Atal EPND-T-INT-POE User manual

2
Inhoud
Notice, installation and troubleshooting.................................................................................................1
Introduction......................................................................................................................................................4
General safety rules .....................................................................................................................................4
Device description and important notices...................................................................................................5
Getting started .................................................................................................................................................6
What is needed for operation......................................................................................................................6
Mounting the device....................................................................................................................................6
Device settings .............................................................................................................................................8
Checking functions.......................................................................................................................................9
Device setup ...................................................................................................................................................10
Setup using web interface..........................................................................................................................10
General ......................................................................................................................................... 11
Network ........................................................................................................................................ 11
Alarm limits................................................................................................................................... 12
Channels ....................................................................................................................................... 13
SOAP protocol............................................................................................................................... 13
Email ............................................................................................................................................. 14
Modbus a Syslog protocols........................................................................................................... 15
SNMP ............................................................................................................................................ 15
Time .............................................................................................................................................. 15
WWW and security....................................................................................................................... 16
Setup using TSensor software....................................................................................................................16
Factory defaults .........................................................................................................................................17
Communication protocols..............................................................................................................................19
Website ......................................................................................................................................................19
SMTP –sending e-mails .............................................................................................................................19
SNMP 19
Modbus TCP ...............................................................................................................................................20
SOAP 22
Syslog 24
SNTP 24
Troubleshooting .............................................................................................................................................25
I forgot the device IP address ....................................................................................................................25
I cannot connect to the device...................................................................................................................25
In search window is only IP and MAC address displayed ............................................................. 25
Device IP address is not displayed in window Find device........................................................... 25
The device is not found even after manually setting MAC address......................................... 25
Error or n/a is displayed instead the measured value...............................................................................25
Error2 is displayed on all channels instead the measured value...............................................................26
I forgot the password for setup .................................................................................................................26
Factory defaults .........................................................................................................................................26
Technical specifications..................................................................................................................................27
Dimensions.................................................................................................................................................27
Basic parameters........................................................................................................................................28
Operating terms.........................................................................................................................................30
End of operation ........................................................................................................................................30
Technical support and service....................................................................................................................30
Preventive maintenance ............................................................................................................................31

3
Optional accessories.......................................................................................................................................32
Temperature probe TEDX-102E.................................................................................................... 32
Temperature probe TED-102E...................................................................................................... 32
Temperature probe TEPD-102C.................................................................................................... 32
Humidity probe TRHD-101............................................................................................................ 32
Power supply Adapter EPND-AC01............................................................................................... 32
Device case holder for RACK 19" ATS-AC20 ................................................................................. 33
Probes holder for RACK 19" ATS-AC21 ......................................................................................... 33
CDB-WIN database ....................................................................................................................... 33

4
Introduction
This chapter provides basic information about device. Before
starting please read this manual carefully.
Thermometer Web Sensor EPND-T-INT-POE, Web Sensor EPND-3T-POE and Web Sensor EPND-
4S-POE are designed to measure temperature or relative humidity. Temperature can be
displayed in °C or °F. Relative humidity have unit %RH. Communication with the device is
realized via Ethernet network. Device can be powered from external power supply adapter or
by using power over Ethernet –PoE.
Thermometer Web Sensor EPND-T-INT-POE has compact design and measures the
temperature in place of installation. To Web Sensor EPND-3T-POE is possible connect up to
three probes. Web Sensor EPND-4S-POE supports up to four probes. Temperature or humidity
probes are available as optional accessories.
General safety rules
The following summary is used to reduce the risk of injury or
damage the device. To prevent injury, please follow instructions
in this manual.
The device can be services only by a qualified person. The device contains no
serviceable parts inside.
Don’t use the device, if it doesn’t work correctly. If you think, that the device is not
working correctly, let check it by qualified service person.
Don’t disassemble the device. It’s forbidden to use the device without the cover. Inside the
device can be a dangerous voltage and may be risk of electric shock.
Use only the appropriate power supply adapter according to manufacturer specifications and
approved according to relevant standards. Make sure, that the adapter does not have
damaged cables or covers.
Connect the device only to network parts approved according to relevant standards. Where
power over Ethernet is used, the network infrastructure must be compatible with IEEE 802.3af
standard.
Connect and disconnect the device properly. Don’t connect or disconnect Ethernet cable or
probes, if the device is powered.
The device may be installed only in prescribed areas. Never expose the device to higher or
lower temperatures than is allowed. The device has not improved resistance to moisture.
Protect it from dripping or splashing water and do not use at areas with condensation.
Chapter
1

5
Don’t use device in explosive environments.
Don’t stress the device mechanically.
Device description and important notices
This chapter contains information about basic features. Also
there are important notices concerning to functional safety.
Values from the device can be read using an Ethernet connection. The following formats are
supported:
Web pages with user changeable look and XML files
Modbus TCP protocol
SNMPv1 protocol
SOAP protocol
The device can also be used to check measured values and if the limit is exceeded, device
sends warning messages. Possible ways to sending warning messages:
Sending e-mails up to 3 e-mail addresses
Sending SNMP traps up to 3 configurable IP addresses
Displaying the alarm status on web page
Sending messages to Syslog server
The device setup can be made by the TSensor software or web interface. TSensor software can
be free downloaded from the manufacturer’s website. Also you will find there latest firmware
for your device. Do not upload to your device firmware which is not designed for it.
Unsupported firmware can damage your device.
If you want to use PoE, you must use PoE switch compatible with IEEE 802.3af standard.
Reliability of warning messages delivering (e-mail, trap, syslog), depends on actual
availability of necessary network services. The device should not be used for
critical applications, where malfunction could cause to injury or loss of human life.
For highly reliable systems, redundancy is essential. For more information please
see standard IEC 61508.
Never connect the device directly to the Internet. If it is necessary connect the
device to the Internet, properly configured firewall must be used. Firewall can be
partially replaced with the NAT.

6
Getting started
Here you can find information necessary to put newly purchased
equipment to operation. This procedure is only informative.
What is needed for operation
To install the unit you need to the following equipment. Before installation check if it’s
available.
thermometer Web Sensor EPND-T-INT-POE, Web Sensor EPND-3T-POE or EPND-
4S-POE
power supply adapter 5V/250mA or switch with PoE. Before using the device is
necessary to decide which way of powering will be used.
RJ45 LAN connection with appropriate cable
free IP address in your network
for Web Sensor EPND-4S-POE up to 4 temperatures probes type TEDX-102E, TED-
102E, TEPD-102C or relative humidity probe TRHD-101. Web Sensor EPND-3T-
POE supports up to 3 probes.
Mounting the device
check if the equipment from previous chapter are available
install the latest version of TSensor software. This software is used to all device
settings. TSensor software can be free downloaded from the manufacturer’s
website. Software can be also supplied on CD. Device configuration can be made
using web interface. For web configuration is not TSensor software necessary.
contact your network administrator to obtain following information for the
connection to the network:
Chapter
2

7
IP address:
_____._____._____._____
Gateway:
_____._____._____._____
DNS server IP:
_____._____._____._____
Netmask:
_____._____._____._____
check if there is no IP address conflict when you connect the device into network
for the first time. The device has from factory set the IP address to
192.168.1.213. This address must be changed according to information from the
previous point. When you installing several new devices, connect them to the
network one after another.
connect probes to Web Sensor EPND-3T-POE or Web Sensor EPND-4S-POE
connect the Ethernet connector
if the power over Ethernet (PoE) is not used, connect the power adapter
5V/250mA
LEDs on LAN connector should blink after connecting the power
Web Sensor EPND-T-INT-POE connection (power supply adapter, Power over Ethernet):
Web Sensor EPND-4S-POE and EPND-3T-POE connection (power supply adapter, Power over
Ethernet):

8
Device settings
run configuration software TSensor on your PC
switch to an Ethernet communication interface
press button Find device...
the window shows all available devices on your network
click to Change IP address to set new address according to network
administrator instructions. If your device is not listed, then click Help! My device
wasn’t found! Then follow the instructions. MAC address is on product label. The
device is factory set to IP 192.168.1.213.
gateway may not be entered if you want to use the device only in local network.
If you set the same IP address which is already used, the device will not work
correctly and there will be collisions on the network. If the device detects a
collision of IP address then reboot is performed automatically.

9
after changing IP address device is restarted and new IP address is assigned.
Restart of the device takes about 10 seconds.
connect to device using TSensor software and check the measured values. If Web
Sensors EPND-3T-POE and EPND-4S-POE values are not displayed, it’s necessary
to find probes using button Search probes (Find probes).
set the other parameters (alarm limits, SMTP server, etc.). Settings are saved
after click on button Save changes.
Checking functions
The last step is to check measured values on the device website. In the address bar of the web
browser, enter the device IP address. If the default IP address was not changed, then insert
http://192.168.1.213.
Displayed web page lists actual measured values. If the web pages are disabled, you can see
text Access denied. If the measured value exceeds the measurement range or probe is not
correctly installed, then is shown Error message. If the channel is switched off, the web site
displayed n/a instead of the value.

10
Device setup
This chapter describes basic device configuration. There is a
description of settings using web interface.
Setup using web interface
Device can be setup using web interface or TSensor software. Web interface can be managed
by the web browser. Main page will be shown when you insert device address into address bar
of your web browser. There you find actual measured values. Page with history graphs is
shown when you click to tile with actual values. Access to device setup is possible via tile
Settings.
Chapter
3

11
General
Device name can be changed using item Device name. Measured values are stored into
memory according History storage interval field. After changing of this interval all history
values will be cleared. Changes must be confirmed by Apply settings button.
Network
Network parameters can be obtain automatically from DHCP server using option Obtain an IP
address automatically. Static IP address is configurable via field IP address. It is not necessary
setup Default gateway while you use device inside one subnet only. DNS server IP is required
to set for proper function of DNS. Option Standard subnet mask sets network mask
automatically according A, B or C network class. Subnet mask field must be set manually, when
network with non-standard range is used. Periodic restart interval enables to restart device
after selected time since device start.

12
Alarm limits
For each measurement channel is possible to set upper and lower limits, time-delay for alarm
activation and hysteresis for alarm clearing.
Example of setting the limit to the upper alarm limit:
In Point 1 the temperature exceeded the limit. From this time, the time-delay is counting.
Because at point 2 the temperature dropped below the limit value before the time delay
expired, alarm was not set.
In Point 3 the temperature has risen over limit again. During the time-delay the value does not
drop below the set limit, and therefore was in Point 4 caused alarm. At this moment were sent
e-mails, traps and set alarm flag on website, SNMP and Modbus.
The alarm lasted up to Point 5, when the temperature dropped below the set hysteresis
(temperature limit –hysteresis). At this moment was active alarm cleared and e-mail send.
When alarm occurs, alarm messages will be sent. In case of power failure or device reset (e.g.
changing the configuration) will new alarm state evaluated and new alarm messages will be
send.

13
Channels
Channel can be enabled or disabled for measuring using item Enabled. Channel can be
renamed (max. 14 characters) and it is possible select unit of measured value according
connected probe type. When channel is not used, it is possible copy to it one of other channels
–option Clone channel. This option is not available at fully occupied device. Find sensors
button starts searching for connected probes. All changes must be confirmed using Apply
settings button. History values are cleared after changing of channel settings.
SOAP protocol
SOAP protocol can be enabled by option SOAP protocol enabled. Destination SOAP server can
be set via SOAP server address. For setup of server port can be used option SOAP server port.
Device sends SOAP message according selected Sending interval. Option Send SOAP message
when alarm occurs sends message when an alarm on channel occurs or alarm is cleared. These
SOAP messages are sent asynchronously to selected interval.

14
Email
Email sending enabled option allows email features. It is necessary set address of the SMTP
server into SMTP server address field. Domain name for SMTP server can be used. Default port
of the SMTP server can be changed using item SMTP server port. SMTP authentication can be
enabled using SMTP authentication option. When authentication is enabled Username and
Password must be set.
For successfully email sending it is necessary insert Email sender address. This address is
usually same as username of the SMTP authentication. Into fields Recipient 1 to Recipient 3 it
is possible set address of email recipients. Option Short email enable sending emails in short
format. This format is usable when you need to forward emails into SMS messages.
When option Alarm email repeat sending interval is enabled and there is active alarm on
channel, then emails with actual values are sent repeatedly. Info email sending interval option
enables sending emails at selected time interval. CSV history file can be sent together with the
repeat/info emails. This feature can be enabled by Alarm and Info emails attachment option.
It is possible to test email function using button Apply and test. This button save a new
settings and send a testing email immediately.

15
Modbus a Syslog protocols
ModbusTCP and Syslog protocol settings are configurable via menu Protocols. Modbus server
is enabled by default. Deactivation is possible via Modbus server enabled option. Modbus port
can be changed via Modbus port field. Syslog protocol can be enabled using item Syslog
enabled. Syslog messages are sent to IP address of the Syslog server - field Syslog server IP
address.
SNMP
For reading values via SNMP it is necessary to know password - SNMP read community. SNMP
Trap can be delivered up to three IP address - IP address of the Trap recipient. SNMP Traps are
sent at alarm or error state on the channel. Trap feature can be enabled by option Trap
enabled.
Time
Time synchronization with SNTP server can be enabled by Time synchronization enabled
option. IP address of the SNTP is necessary to set into SNTP server IP address item. List of free
NTP servers is available at www.pool.ntp.org/en. SNTP time is synchronized at UTC format,

16
and due to be necessary set corresponding time offset - GMT offset [min]. Time is
synchronized every 24 hours by default. Option NTP synchronization every hour decrease this
synchronization interval to one hour.
WWW and security
Security features can be enabled by the Security enabled option. When security is enabled it is
necessary to set administrator password. This password will be required for device settings.
When secured access is required even to actual values reading it is possible to enable User
account only for viewing. Port of the www server can be changed from the default value 80
using filed WWW port. Web pages with actual values are refreshed according to Web refresh
interval field.
Setup using TSensor software
TSensor software is an alternative to web configuration. Some less important parameters are
configurable only by the TSensor software.

17
Parameter MTU size can reduce size of the Ethernet frame. Lowering of this size can solve
some communication problems mainly with Cisco network infrastructure. TSensor software
can set offset of values at temperature probes. At TRHD-101 humidity probe is possible set
correction of the humidity and temperature.
Factory defaults
Factory defaults button set the device into factory configuration. Network parameters (IP
address, Subnet mask, Gateway, DNS) are left without changes.
Network parameters are changed while you press button at left side of device during power
supply connection. Factory defaults have no effect to user correction inside probe.

18
Parameter
Value
SMTP server address
example.com
SMTP server port
25
Alarm email repeat sending interval
off
Info email repeat sending interval
off
Repeat email attachment
off
E-mail recipients addresses
cleared
E-mail sender
sensor@IP
SMTP authentication
off
SMTP user/SMTP password
cleared
E-mail sending enabled
off
IP addresses SNMP traps recipients
0.0.0.0
System location
cleared
Password for SNMP reading
public
Sending SNMP Trap
off
Website refresh interval [sec]
10
Website enabled
yes
Website port
80
Security
off
Administrator password
cleared
User password
cleared
Modbus TCP protocol port
502
Modbus TCP enabled
yes
History storage interval [sec]
60
SOAP message when alarm occurs
yes
SOAP destination port
80
SOAP server address
cleared
SOAP sending interval [sec]
60
SOAP protocol enabled
off
Syslog server IP address
0.0.0.0
Syslog protocol enabled
off
SNTP server IP address
0.0.0.0
GMT offset [min]
0
NTP synchronization every hour
off
SNTP synchronization enabled
off
MTU
1400
Periodic restart interval
off
Demo mode
off
Turn on all channels
yes
Upper limit
50
Lower limit
0
Hysteresis –hysteresis for alarm clearing
1
Delay –time-delay of alarm activation [sec]
30
Channel enabled
all channels
Unit on the channel
°C or %RH according used probe
Channel name
Channel X (where X is 1 to 5)
Device name
Web sensor

19
Communication protocols
Short introduction to communication protocols of the device. To
use some communication protocols is necessary software, which
can use the protocol. This software is not included.
Website
The device supports displaying of measured values, history graphs and configuration using web
browser. History graphs are based on HTML5 canvas. Web browser must support this feature
for proper function of graphs. Firefox, Opera, Chrome or Internet Explorer 9 can be used. If the
device has IP address 192.168.1.213 type into your browser http://192.168.1.213. Using
TSensor software or web interface can be set automatic webpages refresh interval. The default
value is 10sec. Actual measured values can be obtained using XML file values.xml.
Values from history can be exported in CSV format. History storage interval can be set using
TSensor software or web interface. History is erased after every reboot of the device. Reboot
of the device is performed when the power supply is disconnected and also after configuration
change.
The device allows you to customize the design of web pages. Details of how to change the
design, you can find in the manual appendix. Manual appendix can be obtained from the
manufacturer’s website.
SMTP –sending e-mails
When measured values are over the set limits, the device allows send e-mail to a maximum of
3 addresses. E-mail is send when alarm condition on the channel is cleared or a measuring
error occurs. It is possible to set repeat interval for email sending. For correct sending of e-
mails it is necessary to set address of SMTP server. Domain address can be used as SMTP
server address too. For proper function of DNS is required to set DNS server IP address. SMTP
authentication is supported but SSL/STARTTLS not. Standard SMTP port 25 is used by default.
SMTP port can be changed. Contact your network administrator to obtain configuration
parameters of your SMTP server. E-mail sent by the device cannot be answered.
SNMP
Using SNMP protocol you can read actual measured values, alarm status and alarm
parameters. Via SNMP protocol is also possible to get last 1000 measured values from history
table. Writing via SNMP protocol is not supported. It is supported SNMPv1 protocol version
only. SNMP used UDP port 161. OID keys description can be found in the MIB table, which can
be obtained from device website or from your distributor. The password for reading is factory
set to public. Filed System location (OID 1.3.6.1.2.1.1.6 - sysLocation) is blank by default. The
changes can be made using web interface. Some SNMP manager software requires add .0 at
end of OID key (e.g.: .1.3.6.1.4.1.22626.1.5.2.1.2.0). OID keys:
Chapter
4

20
OID
Description
Type
.1.3.6.1.4.1.22626.1.5.1
Device information
.1.3.6.1.4.1.22626.1.5.1.1
Device name
String
.1.3.6.1.4.1.22626.1.5.1.2
Serial number
String
.1.3.6.1.4.1.22626.1.5.1.3
Device type
Integer
.1.3.6.1.4.1.22626.1.5.2.ch
Measured value (where ch is channel number)
.1.3.6.1.4.1.22626.1.5.2.ch.1
Channel name
String
.1.3.6.1.4.1.22626.1.5.2.ch.2
Actual value –text
String
.1.3.6.1.4.1.22626.1.5.2.ch.3
Actual value
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.4
Alarm on channel (0/1/2)
Integer
.1.3.6.1.4.1.22626.1.5.2.ch.5
High limit
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.6
Low limit
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.7
Hysteresis
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.8
Delay
Integer
.1.3.6.1.4.1.22626.1.5.2.ch.9
Unit
String
.1.3.6.1.4.1.22626.1.5.2.ch.10
Alarm on channel –text
String
.1.3.6.1.4.1.22626.1.5.2.ch.11
Minimal value on channel
String
.1.3.6.1.4.1.22626.1.5.2.ch.12
Maximal value on channel
String
.1.3.6.1.4.1.22626.1.5.3.1.0
SNMP Trap text
String
.1.3.6.1.4.1.22626.1.5.4.1.1.ch.nr
History table value
Int*10
When alarm occurred a warning messages (trap) can be sent to selected IP addresses.
Addresses can be set using TSensor software or web interface. Traps are sent via UDP protocol
on port 162. The device can send following traps:
Trap
Description
0/0
Reset of the device
1/0
Settings was changed
6/0
Testing Trap
6/1
NTP synchronization error
6/2
E-mail sending
error
SMTP server login error
6/3
SMTP authentication error
6/4
Some error occurred during SMTP communication
6/5
TCP connection to server cannot be opened
6/6
SMTP server DNS error
6/7
SOAP message
sending error
SOAP file not found inside web memory
6/8
MAC address can't be obtained from address
6/9
TCP connection to server cannot be opened
6/10
Wrong response code from the SOAP server
6/11 –6/15
Upper alarm on channel
6/21 –6/25
Lower alarm on channel
6/31 –6/35
Clearing alarm on channel
6/41 –6/45
Measuring error
Modbus TCP
Device supports Modbus protocol for communication with SCADA systems. Device use Modbus
TCP protocol. TCP port is set to 502 by default. Port can be changed using TSensor software or
web interface. Only two Modbus clients can be connected to device at one moment. Modbus
device address (Unit Identifier) can be arbitrary. Modbus write command is not supported.
Other manuals for EPND-T-INT-POE
1
Table of contents
Other Atal Measuring Instrument manuals