
2601 Stanley Gault Pkwy, Suite 101 •Louisville, KY 40223 USA
800-923-8413 •502-244-6345 •FAX: 502-244-0369 www.avantevet.com
DISPOSABLE PATIENT CIRCUIT BATTERY BACK-UP MODE
Attach High or Low Flow Vapor Transfer
Cartridge to Disposable Water Path.
Attach patient delivery tube to
Disposable Water Path.
Open door and install Disposable Water Path into
docking station.
Hang sterile water bag.
Wipe water spike with alcohol pad and insert into
sterile water bag.
Allow a minimum of 200ml of water to ll into the
Disposable Patient Circuit.
Volumax VOS is ready for start up.
To activate the internal backup battery mode,
the Run/Standby LED must be green before
disconnecting the power cord.
The unit will enter into battery mode and will maintain
ow and oxygen percentage for at least 15 minutes
The battery icon will ash.
CONNECT TO PATIENT
The ashing green LED becomes steady when the
set variables are reached.
Place the cannula on the delivery tube.
After cannula is attached, it is ready to be placed
on patient.
START UP AND ADJUSTING PARAMETERS
SHUT DOWN MODE
Install oxygen sensor.
Install gass inlet lters on back with lter
bowls vertical (glass side down).
Attach air & O2 hoses. Plug in power cord.
Rotate the blue Setting Control Knob to
illuminate display.
Press in Setting Control Knob to select the parameter
and rotate to adjust the value.
Press and release (do not hold) the Run/Standby button
once to start. A GREEN light indicates RUN mode
(AMBER light indicates STANDBY mode, No Flow).
Green light will stop ashing once temperature
is reached.
Press the Run/Standby button. Unit will enter
Standby mode.
Clamp the water inlet tube and disconnect it by
pulling out the spike.
Open the door, remove the Disposable Patient Circuit
(includes delivery tube & cartridge) by sliding it upwards
out of the docking station
Discard all disposables according
to hospital guidelines.
Disconnect unit from AC power.