Axeon Water Technologies L1-Series User manual

Membrane System
User’s Manual
L1-Series

L1-Series User’s Manual
MKTF-258 2 9/14
This Page
Left Blank

L1-Series User’s Manual
MKTF-258 3 9/14
TABLE OF CONTENTS
INTRODUCTION ....................................................................................................................................... 4
SAFETY ..................................................................................................................................................... 5
FEED WATER AND OPERATION SPECIFICATIONS .................................................................................... 6
REJECTION, RECOVERY AND FLOW RATES .............................................................................................. 7
SYSTEM INSTALLATION AND START-UP PROCEDURES ............................................................................ 8
MEMBRANE ELEMENTS ......................................................................................................................... 10
SYSTEM IDENTIFICATION ....................................................................................................................... 11
MEMBRANE INSTALLATION, REMOVAL AND REPLACEMENT ............................................................... 1
OPERATION DO’S AND DON’TS ............................................................................................................. 14
SPECIFICATIONS ..................................................................................................................................... 15
OPERATION AND MAINTENANCE .......................................................................................................... 16
PREPARING UNIT FOR STORAGE OR SHIPMENT.................................................................................... 17
REVERSE OSMOSIS TROUBLESHOOTING ............................................................................................... 18
TEMPERATURE CORRECTION FACTORS FOR MEMBRANE .................................................................... 0
SERVICE ASSISTANCE ............................................................................................................................. 1
OPERATION LOG ....................................................................................................................................
SYSTEM DRAWING ................................................................................................................................. 3
L1- 00 SERIES FLOW DIAGRAM ............................................................................................................. 4
L1 - SERIES ELECTRICAL SCHEMATIC 110V/60HZ .................................................................................. 4
L1 - SERIES ELECTRICAL SCHEMATIC 0V/50HZ .................................................................................. 6
REVERSE OSMOSIS SYSTEM WARRANTY ............................................................................................... 7

L1-Series User’s Manual
MKTF-258
Your L1 - Series RO S
ystem is a durable piece of equipment which, with proper care,
will last for many years. This User’s Manual outlines installation, operation,
maintenance and troubleshooting details vital to the sustained performance of your
system.
If your system is altered at the site of operation
change, please
consult your local dealer or distributor for assistance.
NOTE: IN ORDER TO MAINTAIN THE MANUFACTURER’S WARRANTY,
AN OPERATING LOG MUST BE MAINTAINED AND COPIES WILL NEED TO BE
SENT TO YOUR LOCAL DEALER OR DISTRIBUTOR FOR REVIEW.
NOTE: PRIOR TO OPERATING OR SERVICING THE REVERSE
OSMOSIS SYSTEM, THIS USER’S
UNDERSTOOD. KEEP THIS AND OTHER ASSOCIATED INFORMATION FOR
FUTURE REFERENCE AND FOR NEW OPERATORS OR QUALIFIED
PERSONNEL NEAR THE SYSTEM.
4
INTRODUCTION
ystem is a durable piece of equipment which, with proper care,
will last for many years. This User’s Manual outlines installation, operation,
maintenance and troubleshooting details vital to the sustained performance of your
If your system is altered at the site of operation
,
or if the feed water conditions
consult your local dealer or distributor for assistance.
NOTE: IN ORDER TO MAINTAIN THE MANUFACTURER’S WARRANTY,
AN OPERATING LOG MUST BE MAINTAINED AND COPIES WILL NEED TO BE
SENT TO YOUR LOCAL DEALER OR DISTRIBUTOR FOR REVIEW.
NOTE: PRIOR TO OPERATING OR SERVICING THE REVERSE
OSMOSIS SYSTEM, THIS USER’S
MANUAL MUST BE READ AND FULLY
UNDERSTOOD. KEEP THIS AND OTHER ASSOCIATED INFORMATION FOR
FUTURE REFERENCE AND FOR NEW OPERATORS OR QUALIFIED
PERSONNEL NEAR THE SYSTEM.
9/14
ystem is a durable piece of equipment which, with proper care,
will last for many years. This User’s Manual outlines installation, operation,
maintenance and troubleshooting details vital to the sustained performance of your
or if the feed water conditions
NOTE: IN ORDER TO MAINTAIN THE MANUFACTURER’S WARRANTY,
AN OPERATING LOG MUST BE MAINTAINED AND COPIES WILL NEED TO BE
NOTE: PRIOR TO OPERATING OR SERVICING THE REVERSE
MANUAL MUST BE READ AND FULLY
UNDERSTOOD. KEEP THIS AND OTHER ASSOCIATED INFORMATION FOR
FUTURE REFERENCE AND FOR NEW OPERATORS OR QUALIFIED

L1-Series User’s Manual
MKTF-258
The Safety section of this User’s Manual outlines the various safety headings
throughout this manual’s text and are enhanced and defined below:
NOTE: INDICATES STATEMENTS THAT PROVIDE FURTHER INFORMATION
AND CLARIFICATION.
CAUTION: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY
CONDITIONS OR PRACTICES THAT COULD RESULT IN
OTHER PROPERTY DAMAGE.
WARNING: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY
CONDITIONS OR PRACTICES THAT COULD RESULT IN INJURY OR LOSS OF
LIFE. FAILURE TO FOLLOW WARNINGS COULD RESULT IN SERIOUS INJURY
OR EVEN DEATH.
DO NOT UNDER ANY
WARNING, OR OTHER DESCRIPTIVE LABELS FROM THE SYSTEM.
PLEASE READ T
H
THE INSTALLATION AN
D
ATTACHED INSTRUCTIO
N
THE PRODUCT’S FAILU
R
AND/OR PERSONAL I
NJU
•
DO NOT USE
W
UNSAFE OR O
F
DISINFECTIO
N
•
PRETREATM
EN
CHEMICALS
, ORGANICS OR INORGA
THE MEMBRA
N
•
ALWAYS TU
RN
5
SAFETY
The Safety section of this User’s Manual outlines the various safety headings
throughout this manual’s text and are enhanced and defined below:
NOTE: INDICATES STATEMENTS THAT PROVIDE FURTHER INFORMATION
CAUTION: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY
CONDITIONS OR PRACTICES THAT COULD RESULT IN
EQUIPMENT OR
OTHER PROPERTY DAMAGE.
WARNING: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY
CONDITIONS OR PRACTICES THAT COULD RESULT IN INJURY OR LOSS OF
LIFE. FAILURE TO FOLLOW WARNINGS COULD RESULT IN SERIOUS INJURY
DO NOT UNDER ANY
CIRCUMSTANCE REMOVE ANY CAUTION,
WARNING, OR OTHER DESCRIPTIVE LABELS FROM THE SYSTEM.
H
E ENTIRE MANUAL BEFORE PROCE
EDI
D
STARTUP. YOUR FAILURE TO FOLLO
W
N
S OR OPERATING PARAMETERS MA
Y
R
E, WHICH CAN CAUSE PROPERTY DA
M
NJU
RY.
W
HERE THE WATER IS MICROBIOLOGI
C
F
UNKNOWN QUALITY WITHOUT ADEQ
UA
N
BEFORE OR AFTER THE SYSTEM.
EN
T MUST BE SUFFICIENT TO ELIMINAT
E
, ORGANICS OR INORGA
NICS THAT
COULD
N
E MATERIAL.
RN
OFF THE UNIT, SHUT OFF THE FEED
W
9/14
The Safety section of this User’s Manual outlines the various safety headings
used
NOTE: INDICATES STATEMENTS THAT PROVIDE FURTHER INFORMATION
CAUTION: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY
EQUIPMENT OR
WARNING: INDICATES STATEMENTS THAT ARE USED TO IDENTIFY
CONDITIONS OR PRACTICES THAT COULD RESULT IN INJURY OR LOSS OF
LIFE. FAILURE TO FOLLOW WARNINGS COULD RESULT IN SERIOUS INJURY
CIRCUMSTANCE REMOVE ANY CAUTION,
WARNING, OR OTHER DESCRIPTIVE LABELS FROM THE SYSTEM.
EDI
NG WITH
W
Y
LEAD TO
M
AGE
C
ALLY
UA
TE
E
COULD
ATTACK
W
ATER, AND

L1-Series User’s Manual
MKTF-258 6 9/14
DISCONNECT THE ELECTRICAL POWER BEFORE WORKING ON
THE UNIT.
•
NEVER ALLOW THE PUMP TO RUN DRY.
•
NEVER ALLOW THE UNIT TO FREEZE OR OPERATE WITH A FEED
WATER TEMPERATURE ABOVE
100
°
F
.
FEED WATER AND OPERATION SPECIFICATIONS
Nothing has a greater effect on a reverse osmosis system than the feed water
quality.
NOTE: IT IS VERY IMPORTANT TO MEET THE MINIMUM FEED WATER
REQUIREMENTS. FAILURE TO DO SO WILL CAUSE THE MEMBRANE TO
FOUL AND VOID THE MANUFACTURER’S WARRANTY.
OPERATING LIMITS
Operating Parameters:*
Feed Temperature
40
-
85°F
System Inlet Pressure
45
-
85 PSIG
Maximum Operating Pressure
(@77°F)
1
0
PSIG (Nominal)
*If any of the feed water parameters are not within the limits given, consult your local dealer or
distributor for assistance.
Feed Water Requirements:**
Maximum SDI Rating
< 3
Maximum
Turbidity
1 NTU
Maximum Free Chlorine and/or
C
hloramines
0 PPM
pH (continuous)
3
-
11
pH (cleaning
-
30 min.)
-
1
**If any of the feed water parameters are not within the limits given, consult your local dealer or
distributor for assistance.
NOTE: HIGHER TDS AND/OR LOWER TEMPERATURES WILL REDUCE
THE SYSTEM’S PRODUCTION.

L1-Series User’s Manual
MKTF-258 7 9/14
REJECTION, RECOVERY AND FLOW RATES
L1 - Series Reverse Osmosis Systems are designed to produce permeate water at
the capacities indicated by the suffix in the system’s name under the conditions listed
above. For example, the L1-300 produces 300 gallons per day of permeate water at
the listed operating test conditions.
The amount of total dissolved solids (TDS) rejected by the membrane is expressed
as a percentage. For example, a 98.5% rejection rate means that 98.5% of total
dissolved solids do not pass through the membrane. To calculate the % rejection,
use the following formula:
% Rejection = [(Feed TDS – Product TDS) / Feed TDS] x 100
Example:
98.5% = [(550-8.25)/550] x 100
NOTE: ALL TDS FIGURES MUST BE EXPRESSED IN THE SAME UNITS,
TYPICALLY PARTS PER MILLION (PPM) OR MILLIGRAMS PER LITER (MG/L).
L1 - Series Reverse Osmosis Systems are designed to reject up to 98.5% NaCl,
unless computer projections have been provided or stated otherwise.
The amount of permeate water recovered for use is expressed as a percentage. To
calculate % recovery, use the following formula:
% Recovery = (Product Water Flow Rate / Feed Water Flow Rate) x 100
Example:
28% = (0.14/0.50) x 100
NOTE: ALL FLOW RATES MUST BE EXPRESSED IN THE SAME UNITS,
TYPICALLY GALLONS PER MINUTE (GPM).

L1-Series User’s Manual
MKTF-258 8 9/14
SYSTEM INSTALLATION AND START-UP PROCEDURES
1. Inspect the system for any damage that could have occurred during shipment. Although
our systems have been individually inspected, complete a quick inspection of the fittings,
tubing and other components.
2. Please provide a reasonable amount of space for installation and leave 6 inches of space
below the filter housings for ease of maintenance.
NOTE: THE REVERSE OSMOSIS SYSTEM SHOULD BE INSTALLED
INDOORS AND IT IS SUGGESTED THAT IT NOT BE IN DIRECT SUNLIGHT OR
EXTREME COLD.
3. Connect the 3/8” tube fitting to an incoming water source. The minimum water pressure
should be at least 30 psi. The system’s minimum operating pressure is 80 PSI, but the
optimum operating pressure is 100 psi.
NOTE: DO NOT OPERATE AT A PRESSURE EXCEEDING 125 PSI. The
operating pressure can be increased on the face of the booster pump by turning the hex
screw clockwise.
4. Connect the concentrate 3/8” tubing (waste) line to drain.
5. Plug the booster pump transformer into a power supply of 110 volts 60 Hz.
6. This system has been designed with an auto-flush restrictor. This restrictor automatically
flushes the reverse osmosis system for 30 seconds every time it starts up and once every
hour when the system is producing water.
NOTE: THE TANK PRESSURE SWITCH WILL SHUT THE SYSTEM OFF
AUTOMATICALLY WHEN THE BLADDER TANK IS FULL.

L1-Series User’s Manual
MKTF-258 9 9/14
7. The sediment filter and carbon must be serviced regularly for optimal performance. The
filters and water quality should be checked every two weeks minimum.
8. Dispose of the product water until the conductivity of the product water reaches your
desired level. Use any TDS or Conductivity meter to monitor the product water quality. A
minimum quality of 96% NaCl rejection is recommended.
NOTE: ANY CHLORINE EXPOSURE WILL DAMAGE THE MEMBRANE
PERMANENTLY.
9. This system has been factory wired and preset with a pressure switch at 20 - 40 psi,
which is only to be used with a pressurized bladder tank. If using an atmospheric storage
tank, a float switch will be required to turn the system on and off.

L1-Series User’s Manual
MKTF-258 10 9/14
MEMBRANE ELEMENTS
L1 - Series Reverse Osmosis Systems equipped with Thin Film Composite (TFC)
HF4 Extra Low Energy membranes, unless otherwise specified. General membrane
element performance characteristics are listed below.
HF4 – Extra Low Energy Membranes

L1-Series User’s Manual
MKTF-258 11 9/14
SYSTEM IDENTIFICATION

L1-Series User’s Manual
MKTF-258 12 9/14
MEMBRANE INSTALLATION, REMOVAL AND REPLACEMENT
Installation and replacing membranes in the pressure vessels is an easy process if
you have the proper information and tools at hand. Please refer to the following
instructions when removing and replacing membrane elements:
WARNING: ALL PRESSURE GAUGES MUST READ ZERO BEFORE
PROCEEDING. BEFORE ATTEMPTING, DISCONNECT THE POWER FROM THE
SYSTEM AND BLEED ALL WATER PRESSURE FROM THE SYSTEM.
1. Remove the end cap from the right (feed water) side of the membrane
housings. This is done by removing the two half-moon retaining disks using a
#5 Allen wrench, the end plug should then freely slide out of the pressure
vessel.
2. Remove the membrane bag containing the membrane element from the
shipping box. The membrane should be contained within a plastic oxygen
barrier bag.
NOTE: WEAR GLOVES FOR THE FOLLOWING STEPS IN ORDER NOT TO
CONTAMINATE THE MEMBRANE.
3. Cut the bag open as close as possible to the seal at one end of the bag, so the
bag may be re-used if necessary.
4. Make sure that all parts are clean and free from dirt. Examine the brine seal,
and permeate tube for nicks or cuts. Replace the O-rings or brine seal if
damaged.
5. Flow directions should be observed for installation of the membrane element
into the pressure vessels.
REPLACING THE MEMBRANE ELEMENT:
WARNING: THE BRINE SEAL MUST BE IN THE SAME POSITION AS IT WAS
FOR FACTORY PRE-INSTALLED MEMBRANE ELEMENT. THE BRINE SEAL IS A

L1-Series User’s Manual
MKTF-258
RUBBER SEAL THAT PROTRUDES ON ONE SIDE OF THE MEMBRANE AND IS
ALWAYS ON THE FEED SIDE OF THE MEMBRANE ELEMENT.
SYSTEMS THE
BRINE SEAL SHOULD BE ON THE RIGHT
HOUSING.
1.
Remove membrane element from the
housing. Long nose pliers may be necessary to pull the old membrane
element ou
t of the membrane element housing.
2.
Lubricate the brine seal
such as Dow Corning® 111.
3.
Install membrane through the right side (feed water) of the membrane housing
with brine seal located on the feed water side.
4.
With a smooth and constant motion, push the membrane element into the
housing so the brine seal enters the housing without coming out of the brine
seal groove.
5. Re-install the
end plug
end cap,
while pushing it onto the housing. Ensure that you do not pinch or
fatigue any o-
rings while re
the outer diameter of the plug is flush with the outer diameter
vessel.
6. Insert the two half-
moon retaining disks until they are
Subsequently fasten using a #5 Allen wrench
7.
Reconnect any fittings that may have been disconnected when the membrane
pressure vessels were disassembled.
8. To Start-
Up the system, please refer to the Initial Start
manual.
CAUTION: WET MEMBRANES ARE SHIPPED IN A PRESERVATIVE
SOLUTION. THE MEMBRANES MUST BE FLUSHED FOR AT LEAST 30
MINUTES TO REMOVE THE PRESERVATIVE FROM THE MEMBRANE.
DI
SCARD ALL OF THE PERMEATE, WHICH IS PRODUCED DURING THE
FLUSH PERIOD.
13
RUBBER SEAL THAT PROTRUDES ON ONE SIDE OF THE MEMBRANE AND IS
ALWAYS ON THE FEED SIDE OF THE MEMBRANE ELEMENT.
FOR
L1
BRINE SEAL SHOULD BE ON THE RIGHT
SIDE OF THE MEMBRANE
Remove membrane element from the
right (feed water) side of
the membrane
housing. Long nose pliers may be necessary to pull the old membrane
t of the membrane element housing.
Lubricate the brine seal
and o-rings with a non-
petroleum based lubricant,
such as Dow Corning® 111.
Do not use a petroleum-
based lubricant.
Install membrane through the right side (feed water) of the membrane housing
with brine seal located on the feed water side.
With a smooth and constant motion, push the membrane element into the
housing so the brine seal enters the housing without coming out of the brine
end plug
(6, Figure 1.A) on the right side
by gently twisting the
while pushing it onto the housing. Ensure that you do not pinch or
rings while re
-
installing the end plug. Push the end plug on until
the outer diameter of the plug is flush with the outer diameter
of the pressure
moon retaining disks until they are
fully seated.
Subsequently fasten using a #5 Allen wrench
.
Reconnect any fittings that may have been disconnected when the membrane
pressure vessels were disassembled.
Up the system, please refer to the Initial Start
-
Up section of this
CAUTION: WET MEMBRANES ARE SHIPPED IN A PRESERVATIVE
SOLUTION. THE MEMBRANES MUST BE FLUSHED FOR AT LEAST 30
MINUTES TO REMOVE THE PRESERVATIVE FROM THE MEMBRANE.
SCARD ALL OF THE PERMEATE, WHICH IS PRODUCED DURING THE
9/14
RUBBER SEAL THAT PROTRUDES ON ONE SIDE OF THE MEMBRANE AND IS
L1
SERIES RO
SIDE OF THE MEMBRANE
the membrane
housing. Long nose pliers may be necessary to pull the old membrane
petroleum based lubricant,
based lubricant.
Install membrane through the right side (feed water) of the membrane housing
With a smooth and constant motion, push the membrane element into the
housing so the brine seal enters the housing without coming out of the brine
by gently twisting the
while pushing it onto the housing. Ensure that you do not pinch or
installing the end plug. Push the end plug on until
of the pressure
fully seated.
Reconnect any fittings that may have been disconnected when the membrane
Up section of this
CAUTION: WET MEMBRANES ARE SHIPPED IN A PRESERVATIVE
SOLUTION. THE MEMBRANES MUST BE FLUSHED FOR AT LEAST 30
MINUTES TO REMOVE THE PRESERVATIVE FROM THE MEMBRANE.
SCARD ALL OF THE PERMEATE, WHICH IS PRODUCED DURING THE

L1-Series User’s Manual
MKTF-258 14 9/14
OPERATION DO’S AND DON’TS
DO:
•Change the cartridge filters regularly
•Monitor the system and keep a daily log
•Run the system as much as possible on a continuous basis
•Always feed the pump with filtered water
DON’T
•Permit chlorine to enter membrane housing with the feed water
•Shut down the system for extended periods
•Operate the system with insufficient feed flow
•Operate the pump dry

L1-Series User’s Manual
MKTF-258 15 9/14
SPECIFICATIONS

L1-Series User’s Manual
MKTF-258 16 9/14
OPERATION AND MAINTENANCE
The reverse osmosis process causes the concentration of impurities. The impurities
may precipitate (come out of solution) when their concentration reaches saturation
levels.
NOTE: PRECIPITATION CAN SCALE OR FOUL MEMBRANES AND MUST
BE PREVENTED.
1. Periodically observe the quality and quantity of product water from the system.
NOTE: CHECK THE FEED WATER PRESSURE GOING INTO THE REVERSE
OSMOSIS MEMBRANE, A SIGNIFICANT DROP IN PRESSURE COULD INDICATE A
FOULED PRE-FILTER.
2. A 20% increase in TDS when checking the permeate water indicates possible membrane
damage, and the membrane may need to be replaced.
3. It is suggested that a hand held TDS digital meter is used once per week to monitor the
water quality.
NOTE: IF THE TDS OF THE FEED WATER EXCEEDS 1000 PPM, A LARGER
FLOW RESTRICTOR SHOULD BE USED TO EXTEND THE MEMBRANE LIFE.
4. It is important to maintain and/or replace the carbon block regularly since the Thin Film
Composite membranes are chlorine sensitive. Irreversible damage will occur with any
chlorine present in the feed water. For additional information, please review the
manufacturer’s membrane specification sheets.
5. The product line has a one way check valve installed. The check valve should be
checked regularly and replaced if it is not properly sealing.
6. Keep the feed water temperature above 4°C (36°F).
NOTE: EXTREMELY COLD FEED WATER WILL LOWER THE PRODUCT
WATER OUTPUT AND INCREASE PUMP PRESSURE.

L1-Series User’s Manual
MKTF-258 17 9/14
PREPARING UNIT FOR STORAGE OR SHIPMENT
Prior to shipping or storing your system, the system should be cleaned
with an appropriate cleaner, flushed with water and protected from biological
attack with an appropriate solution for membrane elements. The membrane
housing(s) and plumbing lines of the system must be completely drained. Any
water remaining in the plumbing of a system may freeze, causing serious
damage.
Preparing system for storage:
1. Totally fill membrane elements in the membrane housing with 2 % M-100
solution, venting the air outside of the pressure vessels. Use the overflow
technique: circulate the M-100 solution in such a way that the remaining air in
the system is minimized after the recirculation is completed.
2. Check the pH once a week. When the pH becomes 3 or lower, change the
preservation solution.
3. Repeat this process at least once a month.
During the shutdown period, the plant must be kept frost-free, or the temperature
must not exceed 113°F (45°C).
Preparing unit for shipment:
4. Disconnect the inlet, concentrate, pre-filter, and permeate plumbing.
5. Drain all water from the pre-filter cartridge housings by unscrewing the
housings, removing the pre-filter cartridges, and drain the water from the
housings.
6. Disconnect the tubing from the connectors on the permeate and concentrate
inlets and outlets.
Allow the system to drain for a minimum of eight hours or until the opened ports quit
dripping.
After draining is complete, reconnect all of the plumbing.

L1-Series User’s Manual
MKTF-258 18 9/14
REVERSE OSMOSIS TROUBLESHOOTING
SYMPTOMS POSSIBLE CAUSES CORRECTIVE ACTION
Low Inlet Pressure
Low supply pressure Increase inlet pressure
Cartridge filters plugged Change filters
Solenoid valve malfunction Replace sol. valve and/or coil
Motor may not be drawing correct current Use clamp-on amp meter to check the
motor amp draw.
Leaks Fix any visible leaks
Low Permeate Flow
Cold feed water See temperature correction sheet
Low operating pressure See low inlet pressure
Defective membrane brine seal Inspect and/or replace brine seal
Fouled or scaled membrane Clean membranes
High permeate flow
Damaged product tube o-rings Inspect and/or replace
Damaged or oxidized membrane Replace membrane
Exceeding maximum feed water
temperature See temperature correction sheet
Poor permeate quality
Low operating pressure See low inlet pressure
Damage product tube o-rings Inspect and/or replace
Damaged or oxidized membrane Replace membrane
Membrane fouling
Metal Oxide Fouling Improve pretreatment to remove metals.
Clean with acid cleaners.
Colloidal Fouling
Optimize pretreatment for colloid
removal. Clean with high pH anionic
cleaners.
Scaling (CaSO4, CaSO3, BaSO4, SiO2)
Increase acid addition and antiscalant
dosage for CaVO3 and CaCO4. Reduce
recovery. Clean with acid cleaners
Biological Fouling
Shock dosage of Sodium Bi-Sulfate.
Continuous feed of Sodium Bi-Sulfate at
reduced pH.
Chlorination and de-chlorination.
Replace cartridge filters.
Organic Fouling
Activated Carbon or other pretreatment.
Clean with high pH cleaner.
Chlorine Oxidation
Check Chlorine feed equipment and de-
chlorination system.
Abrasion of membrane by Crystalline
Material
Improve pretreatment. Check all filters
for media leakage.

L1-Series User’s Manual
MKTF-258 19 9/14
ABNORMAL PERMEATE FLOW
Permeate flow should be within 20% of the rated production, after correcting the feed
water temperatures above or below 77°F. Check your permeate flow meter to
determine the permeate flow rate.
NOTE: TO DETERMINE THE TEMPERATURE CORRECTION FACTOR,
LOCATE THE TEMPERATURE CORRECTION TABLE IN THIS USER’S MANUAL
AND FOLLOW THE DIRECTIONS.

L1-Series User’s Manual
MKTF-258 20 9/14
TEMPERATURE CORRECTION FACTORS FOR MEMBRANE
Find the temperature correction factor (TCF) from the table below. Divide the rated
permeate flow at 77°F by the temperature correction factor. The result is the
permeate flow at the desired temperature. (See example on the next page)
If a system is rated to produce 5 gpm of permeate water @ 77˚ F, the same system
will produce more water at a higher temperature. It will also produce less water at a
lower temperature. Use the temperature correction table to obtain the correct flow.
This manual suits for next models
3
Table of contents
Other Axeon Water Technologies Water Filtration System manuals