Ascon Tecnologic - TRH514 - USER MANUAL - PAG. 1
TRH514
RELATIVE HUMIDITY
AND TEMPERATURE PROBE
WITH 4 ÷ 20 mA OUTPUT
User manual
22/10 - Code: ISTR_M_TRH514_E_01_--
ASCON TECNOLOGIC S.r.l.
Viale Indipendenza 56, 27029 - VIGEVANO (PV) ITALY
TEL.: +39 0381 69871 - FAX: +39 0381 698730
sito: http:\\www.ascontecnologic.com
e-mail: [email protected]
PREFACE
D This manual contains the information necessary for the
product to be installed correctly and also instructions for
its maintenance and use; we therefore recommend that
the utmost attention is paid to the following instructions
and to save it.
This document is the exclusive property of Ascon Tecnologic
S.r.l. which forbids any reproduction and divulgation, even par-
tially, of the document, unless expressly authorized.
Ascon Tecnologic S.r.l. reserves the right to make any formal or
functional changes at any moment and without any notice.
Ascon Tecnologic S.r.l. and its legal representatives do not
assume any responsibility for any damage to people, things or
animals deriving from violation, wrong or improper use or in
any case not in compliance with the instrument features.
D Whenever a failure or a malfunction of the device may
cause dangerous situations for persons, thing or animals,
please remember that the plant has to be equipped with
additional devices which will guarantee safety.
1. INSTRUMENT DESCRIPTION
1.1 General description
The probes of the 500 series are devices characterized by a
very low energy consumption.
2. USAGE WARNINGS
2.1 Admitted usage
D The instrument has been projected and manufactured
as a measuring and control device to be used accord-
ing to EN60730-1 at altitudes operation below 2000 m.
Using the instrument for applications not expressly permitted
by the above mentioned rule must adopt all the necessary
protective measures.
The instrument must not be used in dangerous environ-
ments (flammable or explosive) without adequate protections.
D The installer must ensure that the EMC rules are
respected, also after the instrument installation, if
necessary using proper filters.
3. INSTALLATION WARNINGS
3.1 Mounting requirements
Select a mounting location having the following characteristics:
• It should be easily accessible;
• It must not be subjected to vibrations or impacts;
• Must be free from corrosive gases;
• Must be free from water or other fluids (condensation).
3.2 Installation position
Wall fixing is foreseen, horizontally with the sensor and cable
facing downwards.
Designing the probe 500, an attempt was made to limit as
much as possible the possibility of self-heating of the sensor,
minimizing energy consumption and separating the sensor
from the dissipative element.
Any error in humidity measurement may be due to the settling
time of the probe, in this case it will be necessary to wait for
the time necessary to achieve a stable measurement.
Steam jets, water sprays, drafts, direct sun exposure, conden-
sation on the sensor or mounting on a wall made cold by the
outside temperature can also cause important measurement
errors; due to these conditions, during the installation, the
probe must be protected from any possible external influence.
Ascon Tecnologic - TRH514 - USER MANUAL - PAG. 2
3.3 Electrical Connections
General notes on electrical connections
• The probe cables must be kept separated from high
power voltage wiring;
• If shielded cables are used, the protection shield must be
connected to ground at one side only.
3.3.1 Electrical wiring
VDC
GND
Out-RH
Out-T
Control
i
Power supply voltage: 9 ÷ 30 VDC;
Current consumption: 20 mA + 20 mA max.;
Reverse polarity protection: Diode.
Sensor reading
The output current is normally passed through the input
resistance of the measuring instrument and transformed into
a voltage that can be read by the instrument itself.
4. PROBLEMS AND MAINTENANCE
4.1 Cleaning
We recommend to clean the instrument with a slightly wet
cloth using water and not abrasive cleaners or solvents only.
Should it be necessary to remove the sensor, avoid to me-
chanical stress it and in particular avoid touching the humid-
ity sensor.
4.2 Disposal
The appliance (or the product) must be
disposed of separately in compliance
with the local standards in force on
waste disposal.
5. WARRANTY AND REPAIRS
The instrument is under warranty against manufacturing
flaws or faulty material, that are found within 18 months from
delivery date. The warranty is limited to repairs or to the
replacement of the instrument.
The eventual opening of the housing, the violation of the instru-
ment or the improper use and installation of the product will
bring about the immediate withdrawal of the warranty effects.
In the event of a faulty instrument, either within the period of
warranty, or further to its expiry, please contact our sales de-
partment to obtain authorisation for sending the instrument
to our company.
The faulty product must be shipped to Ascon Tecnologic with
a detailed description of the faults found, without any fees or
charge for Ascon Tecnologic, except in the event of alterna-
tive agreements.
6. TEHNICAL DATA
6.1 Sensor characteristics
Sensor type: CMOS-C.
6.1.1 Relative humidity
Parameter Condition Typical Units
Accuracy/Tolerance Typical ±1.8 %RH
Max. See figure %RH
Repeatability
High 0.08 %RH
Medium 0.15 %RH
Low 0.25 %RH
Resolution - 0.01 %RH
Hysteresis - ±1 %RH
Operative range Extended 0 ÷ 100 %RH
Response time t63% 6 s
Long-term drift Typical <0.25 %RH/year
Humidity measurement tolerance
Relative Humidity (%RH)
Temperature (°C)
|typical %RH|
01020304050607080
20
10
0
30
30
40
60
80
70
90
100
1.8
3
2
2
3
Ascon Tecnologic - TRH514 - USER MANUAL - PAG. 3
6.1.2 Temperature
Parameter Condition Typical Units
Temperaure accuracy Typical ±0.2 °C
Max. See figure °C
Repeatability
High 0.04 °C
Medium 0.07 °C
Low 0.1 °C
Resolution - 0.01 °C
Specified range - -40 ÷ +125 °C
Response time t63% 2 s
Long-term drift Normal <0.03 °C/year
Accuracy of temperature measurement
Temperature (°C)
-40 -20 020406080 100 120
±0.0
±0.2
±0.4
±0.6
±0.8
±1.0
T (°C)
Maximum tolerance
Typical tolerance
6.2 Functional features
Sensor protection degree: IP67 ((protected against dust
and against the effects of temporary immersions);
Air filter: PTFE air filter with two sealing o-rings and thrust
spring;
Installation method: Wall mount;
Electrical connections: Removable connector with screw
terminals;
Dimensions (mm): L = 105 mm H= 70 mm e D = 40 mm
(the box only, fixing parts excluded);
Operating temperature: -30 ÷ +70°C;
Ambient humidity: 0 ÷ 100% RH;
Sensor type: CMOS;
Humidity measuring range: 0 ÷ 100% RH;
Response time at constant conditions (63%) at 23°C: 30 s;
Recovery time from saturation: 90 s
Storage temperature: -30 ÷ +80°C
Power supply: 9 ÷ 30 VDC;
Power consumption: 40 mA max.;
RH calibration: 0% @ 4 mA,
100% @ 200 mA;
Temperature calibration:
-30°C @ 4 mA,
+70°C @ 20 mA;
Typical load: 100 ÷ 200 Ω.
Ascon Tecnologic - TRH514 - USER MANUAL - PAG. 4

Other Ascon tecnologic Measuring Instrument manuals

Ascon tecnologic TRH22 User manual

Ascon tecnologic

Ascon tecnologic TRH22 User manual

Ascon tecnologic TRH42 User manual

Ascon tecnologic

Ascon tecnologic TRH42 User manual

Ascon tecnologic TRH324 User manual

Ascon tecnologic

Ascon tecnologic TRH324 User manual

Ascon tecnologic TRH200 User manual

Ascon tecnologic

Ascon tecnologic TRH200 User manual

Ascon tecnologic K V Series User manual

Ascon tecnologic

Ascon tecnologic K V Series User manual

Ascon tecnologic TLI40 User manual

Ascon tecnologic

Ascon tecnologic TLI40 User manual

Ascon tecnologic TRH301 User manual

Ascon tecnologic

Ascon tecnologic TRH301 User manual

Popular Measuring Instrument manuals by other brands

Minox Windwatch brochure

Minox

Minox Windwatch brochure

Pfeiffer Vacuum GSD 350 OMNISTAR operating instructions

Pfeiffer Vacuum

Pfeiffer Vacuum GSD 350 OMNISTAR operating instructions

Dostmann ScanTemp 485 Operating	 instruction

Dostmann

Dostmann ScanTemp 485 Operating instruction

Laser Technology TruPulse 200X user manual

Laser Technology

Laser Technology TruPulse 200X user manual

JUMO 902020/10 operating manual

JUMO

JUMO 902020/10 operating manual

PCB Piezotronics IMI SENSORS 604B61 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS 604B61 Installation and operating manual

Novanta v40 user manual

Novanta

Novanta v40 user manual

Futech MULTICROSS 3 Super Vision user manual

Futech

Futech MULTICROSS 3 Super Vision user manual

OHAUS MB90 quick start guide

OHAUS

OHAUS MB90 quick start guide

LOVATO ELECTRIC DME D122MID installation manual

LOVATO ELECTRIC

LOVATO ELECTRIC DME D122MID installation manual

Blue-White FLEXFLO A1 Series operating manual

Blue-White

Blue-White FLEXFLO A1 Series operating manual

LIMIT MWA 10 operating manual

LIMIT

LIMIT MWA 10 operating manual

RMG TERZ 94 operating instructions

RMG

RMG TERZ 94 operating instructions

PCB Piezotronics IMI SENSORS EX628F01 Installation and operating manual

PCB Piezotronics

PCB Piezotronics IMI SENSORS EX628F01 Installation and operating manual

Haldex L31298W Installation and service manual

Haldex

Haldex L31298W Installation and service manual

Eltako electronics DSZ12DZ-3x65A operating instructions

Eltako electronics

Eltako electronics DSZ12DZ-3x65A operating instructions

SD Biosensor F200 user manual

SD Biosensor

SD Biosensor F200 user manual

INNO Instrument VIEW600 user manual

INNO Instrument

INNO Instrument VIEW600 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.