Aztec GEM II User manual

www.aztecmedical.co.uk

www.aztecmedical.co.uk
1
GEM II Battery Powered Ring Cutter Kit

www.aztecmedical.co.uk
2
Operational Guide
1. SelectandmountappropriatecungdiscforGEMIIringcuer.Move
‘lockingswitch’tolockposion.Insertcorrectlyloadedbaerycasseein
toGEMIIbody.
2.Applyagenerousquantyofwaterbasedgeltoringsurface.
3. Slidengerguardunderring.
4. InsertGEMIIguidearmdiscintongerguardguideblock.Restcungdisc
onringsurface.AcvatemotorallowingtheweightoftheGEMIIcung
headalonetoprovidethenecessarydownwardpressureontothecung
disc.
5. MoveGEMIIbackandforthalongguideblocksothatcung
disctraversestheenrewidthofthering.Asuddenincrease
inrotaonspeedindicatesringhasbeencutthrough.
6. Ifpaentwarnsofheatbuild-upstopcungimmediately,
cleangelfromringsurfaceandre-applyfreshgelandconnue
withcungprocess.
7. UseGEMIIringspreaderpliersorforcepstoremovering.
Foramoredetaileddemonstraonofthe
cungprocesswatchthe
GEMIIBaeryPoweredRingCuerDVD

www.aztecmedical.co.uk
3
Safety Instructions
Maintenance Instructions
READ INSTRUCTIONS BEFORE USE
Light eye protecon should be worn by operator and paent.
• NeveruseringcuerwithoutngerguardorPerspexshieldinplaceorifeithercomponentis
damaged.
• NeveruseanypowersourceotherthandomescAAsizebaeriescapableofdeliveringa
maximumvoltageof6vDC.
• Nevermodifytheringcuertorunonmainspower.
• NeverexaminearotangcungdiscexceptthroughthePerspexshield.
• Neverreusedamagedorbadlyworndiscs.
• Never start cung through a ring without rst applying a signicant quanty of water based
gel to the ring surface. The gel acts as a heat sink, traps cut ring debris and impedes
trajectory of parcles from a rotang disc.
• Neveroperatetheringcuerinanatmospherewhereasparkcouldcauseanexplosion.
• Neveruseorre-useacungdiscifeitheroneorbothplascwasherssupporngthediscare
missing.
• Alwaysengagethe“Lock”immobiliserswitchtodeacvatethemotorwhen:
a) Mounngordisconnecngacungdisc.
b) Engagingthengerguard.
c) Pausingduringcungtoexaminethecungprogress.
d) Wipingawayusedgelandreapplyingfreshgel.
e) Whenringcuerisnotbeingacvelyusedorleunaendedwhenstored.
Always:
• Examinecungdiscspriortouse.Discardanydamagedorbadlyworndiscs.
• Restthecungbladeontheringsurfacebeforeacvangthemotor.
• Warn the paent that the cung process generates heat and that the paent should warn
the operator if the heat build-up is causing discomfort. If this happens the cung process
must stop. The water based gel wiped from the ring surface and fresh cold gel reapplied
before re-starng the cung process.
• Cleanngerguardunitaeruse.
• RemovebaeryholderfromGEMbodyaeruse.
• Stop motor aer 10 minutes of connuous use and allow motor to cool down for 15 minutes
before re-starng.

www.aztecmedical.co.uk
4
Maintenance Instructions
The ring cuer may be wiped down with any decontaminaon soluon but must not be soaked
in uid or autoclaved.
DISC CARE
Removedisc(RETAINLOCKINGSCREW).Examinediscfordamageorexcessivewear-ifthisisthe
case,replacedisc.
FINGER GUARD CARE
Cleanupngerguardandguidechannelofngerguard.Examinemetalguard;makesuredisc
hasn’tcutthrough.Ifsoreplacemetalguard.
PERSPEX SHIELD CARE
WipeanyresiduefromPerspexshield.
Aer use, examinaon and cleaning, always return the GEM II Ring Cuer and its components
back to their alloed spaces in the GEM II presentaon case.

www.aztecmedical.co.uk
5
Technical Specications
Declaration of Conformity
Model A-Z 300 Series 101
MotorRe280/3DryBearingWeight:
HandPieceWeight:
PowerSource:
NoLoadSpeed:
Torque:
StallTorque:
SoundPressure:
SoundPower:
VibraonLevels:
44gms
127gms
6voltDC(4xAA)Baeries
3v8,800rpm
6v16,000rpm
3v63gcm
6v120gcm
3v126gcm
6v240gcm
61.08dB(A)
73.58dB(A)
3.2m/s2
Recycling & Disposal

www.aztecmedical.co.uk
6
Declaration of Conformity
THE DECLARATION OF CONFORMITY
Wedeclareunderoursoleresponsibilitythatthisproductisinconformitywiththefollowingstan-
dardsorstandardiseddocuments.EN55014inaccordancewithDirecves2004/108/EC,2006/42/
ECandEN60745
ChristopherMacGregor
HeadofQualityAssurance
WEEE compliant registraon no. WEEE/DG 3326 WV
Recycling & Disposal
Onlydisposeofelectrical/electronic/baeryitemsinseparatecolleconschemes,whichcaterfor
therecoveryandrecyclingofmaterialscontainedwithin.
Yourcooperaonisvitaltoensurethesuccessoftheseschemesandfortheproteconofthe
environment.

www.aztecmedical.co.uk
7
Guarantee
DearCustomer,
Intheunlikelyeventthatyourdevicedevelopsafault,pleasecontactourcustomerservice
department.
1. Theseguaranteetermscoveraddionalguaranteerightsanddonotaectyourstatutory
warrantyrights.
2. Ourguaranteecoversproblemscausedbymaterialormanufacturingdefects,andwillresultin
therepairofthesedefectsorreplacementofthedevice.
3. Thefollowingarealsoexcludedfromourguarantee:
a) Faultsduetoaccidents,customermisuse,orunauthorisedrepairs.
b) Consumablepartssuchasblades.
c) Failureduetolackofrounemaintenance.
d) Failureasaresultofnotusingtheequipmentinaccordancewiththemanualandsafety
instrucons.
e) Anyunauthorisedmodicaonstothedeviceoritscomponents.
4. Theguaranteeisvalidforaperiodof2yearsstarngfromthepurchasedateofthedevice.
Guaranteeclaimsshouldbesubmiedbeforetheendoftheguaranteeperiodwithintwo
weeksofthedefectbeingnoced.Noguaranteeclaimswillbeacceptedaertheendofthe
guaranteeperiod.Theoriginalguaranteeperiodremainsapplicabletothedeviceevenif
repairsarecarriedoutorpartsarereplaced.Insuchcases,theworkperformedorpartsed
willnotresultinanextensionoftheguaranteeperiod,andnonewguaranteewillbecome
acvefortheworkperformedorpartsed.
A-Z Tec Medical Ltd
Email:inf[email protected]o.uk
www.aztecmedical.co.uk

www.aztecmedical.co.uk
8
Guarantee
Trouble Shooting & F.A.Q’s
TROUBLE SHOOTING
Motorfailstooperate
Motorstallsduringcung
Cungdiscmakeslileprogressincung
throughring
Check baeries are loaded into the baery
cassee correctly.
Check that retaining springs in baery cas-
see are sound and holding baeries rmly
in place.
Check baeries are in a charged condion.
Check that ‘Lock’ switch on GEM II unit is
not engaged.
Operator applying excessive force during
cung process.
Baeries are depleted, change for fully
charged baeries.
Wrong disc selected for ring metal type
• UseDiamondcungdisc
(Ref.A-Z223)forhardmetalsi.e.
Iron,Steel,Planum.
• UseCarbidecungdisc
(Ref.A-Z224)forsometalsi.e.
Gold,Silver,Copper.
Disc is badly worn, change for new disc
Diamondcungdisc(Ref.A-Z223)
Tocheckifdiscisworn,runngeraround
cungdiscedge.Ifsmoothtotouchthe
diamondgritiswornaway,replacedisc.

www.aztecmedical.co.uk
9
Cungdiscmakeslileprogressincung
throughring(Connued)
Lockingscrewunwindsduringoperaon
Carbidecungdisc(Ref.A-Z224)
Ifcungdiscisworndowntoplasc
spacers,replacedisc.
Baeries have been loaded into baery
cassee incorrectly. Re-load baeries into
cassee in the correct order and make sure
the arrows on the baery cassee and the
GEM II unit are aligned.
F.A.Q’s
1. Howmanyringswillacungdisccut
through?
2. Whyisitnecessarywhencungthrough
hardmetalstomakeacuthalfway
throughtheringononesideandthen
allthewaythroughinaposiondirectly
oppositetherstcut?
A carbide disc acng on a so metal, i.e.
Gold wedding ring approximately 4mm
wide by 1mm thick will cut through 10-15
rings.
A diamond disc acng on a hard metal, i.e.
steel 4mm wide by 1mm thick would be
capable of cung through approximately 6
rings allowing for the fact that a hard metal
requires cung in 2 separate locaons on
the ring.
N.B. The cung surface of the ring must be kept as cool as pos-
sible to preserve the diamond grit on the cung edge of the disc.
This is achieved by cung for 3 or 4 seconds, pausing for a second
and repeang the process.
Rings made of hard metals when cut
through tend to close up trapping the
cung disc. This is due to the high tensile
strength in the metal. Parally cung
through a ring signicantly reduces the ten-
sion in the ring so that when the ring is cut
through on the opposing side to the rst cut
the ring won’t collapse in on itself and trap
the cung disc.

www.aztecmedical.co.uk
10
3. Whyisitnecessarytosupplytwotypes
ofcungdiscwhynotjustusea
diamonddiscforallringcung?
4. ThemotorkeepsstallingandIamcertain
thatIamfollowingthecungprocedure
outlinedintheinstruconsi.e.not
pressingdowntoohardonthecung
discasImoveitbackwardsandforwards
acrosstheringsurface.Alsothealkaline
baeriesusedwerevirtuallynew.
5. HowdoIcheckauseddiamonddiscto
ensurethatisserviceable?
6. Althoughwehaveusedwaterbased
gelaspertheinstruconswesllreceive
complaintsfromafewpaentsabout
heatbuild-upinthering.
In fact either disc could be used for all the
cung procedures. However if a carbide
disc is deployed to cut through a hard metal
i.e. Steel it will wear down rapidly and
require to be replaced several mes during
the process of cung through a single ring.
If a diamond disc is used on so metal the
metal starts to clog up the diamond cung
facets eecvely blunng the discs cung
edge.
If you are not forcing the cung process
the fault lies with the power source.
Although Alkaline baeries are an excel-
lent power source when fresh the GEM II
ring cuer does drain the power from them
quite rapidly and if one of the four baer-
ies is below strength it will have a marked
eect on the performance of the ring cuer
motor. A more consistent and reliable
power source these days are provided by
re-chargeable NiMh baeries which are
widely available and inexpensive. They can
store and deliver more power over a longer
period of me than their Alkaline equiva-
lent. Re-chargeable NiMh hold their charge
for up to 6 months and can be re-charged
at any me as they do not suer from
the memory eect shown by the old NiCd
rechargeable baeries.
Simply run a nger around the outer rim of
the disc. If 50% of the surface feels rough to
the touch the disc has just enough diamond
grit to be serviceable. If not it must be
replaced. A worn disc will cut very lile but
generate a great deal of friconal heat.
Cung generates friconal heat. It can’t
be prevented but it can be controlled to
acceptable levels.

www.aztecmedical.co.uk
11
First if using a diamond disc check condi-
ons of the disc (See Queson 6).
It is important to warn the paent that
there will be a heat build-up in the ring and
that as soon as they feel the ring geng
warm to nofy the operator. The operator
should stop immediately wipe away the
gel from the ring and re-apply fresh cold
gel before connuing the cung process.
Normally by this stage the ring will be virtu-
ally cut through so that the second phase of
cung will be short lived.
According to a research study in the brain
journal of Neurology 125 issue 3 pp. 501-
510 on ‘Pain Threshold on Normal Skin!’
The range of warm sensaon is 37.8 ±
28°C. Redness of the skin equivalent to sun
burn occurs around 44-45°C. In tests we
have carried out on rings that have water
based gel applied to the ring surface prior
to cung aer 60 seconds of connuous
cung temperature on the surface of a ring
made of brass 4mm x 1mm thick measured
38-39°C. Aer a further 60 seconds the
surface temperature had reached 44-45°C.
A sensible precauon to take would be to
cut for 60 seconds, stop, wipe away old gel
and replace with fresh cold gel and re-start
the cung process for a further 60 seconds.
As the cung process has been interrupted
the ring surface temperature would not
reach 44°C. It is unlikely a 3rd phase of cut-
ng would be required.
We have found that liquid soap from a soap
dispenser works well. It helps to lubricate
the ring area and makes inserng the metal
guard secon of the nger guard much
easier. It also collects the debris from the
cung process.
7. Whatifaringhastoberemovedfroman
unconsciouspaent?
8. Whatifwaterbasedgelisnotavailable.
Isthereanotherproductcommonly
availablethatcanbeused?

www.aztecmedical.co.uk
12
The guard is designed to be exible.
With usage, scu and gouge marks natu-
rally develop. The nger guard is safe to use
as a barrier sll exists between the ring and
the paents skin. However if the guard has
been cut through it must be replaced.
No. This indicates that dried gel and metal
debris have formed a coang on the walls
of the groove in the guide block and need
removing by cleaning thoroughly.
This is a precauonary requirement. During
use heat builds up in the motor and over
me can reach a temperature where the
insulaon on the motor windings starts to
break down. This can cause a short circuit
to develop resulng in the heang up of
the retaining springs in the baery cassee,
which at a certain temperature will collapse
and cut o the power to the motor.
A noisy motor does not automacally
indicate that the motor needs replacing. A
drop of lubricant on the drive sha bearing
at the front of the motor and another on
the bearing at the back of the motor should
help to reduce noise.
9. Isitacceptabletobendthemetalnger
guardtogetitunderaring?
10.Inocedthatonthemetalseconofthe
ngerguardthatisinsertedunder
theringthatanumberofscumarks
andgougeshaveappeared.Isitsafeto
usethengerguardinthiscondion?
11.Inocedthattheguidediscdoesn’trun
assmoothlyalongthegrooveinthe
ngerguardsblackplascblockasitdid
whennew.Doesthisindicatethatthe
guardneedsreplacing?
12.FromthesafetynoceIseethataer10
minutesofconnuoususagethering
cuerhastobeletocooldownfor15
minutesbeforeitcanbestartedup
again.Whyisthisnecessary?
13.Wehavenocedthatourringcuer
hasbecomerathernoisycompared
towhenitwasnew.Doesthismeanthe
ringcuermotoriswearingoutand
needsreplacing?Isthereanythingwe
candotoreducethenoise?
Table of contents
Popular Medical Equipment manuals by other brands

HoMedics
HoMedics CMS50D1A Instructions for use

AccuVein
AccuVein AV400 user manual

Symmetrikit Service
Symmetrikit Service Buddy Roamer Service and repair manual

Libelium
Libelium MySignals user manual

Stryker
Stryker SPR Plus Operation manual

Otto Bock
Otto Bock Genium Protector 4X880 Series Instructions for use