BCI 1884SS User manual

Pulse Oximeter for Sleep
Screening
Operation Manual
Catalog Number 1884SS
Version 1, April 2002
Copyright BCI, Inc. - 2002


Table of Contents
Pulse Oximeter for Sleep Screening Operation Manual i
Clinician’s Operation Manual
Warranty & Service Information ................................................................................................... iii
Proprietary Notice ....................................................................................................................... iii
Limited Warranty ........................................................................................................................ iii
Service Support ........................................................................................................................... iii
CE Notice....................................................................................................................................... iii
Warnings, Cautions, & Notes iv
Symbol Definitions ........................................................................................................................ iv
Warnings ........................................................................................................................................ iv
Cautions ...........................................................................................................................................v
Notes ................................................................................................................................................v
Chapter 1: Introduction 1-1
About the Manual......................................................................................................................... 1-1
Theory of Operation..................................................................................................................... 1-1
Chapter 2: Intended Use and Monitor Features 2-1
Intended Use ................................................................................................................................ 2-1
Monitor Features .......................................................................................................................... 2-1
Chapter 3: Controls & Features 3-1
Monitor Front Panel ..................................................................................................................... 3-1
Sensor/Serial Connector............................................................................................................ 3-1
SpO2 Numeric Display .............................................................................................................. 3-1
Pulse Rate Numeric Display......................................................................................................3-1
Pulse Strength Bar Graph.......................................................................................................... 3-1
Low Battery Indicator ............................................................................................................... 3-1
xON OFF Key......................................................................................................................... 3-2
FPRINT Key ........................................................................................................................... 3-2
Chapter 4: Operating Instructions 4-1
Unpacking the Monitor ................................................................................................................ 4-1
Installing the Batteries.................................................................................................................. 4-1
Attaching the Sensor to the Patient .............................................................................................. 4-2
Choosing the Sensor.................................................................................................................. 4-2
Care and Handling of the Sensor............................................................................................... 4-3
Checking the Sensor and Oximetry Cable ................................................................................ 4-4
Cleaning or Disinfecting the Sensors ........................................................................................ 4-4
Turning On the Monitor............................................................................................................... 4-5
Low Battery Indicator .................................................................................................................. 4-6
Turning Off the Monitor .............................................................................................................. 4-6
Checking the Monitor’s Performance .......................................................................................... 4-6
Operating Modes.......................................................................................................................... 4-6
Verification Mode ........................................................................................................................ 4-7
Data Collection Mode .................................................................................................................. 4-7
Trend Printing Mode.................................................................................................................... 4-8
Chapter 5: Instructions for Training the Home User 5-1
Equipment and Supplies Checklist for Home Use ....................................................................... 5-1
Training the Home User............................................................................................................... 5-1
Chapter 6: Trend Data 6-1
Description................................................................................................................................... 6-1
Clearing Trend Data..................................................................................................................... 6-1

Table of Contents
ii Pulse Oximeter for Sleep Screening Operations Manual
Chapter 7: Printer 7-1
Description................................................................................................................................... 7-1
Trend Summary Printout ............................................................................................................. 7-1
What You’ll Need for Printing .................................................................................................... 7-3
Trend Printouts ............................................................................................................................ 7-4
Collecting Trend Data............................................................................................................... 7-4
Paper Feed................................................................................................................................. 7-4
Chapter 8: PC Communication Setup 8-1
Description................................................................................................................................... 8-1
PC Communication Setup............................................................................................................ 8-1
Chapter 9: Maintenance 9-1
Schedule of Maintenance............................................................................................................. 9-1
Storage ......................................................................................................................................... 9-1
Chapter 10: Troubleshooting 10-1
EMI Interference........................................................................................................................ 10-2
Chapter 11: Optional Supplies & Accessories 11-1
Ordering Information:................................................................................................................ 11-1
Chapter 12: Specifications 12-1
Equipment Classification ........................................................................................................... 12-1
Displays, Indicators, & Keys ..................................................................................................... 12-1
SpO2........................................................................................................................................... 12-1
Pulse Rate .................................................................................................................................. 12-2
Printer ........................................................................................................................................ 12-2
Power Requirements .................................................................................................................. 12-2
Battery Life................................................................................................................................ 12-2
Dimensions ................................................................................................................................ 12-2
Environmental Specifications .................................................................................................... 12-2
Appendix A: Data Format Appendix-1
Data format. .................................................................................................................Appendix-1
Appendix B: Setting the Optional Real Time Clock Appendix-2

Warranty & Service Information
Pulse Oximeter for Sleep Screening Operation Manual iii
Warranty & Service Information
Proprietary Notice
Information contained in this document is copyrighted by BCI, Inc. and may not be duplicated
in full or part by any person without prior written approval of BCI, Inc. Its purpose is to
provide the user with adequately detailed documentation to efficiently install, operate,
maintain and order spare parts for the device supplied. Every effort has been made to keep the
information contained in this document current and accurate as of the date of publication or
revision. However, no guarantee is given or implied that the document is error free or that it is
accurate regarding any specification.
Limited Warranty
BCI, Inc. warrants each new device to be free from defects in workmanship and materials
under normal use and service for a period of two (2) years from the date of shipment. The sole
obligation of BCI, Inc. under this warranty will be to repair or replace, at its option, products
that prove to be defective during the warranty period. The foregoing shall be the sole
warranty remedy. Except as set forth herein, seller makes no warranties, either expressed or
implied, including the implied warranties of merchantability and fitness for a particular
purpose. No warranty is provided if the products are modified without the express written
consent of BCI, Inc., and seller shall not be liable in any event for incidental or consequential
damage. This warranty is not assignable.
Service Support
Repairs for devices manufactured by BCI, Inc. under warranty must be made at authorized
repair centers. If the device needs repair, contact your local distributor or the BCI, Inc. service
department. When calling, have the device’s model and serial number ready.
If you need to ship the device, pack the device and accessories carefully to prevent shipping
damage. All accessories should accompany the device.
BCI, Inc.
N7 W22025 Johnson Road
Waukesha, WI USA 53186-1856
Phone: (262) 542-3100
Fax: (262) 542-3325
Toll Free; (800) 558-2345
CE Notice
Marking by the symbol C0473 indicates compliance of this device to the Medical Device
Directive 93/42/EEC.
Authorized Representative (as defined by the Medical Device Directive):
Graseby Medical Ltd.
Colonial Way, Watford, Herts,
UK WD2 4LG
Phone: (44) 1923 246434
Fax: (44) 1923 240273

Warnings, Cautions, & Notes
iv Pulse Oximeter for Sleep Screening Operation Manual
Warnings, Cautions, & Notes
KEYWORD DEFINITION
WARNING Tells you about something that could hurt the patient or hurt the operator
CAUTION Tells you something that could damage the monitor
NOTE Tells you other important information
Symbol Definitions
SYMBOL DEFINITION
r Type BF equipment
2Attention, consult accompanying documents
1 Refer servicing to qualified service personnel
SN Serial Number
REF Catalog Number
F Printer On/Off
y Date of Manufacturing
2 Do not reuse
x On/Off
1 Non AP device
l Lot Number
C Use By
IPX1 Drip Proof
Warnings
WARNING! Federal law (USA) restricts the use or sale of this device by, or on the order
of a physician.
WARNING! Do not use this device in the presence of flammable anesthetics.
WARNING! Do not use this device in the presence of magnetic resonance imaging (MR
or MRI) equipment.
WARNING! This device must be used in conjunction with clinical signs and symptoms.
This device is only intended to be an adjunct in patient assessment.
WARNING! Prolonged use or the patient’s condition may require changing the sensor
site periodically. Patients setting up the device at home should be instructed of the
signs/symptoms of sensor misuse (e.g., pressure sores from prolonged application) and
should be advised to move the sensor to another finger if necessary.
WARNING! When attaching sensors with Microfoam®1tape, do not stretch the tape or
attach the tape too tightly. Any device (tape, Velcro™) applied too tightly may cause
inaccurate readings and blisters on the patient’s skin (lack of skin respiration, not heat,
causes the blisters).
1Microfoam® is a registered trademark of the 3M Company

Warnings, Cautions, & Notes
Pulse Oximeter for Sleep Screening Operation Manual v
WARNING! When connecting this monitor to any instrument, verify proper operation
before clinical use. Refer to the instrument’s user manual for full instructions.
Accessory equipment connected to the monitor’s data interface must be certified
according to the respective IEC standards, i.e., IEC 950 for data processing equipment or
IEC 601-1 for electromedical equipment. All combinations of equipment must be in
compliance with IEC 601-1-1 systems requirements. Anyone connecting additional
equipment to the signal input port or the signal output port configures a medical system,
and, therefore, is responsible that the system complies with the requirements of the
system standard IEC 601-1-1.
WARNING! IEC 950 approved equipment must be placed outside the “patient
environment.” The patient environment is defined as an area 1.5 m (4.92 feet) from the
patient.
PATIENT ENVIRONMENT
R 1.5m (4.9 ft)
Cautions
CAUTION! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in liquid.
Evidence that liquid has been allowed to enter the monitor voids the warranty.
CAUTION! This device is intended for use by persons trained in professional health care
or those who have access to the oversight of a professional health care provider. The
operator must be thoroughly familiar with the information in this manual and/or the
“Home User’s Guide” before using the device.
CAUTION! Connect only the PC cable specifically intended for use with this device (see
Optional Supplies and Accessories).
CAUTION! When printing, or advancing the paper, place the oximeter on its’ side to
prevent paper jams.
Notes
NOTE! Operation of this device may be adversely affected in the presence of strong
electromagnetic sources, such as electrosurgery equipment.
NOTE! Operation of this device may be adversely affected in the presence of computed
tomograph (CT) equipment.
NOTE! Use only SpO2sensors supplied with, or specifically intended for use with, this
device.
NOTE! SpO2measurements may be adversely affected in the presence of high ambient
light. Shield the sensor area (with a towel, for example) if necessary.

Warnings, Cautions, & Notes
vi Pulse Oximeter for Sleep Screening Operation Manual
NOTE! Dyes introduced into the bloodstream, such as methylene blue, indocyanine
green, indigo carmine, patent blue V (PBV), and fluorescein, may adversely affect the
accuracy of the SpO2reading.
NOTE! Any condition that restricts blood flow, such as use of a blood pressure cuff or
extremes in systemic vascular resistance, may cause an inability to determine accurate
pulse rate and SpO2readings.
NOTE! Remove fingernail polish or false fingernails before applying SpO2sensors.
Fingernail polish or false fingernails may cause inaccurate SpO2readings.
NOTE! Batteries are designed to be replaceable by individuals trained in this procedure.
Follow local governing ordinances and recycling instructions regarding disposal or
recycling of device components, including batteries.
NOTE! SpO2averaging is the number of pulse beats over which the SpO2value is
averaged; pulse averaging is the number of seconds over which the pulse value is
averaged.
NOTE! Hazards arising from software errors have been minimized. Hazard analysis was
performed to meet EN60601-1-4:2000.
NOTE! Optical cross-talk can occur when two or more sensors are placed in close
proximity. It can be eliminated by covering each site with an opaque material.
NOTE! All user and patient accessible materials are non-toxic.

Chapter 1: Introduction
Pulse Oximeter for Sleep Screening Operation Manual 1-1
Chapter 1: Introduction
About the Manual
The Operation Manual provides installation, operation, and maintenance instructions for
health-care professionals and other users, trained in monitoring respiratory and cardiovascular
activity.
Theory of Operation
The oximeter determines SpO2and pulse rate by passing two wavelengths of low intensity
light, one red and one infrared, through body tissue to a photodetector. During measurement,
the signal strength resulting from each light source depends on the color and thickness of the
body tissue, the sensor placement, the intensity of the light sources, and the absorption of the
arterial and venous blood (including the time varying effects of the pulse) in the body tissues.
Figure 1.0: Theory of Operation
1. Red and infrared LED light sources
2. Detector
The oximeter processes these signals, separating the time invariant parameters (tissue
thickness, skin color, light intensity, and venous blood) from the time variant parameters
(arterial volume and SpO2) to identify the pulse rate and calculate oxygen saturation. Oxygen
saturation calculations can be performed because oxygen saturated blood predictably absorbs
less red light than oxygen depleted blood.
Since measurement of SpO2depends on a pulsating vascular bed, any condition that restricts
blood flow, such as use of a blood pressure cuff or extremes in systemic vascular resistance,
may cause an inability to determine accurate pulse and SpO2readings.


Chapter 2: Intended Use & Monitor Features
Pulse Oximeter for Sleep Screening Operation Manual 2-1
Chapter 2: Intended Use and Monitor Features
Intended Use
This device is a handheld pulse oximeter that measures SpO2, pulse rate, and pulse strength. It
may be used as a spot check device in the hospital or clinical environments. Additionally, it
may be used to collect long term data in overnight sleep screening studies in the sleep lab or
home. The oximeter will provide reliable measurements on patients ranging from neonate to
adults for spot checking applications and from pediatric to adult for sleep screening
applications when using the appropriate catalog accessories. The oximeter will operate
accurately over an ambient temperature range of 32 to 131°F (0 to 55°C).
This device is not intended for continuous patient monitoring, but can be used for spot
checking and/or long-term data collection applications. There are no audible or visible patient
alarms.
Monitor Features
• Provides fast, reliable SpO2, pulse rate, and pulse strength measurements on any patient,
from neonates to adults for spot-checking applications and on pediatric to adults for sleep
screening applications.
• Ideally suited for use in overnight sleep screening applications allowing printing of
summated data results.
• Three modes of operation: Verification Mode for spot checking, Data Collection Mode
for long term data collection in sleep screening environments and Trend Printing Mode
for outputting summarized trend data to the built-in printer or to a PC.
• Device automatically enters Data Collection Mode after 10 minutes of continuous use.
• Minimal LED display during Data Collection Mode lengthens battery life (I.e. – display is
reduced to a single flashing bar).
• Portable and lightweight. Weighs only 13 ounces (369 grams)., with batteries.
• Ergonomically designed to fit comfortably in the palm of your hand.
• Uses four (4) standard “AA” (type IEC LR6) alkaline cells.
• Battery life (Alkaline Cells):
- 2 sleep screening cycles (1 cycle = 10 minutes with display on, 8 hours with
display off, 5 minutes printing)
or
- Intermittent monitoring: 1 min on, 2 min off (= 1 cycle) 80 hours or 1600 cycles
• Bright, easy-to-read LED displays indicate SpO2and pulse rate measurements.
• An eight-segment LED bar graph indicates pulse strength in the Verification Mode.
• Low Battery Indicator flashes when about 30 minutes of battery use remains.
• Capable of 12 hours of data storage.


Chapter 3: Controls & Features
Pulse Oximeter for Sleep Screening Operation Manual 3-1
Chapter 3: Controls & Features
Monitor Front Panel
Figure 3.1: Monitor Front Panel
1. Sensor/Serial Connector
The sensor connects here, or an oximetry cable can be connected between the monitor and the
sensor. The serial printer or PC communication cable is also connected here.
2. SpO2 Numeric Display
A number shows the patient’s SpO2value in percent. Dashes (--) mean the monitor is not
able to calculate the SpO2value in Verification Mode. SpO2display in Data Collection Mode
will be one flashing bar to conserve battery life.
3. Pulse Rate Numeric Display
A number shows the patient’s pulse rate value in beats per minute. Dashes (---) mean the
monitor is not able to calculate the pulse rate value in Verification Mode. In Data Collection
Mode, the Heart Rate display will be blank to conserve battery life.
4. Pulse Strength Bar Graph
In Verification Mode, the pulse strength bar graph “pulses” with the patient’s pulse beat. The
height of the bar graph shows the patient’s pulse strength.
5. Low Battery Indicator
A Low Battery Indicator will be indicated by the top bar graph LED quickly flashing about
once every second.

Chapter 3: Controls & Features
3-2 Pulse Oximeter for Sleep Screening Operation Manual
Figure 3.2: Monitor Operating Keys
6. xON OFF Key
Pressing this key turns the monitor ON and OFF.
7. FPRINT Key
See Chapter 4 for more details on the three modes.
Verification Mode: In this mode, momentarily pressing the Fkey will not activate the
printer as the printer function is suppressed in this mode. Pressing and holding the Fkey in
the Verification Mode will advance the paper. Reloading the paper is allowed in this mode.
Pressing and holding the Fkey at power up will erase trends, even if the trend memory is
clear.
Data Collection Mode: In this mode, momentarily pressing the Fkey will not activate the
printer as the printer function is suppressed in this mode. However, in the Data Collection
Mode, pressing and holding this key will advance the paper in the printer. Reloading the
paper is allowed in this mode.
Trend Printing Mode: Momentarily pressing the Fkey will activate a printout (statistical
sleep screening information). Pressing the Fkey during the printout will terminate the
printout. Pressing and holding the Fkey will advance the paper. Holding the Fkey for 6 to
8 seconds at power up will clear the trend and switch the unit into Verification Mode.
Reloading the paper is allowed in this mode.
NOTE! All data will be irretrievably lost when the trend data is cleared.

Chapter 4: Operating Instructions
Pulse Oximeter for Sleep Screening Operation Manual 4-1
Chapter 4: Operating Instructions
2
WARNING: This device must be used in conjunction with clinical signs and
symptoms. This device is only intended to be an adjunct in patient assessment.
Unpacking the Monitor
1. Carefully remove the monitor and its accessories from the shipping carton. Save the
packing materials in case the monitor must be shipped or stored.
2. Compare the packing list with the supplies and equipment you received to make sure you
have everything you’ll need.
Installing the Batteries
The oximeter uses 4 standard “AA” alkaline cells, IEC Type LR6.
To install/replace the batteries:
1. Depress the battery door tab and lift up.
2. Install the negative end of each battery first, compressing the battery terminal spring until
the positive terminal clears the positive tab. Press the battery down into place.
3. Place battery door tabs into the slots of the monitor back panel, depress the door tab and
press the door into place.
NOTE: If you install disposable batteries, be sure to dispose of them in compliance with
your institution’s guidelines and local ordinances.
NOTE: The unit will hold data for about one and a half minutes with no battery power.
This allows for battery replacement without loss of stored data.
Figure 4.0: Installing the Batteries.

Chapter 4: Operating instructions
4-2 Pulse Oximeter for Sleep Screening Operation Manual
Attaching the Sensor to the Patient
What you need to know about attaching the sensor to the patient:
2
WARNING! Prolonged use or the patient’s condition may require changing the
sensor site periodically. Patients setting up the device at home should be
instructed of the signs/symptoms of sensor misuse (e.g., pressure sores from
prolonged application) and should be advised to move the sensor to another finger
if necessary.
2
WARNING! When attaching sensors with Microfoam® tape, do not stretch the tape
or attach the tape too tightly. A sensor applied too tightly by any method (tape,
Velcro™) may cause inaccurate readings and blisters on the patient’s skin (lack of
skin respiration, not heat, causes the blisters).
Attaching the patient to the monitor requires these steps:
1. Choose the sensor.
2. Check the sensor and oximetry cable.
3. Clean or disinfect the sensor if using the reusable type. (Disposable sensors are for single-
patient use and do not require cleaning or disinfecting.)
4. Attach the sensor to the patient.
Choosing the Sensor
Choose the appropriate sensor from the following chart.
PATIENT SITE DESCRIPTION
Adult
> 45 Kg
Finger
Finger or Toe
Ear
3044: Sensor, Adult (reusable)
3444, Comfort Clip™ Finger Sensor (reusable)
3043: Sensor, Universal “Y” (reusable)
1300: Sensor, Disposable, Adult Finger
1310: D.O.T Sensor
1311: D.O.T Sensor
3078: Sensor, Ear (reusable)
Pediatric
15-45 Kg
Finger
Finger or Toe
Ear
3044: Sensor, Adult (reusable)
3444, Comfort Clip™ Finger Sensor (reusable)
3043: Sensor, Universal “Y” (reusable)
1301: Sensor, Disposable, Ped. Finger
3078: Sensor, Ear (reusable)
Infant
3-15 Kg (for
spot-check
only)
Hand or Foot
Toe
Finger or Toe
3043: Sensor, Universal “Y” (reusable)
3025: Sensor, Wrap, Infant (reusable)
1303: Sensor, Disposable, Infant
Neonate
< 3 Kg (for
spot-check
only)
Hand or Foot
Foot
1302: Sensor, Disposable, Neonate
3026: Sensor, Wrap, Neonate (reusable)

Chapter 4: Operating Instructions
Pulse Oximeter for Sleep Screening Operation Manual 4-3
Care and Handling of the Sensor
2
WARNING! Misuse or improper handling of the sensor and cable could result in
damage to the sensor. This may cause inaccurate readings.
Hold the connector rather than the cable when connecting or disconnecting the sensor to the
device as shown in Figure 4.1.
Figure 4.1: Disconnecting or connecting the sensor.
1. Connector
2. Cable
Do not use excessive force or unnecessary twisting when connecting, disconnecting, storing,
or using the sensor.
When placing the sensor on the patient, allow the cable to lay across the palm of the hand and
parallel to the arm of the patient as shown in Figure 4.2.
Figure 4.2: Positioning the cable of the finger sensor.
Upon completion of patient monitoring, detach the sensor as shown in Figure 4.1 and loosely
coil the finger sensor cable.

Chapter 4: Operating instructions
4-4 Pulse Oximeter for Sleep Screening Operation Manual
Checking the Sensor and Oximetry Cable
Follow these instructions each time before you attach the sensor to the patient. This helps
ensure the sensor and oximetry cable is working properly.
2
WARNING! Using a damaged sensor may cause inaccurate readings. Inspect each
sensor. If a sensor appears damaged, do not use it. Use another sensor or contact
your authorized repair center for help.
2
WARNING! Using a damaged oximetry cable may cause inaccurate readings.
Inspect the oximetry cable. If the oximetry cable appears damaged, do not use it.
Contact your authorized repair center for help.
1. Carefully inspect the sensor to make sure it does not appear damaged.
2. If using the oximetry cable:
a. Carefully inspect the oximetry cable to make sure it does not appear damaged.
b. If the sensor is not already connected to the oximetry cable, connect the sensor to the
oximetry cable. Carefully align and push the connectors together firmly and close the
latch to secure the connectors.
c. If the oximetry cable is not already connected to the monitor, connect the oximetry
cable to the monitor. Carefully align and push the connector firmly into the monitor.
3. If not using the oximetry cable, connect the sensor to the monitor. Push the connector
firmly into the monitor.
4. If the monitor is not already on, press the xkey to turn on the monitor.
2
WARNING! If any of the integrity checks fail, do not attempt to monitor the patient.
Use another sensor or oximetry cable, or contact the equipment dealer for help if
necessary.
5. Before the sensor is attached to the patient, check the integrity of the sensor, oximetry
cable, and oximeter by making sure the red light in the sensor is illuminated.
NOTE! Obstructions or dirt on the sensor’s red light or detector may cause the integrity
checks to fail. Make sure there are no obstructions and the sensor is clean.
6. You are now ready to attach the sensor to the patient.
Cleaning or Disinfecting the Sensors
Clean or disinfect reusable sensors before attaching to a new patient.
2
WARNING! Do not autoclave, ethylene oxide sterilize, or immerse the sensors in
liquid.
2CAUTION! Unplug the sensor from the monitor before cleaning or disinfecting.
Clean the sensor with a soft cloth moistened in water or a mild soap solution. To disinfect the
sensor, wipe the sensor with isopropyl alcohol.

Chapter 4: Operating Instructions
Pulse Oximeter for Sleep Screening Operation Manual 4-5
Turning On the Monitor
1. To turn on the monitor, press the xkey. When turned ON, the monitor does the
following:
• The pulse strength bar graph segments rapidly light the bottom and top LED.
• The monitor's software revision is momentarily displayed.
Figure 4.3: SpO2, pulse rate, and pulse strength bar graph.
1. Patient’s SpO2Shown Here
2. Patient’s Pulse Shown Here
3. Patient’s Pulse Strength Shown Here

Chapter 4: Operating instructions
4-6 Pulse Oximeter for Sleep Screening Operation Manual
Low Battery Indicator
1. The Low Battery Indicator is the top LED segment of the pulse strength bargraph.
2. The Low Battery Indicator will flash rapidly when the batteries need to be replaced.
3. The Low Battery Indicator and “trn” will flash rapidly if the monitor is trying to print
and there is not enough battery life for printing. This indicator will go away when the
printing stops.
2 WARNING! When the LOW BATTERY INDICATOR flashes, you must immediately
replace the monitor’s batteries. Otherwise, the monitor turns itself off about 30
minutes after the highest bargraph LED begins to flash in any mode.
Turning Off the Monitor
Turn off the monitor when you are not monitoring a patient.
To turn off the monitor, press the xkey. To turn off the monitor in the Data Collection
Mode, the xkey must be pressed and held for 6 seconds.
NOTE! Once the monitor is powered off in the Data Collection Mode, it cannot resume
collection of trend data without clearing all previously obtained data.
Checking the Monitor’s Performance
Pulse oximeters do not require user calibration. If checking the function of the device is
desired, an optional Oximeter/ECG Patient Simulator (Catalog No. 1606HH) is available as an
accessory. The simulator attaches to the oximeter via the 3311 oximetry cable. It provides a
known SpO2and pulse rate signal to the oximeter. This allows the oximeter's performance to
be checked.
NOTE! The 1606HH Oximeter/ECG Patient Simulator does not calibrate the monitor; the
monitor does not require calibration. The 1606HH provides a known SpO2and pulse rate
to the monitor that allows you to check the monitor’s performance.
NOTE! Follow the instructions included with the 1606HH Oximeter/ECG Patient
Simulator.
Operating Modes
The monitor has three operating modes: Verification Mode, Data Collection Mode, and Trend
Printing Mode.
• Verification Mode is intended for use in spot checking and for verification of correct
sensor placement.
• Data Collection Mode is intended for use in long-term data collection in sleep screening
environments.
• Trend Printing Mode is intended for use in the output of the trend data summation to the
printer or a full computer data download.
Table of contents
Other BCI Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Graham Field
Graham Field GENDRON TRC28750 Series user manual

Medirol
Medirol VIVERA FASTENER F401 Operation and maintenance manual

Braun
Braun Aesculap Eccos Instructions for use/Technical description

Lanaform
Lanaform PULSE OXIMETER S1 Instructions for use

Anspach
Anspach Black Max operating manual

Pic Solution
Pic Solution AIRFamily evolution manual

Merit Medical
Merit Medical Q50X Instructions for use

bleep
bleep DreamPort Application Instructions

Cardinal Health
Cardinal Health CATALYST Patient Quick Reference Guide

Infoscan
Infoscan MED Recorder Patient manual

Park House Healthcare
Park House Healthcare RICHMOND ULTRA LOW BED Service manual

inbody
inbody 170 user manual