Bonart ART-SP2 User manual

ART-SP2 Scaler /Polisher Combo Unit
Instruction Manual
Bonart Co., Ltd.
Jorgen Kruuse A/S
No.80, Wuxun St., Anle Dist.,
Keelung City, 20446, Taiwan, R.O.C.
Tel: 886-2-22983980
Fax:886-2-22983981
Email:[email protected]
Havretoften 4, DK-5550 Langeskov,
Denmark
2460
COPYRIGHT© 2018 BONART CO., LTD.

- 1 -
BONART Co., Ltd. owns all rights to this publication. This publication shall be used solely as a
reference for operation, maintenance, and repair on BONART equipment. No part of this document
may be reproduced or distributed in any form for any other purpose.
In the event of inadvertent or deliberate publication, BONART intends to enforce its rights to this
manual under the copyright law. Reproduction of the materials presented in this manual without the
express written permission of BONART Co., Ltd. is prohibited.
The contents of this manual are subject to change without prior notice.
PROPERTY OF BONART CO., LTD.
ALL RIGHTS RESERVED
ART is a registered trademark of BONART Co., Ltd. & its affiliates.
Document No. RD-02-14-05 Rev.1.1 (2018/03)
Printed in Taiwan, R.O.C.

- 2 -
Operator Safety
Please read the entire manual thoroughly before operating the unit.
ART-SP2 Scaler /Polisher Combo Unit
The dental equipment described in this manual should be used only by properly trained
personnel. Only qualified personnel shall perform any adjustments, maintenance and repair of
the equipment.
NOTE, CAUTION, AND WARNING STATEMENTS
NOTE:Provide tips and advice.
CAUTION:Provide correct operating or maintenance procedures.
WARNING:Alert user of danger of injury or damage.
SYMBOLS
Complies with MDD 93/42/EEC
BF equipment
Grounding terminal
Attention! Please read instructions

- 3 -
Warnings
Important!
The ART-SP2 should be operated solely by qualified personnel (trained doctors,
dentist, and hygienists, etc.).
Patients with pacemakers should not be treated with the ART-SP2
Do not immerse the ART-SP2 in water!
If the patient or operator is pregnant, consult a physician prior to performing
dental scaling for safety purposes.
The ART-SP2 requires a shielded AC power cord.
Plug the power cord into a well-grounded outlet.
Attention to users with cardiac pacemakers!
Patients with pacemakers should avoid treatment with the ART-SP2. Studies have shown
that electronic appliances including razors, hair dryers, microwave ovens, TV receptors, and
other medical equipment, such as the ART-SP2, may interfere with the performance of
pacemakers.
For more information on this subject, please refer to the following articles:
"Advances in Cardiac Pacemaker", The New York Academy of Sciences, Vol. 167,
Article 2, pp. 515-1075
"Electromagnetic Radiation Interference with Cardiac Pacemaker", U. S. Department of
Health, Education and Welfare
"The Individual with a Pacemaker in the Dental Environment", Journal of the American
Dental Association, Vol. 91, No. 6, pp. 1224-1229

- 4 -
Please Read Before Operating:
Make sure the device is connected to a grounded outlet. Failure to
meet this requirement may severely damage the user and device.
Place the device on a flat and stable surface. Placing the device
on a tilted or unstable surface may degrade performance or damage
the device.
Do not modify the device. Modifications to the device will violate
safety codes and invalidate the warranty while endangering the
patient and operator.
Do not place heavy objects on top of the power cord and keep the
power cord away from intense heat.
Unplug the device and call Bonart for service and instructions if
you observe any abnormalities while operating the device.

- 5 -
Preface
About the ART-SP2 Scaler/Polisher Combo Unit
The ART-SP2 is CE 2460 certified in compliance with the applicable requirements of the Council
Directive 93/42/EEC. The reviewing council also classifies the ART-SP2 in accordance with
EN60601-1 and EN60601-1-2 safety requirements.
How it works
The ART-SP2 Piezoelectric Ultrasonic Scaler generates ultrasonic waves in the hand piece to
vibrate the tip, which allows the ART-SP2 to remove calculus and tartar easily. These ultrasonic
waves are created by 4 ceramic piezoelectric plates subjected to high frequency alternating
currents.
The piezoelectric transducer operates at a frequency of 29khz with minimal noise and heat, and
efficiently reduces the amount of cooling water required during treatment. This
electric-mechanical feature provides greater efficiency than traditional magnetostrictive systems.
Features:
Scaler and polisher in one system
Automatic phase lock and power control
Automatic feedback control
Micro-motor speed control of 2,000 to 30,000 RPMS
Supplies & Replacement Part
To order supplies or replacement parts in the U.S., please contact one of our authorized
distributors listed on www.askmagpie.com . For outside of U.S., please email to
CAUTION:Since the scaling tip is made of stainless steel, avoid direct contact with teeth.
Direct contact will cause damage to teeth enamel. For optimum performance, we suggest
practicing on a model or aluminum plates to become more familiar with the scaling tips before
performing on patients.

- 6 -
Table of Contents
Copyrights ……………………………...
1
Operator Safety ..........……………………….
2
Warnings ......………………………….
3
Preface ……..……………………….
5
Contents ..…………………………….
6
Section I. Indications for Use ..…………………………….
7
Section II. Contraindications and Warnings ……………………..
7
Section III. Precautions .……………………………..
7
Section IV. Infection Control ……………………………...
9
Section V. Installation …..………………………………………….
11
Section VI. General Description and Information of Parts…………….
16
Section VII. Techniques ……………………………….
18
Section VIII. System Maintenance …………………………..…
19
Section IX. Troubleshooting ……………………………….
20
Section X. Specifications ……………………………….
21
Section XI. Disposal ..……………………………..
21

- 7 -
Section Ⅰ: Indications for Use
1-A Ultrasonic scaling procedures:
Removal of calculus and plaque during dental prophylaxis.
General supra and sub-gingival scaling applications.
Periodontal debridement for all types of periodontal diseases.
Endodontic procedures.
1-B Polishing procedures:
For removal of calculus and plaque during dental polishing.
All general supra and subgingival polishing applications.
Periodontal debridement for all types of periodontal diseases.
Endodontic procedures with optional bur tools.
Section Ⅱ: Contraindications and Warnings
Do not use the ART -SP2 for amalgam restorative dental procedures.
Do not use the ART-SP2 if the patient or operator is wearing a pacemaker.
Do not immerse the ART-SP2 in water or liquid. If the ART-SP2 has water damage, return
the machine to Bonart for servicing.
Do not modify the ART-SP2. Modifications will invalidate the warranty on the machine,
violate safety codes, and endanger the patient and operator.
Section Ⅲ: Precautions
3-A Precautions for the ART-SP2 and all Ultrasonic Scaler Systems
Ensure that an adequate amount of water flows to the piezo tip during use in order to cool
the hand piece and tip.
Take precaution to prevent transmission of contagious diseases as all ultrasonic scalers
produce aerosols.
Use tips manufactured by Bonart for optimal performance.
Keep the ART-SP2 away from heat sources to prevent overheating and damaging of
electrical components.
Ensure that the water valve is closed when the ART-SP2 is not in use.
Avoid treating patients with pacemakers due to the magnetic field produced by the ART-SP2.

- 8 -
3-B:Precautions for Ultrasonic Prophylaxis Procedures
Discard and replace piezo tips as needed. Scaling tips wear out with use and should be
replaced after 2mm of wear (approximately 3 to 4 months) to maintain optimal performance
Discard and replace piezo tips that are bent, damaged, or reshaped. Do not reuse deformed tips
BONART piezo tips are made of stainless steel. Avoid direct contact with teeth to prevent
damage to the enamel. We strongly advise practicing on models and aluminum plates prior to
operating on a live subject with the ART-SP2.
Avoid direct contact of piezo tips with the subject’s lips, cheeks, and tongue.
Wipe accretions from the tooth using the long axis of the scaler tip. DO NOT gouge the tooth
3-C:Precautions for hand-piece
Do Not Twist and Pull Cable
Do Not Bend
Do Not Twist
Do Not Bundle
Pull To Disconnect

- 9 -
Section IV: Infection Control
4-A General Infection Control
Always use standard personal protective equipment such as facemasks, eyewear, face shields,
gloves, and protective gowns.
4-B Water Supply Recommendations
In addition to using protective equipment, ensure that your water supply system conforms to
applicable Centers for Disease Control and Prevention (CDC) and American Dental Association
(ADA) standards, and that all required flushing, chemical flushing, and general infection control
procedures are followed.
To further protect against infection, using distilled or purified water is recommended.
4-C Cleaning and Sterilization:
All sterilization procedures must be followed in accordance to the EN ISO17664 standards.
In order to avoid bacterial or viral infections, always clean and sterilize the following components
after each treatment:
1. Hand piece
2. Tips
3. Metal Wrench
4. Micro-motor
5. Straight hand piece (Straight nose cone)(optional accessory)
6. Prophy angle (unless disposable)(optional accessory)
The hand piece, tips, wrench, straight handpiece and prophy angle can be autoclaved and can
withstand a maximum temperature of 135℃or (275℉) for a maximum of 3 minutes. After
completing the sterilization cycle, air-dry for about 20-30 minutes before reusing.
WARNING: Do not use any other method of sterilization (dry heat, radiation, ethylene oxide,
gas, low-temperature plasma, etc.)
NOTE:The SterilizationAssurance Level (SAL) of steam autoclave should be 10-6 (According
to ISO 13683:Sterilization of health Care products)
NOTE: Prior to cleaning and sterilizing your ART-SP2 unit, always remember to turn off the
device by using the power switch and to disconnect the power plug from the outlet.
Cleaning the ART-SP2 unit
Remove the hand piece from the cable, and the tip from the hand piece.
Clean and disinfect the surface of the outer casing, cords and connectors with a cloth and a
gentle disinfectant.

- 10 -
Cleaning and sterilizing the hand piece
Remove the hand piece from the hand piece cable and the tip from the hand piece.
Take special care not to break the threading pin of the hand piece. The threading pin is
where the tip attaches to the hand piece.
Disinfect the hand piece with a cloth and mild disinfectant containing a neutral pH.
Dry the electric contacts by blowing air onto them with the syringe.
Seal the hand piece in a sterilization pouch (without any tips). If you are short on
sterilization pouches, you may use the surgical tray in place of the sterilization pouch.
Autoclave the hand piece.
WARNING: Do not immerse the hand piece cable in sterilizing fluid. The ART-SP2 can be
severely damaged if sterilization fluid gets inside the unit.
Cleaning Piezo Tip or Wrench
Clean the tip or wrench preferably in an ultrasonic tank and rinse in distilled water.
Dry the tip or wrench.
Disinfect the tip with a cloth and mild disinfectant containing a neutral pH.
Seal the tip or wrench inside a sterilization pouch. If you are autoclaving more than 1 tip,
place each tip or wrench in an individual bag. You may also use the tip holder that comes
with the ART-SP2 in place of the sterilization pouch.
Autoclave the tip or wrench.
WARNING:Before starting the sterilization cycle, make sure that the inside of the tip is
completely dry by blowing air through the internal hole with the syringe. This will prevent
stains and patches from appearing on the surface of the tip.
Cleaning the micro-motor
Remove all attachments to the micro-motor prior to cleaning. The outer surface of the
micro-motor should be cleaned with an antiseptic solution. Never oil the micro-motor.
Ensure that no moisture gets inside the micro-motor.
WARNING: Do not dip the micro-motor and cord directly into sterilization fluids. The
chemicals will interfere with normal operations if it reaches the inside of the system.
WARNING: Do not leave disinfectants on parts longer than recommended by the
manufacturer of the disinfectant to avoid damaging materials.
NOTE: The micro-motor should be cleaned after each use between patients.

- 11 -
Cleaning the straight hand piece (straight nose cone) attachment (optional accessory)
Clean the straight hand piece preferably in an ultrasonic tank and rinse in distilled water.
Dry the straight hand piece.
Disinfect the straight hand piece with a cloth and mild disinfectant containing a neutral pH.
Seal the straight hand piece inside a sterilization pouch. If you are autoclaving more than 1
tip, place each straight hand piece in an individual bag. You may also use the tip holder that
comes with the ART-SP2 in place of the sterilization pouch.
Autoclave the straight hand piece .
Cleaning the ProphyAngle(optional accessory)
The prophy angle should be cleaned to remove hair and debris approximately after every 10 dental
polishing.
1. Remove prophy angle from handpiece and rubber cup.
2. Remove head gears by turning knurled knob clockwise ( threads are reversed) turn to your right.
3. Clean with fine brush and lubricate head gears with mineral oil, Vaseline or any non-toxic
lubricant.
4. Place head gears back and tighten by turning Left or counterclockwise.
5. Normal polishing speeds are around 3,000 to 3,000 RPM, exceeding this may damage your
prophy angle.
Section V: Installation
5-A General Information
If the installation of your ART-SP2 system is performed by non-Bonart dealer personnel, check to
see that the requirements below are followed.
5-B Water Line Requirements
Incoming water supply line pressure to the Scaler must be 25 psi (172 kPa minimum) to 60
psi (414Kpa maximum). If your dental water supply line pressure is above 60 psi, install a
water regulator on the water supply line to your ART-SP2 unit.
After installing the water supply system, flush the water thoroughly prior to connecting to
the scaler.
Manually shut off the water valve of your dental water supply system prior to leaving the
office unoccupied and unattended.
5-C Electrical Requirements
Refer to Section X: Specifications

- 12 -
Unpacking the Unit
Carefully unpack your ART-SP2 unit and verify that all components and accessories are included:
(Figure 1)All components and accessories
Item
Quantity
Item
Quantity
(1) Main Unit
1
(2) Foot Control / Foot Switch
1
(3) AC Power Cord Set.
1
(4) PU Water Tubing
1
(5) User Manual & Literature
Packet
1
(6) Piezo Hand piece.
1
(7) Metal Torque-Limited
Wrench
1
(8)
ART-BS1 Universal Pointed Tip
ART-BS2 Chisel Head Flat Tip
ART-BS3 Round Flat Tip
1
1
1
(9) Micro-motor
1
(10) Micro-motor seat
Optional
(11) Straight Hand Piece
Optional
NOTE: Other piezo tips are available and can be purchased separately through dealers.
Included items are subject to change depending on location.
NOTE:While unpacking, check your ART-SP2 unit for any damages. If any damages
are found, please contact your dealer immediately. Enter the unit serial number on your
warranty card and mail it within 14 days from the date of purchase.

- 13 -
5-E Power Cord / Power Connection
Always make sure that the power switch is set to the OFF position before performing the
following tasks:
oPlugging and unplugging theAC cord into the back of the unit.
oPlugging and unplugging the 3-prong plug into a grounded electrical outlet.
Safety Instructions
A. Grounding:
Verify that the main unit is connected to a grounded outlet prior to connecting other parts to
the main unit.
B. Main voltage range and fuse:
Check the voltage of the ART-SP2 for compatibility with your local voltage before plugging
the power cord into the outlet.
5-F Foot Control Cable Assembly Connection
Align the pins of the foot control plug with the receptacle on the back of the device and push firmly
until seated.
5-G Water Supply Line Connection
Push the blue water tube (hose) over the stainless steel receptacle on the back of the unit. Then
tighten the screw over it.
Connect the quick connector to the water supply line. The ART-SP2 standard package comes with a
male quick connector. If a female quick connector is necessary, please contact a Bonart authorized
dealer to purchase it.
Inspect all connections for leaks.
To remove the water line from the ART-SP2 scaler, turn off the water supply or disconnect the water
supply line from the water tube. Then loosen the screw on the water tube from the receptacle of
the unit and gently pull the water tube off.
5-H Hand piece Cable Assembly Connection
The hand piece is the housing for the piezo tips. It is the part that you hold during operation. The
hand piece must be attached to the hand piece cable prior to using. To attach, connect the free end
of the cable to the matching receptacle on the back end of the hand piece. Do not twist or turn.

- 14 -
5-I: Polishing Hand piece Assembly and Start Up. (Optional accessories)
To detach, pull the cable from the hand piece in a straight motion. Do not twist, turn, or yank the
hand piece from the cable.
Polishing Hand Piece Assembly and Start Up. (Optional accessories)
(Figure 2)
To assemble the polishing hand piece, connect the Prophy Angle, Straight Hand Piece, and
Micro-motor by following the steps below:
1. Connect the straight hand piece and the micro-motor by pushing the straight hand piece over the
rod of the micro-motor. It should snap into place.
(Figure 3)
2. Attach the prophy angle to the straight hand piece by following the steps below:
A) Unlock the prophy angle from the straight hand piece by turning the locking ring clockwise.
This should open the lock. Note: When you first unpack the straight hand piece, there may
be a metal rod at the top that serves as a placeholder. Remove the rod after you open the
lock.
(Figure 4)

- 15 -
B) Match the groove at the base of the prophy angle with the top of the straight hand piece and
connect them together.
(Figure 5)
C) Close the lock by turning the locking ring counter-clockwise. The prophy angle should now
be locked onto the straight hand piece and should not fall off during operation.
(Figure 6)
D) If you wish to remove the prophy angle, follow step Ato release the lock on the prophy angle.
WARNING:The prophy angle should be cleaning and lubricated on a regular basis.
To polish, follow the steps below:
1. Place a rubber cup on the prophy angle and put paste in the cup to begin polishing.
2. Press the Micro-motor (mode) button to switch from scaling to polishing mode.
3. Adjust the Oscillating dial to a low speed, 2,000 rpm.
4. Press on the foot switch to engage the polishing hand piece and begin polishing.
5. Add paste as needed.
Assembling the Scaler Unit
1 Ensure that the power switch is set to off. If plugged in, the power indicator light should not be
lit.
2. Plug the power cord of the main unit into a grounded AC power outlet.
3. Insert the tip into the hand piece and lock it into place by using the enclosed wrench. Be
careful not to overturn the tip as it may cause the tip and hand piece to break.
4. Ensure that the water tube connection to the water supply source and main unit is secure.

- 16 -
Section VI: General Description and Information of Parts
1. Main unit
5. Water control knob
9. Water hose input
2. Hand piece
6. Power control knob
10. Power cord socket
3. Power button
7. Micro-motor socket
11. Forward/Reverse switch
4. Micro-motor (Mode) button
8. Foot switch socket
Figure 7.ART-SP2 diagram
1. Main Unit
The main unit generates power and produces a signal that is passed to the hand piece. The
hand piece is then powered up and vibrates the installed tip.
2. Hand piece
The hand piece is the housing for the piezo tip. It consists of four piezoelectric plates that
vibrate. The plates are driven by the electrical signal from the Main unit. The ART-SP2 unit
is compatible with Bonart BS-type piezo tips and other brands of piezo tips with a comparable
thread. Do not use Bonart BS-type piezo tips with the ART-SP2.
3. Power button (ON/OFF)
The power button is used to turn the ART-SP2 main unit on and off. When the power is on,
the LED indicator should be lit. When turned off, the LED power indicator should not be lit.
Do not press the foot switch while turning on the ART-SP2.
4. Micro-motor (Mode) button
The micro-motor (Mode) button is used to switch between polishing and scaling mode.
When pushed, the LED indicator should be lit, indicating that the system is operating in
polishing mode. Under polishing mode, the micro-motor is engaged. To return to scaling
mode, push the button again.

- 17 -
5. Water control knob
The water control knob allows you to control the amount of water that flows through the hand
piece. If turned clockwise, the amount of water will decrease. If turned counter-clockwise,
the amount of water will increase. Please note that water is necessary to cool the tip and
prevent overheating so it is necessary to ensure that an adequate amount of water flows
through the hand piece while operating the device. The larger the water stream, the lower the
temperature, and vice versa.
WARNING:Do not turn the water knob counter-clockwise more than three full turns.
Doing so will damage the internal water line needle.
WARNING:If needed, gently adjust the water control knob to prevent any serious
damages to the water valve.
6. Power control knob
The Power Control Knob allows you to control the power intensity. When turned clockwise,
the power intensity increases. When turned counter-clockwise, the power intensity decreases.
7. Micro-Motor Socket
The micro-motor socket is where you connect the micro-motor to the main unit. It is located
on the back of the main unit at the bottom left.
8. Footswitch Socket
The foot switch socket is where you connect the foot switch to the main unit. It is located
next to the water hose input on the back of the main unit.
9. Water hose input connector
The water hose input connector is the contact point for the water tube and the main water
supply line. Use either a male or female quick connect to connect the water tube to the water
supply. The ART-SP2 comes with a male quick connect attached to the water tube.
10. Power cord socket
The power cord socket is where you connect the power cord to the main unit.
11. Forward / Reverse switch
The Forward/Reverse switch allows you to control the direction of the micro-motor. It is
located on the back of the main unit.

- 18 -
Section VII: Techniques
7-A Performing Ultrasonic Scaling Procedures
1. Use purified or distilled water to prevent infection when patients experience tissue laceration
during treatment
2. Keep the power cord tidy to avoid tripping and other accidents.
3. Position the footswitch in an easily accessible spot for the user. Keep the footswitch cord
tidy to avoid tripping and other accidents.
4. Lock the tip in the hand piece securely using the enclosed wrench.
CAUTION: DO NOT FORCE OR OVERTURN THE TIPINTO PLACE! Doing so will
damage the tip, hand piece, and wrench.
NOTE: The scaler will not operate if the tip is either removed from the hand piece or
improperly installed.
5. Hold the hand piece over a sink or drain with a tip installed. Verify that water is reaching the
tip.
6. Check your Bonart piezo tip for wear and replace as needed.
7. Use the foot control to regulate water flow to the hand piece. Hold down the foot control to
allow water to flow through. Release the foot control to stop.
8. Adjust the water flow to the desired level using the water-adjustment knob. Increase the
water flow by turning the water control knob clockwise. Decrease the water flow by turning
the water control knob counter-clockwise.
NOTE: Increased water flow results in a cooler hand piece temperature and vice versa.
CAUTION: During operation, a continuous flow of water is required to keep the hand
piece cool.
7-B Applications of the TIP
A. Tangential application (BS1 TIP)
Do not apply the BS1 tip directly to the tooth. Doing so
may damage the enamel. Control the hand piece with a
slow and steady motion.

- 19 -
B. Frontal application (BS2 TIP)
Apply the BS2 tip directly against the tartar, but not
directly against the tooth. Use a slight amount of
pressure.
C. Tangential application (BS3 TIP)
Do not apply pressure when applying the BS3 tip to the
enamel.
Section VIII: System Maintenance
8-A Hand piece maintenance
1. After each use the hand piece and tip should be rinsed with clean water for about 20 to 30
seconds.
2. Inspect the hand piece cable daily to ensure it is in good condition.
3. For sterilization of the hand piece please refer to Section IV.
8-B TIP maintenance
Worn out tips can adversely affect performance, resulting in insufficient power and vibration.
Check tips regularly for wear and tear and replace as necessary.
8-C Main Unit Maintenance
The housing of the ART-SP2 as well as the hand piece cord, should be cleaned and disinfected
(using alcohol or soap and warm water) on a daily basis.
Table of contents
Other Bonart Media Converter manuals