Armstrong Steam QM-3 Use and care manual

Armstrong International
Parc Industriel des Hauts-Sarts, 2ème avenue 4, 4040 HERSTAL - BELGIUM
Tél: +32(0)4 240 90 90 Fax: +32 (0)4 240 40 33 armstronginternational.com
Keep this manual with equipment
for future reference.
STEAM QM-3
Steam Quality Monitor
CALIBRATION MANUAL
Certified by
IOM-246 V2-0

Armstrong International
Parc Industriel des Hauts-Sarts, 2ème avenue 4, 4040 HERSTAL - BELGIUM
Tél: +32(0)4 240 90 90 Fax: +32 (0)4 240 40 33 armstronginternational.com
Armstrong
Contents
Revision History ........................................................... 1
Safety.........................................................................2
Abbreviations and Acronyms............................................. 3
General Information ....................................................... 4
Recalibration Procedure...................................................6
Calibrating the Temperature Sensors .............................. 6
Calibrating the Pressure Transmitter ............................... 8
Calibrating the Heater ................................................ 9
Calibrating the Condensate Flow Meter.......................... 10
Calibrating the NCG Flow Meter................................... 12
Return to Service ........................................................ 14

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 1 of 14
Revision History
Version Release Date Description of Changes
1.0 20/02/2014 Initial
1.1 21/03/2014 Temperature transmitter T (T1, T2, T3) - Design Modification
2.0 17/03/2016 Tee and Adapter - Design Modification: pages 5, 6, 8, 9.
Page 4, General Information Section. Update of the list of material needed
The following have been removed: Decade box (calibrated), 5 volt generator or adjustable power
supply (0-10VDC), M12 digital for decade box
The following have been added: Oil bath, Thermometer (calibrated), Pressure generator,
Manometer (calibrated).
Page 6, Recalibration Procedure Section
Point 3 has been added (Remove the Temperature Sensor T1 after unplugging it and plug it back
in before install it in oil bath).
- "Calibrating the Temperature Sensors (Simulated Measurement Method)" has been updated as
"Calibrating the Temperature Transmitter".
- Chart with the data of the 3 temperature sensors is deleted.
Pages 6 - 8: Calibrating the temperature transmitter and the pressure transmitter procedures have
been updated.
Page 12: Text modifications, Point 1(Disconnect the expansion coil from the bottom of the NCG
burette. To avoid damaging the expansion coil, loosen bottom nut on expansion coil and remove
top nut). Point 13 (o force open the EV1 valve to allow discharge move the arrow to the EV1 line
then move the arrow to left of 1 and press OK).

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 2 of 14
Safety
Icon Legend
Burn hazard! Uninsulated components upstream of cabinet may be hot.
• Do not touch when unit is working.
• Allow to cool before moving or servicing unit.
Live steam will cause burns; condensate water may cause them. Skin exposure
to 140 °F (60 °C) water for only five seconds may cause a second degree burn.
This device must be installed in accordance with appropriate local,
national, and international standards, codes, and practices.
Read this manual. It contains important information.
Service must be performed by a qualified person.
Installation should always be accompanied by competent
technical assistance.
Improper installation, start-up, operation, maintenance, or service
may void warranty.
You are encouraged to contact Armstrong International or its local
sales representative for additional information.
Shock hazard! High voltages present inside equipment.
• Electrical installation must be performed by qualified personnel.
• Disconnect power before performing any electrical service.
Equipment must be disposed of according to applicable
environmental requirements.
Indicates Power On
Indicates Power Off
Indicates important information concerning potential
for personal injury or damage to equipment
Indicates electrical hazard
Indicates hot surface
Keep unit away from heat-sensitive equipment and installations.

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 3 of 14
Abbreviations and Acronyms
Term Meaning Explanation
∆P Differential Pressure ∆P1: Difference between water column in NCG vessel and atmosphere.
∆P2: Difference between water column in condensate vessel and atmosphere.
Al Alarm Indicates an out-of-limit situation, but has no impact on operation.
Al1: Dryness below user-defined set point longer than two seconds.
Al2: Four consecutive calculations of NCGs are over the user-defined limit. This calculated
value is displayed on the mais screen and updated every 30 seconds.
Al3: T1 above 257 °F (125 °C) longer than two seconds.
C Celsius
cm Centimeter
Df Default Indicates failure. Turns off power to heating element and opens EV0to drain.
Df4: T3 above 185 °F (85 °C) longer than two seconds.
Df5: No condensate from condenser in last ten minutes.
Df6: T2 above 356 °F (180 °C) longer than two seconds.
DIN Deutsches Institut für Normung eV
dP Differential Pressure
EC European Community
EEC European Electrotechnical Commission
EN European Norm
EV Electronic Valve
F Fahrenheit
gal Gallon
h Hour
imp Imperial [measure]
in Inch
kg Kilogram
L Liter
lb(s) Pound(s)
max Maximum
min Minimum
mm Millimeter
NCG Non-Condensable Gases NCGmax is the limit of the NCG rate. Alarm 2 indicates the limit has been exceeded. Range is
0–15%. Default is 3.5%.
P Pressure P is steam pressure upstream of calibrated orifice. Modbus sends data as bar even with
imperial measure selected.
ppm Parts per Million
psi(g) Pounds per Square Inch (gauge)
Q Steam Flow
QM Quality Monitoring
R Resistance Shown as watts.
R/O Reverse Osmosis
sec(s) Second(s)
SI International System of Units
STSuperheat
T Temperature T1: Temperature after pressure reduction to atmosphere.
T2: Temperature after heating resistance.
T3: Temperature after condenser.
X Dryness Fraction (sometimes called steam
quality or moisture content)
Xmin is the lower dryness limit. Alarm 1 indicates the limit has been exceeded. Range is
0.85–0.95. Default is 0.95.

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 4 of 14
General Information
This document will walk a technician through the steps necessary to calibrate/recalibrate the Steam QM-3. It is estimated
this process will take between two to four hours. Once recalibration has begun, all components must be recalibrated before
returning the unit to service.
Recalibration is required annually or whenever any component is changed.
What to expect:
Temperature sensors:
There are three temperature sensors which can be calibrated either by simulated measurement method or true
measurement method.
•Simulatedmeasurementmethod—simulatestheresistancethatshouldbecreatedbythetemperaturesensor.
•Truemeasurementmethod—placethetemperaturesensorsintheoilbathandrecreatetheactualtemperaturesfor
calibration. This method is not covered in this document.
Pressure transmitter:
The pressure sensor is calibrated by simulating the voltage created at a specific pressure.
Heating Element:
It is important to calibrate the natural resistance of the heating element and cable with the Steam QM-3 computer. This is done
by measuring the actual resistance.
Condensate Flow Meter:
Measure the average amount of condensate discharged per cycle of the EV2 valve.
Gas Flow Meter (NCG flow meter):
Measure the average amount of non-condensable gasses (NCGs) discharged per cycle of the EV1 valve.
Materials needed:
•Balance(calibrated)
•Ohmmeter(calibrated)
•10mmwrench
•11mmwrench
•14mmwrench
•Adjustablewrench
•NCGcalibrationkit
Open vessels for collecting liquids
500ml vessel
50ml vessel
Draw vessel cap 1L
NCG injector syringe
Multimeter
•Oilbath
•Thermometer(calibrated)
•Pressuregenerator
•Manometer(calibrated)

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 5 of 14
There are a total of nine instruments that need to be calibrated on the Steam QM-3. The instruments are:
•3temperaturesensors
•1pressuretransmitter
•1heatingelement
•1condensateowmeter
•1gasowmeter(NCGmeter)
•2differentialpressuresensors
Heating Element
T1
T2
T3
Differential Pressure Sensors
∆P2
∆P1
Temperature Sensors
Gas Flow Meter
Condensate
Flow Meter
Pressure transmitter

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 6 of 14
Calibrating the Temperature Sensors
Calibration #1: Calibrating T1, T2, T3 Temperature Sensors1
1. Turn off the Steam QM-3 and be sure it is cold. Check that the voltage switch is in the correct position for the voltage
being used.
2. Unplug the Temperature Sensor T1 from the QM-3.
3. Remove the Temperature Sensor T1 after unplugging it and plug it back in before install it in oil bath.
4. Install the oil bath on stable area close to the device.
5. Place the probe and the calibrated thermometer in the oil bath (as close as possible)
6. Set the oil bath at 80 °C
7. Wait that the oil bath is stabilized. And adjust its temperature to have 80 °C on the calibrated thermometer.
8. Turn on the QM-3.
9. On the Steam QM-3 display pad, press up and down arrows simultaneously to access the Debug sensor menu.
10. Write the value of T1 on the calibration certificate. If T1 is +- 0,7°C regarding the set point (80°C), the sensor is
considered as valid (go to the point 14). If not, you must adjust the sensor.
11. On the Steam QM-3 display pad, press left and right arrows simultaneously to access the code menu.
12. Enter 234 using the up and down arrows to change numbers and the left and right arrows to change the position to
access temperature calibration menu, and press OK.
13. Move the arrow to left of 80 °C on T1, press OK and wait for check mark.
14. Write 80 °C on the adjusted column of the calibration certificate.
15. Restart the procedure from point 5 with 140 °C instead of 80 °C.
16. Control the calibration with random temperature between 80 °C and 140 °C (restart the procedure from point 5) and
write it on the calibration certificate.
17. If the three calibrations are correct, check the valid mark on the calibration.
1 To reduce the time, part of calibration can be done simultaneously.
Recalibration Procedure
-> 0 0 0
x.xx
Code Menu
Access menu by pressing and
at same time. Navigate between digits by
pressing or .
Change value for digit by
pressing or .
Press «OK».
Turn off the steam to the Steam QM-3 and allow to cool
Version
220
1
On/Off Switch
Power cord
2
Temperature
transmitters
T1 and T2
Calibration T Sensor
T1 80 °C 140 °C
T2 80 °C 140 °C
T3 40 °C 80 °C
Temperature Calibration Menu 234

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 7 of 14
Calibration #2: Calibrating T2 Temperature Sensor1
Repeat previous procedure in step #1 for T2 Temperature sensor.
Calibration #3: Calibrating T3 Temperature Sensor2
Repeat previous procedure in step #1 for T3 Temperature sensor.
NOTE: Any corrections to the calibration certificate must be done following good documentation practices. The error is crossed
through by a single line. The correction is written alongside with an explanation if necessary. The correction is signed and dated.
1 To reduce the time, part of calibration can be done simultaneously.
2To reduce the time, part of calibration can be done simultaneously.
Temperature Sensor T3

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 8 of 14
Calibrating the Pressure Transmitter
Calibration #4: Calibrating the Pressure Transmitter
1. Remove the pressure sensor at the sanitary fitting.
2. On the Steam QM-3 display pad, press up and down arrows simultaneously to access the Debug sensor menu.
3. Write the value of P1 on the calibration certificate. If P1 is +- 0,1 bar regarding the set point (0 barg), the calibration point is
considered as valid. If not, you must adjust the sensor (see point 8 to 10).
4. Connect the pressure sensor and the calibrated manometer on the pressure generator.
5. Generate a pressure of 3 barg (control it with the calibrated manometer).
6. On the Steam QM-3 display pad, press up and down arrows simultaneously to access the Debug sensor menu.
7. Write the value of P1 on the calibration certificate. If P1 is +- 0,1 bar regarding the set point (0 barg), the calibration point is
considered as valid. If not, you must adjust the sensor (see point 8 to 10).
8. If you must adjust the value : On the Steam QM-3 display pad, press left and right arrows simultaneously to access the code
menu.
9. Enter 152 using the up and down arrows to change numbers and the left and right arrows to change the position to access
pressure calibration menu, and press OK.
10. Move the arrow to left of 3 barg, press OK and wait the check mark.
11. Confirm calibration via the debug sensor menu around 1 barg and 2 barg and write the value on the calibration certificate.
Calibration P Sensor
P1 0 barg 3 barg
EV1 0 1
115 230
Pressure calibration menu 152
1
DIN connector

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 9 of 14
Calibrating the Heater
Calibration #5: Calibrating the Heater
1. Turn off Steam QM-3 and unplug from power source.
2. Release latch on Harting connector and unplug from the Steam QM-3 cabinet.
3. With a calibrated ohmmeter measure the resistance between pin 8 and 16 on the Harting plug. The reading should
be between 100Ω and 200Ω (normal is about 150Ω). If the reading is lower than 100Ω or higher than 200Ω replace
the heater.
4. If the reading is in accepted range, record it. The reading will need to be entered after the Steam QM-3 is powered up.
5. Plug the Harting plug into the Steam QM-3 cabinet, plug in the power cord, and turn on the Steam QM-3.
6. On the Steam QM-3 display pad, press left and right arrows simultaneously to access the code menu.
7. To access the resistance calibration menu, enter 069 using the up and down arrows to change numbers,
and the left and right arrows to change the position. 65 ß
8. Move the arrow to the Rheater line ( using up and down arrows).
9. Move the arrow to left of value and enter measured value rounded to nearest whole ohm value
(example 153.3 is entered as 153).
10. Confirm the re-calibration of the resistance of the heater by signing the relevant box on the re-calibration certificate.
NOTE: Any corrections to the calibration certificate must be done following good documentation practices. The error is crossed
through by a single line. The correction is written alongside with an explanation if necessary. The correction is signed and dated.
Harting plug being removed
Harting plug pin 8 and 16
Resistance calibration menu 069
Calibration Rheater
P 10W 30W
Rheater 146
T3`m65 ß01
Heater

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 10 of 14
'3
Calibrating the Condensate Flow Meter
Calibration #6: Calibrating the Condensate Flow Meter
1. Locate the 6mm tube coming from DP2 going inside the condensate vessel on top of EV2 and pull it out.
2. Confirm the 6mm tube is cut on an angle.
3. Replace tube back inside condensate vessel.
4. Ensure Steam QM-3 cabinet is level.
5. Locate the 6 mm tube coming out of the bottom of EV2.
6. Follow the tube to the end and remove it where it connects to the tee on the condensate return line.
Place the end of the tube in the waste water vessel.
7. Disconnect the 6mm tube from the inlet of EV2 (coming out of the NCG burette).
8. Placethe1literdrawvessellledwithtapwaterontopoftheSteamQM-3soastohavegoodgravityow.
9. Connect the 6mm tube from the 1 liter draw vessel filled with tap water to the inlet to EV2.
10. Position and level a calibrated balance (scale).
11. Record the dry weight of a 500ml container.
12. Turn on the power to the Steam QM-3.
13. Open the valve from the draw vessel (slowly) to establish a cycle rate of the EV2 valve of approximately 30 seconds.
One cycle is from opening to closure of valve, you should be able to hear this.. If the environment prohibits
hearing go to the control panel and press up and down arrows to access sensor menu and watch ∆P2 value. it will
climb from 0 to about 400 then when it opens it will drop back to 0.
14. Aftertheproperowrateisestablished,letitcycleonemoretimeandremovethetubefromthewastecontainer
(be careful not to shake out any remaining water from the tube).
15. Place the tube in the 500ml container (from step 11).
16. Allow the EV2 valve to cycle exactly 20 times (keep track of the number of cycles).
17. After exactly 20 cycles remove the tube (be careful not to shake out any remaining water from the tube.
18. Record the mass of the water and container together to nearest tenth of a gram, and calculate using the following
formula: (water + container) - the dry weight of container(step 11)
20 = average mass of water per cycle
19. On the Steam QM-3 display pad press left and right arrows simultaneously to access the code menu.
6mm tube from DP2
Condensate vessel
EV2
1 liter draw vessel
valve
6mm tube
500ml container
calibrated balance

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 11 of 14
20. Toaccesstheowsensorcalibrationmenuentercode123usingtheupanddownarrowstochangenumbers,
and the left and right arrows to change the position.
21. Move arrow to the right of Q and enter mass per cycle (using up and down arrows) then move the arrow
back to left of Q.
22. Conrmthere-calibrationofthecondensateowburettebysigningtherelevantboxontherecalibrationcerticate.
NOTE: Any corrections to the calibration certificate must be done following good documentation practices. The error is
crossed through by a single line. The correction is written alongside with an explanation if necessary. The correction is
signed and dated.
23. Place the 1 liter draw vessel below the EV2 valve and disconnect the 6mm tube from the inlet to the EV2 valve.
24. Reconnect the 6mm tube coming from the NCG Burette to the inlet on the EV2 valve and leave the 6mm tube
coming out of the EV2 valve in the waste container as we will use this for the calibration of the NCG Flow Meter.
Calibration Q Sensor
-> Q 20.3
NCG 1.3
Flow sensor calibration menu 123

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 12 of 14
Calibration #7: Calibrating the NCG Flow Meter
1. Disconnect the expansion coil from the bottom of the NCG burette. To avoid damaging the expansion coil Loosen
bottomnutonexpansioncoilandremovetopnut.—usetwowrenches10mmand11mmto
ensure nodamagetotheNCGburetteoccurs—andswingoutoftheway.
2. Placethe1literdrawvessellledwithtapwaterontopoftheSteamQM-3soastohavegoodgravityow.
3. Connect the 6mm tube coming from the 1 liter draw vessel to the bottom of the NCG burette (where the expansion.
coilwasconnected—usetwowrenches10mmand11mmtoensurenodamagetotheNCGburetteoccurs).
4. Disconnect the 6mm tube coming from P1 where it connects to the top of the NCG burette.
5. Connect the NCG injector assembly tee to the top of NCG burette.
6. Connect the 6mm tube that was disconnected from the top of the NCG burette to the top of injector tee.
7. To avoid damage to P1 by creating a vacuum, always remove the NCG injector syringe before resetting
to 2ml.
8. With the NCG injector syringe disconnected, reset the syringe to 2ml.
9. Connect the NCG injector syringe to the middle of the NCG injector assembly tee with a 6mm tube.
10. Force open the EV1 valve using the pressure calibration menu 152.
11. On the Steam QM-3 display pad, press left and right arrows simultaneously to access the code menu.
12. To access the pressure calibration menu, enter code 152 using the up and down arrows to change numbers, and the
left and right arrows to change the position.
Calibration P Sensor
P1 0 barg 3 barg
EV1 0 1
115 230
Pressure calibration menu 152
13. To force open the EV1 valve to allow discharge move the arrow to the EV1 line then move the arrow to left of 1 and
press OK.
14. Open the valve on the 1 liter draw vessel to fill NCG burette and allow water to run into the waste container.
15. Close the valve on the 1 litter draw vessel.
16. Close the EV1 valve using menu 152 on the control panel, move the arrow to left of 0 and press OK.
17. To avoid damage to P1 by creating a vacuum, always remove the NCG injector syringe before resetting to 2ml.
Calibrating the NCG Flow Meter
'3
Expansion coil swung out of the way
6 mm tube
1
EV1
syringe
valve
valve
1 liter draw vessel

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 13 of 14
18. Using the NCG injector syringe set at 2ml, very slowly (at a rate 1 tenth of a ml every second) inject air until EV1 valve opens.
Tip: use a stethoscope to hear the EV1 valve cycle.
19. Record amount of air injected to the nearest tenth ml. If your readings are below 0.5 or above 1.5, the QM3 cannot
be calibrated. Contact an Armstrong representative for assistance.
20. Move the arrow to left of 1 and press OK.
21. Repeat this process for exactly 10 cycles.
22. Add the 10 cycles together and divide by 10 this will give you the average ml injected.
23. Recordtheaveragenumberofmlinjectedthatwillbeenteredintheowsensorcalibrationmenu.
24. On the Steam QM-3 display pad, press left and right arrows simultaneously to access the code menu.
25. Toaccesstheowsensorcalibrationmenuentercode123usingtheupanddownarrowstochangenumbers,and
the left and right arrows to change the position.
26. scroll to the NCG line (using up and down arrows) and move the arrow to right of value and enter your average
number of ml injected.
27. Move the arrow back to left of NCG.
28. Write the average volume injected on the calibration certificate and confirm by signing the relevant box.
NOTE: Any corrections to the calibration certificate must be done following good documentation practices. The error is crossed
through by a single line. The correction is written alongside with an explanation if necessary. The correction is signed and dated.
Calibration Q Sensor
-> Q 20.3
NCG 1.3
Flow sensor calibration menu 123

Armstrong International
IOM-246-V2.0
Steam QM-3 Calibration
Page 14 of 14
The calibration procedure for the Steam QM-3 is now complete. Please follow the steps below to insure the Steam QM-3
back in operating condition.
1. Place the 1 liter draw vessel below the NCG burette and open the valve on the 1 litter draw vessel to drain the water
from the NCG burette and injection assembly.
2. When all water has drained from system, close the 1 liter draw vessel valve.
3. Removethe6mmtubecomingfromthe1litterdrawvesselconnectingtothebottomoftheNCGburette—use
two wrenches 10mm and 11mm to ensure no damage to the NCG burette occurs. Reconnect the expansion coil
tothebottomoftheNCGburette—usetwowrenches10mmand11mmtoensureno damage to the NCG
burette occurs.
4. Remove the NCG injector assembly from the top of the NCG burette.
5. Reconnect the 6mm tube coming from ∆P1 to the top of the NCG burette.
6. Reconnect the 6mm tube coming from the EV2 valve out to the tee on the condensate return line.
7. Recheck all connections are properly tightened.
8. Slowly turn on the steam to the Steam QM-3 and check that there are no leaks.
9. The Steam QM-3 is now ready to resume operation.
Return to Service
Other manuals for Steam QM-3
1
Table of contents
Other Armstrong Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Gill Instruments
Gill Instruments 1360-PK-022 user manual

Blue-White industries
Blue-White industries Sonic-Pro S6A Series operating manual

Hanna Instruments
Hanna Instruments HI702 instruction manual

Mahr
Mahr Millimar S 1840 operating instructions

Vaisala
Vaisala WAA151 user manual

Servomex
Servomex 2500 Series installation manual

BiOptic
BiOptic Qsep100 Operation Quick Start Guide

Clarke
Clarke CM265 User instructions

Milwaukee
Milwaukee MW101 user manual

Regan Scientific Instruments
Regan Scientific Instruments BOT-3000E operating manual

Crystal Instruments
Crystal Instruments Coco-80 quick reference

ACR Electronics
ACR Electronics Vecta 3 Product support manual