Biobase BK-BTS1 User manual

Blood Bag Tube Sealer
BK-BTS1
User Manual
Warnings and Precautions
Note:
Please read this manual carefully before use.
Do not use in a humid, hot or dusty location to avoid instrument malfunction.Do not
move the instrument while the instrument is running, and do not open the instrument
door.
Automatic welding machine using the classical oscillator to generate high frequency
oscillation source, the circuit is mature, stable and reliable, fully automated welding
process, no voltage adjustment and time adjustment. Has the advantages of simple
operation, rapid heating, the advantages of accurate and stable, is an ideal instrument.
1.Start machine
1.1Insert the three-hole end of the power cord into the socket with the fuse box (2A) at
the bottom of the instrument and the AC end into the AC 220V power socket.
1.2 Press the power switch in the upper right corner of the front of the instrument. The
indicator on the power switch is red. When the instrument is turned on, it can be heat
sealed.
2 Heat sealing
2.1 The plastic tube that needs to be heat-sealed is placed in the heat sealing groove, and
the heat-closing head of the heat-closing groove is automatically clamped, and the LED
indicator on the instrument emits red light to enter the heat-closing state, and the LED
indicator turns yellow and then goes out. The heat sealing head is loosened to complete
the heat sealing.
2.2 The plastic tube taken out can be broken or cut.
6
5
3 adjustment of heat sealing gap
3.1 Remove the heat sealer mask and the M3 screw next to the heat sealing head as the
gap adjustment screw. The instrument has been adjusted to the ideal state before leaving
the factory. Under normal circumstances, there is no need to adjust.
3.2 The method of adjusting the gap is as follows: the counterclockwise swirling gap
becomes smaller, and the heat sealing of the plastic tube after heat sealing becomes
thinner and relatively easy to break. The clockwise swirling gap becomes larger, and the
heat sealing of the plastic tube after heat sealing becomes thicker, which is relatively
difficult to break or needs to be cut.
3.3 When adjusting, the rotation should be no more than 1/8 turn.
4 Maintenance
4.1 In order to make the instrument work properly, the instrument must be placed in a
clean, dry, non-erosion environment and cleaned frequently. When not in use, turn off
the power and unplug the AC connector.
In order to understand the quality of our products and to provide better services to users,
please note:
Certification
Name:Blood Bag Tube Sealer
Model:BK-BTS1
The product is inspected to meet the technical standards and is
allowed to leave the factory.
2
4.2 Troubleshooting
5 Performance
Power
AC110V~260V 2A
Heat sealing time
0.8~1.2s
Adaptation
material
Outer diameter 2mm~6mm Wall thickness is not than 1mm
Package size
320mm*180mm*130mm
Gross weight
6kg
3
Packing List
Name
Qty
Main machine
1set
fuse
2只(2A)
Power line
1pc
certificate
1pc
User manual
1pc
phenomenon
reason
method
No display
The power is not
plugged in
Fuse does not pass
Power cord is damaged
Plug in the power
Replace the fuse
Replace the power cord
Heat sealing effect is
not ideal
Adjust the heat sealing gap

Popular Food Saver manuals by other brands

4titude 4s2 Service manual

4titude

4titude 4s2 Service manual

CASO DESIGN VC15 Original operating manual

CASO DESIGN

CASO DESIGN VC15 Original operating manual

Avery Dennison Monarch 9415 Maintenance instructions

Avery Dennison

Avery Dennison Monarch 9415 Maintenance instructions

PowerXL HR1014-0BTV owner's manual

PowerXL

PowerXL HR1014-0BTV owner's manual

DoCash 2045 user manual

DoCash

DoCash 2045 user manual

Kogan KAVACSLXXXB user manual

Kogan

Kogan KAVACSLXXXB user manual

Weston SPORTSMAN'S ADVANTAGE 65-0501-RT user guide

Weston

Weston SPORTSMAN'S ADVANTAGE 65-0501-RT user guide

Insta 228 Operation and maintenance manual

Insta

Insta 228 Operation and maintenance manual

Seal-a-Meal VS107 reference guide

Seal-a-Meal

Seal-a-Meal VS107 reference guide

ARENDO 303059 user manual

ARENDO

ARENDO 303059 user manual

Sunbeam VS5500 Service manual

Sunbeam

Sunbeam VS5500 Service manual

Paitec ES2500 user manual

Paitec

Paitec ES2500 user manual

Thermopatch Auto Maxi NL-33 user manual

Thermopatch

Thermopatch Auto Maxi NL-33 user manual

PFE AS700 operating instructions

PFE

PFE AS700 operating instructions

3M Matic 700r-s Instructions and parts list

3M

3M Matic 700r-s Instructions and parts list

basic coatings HYDROLINE B0656 Technical data

basic coatings

basic coatings HYDROLINE B0656 Technical data

Lava V.300 Premium operating instructions

Lava

Lava V.300 Premium operating instructions

FoodSaver V2240-I quick start

FoodSaver

FoodSaver V2240-I quick start

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.