27
PARAMETER 05 WORKING FREQUENCY OF PUMP 1
If the pump is configured for ON-OFF mode, this parameter is the working frequency of the pump.
If the pump is configured for proportional functioning, this parameter states the maximum working frequency of the pump
(when the measurement is far from the set threshold).
PARAMETER 06 FUNCTIONING TYPE OF THE PUMP 2
This parameter allows to set the functioning type of pump 2, typically used for controlling the redox potential level. See
description of parameter P01.
PARAMETER 07 SET-POINT PUMP 2
See parameter P02, but referred to redox measurements.
PARAMETER 08 HYSTERESIS PUMP 2
See parameter P03, but referred to redox measurements.
PARAMETER 09 DOSAGE DIRECTION OF PUMP 2
See parameter P04, but referred to redox measurements and dosing directions “Chlorination / De-chlorination”.
PARAMETER 10 WORKING FREQUENCY OF PUMP 2
See parameter P05, but referred to redox measurements.
PARAMETER 11 DOSAGE ALARM OF PUMP 1
This parameter allows to generate an alarm when the measurement does not return to the setpoint value within a set time
interval, from 0 (function disabled) to 240 minutes. The counter starts when the unit detects a measurement that exceeds
the setpoint value, and automatically resets to zero when measurement returns to an acceptable. If measurement remains
outside of the setpoint for a time longer than that set, an alarm is generated, the display shows the message “AL.1” and the
pump stops dosing. To reset the alarm and resume to normal operations, press the ON/OFF button or shutdown and
restart the unit. This condition can occur when the dosage is insufficient to reach the setpoint (P02).
PARAMETER 12 DOSAGE ALARM OF PUMP 2
See parameter P11, but referred to pump 2.
PARAMETER 13 START-UP DELAY
This parameter allows to set a start-up delay, which is a time of waiting before starting measurements to allow the correct
polarization and stabilization of the measuring electrodes. This prevents unreliable initial readings. Typically in the case of
pH measurements one minute is enough, while for redox electrodes may be necessary a delay of 30 minutes. This waiting
time also allows to compensate for any hydraulic delays which may occur at the start-up of the plant.
Set a time (in minutes) during which, after turning on the electronic unit, the system waits, measurements blink on the
display and the pumps are not active. After this time, the system begins normal operation
.
PARAMETER 14 LANGUAGE
This parameter allows to select the display language.
PARAMETER 15 RESTORE THE DEFAULT SETTINGS
This feature allows to restore the factory configuration if you want to delete incorrect or undesired settings. Once confirmed
this option, all custom setting will be lost. To enable the feature, enter the value “12” .
PARAMETER 16 PASSWORD
This parameter allows to enter a password (numeric value within 1 and 255) to protect the system from unauthorized
access. Once set and confirmed, the password will be requested to access the menus “Advanced configuration” and
“Manual mode”.
The instrument is supplied with no password set (P16=0).
Warning! If you forget the password, you need to send back the unit to the factory for a complete reconfiguration (out of
warranty)!
PARAMETER 17 FUNCTIONING OF THE ALARM OUTPUT
This parameter allows to select the functioning mode of the alarm relay, normally open “NO” (factory setting) or normally
closed “NC”.
PARAMETER 18 FLOW FUNCTIONING
This parameter allows to select the functioning mode of the “flow” input, normally open “NO” (factory setting) or normally
closed “NC”.