Biocare iM 12 User manual

User Manual
Shenzhen Biocare Bio-Medical Equipment Co., Ltd. P/N: 02111717-01
Veterinary Monitor
Model: iM 12&iM 15


About the User Manual
This user manual is only used for the operation on this series of equipments. The company will not take any
responsibility for consequences caused by misuse of the manual.
This user manual contains proprietary information whose copyright is protected and reserved. Without license
from the company, any photocopying, replication or translation of the manual is prohibited.
For technical development and updating of the present product, the company reserves the rights to amend and
alter the contents of the manual. If the alternation does not involve the safety of the product having been sold, the
customer will not be informed individually.
For reasons of technical updating and customers‘special requirements, if the product performance is unaffected,
some assembled units may not be in accordance with the standard configurations indicated in the manual.
Version: V2.1 Date: 2018-12
All rights reserved © Shenzhen Biocare Bio-Medical Equipment Co., Ltd.

Shanghai International Holding Corp. GmbH (Europe)
Eiffestraße 80
20537 Hamburg
GERMANY
Shenzhen Biocare Bio-Medical Equipment Co., Ltd.
#16-1, Jinhui Road, Jinsha Community, Kengzi Sub-District, Pingshan New
District, 518122 Shenzhen, PEOPLE'S REPUBLIC OF CHINA
Tel: 86-755-33005899 Fax: 86-755-27960643
Website: http://www.biocare.com.cn

Content
User Manual of Veterinary Monitor --I--
Content
Preface...................................................................................................................................................................... 1
Customer Required Reading.................................................................................................................................. 3
Chapter 1 Operation Safety Information.............................................................................................................. 5
1.1 Safety Information .......................................................................................................................................... 5
1.2 Influence On the Environment and Energy Sources....................................................................................... 7
1.3 Safety Classification........................................................................................................................................ 7
1.4 Safety Requirements ....................................................................................................................................... 7
1.5 Device Identifications ..................................................................................................................................... 9
Chapter 2 Overview............................................................................................................................................... 11
2.1 Brief Introduction.......................................................................................................................................... 11
2.1.1 Applicability.......................................................................................................................................... 11
2.1.2 Contraindications................................................................................................................................... 11
2.2 Configuration Composition........................................................................................................................... 11
2.3 Main Unit...................................................................................................................................................... 12
2.3.1 Front View............................................................................................................................................. 12
2.3.2 Side View.............................................................................................................................................. 15
2.3.3 Back View............................................................................................................................................. 17
2.4 Modules......................................................................................................................................................... 17
2.5 Screen Display .............................................................................................................................................. 18
2.5.1 Main Interface........................................................................................................................................ 18
2.5.2 Interface Explanation............................................................................................................................. 18
2.6 Shortcut Key ................................................................................................................................................. 22
Chapter 3 Basic Operation ................................................................................................................................... 23
3.1 Installation..................................................................................................................................................... 23
3.1.1 Unpacking and Checking....................................................................................................................... 23
3.1.2 Environmental Requirements ................................................................................................................ 23
3.1.3 Normal Operation Conditions................................................................................................................ 23
3.2 Operation Preparation ................................................................................................................................... 24
3.2.1 A.C. Power Supply Connection............................................................................................................. 24
3.2.2 Turn On ................................................................................................................................................. 24
3.2.3 Sensor Connection................................................................................................................................. 25
3.2.4 Recorder Checking................................................................................................................................ 25
3.3 Monitoring .................................................................................................................................................... 25
3.4 Turn Off........................................................................................................................................................ 25
3.5 Input Board ................................................................................................................................................... 26
3.6 Touch Screen (Optional)............................................................................................................................... 26
3.7 Interface Setup .............................................................................................................................................. 26
3.8 Main Menu.................................................................................................................................................... 28
3.9 General Setup................................................................................................................................................ 28
3.9.1 Monitor Definition................................................................................................................................. 28
3.9.2 Screen Type Setup................................................................................................................................. 28
3.9.3 Screen Brightness and Panel Backlight Adjustment.............................................................................. 28

Content
--II-- User Manual of Veterinary Monitor
3.9.4 Time and Date Setup ............................................................................................................................. 28
3.9.5 Audio Setup........................................................................................................................................... 29
3.9.6 Help Menu............................................................................................................................................. 29
3.10 Configuration Management ........................................................................................................................ 29
3.10.1 Recent Configuration Self-recovery.................................................................................................... 29
3.10.2 Turning-on Default Configuration Setup............................................................................................. 29
3.10.3 User Configuration Storage................................................................................................................. 30
3.10.4 User Configuration Deletion................................................................................................................ 30
3.10.5 Default Configuration Manual Restore................................................................................................ 31
Chapter 4 Animal Management........................................................................................................................... 33
4.1 Admit Animal ............................................................................................................................................... 33
4.2 Quick Admit.................................................................................................................................................. 33
4.3 Edit Animal Demographics........................................................................................................................... 33
4.4 Discharge Animal.......................................................................................................................................... 33
4.5 Data Management......................................................................................................................................... 34
4.6 Central Monitoring System........................................................................................................................... 34
Chapter 5 User Interface....................................................................................................................................... 35
5.1 Interface Style Setup..................................................................................................................................... 35
5.2 Standard Interface......................................................................................................................................... 35
5.3 Minitrends Interface...................................................................................................................................... 36
5.4 BigNumerics Interface.................................................................................................................................. 37
5.5 OxyCRG Interface ........................................................................................................................................ 37
5.6 View Other Bed Interface ............................................................................................................................. 38
5.7 C.O. Measure Screen (Optional)................................................................................................................... 40
5.8 7 Lead Half Screen........................................................................................................................................ 41
5.9 7 Lead Full Screen ........................................................................................................................................ 41
Chapter 6 Parameter Monitoring ........................................................................................................................ 43
6.1 ECG............................................................................................................................................................... 43
6.1.1 ECG Measuring Principle...................................................................................................................... 43
6.1.2 Definition of ECG Monitoring .............................................................................................................. 44
6.1.3 ECG Intended Use................................................................................................................................. 44
6.1.4 Safety Information................................................................................................................................. 44
6.1.5 Monitoring Procedures.......................................................................................................................... 45
6.1.6 ECG Display.......................................................................................................................................... 47
6.1.7 ECG Setup............................................................................................................................................. 47
6.1.8 ST Analysis ........................................................................................................................................... 49
6.1.9 Arrhythmia Monitoring ......................................................................................................................... 50
6.1.10 Arrhythmia Alarm ............................................................................................................................... 51
6.1.11 ECG Relearn........................................................................................................................................ 55
6.2 Resp............................................................................................................................................................... 56
6.2.1 Resp General Description...................................................................................................................... 56
6.2.2 Resp Display.......................................................................................................................................... 56
6.2.3 Placing Respiration Electrodes.............................................................................................................. 57
6.2.4 Resp Setup............................................................................................................................................. 57

Content
User Manual of Veterinary Monitor --III--
6.3 PR.................................................................................................................................................................. 58
6.3.1 PR General Description......................................................................................................................... 58
6.3.2 PR Source.............................................................................................................................................. 59
6.3.3 Alarm Source Setup............................................................................................................................... 59
6.3.4 Pulse Volume Setup............................................................................................................................... 59
6.4 SpO2.............................................................................................................................................................. 59
6.4.1 SpO2General Description ..................................................................................................................... 59
6.4.2 Safety Information................................................................................................................................. 61
6.4.3 SpO2Module......................................................................................................................................... 62
6.4.4 Monitoring Steps................................................................................................................................... 62
6.4.5 SpO2Setup............................................................................................................................................. 63
6.4.6 Influencing Factors of Measure............................................................................................................. 65
6.4.7 Masimo Information.............................................................................................................................. 65
6.4.8 Nellcor Information............................................................................................................................... 65
6.5 NIBP ............................................................................................................................................................. 66
6.5.1 NIBP General Description..................................................................................................................... 66
6.5.2 Safety Information................................................................................................................................. 67
6.5.3 Measure Restriction............................................................................................................................... 69
6.5.4 Measure Mode....................................................................................................................................... 69
6.5.5 Measure Procedure................................................................................................................................ 69
6.5.6 NIBP Display......................................................................................................................................... 70
6.5.7 NIBP Setup............................................................................................................................................ 71
6.5.8 NIBP Leakage Test................................................................................................................................ 72
6.5.9 NIBP Accuracy Test.............................................................................................................................. 73
6.6 Temp............................................................................................................................................................. 74
6.6.1 General Description............................................................................................................................... 74
6.6.2 Safety Information................................................................................................................................. 75
6.6.3 Measure Procedure................................................................................................................................ 75
6.6.4 Measure Display.................................................................................................................................... 76
6.6.5 Temp Unit Setup.................................................................................................................................... 76
6.6.6 Alarm Setup........................................................................................................................................... 76
6.7 IBP (Optional)............................................................................................................................................... 76
6.7.1 General Description............................................................................................................................... 76
6.7.2 Safety Information................................................................................................................................. 77
6.7.3 Monitoring Steps................................................................................................................................... 77
6.7.4 IBP Display ........................................................................................................................................... 78
6.7.5 IBP Setup............................................................................................................................................... 78
6.7.6 Sensor Zero-Calibration ........................................................................................................................ 79
6.8 CO2(Optional).............................................................................................................................................. 79
6.8.1 General Description............................................................................................................................... 79
6.8.2 Side-stream CO2Module....................................................................................................................... 81
6.8.3 Main-stream CO2Module ..................................................................................................................... 83
6.8.4 CO2Setup.............................................................................................................................................. 85
6.8.5 Zero........................................................................................................................................................ 86
6.8.6 Calibrate ................................................................................................................................................ 87
6.8.7 Influencing Factors of Measure............................................................................................................. 88
6.8.8 Faulty Handling..................................................................................................................................... 88

Content
--IV-- User Manual of Veterinary Monitor
6.8.9 Emissions............................................................................................................................................... 88
6.9 C.O. (Optional) ............................................................................................................................................. 89
6.9.1 General Description............................................................................................................................... 89
6.9.2 Influencing Factors of Measurement..................................................................................................... 89
6.9.3 Setting Up the C.O. Measurement......................................................................................................... 90
6.9.4 Understanding the C.O. Display............................................................................................................ 92
6.9.5 Changing C.O. Settings......................................................................................................................... 92
6.9.6 Measuring the Blood Temperature........................................................................................................ 93
6.10 AG anesthetic gas (optional)....................................................................................................................... 93
6.10.1 Overview ............................................................................................................................................. 93
6.10.2 AG Display.......................................................................................................................................... 93
6.10.3 MAC Value ......................................................................................................................................... 94
6.10.4 Mainstream AG module ...................................................................................................................... 94
6.10.5 Sidestream AG module........................................................................................................................ 99
6.10.6 AG Setup........................................................................................................................................... 105
6.10.7 Replacing Anesthetic......................................................................................................................... 105
6.10.8 Measurement Influencing Factors ..................................................................................................... 106
6.10.9 Emissions........................................................................................................................................... 106
Chapter 7 Alarm.................................................................................................................................................. 107
7.1 General Description .................................................................................................................................... 107
7.2 Alarm Type ................................................................................................................................................. 107
7.3 Alarm Level ................................................................................................................................................ 107
7.4 Alarm Mode................................................................................................................................................ 107
7.5 Alarm Setup ................................................................................................................................................ 108
7.5.1 Global Alarm Interface........................................................................................................................ 109
7.5.2 Parameter Alarm Setup........................................................................................................................ 109
7.6 Alarm Configuration................................................................................................................................... 110
7.7 Alarm Setup ................................................................................................................................................ 111
7.7.1 Alarm Setup Interface.......................................................................................................................... 111
7.7.2 Parameter Alarm Setup........................................................................................................................ 112
7.8 Alarm Pause................................................................................................................................................ 113
7.9 Alarm Silence.............................................................................................................................................. 113
7.10 Alarm Detection and Counter Measures................................................................................................... 114
7.11 Other Bed Alarm....................................................................................................................................... 114
7.11.1 Other Bed Alarm Auto Prompting..................................................................................................... 114
7.11.2 Other Bed Alarm Silence................................................................................................................... 114
Chapter 8 Freeze and Review............................................................................................................................. 115
8.1 Enter Freeze ................................................................................................................................................ 115
8.2 Remove Freeze............................................................................................................................................ 115
8.3 Record Frozen Waveforms ......................................................................................................................... 115
8.4 Review ........................................................................................................................................................ 115
8.4.1 Review Window.................................................................................................................................. 115
8.4.2 Graphic Trends.................................................................................................................................... 116
8.4.3 Tabular Trends..................................................................................................................................... 117
8.4.4 Events.................................................................................................................................................. 118

Content
User Manual of Veterinary Monitor --V--
8.4.5 NIBP List............................................................................................................................................. 119
8.4.6 Long ECG............................................................................................................................................ 120
Chapter 9 Calculations........................................................................................................................................ 123
9.1 General Description .................................................................................................................................... 123
9.2 Medication Calculation............................................................................................................................... 123
9.3 Hemodynamic Calculation.......................................................................................................................... 125
9.3.1 Review................................................................................................................................................. 125
9.3.2 Output.................................................................................................................................................. 126
9.4 Renal Function Calculation, Oxygenation Calculation, Ventilation Calculation........................................ 126
Chapter 10 Recording......................................................................................................................................... 127
10.1 Recorder.................................................................................................................................................... 127
10.2 Record Setup............................................................................................................................................. 127
10.3 Start and Stop Recording .......................................................................................................................... 128
10.4 Install Recording Paper............................................................................................................................. 128
10.5 Cleaning of the Thermal Print Head ......................................................................................................... 129
Chapter 11 Other Functions............................................................................................................................... 131
11.1 Power-On.................................................................................................................................................. 131
11.2 Colors of the Measured Physiological Parameters.................................................................................... 131
11.3 Analog Output Setup................................................................................................................................. 131
11.4 Nurse Call System..................................................................................................................................... 131
11.5 Defibrillation synchronization................................................................................................................... 132
11.5 Manual Event............................................................................................................................................ 133
11.6 Defaults..................................................................................................................................................... 133
11.7 System State Indicator............................................................................................................................... 133
11.8 Standby Mode........................................................................................................................................... 134
Chapter 12 Battery.............................................................................................................................................. 135
12.1 General Description .................................................................................................................................. 135
12.2 Battery Installation.................................................................................................................................... 136
12.3 Battery Recycling...................................................................................................................................... 136
Chapter 13 Cleaning and Maintenance............................................................................................................. 137
13.1 Cleaning of Monitor.................................................................................................................................. 137
13.2 Disinfecting of Monitor............................................................................................................................. 137
13.3 Fan Cleaning............................................................................................................................................. 137
13.4 Storage of Monitor.................................................................................................................................... 138
13.5 Transport................................................................................................................................................... 138
13.6 Inspection of Monitor................................................................................................................................ 138
Chapter 14 Maintenance..................................................................................................................................... 139
14.1 Safety Information .................................................................................................................................... 139
14.2 NIBP Accuracy Test ................................................................................................................................. 139
14.3 NIBP Overpressure Test ........................................................................................................................... 139
14.4 NIBP Leakage Test................................................................................................................................... 139
14.5 User Maintain............................................................................................................................................ 139

Content
--VI-- User Manual of Veterinary Monitor
14.6 Demo Model ............................................................................................................................................. 141
14.7 Monitor System Information..................................................................................................................... 141
Chapter 15 Troubleshooting and Solutions....................................................................................................... 143
15.1 Check before Use...................................................................................................................................... 143
15.2 The Monitor cannot be Turned On............................................................................................................ 143
15.3 The Monitor cannot be Shut Down Normally with ON/OFF Switch........................................................ 143
15.4 No Display on Screen................................................................................................................................ 143
15.5 Interference to ECG Signal Too High or Baseline Too Coarse................................................................ 143
15.6 No Measured Result of NIBP.................................................................................................................... 144
15.7 No Measured Result of SpO2.................................................................................................................... 144
15.8 Measure Result of EtCO2is Low (Optional)............................................................................................. 144
15.9 The Sound of Sidestream CO2Pump Becomes High (Optional).............................................................. 144
15.10 Body Temperature without Numerical Value or Inaccurate Readings.................................................... 144
Appendix A Packaging and Accessories............................................................................................................ 145
A.1 Packaging................................................................................................................................................... 145
A.2 Accessories................................................................................................................................................. 145
Appendix B Product Specifications.................................................................................................................... 147
B.1 Safety Specifications.................................................................................................................................. 147
B.1.1 Product Classification......................................................................................................................... 147
B.1.2 Environment Specifications................................................................................................................ 148
B.1.3 Power Specifications........................................................................................................................... 148
B.2 Physical Specifications............................................................................................................................... 148
B.3 Hardware Specifications............................................................................................................................. 148
B.4 Data Storage............................................................................................................................................... 150
B.5 Wireless Network....................................................................................................................................... 150
B.6 Measuring Specifications ........................................................................................................................... 150
B.6.1 ECG Monitoring ................................................................................................................................. 150
B.6.2 Respiration (Resp) Monitoring ........................................................................................................... 152
B.6.3 SpO2 Monitoring................................................................................................................................. 153
B.6.4 PR Specifications................................................................................................................................ 154
B.6.5 NIBP Monitoring................................................................................................................................ 155
B.6.6 Temperature (Temp) Monitoring........................................................................................................ 156
B.6.7 IBP Monitoring................................................................................................................................... 156
B.6.8 CO2Monitoring (Optional)................................................................................................................. 157
B.6.9 C.O. Specifications(Optional)............................................................................................................. 160
B.6.10 AG Specifications (Optional)............................................................................................................ 161
B.6.11 Recorder Specifications.................................................................................................................... 162
Appendix C Alarm Information......................................................................................................................... 163
Appendix D Factory Default Setup.................................................................................................................... 169
D.1 Animal Demographics................................................................................................................................ 169
D.2 Alarm.......................................................................................................................................................... 169
D.3 Alarm Limit................................................................................................................................................ 170
D.3.1 >20kg............................................................................................................................................... 170

Content
User Manual of Veterinary Monitor --VII--
D.3.2 10~20kg............................................................................................................................................ 171
D.3.3 <10 kg................................................................................................................................................. 172
D.4 Screen Setup............................................................................................................................................... 173
D.5 User Maintain............................................................................................................................................. 173
D.6 ECG............................................................................................................................................................ 173
D.7 NIBP........................................................................................................................................................... 175
D.8 SpO2........................................................................................................................................................... 175
D.9 Resp............................................................................................................................................................ 176
D.10 IBP ........................................................................................................................................................... 176
D.11 CO2........................................................................................................................................................... 176
D.12 C.O........................................................................................................................................................... 177
D.13 AG............................................................................................................................................................ 177
D.14 PR............................................................................................................................................................. 177
D.15 Other Setup............................................................................................................................................... 178
Appendix E EMC- Guidance and Manufacture’s Declaration..................................................................... 179
E.1 Guidance and manufacture’s declaration-electromagnetic emissions for all EQUIPMENT and SYSTEMS
........................................................................................................................................................................... 179
E.2 Guidance and manufacture’s declaration-electromagnetic immunity for all EQUIPMENT and SYSTEMS
........................................................................................................................................................................... 179
E.3 Guidance and manufacture’s declaration-electromagnetic immunity for EQUIPMENT and SYSTEMS
those are not LIFE-SUPPORTING................................................................................................................... 180
E.4 Recommended separation distance between portable and mobile RF communications equipment and the
EQUIPMENT or SYSTEM for EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING...................... 181

Content
--VIII-- User Manual of Veterinary Monitor
-- The Blank Page --

Preface
User Manual of Veterinary Monitor --1--
Preface
Thank you for using the monitor produced by our company.
In order to help you master the operation on this series of monitors as soon as possible, a user manual (the current
manual) has been attached to this series of products. It is strongly recommended you read it before you install and
use the product for the first time.
For performance and reliability improvements, some alternations will be made to the equipment (including the
hardware and software) by the manufacturer at times. On that occasion, although some information will be altered
or added, there is still a possibility of mismatch between the description in the manual and the product. Thank you
for understanding. For any error and omission in the manual, your notification is welcomed.
This series of monitors contain iM 12 (Named as 12 inches monitor in the following) and iM 15 (Named as 15
inches monitor in the following).
Manual Abstract
[Main composition and performance]
This series of monitors are mainly composed of a host processor and other corresponding functional accessories
(including electrocardiograph (ECG) lead cables, blood pressure cuffs and catheters, invasive pressure monitoring
sensors (optional modules), blood oxygen probes, body temperature probes, EtCO2measurement components
(optional modules), and anesthesia gas measurement components (optional modules).)
[Scope of application]
This series of monitors application is for monitoring animals‘electrocardiogram, respiration, pulse rate, heart rate,
pulse oxygen saturation, body temperature, non-invasive blood pressure, invasive blood pressure (optional),
EtCO2(optional), and Anesthetic gas (optional) in a medical therapy unit.
[Cautions, warnings and suggestion]
1) This series of monitors do not have components for the customer‘s self-maintenance. When something is out
of order, please don‘t disassemble on your own.
2) This series of monitors do not belong to treatment facilities, and cannot be applied for household use.
3) The optional modules involved can be equipped according to the customers‘need. The required equipment
has been preset by the manufacturer before this series of monitors come from the factory.
4) Don‘t allow contact to the animal, hospital bed or monitor during defibrillation.
5) Please turn off the power before cleaning this series of monitors.
6) Don‘t use this series of monitors under the condition of a high temperature, high humidity, high flammability,
high dust, or electromagnetic radiation.
7) Please keep the mains power source and grounding situations safe and stable.
8) (For other information, please refer to the manual.)

Preface
--2-- User Manual of Veterinary Monitor
-- The Blank Page --

Customer Required Reading
User Manual of Veterinary Monitor --3--
Customer Required Reading
This section will tell you what operative procedures should be paid close attention to, how to avoid abnormal
operation, and what possible detrimental risks might occur to this series of monitors or animal when you use this
series of monitors.
The company‘s Imperatives: please read the manual thoroughly before using this series of monitors, and perform
the operative procedures according to the instructions described in it; The company won‘t take any safe, reliable
or performance guarantee responsibility for abnormal monitor phenomena or animal body injures caused by
violation of the requirements concerning monitor application or maintenance indicated in this section and the user
manual, nor offer free maintenance for such breakdowns. Once again, the company reminds you to read the
contents of the current section and the manual before use.
For <10 kg animals and >20 kg animals model is prohibited for blood pressure measurement. Or,
the pressure may cause limb injuries, or even limb necrosis.
This series of monitors can only be used for one animal at a time.
Customer required reading blood pressure monitoring is prohibited for animals with a serious
hemorrhage tendency or sickle cells. Or, local hemorrhage may occur.
A cuff is prohibited for an infused or intubated limb or an area with local skin injuries. Or, it may lead
to limb injuries.
Continuous use of the ear-clamping or tongue-clamping pulse oxygen sensor may cause discomfort
or a pressure pain, especially for animals with microcirculation disturbance. No more than two-hour
clamping for the same finger is recommended.
More careful inspection of the pulse oxygen sensor measurement site should be done for animals
with special needs. The sensor cannot be placed on edematous or fragile tissues.
This series of monitors must be well grounded in order to prevent possible electrical danger as well
as to secure a good ECG signal quality.
Although all animal contact parts of this series of monitors have been approved by bio-compatibility
tests, some individuals may still have allergies to monitors parts. The application of this series of
monitors must be stopped for animals that have allergies to monitors.
All measurement cords and plastic tubes should be kept away from the animal‟s neck in order to
avoid asphyxia caused by neck winding.
Accessories cannot be replaced indiscriminately. When accessory replacement is necessary, an
accessory of the same type provided by the manufacturer or approved for this series of monitors
should be used. Only the accessories from the same manufacturer and of the same type can be
used for accessory replacement. Or, adverse consequences of safety and bio-compatibility may
occur.
Do not open the pulse oxygen sensor and look directly at the light device (as the infrared light can‟t
be detected by eyes). The warning also applies to maintenance persons. The light may harm to your
eyes.
If this series of monitors falls accidentally, its use must be stopped. Only after safety and technical
index tests prove this series of monitors is still operational, it can go on to being used.
When blood pressure is measured, the manual mode is recommended by the manufacturer. If the
automated or continuous model is selected, a qualified observer should be present.
For animals with pacemakers, heart rate meter may be in asystole or arrhythmia the pacemaker

Customer Required Reading
--4-- User Manual of Veterinary Monitor
pulse count. Do not rely solely on heart rate alarm. Should be closely monitoring animals with
pacemaker.
Do not modify this series of equipments without authorization of the manufacturer. If this series of
equipments are modified, appropriate inspection and testing must be conducted to ensure
continued safe use of the equipment.
Please read clinical limitations and contraindication information carefully.

Chapter 1 Operation Safety Information
User Manual of Veterinary Monitor --5--
Chapter 1 Operation Safety Information
1.1 Safety Information
Warning: emergencies may be caused which may lead to death, severe bodily injuries or property
loss if you do not follow this advice.
Explanation: instructions or explanations are provided for better use of this series of product.
Attention: important information and prompts are included, which may lead to slight bodily injuries or
breakdowns of the product if you do not follow them.
Attention
For the sake of safety as well as more effective use of this series of monitors, please read the user
manual carefully to thoroughly know the correct operative method.
Warning
This series of monitors must be placed on a smooth and flat worktable. Strong vibration or impact
should be avoided when being moved. Please check the device, connection wires and accessories
before use to be sure that they work normally and safely.
Make sure that the frequency and voltage of the A.C. power source satisfies the requirements, and
has enough capacity. This series of monitors can only be connected to an outlet with a grounding
wire. If the outlet is not connected to a grounding wire, please use the battery for power supply
instead of the outlet.
Make sure that the room has a good power supply system and a good ground circuit, or, injuries
may occur to the animal.
The electrodes and their connectors as well as the accessories should not have contact with other
conductoring wires including the ground.
Do not open the outer shell of the device, or, electric shock may occur. The maintenance and
upgrading of this series of monitors can only be performed by the maintenance persons trained or
authorized by the company.
When there is a doubt about the integrity of the grounding wire, the battery (DC power) in the
machine should be used.
Do not touch the animal or the hospital bed when this series of monitors and a cardiac defibrillator
are being used. All electrodes connected and unconnected to the animal, as well as the animal
himself, need not be grounded. For the protection from the defibrillator discharge procedure, please
use the cable provided by the company. This series of monitors are not recommended to be
concurrently used with other electrical stimulators. If this is necessary, it should be done under
direct guidance of specialized technicians.
Be cautious when the animal is connected with more than one instrument, because the total leak
current may be harmful to the animal. Devices in compliance with the standard of IEC60601-1 are
allowed to be connected to this instrument, and the total leak current should be measured by the
users to determine that if it meets the requirement and can be used after connection.
The signal input/output ports (when needed to use) are only permitted to be connected with devices
which compliance with the standard of IEC 60601-1 when used within the animal environment, and
compliance with the standard of other IEC or ISO standard when used outside the animal
environment, the composition of the system should comply with the requirements of IEC 60601-1-1.

Chapter 1 Operation Safety Information
--6-- User Manual of Veterinary Monitor
In order to prevent burns, a high frequency electrosurgical should be kept far away from the
electrodes. The electrical resistance between the electrosurgical and the animal‟s body should be
as small as possible and great caution should be used.
The alarm sound volume and limits should be set up according to the animal‟s actual status.
Veterinary Monitoring cannot only depend on the sound alarm system. When the sound volume is
tuned down to the minimum, it may place the animal in danger. Therefore, close attention should be
paid to the animal‟s actual clinical status.
The physiological waveform, physiological parameters and alarm information displayed by this
series of monitors can only be used as reference by the physician. They cannot be directly used as
the basis for clinical treatment.
If there are any anomalies during use, please turn off this series of monitors immediately for
examination.
Please place the power source and all types of accessory electric cables carefully, in case of they
entangle the animal; the winding may even cause the animal‟s asphyxia, and electrical disturbance
between them.
Handling of the packing materials should follow the associated local regulations, or the hospital
waste treatment rules. The packing materials should be placed out of reach of children.
Explanation
For the sake of the animal‟s safety, please use the accessories specified in the user manual.
When the device and its accessories are nearing the expiration date for use, they should be
disposed of according to associated local regulations or hospital rules.
Electromagnetic fields can influence the performance of this series of monitors. Therefore, any
device used nearby should meet the corresponding EMC requirement. A mobile telephone, X rays
and MRI equipment are all likely to be an interference source as they can emit high-intensity
electromagnetic radiation.
Before the power source is switched on, please be sure that the voltage and frequency satisfy the
requirements indicated on the label attached to the device or in the user manual.
Please install or carry the device appropriately to avoid device damage caused by falls, collision,
strong vibration and other external mechanical forces.
The device and its accessories should be checked and calibrated regularly, or, the technical
specifications in the user manual may not be obtained.
Attention
Please install the device at a place where the observation, manipulation and maintenance of the
device is convenient.
Please put the user manual near the device for convenient and quick reference when necessary.
It should not to position the equipment so that it is difficult to operate the disconnection device from
the supply mains.
The software of this device is developed according to the IEC60601-1-4 standards, which has

Chapter 1 Operation Safety Information
User Manual of Veterinary Monitor --7--
minimized the possibility of the risks caused by programming errors.
The user manual introduces the product according to its most complete configurations. Therefore,
your purchased product may lack some configurations or corresponding functions.
1.2 Influence On the Environment and Energy Sources
Handling of the packing materials, exhausted battery and scrap materials should be carried out according to local
regulations. The user should carry out reasonable handling for the scrap product and materials according to the
local laws and regulations, and offer possible help for waste classification and recycling.
1.3 Safety Classification
This series of monitors belong to the following types:
1) Based on shockproof types:
Class I, internal power supply.
2) Based on shockproof levels:
Type BF (*) applied parts: EtCO2measurement module (optional), and anesthetic gas module (optional).
Type CF (*) applied parts: ECG (Respiration) measurement, IBP measurement module (optional), the NIBP
measurement module, Temp measurement module, SpO2 measurement module and C.O. measurement
module.
(Attention: * represents an anti-defibrillation function.)
3) Based on the levels of protection from noxious liquid infiltration:
A common closed device which cannot prevent noxious liquid infiltration.
4) Based on safety levels in an atmosphere of easily flammable anesthetic gas mixed with air or with oxygen or
nitrous oxide:
This series of monitors cannot be used in an atmosphere of easily flammable anesthetic gas mixed with air or
with oxygen or nitrous oxide.
5) Based on duty:
A continuously-running equipment.
1.4 Safety Requirements
Animal Number
This series of monitors can only monitor one animal at a time.
Interference
Do not use a mobile telephone near this series of monitors as the high-intensity electromagnetic interference
emitted from it may strongly influence the normal operation of this series of monitors.
Water Exposure Prevention
This series of monitors must be protected from water exposure in order to prevent electrical shock and to
reduce equipment breakdown. If water enters accidentally, the use of this series of monitors should be
stopped immediately. It can only be used again after maintenance by specialized technicians.
Accuracy
When there is a doubt about any parameter displayed or printed, please adopt another method to determine
the animal‘s physiological parameter. Insure that your monitor works accurately.
Alarm
The monitoring process cannot only depend upon the sound alarm. Tuning-down or turning-off of the alarm
sound volume may place the animal in danger. Caution: the most reliable monitoring can only be done with
close monitoring of the animal combined with correct use of the monitoring device.

Chapter 1 Operation Safety Information
--8-- User Manual of Veterinary Monitor
Attention: the alarm function of monitoring devices should be checked regularly.
Before Use
All connection cables should be carefully checked before use. Any damaged cable or connector should be
replaced immediately.
Cable
The cables should be kept away from the animal‘s neck in case of entanglement.
Data Clearing
When this series of monitors are used for another animal, the preceding animal‘s data should be cleared. You
can clear the data by selecting [Main Menu], [Animal Management] and [Data Clearing], sequentially, and
then pressing Confirmation.
Packing Materials Handling
Packing materials handling should observe local rules on waste management. They should be kept out of
reach of children.
Explosion Hazard
Do not use this series of monitors under the condition of flammable gas, vapor or liquid.
Leakage Current Test
When this series of monitors monitor are concurrently used with other devices, the leakage current should be
tested by specialized technicians. Only after safety is secured, can it be used for the animal.
Battery
This series of monitors have been equipped with batteries, which lose electricity even after the monitor has
been turned off. Therefore, the battery should be fully charged and then taken out before the monitor is stored
so that the service life of the battery is not shortened.
Accessory and Equipment Handling
Disposable accessories can only be used once. Repeated use can lead to performance reduction and
cross-contamination.
Service Life
The service life for this series of monitors is five years. After the service life, this series of monitors and its
accessories should be disposed of according to the associated laws and regulations. If you have any question
about their handling, please contact the manufacturer or the agency.
Operation Instructions
For the sake of continuously safe use of this series of monitors, please operate this series of monitors
according to the instructions. However, these operation instructions can by no means substitute for the
accepted medical practical experience in animal nursing.
Data Loss
This series of monitors have the possibility of data loss at any time. Before this series of monitors return to
normal, please monitor the animal closely, or use other equipment. If the monitor cannot return to normal
within 60 seconds, please turn off the power and restart the monitor. After the monitor returns to normal,
please check its monitoring status and alarm functions.
This manual suits for next models
1
Table of contents
Other Biocare Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Iron Bow
Iron Bow Poly CLINiC Gen 2 Installation and user guide

Mizuno
Mizuno MOT-VS600 Operator's manual

Gima
Gima OXY-PED Use and maintenance book

EMDR Kit
EMDR Kit CLASSIC English Manual

Top shelf Orthopedics
Top shelf Orthopedics Wrist/Thumb Lace-Up Supports Instructions for use

Mizuho
Mizuho Evans DT Series Instructions for use