Manual for use of thermometer-hygrometer ATAL ATC11 with surface
temperature probe
Instrument is designed for measurement of temperature and relative humidity (attached sensors)
and temperature from the built in magnetic surface probe. It enables to display the dew point
temperature and the difference between surface temperature and dew point temperature. Measured
values are displayed on a dual line LCD display. Temperature is measured by RTD sensor
Ni1000/6180ppm. Instrument compares measured values of temperature, humidity, dew point and
temperature difference with two adjustable levels for each measured quantity. Breaking the level is
indicated by blinking the proper value on display and by audio indication (switchable). Instrument
is equipped with minimum and maximum memory and Hold function. Minimum and maximum
values and Hold value are possible to display on the LCD anytime.
Technical parameters:
Parameters of measurement:
Ambient and surface temperature: Range of measurement: -10 to +60 °C
Resolution: 0.1 °C
Accuracy: ±0.4 °C
Humidity: Range measurement: 5 to 95 %RH
Resolution: 0.1 %RH
Accuracy: ±2.5 % RH at the range of 5 to 95 %RH at 23°C
Dew point temperature (calculated from air temperature and humidity):
Range: -50 to +60 °C
Resolution: 0.1 °C
Accuracy: ±0.5 °C at the range of 30 to 95 % RH
Measuring interval and display reading refresh: approximately 0.7 s in FAST mode
0.7 to 5 s in dynamic mode
Response time (air flow approximately 1 m/s):
air temperature: t63 < 35 s, t90 < 60 s (temperature step 20 °C)
relative humidity: t63 < 30 s, t90 < 60 s (humidity step 30 %RH, constant temperature)
Power: battery 9V or ac/dc adapter 12V with NiCd accumulator 9V
Average current consumption: 0.2 to 0.8 mA (depending on operation mode)
Typical life of Zinc-Chloride battery: 4 months
Typical life of Alkali-Mangan battery: 6 months
Protection: IP20
Operation conditions:
Ambient temperature range: -10 to +60 °C
Ambient relative humidity range: 5 to 95 %RH (non condensing)
Outer characteristics in accordance with EN 33-2000-3: normal environment with
characteristics AD1, AE1, AF1, AG1, AH1, AK1, AL1, AN1, AP1, AQ1, AR1,
AS1, BA1, BE1
Not allowed manipulation: it is not allowed to touch sensors under the cover to avoid
sensors damaging or to effect calibration.
The sensors (under the cover) should not be exposed to direct contact with water or other
liquids.
Storing conditions: temperature -10 to +60 °C relative humidity 5 to 95 %RH
Dimensions without probes: 141 x 71 x 27 mm
Weight including battery: approximately 200 g
Material of the case: ABS Forsan