axing OTX 1310-10 User manual

OTX 1310-10
Optical Transmitter
Operation Instructions

Operation Instructions | OTX 1310-10 | Optical Transmitter
22017-10-03 | © AXING AG Switzerland | Technical data are subject to change, errors and omissions excepted
Table of contents
1.
Description...........................................................................................................................................................4
1.1. Block Diagram .............................................................................................................................................4
2.
Front and rear panel ............................................................................................................................................5
2.1.
Operating elements on the front panel........................................................................................................5
2.2.
Connectors on the rear panel.......................................................................................................................6
3. Installation...........................................................................................................................................................7
3.1.
Menu Interface ............................................................................................................................................8
4. WEB Management .............................................................................................................................................10
4.1.
Parameters ................................................................................................................................................10
4.2.
Settings......................................................................................................................................................11
4.3.
Information................................................................................................................................................11
4.4.
Network configuration...............................................................................................................................12
4.5.
Change password......................................................................................................................................12
5. Troubleshooting.................................................................................................................................................13
6. Specification.......................................................................................................................................................14
6.1.
Optical characteristics................................................................................................................................14
6.2.
RF characteristics.......................................................................................................................................14
6.3.
Other characteristics ..................................................................................................................................14
6.4.
Outlines Drawing.......................................................................................................................................15

2017-10-03 | Technical improvements, changes in design, printing- and other errors expected. 11
Safety instructions
∂The core part of OTX 1310-10 is a ESDlaser.
∂Please obey the operation rule strictly and carefully when disassemble and assemble. The wrist belt and
the union suit should be also grounded.
∂All electric power and optical transmitter should be given a good grounding connection.
∂The electric power of the optical transmitter should be 90…265VAC and a regulated AC power is
suggested.
∂The RF signal must not be connected until the optical transmitter arrives at it’s normal working point. The
amplitude of the RF signal should follow the instructions in the test report of the optical transmitter. An
abnormal RF signal power will overload and damage the laser.
∂The optical transmitter should be stored in a ESD protection environment (such as a ESD protection
container) and cannot be stored with corrosive cargo. The store temperature should be within -20℃ ~
+50℃.
∂Forced cooling should be added when multiple optical transmitters are mounted on the same rack.
∂Please don’t open or repair any part of the transmitter, otherwise the producer don’t pay the fiddler.
∂Eyes may be hurt if you watch the optical outputdirectly.
∂The heat elimination hole should not be blocked. The transmitter need good airing to work properly.
∂The optical output should be covered with our dust-proof if the transmitter don’t work for a long period.
∂When failure occurs,the transmitter should be sent to the producer in time. And please don’t open or
repair any part of the transmitter, otherwise the producer don’t pay the fiddler
Direct eye exposure to laser beam may cause physical damage

Operation Instructions | OTX 1310-10 | Optical Transmitter
4
2017-10-03 | Technical improvements, changes in design, printing- and other errors expected.
1.
Description
The OTX 1310-10 is a high performance optical transmitter, based on ORTEL and AOI transmission
technology. The optical transmitter has following functions:
∂AGC/MGC mode
∂Low noise DFB laser
∂Adjustable OMI
∂SNMP (Simple Network Management Protocol) for administration and monitoring
∂
Furthermore, it has two redundant plug-in power supply modules.
OTX 1310-10 is used for long-distance optical fiber transmission of television TV signals. It uses a high-
performance DFB laser and RF power automatic processing technology, which assures excellent performance. It
can be used for large-scale HFC networks.
1.1. Block Diagram

2017-10-03 | Technical improvements, changes in design, printing- and other errors expected. 5
2.
Front and rear panel
2.1.
Operating elements on the front panel
12 3 4
1. LCD Display
∂Output power P(mW)
∂Laser BIAS (mA)
∂Laser Temperature T(℃)
∂COOL CURRENT (mA)
∂Power supply +24.4V, +5.4V, -5.4V’s Voltage detecting
∂IP address
∂Gateway
∂OMI
∂Inner temperature
2. LEDs (green means that it works fine, red meansfailed):
∂RF is for input level status
∂LASER is for laser working status
∂PWR1 is for power supply one workingstatus
∂PWR2 is for power supply two workingstatus
3. Back button, UP and DOWN Button, Enter button
4. RF TEST: Output with -20dBuV outputlevel

Operation Instructions | OTX 1310-10 | Optical Transmitter
62017-10-03 | Technical improvements, changes in design, printing- and other errors expected.
2.2.
Connectors on the rear panel
6 6
1 2
34
57 8
1. RF Input Port
2. Grounding
3. RJ45 Web Management Interface
4. RS-232
5. Output Optical Connector
6. Fans
7. Power Supply Two
8. Power Supply One

2017-10-03 | Technical improvements, changes in design, printing- and other errors expected. 7
3. Installation
Check and make sure that the equipment, enclosure and electric power have a good grounding connection.
Check and make sure that the electric power is within 90…265VAC.
Connect the optical fibre with the optical output. Make sure the connection is cleanded before use.
Connect the electric power and turn on the power switch.
The power supply LEDs are displaying the working status.
Touch the display choice button, check the index such as the output laser power, the laser’s offset current,
temperature, cooling current, etc.
If the status is normal and the RF input level matches the requirements, connect the RF signal to the RF
input.

Operation Instructions | OTX 1310-10 | Optical Transmitter
82017-10-03 | Technical improvements, changes in design, printing- and other errors expected.
RF Menu
Press
RF Gain
Mode
RF Gain
Mode
OMI
Level
OMI Level
Normal
OMI
Adjust
RF
Attenuation
OMI
Adjust
RF
Atte
nuation
OMI
Adjust
RF
Attenuation
OMI
Adjust
RF
Attenuation
OMI
Adjust
RF
Attenuation
OMI
Adjust
RF
Attenuation
3.1.
Menu Interface
Press
System Main
OTX
Press
System
Press
Optical
Press
RF Menu

2017-10-03 | Technical improvements, changes in design, printing- and other errors expected. 9
Optical Menu
Laser Power
10.5mW
Laser Bias Current
53mA
Laser Temp
25.3℃
Laser Cool Current
+100mA
Optical Wave Length
1310.00nm
System Menu
Firmware Version
V1.0
Series Number
System Temp
32℃
Power1
+24.4V
+24.4V
Power1
+5.4V
+5.4V
Power1 -5.4V
-
5.4V
DNS
192.168.1.1
Power2 +24.4V
+24.4V
GATE WAY
192.168.1.1
Power2 +5.4V
+5.4V
MASK
255.255.255.0
Power2 -5.4V
-5.4V
IP Address
192.168.1.10

Operation Instructions | OTX 1310-10 | Optical Transmitter
10 2017-10-03 | Technical improvements, changes in design, printing- and other errors expected.
4. WEB Management
Enter 192.168.1.1 into your web browser to get the following web interface.
The default password is admin
After the login, 5 items are displayed on the web interface:
∂“Parameters”
∂“Settings”
∂“Information”
∂“Network configuration”
∂“Change password”
4.1.
Parameters
Here all parameters you get from the optical transmitter with their different working status are shown.

2017-10-03 | Technical improvements, changes in design, printing- and other errors expected. 11
4.2.
Settings
The parameters, which are displayed here can be changed:
AGC and MGC mode can be selected.
RF Attenuation from 0~12dB can be set , when MGC mode is activated
OMI can be set from -5~5dB
4.3.
Information
Here general information about the model, serial number, firmware version, SNMP private OID, admin contact,
system name and location are displayed.

Operation Instructions | OTX 1310-10 | Optical Transmitter
12 2017-10-03 | Technical improvements, changes in design, printing- and other errors expected.
4.4.
Network configuration
Here the following parameters according to customer network can be set.
4.5.
Change password
Here the password can be changed. The default password is admin.

2015-12-04 | Technical improvements, changes in design, printing- and other errors expected. 13
5. Troubleshooting
General Recovery Processing
Failure Phenomenon Reason Settle Method
LED does not work no electric power or fuse
burn out Power ON or Fuse replace
LED is black Temperature maybe too
high Force Cooling
The measurement of the
LED displays the status is normal,
but there is no laser output or the
output does not match the display
output laser power is not
correct.
Ring flange or test jumper
connection is dirty.
Change the output laser power sensor
Clean the connection
Change the test jumper
Test jumper connection
mismatches.
Both, the laser power of the
optical transmitter and receiver
are normal, but there is no RF
output signal at the receiver or the
output level is too low.
No input RF signal or RF
connection is not good.
Check the RF level and reconnect the
RF input.
If you checked everything mentioned above and the failure phenomenon still exists, the transmitter should be
shut down immediately. Please contact the producer directly.

Operation Instructions | OTX 1310-10 | Optical Transmitter
14 2017-10-03 | Technical improvements, changes in design, printing- and other errors expected.
6. Specification
6.1.
Optical characteristics
Parameter Unit Value
Working Wavelength nm 1290~1330
Output Power mW 10
Optical Fiber Type -- SM
Optical Connector -- SC/APC
6.2.
RF characteristics
Parameter Unit Value
Frequency Range MHz 47…1002
RF Input Return Loss dBm ≥16
RF Input Level dBμV75~85
Flatness dB ±0.75
CNR1 dB >50
CSO1 dBC ≤-60
CTB1 dBC ≤-65
Testing condition: OIM=3.5%, optical input power @-1dBm, 98chs signal output, above 550MHz at -10dB
offset.
6.3.
Other characteristics
Parameter Unit Value
Operating Voltage V 90~260, 50Hz
Power Consumption W ≤30
Operating Temperature °C 0~50
Dimension(L*W*H)mm 484*370*44
Weight Kg <3.5

2017-10-03 | Technical improvements, changes in design, printing- and other errors expected. 15
6.4.
Outlines Drawing
Dimensions are in mm.

Operation Instructions | OTX 1310-10 | Optical Transmitter
16
2015
-
12
-
04
|
©
AXING
AG
Swizzerland
|
Technical
data
are
subject
to
change,
errors
and
omissions
excepted
Table of contents
Other axing Transmitter manuals