OPERATING INSTRUCTIONS
Place the B8572 instrument on table top or
flat surface, test button a if low battery
symbol appears , replace batteries.
Take 5lbs weights from case (use care)
Plug red/black test leads into test weights
Plug other end of leads into B8572
Accessories: B8565/20 (one 20’ test lead)
(For testing surface to ground on floors)
TESTING:
Types of testing as specified in ESD-S4.1
1. Periodic testing on installed static safe
workstations.
2. Qualification of workstation materials
3. Performance of materials
4.
PERIODIC TESTING OF INSTALLED
PRODUCTS
Measure RTG (Resistance to Ground)
Remove items from surface to be tested
Remove ESD sensitive devices
Clip one end of test lead from B8572 to
grounded point (see fig 2.1)
Use one 5lb electrode on other test lead to
surface to be tested and push green button
TESTING FLOORS
Place one 5lb electrode of floor surface and with
B8565/20 20’
lead connect to known ground point.
(check with your facility staff)
.
Note: ESD 7.1 is designed to measure floor materials
with resistance of 2.5 x 104to 1.0 x 1011 ohms.
Note: ESD-S4.1 suggests that a static control surface
in the test range of 1 x 106 to 1.0 x 1010 ohms is
acceptable.
QUALIFICATION OF INSTALLED
STATIC SAFE SURFACE RTT
Measurements of resistance of Top to Top
and resistance point to point are the same.
Place both 5lb electrodes on surface to be tested.
Place no closer than 2” from edge of surface to be
tested.
Place no closer than 3” from any groundable point.
(snaps are a groundable point).
Place in the center of mat area it would be area
most worn. (about 10” apart)
For RTG Fig 2.1 place one lead with insulated plug
to groundable point.
REPORTING AND USING TEST RESULTS
Different standards have different requirements please
review the standards that you are testing too.
Per ESD 4.1 (work surfaces)
RTG maximum and minimum values
measured for resistance-to ground
RTG maximum and minimum values measured for
point-to-point resistance in ohms.
Per ESD 7.1 (floors)
RTG all values in ohms for resistance to ground
RTT all values in ohms for point-to-point resistance.
Voltage levels tested
<106 @ 10 volts
>106 @ 100 volts
Note: Temperature & Humidity
Test dates (work surfaces) quantity see ESD
handbook TR 20.20 phg/ 5.3.1.3
Fig 2.1
10” apart
Test leads
Fig 2.2
Fig. 2.3
To known
ground
RTG Test
Mat surface
Mat surface
RTT test
Always clean bottom of electrodes with alcohol pads
Always clean mat surface with approved cleaner
SPECFICITIONS
Accuracy range: +/-3 % per gegohm
Test voltage: 10,25,50,100,250 & 500volts
Electrodes: Two (2) 5lb weights
Power: Six (6)AA
Auto shut off: Conserve batteries
Lock test : As needed
Resistance range: 100G/100v , 100G/500v analog
20G/100v digital
Continuity test: .01meg-220G
Diode test: .01kohm-10000kohm Size:
Voltage: +/-1 V-500V
Test leads: 2-4’ long banana to banana
(red & black)
Millivolt test: +/- 1m-1999mV
Low battery: Symbol
Water& dust proof: IP54
Weight: 775gms with boot
Size: 11mm x 220mm x 45mm
Designed to: IEC1010-1
Case: Foam filled
Warranty: Three (3) years
ACCESSORIES:
B8563/20: 20’ lead for floor testing
B48775: Electrostatic field meter
B48775PVS: Complete periodic verification system for testing
ionization. Field meter, charger, plate & case.
Locking
Auto-shut off
Digital scale
Analog scale
Footwear: Incoming testing on sampling basis should be
performed for all static footwear products.
(TR 20.20 pgh. 5.3.3.4)
Floor: Per (TR 20.20 pgh. 5.34.13)
Testing and monitoring are required
for performance.
Seating: ESD STM 12.1 and (TR 20.20 pgh. 5.3.5.3)
recommend testing should be preformed. With a reading
of 1 x109 ohms.
Garments: A process of testing garments per ESD STM
2.1, point-to point and sleeve-to-sleeve resistance should
be done.
see (TR 20.20 pgh. 5.3.3.4)
MODEL # B8572
SERIAL # 1000-351/110111/1335
DATE 06-21-11
Must call to activate warranty
1-888-257-3680
Information sent by mail
M ohm
EXAMPLE:
Reading (115.)= 1.15 x 108
Reading (.096.)= 9.6 x 104
Reading ( .96)= .96 x 105
Reading ( 6.3)= 6.3 x 106
Reading (16.3)= 1.63 x 107
Reading (6.13 G)= 6.13 x 109
Reading (>20G)= 2 x 1010
Reading (100G)= 1 x 1011 analog
Backlight
Test button
Test voltage at probes
BOTRON CO INC.
325 W. Melinda Ln.
Phoenix AZ 85027
Ph 623-582-6700
Fax 623-582-6776
[email protected] / www.botron.com
Caution, risk of shock.
Caution, refer to user guide.
Equipment protected throughout by
Double insulation case (Class II).
Equipment complies with current EU directives.
Equipment must not be connected to installations >500V.
Test voltage
Locking
Voltage Test
voltage
Locking
Continuity Test
Resistance in K ohms
Buzzer
DIGITAL/ ANALOG
AUDIT MEGOHM
METER B8572

Other manuals for B8572

1

Other Botron Measuring Instrument manuals

Botron OMNIGND B92700 User manual

Botron

Botron OMNIGND B92700 User manual

Botron B9201 Instruction Manual

Botron

Botron B9201 Instruction Manual

Botron B8572 User manual

Botron

Botron B8572 User manual

Popular Measuring Instrument manuals by other brands

AEP transducers MP2C operating manual

AEP transducers

AEP transducers MP2C operating manual

Sony BVU-800 Operation and maintenance manual

Sony

Sony BVU-800 Operation and maintenance manual

Wireflow WF 3169 user manual

Wireflow

Wireflow WF 3169 user manual

IMT Analytics FlowAnalyser PF-300 PRO user manual

IMT Analytics

IMT Analytics FlowAnalyser PF-300 PRO user manual

Fluke 922 user manual

Fluke

Fluke 922 user manual

ITRON OpenWay Riva IRM-STAR installation guide

ITRON

ITRON OpenWay Riva IRM-STAR installation guide

PCB Piezotronics 080A57 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 080A57 Installation and operating manual

CSE Universe UMG 96L operating instructions

CSE Universe

CSE Universe UMG 96L operating instructions

Circutor AR5-L instruction manual

Circutor

Circutor AR5-L instruction manual

Endress+Hauser Proline t-mass A 150 operating instructions

Endress+Hauser

Endress+Hauser Proline t-mass A 150 operating instructions

Agilent Technologies E4980A Operation manual

Agilent Technologies

Agilent Technologies E4980A Operation manual

Omega HHP350 Series user guide

Omega

Omega HHP350 Series user guide

VDO CONTISYS brochure

VDO

VDO CONTISYS brochure

Sensor Electronics AirDistSys 5000 Operator's manual

Sensor Electronics

Sensor Electronics AirDistSys 5000 Operator's manual

Keysight E5071C ENA Series Service guide

Keysight

Keysight E5071C ENA Series Service guide

Lovato DMG900 instruction manual

Lovato

Lovato DMG900 instruction manual

EA Technology UTL2 operating manual

EA Technology

EA Technology UTL2 operating manual

Diehl Metering Sharky 775 installation guide

Diehl Metering

Diehl Metering Sharky 775 installation guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.