
Model B12-70WetAcidGasTransmitter
O&M Manual-15 -
-Release E, 7/15
CALIBRATION
Transmittercalibration requiresadjustment ofboth zeroand span. Zeroisadjusted when the
sensorisexposed toaircontaining no acidgases.Spanisadjustedwhen agasstreamcontaining a
known concentrationofacid gasispassed through the lowvolume flowcell.
ZEROADJUSTMENT
Aspreviouslymentioned, adjusting thetransmitterzerorequiresthat the sensorbe exposed to
acidgasfree air.The simplestapproachistosimplyexposethesensorto ambient airthat isfree of
gasestowhichthe sensorwillrespond.Threadingthelowvolumeflowcellontothesensorwill isolate
the sensorfromsurrounding airforzeroing purposes.
Ifthe areainwhichthe sensorisoperatingisknownto begasfree, thenthetransmittercan be
zeroedwithout furtherequipment. Ifnot, useof“zeroair” fromagascylinderisrecommended. Zeroair
isavailableaspartofallATIcalibration kits,ormaybeobtained fromanyspecialtygassupplier.When
zeroairisto beused, alowvolumeflowcell(partno.00-0422)mustbe used.The lowvolumeflowcell
providesaconfined spacearoundthesensorintowhichthe zeroaircanflow.Lowvolumeflowcells
provide tube fittingsatthe bottomto connect airtubing asshown in Figure 2.
Tozerothetransmitter,removethe coverandconnectaDVMtothe testpointsshowninFigure
6. Observethe DVMvalueto be surethat itisnolongerdeclining, andthat itisstable±4mv.Usethe
zeropotentiometer(Figure6)to adjustthe testpoint voltage to0.040 VDC (40mv.).Ifthe transmitteris
supplied with the LCD option, noDVM isrequired.
SPANADJUSTMENT
Oncethe zerohasbeen set, threadthe lowvolumeflowcell onto the sensor(ifitisnot already
therefromzeroing).Attachasourceofgaswithaknowacidgasconcentrationandadjustthe sample
flowrateto 500cc/min. Forconvenience, chlorinegasmaybeused asasimulant forcalibration. Atypical
chlorine gassourceistheelectrochemicalchlorinegeneratormanufactured byAdvancedCalibration
Designsin Tucson, AZ. Allowsample to flow for10 minutes.
ObservetheDVMvalue. The reading shouldbe relativelystable±4mv.Adjustthe span
potentiometer(Figure6)untilthe propertestpoint voltageisobtained. The propervoltage to set when
adjustingtransmitterspanwillvarydepending on therangeofthetransmitter.The following formulais
used tocalculatetheproperspan voltage foranytransmitter.
V = 0.040 V + [0.160X (Measured Concentration ÷Transmitter Range)]
Asanexample,supposeyou arecalibrating atransmitterwitharangeof0-10 PPM,andthe
chlorine gasstandard used forcalibrationhasavalue of5 PPM. The above formulawould thenbecome:
V = 0.040 V + [0.160X (5 ÷10)] = 0.120V.
Therefore, adjusting the transmitterto a reading of120 mv. at the test point wouldproperly
calibrate the transmitter.