7.1 Troubleshooting and notes
PA-X1-M-E-1 A2
8 Storage and transportation
Possible causeFault
There is
no water
flowing
out after
starting
the device.
Air pressurization
doesn't complete.
Solutions
Please wait for
one minute.
Refer to Picture 1.
Water pipe and
air pipe connect
incorrectly.
Pump doesn't
work.
Don't connect the
power or the
power switch is
off.
Pump keeps
working.
6.1 Please make sure all pipes are connected
accurately and tightly before operating.
6.2 Please cut off the power source when inject
or change water.
6.3 Do not exceed the maximum water volume
of the bottle when inject water.
6.4 When the model is working, please don't pull
the pipes forcibly in order to avoid damaging the
air pipe and the water pipe.
6.5 Because there is air pressure in the bottle,
please pull out the air pipe and the water pipe
before open the bottle cap, and then inject water.
6.6 Please screw tightly the bottle cap after change
water, and then turn on the power switch.
6.7 Please do not pull out the water pipe after
completing pressurization in order to avoid water
spraying from the bottle and cause secondary risk.
6.8 Please use the special configuration power
supply of our company.
6.9 Repair of the equipment should be carried out
by the manufacturer or by the authorized dealer.
6 Precautions
7 Maintenance
The bottle cap isn't
screwed tightly.
Please screw tightly
the bottle cap.
The suction pipe
isn't connected or
it isn't connected
tightly.
Please reconnect
the suction pipe.
Please connect the
power or turn on the
power switch.
There is air
leakage at the air
pipe junction.
Please reconnect
the air pipe.
If the problem still can't be solved, please contact with
local dealer or manufacturer.
8.1 Environmental conditions of storage and
transportation:
a) Relative humidity: 0 to 80%
b) Atmospheric pressure: 50kPa to 106kPa
c) Environmental temperature: -10℃ to +50℃
8.2 Prevent excessive shock and vibration in
transportation, be sure to handle with care and
avoid inversion.
8.3 Don't mix with dangerous goods during
transportation.
8.4 Avoid the sun, rain or snow during
transportation.
8.5 The equipment should be handled carefully
and lightly. Be sure that it's far from the vibration,
and install or store in a cool, dry and ventilated
place.
8.6 Don't store the machine with the articles
that are combustible, poisonous, caustic
and explosive.
9 After-service
9.1 We offer 12 months free repair to the
equipment based on warranty card from the
date when it is sold to the end user.
9.2 The repair of the equipment should be
carried out by our professional technician,
irretrievable damage caused by the
nonprofessional technician, and damage
accidentally or deliberately caused by operators,
are out of the range of warranty.
10 Environmental protections
10.1 You can deal with it based on the local law.
10.2 We reserve the right to change the design
of the equipment, product technique, accessories,
instruction manual and the content of original
packing at any time without notice. If there are
some differences between picture and real
equipment, take the real equipment as the norm.
Using the wrong
power supply.
Please check
whether you have
used the special
configuration
power supply of
our company.