manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Bard
  6. •
  7. Medical Equipment
  8. •
  9. Bard Duet User manual

Bard Duet User manual

CARE AND MAINTENANCE
The care and maintenance of the catheter requires well trained, skilled personnel following a detailed protocol. The protocol should include a
directive that the catheter is not to be used for any purpose other than the prescribed therapy.
ACCESSING CATHETER, CAP CHANGES, DRESSING CHANGES3
• Experiencedpersonnel
• Useaseptictechnique
 •Properhandhygiene
 •Cleanglovestoaccesscatheterandremovedressingandsterileglovesfordressingchanges
 •Surgicalmask(1forthepatientand1forthehealthcareprofessional)
• Catheterexitsiteshouldbeexaminedforsignsofinfectionanddressingsshouldbechangedateachdialysistreatment.
• CatheterLuer-lockConnectorswithendcapsattachedshouldbesoakedfor3to5minutesinpovidoneiodineandthenallowedtodry
before separation.
• Carefullyremovethedressingandinspecttheexitsiteforinammation,swellingandtenderness.Notifyphysicianimmediatelyifsignsof
infection are present.
EXIT SITE CLEANING9
• Useaseptictechnique(asoutlinedabove)
• Cleantheexitsiteateachdialysistreatmentwithchlorhexidinegluconateunlesscontraindicated.Applyantisepticpermanufacturer’s
recommendations.Allowtoairdrycompletely.
• Covertheexitsitewithsterile,transparent,semipermeabledressingorperhospitalprotocol.
RECOMMENDED CLEANING SOLUTIONS
Catheter Luer-lock Connectors/End Caps:
•Povidoneiodine(allowconnectors/endcapstosoakfor3to5
minutes)3
WARNING:Alcoholshouldnotbeusedtolock,soakordeclotpolyure-
thaneDialysisCathetersbecausealcoholisknowntodegradepoly-
urethanecathetersovertimewithrepeatedandprolongedexposure.
Hand cleaner solutions are not intended to be used for disinfecting
ourdialysiscatheterLuer-lockConnectors.
Exit Site:
•Chlorhexidinegluconate2%solution(preferred)3,7,8,9,10
•Chlorhexidinegluconate4%solution
•Diluteaqueoussodiumhypochlorite
•0.55%sodiumhypochloritesolution
•Povidoneiodine
•Hydrogenperoxide
•Chlorhexidinepatches
•Bacitracinzincointmentsinpetrolatumbases
WARNING:AcetoneandPEG-containingointmentscancausefailure
of this device and should not be used with polyurethane catheters.
Chlorhexidinepatchesorbacitracinzincointments(e.g.Polysporin*
ointment)arethepreferredalternative.
POST DIALYSIS
Useaseptictechnique(asoutlinedabove).
1. Flusharterialandvenouslumenswithaminimumof10mLofsterilesaline.
WARNING: Toavoiddamagetovesselsandviscus,infusionpressuresmustnotexceed25psi(172kPa).Theuseofa10mLorlargersyringeis
recommended because smaller syringes generate more pressure than larger syringes.
2. Injectheparinsolutionintoboththearterialandvenouslumensofthecatheter.Theappropriateheparinsolutionconcentrationandushing
frequencyshouldbebasedonhospitalprotocol.Heparinsolutionof1,000to5,000units/mLhasbeenfoundtobeeectiveformaintaining
thepatencyofhemodialysisandapheresiscatheters.Wheninjectingheparinsolution,injectquicklyandclampextensionwhileunderposi-
tivepressure.Heparinsolutionvolumetolockeachlumenmustbeequaltotheprimingvolumeofeachlumen.Primingvolumesaremarked
on each lumen.
3. CleancatheterLuer-lockconnectorsperhospitalprotocol.Attachsterileendcapstoboththearterialandthevenousclampingextension
pieces.
WARNING: Topreventsystemicheparinizationofthepatient,theheparinsolutionmustbeaspiratedoutofbothlumensimmediatelyprior
tousingthecatheter.Inmostinstances,nofurtherheparinsolutioninjectionisnecessaryfor48-72hours,providedthecatheterhasnot
beenaspiratedorushed.
CATHETER REMOVAL
Evaluatethecatheterroutinelyandpromptlyremoveanynonessentialcatheter9perphysician’sorders.Afterremovingthecatheter,apply
manualpressuretothepuncturesitefor10-15minutesuntilnosignsofbleedingarepresent.Thenapplysterile,transparent,semipermeable
dressingordressingperhospitalprotocolforaminimumof8hours.Followhospitalprotocolregardingbedrestaftercatheterremoval.
DISPOSAL
Afteruse,thecathetersandaccessoriesmaybeapotentialbiohazard.Handleanddisposeofinaccordancewithacceptedmedical
practice and all applicable laws and regulations.
TROUBLESHOOTING
PATIENT WITH FEVER
Patientwithfeverandchillsfollowingtheproceduremaybeindicativeofcatheter-relatedsepsis.Ifthisresults,removalofthecathetermaybe
indicated.
INSUFFICIENT FLOW
Excessiveforcemustnotbeusedtoushanobstructedlumen.Insufficientbloodowmaybecausedbyanoccludedtipresultingfromaclot
orbycontactingthewallofthevein.Ifmanipulationofthecatheterorreversingarterialandvenouslinesdoesnothelp,thenthephysicianmay
attempttodissolvetheclotwithathrombolyticagent(e.g.TPA,Catho*Activase*thrombolytic).Physiciandiscretionadvised.
CATHETER EXCHANGE
Donotroutinelyreplacedialysiscatheterstopreventcatheter-relatedinfections9.Itmaybecomenecessarytoexchangetheindwellingcatheter
duetoapersistentriseinpressuresordecreaseofowrateswhichcannotberectiedthroughtroubleshooting.Catheterexchangesshould
be performed under strict aseptic conditions in which the physician should wear a cap, mask, sterile gown, sterile gloves, and use a large sterile
drape to cover the patient.
REFERENCES
1. Aitken,D.R.andMinton,J.P.“ThePinch-OSign:AWarningofImpendingProblemswithPermanentSubclavianCatheters”,AmericanJournal
ofSurgery,Vol.148,Nov.1984,pp.633-638.
2. Mickley,V.,“Centralvenouscatheters:manyquestions:fewanswers”,NephrolDialTransplant,(2002)17:1368-1373.
3. NationalKidneyFoundationK/DOQIGUIDELINES2006.
4. Sulek,CA.,Blas,ML.,Lobato,EB,“Arandomizedstudyofleftversusrightinternaljugularveincannulationinadults.”JClinAnesth.2000
Mar;12(2):142-5
5. Tan,P.L.,Gibson,M.,“CentralVenousCatheters:theroleofradiology”,ClinRad.2006,61:13-22
6. Octavio,Bella,Colemenares,Garcia,andFlores;“RightVersesLeftInternalVeinCatheterizationforHemodialysis:ComplicationsandImpact
onIpsilateralAccessCreation”;ArticalOrgans;2004,28(8):728-733.
7. TheInstituteforHealthcareImprovement,“How-to-Guide:PreventCentralLineInfections,”2006.
8. TheJointCommissionHospitalAccreditationOrganization,NationalPatientSafetyGoals,2009.
9. CenterforDiseaseControlandPrevention,“GuidelinesforthePreventionofIntravascularCatheter-RelatedInfections,”MorbidityandMortal-
ityWeeklyReport,Aug.9,2002,51(RR-10),1-32.
10.TheSocietyforHealthcareEpidemiologyofAmerica,“StrategiestoPreventCentralLine-AssociatedBloodstreamInfectionsinAcuteCare
Hospitals,”InfectionControlandHospitalEpidemiology,Oct.2008,29(S1):S22-S30.
Otherreferencesavailableuponrequest
MRConditional
Non-clinicaltestinghasdemonstratedthedeviceisMRConditional.Itcanbescannedsafelyunder:
• staticmagneticeldof3Teslaorless
• spatialgradienteldof360Gauss/cmorless
• maximumspecicabsorptionrate(SAR)or4W/kgfor15minutesofscanning.
• Innon-clinicaltesting,thedeviceproducedatemperatureriseoflessthan1.4°Catamaximumspecicabsorptionrate
 (SAR)of4W/kgfor15minutesofMRscanningina3TeslaSiemensTriowithsoftwareversionB15.
ForMinimalImageArtifact
• MRimagequalitymaybecompromisediftheareaofinterestisinthesameareaorrelativelyclosetothepositionoftheextensionlegs.
Therefore,itmaybenecessarytooptimizeMRimagingparametersforpresenceofthemetallicportionoftheextensionlegsormoveexten-
sion legs away from area of interest.
Anissuedorrevisiondatefortheseinstructionsisincludedfortheusersinformation.Intheeventtwoyearshaveelapsedbetweenthisdateand
product use, the user should contact Bard Access Systems, Inc. to see if additional product information is available.
Revisiondate:March2010.
*Bard, Duet and StatLockaretrademarksand/orregisteredtrademarksofC.R.Bard,Inc.
Allothertrademarksarethepropertyoftheirrespectiveowners.
©2010C. R. Bard, Inc. Allrightsreserved.
Bard Access Systems, Inc.
605North5600West
SaltLakeCity,UT84116U.S.A.
1-801-522-5000
CustomerService:800-545-0890
ClinicalInformation:800-443-3385
www.bardaccess.com
InstructionsForUse
For
Duet*
Long-TermHemodialysisCatheterSystem
Access Systems
DESCRIPTION
The Duet* Long-Term Hemodialysis Catheter System is made of Carbothane* radiopaque polyurethane. The catheter system is comprised of
twosingle10Frenchlumensthatallowforowratesashighas400mL/min.Eachcathetercomeswithanextensionlegandaretentioncufor
tissueingrowthtoanchorthecatheter.Thearterial(red)andvenous(blue)lumensaredistinguishedbythecolorcodingoftheprintingonthe
catheterandthecolor-codedLuer-lockconnector.
STERILIZATIONBYETHYLENEOXIDE.
Sterile,
Non-Pyrogenic
Unless Package is
Damaged or Opened.
Do not use if package is damaged
MRConditional  Non-pyrogenic
Sterilize
This product does
notcontainDEHP This product and packaging do not
containnaturalrubberlatex.
INDICATIONS FOR USE
The Duet*Long-termHemodialysisCatheterSystemisindicatedforuseinattainingshort-termorlong-termvascularaccessforhemodialysis,
hemoperfusion,orapheresistherapy.Thecathetersmaybeinsertedpercutaneouslyintotheinternaljugularvein,subclavian,orexternaljugular
vein.
CONTRAINDICATION
Thisdeviceiscontraindicatedforpatientsexhibitingsevere,uncontrolledthrombocytopeniaorcoagulopathy.
WARNINGS
First Rib
Subclavian Vein
Clavicle
Vertebra
Internal Jugular Vei
n
Superior Vena Cava
Sternum
Pinch-off Area
Infraclavicular Fossa
Axillary Vein
WARNING:Ifthesubclavianveinisused,percutaneousinsertionofthecatheter
shouldbemadeintotheaxillary-subclavianveinatthejunctionoftheouterand
mid-thirdsoftheclaviclelateraltothethoracicoutlet.Thecathetershouldnotbe
insertedintothesubclavianveinmediallybecausesuchplacementcanleadtocom-
pressionofthecatheterbetweentherstribandclavicleandcanleadtodamageor
fractureandembolizationofthecatheter.1Fluoroscopicorradiographicconrmation
of catheter tip placement should be helpful in demonstrating that the catheter is not
beingpinchedbytherstribandclavicle.1
•Useofthesubclavianveinforcatheterplacementmayresultin
subclavianveinstenosis.Subclavianveinstenosismaypreventfuture
useoftheipsilateralextremityforpermanentaccess.Useofthejugular
vein is preferred.
•Alcoholoralcohol-containingantiseptics(suchaschlorhexidine)maybeusedto
cleanthecatheter/skinsite;however,careshouldbetakentoavoidprolongedorexcessivecontactwiththesolutions(s).
• AcetoneandPEG-containingointmentscancausefailureofthisdeviceandshouldnotbeusedwithpolyurethanecatheters.Chlorhexidine
patchesorbacitracinzincointments(e.g.,Polysporin* ointment)arethepreferredalternative.
• FollowUniversalPrecautionswheninsertingandmaintainingthisdevice.
• Cardiacarrhythmiasmayresultiftheguidewireisallowedtopassintotherightatrium.
• Donottunnelthroughmuscle.
• Clampingthecatheterrepeatedlyinthesamespotcouldweakenthetubing.Changethepositionoftheextensionclampsregularlyto
prolongthelifeoftheextensiontubing.AvoidclampingnearorontheLuer-lockconnectors,asthismaycausetubingfatigueandpossible
disconnection.Extensiontubingmaydeveloptearsorcutsifsubjectedtoexcessivepullingorcontactwithroughedges.Examinetubingfor
damage both prior to and after each treatment.
• Cathetersshouldbeimplantedcarefullytoavoidanysharporacuteangleswhichcouldcompromisetheopeningofthecatheterlumens.
• Topreventairembolismand/orbloodloss,placethumbovertheexposedoriceofthesheathintroducer.
• Toavoiddamagetovesselsandviscus,infusionpressuresshouldnotexceed25psi(172kPa).Theuseofa10mLorlargersyringeisrecom-
mended because smaller syringes generate more pressure than larger syringes. Note:Athreepound(13.3Newton)forceontheplungerof
a3mLsyringegeneratespressureinexcessof30psi(206kPa)whereasthesamethreepound(13.3Newton)forceontheplungerofa10mL
syringegenerateslessthan15psi(103kPa)ofpressure.
• AccessoriesandcomponentsusedinconjunctionwiththiscathetershouldincorporateLuer-lockconnectors.
• Theheparinsolutionmustbeaspiratedoutofbothlumensimmediatelypriortousingthecathetertopreventsystemicheparinizationofthe
patient.
• Failuretoclampextensionswhennotinusemayleadtoairembolism,bleeding,andpossibleocclusions.
• Intherareeventofaleak,thecathetershouldbeclampedimmediately.Necessaryremedialactionmustbetakenpriortoresumingdialysis,
hemoperfusion, or apheresis procedure.
• Donotresterilizethecatheterorcomponentsbyanymethod.Themanufacturerwillnotbeliableforanydamagescausedbyreuseofthe
catheter or accessories.
• Cannulationoftheleftinternaljugularveinisreportedlyassociatedwithahigherincidenceofcomplicationscomparedtocatheterplace-
mentintherightinternaljugularvein.4
• Donotadvanceguidewireorcatheterifunusualresistanceisencountered.
• TheblueCompressionSleeveisanecessarycomponentoftheExtensionLegAssembly.AlwaysvisuallyconrmthattheCompressionSleeve
remainsintheFemaleConnectorduringassembly.
• Alcoholshouldnotbeusedtolock,soakordeclotpolyurethaneDialysisCathetersbecausealcoholisknowntodegradepolyurethane
cathetersovertimewithrepeatedandprolongedexposure.
• Reusemaycreateariskofpatientinfectionofuserinfection,compromisethestructuralintegrityofthedevice,leadtodevicefailureand/or
leadtoinjury,illnessordeathofthepatient.
• TheSlideClamp,ThumbClamp,andPlugareprovidedforuseduringcatheterplacementonly.Donotreuse.
• Notintendedforpediatricorneonataluse.
• IntendedforSingleUse.DONOTREUSE.Reuseand/orrepackagingmaycreateariskofpatientoruserinfection,compromisethestructural
integrityand/oressentialmaterialanddesigncharacteristicsofthedevice,whichmayleadtodevicefailure,and/orleadtoinjury,illnessor
death of the patient.
CAUTIONS
• Repeatedovertighteningofbloodlines,syringesandcapswillreduceLuer-lockconnectorlifeandcouldleadtopotentialconnectorfailure.
Incaseofdamage,clampthecatheterbetweenthepatientandthedamagedareawithasmooth-edged,atraumaticclamp.
• Sterileandnon-pyrogeniconlyifpackagingisnotopened,damagedorbroken.
• Readtheinstructionsforusecarefullybeforeusingthisdevice.
• Federal(USA)lawrestrictsthisdevicetosalebyorontheorderofaphysician.
• CareshouldbetakenNOTtoforcethedilatorsheathintroducerintothevesselduringinsertionasvesseldamageincludingperforationcould
result.
• Leftsidedcatheterplacementmayprovideuniquechallengesduetotherightanglesformedbytheinnominateveinandattheleftbrachio-
cephalicjunctionwiththeSVC.2,5
• Ensurethatthesheathintroducerisonlytornexternally.Cathetermayneedtobefurtherpushedintothevesselassheathintroduceristorn.
• Foroptimalproductperformance,donotinsertanyportionofthecuintothevein.
• Careshouldbetakennottoadvancethesplitsheathtoofarintovesselasapotentialkinkwouldcreateanimpassetothecatheter.
• PlacementandcareoftheDuet*Cathetersshouldbeperformedbypersonsknowledgeableoftherisksinvolvedandqualiedintheproce-
dures.
• Donotcreateasharpbendincathetertunnelasthismaycausekinkingandimpactow.
• Iftheguidewiremustbewithdrawnwhiletheneedleisinserted,removeboththeneedleandguidewireasaunittopreventtheneedlefrom
damaging or shearing the guidewire.
• Cuttingthecatheteranywherebutthepre-printedmarkswillresultintheinabilitytoreadthecatheterlengthandprimingvolumes.
• Cathetertipsmustbestaggeredby4cmtominimizerecirculation.
• BeforeattemptingtheinsertionofDuet*Catheters,ensurethatyouarefamiliarwiththefollowingcomplicationsandtheiremergencytreat-
ment should any of them occur.
POSSIBLE COMPLICATIONS
These and other complications are well documented in medical literature and should be carefully considered before placing the catheter.
Theuseofanindwellingcentralvenouscatheterprovidesanimportantmeansofvenousaccessforcriticallyillpatients;however,thepotential
existsforseriouscomplicationsincludingthefollowing:
• AirEmbolism
• Bleeding
• BrachialPlexusInjury
• CardiacArrhythmia
• CatheterorCuErosionthroughtheSkin
• CatheterEmbolism
• CatheterOcclusion
• CatheterOcclusion,DamageorBreakage
duetoCompressionbetweentheClavicle
andFirstRiborKinking
• Catheter-relatedSepsis
• CardiacTamponade
• Endocarditis
• ExitSiteInfection
• ExitSiteNecrosis
• Extravasation
• FibrinSheathFormation
• Hemothorax
• Hemotoma
• Hydrothorax
• Inammation,Necrosisorscarringofskin
over implant area
• IntoleranceReactiontoImplantedDevice
• LacerationofVesselsorViscus
• Mediastinalwidening
• PerforationofVesselsorViscus
• Pneumothorax
• RetroperitonealBleed
• RisksNormallyAssociatedwithLocaland
GeneralAnesthesia,Surgery,andPost-
OperativeRecovery
• ThoracicDuctInjury
• Thromboembolism
• VenousThrombosis
• VentricularThrombosis
• VesselErosion
0724304/1003R
F. Assembly of Catheter and Extensions
31.(Asapplicable)usingascalpelorscissors,cutcathetersquarelyatdesiredpre-printedmarktoproduceaclean,smoothsurface(Figure5).
Ensureatleast6cmofcatheterextendfromexitsitetoallowforconnection.
35 cm/2.0ml
35 cm/x.xml 40 cm/2.2ml 45 cm/2.4ml
Figure 5
CAUTION: Cuttingthecatheteranywherebutthepre-printedmarkswillresultintheinabilitytoreadthecatheterlengthandpriming
volumes.
Figure 6
Female Connector
Figure7
Blue
Compression
Sleeve
Female Connector
WARNING: TheblueCompressionSleeveisanecessarycomponentoftheExtensionLegAssem-
bly.AlwaysvisuallyconrmthattheCompressionSleeveremainsintheFemaleConnectorduring
assembly(Figure7).
32.HoldunclampedExtensionLegAssemblyandattachsterile10mLsyringeandprimewithnormalsterilesaline,leavingsyringeattached
(Figure8).WARNING: Toavoiddamagetovesselsandviscus,infusionpressuresshouldnotexceed25psi(172kPa).Theuseofa10mLor
larger syringe is recommended because smaller syringes generate more pressure than larger syringes.
Figure 8
33.With catheters cut per Step 31, slip Female
Connector over catheter visually conrming
blue Compression Sleeve remains inside
Female Connector (Figure9).
Note:Attachtheappropriatecolor-coded
(blue-venous,red-arterial)ExtensionLegassembly.
Figure 9
34.SlideproximalendofcatheterovermetalcannulaontheExtensionLegBody,ensuringthatcatheterisattachedentirelengthofmetal
cannulaandnometalisvisible(Figure10).
Figure 10
35.SlideFemaleConnectortowardsExtensionLegBodyandassemblethetwotogetheruntilthetwoarefullyseated(Figure11).
WARNING: TheblueCompressionSleeveisanecessarycomponentoftheExtensionLegAssembly.AlwaysvisuallyconrmthattheCom-
pressionSleeveremainsintheFemaleConnectorduringassembly(Figure7andFigure11).
Figure 11
Blue Compression Sleeve must be present during assembly.
36.AfterExtensionLegAssemblyisoncatheter,ensurecatheterlengthandprimingvolumeprintingisvisible.
37.Graspingtheconnectorinonehand,andthecathetertubingintheother,gentlytugonthejointtotestthesecurityoftheconnector.Ifthe
connectorpullsoutofcatheter,repeattheattachmentprocedure.Aconnectionfailuremaybeduetoone,oracombinationofthefollowing:
 •TheExtensionLegConnectormetalcannulaisnotfullyinsertedintothecatheter.
 •Missingbluecompressionsleeveinfemaleconnector.
38.Withthe10mLsyringeattachedtotheExtensionLegAssembly,removeSlideClampfromcatheter.
WARNING:TheSlideClamp,ThumbClampandPlugareprovidedforuseduringcatheterplacementonly.Donotreuse.
39.Verifycatheterfunctionbyaspiratingtoensureadequatebloodow.Notethatextensionclampsmustbeunclampedtoaspirate.
40.Onceowissatisfactory,ushcatheterwithaminimumof10mLofsterilesaline.
WARNING:Toavoiddamagetovesselsandviscus,infusionpressuresmustnotexceed25psi(172kPa).Theuseofa10mLorlargersyringeis
recommended because smaller syringes generate more pressure than larger syringes.
41.Injectheparinsolutionof1,000to5,000units/mLinamountsequaltotheprimingvolumeasdenotedonthecatheter.Injectquicklyand
clampextensionwhileunderpositivepressure.Attachasterileendcap.WARNING:Failuretoclampextensionswhennotinusemayleadto
air embolism, bleeding, and possible occlusions.
42.Repeatsteps31-41forsecondcatheter(Figure12).
Figure 12
Cu
G. Placement Verication and Securement
43.Afterinsertion,conrmplacementofthecatheterusingx-rayoruoroscopy.
44.Foradditionalsecurity,suturetheinsertionsite,orifpreferred,useaStatLock*
CatheterStabilizationDeviceSecurementdevicetoanchorthecatheter.
45.Managetheexitsiteperhospitalprotocol.
46.Dressthecatheterperhospitalprotocol.
WARNING:AcetoneandPEG-containingointmentscancausefailureofthisdevice
andshouldnotbeusedwithpolyurethanecatheters.Chlorhexidinepatchesor
bacitracinzincointments(e.g.,Polysporin* ointment)arethepreferredalterna-
tive.
47.Recordindwellingcatheterlengthandinsertionsiteonpatient’schart.
PERFORMANCE GUIDELINES
Priming Volumes
Refertoindividualcathetersforprimingvolumeinformationprintedatpre-denedcentimetermarkings.
Assuggestedby In Vitro data
Flow Rate vs Lumen Pressure
Venous
Arterial
35 cm Length (400 ml/min) 50 cm Length (350 ml/min)
199 mmHg 229 mmHg
-208 mmHg -235 mmHg
Note: Reverse ow will result in higher recirculation
INSERTION TECHNIQUE
Forpercutanenousplacement,thecatheterisinsertedthroughasheathintroducerintothesuperiorvenacavaviatheinternaljugularvein3,6
(preferred),externaljugularvein,orsubclavianvein.ThepatientshouldbeplacedinTrendelenburgpositionwiththeheadturnedtotheop-
posite side of the entry site.
CATHETERS MUST BE INSERTED UNDER STRICT ASEPTIC CONDITIONS.
WARNING:Cannulationoftheleftinternaljugularveinwasreportedlyassociatedwithahigherincidenceofcomplicationscomparedtocath-
eterplacementintherightinternaljugularvein.4
CAUTION:Leftsidedcatheterplacementmayprovideuniquechallengesduetotherightanglesformedbytheinnominateveinandattheleft
brachiocephalicjunctionwiththeSVC.2,5
A. Sterile Field and Skin Preparation
1. Provideasterileeldthroughouttheprocedure.Usesterilegloves,masks,caps,sterilegowns,andusealargesteriledrapetocoverthe
patient.Ifhairremovalisneededuseclippersordepilatories.
2. Preparetheaccesssiteusingstandardsurgicaltechniqueanddrapethepreppedareawithsteriletowels.
3. (Ifapplicable)administerlocalanaesthesiatotheinsertionsite.
B. Pre-ush the Catheters
Slide Clamp Thump Clamp Plug
4. Irrigate,prime,andclampbothcatheterlumenswithheparinsolutionoraccordingtohospitalproto-
col.Withcathetersfullofuid,clampeachcatheterusingSlideClamporThumbClampbetweenthe
proximalendofthecatheterandthe50cmmarkingoruseplug.
C. Perform Venipuncture
5. Inserttheintroducerneedlewithanattachedsyringetothedesiredlocationusingultrasoundguidance(preferred).Aspirategentlyasthe
insertion is being made.
6. Whentheveinhasbeenentered,removethesyringeleavingtheneedleinplaceandoccludeendofneedlewiththumb.
7. InserttheexibleJendofthestandardguidewirethroughtheintroducerneedlehubthatwillpermitpassageofa0.038in.(0.97mm)guide-
wire.Advancethestandardguidewiretothedesiredlocationinvessel.
WARNING:Cardiacarrhythmiasmayresultiftheguidewireisallowedtopassintotherightatrium.
WARNING: Do not advance guidewire or catheter if unusual resistance is encountered.
8. Removetheneedlewhileholdingtheguidewireinplace.Wipetheexposedguidewirecleanandsecureitinplace.
CAUTION:Iftheguidewiremustbewithdrawnwhiletheneedleisinserted,removeboththeneedleandguidewireasaunittopreventthe
needle from damaging or shearing the guidewire.
9. Repeatsteps5-8abovetointroducesecondneedleandguidewireintosametargetveinapproximately1-2cmadjacenttotherstincision
site.
10.Aftersecondwirehasbeenplaced,makeasmallincisionateachvenousinsertionsitetoremoveanydermalbridgeandeaseinsertion.
D. Insert Sheath and Advance Catheter
11.Advancethedilatorsheathintroducerovertheexposedguidewireintothevessel.
CAUTION:CareshouldbetakenNOTtoforcethedilatorsheathintroducerintothevesselduringinsertionasvesseldamageincluding
perforation could result.
WARNING:Cardiacarrhythmiasmayresultiftheguidewireisallowedtopassintotherightatrium.
12.Withdrawthedilatorandguidewire,leavingthesheathintroducerinplace.
CAUTION:Careshouldbetakennottoadvancethesplitsheathtoofarintovesselasapotentialkinkwouldcreateanimpassetothecath-
eter.
WARNING:Topreventairembolismand/orbloodloss,placethumbovertheexposedoriceofthesheathintroducer.
13.Removethumbandfeeddistalsectionofcatheterintothesheathintroduceruntiltipiscorrectlypositioned.Thedepthmarkingsinonecm.
increments may be used to determine insertion length.
WARNING:Ensurethecatheterislledwithheparinizedsaline,isclampedandisfreeofairbubblesbeforeinsertingitintothevein.
WARNING: Do not advance guidewire or catheter if unusual resistance is encountered.
14.Withthecatheteradvanced,peelawaythesheathintroducerbygrippingthe“T”handleandbreakingitapartwithadownwardandoutward
motion to initiate separation and withdrawal of the sheath introducer.
CAUTION:Ensurethatthesheathintroducerisonlytornexternally.Cathetermayneedtobefurtherpushedintothevesselassheathintro-
ducer is torn.
CAUTION:Foroptimalproductperformance,donotinsertanyportionofthecuintothevein.
15.Repeatsteps11-14aboveforsecondcatheter.
16.Verifycatheterpositionunderuoroscopyandmakenecessaryadjustments.The venous (infusion) tip should be located at the level of
the caval atrial junction or into the right atrium and approximately 4 cm past the arterial (withdrawal) catheter tip.
CAUTION:Cathetertipsmustbestaggeredby4cmtominimizerecirculation.
E. Subcutaneous Tunneling of Catheters and Cu Placement
Tu nnel Tr act
Catheter
Exit Site Catheter
Insertion
Site
Figure 1
17.Identifydesiredtunnellocationandexitsitetoinsureagentlearc.Ensure
thattheexposedproximalportionofcatheterislongenoughtocutthe
cathetertothedesiredlengthandtoreadlengths/primingvolumes.
18.Makenoteofthedesiredlocationatwhichthecuwillresideinthetunnel.
19.(Ifapplicable)administerlocalanaesthesiatotunneltract.
20.Makeanincisionatthecatheterexitsite(Figure1).
21.Usingthemetaltunnelerprovided,createasubcutaneoustunnelfromthe
insertionsitetoemergeatthecatheterexitsite(Figure2b).
WARNING: Do not tunnel through muscle.
22.Oncemetaltunneleremergesfromcatheterexitsite,placetaperedend
of tunnel slide dilator over tip of tunneler at insertion site. While holding
endofmetaltunneler,expandtunneltractfrominsertionsitetohalfway
throughthetunneltodesiredculocationorwithin2cmofthecatheter
exitsiteusingaback-and-forthmotion(Figure3aand3b).
Figure 2bFigure 2a
Tunneler
Tunnel Slide Dilator
Figure 2c
23.Removetunnelslidedilatorwhileholdingmetaltunnelerinplace.Ifused,removeplug.
24.Attachcathetertotunnelersothatcatheter’sproximalendslidesoverbarbedendofmetaltunneler.
Tunneler
Tunnel
Slide
Dilator
Figure 3a
Figure 3b
25.RemoveSlideClamporThumbClamp.   
Figure 4
Cu
WARNING:TheSlideClamp,ThumbClampandPlugareprovidedforuse
during catheter placement only. Do not reuse.
26.Pullmetaltunnelercarefullyuntilcatheteremergesfromcatheterexitsite.
The catheter should not be forced through the tunnel.
CAUTION: Do not create a sharp bend in catheter tunnel as this may cause
kinkingandimpactow.
27.EngageSlideClampbetweendesiredcutlengthandcatheterexitsiteto
prevent blood loss and air embolism.
28.Removetunnelerbycuttingcathetersquarelyatdesiredpre-printedmarkto
produceaclean,smoothsurface(Figure5).
CAUTION:Cuttingthecatheteranywherebutthepre-printedmarkswillresult
in the inability to read the catheter length and priming volumes.
29.Rechecktippositioningusingx-rayoruoroscopy.
30.Repeatsteps17-29forsecondcatheter(Figure4).

Other Bard Medical Equipment manuals

Bard DuoGlide User manual

Bard

Bard DuoGlide User manual

Bard PowerPICC SOLO 2 User manual

Bard

Bard PowerPICC SOLO 2 User manual

Bard PowerMidline User manual

Bard

Bard PowerMidline User manual

Bard StatLock User manual

Bard

Bard StatLock User manual

Bard E-Luminexx Vascular Stent User manual

Bard

Bard E-Luminexx Vascular Stent User manual

Bard Site-Rite 5 User manual

Bard

Bard Site-Rite 5 User manual

Bard Onflex User manual

Bard

Bard Onflex User manual

Bard DAVOL 100cc User manual

Bard

Bard DAVOL 100cc User manual

Bard QuickLink 70310QCA1 User manual

Bard

Bard QuickLink 70310QCA1 User manual

Bard Site-Rite IV User manual

Bard

Bard Site-Rite IV User manual

Bard Groshong Midline User manual

Bard

Bard Groshong Midline User manual

Bard EnCor Ultra User manual

Bard

Bard EnCor Ultra User manual

Bard Magic3 Go User manual

Bard

Bard Magic3 Go User manual

Bard Pinpoint GT User manual

Bard

Bard Pinpoint GT User manual

Bard Groshong PICC User manual

Bard

Bard Groshong PICC User manual

Bard PowerPICC SOLO 2 User manual

Bard

Bard PowerPICC SOLO 2 User manual

Bard Site-Rite 6 User manual

Bard

Bard Site-Rite 6 User manual

Bard vas-cath User manual

Bard

Bard vas-cath User manual

Bard ENCOR ECMRINTLOC User manual

Bard

Bard ENCOR ECMRINTLOC User manual

Bard retrO User manual

Bard

Bard retrO User manual

Bard Ultrasound System User manual

Bard

Bard Ultrasound System User manual

Bard MAX-CORE MC1410 User manual

Bard

Bard MAX-CORE MC1410 User manual

Bard Atlas Gold PTA Dilatation Catheter User manual

Bard

Bard Atlas Gold PTA Dilatation Catheter User manual

Bard PowerGlide User manual

Bard

Bard PowerGlide User manual

Popular Medical Equipment manuals by other brands

ZacUrate 430P user manual

ZacUrate

ZacUrate 430P user manual

Stryker Model 1004 Operation manual

Stryker

Stryker Model 1004 Operation manual

laerdal Thomas Select Tube Holder user guide

laerdal

laerdal Thomas Select Tube Holder user guide

Gima 34327 manual

Gima

Gima 34327 manual

PARAMOUNT BED A5 Series instruction manual

PARAMOUNT BED

PARAMOUNT BED A5 Series instruction manual

VOCARE vitals360 user manual

VOCARE

VOCARE vitals360 user manual

Datex-Ohmeda S/5 M-COPSv Technical Reference Manual Slot

Datex-Ohmeda

Datex-Ohmeda S/5 M-COPSv Technical Reference Manual Slot

Barco Nio Color 5MP user guide

Barco

Barco Nio Color 5MP user guide

Welch Allyn Connex Vital Signs Monitor 6000 Series Service manual

Welch Allyn

Welch Allyn Connex Vital Signs Monitor 6000 Series Service manual

laerdal SimBaby user guide

laerdal

laerdal SimBaby user guide

Atmos S 201 Thorax Short operating instructions

Atmos

Atmos S 201 Thorax Short operating instructions

Aspen Vista quick start guide

Aspen

Aspen Vista quick start guide

O2 Concepts OXLIFE LIBERTY 301-0005 user manual

O2 Concepts

O2 Concepts OXLIFE LIBERTY 301-0005 user manual

Optomed Aurora user manual

Optomed

Optomed Aurora user manual

AMO iDesign G300 Operator's manual

AMO

AMO iDesign G300 Operator's manual

VeinSight VS500 user manual

VeinSight

VeinSight VS500 user manual

GE MAC 600 SPY Service manual

GE

GE MAC 600 SPY Service manual

NeoMed GravityPro Instructions for use

NeoMed

NeoMed GravityPro Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.