manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. bluelab
  6. •
  7. Measuring Instrument
  8. •
  9. bluelab Guardian Monitor Wi-Fi User manual

bluelab Guardian Monitor Wi-Fi User manual

Guardian Monitor Wi-Fi
Manual
Your key growing parameters - at a glance,
anywhere, any time.
English
2
Contents
Features
2
Important –pH probe care
4
Installation / Mounting
5
Selecting units
6
Setting brightness
6
Connecting to Wi-Fi
7
Connecting to a new network
8
Enable / Disable Wi-Fi
8
pH Calibration
9
Placement of probes
12
Setting alarms
13
Cleaning the conductivity probe
15
Cleaning the pH probe
16
Hydrating the pH Probe
17
Trouble shooting guide
18
Technical Specifications
19
Features
Connected mode via Wi-Fi with Bluelab
Edenic software
Easy connection onboarding using Bluetooth
from your mobile device
Measure pH
Measure conductivity and temperature
2 metre / 6.6 foot standard cables
110 - 240 VAC (Mains powered)
4 interchangeable plug types (USA, Euro, UK,
NZ/AUS)
Replaceable Bluelab pH Probe (BNC connection)
Includes screws for wall or rack mounting
Non-volatile memory - keep settings through
power loss
Plant-safe green LEDs with adjustable
display brightness
Flashing high and low alarms
English
3
What is Plant Safe? Green LED’s are safe for continued growth during a plant’s
fruiting stage when hours of darkness are required.
Guardian Monitor Wi-Fi
ATTENTION.
Keep your pH probe tip wet
to avoid permanent damage
Conductivity Units
Temperature Units
pH and Calibration
Settings
Alarms
English
4
2.0 IMPORTANT - Bluelab pH Probe care
pH probes DO NOT last forever. They age through normal use and will eventually fail. The
life time of a pH probe depends on the environment it is used in and the way that it is
treated. To receive a long life from your pH probe, ensure you follow the guide below.
pH probes contain glass and are therefore FRAGILE. With good care, they will give a long
service life.
Bluelab pH Probe
DO NOT let the pH probe tip dry.
DO NOT bend the probe; this will break its internal glass tube.
DO NOT knock the probe; this will break its internal glass tube or external glass bulb.
DO NOT plunge a cold pH probe into a hot liquid - sudden temperature changes can crack
the glass and permanently damage the probe.
DO NOT immerse in oils, proteins or suspended solids that will leave a coating on the glass
bulb.
Always remove pH probe storage cap before use
1. Grip the top of the cap and gently twist the base one rotation
clockwise to loosen slightly.
2. Next slowly slide the cap off the pH probe. DO NOT
completely remove the base of the cap from the top of the
cap.
Storing the pH probe
When storing the pH probe, the pH probe tip must be kept moist.
To prepare the pH probe for storage, add enough Bluelab pH
Probe KCl Storage Solution to the storage cap so the probe tip is
covered. Then replace the cap and store in a secure place. DO
NOT use RO (Reverse Osmosis), Distilled or De-ionized water. Pure
water changes the chemistry in the reference, causing the probe
to die.
If the pH probe has been accidentally allowed to dry out;
The pH probe must be ‘hydrated’ for 24 hours in KCl storage solution.
Removing pH probe
storage cap
Ensure probe tip is
covered
by the KCl storage solution
BNC Adaptor
pH probe
storage cap
Glass tube inside
plastic barrel
Glass bulb
pH probe holder
suction cup
pH probe tip
English
5
3.0 Mounting on a wall
Select a suitable location that is less than 2 meters
from your reservoir, and less than 1.5 meters from an
electrical mains outlet. NOTE: Avoid placing the
Guardian Monitor where it can be damaged by direct
sunlight, water or nutrient salts.
Option 1
Mounting keyholes top and bottom
Option 2
Mounting holes top and bottom
Option 1
a) Install two fasteners (top and bottom) to the wall
172mm / 6.¾” apart.
b) Slide the Guardian Monitor onto the fasteners
using the keyholes.
Option 2
a) Fix the four fasteners through each mounting
hold in each corner of the case.
3.1 Mounting on a support or bar
Select a suitable location that is less than 2 meters from your reservoir, and less than 1.5
meters from an electrical mains outlet. NOTE: Avoid placing the Guardian Monitor where it can
be damaged by direct sunlight, water or nutrient salts.
Fix each of the cable ties provided through the small mounting holes in the top of the case
and wrap around the support or bar and secure.
4.0 Install plug adaptor
Select the appropriate mains plug adaptor for your
country.
Connect the plug adaptor to the power supply.
5.0 Connect power adaptor and Bluelab pH Probe
Connect the power adaptor into the base of the
Guardian Monitor where ‘Power Input’ is labelled. Plug
the adaptor into mains power supply.
Connect the pH probe to the Guardian Monitor by lining
up the lugs of the BNC fittings.
Fasten securely by pushing the pH probe
connector on and twisting one quarter turn.
English
6
6.0 Select conductivity unit
Conductivity readings can be displayed in EC, CF, TDS or ppm 700. The selected unit is
indicated by one of the 3 LEDs next to the conductivity window. When the EC/CF LED is on,
the Guardian Monitor is displaying conductivity in units of EC or CF.
If there is a decimal point present in the conductivity display, the units are EC. If there is
no decimal point the units are CF.
Press and hold the conductivity unit button
and after 3 seconds it will change to the next
unit indicated by the glowing LED.
Release.
Repeat steps 1 and 2 until the desired unit is
selected.
7.0 Select temperature unit
Temperature readings can be displayed in either degrees celsius (°C) or degrees
fahrenheit (°F). The selected unit is indicated by one of the two LEDs next to the
temperature window.
Press and hold the temperature unit button
and after 3 seconds it will change to the next
unit indicated by the glowing LED.
Release.
Repeat steps 1 and 2 until the desired unit is
selected.
8.0 Select brightness level of the LED displays
Brightness can be adjusted to best suit the light environment. There are 10 levels of
brightness that can be selected.
Press the Settings button.
Then press the pH/Cal button which
corresponds to the [br] brightness menu.
Press the up and down arrows to select the
required brightness. Brightness levels are
shown in the pH window.
When happy with brightness, press the
settings button to complete the change.
English
7
9.0 Connect Guardian Monitor to Wi-Fi for the first time
To connect the device to Wi-Fi, it is helpful to have the onboarding QR code available for
scanning. This can be found on the back of the device or on the packaging.
Have the security code QR image handy.
When first powered on, the Guardian Monitor Wi-Fi automatically starts in Provisioning mode.
This allows Wi-Fi settings to be configured from a mobile device. Once the Guardian Monitor
has been connected to a Wi-Fi network it will no longer start in Provisioning mode. If you need
to connect to a new network see section 10.
1. Ensure Bluetooth is enabled on your mobile device and the Edenic app is installed.
2. Open the Edenic app on your mobile device and log in.
3. Remove the power cable for a few seconds then reconnect.
4. In the Edenic app, click the add device button in the bottom right.
5. Select scan a QR code.
6. Scan the QR code provided on the back of your Guardian / packaging.
7. Enter a device name to help you recognise this Guardian Monitor.
8. Wi-Fi details –Click the view Wi-Fi networks to see available networks.
9. Select the Wi-Fi network you want to connect to.
10. Enter the Wi-Fi network password.
11. Select continue, your device will now be ready and connected.
English
8
10.0 Connect to a different Wi-Fi network
On your Guardian device, enter provisioning mode:
1. Press the settings button.
2. Press the Temperature unit button which
corresponds with the Prov (Provisioning) menu.
3. The device is now ready to accept new Wi-Fi
settings from your mobile device.
On your mobile device (while Prov and the security code is displayed on your Guardian):
4. Ensure Bluetooth is enabled on your mobile device and the Edenic app is installed.
5. Open the Edenic app on your mobile device and log in.
6. In the Edenic app, click the add device button in the bottom right.
7. Select search Bluetooth and pick the Guardian from available devices listed.
8. Enter a device name to help you recognise this Guardian Monitor.
9. Enter the 4-digit security code, the code can be found in the following three ways:
I. Shown in the Temperature window of the Guardian (while in PROV mode).
II. On the QR code sticker on the back of your Guardian
III. On the QR code on the packaging.
10. Select the Wi-Fi network you want to connect to.
11. Enter the Wi-Fi network password.
12. Select continue, your device will now be ready and connected.
11.0 Enable / Disable Wi-Fi and data
By default, Wi-Fi is enabled. If you want to disable Wi-Fi follow these instructions.
1. Press the settings button.
2. Press the Conductivity units button.
3. Use up and down arrows to toggle Wi-Fi on/off.
4. If turned on, the connection status is shown in
the third window.
a. conn –Wi-Fi is connected
a. Err1 - Wi-Fi not connected
b. Err2 - No internet available
c. Err3 - Edenic service not reachable
d. Err4 - Time sync, please wait
5. Once On or Off is selected, press the settings
button to complete.
English
9
12.0 pH calibration
pH calibration is required before first use to ensure that the first reading is accurate.
•The Bluelab Conductivity/Temperature Probe DOES NOT require calibration.
•The pH of the Guardian Monitor Wi-Fi DOES require calibration.
For accurate pH readings the pH probe is cleaned and recalibrated when:
•The reading is different to what you were expecting.
•The Guardian Monitor Wi-Fi is reset to factory default.
•The pH probe is replaced with a new one.
•It has been a month since the last successful calibration and the calibration indicator
lights are flashing.
When calibrating the pH after first use the pH probe needs to be cleaned. See pH probe
cleaning in section 14.0.
For best pH calibration
pH reading accuracy is dependant on the accuracy and age of the calibration solutions used,
and use and cleanliness of the pH probe tip.
•Ensure the pH probe has been cleaned and rinse the pH probe with clean water between
calibration solutions to reduce contamination of the pH solutions.
•Only fresh uncontaminated solutions should be used.
•Calibrate the pH at the same temperature as the solution to be measured.
•ALWAYS calibrate the pH probe with pH 7.0 then pH 4.0 or pH 10.0.
NOTE: The conductivity/temperature probe does not need to be calibrated, but must be
cleaned to remove any build up of nutrient salts. See Section 13.0.
The pH calibration involves cleaning the pH probe tip and then calibrating in TWO SOLUTIONS.
If a reading below pH 7.0 is expected, use pH 7.0 and pH 4.0 calibration solutions.
If a reading above pH 7.0 is expected, use pH 7.0 and pH 10.0 calibration solutions.
Follow the steps on the following page for Guardian Monitor pH calibration.
Storage and use of calibration solutions
English
10
•Always place the lid back onto the bottle after use or evaporation will occur rendering the
solution useless.
•DO NOT measure directly into the bottle. Tip a small
amount into a clean container and discard after use.
•Never add water to solutions.
•Store in a cool place.
pH reading accuracy is dependant on the accuracy and age of the
calibration solutions used, and use and cleanliness of the pH probe tip.
12.0 pH calibration cont.
To calibrate the pH probe
Clean EC and pH probe tips.
See section 15.0 (the pH probe does not require cleaning
before the first use).
pH 7.0 calibration
a) Prepare a small amount of pH 7.0 calibration solution
in a container.
b) Ensure the Guardian Monitor Wi-Fi
is plugged in and in operation mode.
c) Rinse EC and pH probe tips in fresh water, shake off
excess water and place them in pH 7.0 calibration
solution. Wait
at least one minute for reading to stabilize.
d) Press and hold the pH cal button for a few seconds
until ‘PH’ and ‘CAL’ appear in the windows, then
release the pH cal button. The ‘7’ LED will glow green.
All three windows will return to monitoring mode
with all current values being displayed.
e) If ‘Err’ is displayed see the troubleshooting guide on
section 18.0.
pH 4.0/10.0 calibration
a) Prepare a small amount of pH 4.0 or
pH 10.0 calibration solution in a container.
b) Rinse EC and pH probe tips in fresh water, shake off
excess water and place them in either pH 4.0 or pH
Clean the pH probe
pH 7.0
calibration/
solution
pH
Rinse in tap water
pH 4.0
calibration/
solution
pH
English
11
10.0 calibration solution. Wait at least one minute for
reading to stabilise.
c) Press and hold the pH cal button for a few seconds
until ‘PH’ and ‘CAL’ appear in the windows, then
release the pH cal button. The 4/10 LED will glow
green. All three windows will return to monitoring
mode with all current values being displayed.
d) If ‘Err’ is displayed see the troubleshooting guide on
section 18.0.
e) The Guardian Monitor Wi-Fi is now calibrated and
ready for use.
Successful pH 7 and pH 4
calibration
12.0 pH calibration cont.
The LEDs next to the pH window will let you know the status of calibration. See the table
below.
NOTE: The pH probe and hence calibration will eventually fail due to;
•Contamination and age of pH probe.
•pH probe used to measure solutions at temperatures above 50 °C (122 °F) or below
0 °C (32 °F).
•pH probe exposed to aggressive chemicals.
•Internal damage to the pH probe from rough treatment.
•Damage to the cable of the pH probe from rough treatment.
•pH probe repeatedly drying out.
•Moisture getting inside the BNC connector on the pH probe cable.
pH LEDs
Using factory default calibration values. Both LEDs off. Readings may be
unreliable.
pH 7 calibrated OK. Using factory default for pH 4/10. pH 7 calibrated OK. Using
factory default for pH 4/10. Readings may be unreliable.
pH 7 and pH 4 or pH 10 calibrated OK.
Flashing
30 days passed since last full calibration - calibration due.
Flashing
If 7 is flashing and 4/10 is off, calibration is also required as 4/10 was never
calibrated.
English
12
13.0 Placement of probes
Both the Bluelab pH Probe and Bluelab Conductivity/Temperature Probe require
submersion in the liquid for a measurement to occur.
•Do not pour concentrated nutrient solution or pH adjuster directly onto probes when in
the reservoir, as very strong acid may damage the probes or your alarms (if set) may
trigger.
Fit the (optional) pH probe holder to the stem of the pH probe
using a gentle twisting motion.
Place the pH probe into the reservoir and push the suction cup
onto the side of the reservoir but far enough down so the pH
probe tip is in the solution. The holder prevents damage to the
pH probe from banging onto the side of the reservoir with
movement of the solution.
Place the conductivity/temperature probe into the reservoir
selected.
English
13
14.0 Set alarms (optional)
The alarm function warns you when the solution
deviates from the desired levels you have chosen
for each measurement.
When an alarm condition is present, the display
of the affected measurement will flash.
If the measurement changes back to within the
limits you have chosen, the flashing will stop.
Even though the resolution of the Guardian is 10 ppm / TDS, the alarms will set only in 50
ppm and 70 ppm increments.
There are two ways to set the alarm, ‘quick-set’ or ‘detailed-set’.
Alarm ‘quick-set’
This allows you to quickly set the ‘alarm HIGH’ and ‘alarm LOW’ values for all of the three
measurements. The table below shows the values that are pre-set when ‘quick-set’ is used.
NOTE: Before you use the quick-set function, the solution in the tank/reservoir must be
adjusted to the desired/actual levels for all three parameters and the levels displayed in
each of the windows. If this is not done you should use the alarm ‘detailed-set’.
To activate the ‘quick-set’ alarm;
Press and hold the alarm button until ‘AL H’ is displayed in
all three windows.
Release the alarm button.
Press the Settings button once.
All three windows will display Auto for
1 second and then SAVE will be displayed in the conductivity
window. The alarms are now auto set and the alarm mode
will be turned on. The alarm LED will glow.
English
14
14.0 Set alarms (optional) cont.
Turn alarm ON/OFF
Press the alarm button to change between alarm ON and alarm OFF.
When the alarm is ON the alarm LED will glow.
NOTE: If you press and hold the alarm button you will enter the
alarm setting mode.
Alarm ‘detailed-set’
To set the detailed ‘alarm HIGH’ and ‘alarm LOW’ values;
Press and hold the alarm button until ‘AL H’ is
displayed in all three windows.
Release the button and the current ‘alarm HIGH’
values are displayed in each window.
To set conductivity alarm high value, press the
conductivity unit button then the up and down
buttons to change its value.
To set temperature alarm high value, press the
temperature unit button then the up and down
buttons to change its value.
To set pH alarm high value, press the pH unit button
then the up and down buttons to change its value.
Press the alarm button to now select the low alarm
values.
Set the low alarm values the same way as you did the
high alarms.
Press the alarm button. SAVE is briefly displayed in
the conductivity window and the Guardian Monitor
will return to its normal monitoring mode.
NOTES: To review current alarm settings without changing the values, repeatedly press the
alarm button to step through the high and low alarms.
If you want to exit the alarm settings mode without keeping any changes you have made
DO NOT press any keys. The Guardian Monitor will ‘timeout’ and return to its normal
monitoring mode without saving any changes after 1 minute.
English
15
15.0 Cleaning the conductivity/temperature probe
Cleaning the Bluelab Conductivity/Temperature Probe
Periodically cleaning ensures accurate readings.The conductivity/temperature probe is
cleaned using the Bluelab Conductivity Probe Cleaner, or “Jif” a trade name for a liquid
scourer cream used in home bathrooms and kitchens. Never use scented varieties as they
contain oils that contaminate the conductivity/temperature probe.
Remove shroud.
Warm the shroud in your hand for a few seconds
to help with removal.
Clean the conductivity probe face.
Place one or two drops of Bluelab Conductivity
Probe Cleaner onto the probe face and rub with
the Bluelab Chamois or your finger firmly and
vigorously.
Rinse the conductivity probe face.
Rinse off all traces of cleaner under running tap
water while scrubbing the probe face with the
other side of the Bluelab Chamois or the same
finger.
Check that the water forms a smooth film on the
probe face. Ensure you have a clean, smooth film
without any beads of water. If you have beads of
water, repeat steps 2 and 3.
Refit the shroud and test in 2.77 EC Conductivity
Standard Solution to ensure adequate cleaning.
Place the probe tip into the solution, wait for the
reading to stabilize to a constant value. This can
take a few minutes while the probe adjusts to the
temperature of the solution.
Repeat the cleaning process if the reading given is
not within 0.1 EC, 1 CF, 50 ppm or 70 ppm of the
values in the table below.
Testing the Bluelab Conductivity/Temperature Probe
The conductivity/temperature probe is tested in either
2.77EC/27.7CF/1385 ppm or 1940 ppm solution
depending on the unit of conductivity chosen.
Use the standard solutions in the table to the
right. Bluelab solutions are recommended.
NOTE: The shroud MUST be left on the probe when taking readings.
2.77 EC Conductivity
Uneven beads of
water
Clean, smooth film
of water
English
16
16.0 Cleaning the Bluelab pH Probe
To ensure accurate readings the pH probe tip needs to be rinsed in water after each use
and cleaned prior to calibration using the following instructions.
The storage cap must always be put back on after cleaning. Always ensure it contains
enough Bluelab pH Probe KCl Storage Solution to cover the probe tip.
Remove storage cap from pH probe.
Hold the top of the storage cap, twist the cap to
loosen then remove.
Rinse pH probe tip under fresh tap water.
Never use RO (Reverse Osmosis), Distilled or De-
ionized water.
Fill a small plastic container with clean tap water.
Add a small amount of Bluelab pH Probe Cleaner
or mild detergent (dishwashing liquid).
Gently stir the probe tip in the mixture.
Ensure that you do not ‘knock’ the soil pH probe
on the side of the container as this may cause
damage to the probe.
If the probe tip requires removal of heavy
contamination: Gently brush around the
glassware with a few drops of Bluelab pH Probe
Cleaner or mild detergent (dishwashing liquid) and
a soft toothbrush.
Rinse well under fresh running tap water to
remove all traces of the detergent mixture.
Calibrate pH probe after cleaning, see section
12.0.After calibration, store pH probe in the
storage cap, ensuring there is enough KCl Storage
Solution to cover the probe tip.
English
17
17.0 Hydrating the pH probe
Hydrate the pH probe in Bluelab pH Probe KCl Storage Solution when:
•the probe tip has not always been stored in KCl storage solution, to improve the
reading response speed.
•
the probe tip has been accidentally allowed to dry out.
Never use RO (Reverse Osmosis), De-ionized or Distilled water. Pure water changes the
chemistry in the reference, causing the probe to die.
Loosen, then remove the storage cap. Place the pH
probe upright in a plastic container.
Bluelab pH
Probe KCI
Storage Solution
Clean the pH probe tip. Ensure the probe tip is
cleaned before hydrating. See section 16.0 for
instructions.
Add enough Bluelab pH Probe KCl Storage Solution to a
plastic container to submerge the pH probe tip.
Leave to soak for at least 24 hours. After hydration,
always calibrate the pH probe to ensure accuracy, see
section 12.0.
English
18
18.0 Troubleshooting guide
Trouble
Reason
Correction
Nutrient reading low
Contaminated conductivity/temperature probe.
Clean conductivity/temperature probe
(see section 15.0).
Temperature of conductivity/temperature probe
and solution different.
Wait 5-10 minutes for conductivity/temperature
probe to reach solution temperature.
Temperature reading
inaccurate
Temperature of conductivity/temperature probe
different to solution temperature.
Wait 5-10 minutes for conductivity/temperature
probe to reach solution temperature.
pH reading inaccurate
Contaminated pH probe / glassware not clean.
Clean pH probe (see Section 16.0); then calibrate
(see Section 12.0).
Using factory default calibration.
Calibrate pH probe (see section 12.0).
Calibration old.
Calibrate pH probe (see section 12.0).
Broken glass bulb, tube or connector.
Check pH probe for damage.
pH probe damaged or old.
Replace pH probe.
Display shows ‘Err’ during
calibration
An attempt was made to calibrate with pH 4.0 or
pH 10.0 calibration solution more than one hour
after calibration with the pH7.0 solution.
Calibrate to pH 7.0 again then do pH 4.0 / pH 10.0
calibration within one hour.
Old or contaminated solutions used for
calibration.
Use fresh calibration solutions.
Dirty or contaminated pH probe.
Clean pH probe (see section 16.0).
pH probe tip been allowed to dry.
Hydrate pH probe (see section 17.0).
pH probe damaged or old.
Replace pH probe.
No display
Mains not switched on.
Switch mains power on.
Power adaptor not plugged into the Guardian
Monitor Wi-Fi.
Plug power adaptor into the base of the Guardian
Monitor Wi-Fi labelled ‘Power Input’.
pH displays ‘or’
pH displays ‘ur’
Over range pH.
Under range pH.
Solution > 14.0 pH.
Solution < 0.0 pH.
Check pH probe connection. pH probe could be
faulty. Guardian Monitor Wi-Fi could be wet inside.
temp displays ‘or’
temp displays ‘ur’
Over range temperature.
Under range temperature.
Solution >51 °C / 122 °F.
Solution <0 °C / 32 °F.
Conductivity/temperature probe or Guardian
Monitor Connect faulty.
conductivity displays ‘or’
Over range conductivity/nutrient.
Over range conductivity >9.9 EC, 99 CF, 1990 ppm.
Conductivity/temperature probe or Guardian
Monitor Wi-Fi faulty.
A display shows _ _ _ _
EC and pH cannot be determined as temperature
range is over/under.
(see technical specification range section 17.0).
Test pH probe in calibration solutions and
conductivity/temperature probe in a known
conductivity standard solution to eliminate these as
a cause of this problem. Check solution
tank/reservoir for problems.
Device doesn’t add to
Edenic app
Device security code incorrect
Enter correct 4-digit security code on QR code found
of rear of Product
Device doesn’t add to
Edenic app
Device not in Provisioning mode, mobile Bluetooth
off.
Ensure Device turned on, in Provisioning mode,
mobile Bluetooth is turned on and within range of
Device
Device doesn’t reconnect
when I move it to its
location
Device outside of Wi-Fi range
Either relocate Wi-Fi router within range or connect
to Wi-Fi network at Device’s location
For customer support visit our Support Centre at
www.bluelab.com
English
19
19.0 Technical specifications
Specifications
pH
Conductivity
Temperature
Measurement Range
0.0 –14.0 pH
0 –5.0 EC,
0 –50 CF,
0 –2500 TDS (ECx500),
0 –3500 ppm (ECx700)
0 –50 °C
32 - 122 °F
Resolution
0.1 pH
0.1 EC,
1 CF,
10 TDS,
10 ppm
1 °C
1 °F
Accuracy at 25 °C / 77 °F
±0.1 pH
±0.1 EC, ±1 CF
±50 TDS, ±70 ppm
±1 °C
±2 °F
Calibration
Two point
(pH 7.0 and
pH 4.0 or
pH 10.0)
Not required
(factory calibrated)
Not required
(factory calibrated)
Automatic Temperature
Compensation
Yes
(if conductivity/temperature
probe is in same solution as
pH probe)
Yes
-
Operating Environment
0 - 50°C / 32 - 122°F
Power Source
Input: 100-240 VAC, 50-60 Hz, 5 VA,
4 interchangeable plug types (USA, Euro, UK, NZ/AUS)
Output: 5 VDC 1 A
Signal Range
Indoor / Urban: 66 feet / 20 meters
Outdoor / RF line-of-sight: 164 feet / 50 meters
Frequency Band
2.4 GHz ISM
System Requirements
Android (8.0 or later) or iOS (12 or later) Mobile phone for Edenic Software.
Internet connection for remote access & data-logging.
Certs
CE, FCC, IC. Contains Wi-Fi/Bluetooth Module
IC: 21098-ESPS3WROOM1, FCC ID: 2ACZ7-ESPS3WROOM1
User Manual Languages
Available
English
Radio
Wi-Fi: 2.4 GHz (802.11 b/g/n)
Bluetooth
5 (Low energy)
Document version BGUSTD02_Manual_2023-04-20

This manual suits for next models

1

Other bluelab Measuring Instrument manuals

bluelab Combo Meter Guide

bluelab

bluelab Combo Meter Guide

bluelab pH Pen Installation guide

bluelab

bluelab pH Pen Installation guide

bluelab Guardian Monitor User manual

bluelab

bluelab Guardian Monitor User manual

bluelab Soil pH Pen Guide

bluelab

bluelab Soil pH Pen Guide

bluelab EC pen Guide

bluelab

bluelab EC pen Guide

bluelab Truncheon User manual

bluelab

bluelab Truncheon User manual

bluelab Commercial trunchelon nutrient meter Guide

bluelab

bluelab Commercial trunchelon nutrient meter Guide

bluelab pH pen Guide

bluelab

bluelab pH pen Guide

bluelab Soil pH Pen Guide

bluelab

bluelab Soil pH Pen Guide

bluelab BLU2210 User manual

bluelab

bluelab BLU2210 User manual

bluelab Combo Meter User manual

bluelab

bluelab Combo Meter User manual

bluelab Truncheon Guide

bluelab

bluelab Truncheon Guide

bluelab pH Pen Guide

bluelab

bluelab pH Pen Guide

bluelab conductivity pen Guide

bluelab

bluelab conductivity pen Guide

bluelab Multimedia pHMeter Guide

bluelab

bluelab Multimedia pHMeter Guide

bluelab pH Pen Guide

bluelab

bluelab pH Pen Guide

bluelab Truncheon 124060 Guide

bluelab

bluelab Truncheon 124060 Guide

bluelab Soil pH Meter Guide

bluelab

bluelab Soil pH Meter Guide

bluelab pH Probe Guide

bluelab

bluelab pH Probe Guide

bluelab Truncheon Guide

bluelab

bluelab Truncheon Guide

bluelab CTRL50110 Guide

bluelab

bluelab CTRL50110 Guide

Popular Measuring Instrument manuals by other brands

horiba U-51 instruction manual

horiba

horiba U-51 instruction manual

AEMC 205 user manual

AEMC

AEMC 205 user manual

Lovibond Scuba instruction manual

Lovibond

Lovibond Scuba instruction manual

Seitron GAS T ONE Use and maintenance

Seitron

Seitron GAS T ONE Use and maintenance

YOKOGAWA AQ6374 Getting started guide

YOKOGAWA

YOKOGAWA AQ6374 Getting started guide

Sibata DM-100S Operation manual

Sibata

Sibata DM-100S Operation manual

Endress+Hauser Promonitor NRF560 operating instructions

Endress+Hauser

Endress+Hauser Promonitor NRF560 operating instructions

iRhythm Zio QX manual

iRhythm

iRhythm Zio QX manual

BEHA Unitest Duplex 93486 instruction manual

BEHA

BEHA Unitest Duplex 93486 instruction manual

Nidek Medical LM-1800P Interface manual

Nidek Medical

Nidek Medical LM-1800P Interface manual

WilTec naturewater 50810 Operation manual

WilTec

WilTec naturewater 50810 Operation manual

Halma SENSITRON SMART S Short installation manual

Halma

Halma SENSITRON SMART S Short installation manual

Endress+Hauser Cerabar M PMC51? Cerabar M PMP51 operating instructions

Endress+Hauser

Endress+Hauser Cerabar M PMC51? Cerabar M PMP51 operating instructions

Epilog Laser Zing 24 quick start guide

Epilog Laser

Epilog Laser Zing 24 quick start guide

KMB SMY 133 operating manual

KMB

KMB SMY 133 operating manual

ADLINK Technology PCIe-9834P quick start guide

ADLINK Technology

ADLINK Technology PCIe-9834P quick start guide

Agilent Technologies B1500A user guide

Agilent Technologies

Agilent Technologies B1500A user guide

Xylem FLYGT MAS 711 Installation, operation and maintenance manual

Xylem

Xylem FLYGT MAS 711 Installation, operation and maintenance manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.