manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. ATAGO
  6. •
  7. Measuring Instrument
  8. •
  9. ATAGO 4555 User manual

ATAGO 4555 User manual

Error Messages
The following messages alert the user when an operation has failed.
The battery is low.
The sensor temperature is below the temperature range.
The ZERO button was pressed with nothing or something
other than water on the sensor.
The sensor temperature is above the temperature range.
The sample measured outside the measurement range.
The instrument is faulty. (Replace the batteries. Contact
ATAGO if this error persists.)
Automatic Temperature Compensation
The readings are corrected, based on the temperature of the sensor, within the automatic temperature compensation range.
【Caution】
◇Measurements may fluctuate with hot or cold samples. Wait for approximately 20 seconds to press the START button. Measurements will stabilize
once the instrument acclimates to the sample temperature.
Storage and Maintenance
Store the instrument in a dry place away from
direct sunlight. Exposure to humidity and heat
may damage the instrument.
Clean and dry the sample stage thoroughly, following the
“Cleaning”instructions.
Store the unit away from direct sunlight at a stable temperature
with as little fluctuation as possible.
Repair and Warranty
The instrument is warranted for one year from the date of purchase. This warranty is void if the instrument shows evidence of the following. Send the
included batteries as well if they are still in use.
Having been disassembled by unauthorized personnel
Damages to the sensor and/or sample stage
Water damage or having been dropped
Having been misused and/or operated outside the environmental specifications
Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires. Contact an ATAGO authorized service center for service and support.
Please have the serial number information ready when contacting a service center.
Specifications
Measurement range
Sodium hypochlorite 50-500ppm
10-40℃
Output NFC Forum Type 4 Tag , ISO/IEC 14443 Type A
Output category : Date Time,CONC[ppm], Temp [degC]
(e.g.) 2017/08/17 09:30:45, 503, 20.4
Resolution
1ppm / 0.1℃
Maximum number of data history 100
Accuracy
About ±50ppm / ±1℃
Sample volume
At least 0.6mL
Automatic temperature compensation range 10 to 40℃
Measurement time
Approx. 2 seconds
Ambient temperature range 10 to 40℃
Power supply
Two (2) AAA alkaline batteries
Backlight
The backlight stays on for 30 seconds
International Protection class IP65
after any button is pressed.
Dimensions and weight
55 (W) x 31 (D) x 109 (H)mm, 100g (main unit only)
Patent Granted in Japan, United States, Germany, China and Taiwan.
Headquarters: The Front Tower Shiba Koen,
23rd Floor 2-6-3 Shiba-koen, Minato-ku,
Tokyo 105-0011, Japan
TEL: 81-3-3431-1943 [email protected]
http://www.atago.net/
TEL: 1-425-637-2107
[email protected]
TEL: 91-22-2854-4915 / 4071-3232
[email protected]
TEL: 66-21948727-9 ,66-21171549
[email protected]
TEL: 55 16 3913-8400
[email protected]
TEL: 39 02 36557267
[email protected]
TEL: 86-20-38108256
[email protected]
TEL: 7-812-777-96-96
[email protected]
TEL: 234-707-558-1552
[email protected]
TEL: 7-727-257-08-95
[email protected]
2106K Printed in Japan
4555-E02
Hypochlorous acid Meter
Instruction Manual
PAL-Hypochlorous Acid (COVID-19) Cat.No.4555
Parts
Contents
Main unit ........ 1
Instruction Manual (this book)..... 1
AAA batteries ...........2
AAA alkaline batteries are included. Remove the tape from the battery compartment before inserting the batteries.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
Introduction
Thank you for purchasing the instrument. Carefully read and follow all instructions. Keep this manual for future reference.
Safety Instructions
Read and follow all safety instructions before operating the instrument. Failure to comply with the following instructions may result in personal injury or
property damage.
◇Ensure safety when handling hazardous materials. Observe precautionary measures and use protective equipment, Be aware of the hazards of such
chemicals and emergency response guidelines.
◇ATAGO may not be held liable for any injury or damage arising in connection with handling of hazardous materials during the use of the instrument.
◇Do not drop the instrument or subject it to strong physical shock.
◇Do not attempt to repair, modify, or disassemble the instrument.
◇Carefully read this manual to have basic knowledge of the function of each component.
◇ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.
◇Some acids may corrode the sensor and/or metal sample stage, which may cause erroneous measurements.
◇Do not use metal tools, such as a spoon, as they may scratch the sensor, resulting in erroneous measurements.
◇Do not use water above 50℃to rinse the instrument.
◇Only use the specified battery type. Observe proper polarities, properly aligning the anodes and cathodes.
◇Store the instrument away from direct sunlight/heat sources and excessive amounts of dust/debris.
◇Do not expose the instrument to a rapid change in ambient temperature.
◇Do not subject the instrument to strong vibration.
◇Do not subject the instrument to extreme cold temperature.
◇Do not place the instrument under anything heavy.
◇Loosen the battery compartment cover for air transportation.
<International Protection Classification IP65>
◇The instrument is water-resistant, not waterproof, and should not be submerged.
<Chemical Resistance of Body Case>
◇The body case is made of resin. Do not expose it to water vapor. Some solvents may compromise the structural integrity of the instrument.
Press to take measurements
and hold down to turn off
the display.
START button
(Power button)
Place and remove batteries from here.
Batterycompartment
ZERO button
Press to perform
zero-setting.
START button &
ZERO button
Press to set date, time,
and delete data history.
Lanyard hole
Sample stage
Apply water and
samples on the glass
prism located in the
center of the sample
stage.
LCD
Measurement results,
prism temperature,
remaining battery charge,
etc., are displayed.
The displayed value is an
example.
R Button
Press to restore default
settings for the Offset
Feature.
Image is for explanation purposes only. It may be
different than the actual product purchased.
Zero-setting and Measurement
ZERO Setting with the water used for the sample dilution.
【Caution】
◇Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries.
◇Let water on the sensor acclimate to the temperature of the instrument before zero-setting.
◇When “AAA”is displayed, wipe the sensor clean, apply water, and press the ZERO button again.
1. Wipe the sensor clean and apply about
0.6mL of water used for dilution.
2. Press the ZERO button while the
water is on the sensor.
3. Zero-setting is complete
when “000”blinks twice and then
stops.
Proceed to measurement.
Wipe the sensor clean with tissue
paper.
Press the ZERO button (with nothing on the sensor).⇒Clear the zero setting.
Measurement
1. Wipe the sensor clean and
apply about 0.6mL of the sample.
2. Press the START
button
3. After “---”appears, the
measurement and the
sensor temperature are
displayed.
【Caution】
◇This instrument displays the concentration of peracetic acidin water as a ppm of total volume.
◇Do not use metal tools to apply samples on the sensor as they may scratch the sensor.
◇Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature,
approximately 20 seconds, to press the START button. Alternatively, press the START button multiple times until
measurements become stable.
◇Do not splash water above 50℃. The plastic may warp, which may compromise the water resistance.
◇The displayed temperature is that of the sensor and may not necessarily match the temperature of the sample.
<LCD Auto Shut-off>
The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START
button for more than 2 seconds.
Cleaning
【Caution】
◇Do not scratch the sensor.
◇The instrument is water-resistant, not waterproof, and should not be submerged.
1. Wipe off the sample.
2. Clean the sensor and sample stage using a mild soap and thoroughly rinse with water.
3. Dry the area with tissues thoroughly.
Offset Function
1.Hold down
ZERO while it
is turned on.
“b”wil appear.
2.Press ZERO
to select either
addition (b) or
subtraction (-b).
3.Press START
to confirm.
4.Enter the addition/subtraction number.
ZERO to change the number:
0, 1, 2, …, 8, 9, 0, 1, 2…
START to confirm and move to the next
decimal place.
5.Press START
to confirm the
addition/subtracti
on number.
Next is to program
a coefficient.
6.Enter the
coefficient.
Refer to the
step 4.
7.Press
START to
confirm the
coefficient.
※Screen images when offset is on
y = ax + b y:Actual concentration
x:PAL readings
a:Coefficient (multiplication)
Range: 0.50 to 5.00
b:Addition/subtraction number
Range: -999 to 999(Factory default 0.00)
For default :Press the Ⓡbutton while setting up the Off-Set feature.
Checking with Reference Solution
When there is any doubt regarding accuracy of measurement results, adjust the reference value according to the following
procedur.
The reference solution is available from ATAGO. Part No. RE-130004 Reference solution (0.04% Citric acid solution)
1. Clean the sensor with water.
Dry the area with tissues
thoroughly. Press the ZERO
button (with nothing on the
sensor).
Clear the zero setting.
2. Apply the reference
solution.
At least 0.6mL
3. Turn on the instrument.
Press the START and ZERO
buttons simultaneously.
4. “CCC”will appear.
Flash ⇒Light up
Replacing the Batteries
【Caution】
◇Fasten the battery compartment cover tightly to prevent water ingress or poor connection, which will cause erroneous
measurements. Push the cover in firmly and turn.
◇When the O-ring on the cover is dirty or damaged, the water resistance may be compromised.
◇When the battery icon indicates the low power level ( ), replace both batteries with a brand new set of AAA alkaline batteries
(1.5V).
◇Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty display, consume the battery
power, or affect the instrument’s performance in any way.
◇Check the expiration dates on batteries before purchase.
◇Zero-set the instrument after the batteries are replaced.
1. Insert a coin in the
groove on the battery
compartment cover. Turn
the coin counterclockwise
to remove the cover.
2. Insert batteries,
observing the correct
polarity.
3. Align the cover
and push it down.
4. Close the battery compartment cover by
pushing the cover in with a coin in the groove
and turning it clockwise until it stops.
*Turning excessively may cause malfunction.
Open Close
Approx. 0.6mL of water
1 2 3
O-ring
Groove
Approx. 0.6mL of sample
For addition, only
“b”will appear.
1 2 3
Water after
zero-setting
Offset “b”=
addition of 30
During
measurement
For coefficient (a)
For addition/subtraction (b)
No plus sign will
be displayed.
Mild
soap
20.4℃
502

Other ATAGO Measuring Instrument manuals

ATAGO PAL-OWE2 User manual

ATAGO

ATAGO PAL-OWE2 User manual

ATAGO DR-M2 Manual

ATAGO

ATAGO DR-M2 Manual

ATAGO PAL-Easy ACID40 User manual

ATAGO

ATAGO PAL-Easy ACID40 User manual

ATAGO PAL-Grape Must User manual

ATAGO

ATAGO PAL-Grape Must User manual

ATAGO PAL-12S User manual

ATAGO

ATAGO PAL-12S User manual

ATAGO RX-5000 i-Pius User manual

ATAGO

ATAGO RX-5000 i-Pius User manual

ATAGO MASTER-M Series User manual

ATAGO

ATAGO MASTER-M Series User manual

ATAGO PAL-102S User manual

ATAGO

ATAGO PAL-102S User manual

ATAGO PAL-Easy ACID8 User manual

ATAGO

ATAGO PAL-Easy ACID8 User manual

ATAGO 4430 User manual

ATAGO

ATAGO 4430 User manual

ATAGO PAL-RI User manual

ATAGO

ATAGO PAL-RI User manual

ATAGO PAL-03S User manual

ATAGO

ATAGO PAL-03S User manual

ATAGO PAL-27S User manual

ATAGO

ATAGO PAL-27S User manual

ATAGO PAL-44S User manual

ATAGO

ATAGO PAL-44S User manual

ATAGO PAL-ORANGE User manual

ATAGO

ATAGO PAL-ORANGE User manual

ATAGO PAL-66S User manual

ATAGO

ATAGO PAL-66S User manual

ATAGO PAL-131S User manual

ATAGO

ATAGO PAL-131S User manual

ATAGO Pocket PAL-1 User manual

ATAGO

ATAGO Pocket PAL-1 User manual

ATAGO PAL-04S User manual

ATAGO

ATAGO PAL-04S User manual

ATAGO PAL-pH User manual

ATAGO

ATAGO PAL-pH User manual

ATAGO PAL-Easy ACID1 User manual

ATAGO

ATAGO PAL-Easy ACID1 User manual

ATAGO PAL-64S User manual

ATAGO

ATAGO PAL-64S User manual

ATAGO PAL-BX/SALT+5 User manual

ATAGO

ATAGO PAL-BX/SALT+5 User manual

ATAGO PAL- HIKARi 12 User manual

ATAGO

ATAGO PAL- HIKARi 12 User manual

Popular Measuring Instrument manuals by other brands

Dräger Polytron 7500 Instructions for use

Dräger

Dräger Polytron 7500 Instructions for use

horiba LAQUA F-71G instruction manual

horiba

horiba LAQUA F-71G instruction manual

Mark-10 EK3-100 quick start guide

Mark-10

Mark-10 EK3-100 quick start guide

MRC PVB-820 Operation manual

MRC

MRC PVB-820 Operation manual

Kreg Precision KMS7101 Instructional manual

Kreg

Kreg Precision KMS7101 Instructional manual

Ametek JOFRA CTC-140 A user manual

Ametek

Ametek JOFRA CTC-140 A user manual

Kipp & Zonen SMP series instruction manual

Kipp & Zonen

Kipp & Zonen SMP series instruction manual

Perel EVM125 quick start guide

Perel

Perel EVM125 quick start guide

Nexus 1500+ quick start guide

Nexus

Nexus 1500+ quick start guide

Zeiss GOM Scan 1 User manual hardware

Zeiss

Zeiss GOM Scan 1 User manual hardware

Tektronix 1L40 instruction manual

Tektronix

Tektronix 1L40 instruction manual

Vacom BARION atm II manual

Vacom

Vacom BARION atm II manual

SOCTESTER SOC AERO operating manual

SOCTESTER

SOCTESTER SOC AERO operating manual

SenseAir aSENSE mIII user manual

SenseAir

SenseAir aSENSE mIII user manual

Extech Instruments EX622 user guide

Extech Instruments

Extech Instruments EX622 user guide

Tegam R1L-B instruction manual

Tegam

Tegam R1L-B instruction manual

Sinmag Europe SE42 manual

Sinmag Europe

Sinmag Europe SE42 manual

Keithley SourceMeter 2400 user manual

Keithley

Keithley SourceMeter 2400 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.