Centrifuge Preventative Maintenance Schedule
Taken from the operator manuals supplied with each centrifuge
Pull the mains plug before cleaning
Using detergents follow the instructions of the Hettich Operating Instructions.
Centrifuge Bowl Seal
Cleaning
regularly
Rubber Maintenance
after cleaning
Maintains long life for seal.
Always inspect seal for damage.
Rotor
Cleaning
at least once a week
Lube
after cleaning
Secured
weekly
Maintains good lab practices for cleanliness. Helps
maintain good results. Check rotor xing screw for
safety. Always inspect rotor for damage.
Buckets
Cleaning
at least once a week
Lube
after cleaning
Maintains good lab practices for cleanliness. Helps
maintain good results. Always inspect buckets for
damage.
Motor Shaft
Cleaning
at least once a month
Lube
at least once a month
Helps to ensure that the rotor can be removed
for maintenance when needed. Does not affect
performance. Always inspect shaft for damage.
Surfaces and Chamber
Cleaning
weekly Maintains good lab practices for cleanliness. Always inspect for damage.
Most Common Error Codes and Clearing Procedures
ERROR CODE – 1 (tachometer error)
Clearing Procedure:
• Wait for rotor stand still (120 seconds after rotor stops)
• Turn the power off and wait 10 seconds
• Turn the power on
• Check to make sure rotor is installed properly
before spinning
ERROR CODE – 2 (mains interrupt)
Clearing Procedure:
• Wait for rotor stand still
• Open the lid
• Press the start button
• Close lid and start again
ERROR CODE – 3 (balance error)
Clearing Procedure:
• Wait for rotor stand still
• Check to make sure the centrifuge is sitting on a level
and sturdy surface
• Make sure the rotor is installed properly and the bucket
weight is distributed evenly before spinning
ERROR CODE – d (lid lock error)
Clearing Procedure:
• Turn the power off and wait 10 seconds
• Turn the power on
• Open and close lid
ERROR CODE – F (tachometer error)
Clearing Procedure:
• Wait for rotor stand still
• Turn the power off and wait 10 seconds
• Turn the power on
• Check to make sure rotor is installed properly
before spinning
Extract from the
Operating Instructions
!
1. Clean with soap & water or a mild cleaning
agent after every use (or at least once a week)
2. If parts are contaminated by pathogenic or biologic
material we recommend:
– For disinfection: Helipur®H plus N from
B. Braun Melsungen
3. Dry immediately after cleaning
4. Lightly grease aluminium parts after drying using
acid-free grease, e.g. vaseline
5. With aerosol-sealed rotors and bio safety systems:
– Clean & check sealing rings regularly (weekly)
– Replace sealing ring immediately upon indication of
crack formation, embrittlement or abrasive wear
– Lightly grease seal ring with acid-free grease,
e.g. vaseline
6. Rotor and Motor Shaft (clean once a month):
– Disassemble the rotor and clean
– Lightly grease the motor shaft
7. Checked all parts on a monthly basis for
corrosion damage
RECOMMENDED CLEANING PROCEDURE
Hettich Centrifuge – Rotors and Accessories
© Arthrex GmbH, 2016. All rights reserved. LI2-00016-EN_A
Rotors and attachments may no longer be used upon indication of wear and tear or corrosion!

Popular Laboratory Equipment manuals by other brands

Parker Research Corp DA750 operating instructions

Parker Research Corp

Parker Research Corp DA750 operating instructions

Conrad Ai202 operating instructions

Conrad

Conrad Ai202 operating instructions

Intra-Lock IntraSpin Series Use

Intra-Lock

Intra-Lock IntraSpin Series Use

Koehler K45 Series Operation and instruction manual

Koehler

Koehler K45 Series Operation and instruction manual

Nakanishi PRESTO II Operation manual

Nakanishi

Nakanishi PRESTO II Operation manual

Biosan Automatic Luminescence Analyser ALA1/4 operating manual

Biosan

Biosan Automatic Luminescence Analyser ALA1/4 operating manual

Datacolor Ahiba Ahiba MULTIPRECISE TC user guide

Datacolor

Datacolor Ahiba Ahiba MULTIPRECISE TC user guide

PIP 120260 operating manual

PIP

PIP 120260 operating manual

Ibx BASIC user manual

Ibx

Ibx BASIC user manual

Ivoclar Vivadent Multilink Instructions for use

Ivoclar Vivadent

Ivoclar Vivadent Multilink Instructions for use

Aqua Ultraviolet SL 100 Watt instruction manual

Aqua Ultraviolet

Aqua Ultraviolet SL 100 Watt instruction manual

POL-EKO SMART ST 1 instruction manual

POL-EKO

POL-EKO SMART ST 1 instruction manual

Vitlab genius2 operating manual

Vitlab

Vitlab genius2 operating manual

CORNING Axygen Axypet instruction manual

CORNING

CORNING Axygen Axypet instruction manual

IKA KS 130 basic operating instructions

IKA

IKA KS 130 basic operating instructions

PASCO PK-9023 Instruction manual and experiment guide

PASCO

PASCO PK-9023 Instruction manual and experiment guide

cytiva 420100 Instructions for use

cytiva

cytiva 420100 Instructions for use

YOKOGAWA FF20 instruction manual

YOKOGAWA

YOKOGAWA FF20 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.