Breathe LIFE2000 User manual

BREATHE TECHNOLOGIES
®
LIFE2000® VENTILATION SYSTEM
Instructions for Use
0 100%

1
CONTENTS
CHAPTER 1: LIFE2000® VENTILATION SYSTEM INTRODUCTION___________________________ 6
Indications for Use _______________________________________________________________________6
Symbols and Conventions__________________________________________________________________6
Safety Information________________________________________________________________________7
Features ______________________________________________________________________________ 10
Packaging Contents _____________________________________________________________________ 11
System Components_____________________________________________________________________ 12
Ventilator _____________________________________________________________________________ 12
Compressor ___________________________________________________________________________ 13
Configurations _________________________________________________________________________ 14
CHAPTER 2: STATIONARY CONFIGURATION __________________________________________15
Introduction to Stationary Configuration_______________________________________________________ 15
Testing the Ventilation System______________________________________________________________ 15
Positioning and Carrying the Compressor _____________________________________________________ 16
Supplying Power to the Compressor _________________________________________________________ 17
Docking the Ventilator into the Compressor____________________________________________________ 18
Connecting an Interface to the Compressor ___________________________________________________ 19
The Outlet Fitting in Stationary Configuration___________________________________________________ 19
Powering On Sequence in Stationary Configuration_____________________________________________ 20
Checking the Compressor’s Internal Battery Status ______________________________________________ 21
Checking the Ventilator Battery Charge_______________________________________________________23
Ventilator Battery Charge Icons, Meanings, and Approximate Time Remaining __________________________23
Silence Alarm Button on the Compressor _____________________________________________________24
Powering O Sequence in Stationary Configuration_____________________________________________ 25
CHAPTER 3: EXTENDED RANGE CONFIGURATION ____________________________________ 26
Introduction to Extended Range Configuration_________________________________________________ 26
Testing the Ventilation System_____________________________________________________________ 26
Positioning and Carrying the Compressor _____________________________________________________27
Supplying Power to the Compressor _________________________________________________________28
Undocking the Ventilator from the Compressor ________________________________________________ 29
Connecting the Ventilator and Compressor in Extended Range Configuration __________________________30
Connecting an Interface to the Ventilator in Extended Range Configuration ____________________________ 31
Powering On Sequence in Extended Range Configuration_________________________________________32

2
Checking the Compressor’s Internal Battery Status ______________________________________________33
Checking the Ventilator Battery Charge_______________________________________________________35
Ventilator Battery Charge Icons, Meanings, and Approximate Time Remaining __________________________35
Assembling the Ventilator Battery Charger and Charging the Ventilator _______________________________36
Securing the Ventilator ___________________________________________________________________37
Belt Clip ______________________________________________________________________________ 37
Ventilator Silence Alarm Button _____________________________________________________________38
Powering O Sequence In Extended Range Configuration_________________________________________39
Disconnecting Extended-Range Configuration__________________________________________________39
CHAPTER 4: STANDALONE CONFIGURATION ________________________________________ 40
Introduction to Stand-Alone Configuration _____________________________________________________40
Testing the Ventilator ____________________________________________________________________ 41
Undocking the Ventilator from the Compressor _________________________________________________ 41
Connecting to a Cylinder__________________________________________________________________42
Connecting an Interface to the Ventilator in Stand-Alone Configuration _______________________________44
Powering On Sequence For the Ventilator_____________________________________________________44
Checking the Ventilator Battery Charge_______________________________________________________45
Ventilator Battery Charge Icons, Meaning, and Approximate Time Remaining___________________________45
Assembling the Ventilator Battery Charger and Charging the Ventilator _______________________________46
Securing the Ventilator ___________________________________________________________________47
Belt Clip ______________________________________________________________________________ 47
Pole Mount ____________________________________________________________________________48
Ventilator Silence Alarm Button _____________________________________________________________49
Powering O Sequence For the Ventilator____________________________________________________ 50
CHAPTER 5: CONNECTING AN INTERFACE ___________________________________________51
The Breathe Interfaces ___________________________________________________________________ 51
Connecting an Interface to the Compressor in Stationary Configuration _______________________________ 51
Connecting an Interface to the Ventilator in Extended Range or Stand-Alone Configuration ________________ 51
The Breathe Universal Circuit™ Connector ____________________________________________________ 52
Connecting the Breathe Universal Circuit™ Connector to Third-Party Patient Masks or Tubes _______________53
Examples of Patient Mask and Tube Connections _______________________________________________53
The Breathe Pillows Interface™ _____________________________________________________________54
Wearing the Breathe Pillows Interface™ ______________________________________________________ 55
Checking the Breathe Pillows Interface™ Positioning ____________________________________________ 55
CONTENTS

3
CHAPTER 6: VENTILATION SETTINGS ______________________________________________ 56
Introduction to Ventilation Settings _________________________________________________________ 56
Home Screen _________________________________________________________________________ 56
Moving Between the Home Screen and Menu Screen____________________________________________57
Menu Screen __________________________________________________________________________57
Touch Screen Energy-Save Mode ___________________________________________________________58
Defining Clinical Settings__________________________________________________________________58
Accessing the Settings Menu ______________________________________________________________58
Accessing the Clinician’s Settings Menu _____________________________________________________ 59
Disabling Access to the Clinician’s Settings Menu ______________________________________________ 59
Prescription Settings ____________________________________________________________________ 60
Editing Prescription Settings ______________________________________________________________ 60
Activating Prescription Settings _____________________________________________________________ 61
Factory Default Prescription Settings ________________________________________________________ 62
Breath Types __________________________________________________________________________63
Ventilation Modes _______________________________________________________________________63
Setting Ventilation Parameters in Control Ventilation Mode ________________________________________64
Setting Ventilation Parameters in Assist/Control Ventilation Mode __________________________________ 66
Setting Ventilation Parameters in Assist Ventilation Mode _________________________________________68
Setting Alarm Limits for Breath Rate and PIP ___________________________________________________ 71
Setting Breath Timeout (Apnea Backup Ventilation Mode) _________________________________________72
Selecting the Source Gas _________________________________________________________________73
Choosing an Activity Button (Patient-Selectable) to Begin Ventilation _________________________________74
“This Prescription Setting is Not Active” Message _______________________________________________75
“Connect Oxygen Source” or “Disconnect Oxygen Source” Message ________________________________75
Adjusting the Trigger Sensitivity (Patient-Adjustable) _____________________________________________76
Accessing the Trigger Sensitivity Screen ______________________________________________________76
Changing Trigger Sensitivity _______________________________________________________________ 77
Accessing the Utilities Menu _______________________________________________________________78
Setting Time and Date ___________________________________________________________________79
Setting Vibration Mode __________________________________________________________________ 80
Setting Audio Loudness __________________________________________________________________ 81
Adjusting Screen Brightness _______________________________________________________________82
Viewing Software Version Information ________________________________________________________83
Summary of Factory Default Settings _________________________________________________________84
CONTENTS

4
CHAPTER 7: ALARMS, ALERTS, AND TROUBLESHOOTING ______________________________ 85
Introduction to Alarms and Alerts____________________________________________________________85
Ventilator On-Screen Alarm Sounds and Message Display_________________________________________86
Active Alarms Window ___________________________________________________________________86
Silencing and Clearing On-Screen Alarms _____________________________________________________87
Ventilator Alarms________________________________________________________________________88
High-Priority Alarms _____________________________________________________________________88
Medium-Priority Alarms __________________________________________________________________ 92
Low-Priority Alarms ______________________________________________________________________98
Compressor Alerts _____________________________________________________________________ 100
Troubleshooting _______________________________________________________________________ 102
CHAPTER 8: MAINTENANCE _____________________________________________________ 108
Cleaning Before First Use ________________________________________________________________ 108
Daily Checks__________________________________________________________________________ 108
Environmental Specifications______________________________________________________________ 108
Alarm Checks _________________________________________________________________________ 109
Cleaning and Disinfecting the Ventilation System_______________________________________________ 109
Cleaning for Single Patient Use ____________________________________________________________ 110
Cleaning for Multi-Patient Use ______________________________________________________________ 111
Cleaning the Breathe Interfaces ___________________________________________________________ 112
Purging the Breathe Universal Circuit™Connector and Breathe Pillows Interface™ ________________________113
Preventive Maintenance __________________________________________________________________114
Battery Replacement_____________________________________________________________________114
Checking and Replacing the Condensation Tray _______________________________________________ 115
Checking and Replacing the Air Inlet Filter____________________________________________________ 116
Checking and Replacing the Cooling Filter Assembly ____________________________________________117
Testing Ventilator Alarms_________________________________________________________________ 118
Recommendations for Frequency of Testing __________________________________________________ 118
Verifying Power-On Self-Test Alarms ________________________________________________________ 118
Verifying Backup Alarm Buzzer ____________________________________________________________ 118
Testing Alarm Conditions_________________________________________________________________ 119
CONTENTS

5
CHAPTER 9: ACCESSORIES ______________________________________________________129
Battery Information _____________________________________________________________________ 129
Ventilator Battery Specifications ___________________________________________________________ 129
Compressor Battery Specifications _________________________________________________________ 129
Ventilator Battery Charger and Power Cord Specifications ________________________________________ 130
Compressor Power Supply and Power Cord Specifications _______________________________________ 130
Oxygen Monitor ________________________________________________________________________131
Exhalation Volume Monitor ________________________________________________________________131
Ventilator Altitude Volume Adjustment Table __________________________________________________ 132
Accessories and Replacement Parts ________________________________________________________ 133
CHAPTER 10: PRINCIPLES OF OPERATION ___________________________________________134
General Overview______________________________________________________________________ 134
Stationary Configuration Operation Summary and Pneumatic Diagram_______________________________ 134
Extended Range Configuration Operation Summary and Pneumatic Diagram__________________________ 134
Stand-Alone Configuration Operation Summary and Pneumatic Diagram _____________________________ 135
CHAPTER 11: PERFORMANCE SPECIFICATIONS _______________________________________136
CHAPTER 12: COMPLIANCE AND IEC CLASSIFICATION _________________________________142
CHAPTER 13: ICONS ____________________________________________________________146
APPENDIX ___________________________________________________________________ 150
Cylinder Duration Information _____________________________________________________________ 150
Cylinder Duration Equation _______________________________________________________________ 151
Replacing the Source Gas Cylinder _________________________________________________________ 151
Potential Tidal Volumes__________________________________________________________________ 152
LIMITED WARRANTY ____________________________________________________________153
CONTENTS

6
CHAPTER 1: LIFE2000® VENTILATION SYSTEM INTRODUCTION
INDICATIONS FOR USE
The Breathe Technologies®Life2000®Ventilation System is intended to provide continuous or intermittent
ventilatory support for the care of individuals who require mechanical ventilation.
The Life2000®Ventilation System consists of the Life2000®Ventilator and the Life2000®Compressor.
The System is intended for use by qualified, trained personnel under the direction of a physician. Specifically, the
System is applicable for adult patients who require the following types of ventilatory support:
• Positive Pressure Ventilation, delivered invasively (via ET tube) or non-invasively (via mask).
• Assist/Control mode of ventilation.
The System is suitable for use in home and institutional settings and is not intended for ambulance or air
transportation.
WARNING:
Use the Breathe Technologies® Life2000® Ventilation System only for patients who meet the Indications for Use.
If the ventilation system is used for patients that do not meet the Indications for Use, patients may not receive
appropriate respiratory therapy.
SYMBOLS AND CONVENTIONS
The following symbols and conventions are used throughout this manual:
THIS MEANS THIS
WARNING: Indicates hazards that, if not avoided, may cause severe injury or death.
CAUTION: Indicates hazards that, if not avoided, may result in minor or moderate injury, or damage to or
impaired performance of equipment.
TIP: and
TIPS: Indicates tips that may be helpful when using the ventilation system.
NOTE: and
NOTES: Indicates additional information about a behavior or feature.
BOLD TEXT The names of menu items and buttons displayed on the touch screen are indicated with
bold text.
For example, the Menu screen has several buttons, including Home Screen, Settings, and
Information.

7
1
LIFE2000 VENTILATION SYSTEM INTRODUCTION
SAFETY INFORMATION
Please read the following safety warnings and cautions in their entirety before using the Breathe Technologies®
Life2000® Ventilation System. Warnings and cautions can also be found throughout this Instructions for Use.
WARNING:
• The Breathe Technologies® Life2000® Ventilation System is a restricted medical device intended for use
by qualified, trained personnel under the direction of a physician.
• Use the Breathe Technologies® Life2000® Ventilation System only for patients who meet the Indications
for Use. If the ventilation system is used for patients that do not meet the Indications for Use, patients may
not receive appropriate respiratory therapy.
• If the Breathe Technologies® Life2000® Ventilation System is not functioning properly, respiratory therapy
may be compromised and may result in patient harm or death. Always have an alternate means of
ventilation or oxygen therapy available.
• The operator of the ventilation system is responsible for reading and understanding this manual before
use.
• Failure to read this Instructions for Use may result in product misuse, which may cause equipment damage
or patient mistreatment.
• The prescription and ventilation settings should only be changed on the order of the supervising physician.
• When the ventilation system is in use, keep it in a well-ventilated area to prevent it from overheating.
The ventilation system may overheat and be permanently damaged if it is used in an area that is not well
ventilated.
• Do not allow smoking near oxygen sources or near the ventilation system and do not place oxygen
sources or the ventilation system near any source of direct heat or open flame because flammable
materials burn more readily in the presence of oxygen.
• Do not submerge the ventilation system in liquids or pour liquids on it. Liquids may cause components in
the system to malfunction.
• Do not use the Breathe Technologies® Life2000® Ventilation System in magnetic resonance imaging (MRI)
environments. MRI equipment may cause electronic components in the system to malfunction. Use of the
ventilation system in an MRI environment may damage the ventilation system or other equipment and may
cause severe injury.
• Do not use the ventilator or compressor in the presence of flammable anesthetics.
• Do not use the ventilation system with oxygen in the presence of flammable anesthetics such as fluroxene,
cyclopropane, divinyl ether, ethyl chloride, ethyl ether, and ethylene, as they may form flammable or
explosive mixtures with oxygen.
• Do not use the ventilator with helium or helium mixtures.
• Do not use the ventilator with nitric oxide.
• Do not use the ventilator in a hyperbaric chamber.
• Do not eat, drink, or chew gum while using the ventilation system. Food or liquids that make contact with
the ventilation system may cause components in the system to malfunction. Eating, drinking, or chewing
gum while using the system may also increase the risk of choking.
• Do not power on or use the compressor without the filters and condensation tray properly installed.
• Do not insert foreign objects into any part of the ventilation system.
• The backside of the ventilator enclosure may reach 49°C in a 40°C environment.
• Unauthorized modifications can result in equipment damage, or patient injury or death.

8
LIFE2000 VENTILATION SYSTEM INTRODUCTION
1
WARNING:
• For any accessories, read the label and accompanying document(s) before use.
• Use only Breathe Technologies® approved accessories and replacement parts with the ventilation system.
If unauthorized accessories or replacement parts are used with the system, the ventilation system may be
damaged and performance may be degraded.
• Do not connect the ventilation system components or accessories to any other equipment that is not
described in this Instructions for Use.
• Adding attachments or other components and/or sub-assemblies to the ventilator breathing system can
cause an increase in expiratory resistance at the patient connection.
• Adding humidification or nebulization can increase the resistance of the breathing circuit. The operator of
the ventilation system needs to monitor the breathing system for increased resistance and blockage.
• Ventilator accuracy can be aected by the gas added by use of a nebulizer.
• To ensure accuracy of oxygen administration and to monitor for the presence of contamination (incorrect
gas connected), use an external oxygen monitor to verify the oxygen concentration in the delivered gas.
• To monitor minute volume, use an external exhaled volume monitor.
• Before beginning ventilation therapy in Stand-Alone Configuration, verify that there is an adequate supply
of source gas supply for the intended duration of the therapy. Otherwise, the patient may not receive
appropriate therapy.
• Use only a Breathe Technologies® source gas supply hose with the ventilation system. If an unauthorized
source gas supply hose is used with the ventilation system, the system may be damaged.
• Only use the ventilator with the compressor or approved medical grade compressed oxygen. Use with
non-approved sources of gas may cause the ventilator to malfunction and the patient may not receive
appropriate respiratory therapy.
• If using the ventilator with an alternate gas source in Stand-Alone Configuration, and the ventilator is not
used with a regulator capable of 41 PSI to 87 PSI (nominal 50 PSI) with greater than 40 LPM capability,
patients may not receive appropriate respiratory therapy.
• To prevent risk of cross-contamination, clean and disinfect the ventilation system before using it on a
new patient, and use a new Breathe Pillows Interface™ or Breathe Universal Circuit™connector. For the
third-party patient mask, refer to the user guide provided by the manufacturer. Replace the oxygen hose
between patients.
• Breathe interfaces are designed for single-patient use. To prevent risk of cross-contamination use a new
Breathe Pillows Interface™ or Breathe Universal Circuit™connector for each new patient. For third-party
masks or tubes, refer to the user guide provided by the manufacturer for replacement and/or cleaning and
disinfection instructions.
• Do not subject Breathe interfaces or source gas supply hoses to heat sterilization, hot water pasteurization,
autoclaving, radiation sterilization, ethylene oxide gas sterilization, or attempt to clean them in a dishwasher
or microwave oven. Doing any of these may damage the interfaces or hoses and impair gas delivery.
• If using the Breathe Pillows Interface™, properly secure the patient interface to the face and route tubing
around the ears to avoid strangulation.
• The interface, source gas supply hose, and power cords should be positioned to avoid restricting
movement, causing a tripping hazard, or posing a strangulation risk.
• Do not cover or block the compressor’s internal alarm buzzer with any object. Covering the buzzer may
make it dicult for a patient or caregiver to hear alarms, which may result in inadequate respiratory therapy.
• Do not cover the ventilator, touch screen, speaker, or backup alarm buzzer with tape or any other object.
Covering the ventilator or any of its parts might cause diculty in hearing alarms and might aect ventilator
performance.
• Ensure that the alarm loudness is set above the loudness of your surroundings.

9
1
LIFE2000 VENTILATION SYSTEM INTRODUCTION
CAUTION:
• No user serviceable components are inside the device; do not attempt to repair any components inside
the device.
• Do not place the battery charger on wet surfaces or use in wet environments. Wet environments may
damage the battery charger and may cause electric shock.
• Use only the Breathe Technologies® approved battery charger and cord set with the ventilation system.
If an unauthorized battery charger or cord set is used with the ventilation system the system may be
damaged.
• The compressor's power supply must be certified to IEC 60601-1, IEC 60601-1-11, and be Class II and IP22.
• If using in Extended Range or Stand-Alone Configuration (wearable configurations), make sure the clip
is securely fastened to the belt and the ventilator. If the clip is not securely fastened to the belt or the
ventilator, the ventilator may fall and be damaged.
• If using in Extended Range or Stand-Alone Configuration (wearable configurations), secure the ventilator to
prevent it from falling or becoming damaged.
• Breathe Technologies recommends a 30-day replacement schedule for the Breathe Universal Circuit™
connector and the Breathe Pillows Interface™.
• Do not use a Breathe Pillows Interface™ or Breathe Universal Circuit™connector that is cracked, odorous,
broken, or kinked. If a damaged interface is used, the patient may not receive adequate respiratory
therapy.
• 70% isopropyl alcohol may damage the touch screen. When cleaning external surfaces of the ventilation
system with 70% isopropyl alcohol, avoid contact with the touch screen.
• Keep in a clean environment to protect the equipment from ingress of dust, lint, and pests.
• Do not leave the ventilation system exposed to the sun or other sources of radiant heat, it may overheat.
• Do not allow children or pets to access the ventilation system; it may become damaged.
• The performance of the Life2000® Compressor has only been validated with the Life2000® Ventilator.
• Do not connect another device to the outlet fitting on the compressor when the ventilation system is in
Stationary Configuration.
• Do not connect the purge tube connector to the outlet fitting on the compressor while the ventilation
system is in Stationary Configuration and in use.
• Using the outlet fitting while the ventilation system is delivering therapy in Stationary Configuration might
decrease flow to the patient.
• The Life2000®Ventilation System provides high flows up to 40 LPM which may cause drying of the airway
passages. Alert the physician if the patient experiences air passageway drying.

10
LIFE2000 VENTILATION SYSTEM INTRODUCTION
1
FEATURES
The Breathe Technologies® Life2000® Ventilation System is a critical care, volume control mechanical ventilation
system designed for a broad range of applications in critical care and home settings.
The modular Life2000® Ventilation System (system) is composed of the Life2000® Ventilator (ventilator) and the
Life2000®Compressor (compressor).
The ventilator:
• Oers three dierent volume control modes of operation:
• Control Ventilation
• Assist/Control Ventilation
• Assist Ventilation.
• Can be used with a variety of commercially available invasive interfaces (such as ET tubes) or non-invasive
masks such as full face, nasal, and pillows masks.
• Enables clinicians to define three prescriptions based on patient need.
• Allows for an adjustable PEEP setting for each prescription.
• Allows for an adjustable trigger sensitivity for each prescription.
• Includes the ability to set various critical alarms for each prescription.
• Has up to four hours of battery-powered operation.
• Displays patient breath rate, Peak Inspiratory Pressure (PIP), average flow, and current volume level.
The compressor:
• Provides a continuous 50-PSI pressure source.
• Is a charging station for the ventilator.
• Has an internal battery with one hour of operation.

11
1
LIFE2000 VENTILATION SYSTEM INTRODUCTION
PACKAGING CONTENTS
WARNING: For any accessories, read the label and accompanying document(s) before use.
TIP:
The ventilation system is shipped in specially designed, protective boxes. Do not throw away the boxes; keep
them for future transportation needs.
Life2000® Ventilator (ventilator)
The ventilator can be used with the Life2000®
Compressor or an alternate 50-PSI pressure source.
Life2000® Compressor (compressor)
The compressor is an electropneumatic power unit
that provides the ventilator with a continuous pressure
source and is a charging station for the ventilator.
Belt clip for ventilator
The belt clip is used to secure the ventilator when it is
used in wearable configurations.
Battery charger and AC power cord for the
ventilator
The battery charger and AC power cord connect the
ventilator to an AC power source.
.
External power supply and AC power cord for
the compressor.
The external power supply and AC power cord
connect the compressor to an AC power source.
Ventilator Carry Pouch
Alternative to belt clip for securing the ventilator when
using in wearable configurations.

12
LIFE2000 VENTILATION SYSTEM INTRODUCTION
1
SYSTEM COMPONENTS
VENTILATOR
TOP
Battery charger connection
Silence Alarm button
For manufacturer’s use only
FRONT
Touch screen
Activity buttons
High Activity button
Medium Activity button
Low Activity button
Power button for ventilator
Power indicator light
Alarm speaker
Backup alarm buzzer
Breath indicator light
SIDE
13 Belt clip sockets
BOTTOM
11 Interface connection
12 Gas inlet connection
2
3
1
4 5 6 7 8 109
11
12
13

13
1
LIFE2000 VENTILATION SYSTEM INTRODUCTION
18
11
12
13
15
14 16 17
COMPRESSOR
FRONT
Ventilator docking cradle
Power source indicator light
Power button for compressor
Locked icon
(ventilator charging indicator light)
Locking knob
Unlocked icon
Battery charge status button
Battery charge indicator lights
and battery charge scale
SIDE
Interface connection
Outlet fitting
BACK
11 Handle
12 Air inlet and filter
13 Cooling filter cover
Cooling vents
Condensation tray
Power supply connection
and cover
SIDE
Alarm buzzer (internal)
Silence Alarm button
0 100%
10
9
2
3
1
7
6
5 8
4
Outlet Air OnlyOutlet Air Only

14
LIFE2000 VENTILATION SYSTEM INTRODUCTION
1
CONFIGURATIONS
The modular Life2000® Ventilation System (system) is composed of the Life2000® Ventilator (ventilator) and the
Life2000® Compressor (compressor). The system can be used in three dierent configurations.
STATIONARY CONFIGURATION
The ventilator is docked into the compressor for ventilation while resting. For information about how to setup the
ventilation system in this configuration, see the following chapter "Chapter 2: Stationary Configuration".
EXTENDED RANGE WEARABLE CONFIGURATION
The ventilator is connected to the compressor with a gas supply hose to enable the activities of daily living. For
information about how to setup the ventilation system in this configuration, see "Chapter 3: Extended Range
Configuration" on page 26.
STANDALONE WEARABLE CONFIGURATION
The ventilator is connected to an alternate pressure source such as a hospital wall source or an air or oxygen
cylinder. For more information about how to setup the ventilator in this configuration, see "Chapter 4: Stand-Alone
Configuration" on page 40.

15
CHAPTER 2: STATIONARY CONFIGURATION
INTRODUCTION TO STATIONARY CONFIGURATION
In Stationary Configuration, the ventilator is docked into the compressor. The compressor provides the ventilator
with a continuous pressure source and functions as a charging station.
STATIONARY CONFIGURATION SETUP CHECKLIST
Refer to the chapter contents for full instructions, warnings, and cautions.
Test the ventilator before using it on a new patient if in multi-patient environment (see section below).
Position the compressor (see page 16).
Connect the compressor to an AC power source using the compressor’s external power supply and AC
power cord (see page 17).
Dock the ventilator into the compressor (see page 18).
TIP: Check to make sure the ventilator is locked in the compressor.
Connect an interface to the compressor (see page 19).
NOTE: For information about the interface on the patient side, see "Chapter 5: Connecting an Interface" on
page 51.
Power on the compressor (see page 20).
Power on the ventilator (see page 20).
Check the compressor’s battery charge status (see pages 21–22).
Check the ventilator’s battery charge (see page 23).
NOTE: Ventilation will not begin until an Activity button is pressed on the ventilator. For more information see
"Choosing an Activity Button (Patient-Selectable) to Begin Ventilation" on page 74.
TESTING THE VENTILATION SYSTEM
In a multi-patient setting, the ventilation system must be tested before it is assigned to a new patient. For
instructions on testing the ventilation system, see "Testing Ventilator Alarms" on page 118.

16
STATIONARY CONFIGURATION
2
POSITIONING AND CARRYING THE COMPRESSOR
Position the compressor upright on a flat, level surface. Make sure that the cooling vents, cooling filter cover, and
air inlet on the back of the compressor are not blocked, and there is sucient clearance from surrounding objects.
Protect the compressor from falling.
The compressor should not be used adjacent to or stacked with other equipment and that if adjacent or stacked
use is necessary, the compressor should be observed to verify normal operation in the configuration in which it
will be used.
When carrying the compressor, make sure to use the handle, keep the compressor in an upright position, and
protect it from falling or dropping.
NOTES:
• The compressor outputs an audible sound and must be at least 3 feet (1 m) away from the user during use.
• When operating in a 40°C (104°F) environment, the gas output by the ventilation system may reach
temperatures up to 48°C (118.4°F). To reduce patient discomfort, operate the ventilation system in a cooler
environment.
• The performance of the compressor may degrade in high temperature, high humidity, or high altitude
environments. If degradation is seen, switch to an alternate means of ventilation. Verify the performance of
the compressor for adequate therapy delivery in the environment(s) in which it will be used and adjust the
volume to compensate for altitude when necessary.
• The ventilator settings might not be achieved
when sourced by the Life2000®Compressor due
to increases in altitude near or above 2500 feet.
Consult the table to ensure that the compressor
can meet the ventilator settings. The tidal volume
delivered to the patient includes the ventilator
set volume + entrainment volume from patient
interface + supplemental oxygen volume (if used).
For additional information see page 152.
WARNING:
• Do not use the ventilation system in the presence of flammable anesthetics.
• Do not cover or block the compressor’s internal alarm buzzer with any object. Covering the buzzer may
make it dicult for a patient or caregiver to hear alarms, which may result in inadequate respiratory
therapy.
• Do not cover the ventilator, touch screen, speaker, or backup alarm buzzer with tape or any other object.
Covering the ventilator or any of its parts might cause diculty in hearing alarms and might aect ventilator
performance.
• When the ventilator is in use, keep it in a well-ventilated area to prevent it from overheating. The ventilator
may overheat and be permanently damaged if it is used in an area that is not well ventilated.
• Do not connect the ventilation system components or accessories to any other equipment that is not
described in this Instructions for Use.
CAUTION:
• Keep in a clean environment to protect the ventilation system from ingress of dust, lint, and pests.
• Do not leave the ventilation system exposed to the sun or other sources of radiant heat, it may overheat.
• Do not allow children or pets to access the ventilation system; it may become damaged.
Simulated Elevation
(in feet)
Observed Maximum
Compressor Output
(in LPM)
0 17
2500 14
4000 12
8000 8

17
2
STATIONARY CONFIGURATION
SUPPLYING POWER TO THE COMPRESSOR
An AC power cord and external power supply are included with the compressor.
Open the compressor's power supply connection cover
Insert the compressor’s power connector into the power supply
connection on the back of the compressor until it clicks into place;
the lock tab on the cord will be on the top.
Connect the AC power cord to the compressor’s external power
supply and cord.
Connect the pronged end of the compressor’s AC power cord to an
AC power source.
Verify that the green LED indicator on the external power supply lights up to indicate the AC connection.
NOTES:
• To remove the power cord from the compressor, press on the end of the lock tab on the external power
supply cord and pull to release the cord from the compressor.
• To isolate the compressor from the supply mains (AC power source), unplug the compressor's AC power
cord from the AC power source.
TIP:
The compressor’s power supply connection is equipped with a protective cover to prevent ingress of water and
debris. Use the cover to keep the power supply connection covered when not in use.
WARNING:
The interface, source gas supply hose, and power cords should be positioned to avoid restricting movement,
causing a tripping hazard, or posing a strangulation risk.
CAUTION:
• Do not place the compressor's AC power cord or external power supply on wet surfaces or use in wet
environments. Wet environments may damage the AC power cord or external power supply and may
cause electric shock.
• Use only the Breathe Technologies® approved AC power cord and external power supply with the
compressor. Using an unauthorized AC power cord or external power supply may damage the compressor.

18
STATIONARY CONFIGURATION
2
DOCKING THE VENTILATOR INTO THE COMPRESSOR
NOTE: When switching between ventilation system configurations, ensure that the patient receives adequate
ventilation therapy; place the patient on an alternate means of ventilation if necessary.
Ensure that the ventilator is powered o.
NOTE: Alarms might be encountered and/or the selected Activity
Button might be inadvertently changed if the ventilator is not
powered o before docking.
To dock the ventilator into the compressor, make sure the locking
knob is in the unlock position as shown.
Position the ventilator into the compressor cradle as shown. Push the
ventilator into the compressor in the direction indicated by the arrow
until it clicks into place.
Gently push the top of the ventilator (in the area indicated) until the
front of the ventilator is flush with the front of the compressor and
clicks into place.
TIP: Avoid pushing directly on the touch screen or buttons.
1
0 100%
2
Push here

19
2
STATIONARY CONFIGURATION
Turn the locking knob clockwise to the locked position to lock the
ventilator in the compressor.
When the compressor is on, check that the locked icon lights up
(green) to indicate the ventilator is properly docked. When the
ventilator is properly docked and locked the ventilator is being
charged by the compressor.
NOTES:
• If the compressor is o and connected to an AC power source,
press the battery charge status button on the compressor to
illuminate the green locked icon and verify that the ventilator is properly docked.
• If the knob does not lock on first attempt, turn knob slightly counterclockwise and then clockwise again.
• To ensure continuous ventilation, make sure the ventilator battery is being charged by the compressor
while the ventilator is docked. For more information see "Checking the Ventilator Battery Charge" on
page 23.
NOTE: For undocking instructions, see "Undocking the Ventilator from the Compressor" on page 29.
VENTILATOR CHARGING BY COMPRESSOR WHEN DOCKED AND LOCKED
Compressor ON Compressor OFF
AC CONNECTED to Compressor Ventilator charging Ventilator charging
AC DISCONNECTED from Compressor Ventilator charging Ventilator DOES NOT charge
NOTE: The ventilator will not charge when the compressor is powered o and not connected to AC power.
0 100%
5
6
CONNECTING AN INTERFACE TO THE COMPRESSOR
Plug the Breathe Pillows Interface™ or Breathe Universal Circuit™connector
into the interface port on the side of the compressor until it clicks. For
more information about wearing interfaces, see "Chapter 5: Connecting
an Interface" on page 51.
0 100%
THE OUTLET FITTING IN STATIONARY CONFIGURATION
Do not connect a gas source to the outlet fitting on the compressor.
CAUTION:
Using the outlet fitting while the Life2000 Ventilation System is delivering therapy in Stationary
Configuration might decrease flow to the patient.
0 100%
Table of contents
Other Breathe Medical Equipment manuals
Popular Medical Equipment manuals by other brands

iMediSync
iMediSync iSyncWave User manual and installation guide

KCI
KCI V.A.C.Via Instructions for use

Nasco
Nasco Life/form LF01007 Adult IV Foot instruction manual

ATsens
ATsens AT-Patch quick guide

Inovus Medical
Inovus Medical Lapar Pro quick start guide

ThermoFisher Scientific
ThermoFisher Scientific invitrogen E-Base EBM03 quick reference