Badger Basket 3000+ User manual

SCOPE OF MANUAL
This manual contains information concerning the
installation, operation and maintenance of the Model
3000+ Micro-Chart Recorder. To ensure proper
performance of the meter, the instructions given in this
manual should be thoroughly understood and followed.
Keep the manual in a readily accessible location for
future reference.
Changes and additions to the original edition of this
manual will be covered by a "CHANGE NOTICE"
supplied with the manual. The change notice will explain
any changes made to the product described in this
manual.
MODEL 3000+ TABLE OF CONTENTS
Page Page
OPERATION
Unpacking and Inspection................................. 1 General Description................................................ 7
General Specifications ...................................... 2 Status Mode............................................................ 8
Calibration Mode.................................................... 9
INSTALLATION PROCEDURE Error Screens.......................................................... 15
Electronics Hexadecimal Conversions...................................... 16
Enclosure Mounting.......................................... 3 Illustrated Parts List ............................................... 18
Wiring Connections.......................................... 4 Troubleshooting Diagram ...................................... 20
UNPACKING
&
INSPECTION
To avoid damage in transit, Eastech Badger products
are shipped to the customer in special shipping
containers. Upon receipt of the product, perform the
following unpacking and inspection procedures:
NOTE: If damage to the shipping container is evident
upon receipt, request the carrier to be present when the
product is unpacked.
a. Carefully open the shipping container following
any instructions that may be marked on the box. Remove
all cushioning material surrounding the product and
carefully lift the product from the container.
Retain the container and all packing material for possible
use in reshipment or storage.
b. Visually inspect the product and applicable
accessories for any physical damage such as scratches,
loose or broken parts, or any other sign of damage that
may have occurred during shipment.
NOTE: If damage is found, request an inspection by the
carrier's agent within 48 hours of delivery and file a claim
with the carrier. A claim for equipment damage in transit is
the sole responsibility of the customer.
941072
5
1

2
GENERAL SPECIFICATIONS
INPUTS 3 Selectable Analog Channels: 0-20 mADC, 4-20 mADC or 1-5 VDC
single input or summation.
Pulse: 0-150 ppm to 0-1000 pps, field adjustable.
4-20 mADC signal into max. 800 ohms
OUTPUTS 2 dry contact SPDT relays assignable to any of the following:
High or low alarms, contact integrator, loss of input signal, EEPROM
failure. Rated for 0.25 amp @ 120 VAC or 0.5 amp @ 24 VDC.
2 line, 24 alphanumeric character per line Liquid Crystal Display.
DISPLAY Displays instantaneous flow rate (or pressure, or temperature), totalization,
alarm indication and is used to program the recorder through menu driven
prompts.
RECORDING 12 inch circular chart with 24 hour, 7 day or 31 day selectable chart speed,
single disposable red ink pen.
PROGRAMMING Menu driven display prompt with 5 key tactile feedback keypad.
ENCLOSURE Indoor: NEMA 12, metallic; Outdoor: NEMA 4X, non-metallic.
TEMPERATURE Indoor: 32° F to 150° F (0° C to 65° C)
LIMITS Outdoor: -40° F to 150° F (-40° C to 65° C)
POWER Standard: 117 VAC, 50/60 Hz +/- 10%
REQUIREMENTS Optional: 230 VAC, 50/60 Hz +/- 10%
Recording Indicating Totalization
Accuracy: +/- 0.5% of span +/- 1% or 1 LSD +/- 0.25%
PERFORMANCE Linearity: +/- 0.5% of span +/- 0.5%
Hysteresis: +/- 0.5% of span
Repeatability: +/- 0.5% of span

INSTALLATION
ENCLOSURE DESCRIPTION
The Model 3000+ Micro-Chart Recorder is housed in
either a NEMA 12 or NEMA 4X enclosure.
The NEMA 12 enclosure is suitable for indoor use and
designed to be panel mounted. It can be supplied with
heater and thermostat for indoor installations in buildings
that are not heated in the winter. Mounting hardware is
provided for panel mounting of the enclosure. Care
should be taken when selecting the mounting location so
that the recorder door can be opened completely for
access when changing charts.
INDOOR ENCLOSURE DIMENSIONS
15-5/8"
19-3/16"
8-13/16"
7 "
The NEMA 4X enclosure is suitable for outdoor use
and can be wall or pole mounted. It comes equipped with
heater and thermostat. The door can be supplied with or
without a window. The door is provided with two
lockable latches. Caution should be used when mounting
the enclosure so that it is not exposed to direct sunlight. A
sun shield is recommended to prevent the temperature inside
the enclosure reaching above the maximum rating of the
electronics (see temperature specification under General
Specifications). Viewing window is optional.
OUTDOOR ENCLOSURE DIMENSIONS
16"
20"
7-7/8"
INDOOR PANEL MOUNTING
Prior to mounting the enclosure, it is recommended that
the knockouts for the conduit fittings be removed. This
should be done with the enclosure resting on a solid
surface.
A hole 17-7/8" high by 14-1/4" wide is required for
inserting the enclosure in a panel. Slide the recorder into
the panel. Install the four mounting brackets into the slots
in the sides of the enclosure as shown below. Tighten the
mounting screws until the enclosure is secure to the panel.
14-1/4 "
17-7/8 " PANEL
OPENING
CUTOUT
MOUNTING
BRACKETS
4 EACH
3

OUTDOOR ENCLOSURE MOUNTING
The NEMA 4X outdoor enclosure is provided with four
mounting feet which are to be installed by the user at the
orientation desired. Pictured below is the installation
detail of the feet and the mounting hole dimensions.
If the outdoor enclosure has the window option, do not
mount facing the sun or install a sun shade to prevent
overheatingoftheelectronicsinsidetheenclosure.
WIRING CONNECTIONS
+4-20 MA
OUTPUT
NO
C SW 1
NC
NO
C SW 2
AN1
AN2
AN3
NOT
USED
+
FREQ IN
PWR
PWR
NC
1
2
3
4
5
6
7
8
1
2
3
4
5
6
1
2
3
117/230VAC
FACTORY
WIRED
TB1
TB2
TB3
TB4
TB5
1
2
3
+
+
+
FUSE 1/4 AMP
HNG
117 / 230 VAC
CONNECTIONS
MOUNTING FEET
14.55 "
17.03 "
10.40 "
20.96 "
4

The indoor case has six 7/8" diameter knockouts for
3/4" conduit fittings. Three are located on the bottom of
the case and three on the back of the case. The three you
use will depend on whether you panel or wall mount the
enclosure.
To reduce the chance of noise on the signal input wires,
the input wires should be routed in separate conduit from
the power wiring.
Local regulations for proper grounding and connection
of the power to the recorder should be strictly followed.
The power wires are connected to the terminals located
on the inside back of the recorder. The AC voltage which
the recorder has been setup will be clearly marked near
the terminal strip. Verify that you are connecting the proper
voltage to the recorder. The terminal labeled H is the High
(black) wire, the terminal labeled N is the Neutral (white)
wire and the terminal labeled G is the Ground (green)
wire. The power wires from the terminal strip to the
circuit board are prewired at the factory.
The terminal blocks on the circuit board have snap on
connectors. To connect the input and output wires, pull
off the connectors from the terminal strips. Route the
signal wires through the flexible wiring raceway between
the back case and front door. Loosen the screws on the
connectors. Strip off about 1/4 inch of insulation from the
wires. Insert the wire into the proper connector and
tighten the screw securely. After connecting all of the
wires to the connector, snap the connector back on the
terminal strip.
The following section describes the proper connections
for the various transmitters that may be used with this
recorder and the connections for the relay outputs and the
4-20 mADC output.
There are three analog input channels and one
frequency input channel. The analog inputs will be setup
for a standard 4-20 mADC input signal unless otherwise
stated on the Data Sheet supplied with the manual. Three
different analog inputs can be connected to the three
analog input channels, but only the selected analog
channel can be recorded on the recorder.
ANALOG TRANSMITTERS
TRANSMITTER
1
2
3
4
5
6
AN1
AN2
AN3
TB4
The above wiring diagram shows the wiring
connections for current or voltage input signals. The Data
Sheet furnished in the front of this manual will indicate
the type of analog input for which your recorder was
setup. If this is different from the output of your
transmitter, contact the factory for corrective action.
TWO WIRE TRANSMITTERS
1
2
3
4
5
6
AN1
AN2
AN3
TRANSMITTER
+
-
1
2
3
FREQ IN
PWR
PWR
TB2
TB4
The above wiring diagram is for transmitters that
require DC voltage to power the 4-20 mADC loop. The
voltage at TB2 terminals 2 and 3 is 40 VDC.
5

TRANSMITTER
PULSE TRANSMITTER
TB2 2
1
3
FREQ IN
PWR
PWR
Accepts a frequency input of 0-150 pulses per minute
to 1-1000 pulses per second. Programmed at the factory,
but field changeable.
NOTE: Because of the high speed counting capability
of the 3000+ recorder, some electro-mechanical contacts
used as a pulse (frequency input can cause "contact
bounce" which the recorder will measure as pulses. If this
occurs, install a .47 uf capacitor (Badger Part # 130739-
0002) between the two frequency input screw terminals.
MAX. 1000
OHM LOAD
+
FREQ IN
PWR
PWR
TB2 1
2
3
+4-20 MA
OUTPUT
NO
C SW 1
NC
NO
C SW 2
NC
1
2
3
4
5
6
7
8
TB5
+
4-20 mADC OUTPUT
The two-wire 4-20 mADC will be a characterized
output. For example, if the 4-20 input is from a DP Cell
and the 3000+ recorder is setup to extract the square root
from the raw signal, the 4-20 output will be linear to flow
not differential. DC power must be connected in series
with the load, as shown above.
DRY CONTACT
RELAYS MAX
LOAD 250 MA
+4-20 MA
OUTPUT
NO
C SW 1
NC
NO
C SW 2
NC
1
2
3
4
5
6
7
8
TB5
RELAY OUTPUTS
Relays are assignable to different functions. See
Operating Section of the Manual.
INSTALLATION OF CHART AND PEN
ALIGNMENT
Open the recorder door and unscrew the know from the
chart hub. Lift the pen arm with the pen lifter off the chart
plate. Remove the protective cap from the fiber tipped
pen. Lower the pen close to the chart plate and adjust the
pen until it aligns with the reference scribe on the chart
plate.
Lift the pen back up with the pen lifter. Place a chart on
the chart hut and screw the knob onto the hub loosely.
Slide the edge of the chart under the five plastic holders.
Rotate the chart until the proper time line is aligned with
the pen. Tighten the knob to the hub assembly. Remove
the backlash by turning the chart knob counterclockwise.
Lower the pen arm until the pen is touching the chart.
PEN ARM
REFERENCE SCRIBE
CHART PLATE
6

OPERATION
GENERAL DESCRIPTION
The Model 3000+ Micro-Chart Recorder is a circular
chart recorder with microprocessor-based electronics. The
unit is supplied with a LCD display for indicating rate
and totalization. The rate is displayed with 4 digits and its
multiplier and unit of measurement. The totalizer is
displayed with 8 digits and its multiplier and unit of
measurement.
7
The display is also used to program the recorder
through menu driven prompts. The programming is
activated via front panel keys. These keys also activate a
Status Mode which will indicate if there are any alarms
tripped and perform a self-test operation of the recorder.
The following will describe in detail the various
functions of the Model 3000+ Recorder and programming
steps to recalibrate the recorder in the field if necessary.
INITIAL POWER UP
After installation and wiring of the Model 3000+
Recorder, power may be turned on by the power switch
on the lower right hand corner of the chart plate.
When power is applied, the display screen will have a
solid line across the top of the display for a few seconds.
Then the following screen will appear:
3000+ RECORDER
BADGER METER INC
This screen informs the user of the Model of the
recorder. This screen will be displayed for a few seconds.
The next screen to be displayed is the following:
SOFTWARE REVISION X.XX
SERIAL NUMBER 12345678
This screen identifies the software revision the recorder
is operating with and the serial number of the unit. This
screen will be displayed for a few seconds.
The next screen shown is the screen that will be
displayed during normal operation of the recorder.
*FLOW 1234 X 10,000 GPM
12345678 X 100,000 GAL
The top line displays the flow rate with its multiplier
and the unit of measurement. The second line displays the
totalized flow (4 or 8 digits) with its multiplier and unit of
measurement.
The asterisk that is to the left of FLOW may or may not
be present. If it is present, it indicates that an alarm has
been tripped. The various alarms will be discussed in the
Status Mode Section of this manual.
The recorder should be fully operational if there is an
input signal from a transmitter. The recorder is
programmed at the factory with the information supplied
by the user and the Zero and Span have been calibrated.
If the flashing asterisk is present proceed to the Status
Mode Section to determine the cause and proper action.
If the user desires to reset the Zero and Span or change
any of the parameters of the recorder, proceed to the
Calibration Section of this manual.

STATUS MODE
The Status Mode contains the Alarms Tripped screen
and the Self Test screen. To enter into the Status Mode
press the MENU key and the following screen will
appear:
PRESS UP FOR CALIBRATION
PRESS DOWN FOR STATUS
Pressing the DOWN arrow key will allow entry into the
Status Mode screens. If you decide not to enter the Status
Mode, press the MENU key and you will return to the
normal operation screen.
Press the DOWN arrow key and the following screen
appears:
- ALARMS TRIPPED -
IN LO HI CHT INT EE
The Alarms Tripped screen will display any conditions
that have occurred.
To clear the alarms press the DOWN arrow key.
Should an alarm still be tripped, it will still be shown on
the display.
The definition of the alarms are as follows:
IN = There is no 4-20 mA signal or the input current
is below the pen arm zero current value.
LO = This indicates that the low alarm set point has
been tripped. This set point may be changed, refer to the
Calibration Section of this manual.
HI = This indicates that the high alarm set point has
been tripped. This set point may be changed, refer to the
Calibration Section of this manual.
CHT = This indicates that the chart has made a
complete rotation and the chart needs to be changed.
INT = Indicates a contact integrator pulse output has
occurred.
EE = EEPROM read/write error which indicates a
failure in the microprocessor. Consult the factory.
To return to the normal operating screen press the
MENU key.
To enter the Self Test screen press the ENTER key and
the following screen will appear:
- SELF TEST -
PRESS UP TO ACTIVATE
If you decide not to enter into the Self Test section,
press the MENU key and you will return to the normal
display screen.
To enter into the Self Test section press the UP arrow
key and the following screen will appear:
- SELF TEST -
EEPROM TEST=> PASSED
This tests the EEPROM and indicates if it passed or
failed. This takes a few seconds and then the following
screen will appear:
- SELF TEST -
VREF TEST=> PASSED
This tests the reference voltage and indicates if it
passed or failed. The following screen will appear:
- SELF TEST -
SERVO TEST=> PASSED
This tests the servo drive system and indicates if it
passed or failed.
Should any of the screens indicate a failure, contact the
factory for the appropriate action to be taken.
Upon completion of the Self Test the unit will return to
the normal operating display screen. This completes the
Status Mode Section.
8

CALIBRATION MODE
The Calibration Mode allows the user to reprogram the
3000+ Recorder to different parameters, calibrate the
Zero and Span and assign and set the relay outputs.
To enter the Calibration Mode, press the MENU key
and the following display screen will appear:
PRESS UP FOR CALIBRATION
PRESS DOWN FOR STATUS
9
If you decide not to enter the Calibration Mode, press
the MENU key and the display will return to the normal
operating screen.
When in the Calibration Mode and the desired change
has been made and no other changes are desired, you may
skip to the end of the calibration screens by pressing the
MENU key.
To enter the Calibration Mode, press the UP arrow key
and the following screen will appear:
SECURITY ID
INPUT 4 DIGIT ID 0000
This screen is the security identification which requires
a security code to be entered before access is allowed into
the Calibration Mode. This code is set to 0000 at the
factory. You may set a new code when in the Calibration
screen.
To enter a security code use the RIGHT arrow key to
move the cursor under the digit to be changed and use the
UP or DOWN arrow keys to change the digit to the
desired value. The values can be from 0 to 9 and A to F.
After entering the security code, press the ENTER key.
If the code is incorrect the display will return to the
normal operating screen. If the code is correct, the
following screen will appear:
MEASUREMENTS UNITS
UNIT INITIALS=GPM
This screen allows the user to select the engineering
unit of measure for the flow rate. To change the units,
press the UP or DOWN arrow keys until the desired unit
is shown. The engineering units available are:
GPM = Gallons per Minute
GPD = Gallons per Day
MGD = Million Gallons per Day
CFS = Cubic Feet per Second
CFM = Cubic Feet per Minute
CFH = Cubic Feet per Hour
CMM = cubic Meters per Minute
OPTION = Any three letter label can be entered
After you have selected the unit desired, press the
ENTER key. If you have selected OPTION, skip to the
section entitled OPTION on Page 14.
The next screen, after selecting a standard
measurement, allows the selection of the multiplier for the
unit of measurement and will appear as follows:
UNIT MULTIPLIER
GPM X 1,000
Press the UP or DOWN arrow key to select the desired
flow multiplier. The following multipliers are available:
X.001
X.01
X.1
X1
X10
X100
X1,000
X10,000
Press the ENTER key to store the multiplier. The
following screen will appear:
FULLSCALE UNITS=GPM
FULLSCALE X 1,000
This screen allows the setting of the full scale range.
Press the UP or DOWN arrow keys to change each digit
to the desired value and the RIGHT arrow key to move
the cursor under each digit. Press the ENTER key and the
following screen will appear:
TOTALIZER MULTIPLIER
X 1,000
This screen allows the selection of the totalizer
multiplier. Press the UP or DOWN arrow keys to select
the desired multiplier and then the ENTER key. The
following multipliers are available:

X.01 X100
X.1 X1000
X1 X10,000
X10 X100,000
The following screen will appear:
CHART REVOLUTION SPEED
1 REVOLUTION IN 1 DAY
This screen allows the selection of the chart rotation
speed. The following speeds are available:
1 REVOLUTION IN 2 MINUTES
1 REVOLUTION IN 24 MINUTES
1 REVOLUTION IN 1 HOUR
1 REVOLUTION IN 1 DAY
1 REVOLUTION IN 7 DAYS
1 REVOLUTION IN 31 DAYS
- - - - -
- - -
The two dash line positions are for future chart
revolution speeds. Press the UP or DOWN arrow keys to
select the desired chart speed and press the ENTER key.
The following screen will appear:
RECORDER DAMPING
DAMPING TIME= 6 SECS.
This screen allows the selection of the amount of
damping in the response of the pen movement. The
damping times available are:
NONE 6
16 40
100 240
600
These times are in seconds. Press the UP or DOWN
arrow keys to select the desired value and then the
ENTER key. The following screen will appear:
- FUNCTION SELECTION -
POWER=> LINEAR
This screen allows the selection of linearization
functions for use with differential pressure transmitters,
open channel flow measurement devices or any definable
function. Check the output specification of the device that
is being connected to the recorder to determine the proper
power selection. The functions available are:
POWER=>LINEAR
POWER=>1.55
POWER=>1.58
POWER=>1.95
POWER=>1/2
POWER=>3/2
POWER=>5/2
POWER=>OPTION
The linear function is used where the transmitter output
is proportional to flow.
The 1.55 power is used with 1", 2", 3", 24" Parshall
flumes and all Lagco flumes.
The 1.58 power is used with a 6" Parshall flume.
The 1.95 power is used with the Badger Meter Manhole
flumes.
The 1/2 power is used with differential pressure
transmitters where the output does not have the square
root extracted.
The 3/2 power is used with Suppressed Rectangular
weirs and Cipoletti weirs.
The 5/2 power is used with all V-Notch weirs.
The OPTION power allows a 16 point HQ curve to be
programmed into the recorder. Consult the factory for
programming details.
Press the UP or DOWN arrow keys to select the desired
function and then the ENTER key. The following screen
will appear:
INPUT SELECTION
CHANNEL=ANALOG #1
This screen allows the selection of the desired input
channels to be used as the input to the recorder. There are
four channels available to connect the input(s) signal(s):
ANALOG #1, ANALOG #2, ANALOG #3, AND
FREQUENCY.
The ANALOG channels can accept 0-20 mA, 4-20 mA,
or 1-5 VDC. The recorder must be set up at the factory
for either a current or a voltage input. These channels can
be selected individually or summed in any combination.
The frequency channel is used for pulsed inputs. The
operating range is 0-1000 pps to 0-150 ppm. Press the UP
or DOWN arrow keys to select the desired input channel
and then the ENTER key.
10

The following describes the programming screen
sequence if the selection is: ANALOG 1, 2 or 3; FRE-
QUENCY; or SUMMATION.
ANALOG INPUT 1, 2 OR 3 IS SELECTED
After selecting one of the ANALOG channels with the
UP or DOWN arrow keys and the ENTER key is pressed,
the following screen will appear:
CHART PEN ZERO
11
This screen calibrates the pen zero point on the chart
paper to the zero of the input signal. It is important that the
pen be properly aligned before making this adjustment. The
zero value of the input signal must be input to the recorder
when making this adjustment. For example, if the input
signal is 4-20 mA for 0 to 100%, then 4 mA must be input to
the recorder to make the pen zero movement.
Press the UP or DOWN arrow keys to move the pen to
the proper position. No changes will be made to the
previously stored zero value unless the UP or DOWN
arrow keys are pressed. So if the user is stepping through
the program to access another programming screen, the
input does not need to be at the zero value if the UP or
DOWN arrow keys are not pressed. When pressing the
UP or DOWN arrow keys, the display will indicate the
direction of the pen movement. Once the pen is
positioned to zero on the chart, press the ENTER key. Be
sure to press the ENTER key before changing the input signal.
The following screen will appear:
-CHART PEN SPAN-
UP OR DOWN TO MOVE PEN
This screen calibrates the pen span point on the chart
paper to the span of the input signal. The span value of
the input signal must be input to the recorder when
making this adjustment. For example, if the input signal is
4-20 mA for 0 to 100%, then 20 mA must be input to the
recorder to make the pen span adjustment.
Press the UP or DOWN arrow keys to move the open
to the proper position. No changes will be made to the
previously stored span value unless the UP or DOWN
arrow keys are pressed. So if the user is stepping through
the program to access another programming screen, the
input does not need to be at the span value if the UP or
DOWN arrow keys are not pressed. When pressing the
UP or DOWN arrow keys, the display will indicate the
direction of the pen movement. Once the pen is
positioned to span on the chart, press the ENTER key. Be
sure
press the ENTER key before changing the input signal. The
next screen to appear will be the Relay Assignment screen.
FREQUENCY INPUT IS SELECTED
After selecting the frequency input channel with the UP
or DOWN arrow keys and pressing the ENTER key, the
following screen will appear:
FREQUENCY INPUT MODE
PPS PPM
This screen selects the frequency rate of Pulses Per
Minute or Pulses Per Second. Press the RIGHT arrow key to
move the cursor under the desired frequency rate and then
depress the ENTER key. The following screen will appear:
FREQUENCY ZERO
0000 PULSES PER SECOND
This screen allows the setting of the frequency which
represents zero. Press the UP or DOWN arrow keys to
change the value of each digit and the RIGHT arrow key to
move the cursor under the digit to be changed. Press the
ENTER key to store this value. The following screen will
appear:
FREQUENCY SPAN
0000 PULSES PER SECOND
This screen allows the setting of frequency which
represents span. Press the UP or DOWN arrow keys to
change the value of each digit and the RIGHT arrow key
to move the cursor under the digit to be changed. Press
the ENTER key to store this value. The following screen
will appear:
- CHART PEN ZERO -
UP OR DOWN TO MOVE PEN
This screen calibrates the pen zero point on the chart
paper to the zero of the input signal.It is important that the
pen be properly aligned and the 4 mA is being applied to the
input before making this adjustment. Press the UP or
DOWN arrow keys to move the pen to the proper
position. No changes will be made to the previously
stored zero value unless the UP or DOWN arrow keys are
pressed. So the user CAN step through the program to
access another programming screen without readjusting
the zero, if the UP or DOWN arrow keys are not pressed.
When pressing the UP or DOWN arrow keys the display
will indicate the direction of the pen movement. Once the
pen is positioned to zero on the chart, press the ENTER

key. The following screen will appear:
- CHART PEN SPAN -
UP OR DOWN TO MOVE PEN
This screen calibrates the pen span point on the chart
paper to the span of the input signal. Change the input
current to 20 mA before continuing. Press the UP or DOWN
arrow keys to move the pen to the proper position. No
changes will be made to the previously stored span value
unless the UP or DOWN arrow keys are pressed. So the user
can step through the program to access another
programming screen without readjusting the span, if the UP
or DOWN arrow keys are not pressed. When pressing the
UP or DOWN arrow keys the display will indicate the
direction of the pen movement. Once the pen is positioned
to span on the chart, press the ENTER key. The next screen
to appear will be the Relay Assignment screen.
SUMMATION INPUT IS SELECTED
The 3000+ Recorder is capable of summing up to three
analog inputs. In the Input Selection screen there is the
following channel combinations that can be selected: #2 +
#2, #1 + #3 and #1 + #2 + #3. Use the UP and DOWN
arrow keys to select the channels to be summed and press
the ENTER key. The following screen will appear:
CHANNEL #01 IS 3
12
0 %
RECORDER FULL SCALE
This screen sets the percent of full scale that the
channel indicated on the screen represents. For example,
if the full scale of the recorder was 10,000 GPM and the
full scale of the transmitter, then the percent for channel
one would be set for 30%. This is to be done for each
channel that is to be summed. After all of the percents
have been set for all of the channels, the sum should total
100%.
Use the UP or DOWN arrow key to adjust the percent
of full scale value. Press the ENTER key to store this
value. After the last channel has been entered the
following screen will appear:
- CHART PEN ZERO -
UP OR DOWN TO MOVE PEN
This screen calibrates the pen zero point on the chart
paper to the zero of the input signal. It is important that the
pen be properly aligned before making this adjustment. The
zero value of the input signals of all the channels to be summed
must be input to the recorder when making this adjustment.
For example, if the input signal is 4-20 mA for 0 to
100%, then 4 mA must be input to all channels to be
summed to make the pen zero adjustment.
Press the UP or DOWN arrow keys to move the pen to
the proper position. No changes will be made to the
previously stored zero value unless the UP or DOWN
arrow keys are pressed. So if the user is stepping through
the program to access another programming screen, the
input does not need to be at the zero value if the UP or
DOWN arrow keys are not pressed. When pressing the
UP or DOWN arrow keys the display will indicate the
direction of the pen movement. Once the pen is
positioned to zero on the chart, press the ENTER key. Be
sure to press the ENTER key before changing the input signal.
The following screen will appear:
- CHART PEN SPAN -
UP OR DOWN TO MOVE PEN
This screen calibrates the pen span point on the chart
paper to the span of the input signal. The span value of the
input signal for all channels to be summed must be input to the
recorder when making this adjustment. For example, if the
input signal is 4-20 mA for 0 to 100%, then 20 mA must
be input to all the channels to be summed, to make the
pen span adjustment.
Press the UP or DOWN arrow keys to move the pen to
the proper position. No changes will be made to the
previously stored span value unless the UP or DOWN
arrow keys are pressed. So if the user is stepping through
the program to access another programming screen, the
input does not need to be at the span value if the UP or
DOWN arrow keys are not pressed. When pressing the
UP or DOWN arrow key the display will indicate the
direction of the pen movement. Once the pen is
positioned to span on the chart, press the ENTER key. Be
sure to press the ENTER key before changing the input signal.
The next screen to appear will be the Relay Assignment
screen.
- RELAY ASSIGNMENT -
RELAY 01 => LO
This screen allows the assignment of the relays for any
of the alarm functions. The available alarm functions are:
IN - Loss of input signal
LO - Low alarm trip
HI - High alarm trip
CHT - Completion of one chart rotation
INT - Contact integrator pulse
EE - Failure of EEPROM
- - -
- - -

There are two blank alarms for future use which are
represented by the dashed lines. Press the UP or DOWN
arrow keys to select the desired alarm and then press the
ENTER key. The second relay assignment will appear
with the same alarm functions to be selected as above.
After assigning the desired alarm the following screen
will appear:
CONTACT INTEGRATOR
0
13
000 X 1,000 GAL
This screen allows the setting of the contact integrator to
the desired amount of totalized flow for each pulse output.
For example, if the desire is to have a contact closure every
10,000 gallons, then the number to be entered is 0010.
Press the UP or DOWN arrow keys to change the value
of the digit and the RIGHT arrow key to move the cursor
under the digit to be changed. Press the ENTER key. The
following screen will appear:
INTEGRATOR PULSE WIDTH
ON FOR 0000 X .0328 SEC.
This screen allows the setting of the pulse width of the
contact integrator output pulse. If the device being
connected to the contact integrator output requires an
input with a pulse width of 50 milliseconds, then the
multiplier would be set to 2. Use the RIGHT arrow key to
move the cursor under the digit to be changed and the UP
or DOWN arrow keys to set the desired value and press
the ENTER key. The following screen will appear:
SET POINT ADJUSTMENT
HI ALARM = 99 PERCENT
This screen allows the setting of the High and Low alarm
set points. Press the UP or DOWN arrow keys to the desired
value and then press the ENTER key. The LO alarm screen
set point will appear. Follow the same procedure above to
set the Low alarm. The following screen will appear:
SETPOINT ADJUSTMENT
HYSTERESIS = 03 PERCENT
This screen sets the hysteresis for the High and Low
alarm set points. If the High alarm is set for 60 percent
and the hysteresis is 3 percent, then the High alarm will
trip closed at 63 percent and open at 57 percent.
Press the UP or DOWN arrow keys to the desired value
and press the ENTER key. The following screen will
appear:
PRESS UP FOR 4-20
CURRENT CALIBRATION
This screen allows the entry into the calibration of the
retransmitted 4-20 mA output. Press the UP arrow key to
enter into the current calibration. The following screen
will appear:
4-20 CURRENT OUT
ZERO WORD=3000
This screen allows the setting for the Zero output of the
4-20 mA current output circuit for 4 mA. Connect a current
meter to the output terminals and monitor the current output.
The Zero word is a four digit word with the digit to the far
left being the coarse adjustment and digits moving to the
right being the finer adjustments. Press the RIGHT arrow
key to move the cursor under the desired digit for
adjustment and press the UP or DOWN arrow keys to adjust
the output to the proper value. Press the ENTER key to store
the new value. The following screen will appear:
4-20 CURRENT OUT
SPAN WORD=A235
This screen allows the setting of the Span output of the 4-
20 mA current output circuit for 20 mA. Connect a current
meter to the output terminals and monitor the current
output. The Span word is a four digit word with the digit to
the far left being coarse adjustment and digits moving to the
right being the finer adjustments. Press the RIGHT arrow
key to move the cursor under the desired digit for
adjustment and press the UP or DOWN arrow keys to
adjust the output to the proper value. Press the ENTER key
to store the new value. The following screen will appear:
SECURITY ID
PRESS UP TO CHANGE
This screen allows the entry into the security code
screen which allows the entry of a new code. Press the
UP arrow key and the following screen will appear:
SECURITY ID
INPUT 4 DIGIT ID 0000

To enter a new security code use the RIGHT arrow key
to move the cursor under the digit to be changed and use
the UP or DOWN arrow keys to change the digit to the
desired value. The values can be from 0 to 9 and A to F.
After entering the security code, press the ENTER key
and the following screen will appear:
-STORING PARAMETERS-
PARAMETERS>PROGRAMMED
This screen indicates that the new parameters have been
stored in the recorder. The display will then return to the
normal operating screen. This completes the Calibration
Section.
OPTION (FOR FLOW UNITS SCREEN)
Selecting OPTION in the Flow Units screen allows the
entering of any three letter label for special engineering
units. Press the ENTER key with OPTION selected and
the following screen will appear:
MEASUREMENT UNITS
OPTION INITIALS=AAA
There will be a cursor under the first letter. Use the UP
or DOWN arrow keys to select the desired letter. Press
the RIGHT arrow key to move the cursor to the second
and third letter and select the desired letter. Press the
ENTER key to enter the OPTION INITIALS. The
following screen will appear:
DISPLAY FLOW LABEL AND
TOTALIZER ? YES
This screen asks if the FLOW label is to be displayed
on the normal operating screen. If the recorder is to be
used to record temperature or some other measurement
not related to flow or requires totalization, then selecting
NO will remove the FLOW label and the totalizer value
from the screen. Use the UP or DOWN arrow keys to
select YES or NO and press the ENTER key. The
following screen will appear:
UNIT MULTIPLIER
AAA X 1,000
This screen allows the selection of the desired flow rate
multiplier. Press the UP or DOWN arrow keys to select
the desired multiplier and then press the ENTER key. The
ollowing screen will appear:f
FULLSCALE UNITS=AAA
FULLSCALE = 0000 X 1,000
This screen allows the entering of the full scale flow
rate desired. Press the UP or DOWN arrow keys to select
the desired value and the RIGHT arrow key to move the
cursor under the digit to be changed. After the desired
value is selected, press the ENTER key and the following
screen will appear:
OFFSET UNITS=AAA
OFFSET=0000 X 1,000
This screen allows an offset value to be used instead of
0 for a 4 mA input. If the input device has a value other
than 0 for 4 mA, then that value would be entered into
this screen. Press the UP or DOWN arrow keys to select
the desired value and the RIGHT arrow key to move the
cursor under the digit to be changed. After the desired
value is selected press the ENTER key.
TOTALIZER UNIT
UNIT INITIALS=AAA
This screen allows the entering of the desired totalizer
unit label. Press the UP or DOWN arrow keys to select
the desired letter and the RIGHT arrow key to move the
cursor under the letter to be changed. After the desired
letters are selected press the ENTER key. The following
screen will appear:
TOTALIZER MULTIPLIER
X 1,000
This screen allows the selection of the desired totalizer
multiplier. Press the UP or DOWN arrow keys to select
the desired multiplier and then depress the ENTER key.
The following screen will appear:
14

-TOTALIZER WORD-
0
15
000 0000 0000
The Totalizer Word is required to setup the time base to
properly totalizer in the selected option units. The
totalizer word is defined as follows:
HEX2,000,000x
ecin Units/SFlowMax. MultiplierTotalizer =
EXAMPLE:
Maximum flow = 1800 x 1 BPM (Barrels/min)
Totalizer multiplier = BAR x 100
Totalizer Word =
[100/(1800/60)] x 2,000,000 = 6666666.6
Converted to Hexadecimal = 65B9AA (see the Decimal
to Hexadecimal Conversion Section).
The TOTALIZER WORD is 0000 0065 B9AA
NOTE: The totalizer word must be greater than Hex
1000.
Use the UP and DOWN arrow keys to change each
digit to the required value and the RIGHT arrow to move
the cursor. The ENTER key must be depressed after each
set of four digits. Return to Page 10 where the next screen
will be the Chart Speed selection.
ERROR SCREENS
There are several Error Screens that will be displayed
by the recorder if the microprocessor detects errors in the
values programmed into the recorder.
-CHART PEN ZERO-
<INPUT ZERO ERROR>
OR
-CHART PEN SPAN-
<INPUT SPAN ERROR>
If either of the two above Error Screens appear during
the chart pen adjusting procedure, one of the following
problems probably occurred:
1. There was not a 4-20 mADC analog signal to the
recorder analog channel you selected while adjusting the
chart zero span.
2. An analog input signal channel was selected
without adjusting the chart pen zero and span of this
channel.
3. The Zero and Span of the recorder have been
adjusted too close to the same input value.
To correct these errors, readjust the Chart Pen Zero and
Span while inputting the 4-20 mADC analog signal to the
proper selected recorder channel. Refer to the Calibration
Mode Section of this manual.
-STORING PARAMETERS-
<TOTALIZER WORD ERROR>
If the above Error Screen appears after pressing the
ENTER key to store the calibration adjustments, this
indicates that you have selected a totalizer multiplier that
is too low for your selected flow rate. To correct this
error, re-enter the Calibration Mode and when the
Totalizer Multiplier Screen appears, increase the totalizer
multiplier and press the ENTER key. Then press the
MENU key to store the new multiplier and return to the
normal operating screen.
If you still have a Totalizer Word Error message in the
Storing Parameters screen, repeat the above procedure
until the error is not present.

16
HEXADECIMAL SYSTEM
The purpose of the hexadecimal number system is
simply to reduce the number of units necessary for
representing any given numerical figure, thereby reducing
the amount of space in memory necessary to retain it.
Herein, a decimal (normal) number is converted to a
correct hexadecimal number which consists of fewer
characters.
The hexadecimal number system consists of only 16
characters which are shown below by the boldface
characters.
0 = 0
1 = 1
2 = 2
3 = 3
4 = 4
5 = 5
6 = 6
7 = 7
8 = 8
9 = 9
10 = A
11 = B
12 = C
13 = D
14 = E
15 = F
These characters may be arranged in various sequences to produce an infinite number of representations of decimal
numbers. For example: 6D4C = 27,980.
CONVERSION OF DECIMAL WHOLE NUMBER TO HEXADECIMAL NUMBER
1. Divide the decimal number by 16.
Example: 57420/16 = 3588.75
2. Multiply only the fractional part of the product by 16
to arrive at the first character in the hexadecimal
equivalent. Remember, all numbers produced I this step
are shown above. Also note that the product may have no
fraction (.000) which would result in zero as the
hexadecimal number.
Example: .75 x 16 = 12 = C
3. Divide the whole number portion of the product by 16
thereby producing yet another number.
Example: 3588/16 = 224.25
4. Repeat steps 2 and 3 in a cyclical fashion until the
numerator to be divided in Step 3 is less than 16. At that
point, the numerator represents the final character in the
hexadecimal sequence.
EXAMPLE: Convert 57,420 into a hexadecimal number:
1.) 57,420/16 = 3588.75
2.) .75 X 16 = 12 = C
3.) 3,588/16 = 224.25
2.) .25 X 16 = 4
3.) 224/16 = 14.00
2.) .00 X 16 = 0
3.) 14 = E
57,420 = E04C

17
CONVERSION OF A WHOLE HEXADECIMAL NUMBER TO A DECIMAL NUMBER
1. Multiply the left most character in the hexadecimal number by 16.
Example: E= 14 X 16 = 224
2. Add to the product previously found the value of the next character.
Example: 224 + 0 = 224
3. Multiply the previously found product by 16.
Example: 224 X 16 = 3584
4. Repeat steps 2 and 3 in a cyclical fashion until you have added the last
hexadecimal character. Do not multiply beyond that point.
EXAMPLE: Convert E04C into a decimal number.
1.) E= 14; 14 X 16 = 224
2.) 224 + 0 = 224
3.) 224 X 16 = 3584
2.) 3584 + 4 = 3588
3.) 3588 X 16 = 57,408
4.) 57,408 + 12 = 57,420
E04C = 57,420

18
INSERT DWG 600662-9999 3000+ RECORDER ASSY DWG

19
CHART PLATE ASSEMBLY PARTS LIST
ITEM PART NUMBER DESCRIPTION U/M QTY
12 543744-0001 SERVO MECH ASSY EA 1.0
13 400001-0088 HEX NUT 2-56 X 7/16 EA 4.0
14 430002-0106 FLAT WASHER EA 4.0
15 501401-0001 SPACER #2 EA 2.0
16 151960-0001 PCB MOTHER BOARD EA 1.0
17 510031-0131 SPRING EA 1.0
18 543075-0001 LCD DISPLAY EA 1.0
19 500050-0043 SPACER EA 4.0
20 410001-0007 HEX NUT 6-32 EA 13.0
22 512645 HINGE EA 1.0
23 512594 CHART PLATE EA 1.0
24 501078 MEMBRANE SWITCH EA 1.0
27 540769 CHAIN ASSY EA 1.0
29 400002-0015 SCREW 6-32 X 5/16 EA 11.0
30 400005-0017 SCREW 2-56 X 3/16 EA 2.0
31 501132-0005 CONNECTOR HOUSING EA 1.0
32 501425-0002 STEPPER MOTOR EA 1.0
33 500050-0015 SPACER #6 EA 2.0
34 400006-0025 SET SCREW 4-40 X 3/16 EA 1.0
35 524720 CHART CLAMP BODY EA 1.0
36 400006-0071 SET SCREW 5-40 X 5/16 EA 1.0
37 512610 COVER W/LOGO EA 1.0
38 524743 CHART KNOB EA 1.0
39 400002-0010 SCREW 6-32 X 5/8 EA 2.0
40 512114 PEN ARM EA 1.0
41 500638-0003 ON/OFF SWITCH EA 1.0
42 430002-0105 FLAT WASHER EA 5.0
43 500983-0001 SNAP FASTENER EA 5.0
44 512224 PEN REPLACEMENT DECAL EA 1.0
45 500724-0001 RED FIBER TIP PEN EA 2.0
46 440005 RETAINER EA 2.0
47 501276-0002 SCREW EA 2.0
48 500146-0005 LOCTITE EA 0.01
49 501239 BELT CLAMP EA 1.0
50 525733 PEN LIFTER EA 1.0
52 430003-0002 WASHER EA 11.0
79 500014 ON/OFF NAMEPLATE EA 1.0
CASE ASSEMBLY PART LIST
51 512599 CASE EA 1.0
52 430003-0002 WASHER EA 1.0
53 500023-0037 TERMINAL BLOCK EA 1.0
54 512633 MTG BRACKET EA 4.0
55 400013-0064 MTG SCREW 5/16-18 X 4.5 EA 4.0
56 501275-0001 CLIP EA 2.0

TROUBLESHOOTING DIAGRAM
YES
NO Display
lit & ok? NO CONTACT
FACTORY
Turn power
off, then
back on
Ok
Now?
Turn power off,
unplug display &
plug it back in.
Turn power back on.
Ok
Now?
NO
YES
YES
Reprogram
for proper flow
& totalizer units
Did unit
program
ok?
Try programming
again. If unit will
not accept changes,
contact factory
Check connector & label
coming from motor on
servo board to recorder
circuit board. Make sure
all connector pins are
properly mated on correct
motor output.
Also check for mechanical
binding of belt, motor and
pen. Make sure that pen
arm transport tie down is
removed from unit.
Increase
damping.
See IOM
page 10.
Ok
Now?
CONTACT
FACTORY
Will
pen arm
move with
input now?
CONTACT
FACTORY
YES
GO TO NEXT PAGE
Do
pen arm &
display agree
with input?
Supply known input
value & compare with
display & pen position.
Recapture pen
arm zero & span.
See IOM pg 11. Does the
display / pen
jump or jerk?
START
ARE THE PEN
ARM, FLOW
DISPLAY &
TOTALIZER
OK?
NO
Display
show legible
characters?
Flow &
total
units
ok?
Pen arm
moves?
Do
pen arm &
display agree
with input?
YES
YES
YES
YES
NO NO
NO
NO
YES
NO NO
YES
YES
Recapture pen arm
zero & pen arm span.
See IOM pg 11.
Will
pen arm
move with
input?
YES
YES
NO
YES
NO
NO
NO
YES
NO
TO NEXT PAGE
Check 117 VAC &
connections, power
switch, & fuse
Display
lit & ok?
20
Table of contents
Other Badger Basket Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Marconi Instruments
Marconi Instruments TF 2502 instruction manual

Haltian
Haltian Thingsee BEAM User and installation guide

wtw
wtw inoLab Oxi Level 2 operating manual

CORNING
CORNING ONE MxU Quick installation sheet

Satel
Satel K-1 2E quick start guide

Agilent Technologies
Agilent Technologies Agilent 4396B user guide