric display alternately
flashes a temperature of 43˚C
or higher and the word “HI.”
Condition
Alarm sounds, alphanumeric
display and alarm indicator
light go dark.
Cause
•Unit is not turned on, plugged
in, or power cord is not plugged
into an appropriate outlet.
•Unit failure.
Cause
Temporary over-temperature
condition because:
•An extreme change in flow
rates occurred (e.g., from 500
ml/min to stop flow).
•Unit was turned on and
reached set point temperature
before warming cassette
was inserted.
•Fluids were prewarmed to
above 42°C before being run
through the warming unit.
Cause
Primary controller failure.
Unit will no longer operate.
Solution
•Turn unit on. Make sure the
power cord is plugged into the
power entry module of the
warming unit. Make sure the
warming unit is plugged into
a properly grounded outlet.
•Call Arizant Healthcare
Customer Service.
Solution
•Open flow to reduce tem-
perature. Alarms will stop
when the display reads 41°C.
The unit is ready to use.
•Alarms will stop when the
display reads 41°C. The unit
is ready to use.
•Turn off unit. Discontinue
infusion of fluids. Do not
warm fluids before infusing
them through the Ranger unit.
Solution
Power to heating plates will
shut off if temperature rises
to 46˚C. Turn unit off and
unplug it. Discontinue use of
unit. Discard disposable set.
Alarm will continue to sound
if you do not unplug unit.
Call Arizant Healthcare
Customer Service.
Ranger Blood/Fluid Warming System
8
English
Troubleshooting

Condition
•Unit alarms soon after plug-
ging it in (unit does not have
to be turned on for this con-
dition to occur).
•Heater temperature rises to
46˚C and unit shuts down
soon after plugging it in (unit
does not have to be turned on
for this condition to occur).
Condition
Alarm sounds but unit has
been turned off.
Condition
Cannot remove warming
cassette from unit.
Condition
Alarm indicator light illumi-
nates and alarm sounds,
alphanumeric display alter-
nately flashes a temperature of
33°C or below and the word
“LO.”
Condition
Alphanumeric display reads
“Er 4” or “Open.”
Condition
Alphanumeric display reads
“Er 5” or “Open.”
Cause
Test screw on bottom of unit
is loose or missing.
Cause
Independent backup safety
system has been activated.
Cause
•Warming cassette is too full,
fluids are still being infused,
or clamp is open proximal to
the warming cassette.
•Warming unit is below
patient level, creating
excessive back pressure.
Cause
Under temperature condition
caused by very high flow
using very cold fluid, or
defective heater/relay.
Cause
Open wire on temperature sensor.
Cause
Electrical interference.
Solution
Make sure test screw is com-
pletely tightened. If it is missing,
turn unit off and unplug it.
Call Arizant Healthcare
Customer Service.
Solution
Unplug unit. Call Arizant
Healthcare Customer Service.
Solution
•Make sure fluid is drained
from warming cassette before
sliding out the cassette, that
fluids are no longer infusing, and
that clamp is closed proximal
to the warming cassette.
•Raise unit above patient level.
Solution
Alarm should stop when
temperature rises above 33°C.
If alarm continues, unplug unit
and discontinue use. Call Arizant
Healthcare Customer Service.
Solution
Do not use unit. Call Arizant
Healthcare Customer Service.
Solution
Remove the unit. Refer to
biomedical technician or
call Arizant Healthcare
Customer Service.
Ranger Blood/Fluid Warming System
English
9

Maintenance and Storage
CAUTION
• Do not immerse the warming unit in cleaning or sterilizing solutions. The unit is not liquid-
proof.
• Do not clean the warming unit with solvents. Damage to the case, label, and internal compo-
nents may result.
Cleaning
1. Disconnect the Ranger warming unit from the power source.
2. Wipe the outside of the unit with warm, soapy water, nonabrasive cleaning solutions, dilute
bleach, or cold sterilants. Do not use abrasive materials.
3. Wipe with a dry, soft cloth.
TO CLEAN THE INSIDE OF THE UNIT:
The Ranger cleaning tool is intended to clean both heating plates of the warming unit.
CAUTION
• Do not allow spills to dry inside the unit, as this may make it more difficult to clean the unit.
•The cleaning tool provides only superficial cleaning–it does not disinfect or sterilize the
interior of the unit.
1. Unplug the warming unit.
2. Unfold the cleaning tool. Wet the foam pads with a nonabrasive solution, such as Alconox*
brand detergent.
3. Insert the tool from the back of the unit and pull the tool all the way out from the front.
4. Rinse the tool with water and repeat 3 times.
5. Wipe off the unit to remove excess fluid.
Note: If the cleaning tool cannot be inserted through the slot of the warming unit because of
excess dried-on fluids, send the unit to a biomedical technician.
Do not use abrasive materials or solutions to clean the heater plates.
Storage
Store all components in a cool, dry place when not in use.
Servicing
There are no user-serviceable parts in the Ranger warming unit. All service must be per-
formed by Arizant Healthcare Inc. or an authorized service technician.
Call Arizant Healthcare Customer Service at 800-733-7775 or 952-947-1200
for service information.
*Alconox is a registered trademark of Alconox, Inc.
Ranger Blood/Fluid Warming System
10
English

Specifications
Physical characteristics
WARMING UNIT
4.5 in. (11 cm) high x 7.5 in. (19 cm) wide x 10 in. (25 cm) long;
wt.: 7 lb. 7 oz. (3.4 kg)
CLASSIFICATION
Classified under IEC 601-1 Guidelines (and other national versions of the Guidelines) as
Class I, Type B, Ordinary Equipment. Not suitable for use in the presence of flammable
anesthetic mixtures with air or with oxygen or nitrous oxide. Classified by Underwriters
Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only, in accordance
with UL 2601-1 and in accordance with Canadian/CSA C22.2 No. 601.1.
Electrical characteristics
INPUT VOLTAGE
100-120 or 220-240 VAC
OPERATING FREQUENCY
100-120 VAC, 50/60 Hz
220-240 VAC, 50 Hz
MAXIMUM HEATING POWER
900 W
FUSE
10 Amp
Leakage current
Meets leakage current requirements in accordance with UL 2601-1 and IEC 601-1.
Environmental conditions
OPERATING TEMPERATURE RANGE
15° to 40°C (59° to 104°F)
STORAGE TEMPERATURE RANGE
-20° to 45°C (-4° to 113°F)
OPERATING HUMIDITY
10 to 85% RH, noncondensing
ATMOSPHERIC PRESSURE RANGE
50 kPa to 106 kPa
Ranger Blood/Fluid Warming System
English
11
Temperature characteristics
SET POINT TEMPERATURE
41°C
OVER-TEMPERATURE ALARM
43°C
UNDER-TEMPERATURE ALARM
33°C
OVER-TEMPERATURE CUTOFF
46°C

Definition of Symbols
ON
OFF
Fuse
Attention - Read accompanying documents
Non Explosion-Proof
Type B Applied Part
Voltage, Alternating Current (AC)
Ranger Blood/Fluid Warming System
12
English

BETRIEBSHandbuch
Deutsch 13
Ranger, und die Bair Hugger und Arizant Logos sind beim Patent & Trademark Office in den USA
und in anderen Ländern eingetragene bzw. angemeldete Markenzeichen von Arizant Healthcare Inc.

Inhalt
Technischer Kundendienst und Bestellannahme
Einführung
Indikationen
Warnhinweise und Vorsichtsmaßnahmen
Produktbeschreibung
Das RangerWärmegerät
Das Ranger Wärmeset zum Einmalgebrauch
Distale Flüssigkeitstemperaturen
Sicherheitsvorkehrungen
Gebrauchsanweisung
Vorbereitung und Aufstellen des Ranger Wärmegeräts
Entfernen des Wärmesets aus dem Ranger Wärmegerät
Transfer des Wärmesets von einem Ranger Wärmegerät in ein anderes
Störungssuche
Wartung und Aufbewahrung
Technische Angaben
Symbole
Ranger Blut-/Flüssigkeitswärmesystem
Deutsch
15
16
17
17
17
18
18
18
19
19
20
20
21
21
22
24
25
26

Technischer Kundendienst und
Bestellannahme
USA, WELTWEIT
TEL.: FAX:
1-800-733-7775 1-800-775-0002
1-952-947-1200 1-952-947-1400
INNERHALB EUROPAS
TEL.: FAX:
+49-4154-9934-0 +49-4154-9934-20
0800-100-1236 (Gebührenfrei innerhalb 0800-100-1324 (Gebührenfrei innerhalb
Deutschlands) Deutschlands)
Reparaturen und Produktrückgabe im Rahmen der Garantie
Ersatzteile zur Behebung eines Problems werden kostenlos geliefert. Zur Rückgabe eines
Gerätes an Arizant Healthcare Inc. benötigen Sie zunächst ein Rückgabe-
Autorisationsnummer (RA-Nummer), die Sie vom Kundendienst erhalten können. Bitte geben
Sie diese RA-Nummer in Ihrem Schriftverkehr an. Ein Versandkarton wird Ihnen, wenn
nötig, kostenlos zugeschickt. Wir werden Ihr Gerät innerhalb von fünf (5) Arbeitstagen nach
Eingang reparieren und zurücksenden. Bitte rufen Sie Ihren örtlichen Händler oder Vertreter
an und erkundigen Sie sich nach Leihgeräten für die Zeit, in denen Ihr Gerät in der
Reparatur ist.
Technische Hilfe per Telefon
Bitte vergessen Sie nicht, daß Sie uns bei einem Anruf die Seriennummer Ihres Geräts mit-
teilen müssen. Diese Nummer befindet sich auf einem Etikett am Boden des Wärmegeräts.
Ranger Blut-/Flüssigkeitswärmesystem
16
Deutsch

Einführung
Das Ranger Blut-/Flüssigkeitswärmesystem besteht aus einem Wärmegerät und einem
Wärmeset zur intravenösen Blut-/Flüssigkeitsverabreichung. Das Wärmeset ist nur zum
Einmalgebrauch bestimmt. Das Ranger Wärmesystem wurde für die Erwärmung und
Verabreichung von Blut, Blutprodukten und Flüssigkeiten entwickelt. Die Strömungs-
geschwindigkeiten reichen vom Offenhalten der Vene bis zu 500ml/min. Bei diesen Strömungs-
geschwindigkeiten kann das Gerät Flüssigkeitstemperaturen von 33º C bis 41º C aufrecht
erhalten. Das Gerät benötigt weniger als 2 Minuten, um eine Temperatur von 41º C zu erreichen.
Zwei unterschiedliche Blut-/Flüssigkeitswärmesets sind erhältlich: Ein Wärmeset für
Anwendungen mit normalen Strömungsgeschwindigkeiten und das andere für Anwendungen
mit hohen Strömungsgeschwindigkeiten. Diese Wärmesets sind sterile, latexfreie
Einwegprodukte zur Verwendung mit dem wiederverwendbaren Wärmegerät.
Das Ranger Wärmegerät kann an einer Infusionsstange montiert werden. Ein Griff oben am
Gerät erleichtert das Transportieren. Wird das Wärmegerät an einer Infusionsstange montiert,
paßt es bequem über ein Bair Hugger Wärmegerät Modell 505 oder über ein Wärmegerät
Modell 500/OP.
Indikationen
Das Ranger Wärmesystem ist zur Erwärmung von Blut, Blutprodukten und Flüssigkeiten
bestimmt.
Warnhinweise und Vorsichtsmaßnahmen
WARNUNG
•Gefahr durch elektrische Spannung. Gehäuse des Wärmegeräts nicht öffnen.
•Keine anderen Geräte oder Produkte anstelle eines Ranger Wärmegeräts oder Ranger
Wärmesets verwenden, da dies zu Verbrennungen, elektrischem Schlag oder Systemschaden
führen kann.
•Keine Flüssigkeit verabreichen, wenn sich Luft in den Schläuchen befindet, da dies zu einer
Luftembolie führen kann.
•Behandlung Abbrechen, wenn der Alarmton eine Übertemperatur anzeigt und die
Temperatur sich nicht auf den Sollwert zurückstellt. Sofort Flüssigkeitsstrom stoppen und
Wärmeset entsorgen. Wärmegerät von Fachpersonal überprüfen lassen oder den Arizant
Healthcare Kundendienst anrufen.
GEFAHR
Explosionsgefahr. Nicht im Bereich von entflammbaren Anästhetika verwenden.
ACHTUNG
Das Blut-/Flüssigkeitswärmegerät nicht in Flüssigkeiten eintauchen. Mit einem sauberen,
leicht angefeuchteten Tuch abwischen.
Ranger Blut-/Flüssigkeitswärmesystem
Deutsch
17

Produktbeschreibung
Das Ranger Blut-/Flüssigkeitswärmesystem besteht aus einem Wärmegerät und einem sterilen Wärmeset
zur intravenösen Blut-/Flüssigkeitsverabreichung. Das Wärmeset ist nur zum Einmalgebrauch bestimmt.
Das RANGER Wärmegerät
Das Ranger Wärmegerät ist kompakt, leicht und flüssigkeitsabweisend. An seiner Seite befindet sich eine
Klemme, mit deren Hilfe das Gerät an einer Infusionsstange montiert werden kann. Der Griff oben am
Gerät erleichtert das Transportieren. Vorn am Gerät befindet sich
•Alphanumerische Anzeige, die die Temperatur der Heizplatten bei normalem Betrieb anzeigt. Bei
Übertemperatur leuchten abwechselnd eine Temperatur über 43°C und das Wort „HI“ (für „HIGH“ =
„HOCH“) auf. Bei Untertemperatur leuchten abwechselnd eine Temperatur unter 33°C und das Wort
„LO“ (für „LOW“ = „NIEDRIG“) auf.
•Warnlicht, das bei Über- oder Untertemperatur aufleuchtet.
RANGER Einwegset
Das Blut-/Flüssigkeits-Administrationsset enthält eine Wärmekassette (Wärmeaustauscher), Schläuche
und Verbindungen die mit krankenhausüblichen Blut-/Flüssigkeitsinfusionssets kompatibel sind. Das Set
enthält ferner eine Luftfalle sowie einen Injektionsport. Das High-Flow-Set enthält zusätzlich 2
Infusionsbeutel-Spike und einen 150 Mikron-Filter, der sich in einer zentralen Tropfkammer befindet.
Einweg-Irrigationssets liefern Durchflussgeschwindigkeiten von 30 l/Std. Das Priming-Volumen
ist 142 ml. Dazu gehören 2 Spikes eine 183 lange Patientenleitung.
Gebrauchsanweisungen können Sie den jedem Einwegset beiliegenden Anweisungen entnehmen.
Ranger Blut-/Flüssigkeitswärmesystem
18
Deutsch
Schlitz für
Wärmetasche
Alphanumerische
Anzeige Halterung für
Luftfalle
Stromschalter
Klemme für
Infusionsstange
Warnlicht
Wärmekassette
Durchflussgeschwindigkeit
Priming-Volumen Länge der Patientenleitung
Standard-Flow-Set 150 ml/min 39-44 ml 76 cm
152 cm (76 cm lange Patientenleitung
plus 76 cm lange Verlängerung)
High-Flow-Set 100-150 ml/min 150 ml 152 cm (76 cm lange Patientenleitung
plus 76 cm lange Verlängerung)
Kinderheilkunde/ 6 l/Std. 20 cc 46 cm
Neugeboren-Set

Distale Flüssigkeitstemperaturen
Meßwerte sind nur nominal. Die Temperaturen wurden bei 18ºC Zimmertemperatur gemessen.
Sicherheitsvorkehrungen
In der folgenden Tabelle werden die Warnfunktionen und Alarmsignale des Ranger
Wärmegeräts beschrieben.
*Vorübergehende Zustände können zu einer Übertemperatur führen. Hierzu gehören u.a.:
•Es kam zu einer äußerst starken Veränderung in der Strömungsgeschwindigkeit
(z.B. von 500 mL/min auf Stillstand).
•Das Gerät wurde eingeschaltet und erreichte eine Temperatur von 41º C, bevor die Wärmetasche
in das Gerät geschoben wurde.
•Flüssigkeiten wurden vor der Infusion auf eine Temperatur über 42º C vorgewärmt.
Ranger Blut-/Flüssigkeitswärmesystem
Deutsch
19
FLÜSSIGKEITSGESCHWINDIGKEIT (l/h)
0,5
1,0
2,0
3,0
4,5
6,0
9,0
15,0
20,0
30,0
BEI 10ºC EINLASSFLÜSSIGKEIT
32˚C
36˚C
39˚C
39˚C
39˚C
39˚C
38˚C
-
-
-
BEI 20ºC EINLASSFLÜSSIGKEIT
32˚C
36˚C
38˚C
40˚C
40˚C
39˚C
38˚C
36˚C
35˚C
33˚C
ART DES ALARMS
Übertemperatur: 43º C
Untertemperatur: 33º C
Sicherheitsalarm: 46º C
WAS WIRD ANGEZEIGT
Warnlicht leuchtet auf
und Alarmton ist zu
hören; alphanumerische
Anzeige zeigt abwech-
selnd eine Temperatur
über 43°C und das Wort
„HI“ (für „HIGH“ =
„HOCH“) an.
Warnlicht leuchtet auf
und Alarmton ist zu
hören; alphanumerische
Anzeige zeigt abwechsel-
nd eine Temperatur
unter 33°C und das
Wort „LO“ (für „LOW“
= „NIEDRIG“) an.
•Warnlicht leuchtet auf
und Alarmton ist zu
hören; alphanumerische
Anzeige zeigt abwechsel-
nd eine Temperatur über
43°C und das Wort „HI“
(für „HIGH“ =
„HOCH“) an.
•Die alphanumerische
Anzeige ist erloschen;
Alarmton ist zu hören
(der Warnton des
Sicherheitsalarms funk-
tioniert auch dann, wenn
die alphanumerische
Anzeige erloschen ist).
BESCHREIBUNG/URSACHE
Die Temperatur der
Heizplatten ist
vorübergehend auf
43º C gestiegen. *
Die Temperatur der
Heizplatten ist auf
33º C gefallen.
Die Temperatur der
Heizplatten ist auf
46º C gestiegen. Das
Sicherheitssystem
schaltet sich bei 46º C
ein und schaltet den
Strom zu den
Heizplatten ab.
ABHILFE
Die alphanumerische
Anzeige beobachten.
Wenn die Temperatur
nicht auf 41º C fällt (was
einige Minuten dauern
kann), Gerät nicht
weiterverwenden. Den
Kundendienst von
Arizant Healthcare
anrufen.
Alarmsignale stellen sich
aus, wenn die Temperatur
über 33º C steigt. Gerät
weiterverwenden. Sollte
die Temperatur nicht
über 33º C ansteigen, den
Stecker des Geräts heraus-
ziehen und den Kunden-
dienst von Arizant
Healthcare anrufen.
GERÄT AUSSCHALTEN,
STECKER HERAUS-
ZIEHEN. Gerät nicht
verwenden. Wärmeset
entsorgen. Den
Kundendienst von
Arizant Healthcare
anrufen.
Table of contents
Languages:
Popular Medical Equipment manuals by other brands

Lifeline
Lifeline SX319 Instructions for using

Sunrise Medical
Sunrise Medical Hoyer HPL600 Installation and instruction manual

Ambu
Ambu Mini Perfit ACE Instructions for use

Porter
Porter DIGITAL MDM Operation and installation manual

Mindray
Mindray BeneHeart D1 Operator's manual

Tridien Medical
Tridien Medical ST-1200 operating instructions

Respironics
Respironics BiPAPautoSV Advanced user manual

Keeler
Keeler Cryomatic MKII Instructions for use

laerdal
laerdal V-Vac manual

Merit Medical
Merit Medical ONE Snare Endovascular Microsnare System Instructions for use

Konica Minolta
Konica Minolta Pulsox-2 Multi Lingual Short Form Instruction Manual

Medline
Medline EntraFlo Operating instruction guide